F:\Ships\lmgould\2003\lmg03_2\surface\LMG03_2_data_report.doc Page 107/27/17 12:55 PM

Similar documents
Methods Used in Measuring Surface Seawater pco 2 Aboard RVIB Nathaniel B. Palmer

REPORT OF UNDERWAY pco 2 MEASUREMENTS IN SURFACE WATERS AND THE ATMOSPHERE DURING February - March (RVIB Nathaniel B. Palmer Cruise 01/2)

REPORT OF UNDERWAY pco 2 MEASUREMENTS IN SURFACE WATERS AND THE ATMOSPHERE DURING July (RVIB Nathaniel B. Palmer Cruise 00/5)

F:\Ships\NBPalmer\2001\nbp01_3\surface\NBP01_3_data_report.doc Page 104/26/17 4:13 PM

F:\Ships\NBPalmer\2002\nbp02_2\surface\NBP02_2_data_report.doc Page 104/25/17 3:33 PM

F:\Ships\NBPalmer\2001\nbp01_4\surface\NBP01_4_data_report.doc Page 106/02/15 11:21 AM

F:\Ships\NBPalmer\2003\nbp03_4a\surface\NBP03_4A_data_report.doc

F:\Ships\NBPalmer\2002\nbp02_1\surface\NBP02_1_data_report.doc Page 103/19/15 11:16 AM

REPORT OF UNDERWAY pco 2 MEASUREMENTS IN SURFACE WATERS AND THE ATMOSPHERE DURING February - April RVIB Nathaniel B. Palmer Cruise 03/4

Taro Takahashi, Colm Sweeney, S. C. Sutherland. Lamont-Doherty Earth Observatory of Columbia University Palisades, NY

Part 1 - Data Processing and Quality Control.

Part 1 - Data Processing and Quality Control.

Part 1 - Data Processing and Quality Control.

Part 1 - Data Processing and Quality Control.

Part 1 - Data Processing and Quality Control.

Part 1 - Data Processing and Quality Control.

Part 1 - Data Processing and Quality Control.

Part 1 - Data Processing and Quality Control.

F:\Ships\NBPalmer\2003\nbp03_1a\surface\NBP03_1a_data_report.doc Page 104/24/17 2:54 PM

Part 1 - Data Processing and Quality Control.

Part 1 - Data Processing and Quality Control.

COMPARISON OF SURFACE WATER pco 2 AND SST DATA DURING AN ENCOUNTER OF RVIB PALMER AND RV GOULD ON MARCH 13, 2009

Under way pco 2 System Users Manual

Uniform format surface fco 2 database Benjamin Pfeil and Are Olsen Bjerknes Centre for Climate Research, Bergen, Norway

Technical Report by Peter Guenther and Charles D. Keeling Carbon Dioxide Research Group Scripps Institution of Oceanography La Jolla, CA

WHP Operations and Methods { July William M. Smethie, Jr. Lamont Doherty Geological Observatory Palisades, New York U.S.A.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Application Note 268 Metabolic Analysis Systems and Propane-based Verification

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

CTD Data Documentation

Sensor Stability. Summarized by Breck Owens David Murphy, Sea-Bird Electronics AST-14 Wellington March 18-21

Generating Calibration Gas Standards

Application Note 183. Page 1 of Aero Camino, Goleta, CA Tel (805) Fax (805)

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

Columbus Instruments

11/22/ (4) Harmonization: <846> SPECIFIC SURFACE AREA

Another convenient term is gauge pressure, which is a pressure measured above barometric pressure.

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Dissolved Oxygen Meter. ino2. Instruction Manual. Serial No..

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

Focus on VOC Emissions Reduction Using an Oxygen Based Inerting Control System For Inert Gas Blanketing of Chemical Process Vessels

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

Operation manual Oxygen sensor

What I have learned about SF 6 gas testing.a Practical explanation

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

Cover Page for Lab Report Group Portion. Head Losses in Pipes

Experiment. THE RELATIONSHIP BETWEEN VOLUME AND TEMPERATURE, i.e.,charles Law. By Dale A. Hammond, PhD, Brigham Young University Hawaii

Cover Page for Lab Report Group Portion. Pump Performance

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow

CHAPTER 3 RESULTS AND DISCUSSION

Lung Volumes and Capacities

Experiment 8 GAS LAWS

Kinetic-Molecular Theory

(12) United States Patent

MODEL GT820 OXYGEN SENSOR

USING THE CS400/CS405 (KPSI SERIES 169/173) SUBMERSIBLE PRESSURE TRANSDUCER WITH CAMPBELL SCIENTIFIC DATALOGGERS INSTRUCTION MANUAL

This test shall be carried out on all vehicles equipped with open type traction batteries.

Module 2, Add on Lesson Depth Sensor. Teacher. 90 minutes

AMT-Ex Dewpoint Transmitter

Under pressure pushing down

Technical Data Sheet MF010-O-LC

Chapter 10 Gases. Characteristics of Gases. Pressure. The Gas Laws. The Ideal-Gas Equation. Applications of the Ideal-Gas Equation

Measurement of Atmospheric Pressure. Atmospheric Instrumentation

Standard Test Methods for Water Vapor Content of Electrical Insulating Gases by Measurement of Dew Point 1

Best Practice Guide, Servomex 2700

States of Matter Review

Figure Vapor-liquid equilibrium for a binary mixture. The dashed lines show the equilibrium compositions.

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300

Constant Pressure Inlet (CCN) Operator Manual

Ultimate Performance in Gas Chromatography

Fluid Machinery Introduction to the laboratory measurements

Accuracy and Stability of Sea-Bird s Argo CTD and Work Towards a Better CTD for Deep Argo. David Murphy Sea-Bird Electronics

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Flow in a shock tube

Pressure Measurement. Introduction. Engr325 Instrumentation. Dr Curtis Nelson 3/12/18

LEAP CO 2 Laboratory CO 2 mixtures test facility

CORESTA RECOMMENDED METHOD N 6

Helium Mass Spectrometric Leak Detection In Large Size Process Plants

New insights of pco2 variability in the tropical eastern Pacific Ocean using SMOS SSS and OSTIA SST C W Brown, J Boutin, L Merlivat, LOCEAN Paris

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

Test Report # Rev 0. Adiabatic Compression With Constant Bleed Valve

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

Application Note AN-107

Laser Spectrometers for Online Moisture Measurement in Natural Gas. Ken Soleyn GE M&C

Boyle s Law: Pressure-Volume Relationship in Gases

Sampling Considerations for Equilibrium Dissolved Oxygen [DO] Sensors

A Journal of Practical and Useful Vacuum Technology. By Phil Danielson

Mitos Fluika Pressure and Vacuum Pumps Datasheet

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

INTRODUCTION Porosity, permeability, and pore size distribution are three closely related concepts important to filter design and filter performance.

INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices

Chapter 13 Gases, Vapors, Liquids, and Solids

MS.RAJA ELGADY/PRESSURE PAPER 3

GCMSD-Headspace Analysis SOP

1. Compute the TWA for the following respirable dust exposures:

ASPECTS OF VACUUM TECHNOLOGY. (a) To see something of the operation of vacuum pumps; Then, we will use the vacuum system for one of two experiments:

Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL -

Yokosuka Cruise Report YK11-02

Transcription:

F:\Ships\lmgould\2003\lmg03_2\surface\LMG03_2_data_report.doc Page 107/27/17 12:55 PM REPORT OF UNDERWAY pco 2 MEASUREMENTS IN SURFACE WATERS AND THE ATMOSPHERE DURING May-June 2003 1. General 1-a) The Cruise: R/V Laurence M. Gould 03/2 S. C. Sutherland, Taro Takahashi, Colm Sweeney Lamont-Doherty Earth Observatory of Columbia University Palisades, NY 10964 27 July, 2017 1) This cruise was part of GLOBEC. The ship departed Punta Arenas, Chile on 13 February 2003, arrived at Palmer Station on 17 February. She departed the next day for local research, returning on 25 March. The northbound transit began from Palmer Station on 25 March, and ended with the ship s arrival in Punta Arenas on 29 March 2003. 2) Regarding SST the thermosalinograph temperature data was find for this cruise. We use 1 scan offset to adjust the pco s timing. 3) There were serious problems with our system during this cruise as witnessed by the 50% rejection rate of observations. The primary culprit was the gas flow from the equilibration vessel. 1-b) The pco 2 data: There were no merged files created during this cruise. We merged data from 4 sources to create a final file. Ship position, course, speed data came from one-second NAV files LMGlpcd.dXXX where XXX = julian date. The thermosalinograph files, named LMGltsg.dXXX contain temperature and salinity data. Meteorological data, wind speed, wind direction, barometric pressure, air temperature, and relative humidity come from LMGlmet.dXXX. pco 2 data come from LMGlpco.dXXX. The latter are the data coming from our pco 2 data acquisition program. The system performed very well, and the editing was obvious, i.e. no flow of equilibrator gas, etc. Thirty seawater and five air observations are made between standard sets. Each observation involves flowing the equilibration gas for 120 seconds. We also flow the air for 120 seconds before stopping for an observation. Details on these and other aspects of the analysis system are explained later.

F:\Ships\lmgould\2003\lmg03_2\surface\LMG03_2_data_report.doc Page 207/27/17 12:55 PM Standards: Cylinder Concentration (ppm) Nitrogen (UHP) 0.0 CC46457 150.19 CC51612 353.77 CC22986 251.79 CC46471 788.0 A 4 th order calibration curve is used with 5 standards, as described in Section 2. Standards flow for about 120 seconds before an observation is made. 1-c) AIR data: There are 1,347 air observations. By applying a mean and standard deviation filter to the data, 631 were retained. These gave a mean of 372.01 ± 0.51 (N = 631). By selecting only those air observations when the wind was within 5 of directly on the bow, we get only 68 observations. Interestingly, these give a mean of 372.31 ± 1.29. This compares with a value from the GlobalView CO 2 database extrapolated from 2001 of 371.2. Our value contains a great deal of noise, but is within 1 ppm of the GlobalView value. Much better than earlier cruises. To calculate pco 2 we use the Global View CO 2 data, extrapolated from the end of 2001 through 2002. The last observations in that file are for year 2001. A longterm trend is fitted to these observations, then extrapolated forward. Naturally, there is some uncertainty in this extrapolation, but it should be less than ±0.5 ppm. This value goes into the vco2_air and "gvvco2_air" variables in the L032sfc database. 1-d) Editing: The group of individual days beginning with 44 and ending with 65 were concatenated together and edited under the name rawdata.txt. Periods where the system was interrupted before a standard run was complete, those were the data are unreadable, etc. are edited at this level. a. The TSG temperature was reliable for most of this cruise b. There were several periods during which the flow of equilibrator air dropped to near or at zero. These are, of course, rejected. Of the 8,137 observations of seawater pco 2, we rejected 4,044, retaining 4,093 records. The program editpco2.prg performs this editing operation. c. To create a L032sfc.dbf surface format file, we use only ACCEPTED pco 2 values, but ALL the data are retained in the pco2data.dbf database in the subdirectory: \LMGOULD\2003\lmg03_3\SURFACE\PCO2DATA. We estimate the air value from the

F:\Ships\lmgould\2003\lmg03_2\surface\LMG03_2_data_report.doc Page 307/27/17 12:55 PM Globalview CO2database (as explained under 1-d) AIR data. The program makel032.prg performs this process. The program also adds salinity from the edited salinity file saldata.dbf in the salinity subdirectory (see other data below). 1-f) Other Data: Salinity: There is a program, procsal.prg in the subdirectory salinity which performs editing of the salinity data. The system uses a mean and standard deviation of 15 data points. There are five passes through the data. Three calculate mean and standard deviation, two perform edits. After each of the first 2 calculation passes, data points are rejected if outside 1 standard deviation. A minimum standard deviation of 0.02 is assigned. After the second editing pass (4th through the data), mean and standard deviation of the remaining data are calculated using only accepted values. Between accepted values the mean is unchanged. This file saldata.dbf is then used as the source for salinity in the master file L032sfc.dbf. Wind speed units are meters seconds -1.

F:\Ships\lmgould\2003\lmg03_2\surface\LMG03_2_data_report.doc Page 407/27/17 12:55 PM 2. MEASUREMENTS OF pco 2 IN SURFACE WATERS 2-a) The LDEO Underway System for Surface Water pco 2 Measurements: The system for underway measurements of pco 2 in surface waters consists of a) a water-air equilibrator, b) a non-dispersive infra-red CO 2 gas analyzer and c) a data logging system. The measurement system is schematically shown in Fig. 1, and is similar with the one described in Bates et al. (1998). Each of these units and the data reduction procedures used will be described below. Figure 1 - The underway pco 2 system used for the measurements of pco 2 in surface waters during the Southern Ocean JGOFS (AESOP) Program.

F:\Ships\lmgould\2003\lmg03_2\surface\LMG03_2_data_report.doc Page 507/27/17 12:55 PM 2-b) Water-air Equilibrator: The equilibrator has a total volume of about 30 liters and is equipped with a standpipe drain which automatically maintains the level of water in the equilibrator at a constant level at about half the height of the equilibrator leaving about 15 liters of headspace. Seawater from the ship s uncontaminated water line is continuously pumped into the equilibrator at a rate of about 10 liters/min, giving a mean residence time of water in the equilibrator of about 1.5 minutes. The headspace above the water serves as an equilibration chamber. A carrier gas (commonly marine air) is drawn into the chamber by a diaphragm pump, and exchanges CO 2 with a continuous flow of seawater sprayed into the chamber through a shower head. Because of large gas-water contact areas created by fine water droplets as well as gas bubbles in the pool of water, CO 2 equilibration between the carrier gas and seawater is achieved rapidly with a e-folding time of 2 to 3 minutes. Under normal operating conditions, the carrier gas in the equilibration chamber is pumped into the infra-red gas analyzer at a rate of about 50 ml/min. At this rate, the residence time of the carrier gas in the equilibration chamber is about 300 minutes, that is about 100 times as long as the equilibration time. Therefore, the carrier gas in the head space is always in equilibrium with water. The over all response time of the equilibrator system has been estimated to be of an order of several minutes. The large volume of water in the equilibrator is chosen in order to have a large thermal inertia of the equilibrator, so that the effects of room temperature changes on the equilibration temperature may be minimized. The temperature of water in the equilibrator is monitored continuously using a Guildline platinum resistance thermometer (readable to 0.05 o C) and recorded on the data logging computer. A calibrated mercury thermometer is also inserted in the equilibrator for testing the performance of the platinum thermometer. At the gas intake end of the equilibrator, a flow indicator based on a U-tube manometer is attached. This gives a visual confirmation for the fact that marine air is taken into the equilibration chamber at a desired flow rate and ensures that two way air exchange does not occur. Since we operate the system with the equilibration chamber at the same pressure as the ambient room pressure, the total pressure, at which the gas was equilibrated, is measured using a precision electronic barometer (Setra Model 270, Action, MA) outside the equilibrator. This equilibration pressure is also logged on the computer. The temperature and salinity of seawater at the in situ conditions were measured using a SeaBird Model SBE-21 thermosalinograph aboard the N. B. Palmer and a SIO/ODF thermosalinograph unit based on Neil Brown sensors aboard the R. Revelle. The precision of the report temperature data has been estimated to be about 0.005 o C. 2-c) Infra-red CO 2 Gas Analyzer: The equilibrated gas was passed through a water trap (to collect aerosols and condensates), a reverse flow naphion dryer (PermaPure flushed with pure nitrogen gas) to remove water vapor (to a level of 20 o C dewpoint), and was introduced into the IR sample cell at a rate of about 50 ml/min for CO 2 determinations. A LI-COR infra-red gas

F:\Ships\lmgould\2003\lmg03_2\surface\LMG03_2_data_report.doc Page 607/27/17 12:55 PM analyzer (Model 6251, Lincoln, NB) was used. After about 2 minutes of purging period, the gas flow was stopped and readings were recorded on the computer. Although an electronic circuit was provided by the manufacturer in order to linearize the CO 2 response, it exhibited a few inflexions that deviated from linearity by a few ppm. Therefore, we chose not to use the outputs from the linearization circuit supplied by the manufacturer. Instead, we used five standard gas mixtures (one pure nitrogen and four CO 2 -air mixtures) during the expeditions, and established response curves using the raw millivolt output from the analyzer. The CO 2 concentrations in the gas mixtures were calibrated using the SIO standards determined by C. D. Keeling s group using the manometric method. The concentrations of CO 2 in the standard gas mixtures were summarized above. During normal operations, each of the standard gas mixtures was passed through the analyzer for 120 seconds at a rate of about 60 ml/min. This replaced the IR analyzer cell completely with the new gas. The flow was stopped for 10 seconds and then a millivolt reading from the analyzer was taken and recorded. Samples of equilibrated air and marine air were pumped through the analyzer for 120 seconds (2 minutes) at a rate of about 50 ml/min to purge the previous sample in the IR cell. The flow was stopped for 10 seconds and a reading for the analyzer output was recorded. This procedure was intended to eliminate errors due to fluctuations of the dynamic pressure within the IR cell by irregular gas flow rates. The slow flow rates used for samples were required for the removal of water vapor using the PermaPure membrane dryer. We analyze 20 seawater and two air samples between two sets of calibration runs using the five standard gases. (the current setup is 30 seawater samples, 5 atmospheric) 2-d) Data Logging System: Our laptop records the following data and writes it via a serial port to the ship s RVDAS system. Date, Time (GMT), Sample ID (standard gas cylinder numbers, seawater CO 2, atmospheric CO 2 ) Barometric pressure in the laboratory (to 0.1 mb) IR cell temperature, Gas flow rate in the IR cell (to 0.1 ml/min), Temperature of equilibration (to 0.01 o C), Analyzer output (millivolts to 0.1 mv) CO 2 concentration in dry gas sample (preliminary based on the last response curve), and pco 2 (preliminary value based on the last response curve). The ships RVDAS system adds several other data elements from various sensors and equipment. The list is: Latitude, Longitude, Thermosalinograph temperature, thermosalinograph salinity, fluorometer voltage, wind speed, wind directiion, and, barometric pressure.

F:\Ships\lmgould\2003\lmg03_2\surface\LMG03_2_data_report.doc Page 707/27/17 12:55 PM 2-e) Data Reduction Procedures: The concentration of CO 2 in the sample was computed by the following way based on the millivolt reading and time of the reading. The millivolt reading taken for each of the five standard gases at the time of sample measurement was computed by linearly interpolating as a function of time using the readings taken before and after the respective standard gases were analyzed. This yields millivolt reading for each of the five standard gases at the time when the sample was analyzed. These five values were fit to a fourth-order polynomial equation (with five constants to be determined). This serves as the response curve. The CO 2 concentration in the sample was computed using the response curve that was established at the time of each sample analysis. This method has been demonstrated to yield more reliable CO 2 values compared with those computed, for example, using a least-squares fit of a quadratic or cubic functions to the five calibration points. The method described above yields atmospheric CO 2 values that are consistent with those reported for the South Pole and the Cape Grim by the Climate Monitoring and Diagnostics Laboratory/NOAA in Boulder, CO. The partial pressure of CO 2 in seawater, (pco 2 )sw, at the temperature of equilibration, Teq, in the unit of microatmospheres (µatm) was computed using the expression: (pco 2 )sw @ Teq = (Vco 2 )eq x (Pb Pw),......[1] (Vco 2 )eq Pb Pw = the mole fraction concentration (ppm) of CO 2 in the dried equilibrated carrier gas; = the barometric pressure (that is equal to the total pressure of equilibration) in atmospheres; and = the equilibrium water vapor pressure at Teq ( o C) and salinity. The water vapor pressure was computed using the following formulation; Pw (atm) = (1/760)x(1 5.368x10-4 x Sal) x EXP{[0.0031476 (1/TK)]/1.8752x10-4 },... [2] where Sal is salinity in PSU measured using the ship s thermosalinograph, and TK is the temperature of equilibration in o K. The (pco 2 )sw at the in situ temperature, T in situ, was computed using a constant value of 0.0423 % per o C for the effect of temperature (Takahashi et al., 1993): (pco 2 )sw @ Tin situ = (pco 2 )sw @ Teq x EXP[0.0423 x (Tin situ Teq)]. The value for Tin situ is taken to be the seawater temperature measured by the ship s thermosalinograph at the time of pco 2 measurements. Teq is generally warmer than

F:\Ships\lmgould\2003\lmg03_2\surface\LMG03_2_data_report.doc Page 807/27/17 12:55 PM Tin situ by 0.5 ~ 0.8 o C. Hence the temperature correction is normally less than 3% of pco 2 values. The over all precision of the reported pco 2 )sw values has been estimated to be about +1.5 uatm. 3. MEASUREMENTS OF pco 2 IN THE ATMOSPHERE 3-a) Measurements: The air measurement system is shown schematically in Fig. 1. Uncontaminated marine air samples were collected about 10 m above the sea surface using a DEKORON tubing (3/8 i.d., Calco Inc., PA), a thin-wall aluminum tubing protected by plastic casing. The intake was located at the middle of the foremast about 10 m above the sea surface. A KNF Neuberger air pump that was located near the IR analyzer was used to pump air through the tubing and into the IR analyzer. Even when air samples were not directed to the analyzer, the pump was on all the time to keep the air flowing through the sampling line. For the analysis, the air sample was passed through a drying column to remove water vapor (the same PermaPure column as used for the equilibrated gas) and introduced into the IR cell for CO 2 analysis at a rate of about 50 ml/min. After 2 minutes of purging the cell, the flow was stopped for 10 seconds and the IR millivolt output reading was recorded. 3-b) Data Processing: The partial pressure of CO 2 in the air, (pco 2 )air, was computed in the unit of microatmospheres (µatm) in the same way as that for seawater using Eq. [3] below: (pco 2 )air = (Vco 2 )air x (Pb Pw),......[3] (Vco 2 )air Pb Pw = the mole fraction concentration (ppm) of CO 2 in the dried air sample; = the barometric pressure at sea surface in atmospheres; and = the equilibrium water vapor pressure at Tin situ ( o C) and salinity given by Eq. [2]. The precision of the atmospheric pco 2 values have been estimated to be about + 1 µatm.

F:\Ships\lmgould\2003\lmg03_2\surface\LMG03_2_data_report.doc Page 907/27/17 12:55 PM 4. REFERENCES CITED Bates, N. R., Takahashi, T., Chipman, D. W. and Knapp, A. H. (1998). Variability of pco 2 on diel to seasonal time scales in the Sargasso Sea. Jour. Geophys. Res., 103, 15567-15585. CO 2 Group, Lamont-Doherty Earth Observatory. (1999) pco 2 Equilibrator Users Manual, LDEO of Columbia University, Palisade, NY, pp.10. Takahashi, T., Olafsson, J., Goddard, J., Chipman, D. W. and Sutherland, S. C., (1993). Seasonal variation of CO 2 and nutrients in the high-latitude surface oceans: A comparative study. Global Biogeochemical Cycles, 7, 843-878.