Oxymax W COS51D. Technical Information. Digital sensor for measuring dissolved oxygen Membrane-covered amperometric sensor

Similar documents
Oxymax W COS41. Technical Information. Potentiostatic amperometric two-electrode sensor Long-term stable sensor for water and wastewater applications

Oxymax H COS21D. Technical Information

Oxymax H COS21. Technical Information. Sensor with long-term stability for frequent sterilization and autoclaving

Oxymax COS61/COS61D. Technical Information

Oxymax COS22/22D. Technical Information

COM280. Technical Information. Oxygen measurement Battery-driven pocket meter for dissolved oxygen measurement

CCS240 and CCS241. Technical Information. Sensors for chlorine dioxide Amperometric, membrane-covered sensors for installation in the CCA250 assembly

Technical Information Oxymax COS22D/COS22

Operating Instructions. Oxymax COS51D. Digital sensor for dissolved oxygen With Memosens technology BA00413C/07/EN/

Operating Instructions. Oxymax W COS41. Dissolved oxygen sensor. BA284C/07/en/

Electropneumatic Positioner and Pneumatic Positioner Type 3760 JIS

High-performance submersible pressure transmitter For level measurement Model LH-10

Sensor for Measuring Dissolved Oxygen

Electropneumatic Positioner Type 4763 Pneumatic Positioner Type 4765

OUS31. Technical Information

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

Operating Instructions Oxymax COS22D

High-performance submersible pressure transmitter For level measurement Model LH-20

High-performance submersible pressure transmitter For level measurement Model LH-20

LMP 307. Stainless Steel Probe. Stainless Steel Sensor. accuracy according to IEC 60770: standard: 0.35 % FSO option: 0.25 % / 0.

Electropneumatic Positioner and Pneumatic Positioner Type 3760 JIS

Technical Information Conducal CLY421

High-performance submersible pressure transmitter For level measurement Model LH-10

IDL01. Battery Powered Precision Digital Gauge for Leak Testing. Stainless Steel Sensor. class 0.05

DM01. Please visit our website: Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

Turbimax CUS31. Technical Information

High-performance submersible pressure transmitter For level measurement Model LH-20

Oxygen Measurement Systems High Reliability and Wide Application Coverage

Torque Tube TB300 Digital Transmitters

L (up to 25 m) Electronic part. working temp.: Ñ IP 66 / IP67. Mounting bracket ÐÑ. Sensor type. working temp.:

Level transmitter LT 100

Technical Information Cleanfit CUA451

ETd 04. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. Type: ETd 04. class 0.05

High-performance submersible pressure transmitter For level measurement in hazardous areas Model LH-20

DM01. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

SITRANS P measuring instruments for pressure

Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter

Expert Hydrostatic Level Transmitters

Pressure Measurement Single-range transmitters for general applications

1 Overview. Pressure Measurement Single-range transmitters for general applications. 1/22 Siemens FI US Edition

No.600,No.21 Supmea Automation Co.,Ltd. Vortex flow meter. Datasheet DY2900

Hidrostatic Level DATASHEET

OPTIBAR P 1010 C Technical Datasheet

2 Overview. SITRANS P measuring instruments for pressure. Transmitters for gage and absolute pressure. Z series for gage pressure

Ultrasonic. 58 Overview VEGASON 60 VEGASON 61, 62, 63

Indumax CLS54D. Operating Instructions

JUMO dtrans O2 01 Two-wire Transmitter for dissolved oxygen (DO)

1 Overview. Pressure Measurement Single-range transmitters for general applications. 1/22 Siemens FI

9700 Transmitter Submersible hydrostatic level transmitter

Operating manual. TA 197 Oxi. Dissolved Oxygen Depth Armature

PCDY01 Dissolved Oxygen Meter

1 Overview. Pressure Measurement Transmitters for basic requirements. 1/16 Siemens FI SITRANS P220 for gauge pressure

Tapped exhaust port. Sturdy, simple, reliable design. Easy to add on Feedback. High gain pilot valve. Simple calibration, external zero adjustment

dtrans O2 01 Two-wire Transmitter for dissolved oxygen (DO)

Dissolved Oxygen and measurement possibilities. Berno Lüpkes, 15 th March 2017

C-Flow Coriolis Mass Flow Meters

Two-Wire In Situ Oxygen Analyzer (550 to 1400 C)

Miniature flowmeters DK 46, DKR 46 DK 47, DK 48, DK 800

JUMO dtrans p30 Pressure Transmitter

Ultrasonic. Continuous level measurement. Overview Page 64 VEGASON series 60 Page 66 AL-EN

VersaFlow Vortex 100 Vortex Flow Meter Technical Datasheet. Specification. The all-in-one solution. Highlights. Industries.

Operating Instructions CCS120. Sensor for total chlorine. BA388C/07/en/

Precision pressure sensor Basic version Model CPT6020

Pressure Measurement Single-range transmitters for general applications

ABB MEASUREMENT & ANALYTICS DATA SHEET series Dissolved oxygen systems

DMK 458. Pressure Transmitter for Marine and Offshore. Ceramic Sensor. accuracy according to IEC 60770: standard: 0.25 % FSO option: 0.

Technical Data. Dimensions

Pressure Measurement. Transmitters for basic requirements SITRANS P Z for gauge pressure. 2/4 Siemens FI Overview

1 Overview. Pressure Measurement Single-range transmitters for general applications. 1/16 Siemens FI US Edition

INSTRUCTION MANUAL. 840 Transmitter. Loop Powered Dissolved Oxygen Transmitter

STERILIZABLE DISSOLVED OXYGEN SENSORS

NivuBar Plus II

Hydrostatic. Continuous level measurement. Overview Page 188 VEGAWELL 52 Page 190 VEGABAR 86, 87 Page 192 Accessory Page 196 AL-EN

Precision digital pressure gauge Model CPG1000

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars.

9408 Series Dissolved Oxygen Systems. Dissolved oxygen systems designed for reliable operation with minimal operating costs

Type N-10, N-11 Hazardous Area Non-incendive Transmitters

Submersible pressure sensor High-performance Model LH-20

Pneumatic high-pressure controller Model CPC7000

Technical Data. Dimensions

SITRANS. Pressure transmitter SITRANS P, Z series for gauge and absolute pressure. Introduction. Safety instructions 2.

Technical Data. Dimensions

Technical Data. Dimensions

Optical Dissolved Oxygen Meter

Dimensions. Technical Data General specifications Switching element function Rated operating distance s n 10 mm

Smart performance with innovative and ever-strong coil drive even under harsh working environments

Dissolved Oxygen Guide

How to specify a product. Process Sensors and Mechanical Instruments

XMO2. Panametrics Smart Oxygen Analyzer. Applications. Features. bhge.com. An oxygen transmitter for use in:

Flow metering in gases Introduction to COMBIMASS family

Technical Data. General specifications Switching element function Rated operating distance s n 5 mm

STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE

Ultrasonic level transmitter and indicator

Hydraulic hand spindle pump Models CPP1000-M, CPP1000-L

Liquid level instruments BM 24 /BM 51

Level transmitters and indicators Series LU Ultrasonic level transmitter and indicator for liquids and solids

x act i Precision- Pressure Transmitter for Food Industry, Pharmacy and Biotechnology Stainless Steel Sensor

ATM.1ST/Ex - High Precision Transmitter

Gas density monitor With integrated transmitter Model GDM-100-TI

ABB MEASUREMENT & ANALYTICS DATA SHEET. AP100 ph/redox (ORP) cartridge sensors with water-wash option

Transcription:

Technical Information Oxymax W COS51D Digital sensor for measuring dissolved oxygen Membrane-covered amperometric sensor Application The continuous measurement of the concentration of dissolved oxygen is very important in many areas of water management: Sewage treatment plants: Oxygen measurement and regulation in the activated sludge basin for a highly efficient biological cleaning process Water monitoring: Oxygen measurement in rivers, lakes or seas as an indicator of the water quality Water treatment: Oxygen measurement for status monitoring of drinking water for example (oxygen enrichment, corrosion protection etc.) Fish farming: Oxygen measurement and regulation for optimum living and growth conditions Your benefits Sensor with digital data transmission: Storage of calibration data in sensor Insensitive to electromagnetic interferences thanks to digital communication to the transmitter Maximum measurement accuracy: Longterm-stable measurement thanks to potentiostatic amperometric three-electrode system Long maintenance intervals Intelligent sensor self monitoring Membrane covered sensor, i.e.: high O 2 selectivity No flow necessary - measurement in still waters possible Minimum maintenance effort Minimum calibration effort thanks to simple calibration in air Further benefits offered by Memosens technology Maximum process safety through contactless inductive signal transmission Data safety through digital data transmission Easy handling due to storage of sensor-specific data Predictive maintenance possible thanks to registration of sensor load data TI413C/07/en/10.06 71033048

Function and system design Measuring principle Potentiostatic amperometric three-electrode system Memosens technology The oxygen molecules diffused through the membrane are reduced to hydroxide ions (OH-) at the cathode. Silver is oxidized to silver ions (Ag+) at the anode (this forms a silver halogenide layer). A current flows due to the electron donation at the cathode and the electron acceptance at the anode. Under constant conditions, this flow is proportional to the oxygen content of the medium. This current is converted in the transmitter and indicated on the display as an oxygen concentration in mg/l, as a saturation index in % SAT or as an oxygen partial pressure in hpa. The high-impedance, current-free reference electrode plays an important role. The formation of a silver bromide or silver chloride coating on the anode uses up the bromide or chloride ions dissolved in the electrolyte In the case of conventional membrane-covered sensors working with the two-electrode system, this causes an increase in signal drift. This is not the case with the three-electrode system: The change in bromide or chloride concentration is registered by the reference electrode and an internal control circuit holds the working electrode potential constant. The advantages of this principle are significantly increased accuracy of the signal and considerably extended calibration intervals. Maximum process safety The inductive and non-contacting measured value transmission of Memosens guarantees maximum process safety and offers the following benefits: All problems caused by moisture are eliminated. The plug-in connection is free from corrosion. Measured value distortion from moisture is not possible. The plug-in system can even be connected under water. The transmitter is galvanically decoupled from the medium. The result: No more need to ask about "symmetrically high-impedance" or "unsymmetrical" (for ph/orp measurement) or an impedance converter. EMC safety is guaranteed by screening measures for the digital measured value transmission. Data safety through digital data transfer The Memosens technology digitalizes the measured value in the sensor and transfers it to the transmitter via a contactless connection. The result: An automatic error message is generated if the sensor fails or the connection between sensor and transmitter is interrupted. The availability of the measuring point is dramatically increased by immediate error detection. The digital signals are suitable for application in hazardous areas; the integrated electronics are intrinsically safe. Easy handling Sensors with Memosens technology have integrated electronics that allow for saving calibration data and further information such as total hours of operation and operating hours under extreme measuring conditions. When the sensor is mounted, the calibration data are automatically transferred to the transmitter and used to calculate the current measured value. Storing the calibration data in the sensor allows for calibration and adjustment away from the measuring point. The result: Sensors can be calibrated unter optimum external conditions in the measuring lab. Wind and weather do neither affect the calibration quality nor the operator. The measuring point availability is dramatically increased by the quick and easy replacement of precalibrated sensors. The transmitter does not need to be installed close to the measuring point but can be placed in the control room. Maintenance intervals can be defined based on all stored sensor load and calibration data and predictive maintenance is possible. The sensor history can be documented on external data carriers and evaluation programs at any time. Thus, the current application of the sensors can be made to depend on their previous history. 2 Endress+Hauser

Measuring system A complete measuring system comprises at least: the digital oxygen sensor Oxymax W COS51D a transmitter, e.g. Liquiline M CM42 a special measuring cable, CYK10 an assembly, e.g. immersion assembly CYA611 or retractable assembly COA451 Optional (see accessories): Universal suspension assembly support CYH101 for immersion operation Junction box RM (with cable extension) Automatic cleaning system Chemoclean with spray head Measuring system (example) 1 Digital oxygen sensor Oxymax W COS51D 2 Retractable assembly COA451 3 Liquiline M CM42 4 Measuring cable CYK10 a0006735 Input Measured variable Measuring range dissolved oxygen [mg/l / % SAT / hpa] 0.01 to 100 mg/l 0.00 to 1000 % SAT 0 to 2000 hpa Wiring Electrical connection The sensor is electrically connected to the transmitter by means of the special measuring cable CYK10. Special measuring cable CYK10 a0003350 Endress+Hauser 3

Performance characteristics Response time COS51D-***0* (black membrane cap for normal response time): t 90 : 3 minutes t 98 : 8minutes (each at 20 C / 68 F) COS51D-***1* (white membrane cap for fast response time): t 90 : 0.5 minutes t 98 : 1.5 minutes (each at 20 C / 68 F) Reference operating conditions Reference temperature: Reference pressure: 25 C (77 F) 1013 hpa (15 psi) Signal current in air 1) Zero current Measured value resolution COS51D-***0* (black membrane cap): approx. 300 na COS51D-***1* (white membrane cap): approx. 1100 na < 0.1 % of the current in air 10 μg/l (10 ppb) Maximum measured error ±1 % of measured value 2) Repeatability ±1 % of measured value Long-term drift Zero-point drift: Measuring range drift: < 0.1 % per week at 30 C (86 F) and under constant conditions < 0.1 % per week at 30 C (86 F) and under constant conditions Influence of medium pressure Polarization time Oxygen intrinsic consumption Pressure compensation not necessary < 60 minutes COS51D-***0*: Approx. 90 ng/h in air at 25 C (77 F) COS51D-***1*: Approx. 270 ng/h in air at 25 C (77 F) Installation Angle of installation Not permissible! Permissible angle of installation 10 10 Not permissible! Permissible angle of installation a0006741-en 1) For the reference operating conditions indicated 2) In accordance with IEC 746-1 at nominal operating conditions 4 Endress+Hauser

Environment Ambient temperature range Storage temperature 5 to 50 C (20 to 120 F) filled with electrolyte: 5 to 50 C (20 to 120 F) without electrolyte: 20 to 60 C (0 to 140 F) Ingress protection IP 68 Process Process temperature Process pressure 5 to 50 C (20 to 120 F) max. 10 bar (145 psi) permissible overpressure Underpressure operation is not permissible. Mechanical construction Design, dimensions Dimensions a0006742 Weight 0.3 kg (0.7 lbs) Endress+Hauser 5

Material Sensor shaft: Membrane cap: Cathode: Anode/Reference electrode: POM POM Gold Silver / silver bromide Process connection G1 and NPT ¾" Membrane thickness Temperature compensation Electrolyte COS51D-***0*: approx. 50 μm COS51D-***1*: approx. 25 μm internal Alkaline electrolyte Ordering information Product structure Approval A Non-hazardous area G ATEX O FM / CSA Head type S NPT 3/4", G1, Memosens connector Cable length 8 Standard version, without cable Membrane cap 1 COY31-WP membrane cap, minimum flow rate 5 mm/s, for normal response time 2 COY31-S-WP membrane cap, minimum flow rate 25 mm/s, for fast response time Accessories 0 No accessories COS51D- Complete order code Scope of delivery! Note! The following items are included in the delivery: Oxygen sensor with transport protection cap for membrane protection Accessories set with the following contents: 2 replacement cartridges (replacement membrane caps) 10 plastic ampoules containing electrolyte 1 sealing kit with 3 O-rings 6 abrasive sheets Operating Instructions (on CD only) Brief Operating Instructions (paper version) With sensor version COS51D-**8** you must order the measuring cable CYK10 separately (see accessories). Ex approval Certificates and approvals Version COS51D-G**** ATEX II 1G EEX ia IIC T3/T4/T6 Version COS51D-O**** IS Class 1 Division 1 ABCD T4/T6 Class 1 Zone 0 AEx ia IIC T4/T6 6 Endress+Hauser

Assemblies (selection)! Note! Accessories In the following sections, you find the accessories available at the time of issue of this documentation. For information on accessories that are not listed here, please contact your responsible service. Retractable assembly Cleanfit COA451 manually driven retractable assembly, stainless steel, with ball valve, for oxygen sensors; ordering acc. to product structure (Technical Information TI368C/07/en) Immersion assembly Dipfit W CYA611 for sensor immersion in basins, open channels and tanks, PVC Ordering acc. to product structure (Technical Information TI166C/07/en) Zero solution Measuring cable 3 units to produce 3 x 1 liter oxygen-free solution order no. 50001041 CYK10 Memosens data cable For digital sensors with Memosens technology Ordering according to product structure, see below Certificates A Standard, non Ex G ATEX II 1G EEx ia IIC T6/T4 Cable length 03 Cable length: 3 m (9.8 ft) 05 Cable length: 5 m (16 ft) 10 Cable length: 10 m (33 ft) 15 Cable length: 15 m (49 ft) 20 Cable length: 20 m (66 ft) 25 Cable length: 25 m (82 ft) 88... m length 89... ft length Ready-made 1 Wire terminals! Note! CYK10- complete order code Ex versions of CYK10 are indicated by an orange-red coupling end. CYK81 measuring cable non-terminated measuring cable for extension of sensor cables of e.g. Memosens, CUS31/CUS41 2 wires, twisted pair with shield and PVC-sheath (2x2x 0.5 mm 2 + shield) Sold by the meter, order no. 51502543 Junction box Transmitter Junction box RM To lengthen the cable for Memosens or CUS31/CUS41 With 2 x PG 13.5 IP 65 (i NEMA 4X) Order no. 51500832 Liquiline M CM42 Modular two-wire transmitter for Ex and non-ex areas Hart, PROFIBUS or FOUNDATION Fieldbus available Ordering acc. to product structure, Technical Information TI381C/07/en Endress+Hauser 7

TI413C/07/en/10.06 71033048 Printed in Germany / FM+SGML 6.0 / DT