MARK III Manometric Perfusion Pump Ve rtical Deck

Similar documents
Portable Pressurized Infusion Pump. and. Air Compressor. Operating and Service Manual

Manual Solvent Purification System. Innovative Technology, Inc.

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

HANDBOOK. Squeeze Off Unit SOU 400. Please refer any queries to:

InFlow Modular perfusion bioreactor

FOR INSTALLING CO 2 BLENDER KIT (P/N IN BEER SYSTEM

BioAerosol Nebulizing Generator. Operation and Maintenance User Manual

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y

MF2M Monnier Miniature Compressed Air Filter

Code AWC20HP Air Compressor

Helium-Oxygen Blender

Design of an Automated Perfusion Pump James J Higham

MEDICAL REGULATOR RANGE (MEDIREG RANGE)

MANUAL GAS MANIFOLDS

AUTOMATIC GAS MANIFOLDS

User s Guide. Vacuum Controller for liquid delivery systems

SERIES-O REGULATOR. Instructions for use

OXY-VIVA 3 RESUSCITATOR OPERATING INSTRUCTIONS

Operators Manual. Supera Floor Stand Anesthesia Machine M1000. Models: M1000 Basic Non-rebreathing M1200 Basic M1200

Assembly Drawing: W-311B-A01, or as applicable Parts List: W-311B-A01-1, or as applicable Special Tools: , , &

Budget Range Operators Handbook

Product Sheet Spraymist Unit

Instruction Manual Updated 7/26/2011 Ver. 2.2

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator

Dialflow Regulator. Instructions for Use

GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP

P5523. Users Manual. Liquid to Gas Separator. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

Beamex PG PRESSURE GENERATORS. Beamex PGM PGV PGC PGHH PGPH PGL

product manual HM-4140, HM-4150, HM-4160 HM-4160A HM-4150 Humboldt FlexPanels

Instructions for Use

Pressure Automated Calibration Equipment

HANDBOOK. Squeeze Off Unit SOU 250. Please refer any queries to:

Operation and Maintenance Manual for GENTEC Model 881VR Continuous/Intermittent Digital Suction Regulators

WELLS JOHNSON HIGH VOLUME CANISTERS

Instructions for use Compressed air injector with water trap

Pressure Dump Valve Service Kit for Series 3000 Units

1020 Industrial Drive, Orlinda, TN fax

Pressure Dump Valve Service Kit for Series 2300 Units

RD(H)20/25 Pressure-Reducing Regulator User Manual

Assembly-, installation- and maintenance instruction manual for bladder accumulators IBV / EBV , top reparable

SPECIFICATIONS TABLE Maximum Air Pressure 115 PSI Air Tank Capacity 20 Gallons

Hi-Force Limited Prospect Way Daventry Northants NN11 8PL United Kingdom Tel: +44(0) : Fax: +44(0) : Website:

LV LOADER DIRECTIONAL CONTROL VALVES INSTALLATION & USER GUIDE

System Pressure Manager Standard & System Pressure Manager Plus

INSTRUCTION MANUAL & PARTS

Tridak Model 1050 User Guide

TJF-Q180V Cleaning and Disinfection Checklist

SEMEM3-73 Maintaining medical gas pipeline systems and equipment

Multi-Probe Sampling System User Guide & Manual

OPERATION MANUAL RTI RHS650 RTI TECHNOLOGIES, INC East Market Street York, PA Manual P/N

NGP-250/500 Nitrogen Generator Quick Start Guide

Installation Operation Maintenance. Bermad Level Control Valve with Modulating Horizontal Float Pilot valve One Way Flow IOM.

User's Manual. MixRite TF 10. Edition 05.08

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor

OPERATING INSTRUCTIONS

Alfred ECMO Circuit Priming and Storage Guide

Installation of Your SprayMaster System

CLEMENTS. ANA Oxygen Regulator with Yoke Connector. User Manual. Manual No. ANA Issue 1

DIAGNOSTIC CARDIOLOGY

NTF-230 OPERATION MANUAL

VACUUM REGULATORS CONTENTS

TJF-Q180V Cleaning and Disinfection Checklist

The NAMIC Dream Kit. Single Operator Control. Reduce Contrast Waste. Less Force to Inject. Compensator * Manifold. How the Compensator Manifold Works:

Inert Air (N2) Systems Manual

Oxygen Dialflow Meter. Instructions for Use

PRESSURE TEST PUMP OPERATING & MAINTENANCE INSTRUCTIONS. Model No: PTP100. Part No: GC01/09

Installation and operating manual. Pneumatic control station LK product no: PCS 1-10

INSTRUCTION MANUAL. January 23, 2003, Revision 0

PERFORM Operating Document

Binks SG-2 Plus TM Rotary 2 QT. PRESSURE CUP

ATD LB PRESSURE BLASTER INSTRUCTION MANUAL

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax

P5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

3:1 High Ratio Oil Pump W. Extn. Kit

1.060 Set-Up and Priming of the Bypass Circuit. Perfusion Technology Department. Perfusionists

SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL

MSC-P and MSC-N Manifolds for Steam Distribution and Condensate Collection

SILENTAIRE TECHNOLOGY

instruction manual highpressure handpump manual

RIX COMPRESSORS. MICROBOOST Owner s Manual CONFIGURATION A MB-115 MB-230. November 21, 2002 APPLICABILITY [ A]

Parker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA USA Tel: Ref Rev.

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

GREIG FILTERS, INC. OPERATION MANUAL

AMS SUPER SILENT COMPRESSOR USER & MAINTENANCE GUIDE

User s Guide. CO2 & O2 Controller for miniature incubators

AMP Oil Free Manual AMP 50-8-TC AMP 50-6-D AMP General User and Maintenance Instructions

Operation & Service Manual

Air-Oxygen Blender. Service Manual. Model No. PM5200 Series PM5300 Series (shown)

LRS(H)4 USER MANUAL. Read the complete manual before installing and using the regulator.

Installation and commissioning instructions 255 series and 256 series

MoveRoll Conveyor Operating and Maintenance Manual

UNITY 2. Uni-DAAMS Manual. Version 1.3 (Changes to Section 1.1) March 2010

Linkam Scientific Instruments USER GUIDE TST350. Temperature Controlled Tensile Stress Testing Stage

easycomposites SETUP & DAILY USE INSTRUCTIONS NEVER SWITCH OFF YOUR PUMP WITHOUT FIRST CLOSING THE VALVE TO THE CHAMBER

HPB25 Hydraulic Paving Breaker

Training. Testor Training Manual

RARS5000 AIR BODY SAW OWNER S OPERATING MANUAL

Lubricated TC Models

Operation and Maintenance of the EPV200 Portable Ventilator

UltRo Dual Flow Reverse Osmosis Water System

Transcription:

MARK III Manometric Perfusion Pump Ve rtical Deck Dentsleeve International Ltd. Manufactured by Mui Scientific 45 Traders Blvd. East, Unit 33-34 Mississauga, Ontario, CANADA L4Z 3L3 Tel: (905) 7-888 Fax: (905) 890-353 Toll Free: (866) 75-6345 Email: mail@dentsleeve.com Website: www.dentsleeve.com Handbook September, 006

Index In Section In Section A A - A - A - 3 Section B Section C C - C - C - 3 C - 4 C - 5 C - 6 C - 7 C - 8 C - 9 C - 0 C - C - C - 3 Section D D - D - D - 3 Section E E - E - E - 3 E - 4 E - 5 E - 6 E - 7 Index Overview Schematic of pump Major pump components Air Water System Precautions & Warnings Normal Use Assembly & filling of perfusate reservoir Check, connect & set air supply Connection & flushing of water manifold Check flow values of each hydraulic resistor Turn on water perfusion to fill each transducer Transducer calibration (if required) Connect catheter to transducers Water injection procedure - each channel Check flow value for air perfusion manometry Observation & refilling of perfusate reservoir Maintain correct perfusate reservoir pressure Prevention, recognition and correction of hydraulic resistor blockage Compressor Steps on completion of measurements Perfusate reservoir Water manifold Air supply Set Up Procedures & Connection of Components Air supply Perfusate reservoir prior to first use Installation of pressure transducers Water perfusion manifold removal Standard hydraulic resistors Compact resistors Air perfusion circuit

In Index Section F F - F - F - 3 F - 4 F - 5 Section G G - G - G - 3 G - 4 Section H H - H - Section I I - I - I - 3 I - 4 Cleaning & Disinfection Cleaning of plastic parts Sterilization of perfusate reservoir Sterilization of manifolds Hydraulic resistors Cleaning of compressor Regular Maintenance Every 3 months (or as needed): perfusate water filter Every year (or as needed): gas filters Service of control box by Dentsleeve Compressor dessicant Problem Solving Diagnosis of abnormal air consumption Abnormal air consumption - perfusate reservoir leakage Specifications, Support, Spart Parts & Accessories Specifications Technical Support Spare parts Regulatory information Notes on the use of this manual References within the manual are shown in brackets eg (C 4.) = section C, part 4, instruction Part numbers (#) given in the text are unique for Dentsleeve The technical information and illustrations in this manual reflect specifications and operating procedures at the time of drafting. Some specifications and operating procedures differ from earlier manuals for Mark III perfusion pump models. Dentsleeve reserves the right to vary specifications and operating procedures as part of its continuous product improvement process. 3

Overview A Schematic of pump A 00 kpa or 5 psi Regulated air supply to perfusate reservoir Water perfusate Air supply Pump control box Manifold purge clamp Manifold side-arm pinch valves 3 Way stop cocks Pressure transducers Catheter connector tubes Catheter Note: Schematic diagram of water perfusion circuit. Only critical components are shown. 4

A Overview A Major pump components Reservoir pressurization line Perfusate reservoir Pump control box Reservoir outflow control device Pressure transducer rack Air inlet Air perfusion line Standard hydraulic resistors Manifold flush valve Calibration bowl Main perfusate manifold Pinch valves Note: Pump length, channel numbers and spacings vary according to individual specifications. Standard deck version shown. Transducer types vary: Abbott Transpac 458-0 transducers shown. 5

Overview A A 3 MKIII Pump Air Water System #ZP-GAU/TR/ #ZPFIT/SW/SS-4CPA--3 #ZP-SEA/OR/5 Mini Female Quick Connect #ZPFIT/SW/SS-QM-B-00 6

B Precautions & Warnings! B!! Never drop. Do not use if visible damage. Only pressurise reservoir with this line (C 3.).! Water level never less than cm (C 0).! Air pressures never more than 300kpa / 45psi! Air perfusion channel. Only use for pharyngeal swallow monitoring (C 9). 7

Normal Use C Assembly & filling of perfusate reservoir C Note: See E for Dentsleeve part # s Fill with particle free, degassed, distilled H 0 Add water, then place perfusion float barrier on top Blow-off valve Perfusate outlet hole O ring Reservoir outflow control device! Always use diffusion float barrier! Always use water filter Part #: APFIL/WA/ 3 Tighten firmly to seal May 003 8

C Normal Use For set-up and first use see E to E 7 C Check, connect & set air supply Click! Air Air Set to 00 kpa / 5 psi To gas supply 9 May 003

Normal Use C Connection & flushing of water manifold C 3 Connect to water manifold Part # ZPFIT/SW/SS-QM-B-B-00 Air line from control box Filter 3 Open 4 Open, vent and close May 003 0

C Normal Use C 4 Check flow values of each hydraulic resistor Standard resistor shown : flow is colour coded Flow ml/min Tube Colour Screw Colour Part # Alternative: Compact hydraulic resistor 0.0 Black White R0HRE/ST/(3) 0.0 Black Black R0HRE/ST/(3) 0.04 Black Red R0HRE/ST/4(3) 0.08 Red White R0HRE/ST/8(3) 0.5 Red Black R0HRE/ST/5(3) 0.3 Green White R0HRE/ST/30(3) 0.45 Green Red R0HRE/ST/45(3) 0.6 Green Black R0HRE/ST/60(3) Water flow on case The flow value is... Correct, go to Changed, go to C 5 E 5 or 6 May 003

Normal Use C Turn on water perfusion to fill each transducer C 5 Open pinch valve or clamp Fill transducers Fill transducers Prefered position for 3 way stopcock - other than PVB DPT-600 Note: PVB DPT-600 Transducer shown Note: Abbott Transpac 458-0 Transducer Shown

C Normal Use C 6 Transducer calibration (if required) Turn perfusion off - (C 5) 3 or use gravity calibration Set tranducer gain by applying standard external pressure Shows APCMA/TR/ 68cm H O, 50mm Hg Connect to transducers Pressurise with syringe Note: Shows Dentsleeve calibration manifold; made according to channel numbers and spacings Part # s APCMA/TR/to5 Note: Catheter must be attached to transducers (C 7) C 7 Connect catheter to transducers Select correct channel Twist & push on firmly Note: Catheter detail is for Dentsleeve product 3

Normal Use C Water injection procedure - each channel C 8 H O - no bubbles Open side-port 3 Inject 5ml of H O 4

C Normal Use C 8 Water injection procedure - each channel (continued) 4 Close side-port then remove syringe Note: Arrangement for PVB DPT-600 transducer shown 5

Normal Use C Check flow value for air perfusion manometry C 9 Identify air perfusion channel pinch valve! Only use air perfusion for detection of swallowing in pharynx Ensure resistor for air perfusion channel is correct! Air flow rate through hydraulic resistor is x00 water flow rate. Air flow ml/min Part # Black Black R0HRE/ST/(3) 4 Black Red R0HRE/ST/4(3) 8 Red White R0HRE/ST/8(3)! Must be installed between manifold and transducer to limit air flow into manometric catheter to less than 0ml/min May 003 Suitable compact resistors give airflow on case Air flow ml/min Water Flow Part # ml/min 0.0 R0HRE/CO/ 4 0.04 R0HRE/CO/4 8 0.08 R0HRE/CO/8 6

C Normal Use C 0 Observation & refilling of perfusate reservoir! If perfusate exhausted, large volumes of gas may be perfused down catheter Refill when at cm Release pressure Pull Back 3 Refill 4 Restart perfusion Particle free, degassed, distilled H 0 Tighten Repressurise 7

Normal Use C Maintain correct perfusate reservoir pressure C Monitor air pressure. Maintain at 00kPa/5psi. 00 5 50 kpa psi 00 DENTSLEEVE INT L Mfd. by Mui Scientific Check if air pressure drops 3 Check that air supply is adequate 00 At least 30 kpa/ 0psi 00 300 kpa psi 400 DENTSLEEVE INT L Mfd. by Mui Scientific OR open At least 000 kpa/ 50psi 8

C Normal Use C Prevention, recognition and correction of hydraulic resistor blockage Prevention of blockages - Always use perfusate filter Minimise perfusate particle load Double-distilled or de-ionised water No particles in reservoir (E-) Perfusate filter Part # APFIL/WA/ 3 Replace water perfusate filter regularly - (G-) 4 Check perfusate flow rates - assemble ml syringe barrel on transducer ml syringe 5 Start perfusion - (C -) Female/Female adapter Part # ZPFIT/VP/FTLLC-9 Transducer on rack 6 Measure time to deliver water Note: for resistor flow rates < 0.5ml/min use 0.5-0.50 ml syringe barrels 9

Normal Use C Prevention, recognition and correction of hydraulic resistor blockage C 7 Flushing of blocked resistors - remove resistor from circuit Transducer end 8 Backflush forcibly 9 Observe water flow 0 Re-install - (E 5) OR Remove resistor from circuit Insert flush support tool - (E 6) 3 Same steps as C.7 to.0 above 0

C Normal Use C 3 Compressor Always ground Turn on power switch. Power is on when switch is illuminated 3 Connect compressor air outlet to pump 4 Plug in compressor to pump control

Steps On Completion of Measurements D Perfusate reservoir D Depressurise 3 Remove lid Close and disconnect from manifold 4 Drain Reservoir 5 Remove float & air dry Lint & particle free surface! Do not transfer particles to internal surfaces of reservoir

D Steps On Completion of Measurements D Water manifold Close D 3 Air supply Turn off compressor OR Close cylinder valve 3

Set Up Procedures & Connection of Components E Air supply E! Installation only by an approved, qualified biomedical engineer Use air Note: Air chosen as:. Air dissolution in perfusate prevented by perfusate float barrier.. More suitable for gas perfusion manometry than N. 3. Usually more available and cheaper than N. 4. Available from wall supply and simple compressors. Set up for compressor, wall or air bottle supply Compressor Wall Air bottle E.3 E.4 E.6 or.9 4

E Set Up Procedures & Connection of Components E Air supply (continued) 3 Install next to pump deck or place on bottom shelf of trolley OR 4 Identify air line Black Part # ZPTUB/NS/ Part # ZPFIT/SW/SS-400-7- Connects to pump Air inlet (C.) Part # ZPFIT/SW/B-QC4-B-400ZN 5 Ensure air source is suitable: 30-40 kpa or 0-35 psi no oil or moisture filtered to 0.5 micron 6 Make appropriate connection to supply 5

Set Up Procedures & Connection of Components E Air supply (continued) E 7 Identify air line with Dentsleeve supplied regulator Factory set by Dentsleeve to deliver 300 kpa / 45 psi Part # ZPREG/HP/ Connect to pump air inlet (C.) Cylinder pressure gauge Medical international air pin index (Size C cylinders only) 8 Identify air line if no high pressure regulator supplied Black! 9 Connect to suitable locally sourced high pressure regulator with lockable outlet pressure control! 0 Adjust regulator to deliver outlet pressure of 300 kpa / 45 psi 6

E Set Up Procedures & Connection of Components E Perfusate reservoir prior to first use Check correct assembly of perfusate reservoir Blow-off valve Part # ZPFIT/SW/SS-4CPA-3 Closure Knob Part # FAS/PR/ O ring Part # ZPSEA/OR/ Reservoir outflow control device Gasket under lid Part # ZPSEA/GA/ Part # ZPFIT/SW/SS-QM-S-PM Always use water filter (C.) Part # APFIL/WA/!! Always use diffusion float barrier Part # FLO/PR/ Gasket in channel Part # ZPSEA/GA/ Outlet O ring Part # ZPSEA/OR/ Flush reservoir with water to remove any particles 7

Set Up Procedures & Connection of Components E Installation of pressure transducers E 3 PVB DPT-600 Abbott Transpac 458-0 Mount bar removable 8

E Set Up Procedures & Connection of Components E 3 Installation of pressure transducers (continued) 3 4 9

Set Up Procedures & Connection of Components E Installation of pressure transducers (continued) E 3 5 30

E Set Up Procedures & Connection of Components E 4 Water perfusion manifold removal Water perfusion manifold! Air perfusion (C 9 and E 7) 3

Set Up Procedures & Connection of Components E Water perfusion manifold removal (continued) E 4 Pinch valve 3 3

E Set Up Procedures & Connection of Components E 5 Standard hydraulic resistors Identify standard resistor with correct flow value. (C 4) Prime resistor with water. Use ml syringe for 0.6-0.5ml/min resistors 0.5ml syringe for lower flow rates Particle-free water 3 Inject till water flows from other end 4 Fill resistor connector as syringe withdrawn No bubbles trapped in luer 33

Set Up Procedures & Connection of Components E Standard hydraulic resistors (continued) E 5 5 Connect resistor to manifold side arm luer Ensure resistor connector remains water filled Manifold clamp Part # ZPFAS/RI/ 6 Push and rotate resistor firmly onto manifold luer 7 Open 34

E Set Up Procedures & Connection of Components E 5 Standard hydraulic resistors (continued) 8 Debubble downstream resistor luer Inject particle-free water into bottom of luer 9 Ensure no bubbles are trapped 0 Connect to pressure transducer Preferred position for 3 way stopcock Use special high pressure stopcock Push and rotate resistor firmly onto 3 way stopcock Abbott transpac 458-0 shown Part # ZPSTO/HP/ Water filled Note PVB DPT-600 transducer has integral stop-cock at transducer outlet 3 Final setup: Follow instructions from (C 7.) 35

Set Up Procedures & Connection of Components E Compact resistors E 6 Check flow values for each hydraulic resistor Prime resistor with water - Use flush tool 3 Place flush tool on silicone rubber connector 4 Use ml syringe for 0.6 0.5 ml/min resistors 0.5 0.5 syringe for lower flow rates 5 Inject until water flows from other end 6 Fill resistor connector as syringe withdrawn 7 Remove flush tool 36

E Set Up Procedures & Connection of Components E 6 Compact resistors (continued) 8 Connect as shown DENTSLEEVE 0.60mL/min Water Only #HRE/CO/60 To water manifold Note: Connectors are self-debubbling. PVB DPT-600 transducer shown. 37

Set Up Procedures & Connection of Components E Compact resistors (continued) E 6 9 Connect resistor to manifold sidearm DENTSLEEVE 0.60mL/min Water Only #HRE/CO/60 Manifold clamp Part # ZPFAS/RI/ 0 Open Final set-up - follow from (C 5.). 38

E Set Up Procedures & Connection of Components E 7 Air perfusion circuit! N perfusion into pharynx may dilute inspired oxygen - use only air for perfusate reservoir pressurisation in small children, as this gas is also used for gas perfusion circuit Identify circuit Locate air manifold and check correctly connected 3 Resistor must always be installed To transducer 39

Set Up Procedures & Connection of Components E Air perfusion circuit (continued) E 7 4 Select resistor that controls airflow to less than 0ml/min! 5 Airflow is x00 waterflow Air flow Waterflow ml/min ml/min Part # 0.0 Black Black R0HRE/ST/(3) 4 0.04 Black Red R0HRE/ST/4(3) 8 0.08 R0HRE/ST/8(3) Red White Appropriate compact resistors show airflow on case DENTSLEEVE Air ml/min WATER 0.0mL/min HRE/CO/ DENTSLEEVE Air 4mL/min WATER 0.04mL/min HRE/CO/4 DENTSLEEVE Air 4mL/min WATER 0.04mL/min HRE/CO/4 Air = ml/min Air = 4ml/min Air = 8ml/min 6 Install air flow resistor between air manifold luer and transducer No 3 way stopcock 7 No water prime required! Air perfusion to be used only for pharyngeal manometry To manifold luer 40

Air 4mL/min WATER 0.04mL/min F Cleaning & Disinfection Cleaning of plastic parts F Clean with cloth and mild detergent Sterilization of perfusate reservoir F Gas sterilize only Sterilization of manifolds F 3 Water manifold Air manifold Autoclavable Hydraulic resistors F 4 DENTSLEEVE HRE/CO/4! Not autoclavable Autoclavable Cleaning of compressor F 5 Wipe with a dry cloth 4

Regular Maintenance G G Every 3 months (or as needed): perfusate water filter Replace Part # APFIL/WA/ 4

G Regular Maintenance Every year or as needed: gas filters G! To be done only by an approved, qualified biomedical engineer Disconnect gas supply then open control box (G 3.) Use correct filter Part # APFIL/GA/ Gas (Black) 3 Replace filter 4 Check for air-tightness Note: Gas filter life depends on quality of air always use medical grade 43

Regular Maintenance G G 3 Service of control box by Dentsleeve Undo screw that holds control box Send control box to Dentsleeve 44

G Regular Maintenance Compressor Dessicant G 4 Check dessicant is orange. When it becomes green, replace. Orange GREEN OR Spread granules evenly onto tray, one granule deep. Heat granules for approximately 5 hours. (or until it turns back to its original orange color) at 5 o C (50 o F) in a conventional oven. Cool dessicant before replacing back into canister. 45

Problem Solving H Diagnosis of abnormal air consumption H Check compressor connection For compressor malfunction, please contact Dentsleeve OR Air Check regulator/cylinder connection Not connected Tighten Pressurise air circuit - Turn on Air 00 50 50 3 Adjust air pressure 4 Turn off Air kpa psi 00 DENTSLEEVE INT L Mfd. by Mui Scientific 46

H Problem Solving H Diagnosis of abnormal air consumption (continued) 5 Recognise leakage - Wait hour either or 6 Drop less than 33 kpa/5psi No significant leak 7 Drop more than 33 kpa/5psi Significant leak 8 Follow (H-.9) below 47

Problem Solving H Diagnosis of abnormal air consumption (continued) H 9 Check for leakage outside control box 00 50 50 0 If there are bubbles, replace fitting - Part # ZPFIT/SW/SS-QM-B-00 kpa psi 00 DENTSLEEVE INT L Mfd. by Mui Scientific! Consult approved, qualified biomedical engineer to do check (H-. to H-.7) below Air Leakage? 3 Open pump control box. 4 Remove control box lid/back 48

H Problem Solving H H Diagnosis of abnormal air consumption (continued) 5 Filter leakage - tighten or replace 6 Check rest of air circuit 49

Problem Solving H Abnormal air consumption - perfusate reservoir leakage H First exclude air circuit leakage - (H ) Set pressure 00 50 50 kpa psi 00 DENTSLEEVE INT L Mfd. by Mui Scientific Tighten 3 Set up reservoir. Remove blow-off valve cap 4 No water 5 Pressurise 6 Locate site of leakage - observe underwater Still pressurised 7 Locate site of bubbling 50

H Problem Solving H H Abnormal Air consumption - perfusate reservoir leakage (cont.) ZPFIT/SW/B-QM-B-00ZN 8 If leak elsewhere, dismantle reservoir 9 Identify cause of leakage at base 0 Gasket clean and intact? Identify cause of leakage at top Underside Flip Check O ring Part # ZPSEA/OR/ Gasket underneath clean and intact? Part # ZPSEA/GA/ 5

Specifications, Support, Spare Parts & Accessories I Specifications I Air High pressure regulators Recommended first stage regulators (*if supplied) Medical grade only CONCOA Medical Air (Air pin index) Inlet pressure Pump inlet pressure (from gas supply first stage regulator) Air 67-300kPa (5-45 psi) Compressor 30-40kpa (0-35 psi) Supply lines (from gas bottles) Air (black) line Air Male Inlet Air female connector to perfusate reservoir Air supply on pump Driving air supply (pump) Air flow rate ± 0% Adjustable 0-50 kpa (0-36psi) Flow restricted to 30ml/min at 00kPa (5psi) Pressure relief valves Perfusate reservoir relief valve Control box inlet overpressure relief valve Preset to 00 kpa (9psi) Preset to 300 kpa (45psi) Filtration Air 0.5 micron male/female luer connection disposable disc filter Contact Dentsleeve for advice See contact details on front cover Technical Support I 5

I Specifications, Support, Spare Parts & Accessories H Spare Parts I 3 Spare Parts Kit x 5x x APFIL/GA/ Gas filters, for installation within pump control box. APFIL/WA/ Water filters, for filtration of water perfusate ZP-OCD/PR/ Reservoir outflow control device, for perfusate outflow. Regulatory Information I 4 EC REP Authorised European Representative SOLAL Rue du Travail 67000 Strasbourg FRANCE Tel: 03 88 3 89 85 Fax: 03 88 75 7 3 FDA Approved CE Marked Health Canada Licensed 53