APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

Similar documents
Multiple Gas#5 GC configuration Jan 2016

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Agilent 490 Micro GC Natural Gas Analyzer

System Overview. TCD Detector temperature setpoint Regulator

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

This GC has two columns (I think new columns can be ordered from Restek)

High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

COAL MINES TECHNICAL SERVICES, GAS DETECTION RELATED SERVICES FOR MINES RESCUE

Evaluation of linear and step gradient performance and retention time precision of the Agilent 1200 Series Rapid Resolution LC

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Experiment GC : Analysis of BTEX by GC-FID

REFINERY GAS ANALYSIS BY GAS CHROMATOGRAPHY

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Advancements in Gas Chromatography Analyzers - Keeping up with New Technology. Chuck Runkle Gas Phase Product Specialist ASTS June 2013

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Retention Time Locking: Concepts and Applications. Application

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1

SRI Multi Gas Analyzer 2016

Rapid and Reliable Detection of Dissolved Gases in Water

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

* Analysis of TCG ( Total Combustible Gas ) and 6 kinds of Gases * High Sensitivity for C 2

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Your local gas generation partner. Precision series Modular gas generation solution for GC.

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

Applied Technology and Best Practices in CEE. Conference

Low-Pressure Retention Time Locking with the 7890A GC. Application. Authors. Abstract. Introduction HPI

Automated Determination of Dissolved Gases in Water Anne Jurek

SRI 8610C Gas Chromatograph Multiple Gas #3 GC configuration

Bernhard Thomas September EnCal 3000 prochain Energy in Gas with ROI

IRwin Methane Leak Detector. Innovation makes the difference

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Challenges of gas monitoring and interpretation in underground coal mines following an emergency

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

ISO INTERNATIONAL STANDARD

Carrier Gases in Capillary GC

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

Gas Chromatographic Applications of the Dielectric Barrier Discharge Detector

Agilent G Rebuilt Electron Capture Detector

ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION

CONNECTIONS. The Thermal Conductivity Detector

May 1, By John Heim,Doug Staples

The Preparation of Low Concentration Hydrogen Sulfide Standards Gulf Coast Conference September 12, 2002 Galveston, TX Paper 050,12:45 pm Moody F

The Role of Gas Monitoring in the Prevention and Treatment of Mine Fires

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Hiden QIC Series Gas Analysers Real-time gas analysis systems

ISO 7941 First edition

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

The analysis of complex multicomponent

Gas Leak Measurement of TRT Barrel Module

A Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction

ANALYSIS OF GASES PRODUCED BY THREE UNDERWATER CUTTING DEVICES

GCMSD-Headspace Analysis SOP

Best Practice for Identifying Leaks in GC and GC/MS Systems

New improved robot for gas kinetics

Addressing the World Shortage of Helium

Quantitative Analysis of Hydrocarbons by Gas Chromatography

Extend the time window for the detection of low level anabolic androgenic steroids and their metabolites

Ultimate Performance in Gas Chromatography

instrumentation Pulsed Discharge Detectors Non-Radioactive, Multiple Mode Electron Capture / Helium Photoionization

Model Description

Agilent 7800 ICP-MS. Specifications and Typical Performance. Fast track metals analysis with Solution-Ready ICP-MS

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

Real-time Analysis of Industrial Gases with Online Near- Infrared Spectroscopy

Organic Elemental Analysis Series II CHNS/O Elemental Analyzer. proven performance. reliable results

Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples

Buck Scientific, Inc.

Svensk läkemedelsstandard

Since the beginning of production of oil for transformers

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems

APPLICATION NOTE. GC Integrated Permeation Device

CORESTA RECOMMENDED METHOD Nº 67

Gases and welding distributors continue

Online DGA-monitoring of power transformers

A Revised Framework for the Transformer DGA Guide

Micro GC Fusion Pre-Installation Checlist

Comparison of Different DGA (Dissolved Gas Analysis) Methods for the Condition Assessment of Power Transformers

Zebron Gas Management

Development of DMS and Acetonitrile, and Formaldehyde gas standards. Gwi Suk Heo, Yong Doo Kim, Mi-Eon Kim, and Hyunjin Jin

Describe Flammable Gas Measurement

no peaks missing peaks too small peaks increasing retention decreasing retention

Super-Clean Gas Filters from Restek

M/s. Thermo fisher Scientific

LFE OEM TCD - Thermal Conductivity Detector

Carrier Gases in Capillary GC

Operation of the Perkin Elmer TGA-GC/MS

3. Coupled Techniques in Thermal Analysis

INSTRUMENT INSTRUMENT YL INSTRUMENT

Dissolved Gas Analysis - Past, Present and Future

Comparative analysis of mass spectral matching for confident compound identification using the Advanced Electron Ionization source for GC-MS

Svensk läkemedelsstandard

Converting Helium Carrier Gas GC Methods Nitrogen and Hydrogen

CALCULATING THE SPEED OF SOUND IN NATURAL GAS USING AGA REPORT NO Walnut Lake Rd th Street Houston TX Garner, IA 50438

Detecting Hydrogen in Helium Streams. Samuel Goodman. Pittsford Mendon High School. Rochester, New York. Advisor: Dr.

CHEMTRACE Fremont Blvd. Fremont, CA 94538, Tel: (510) Fax: (510)

Elimination of Percent Level H 2 S Calibration Gas from Flare Gas Monitoring Systems Measuring Total Sulfur, H 2 S and BTU.

Transcription:

APPLICATION NOTE Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT The INFICON 3000 Micro GC provides two fast and accurate solutions for the analysis of coal mine gas components. When configured with two channels, the GC baseline separates the individual components in less than 60 seconds and within 70 seconds when configured with four channels. The developed methods provide fast, real-time coal mine gas-detection solutions to provide rescuers with critical information before they attempt rescues in the mine. In this application, the trace indicators of coal mine explosion are also detected by the performance-enhanced INFICON 3000 Micro GC by reducing the big tailing of the main component. Evaluation of data from this method can help prevent coal mine explosions. INTRODUCTION Coal mine gases released during mining operations are serious safety hazards. When the released gases are mixed with air to the critical concentration, they are highly explosive and have been responsible for explosions in underground coal mines all over the world. Because the concentration of carbon monoxide, ethylene, and acetylene coupled with the ratio of paraffin and olefin in gas from coal mines can be used to predict a fire or explosion in an early stage, it is important to determinate the concentration of these gases in coal mines in real time. The INFICON 3000 Micro GC is a powerful tool for the fast analysis of gas components [1-3]. Its robust construction and multiple channel configurations make analyzing complex samples quick and easy. Each channel or module is a self-contained GC comprising a micro-machined injector and detector and a highresolution capillary column. Up to four different channels expand separation capabilities without compromising speed, making troubleshooting and repair simple. The performance-enhanced INFICON 3000 Micro GC is specially designed for the analysis of low-concentration components. Analytes in coal mine gas, ethylene and ethane compounds, can be used as indicators of a potential coal mine explosion. When the concentration of these components approaches approximately 1 ppm, an explosion could occur if effective precautions are not taken. However, on a normal µgc these peaks are on the tailing of air peaks, affecting the detection limits of these indicators, as well as the reliability and repeatability of the tests. In this application, two fast solutions are developed to analyze gas components from a coal mine by the INFICON 3000 Micro GC. The trace-level (1 ppm) spontaneous-combustion indicators, such as ethylene, ethane, and acetylene in coal mine gas, are also analyzed on performance-enhanced modules. EXPERIMENTAL Sample The concentration and composition of the coal mine gas standards for the fast solution are shown in Table 1. Table 1 Standard compositions and their concentrations Component Concentration (µl/l) Component Concentration (µl/l) O 2 400 C 2 H 4 200 H 2 300 C 2 H 6 600 CH 4 650 C 3 H 6 500 CO 500 C 3 H 8 450 CO 2 400 C 4 H 10 400 C 2 H 2 350 1 of 6

The concentrations of low-level coal mine gas components are ethylene, 1.1 ppm; ethane, 1.1 ppm; acetylene, 1.1 ppm; and air as the balance. These standards were all provided by Beijing AP BAIF Gases Industry Co. Ltd. (Beijing, China). Configurations and Analytical Conditions Two-Channel Configured Micro GC Solution for Fast Coal Mine Gas Analysis Two channels are used for the fast analysis solution on the Micro GC. The first channel is a molecular sieve column (molecular sieve 5Å, 10 m 0.32 mm 12 µm) for the separation of O 2, N 2, CH 4, and CO. The second channel is a PLOT U column (PLOT U, 8 m 0.32 mm 30 µm) for the separation of CO 2, C 2 H 4, C 2 H 6, C 2 H 2, and C 3 compositions. The injectors used are all variable-volume type and helium is used as carrier gas. The analytical conditions are shown in Table 2. Channel Table 2 Analytical conditions of two-channel Micro GC Molecular Sieve 5Å mm 12 µm) PLOT U (8 m 0.32 mm 30 µm) Sample inlet temperature ( C) 60 60 Injector temperature ( C) 80 80 Column temperature ( C) 100 85 Inject time (ms) 20 20 Run time (s) 100 100 Column pressure (psi) 40 40 Four-Channel Configured Micro GC Solution for Comprehensive and Fast Coal Mine Gas Analysis A four-channel configuration of the Micro GC is listed in Table 3. The first channel is used to analyze H 2 in coal mine gas, and Ar is used as the carrier gas because of the sensitivity problem. The fourth channel is for the separation of C 3 H 6, C 3 H 8, and C 4 H 10 in coal mine gas. The second and third channels are the same as those in the two-channel configured Micro GC. Table 3 Four-channel configuration of Micro GC Channel Molecular Sieve 5Å mm 12 µm) Molecular Sieve 5Å (8 m 0.32 mm 12 µm) Sample inlet temperature ( C) 60 60 Injector temperature ( C) 80 80 Column temperature ( C) 60 100 Inject time (ms) 60 20 Run time (s) 100 100 Column pressure (psi) 40 40 Carrier gas Ar He Channel PLOT U mm 30 µm) PLOT Q mm 10 µm) Sample inlet temperature ( C) 60 60 Injector temperature ( C) 80 100 Column temperature ( C) 80 120 Inject time (ms) 20 20 Run time (s) 100 100 Column pressure (psi) 40 40 Carrier gas He He Low-Level Coal Mine Indicators Analysis on Performance-Enhanced INFICON 3000 Micro GC The low-concentration coal mine gas indicators were detected on the performance-enhanced module using a PLOT U column. The separation conditions are shown in Table 4. Table 4 Separation conditions of coal mine indicators on performance-enhanced module Analytical column PLOT U (8 m 0.32 mm 30 µm) Injector Large volume injector Carrier gas He Sample inlet temperature ( C) 50 Injector temperature ( C) 50 Column temperature ( C) 50 Sampling time (s) 10 Inject time (ms) 200 Run time (s) 90 Column pressure (psi) 30 Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC 2 of 6

RESULTS AND DISCUSSION Two-Channel Configured Micro GC for Fast Analysis of Coal Mine Gas O 2, CO, and CH 4 were separated within 40 seconds and CO 2, C 2 H 4, C 2 H 6, C 2 H 2, and C 3 within 60 seconds. With this configuration, the fast, real-time analysis of the main concerned gas compositions was achieved. Figures 1A and 1B show the chromatograms of channels A and B, respectively. Table 5 shows the run-to-run repeatability by relative standard deviations (RSDs) of the peak area. The RSDs of separated compositions are all less than 1%, which means that the INFICON Micro GC with this configuration is stable and reliable. Figure 1 Chromatograms of coal mine gases O 2 CH 4 CO A: Channel A, Molecular Sieve 5 Å CO 2 C 2 H 4 C 2 H 6 C 2 H 2 B: Channel B, PLOT U Table 5 RSDs for peak area Compound Avg. RT Avg. peak area Peak area SD RSD (n = 10) O 2 0.322 1140.63 4.62 0.41 CH 4 0.519 1269.57 2.42 0.19 CO 0.644 1190.11 10.12 0.85 CO 2 0.270 918.22 3.38 0.37 C 2 H 4 0.295 453.82 2.14 0.47 C 2 H 6 0.317 1478.03 4.43 0.30 C 2 H 2 0.404 695.19 2.53 0.36 C 3 H 6 and C 3 H 8 0.889 2519.28 3.10 0.12 Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC 3 of 6

Four-Channel Configured Micro GC for Fast and Comprehensive Analysis of Coal Mine Gas For Micro GC thermal conductivity detector (TCD) systems, helium is the best choice of carrier gas when analyzing coal mine gas because it has the highest sensitivity for all of the analytes except for hydrogen. When hydrogen is the target analyte, argon, not helium, is chosen as the carrier gas. This is described in the application for the detection of hydrogen in fuel cell reformer gases by Micro GC [1], in which argon shows the most linear performance and the most sensitivity for hydrogen. In this solution for coal mine gas analysis, hydrogen and n-butane were separated by the four-channel configured Micro GC; hydrogen analysis was realized in channel A and argon was used as the carrier gas. Channel D with a PLOT Q column was used for the analysis of C 3 H 6, C 3 H 8, and n-c 4 H 10. Figure 2 shows the chromatograms of the analytes in the four Figure 2 Chromatograms of coal mine gases by a four-channel micro GC H 2 A: Molecular sieve, Argon as the carrier gas O 2 CH 4 CO B: Molecular sieve CO 2 C 2 H 4 C: PLOT U C 2 H 6 C2 H2 C 3 H 6 C 3 H 10 D: PLOT Q, carrier gas in Channels B, C, and D are all Helium Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC 4 of 6

The run-to-run repeatability of the four-channel configured Micro GC is shown in Table 6. This table illustrates that the Micro GC shows excellent precision, with all RSDs below 1%. Table 6 Reproducibility of four-channel micro GC solution Compound Avg. RT Avg. peak area Peak area RSD (n = 10) H2 0.451 606.61 5.89 0.97 O2 0.322 1140.60 4.58 0.40 CH4 0.519 1269.59 2.43 0.19 CO 0.644 1190.11 10.11 0.85 CO2 0.314 918.22 3.37 0.37 C2H4 0.345 453.89 2.24 0.49 C2H6 0.375 1478.07 4.45 0.30 C2H2 0.489 695.19 2.59 0.37 C3H6 0.474 1397.50 5.28 0.38 C3H8 0.498 1283.31 4.49 0.38 C4H10 1.085 1274.80 8.42 0.66 Low-Level Coal Mine Indicators Analysis on Performance-Enhanced INFICON 3000 Micro GC The INFICON 3000 Micro GC demonstrates the ability to produce chromatograms with peaks generated by 1 ppm analytes that are well-resolved and well-shaped with flat baselines to reliably detect and accurately integrate such peaks. This helps make the accurate quantitation of trace-level analytes possible and reliable. Figure 3 shows the 1 ppm ethylene, 1.1 ppm ethane, and 1.1 ppm acetylene analysis on a PLOT U module. The INFICON 3000 Micro GC exhibits highsensitivity performance, as the 1 ppm gas is easily detected with an excellent signal-to-noise ratio (greater than 3). Figure 3 Chromatogram of low-level coal mine indicators analyzed on a performance-enhanced PLOT U column Air Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC 5 of 6

Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC 6 of 6 CONCLUSIONS The INFICON 3000 Micro GC provides flexible and reliable configurations for fast coal mine gas analysis. These configurations achieve fast, dependable results, which ensures the real-time analysis of coal mine gases. By using the performance-enhanced INFICON 3000 Micro GC system, the concentration of spontaneous combustion indicators lower than 1 ppm are welldetected, with a signal-to-noise ratio greater than 3. The result meets the requirement of coal mine safety, which can help mine operators detect explosion potential at an early stage. REFERENCES 1. Analysis of Fuel Cell Reformer Gases in Seconds Using the INFICON 3000 Micro Gas Chromatograph, INFICON publication diae67a1. 2. Complete Analysis of Refinery Gases in Seconds Using the INFICON 3000 Micro Gas Chromatograph, INFICON publication diae60a1. 3. Micro GC Analysis of Natural Gas and LPG in Seconds Using the INFICON 3000 Natural Gas Analyzer, INFICON publication diae59a1. Note: 1 psi = 6.89457293 kpa TM www.inficon.com reachus@inficon.com Due to our continuing program of product improvements, specifications are subject to change without notice. diae68a1-a 2010 INFICON