FMA Small Animal - Rodent

Similar documents
User s manual for the SC32 and SC96 form-fitting recording chamber and microdrive systems

Tridak Model 1060 User Guide. Syringe Filling System

ICE PROTECTION - PROPELLER

TJF-Q180V Cleaning and Disinfection Checklist

Patients Guide for Self Administering Intravenous Antibiotics General points

Vacuum Bagging Wings Instruction Manual Purdue University

The Butterfly. Steep Trendelenburg Gel Bean Bag Positioner USE & CARE GUIDELINES DAVID SCOTT COMPANY

BPS-2, Rev TABLE OF CONTENTS

QE-1, Rev Warner Instruments, Inc Dixwell Avenue, Hamden, CT (800) / (203) (203) fax

Fiber Optic Lighted Bubbler Spillway Pot (DLP-45) Installation Manual

Warner Instruments, Inc Dixwell Avenue, Hamden, CT (800) / (203) (203) fax

Standard. Standard. Venepuncture Arm. Light. Part No: Brown. Part No: Black. Part No: 00332

Ultrasound Compatible Lumbar Puncture/ Epidural Simulator

limbsandthings.com Advanced Venepuncture Arm User Guide For more skills training products visit Limbs & Things Ltd.

BioAerosol Nebulizing Generator. Operation and Maintenance User Manual

TJF-Q180V Cleaning and Disinfection Checklist

NHR Powder Feed Hopper

Gas Lines. Technical Manual. Sumitomo (SHI) Cryogenics of America, Inc Vultee Street Allentown, PA U.S.A.

1. Connect the attached anesthesia mask to your anesthesia system using the included tubing and connectors.

Contents. Stainless Steel Side Block. 1.1 Separating the Side Block. Stainless Steel Side Block Reassembly of. Assembly from the Helmet Shell

[USING THE VITROBOT April 9, Using the Vitrobot

Procedure 85 Attaching The Humidifier To The Oxygen Flow Meter Or Regulator. Procedure 86 Administering Oxygen Through A Nasal Cannula

INSTRUCTION MANUAL. January 23, 2003, Revision 0

Owner s Guide. ElevateSUP.com

EcoCell Vertical Electrophoresis Unit

U STAND INSTALLATION INSTRUCTIONS

MULTICHANNEL. Labnet International, Inc.

Thermocouple Temperature Measuring Probes

200 STRINGING MACHINE

Spray Gun 134-A. 1 Liter High- Pressure Gel Coat. Usage Instructions & Parts List

Standard. Bag & Stand Advanced Venepuncture Arm. Light. Part No: Brown. Part No: Black. Part No: 00442

VAC-U DRIVE Swinging float arm for Acrobat like agility 36 Fin disk with DEEP VEE Scoop

MyoSure Setup Simplified

Hydraulic Punch Drivers

Lectric Cycles Mid-Drive Electric Motor Installation

600 / 600FC OWNER'S MANUAL

COLD WELD COPPER TUBE SAMPLING EQUIPMENT AND PROCEDURES

SERIES 20 CHAMBERS THE RC-25 AND RC-25F CHAMBERS ASSEMBLY

5 Gallon Pressure Pot with HVLP Spray Gun and Hose

Encore PE 50-lb Feed Hopper

DETERMINING OPTIMUM RESIDUAL ASPHALT CONTENT (RAC) FOR POLYMER-MODIFIED SLURRY SEAL (MICROSURFACING) MIXTURES

WELLS JOHNSON HIGH VOLUME CANISTERS

RayPette Plus Autoclavable Pipette. User Manual

COMPACT TRIPLE WALL OUTLET STATION MODEL INSTALLATION AND OPERATING INSTRUCTIONS

Unisense In Situ UniAmp Connector System

Combination Breathing Apparatus

ST 18 Aluminium Gas Model

INSTALLING YOUR CLC RUDDER

Instruction Manual Updated 7/26/2011 Ver. 2.2

Hydraulic Punch Drivers

ST 18 Aluminium Gas Model

Underwater Housing for Canon EOS M

Miniball Coulomb Excitation Target Chamber

Wet Lamination of Dry Film Photoresist for the Hobbyist

Instructions for use: Of adhesive ski skins and cut-to-size sets

PH01 Perfusion Cannula Manual

Aseptic Techniques. Techniques for Sterile Compounding. Pharmacy Technician Training Systems Passassured, LLC

Installation of Your SprayMaster System

QRPGuys Single Lever Paddle w/base

Prestige Endoscopic Graspers Instructions for Use

Standard. Abdominal Aortic Aneurysm (AAA) Repair Trainer. Part No: 60610

Operating Instructions

B. Assemble the Method-25 sample train with a heated probe/filter.

CLEANROOM GEL GRID CEILING SYSTEM PRODUCT DATA Division

The pipettes cover a volume range from 0.1µl to 10ml.

Product Information News

Thanks for choosing EJAI SOLAR product

Hygea Ultrasonic Validation Kit

Pre-Paint>Fuselage>Empennage>Fit vertical tail fin. Objectives of this task: Materials and equipment required: Fit the spar extender

Drive Belt Instructions

Abdominal Aortic Aneurysm (AAA) Repair Trainer User Guide

the typical anatomy of the ear canal. The eartips should point in a forward direction as you insert them into your ear canals.

Written By: Scott Havard

Multi-Probe Sampling System User Guide & Manual

[The Octopus vacuum lifting system] 9

The Outer Seal is highly abrasion resistant and provides mechanical stiffness to the splice in order to better match the cable stiffness.

NHR-X-X Encore Feed Hoppers

STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE

Basic Install Guide. Parts: 1 P a g e

3M Littmann Master Cardiology Stethoscope

INSTRUCTION MANUAL CZ 630/631

Shimano Di2 Installation on S5

USER MANUAL LTH-255A Hand Piece

Safety Huber Needles

O P E R A T O R S M A N U A L

A hose layline contains important information for specifying the replacement assembly: manufacturer, hose trade name, working pressure and hose ID.

MODEL ANALOG ALTIMETER USER'S MANUAL 330KHZ, 1000M DEPTH RATED 0.2 TO 100FT OPERATING RANGE ANALOG OUTPUT

12S 1st Stage. -Maintenance Procedure-

FIRST TEAM SPORTS, INC Storm Portable Series Assembly Instructions

Specialists in HTM Medical Gas Pipeline Equipment AVSU/Module Operating and Maintenance Instructions

1.0 PURPOSE The purpose of this procedure is to describe the use of basic aseptic technique.

Installation Instructions

STERILIZABLE DISSOLVED OXYGEN SENSORS

ESA Elbow Sealing Adapter Kit

iphone 6 Full Display Assembly Replacement

The P-Rod Double by The Airgun Lab Installation Instructions Revision F 08/01/2016 page 1 of 18

Standard and Low-Flow Metric Powder Feed Pumps

Unblocking the MicroSprayer Aerosolizer Model IA-1C With the High Pressure Reverse Cleaning Adapter. Instructions for Use

602 STRINGING MACHINE OWNER'S MANUAL

Brixham Archers Arrow Workshop 2015

Transcription:

MicroProbes for Life Science FMA Small Animal - Rodent User Instructions Version: August, 2015 18247-D Flower Hill Way Gaithersburg, MD 20879 Phone: +1 301 330-9788 Fax: +1 301 330-9667 w w w.microprobes.com support@microprobes.com

TABLE Table Of OF Contents CONTENTS Introduction How to Sterilize Removing the FMA from Plastic Box Removing the Carbo-Wax Implanting the FMA Connector Maintenance Impedance Testing 3 4 4 5 6 9 10 Terms and Conditions 10 Contact Information 11 2

Introduction The Floating Microelectrode Array (FMA) features an array head connected to a connector using a thin flexible silicone cable, allowing the array to accommodate natural tissue motion with minimal restriction. Due to this design, special care must be taken during handling and implantation to ensure proper function. Each FMA is visually and electrochemically inspected before shipping, with specifications and impedances provided within the included data sheet. To avoid accidental damage, it is recommended that the FMA be sterilized within its box and not be removed and handled until implantation. Arrays are shipped coated with a protective droplet of carbo-wax. This wax MUST BE REMOVED before implantation, using the procedure detailed below. 3

How to Sterilize It is recommended that FMAs NOT be autoclaved. While the materials within the FMA are all autoclave compatible, the fragile nature of the array and the special housing required during sterilization make autoclaving a poor alternative to ETO. For gas (ETO) sterilization, it is recommended that the lid of the FMA box be propped open slightly before insertion into a pouch, to maximize air flow. This is often done by positioning gauze pads between the lip of the box and the lid and gently holding the lid shut with surgical tape, creating a small gap. ETO will penetrate the protective carbo-wax. The sterilizer should be set to long cycle (24 hour) mode, to allow for proper out-gassing of the soft materials in the array. Removing the FMA from Plastic Box Polyimide (Kapton) tape is used to secure the FMA within the box, to prevent movement during shipping and handling. We recommend that the following procedure be used to safely remove the FMA before use: 1. 2. Completely remove the tape and wax paper securing the silicone cable. Remove the tape securing the top of the connector. 3. Grasping the connector only (never grasp the connector and cable at the same time), lift it off the double-sided tape by gently tilting the connector forward and twisting it until it comes free. A small squirt of isopropyl alcohol can be used to loosen the adhesive. 4

4. 5. 6. Securely hold the connector in such a way as to minimize tension and torsion on the cable. This is critical for FMAs with short cable lengths. Using a syringe, gently flush the base of the array head substrate with isopropyl alcohol, being careful to not dissolve the carbo-wax (polyethylene glycol) droplet. The alcohol will dissolve the glue on the polyimide tape and release the substrate. If the array head does not lift freely, carefully pry the substrate loose using a small spatula or needle. The FMA assembly may now be lifted from the container. Removing the Carbo-Wax Carbo-wax can be easily dissolved in distilled water or saline at room temperature, though heating the water to 50 C will decrease the time required for this process to 3-5 minutes. If it is necessary to test or inspect the array prior to implantation, the carbo-wax can be dissolved by simply immersing the array head into a warm water bath while carefully holding the connector in place. Once testing is completed, it is recommended that carbo-wax be re-applied to the array head for protection until implantation. Carbo-wax can be re-applied to the array by heating carbo-wax powder in an aluminum dish to 50 C and applying small droplets of the melted wax to the base of the array substrate using a thin (~0.003 diameter) wire, taking care to not touch the electrodes. Repeat adding droplets until the array is fully encapsulated. Methods to dissolve the wax at the time of implantation vary with the animal model and implantation procedure, but a common method is the following: 5

1. 2. 3. Securely grip the connector in place while holding the array head using a suction tip, as described later. While carbo-wax is biocompatible, many prefer protecting the brain surface by laying a sterile silicone or wax sheet or gauze pad beneath the array during the removal process. Draw normal STERILE 50 C saline into a syringe and carefully squirt it over the carbo-wax, collecting the run-off using low suction. Implanting the FMA It is critical that the array electrodes only ever come into contact with soft tissue, as even light contact with hard surfaces or bone has the strong likelihood of bending electrodes or blunting tips. Attempting to implant such an array will likely result in damaged tissue, buckled electrodes, and improper function. This guide is intended for small animal models, which typically feature very small craniotomies and short cable lengths. For implantation into larger animals, a different guide is available: User Instructions for FMA, Large Animal. While the focus of this guide is cortical implantation, those implanting into other locations should be able to adapt much of the procedure. Small animal models can present a challenge during implantation, due to the very small craniotomy and limited amount of space available on the cranial surface. Lateral cortical targets such as barrel present an even greater challenge due to the curvature of the skull and awkward location. The most critical concern is that the tension on the array be minimized during implantation. While the silicone cable is very flexible, if bent it can create enough spring force to pop the array out of the brain if unsecured. The size and shape of the craniotomy is another concern, as care 6

must be taken to ensure that it is cut large enough to accommodate the entire array footprint as well as the cable. Users have reported catching corners of the FMA array on bone edges while implanting into too small of a craniotomy, resulting in a tilted array and severe tissue damage. In most cases it is preferred that the microelectrodes be oriented 180 degrees relative to the top of the connector, which should be the orientation at shipment. However, depending on the curvature of the skull and brain, it may be preferred that the electrodes be oriented at a different angle in order for the array to be perfectly orthogonal to the tissue when the connector is fixed to the bone. To change the orientation of the array, hold the connector with hand, and carefully grip the cable with the thumb and forefinger of the other hand, at least 1 cm away from the array substrate. Twist the cable with your fingertips about 2 complete turns in the direction you wish the array to be tilted. Carefully release the cable, and repeat as necessary until the array is pointing in the correct orientation. When cutting the craniotomy, be sure that the edges extend beyond not just the electrodes but also the edges of the substrate, as it will need to rest completely within the craniotomy and flush with the brain. If needed, Microprobes for Life Science can provide you with a blank substrate template that you can sterilize and use for sizing purposes. Be sure to cut a groove for the cable, which extends directly out of the side of the substrate. Also, be sure that the dura is completely retracted from the target tissue, as thick folds near the edges can catch electrodes and cause buckling and skewing. Due to the short length of the cable, it is critical that the cable remain straight and relaxed throughout the implantation to avoid any twisting or lifting forces on the array. We recommend the use of a second manipulator arm to hold the connector and allow it to be moved and secured at the ideal location until the implantation is completed. 7

The best way to manipulate the array head is by using a vacuum tip which holds the substrate in place through suction on the flat upper surface. Microprobes for Life Science retails a Vacuum Inserter Tool specifically designed for this purpose, which includes an electric vacuum pump and a specially-ground metal cannula. Users have also reported success using low house suction connected to blunt-tipped syringes and steel, glass, or plastic pipettes. The quality of the tip is crucial, as an improper seal could result in low suction force and the unintentional release of the array. Once the array head is captured on the end of the suction tip, the carbo-wax should be dissolved using the method previously described in this guide. Once the carbo-wax is completely dissolved, the FMA array should be lowered and positioned about 5mm above the target area. Very slowly, lower the FMA into the cortex using a micromanipulator. If implanted at a proper speed, the pia should not dimple more than 2 mm. Users have reported success using a step-wise implantation, where they insert a short distance and then wait for the depressed pia to rise up around the electrodes before continuing. Step size varies between users, with some preferring under 0.5 mm per step and others preferring 1 2 mm per step. It is imperative that the array be implanted perpendicular to the surface of the cortex, or the tissue or array may be damaged during insertion. Some users have reported success inserting at very high speeds (10 m/sec) using the rapid insertion tool employed with Utah-style arrays. Despite this, MLS recommends the very low speed method, due to the larger volume of experience. 8

Once the array substrate is flush with the pial surface, the FMA is completely implanted. At this point, it is critical to secure the cable in place at the edge of the craniotomy. We recommend using a skull screw fixed close to the craniotomy edge, as it proves a large and firm surface for cementing the cable with a medical-grade high-viscosity biocompatible cyanoacrylate (such as Loctite #454), or even better with a light-activated dental composite resin. Once the cable is securely cemented to a screw or bony edge, the vacuum can be shut off and the tip carefully retracted. It is possible that tension in the cable will cause the FMA to lift from the brain as the tip is retracted. If so, it may be necessary to begin sealing the craniotomy with the vacuum tip in place. We recommend a dural sealant followed by a fast-curing silicone. Once the array is situated and the craniotomy sealed, the connector can be mounted onto the cranium using the surgeon s preferred method (such as skull screws and bone or dental cement). Connector Maintenance The connector of a chronically-implanted array is susceptible to clogging or damage due to animal manipulation, cage debris, and biological fluids. It is recommended that the connector be routinely cleaned using alcohol and air-dried using canned air, which will remove most contaminants and prevent electrical shunting. Microprobes for Life Science offers the option of a titanium pedestal with integrated sealing cap, but care must be taken as it may be too large to fit properly in rodent, depending on the animal and craniotomy location. Other options are to seal the connector with a rubber cap or male connector, though care must be taken to properly secure these options to prevent unintended removal. 9

Impedance Testing It is recommended that the impedance values of implanted electrodes be tested routinely, which can diagnose many problems. It is expected that electrode impedances will increase rapidly during the initial two weeks of implantation due to inflammatory response. However, a very large and sudden increase in impedance (typically above 10 MOhm) can indicate a broken electrode or connector, while a sudden decrease in impedance can indicate electrical shunting due to connector contamination. Microprobes for Life Science offers a Multi-Channel Impedance Tester (IMP-2MC) which can quickly and easily measure the 1 khz impedance of up to 18 channels. Alternatively, a potentiostat equipped with a multiplexer can be used. Terms and Conditions Please inspect the package carefully upon arrival and report any damage to us within 7 days of receipt of the package. Unused items may be exchanged if items and packaging are undamaged and in good condition. Exchange must be made within 30 days of invoice date and with prior permission from our Customer Service Department. Please call 301-330-9788 or email support@microprobes.com to request a Return Material Authorization (RMA) number. We do not accepted returns after 90 days from invoice date. Custom design products are non-returnable. 10

Contact Information M icroprobes for Life Science 18247-D Flower Hill Way Gaithersburg, MD 20879 Phone: +1 (301) - 330-9788 Fax: +1 (301) - 330-9667 support@microprobes.com www.microprobes.com Microprobes for Life Science Inc. 2015 All rights reserved 11