CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF MASS METROLOGY

Similar documents
CRITERIA FOR LABORATORY ACCREDITATION IN MASS METROLOGY

KENYA ACCREDITATION SERVICE

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF PRESSURE METROLOGY

General Accreditation Guidance. User checks and maintenance of laboratory balances

Beamex. Calibration White Paper. Weighing scale calibration - How to calibrate weighing instruments

Buoyancy Does it Matter?

technical guide Laboratory Balances Calibration Requirements

Specific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities

INFLUENCE OF MEASURING PROCESS AUTOMATION ON UNCERTAINTY OF MASS STANDARD AND WEIGHTS CALIBRATION.

Organisation Internationale de Métrologie Légale

TG GUIDELINES CONCERNING CALIBRATION INTERVALS AND RECALIBRATION

Supersedes: The copy of this document located on Measurement Canada s website is considered to be the controlled copy.

Any use of this material except in accordance with a written agreement with Western Power is prohibited.

OXFORD CALIBRATION WEIGHTS CLASS E1, E2, F1

Traceable calibration of automatic weighing instruments in dynamic operation

INTERNATIONAL OIML R 32 RECOMMENDATION. Rotary piston gas meters and turbine gas meters ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

Guide No. 2/2009. Design, training and equipment Martin Firlus. Physikalisch Technische Bundesanstalt Braunschweig und Berlin

Requirements and Methods for Calibration of Equipment in Clinical Laboratories. NK Chan (HKRCBTS) 24 August 2007

Sartorius CCL kg Prototype Mass Comparator

Gas mixtures. Individual solutions specifically for your application

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

This test shall be carried out on all vehicles equipped with open type traction batteries.

Compiled by: B Beard. Approved by: SH Carstens. Description of requirements and procedures for compact provers to be used as verification standards.

How to Achieve Optimal Weighing Performance

Vaisala Pressure, RH & Temp. Transmitter. The instrument was operational upon receipt. The instrument was adjusted and calibrated.

Determination of Air Density with Buoyancy Artefacts

THE USE OF Z FACTORS IN GRAVIMETRIC VOLUME MEASUREMENTS

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Calibrating a Micropipette

Certificate of Accreditation

CERTIFICATE OF ACCREDITATION ISO/IEC 17025:2005 ANSI/NCSL Z (R2002)

Organisation Internationale de Métrologie Légale. Direct mass flow measuring systems for quantities of liquids Annex C: Test report format

Date of Shipment: Xxxxx 00, 20xx 3402c National Institute of Advanced Industrial Science and Technology National Metrology Institute of

Supplementary Certificate of Approval. No S433

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

CORESTA RECOMMENDED METHOD N 6

FPG8601 Force Balanced Piston Gauge

Deadweight Tester / Pressure Balance High pressure up to 5,000 bar / 70,000 psi CPB5000-HP. Applications. Special features. Description. Rel.

ISO Vacuum gauges Calibration by direct comparison with a reference gauge

Precision pressure sensor Basic version Model CPT6020

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

DESIGN AND PERFORMANCE OF THE NEW SARTORIUS 1 KG VACUUM MASS COMPARATOR AT PTB

Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience -

Equipment Accuracy / Selection

Vehicle-mounted meters, pump supplied

DENSITY CALIBRATIONS AT MIKES. Heikki Kajastie, Jorma Manninen and Kari Riski Centre for Metrology and Accreditation (MIKES) ABSTRACT

General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT

II. MATERIALS REQUIRED FOR PERFORMANCE TESTING

Gas flow calibrations performed at the National Metrology Institute of South Africa (NMISA)

Laboratory Manual Non Automatic Weighing Devices Page: 1 of 15

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010

ISO INTERNATIONAL STANDARD. Natural gas Measurement of properties Volumetric properties: density, pressure, temperature and compression factor

No. 2 TEST WEIGHT ASSESSMENT

INTERNATIONAL STANDARD

Requirements for the certification of movements and mechanical watches resistant to magnetic fields of 1.5 T ( G)

AIR PRESSURE CONTROLLED MASS MEASUREMENT SYSTEM

H&D Fitzgerald Pressure Pyknometry System Mark 5

ENVIRONMENTAL AND POLLUTANTS GAS ANALYZERS

Accreditation for In-Service Inspection of Transportable Pressure Receptacles (TPRs)

JOS 15 JOS 25 JOS 40. Manual and semiautomatic test benches from DN 25 to DN 40. Characteristics. Basic characteristics. Test bench using balances

Volume Determination for 2 kg~20 kg in NIM

UKAS calibration laboratory 8809 Temperature, pressure and electrical signals calibration guide

GMC-I Service GmbH Calibration Center

Light Lab. Reliable Measurements through Regular Calibration. DAkkS

Calibration Procedure. PosiTest DFT Combo

THE PRESSURE SIGNAL CALIBRATION TECHNOLOGY OF THE COMPREHENSIVE TEST

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z TROXLER ELECTRONICS LABORATORIES, INC. James Pratt Phone:

Experiment #2. Density and Measurements

Calculating Total Uncertainty of Temperature Calibration with a Dry Block. Calibration White Paper. Beamex.

Hydraulic pressure balance Model CPB5800

Road and rail tankers with level gauging Part 2 Metrological controls and tests. Page 1 of 42 OIML TC8/SC1/WG3/R 80-2 CD1 Draft of July

Vehicle- or rack-mounted liquefied gas meters, pump supplied

1. Scope: This document describes the procedure for verifying the performance of vaccine vial monitors.

Recommended Practices for Calibration Laboratories

PRESSURE TRANSDUCERS WITH UNIFIED (4-20) ma OR (10-50) ma OUTPUT SIGNAL

Chemical Industry. School experiments. Weighing. Helper Booklet. Tips and Tricks for Weighing. Natural science laws experience live learn easily

A Code of Practice for the Calibration of Industrial Process Weighing Systems

Dead-weight tester High-pressure version Model CPB3800HP

INSPECTION FOCUS ON:

Additional Information

Primary standard differential pressure balance Model CPB6000DP

Measurements. Metric System

THE NEW SARTORIUS 1KG-PROTOTYPE BALANCE FOR HIGH PRECISION MASS DETERMINATION

Density Determination Kit Instruction Manual

CALIBRATION SERVICES HUMIDITY DEW POINT AIR VELOCITY MASS FLOW TEMPERATURE CO 2 PRESSURE. (+34)

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. PRECISION MEASURING CORP Groesbeck Hwy. Fraser, MI Robert Meyer Phone:

Technical Data Sheet MF010-O-LC

Calibrating Piston Pipettes

CHM Introductory Laboratory Experiment (r17sd) 1/13

Operating Instructions METTLER TOLEDO Pipette Check Application for AX and MX/UMX Balances Version 1.xx

SmartCal. SmartCal Troubleshooting Checklist

Keeping It Simple. Gas Mixtures for Stack Emissions Monitoring.

High-speed pressure sensor Model CPT6140

LAB 14 EDITION 5 July 2015

Pneumatic dead-weight tester Model CPB3500

OIML R 140 RECOMMENDATION. Edition 2007 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Measuring systems for gaseous fuel

CORESTA Guide N 6. February A User Guideline for the use of balances for cigarettes and cigarette related products

Gravimetric preparation of NO 2 primary reference gas mixtures at the NMISA. Tshepiso Mphamo 03 September 2012

DX2/DXX Operating Instructions

Transcription:

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF MASS METROLOGY Approved By: Chief Executive Officer: Ron Josias Senior Manager: Mpho Phaloane Revised By: Specialist Technical Committee Members Date of Approval: 2014-11-28 Date of Implementation: 2014-11-28 SANAS Page 1 of 13

TR15-07 CONTENTS: 1. Purpose and Scope... 3 2. References, Definitions and Abbreviations... 3 2.1 References... 3 2.2 Definitions... 3 2.3 Abbreviations... 4 3. Environmental Requirements... 5 4. General Requirements... 5 5. Technical Requirements... 6 5.1 Weights used for the Calibration of Weighing Instruments... 6 5.2 Calibration of Weighing Instruments - Minimum Requirements... 6 5.4 Calibration of Weights - Requirements... 8 5.5 Calibration of Weights Uncertainty of Measurement...10 5.6 Calibration of Force (Newton) or Non-metric Weights...10 5.7 Calibration of a Fruit Penetrometer...10 Appendix 1: Schedule of Accreditation...11 Appendix 2: Source OIML R111-1 Edition 2004(E) Table B7...12 Addendum 1: Amendment Record...13 SANAS Page 2 of 13

1. Purpose and Scope The purpose of this document is to define the specific environmental, general and technical requirements to be met by accredited laboratories in the field of mass metrology. This document is applicable to South African National Accreditation System (SANAS) Accredited Laboratories in this field. This document does not address the requirements for accreditation as a mass verification laboratory. 2. References, Definitions and Abbreviations 2.1 References [1] OHS Act, 1993 Environmental Regulations 1987 - Lighting [2] OHS Act, 1993 General Admin Regulations 1994 Handling of Hazardous Chemical Substances [3] ISO 17025:2005 General requirements for the competence of testing and calibration laboratories. [4] NPL Good Practice Guide Cleaning, Handling and Storage of Weights. [5] AS TG2 IANZ Technical Guide: Laboratory Balances Calibration Requirements International Accreditation New Zealand. [6] Mettler-Toledo Fundamentals of Mass Determination. [7] SANS 1697:2007 South African National Standard Verification standards for the verification of mass-measuring instruments, including commercial standard masspieces. [8] OIML R 111-1 OIML International Recommendation R 111-1 : 2004 (E) Weights of Classes E1, E2, F1, F2, M1, M2, M3 [9] NPL Guide 71 Good Practice Guide Nº 71: The Measurement of Mass and Weight [10] OECD Guidance on objective tests for determining the ripeness of fruit AGR/CA/FVS(1993)11/REV6 [11] PTB-MA-24e Determination of Mass Part 1: Dissemination of the unit of mass Balances, Weights and Test Rooms. (ISBN 3-89429-756-5) 1996 2.2 Definitions 2.2.1 Apparent Mass The value indicated by the balance i.e. the result of a measurement before correction for systematic error (e.g. for buoyancy correction). [11] 2.2.2 Buoyancy Correction A buoyancy correction, is the correction applied when weights of differing densities are compared with each other during the calibration process, the buoyancy being a result of the upward force when the weight is immersed in a fluid (such as air) during the weighing process. 2.2.3 Calibration Authority Body responsible for the national measurement standards (e.g. NMISA) of a country, or a laboratory accredited by an accreditation body that has been appointed as a national accreditation body (e.g. SANAS), or a competent authority acceptable to the national responsible body for legal metrology. 2.2.4 Conventional Mass Conventional value of the result of weighing in Air. For a weight taken at a reference temperature ( t ref ) 20º C, the conventional mass is the mass of a reference weight of a density (ρ ref ) of 8 000 kg/m 3 which it balances in air of a reference density (ρ 0 ) of 1,2 kg/m 3.[8] SANAS Page 3 of 13

TR15-07 2.2.5 Discrimination The smallest change in mass that can be detected by the weighing instrument. For practical purposes discrimination is synonymous with readability. [5]. Discrimination is also synonymous with the sensitivity of a non-self indicating weighing instrument. 2.2.6 Readability (d) The smallest scale division or digital interval of the weighing instrument. For some mechanical weighing instruments the scale marks may be sufficiently far apart for an estimation to be made of the actual weighing instrument reading when the pointer lies between two scale marks. The estimated readability may therefore be lower than the marked readability. [5]. 2.2.7 Resolution The readability expressed as a portion of the capacity. For example a weighing instrument with a capacity of 3 000g and a readability of 0,1g has a resolution of 1 part in 30 000. [5]. 2.2.8 Scale Interval The value expressed in units of mass: a) In the case of analogue indication, the difference between the values that correspond to two consecutive scale marks, or b) In the case of digital indication, the difference between two consecutively indicated values. [7] 2.2.9 True Mass The true mass of a body relates to the amount of material it contains. The prefix true is added to the word mass where it is important to make it clear that a particular value of mass being considered is not a conventional mass value and it is important to avoid potential ambiguity. The international prototype kilogram, on which the international mass scale is realized, is defined as a true mass of exactly 1 kilogram. Most High accuracy comparisons are performed on a true mass basis. [9] 2.2.10 Verification Standard Masspiece A material measure of mass that has been calibrated before by a calibration authority and that is used to verify massmeters or, in conjunction with a verification standard massmeter, to verify commercial standard masspieces. [7] 2.2.11 Weight A material measure of mass, regulated in regard to its physical and metrological characteristics: shape, dimensions, material, surface quality, nominal value, and maximum permissible error. [8] 2.2.12 Fruit Penetrometer The determination of the firmness of fruit using a penetrometer is based on the force necessary to push a plunger of specified size into the pulp of the fruit up to a specified depth. [10] 2.3 Abbreviations d - Readability CMC - Calibration and Measurement Capability mpe - Maximum permissible error NPL - National Physical Laboratory, United Kingdom OECD - Organisation for Economic Co-operation and Development OHS Act - Occupational, Health and Safety Act, Act 85 of 1993 OIML - International Organization of Legal Metrology RH - Relative Humidity SANS - South African National Standard UoM - Uncertainty of Measurement SANAS Page 4 of 13

3. Environmental Requirements 3.1 The temperature and humidity in any laboratory where weights are calibrated must be maintained between the limits as specified in table 1. Table 1 Class of Weight Temperature Range (ºC) T during calibration Relative Humidity Range (rh) (%) rh during calibration Min Max ºC / Hr ºC / 12 Min Max (% / 4 Hr) Hr E1 18 25 ± 0,3 ± 0,5 40 60 ± 5 E2 18 25 ± 0,7 ± 1,0 30 70 ± 10 F1 18 27 ± 1,5 ± 2,0 20 80 ± 15 F2 18 27 ± 2,5 ± 3,5 20 80 ± 15 M1 ӿ ӿ ± 3,0 ± 5,0 ӿ ӿ ӿ M2 ӿ ӿ ӿ ӿ ӿ ӿ ӿ M3 ӿ ӿ ӿ ӿ ӿ ӿ ӿ ӿ No specific requirement A laboratory that has a CMC of ⅓ and less of the mpe for the specific weight class (E1 to M3 as per OIML R111) shall comply with the applicable environmental requirements as specified in Table 1 above. 3.2 Where applicable the laboratory shall maintain appropriate records to demonstrate and confirm the humidity, temperature and temperature gradients within the laboratory. 3.3 Weighing instruments must be calibrated at ambient temperatures. 3.4 Lighting within the laboratory shall be adequate to facilitate the correct performance of the calibration work undertaken. Cognisance shall be taken of the minimum levels of lighting as specified in the environmental regulations of the OHS Act. [1] [3] 3.5 Drafts in the laboratory caused by doors, fans and/or air conditioners should be such that they do not have an adverse effect on the measurement results, where necessary local isolation in the form of draft shield should be utilized. 3.6 Thermometers and/or temperature, relative humidity, and barometric pressure recorders or indicators used for the measurement and recording of the ambient temperature, humidity and/or barometric pressure in the laboratory, and onsite shall be calibrated. 4. General Requirements 4.1 Laboratories shall have a policy and procedure/s that addresses how weights, both their own standards, and weights belonging to customers are to be handled, cleaned and refurbished. [4]. 4.2 All cotton and chamois leather gloves and cloths or other equipment used for the handling of weights shall be kept clean. [4]. 4.3 Raw data shall not be recorded in pencil, or erasable ink. 4.4 The schedule of accreditation shall list the types of weighing instruments that the laboratory is competent to calibrate under the parameter weighing instruments. This list may include the following: digital self indicating; mechanical self indicating; SANAS Page 5 of 13

TR15-07 semi-self indicating; non-self indicating; equal arm balances; automatic scales; hopper scales; conveyor scales; comparators; gram gauges; fruit penetrometers; in-motion weighers; crane scales and weighbridges. Due to the complexity of the calibration of certain types of scale for example equal arm balances, crane scales and weighbridges, these types may be listed separately on the schedule of accreditation with dedicated MC s. 5. Technical Requirements 5.1 Weights used for the Calibration of Weighing Instruments 5.1.1 The weights used to perform the calibration of weighing instruments shall be calibrated to an appropriate level of uncertainty. 5.1.2 Make up weight may not be used for the calibration of weighing instruments having a capacity of 500 kg or less. 5.1.3 For weighing instruments having a capacity of larger than 1 000 kg the mass of the standard weights used shall be at least 500 kg and the make up weight may not exceed 50% of the capacity of the weighing instrument. 5.1.4 For weighing instruments having a capacity of larger than 10 000 kg the mass of the standard weights used shall be at least 5 000 kg and the make up weight may not exceed 80% of the capacity of the weighing instrument. 5.1.5 All make up weight must be of a stable material. 5.1.6 In certain instances it may not be possible to use weights having an uncertainty equal or smaller than the readability of the weighing instrument due to local and international limitations on the uncertainty of weight calibration. For example the maximum uncertainty for a 5 gram Class E1 weight is ± 5 μg ( 1 / 3 of 16μg); whilst micro balances with a capacity of 5 gram and readability of 0,1 μg are available. [8] 5.2 Calibration of Weighing Instruments - Minimum Requirements The following are the minimum tests which must be carried out when calibrating a weighing instrument unless the user or client has requested or agreed to a partial calibration, in which case both the weighing instrument and the calibration certificate/report must be marked accordingly. Laboratories may also carry out any number of additional measurements using various test loads. 5.2.1 Repeatability (or Precision) Measurement a) Take at least five readings using (where practical) a single standard weight (or other suitable artefact) as close as practical to half the capacity of the weighing instrument. b) Where the capacity of the weighing instrument exceeds 30 kg, the number of readings may be reduced to a minimum of three. SANAS Page 6 of 13

c) Report at least the nominal load applied and the standard deviation of the repeated weighings. 5.2.2 Eccentric (or Corner) Measurement a) Place standard weights (or a suitable artefact) of between one quarter and one third of the weighing instrument s capacity alternately in the centre and in the various eccentric positions on the load receptor to determine the deviation (if any) caused by the off-centre loading. Report at least the nominal load applied and the maximum deviation from the centre, caused by the off-centre loading. b) Where a manufacturer of a weighing instrument specifies a particular load for this test, that load may be used. c) The eccentric test is not carried out in the case of a weighing instrument with a suspended load receptor. d) Where the load receptor is supported on more than four points, a load as close as practicable to the capacity of the instrument divided by the number of supporting points minus one, is used. e) Where a load receptor has a single point of support, the load must be placed in any position on the load receptor, at a distance of approximately two thirds of the distance from the centre to the edge of the load receptor. In the case of a square or rectangular load receptor, the receptor is divided into quarters and the load placed in the centre of the quarter. f) Where a load receptor has four or more supporting points, the surface area is divided by the number of supporting points and the load is placed in the centre of (or evenly spread over) each segment in turn. g) Where a load receptor is in the form of rails or tracks (e.g. rail vehicle scales or overhead track scales) the load must be applied at the beginning, in the middle and at the end of the rail or track 5.2.3 Accuracy and Linearity Test the accuracy and linearity using at least five different test loads (excluding zero), approximately evenly spread over the weighing range of the instrument. Additional measurements, such as treating each range as a separate instrument, must be carried out in the case of: a) Multi-range or multi interval weighing instruments; b) Weighing instruments with multi revolution indicators; c) Weighing instruments with optically projected indication together with built in unit weights to increase the capacity; In the case of non-self-indicating weighing instruments, the accuracy of any travelling or sliding poises and proportional counterpoise weights must be tested. Record also the actual load applied to reach equilibrium of the beam or steelyard. Report both the actual values of the loads applied and the resultant indication. Where the total of the actual errors or inaccuracies of the standard weights used is less than one quarter of the value of the smallest graduation or readability of the weighing instrument, the nominal values of the loads may be stated instead of the actual values. SANAS Page 7 of 13

TR15-07 5.3 Weighing Instruments Uncertainty of Measurement. 5.3.1 Manufacturers of weighing instruments including electronic, electro-mechanical and mechanical balances have traditionally specified repeatability, eccentricity and linearity separately. In order for the calibration laboratory to claim compliance with specification they need provide objective evidence based on measurements that the sum of the measurement error and uncertainty does not exceed the stated accuracy. 5.3.2 The need to claim compliance is further complicated as specifications are not necessarily stated as outside limits, for example the repeatability is often specified at ±1 σ, and balance specifications are established under ideal conditions, whilst the balance submitted for calibration is located in a working environment, in which the operating conditions may be within specification but not necessarily ideal. 5.3.3 Laboratories may adopt a simplified approach in practice in the estimation of the measurement uncertainty for the linearity measurement of weighing equipment having fewer than 30 000 divisions, whereby the uncertainty is estimated by taking into account the uncertainty of the weights, the eccentricity (corner load), repeatability, and/or the resolution of the weighing instrument (whichever is larger). 5.3.4 The standard uncertainty due to the eccentricity(u e ) is estimated as u e = [d 1 /d 2 * D] * 1/ 3 (adapted from R111 / R 76) Where: d 1 = the distance from the centre of the load receptor to the centre of the weights; d 2 = the distance from the centre of the load receptor and one of the corners; D = the deviation between the maximum and minimum measured values; The ratio of d 1 /d 2 will typically be either 2 / 3 or 2 / 4 (see 5.2.2) 5.3.5 Laboratories shall supply (and have available) for evaluation all relevant data of the weighing instrument, such as specification sheets, used in the determination of the calibration and measurement capability (CMC), along with copies of the Uncertainty of Measurement estimation. All relevant factors shall be considered, and only after confirmation that the contribution is insignificant may they be omitted from the estimation. This process shall be documented. 5.3.6 Laboratories are required to consider as a minimum the following contributors when establishing the uncertainty of measurement of weighing instruments having in excess of 2 000 000 divisions (typically analytical, semi and micro balances): The uncertainty of the weights; The stability of the weights; Buoyancy of the weights; The repeatability of the weighing instrument; The resolution of the weighing instrument; Temperature coefficient of sensitivity; The eccentricity (corner load). Unless records are available the temperature drift assumed shall not be less than 0,5ºC. Where the repeatability of the balance is found to be zero, the repeatability stated by the manufacturer shall be assumed. 5.4 Calibration of Weights - Requirements 5.4.1 Weighing in the laboratory must take place on a weighing bench of adequate construction, which will facilitate the stable and correct operation of the weighing SANAS Page 8 of 13

instrument. The weighing bench design shall take cognisance of vibration, stability, magnetic susceptibility, static charging, and support. 5.4.2 Laboratory procedures shall address stabilization times for the calibration of weights. 5.4.3 Where an uncertainty of equal to or less than equal to ± 0,000 02% (± 2.10-7 M) is quoted, the density of the weights must either be known (from a prior determination), or determined and a buoyancy correction must be applied.(typical OIML Class E1) 5.4.4 Where an uncertainty of between ± 0,000 50 % (± 5.10-6 M) and ± 0,000 02% (± 2.10-7 M) is quoted, the density of the weights may be measured or assumed and a buoyancy correction applied. Refer to Appendix 2 for a table of expected material densities and uncertainties. (Typical OIML Classes E2 to F2) 5.4.5 No buoyancy correction need be applied when weights are calibrated against standards of similar material, (i.e. Stainless Steel and Stainless Steel), and uncertainties of between ± 0,000 2% (2.10-6 M) and ± 0,005 % (5.10-5 M) are claimed. (Typical OIML Classes M1 to M2) 5.4.6 No buoyancy correction need be applied when uncertainties of 0,005 % and larger (5.10-5 M) are claimed. (Typical OIML Class M3) 5.4.7 When cast iron weights are calibrated, laboratories may not quote uncertainties of less than 0,001 6 % (± 1,6 10-5.M). 5.4.8 The calibration laboratory shall where applicable indicate on the calibration certificate the prevailing measurement conditions such as air density, weight density, temperature, and indicate whether these values are assumed or have been measured. 5.4.9 A SANAS accredited mass laboratory may calibrate verification standard masspieces for use by an accredited verification laboratory. The metrological characteristics of these masspieces should conform to the maximum permissible tolerance requirements as specified in Tables 1 or 2 of SANS 1697:2007. [7] In order to claim metrological compliance the sum of the error and the uncertainty of measurement may not exceed the tolerances as specified in these tables. The SANS 1697:2007 standard includes both physical (material of manufacture, shape, adjustment, surface finish, nominal values, identification, calibration mark) and metrological (maximum tolerance) specification. [7] Any statement of compliance on the calibration certificate shall indicate which clauses or sub-clause of the standard have been complied with. 5.4.10 Laboratories wishing to certify weights according to the accuracy classes as per OIML R111 shall ensure that the UOM does not exceed one third of the tolerance specification. As R111 contains both metrological and physical requirements, general class compliance cannot be claimed unless both, physical and metrological requirements have been satisfied. Where a statement of compliance is made to an accuracy class and the physical requirements have not been evaluated this shall be clearly indicated on the calibration certificate. 5.4.11 When reporting the results the laboratory may not report on the calibration certificate more significant digits than it is able to measure on the weighing instrument, or more significant digits than have been specified on the certificate of calibration of their standard masspiece used during the calibration. 5.4.12 No laboratory shall be accredited to perform on site calibration of weights of denomination 50 kg and above unless they also have the capability to perform the calibration in their own laboratory, and their measurement capability is based on the equipment which they have in their own laboratory. This equipment shall include but is not limited to standard masspieces (weights), balances and/or mass comparators, except where the laboratory SANAS Page 9 of 13

TR15-07 has entered into a written agreement with a customer, confirming that the laboratory has been granted access to the necessary equipment. This agreement could be in the form of a letter granting the necessary authorization. In such cases the equipment shall be treated as equipment outside the direct control of the laboratory, in accordance with ISO/IEC 17025 clauses 5.5.1 & 5.5.9. 5.4.13 Laboratories accredited to perform the calibration of weights the nominal mass of which exceeds 100 kg shall have access to suitable ancillary equipment as may be necessary to load and move these weights onto and from the weighing instrument. Such ancillary equipment may include slings, crawl beam, block and tackle, and a fork lift, as appropriate. 5.4.14 Laboratories who wish to be accredited for the calibration of weights the nominal mass of which exceed 500 kg, and where circumstances prevent the removal of these weights from their place of use, shall provide evidence of the calibration of the weighing instrument/s used to perform the calibration of the weight. Such calibration may be limited to a particular site. 5.5 Calibration of Weights Uncertainty of Measurement 5.5.1 The estimation of the uncertainty of measurement for weights calibrated by the laboratory shall consider at least the following contributors: the calibration of the standard masspiece or weight; the air buoyancy; the uncorrected drift of the standard masspiece; the resolution of the balance; the rounding off of the measurement result; the eccentricity (corner load) and the repeatability. 5.5.2 The uncertainty of measurement shall not be reported to more than 2 significant digits. The uncertainty of measurement shall always be rounded up, unless the rounding reduces the uncertainty by less than 5%. 5.6 Calibration of Force (Newton) or Non-metric Weights 5.6.1 The laboratory may calibrate weights in non-metric or weights marked in a nominal force in Newton s, the results and the associated uncertainty shall be reported in metric units of mass (kg, gram, mg, etc.) together with the conversion factor, and gravity (standard or local) used optionally along with the calculated equivalent value in the non-metric unit. The certificate shall include a statement indicating the source (reference) of the conversion factor used. 5.7 Calibration of a Fruit Penetrometer 5.7.1 The calibration of a fruit penetrometer is achieved by applying a stable known force/s (in kgf) against the probe of the penetrometer. 5.7.2 The laboratory shall report in addition to a table of measurement results the nominal diameter of the penetrometer. 6. Authorship The revision of this Technical Requirement document has been prepared by members of the Mass Metrology Specialist Technical Committee. SANAS appreciates the contribution made by the committee members in the preparation of this document. SANAS Page 10 of 13

Appendix 1: Schedule of Accreditation S C H E D U L E O F A C C R E D I T A T I O N MASS METROLOGY Laboratory Accreditation Number : 14XX Permanent Address of Laboratory: The Weighing Shop (Pty) Ltd 63 West Street Over the Hill Johannesburg Postal Address: P O BOX 32222 Johannesburg 2000 Tel: (011) 499-8778 Fax: (011) 499-8879 Technical Signatory Nominated Representative Issue No. Date of Issue Expiry Date : Mr C Heavy : Ms C Light : 03 : 20 August 2013 : 31 July 2019 ITEM MEASURED QUANTITY OR TYPE OF GAUGE OR INSTRUMENT RANGE OF MEASURED QUANTITY MEASUREMENT CAPABILITIES EXPRESSED AS AN UNCERTAINTY (±) OIML R111 Weight Class 1 Mass pieces (Weights) 1 mg to 10 g 10 g to 200 g 200 g to 20 kg 20 kg to 500 kg 0,005 mg 0,0001% 0,002 % 0,003 % F1 F1 M2 M2 2 Weighing Instruments (Type specify) 0 g to 6 g 6 g to 200g 200 g to 2 kg 2 kg to 5 000 kg 0,0005 % + 3 µg 0,0015 % + 0,1 mg 0,003 % 0,01 % 3 Fruit Penetrometer 0 to 20 kg.f 2% + 1 kg.f 4 On-site calibration of weighing instruments Original date of accreditation:1994 Page 1 of 1 The CMC, expressed as an expanded uncertainty of measurement, is stated as the standard uncertainty of measurement multiplied by a coverage factor k = 2, corresponding to a confidence level of approximately 95% ISSUED BY THE SOUTH AFRICAN NATIONAL ACCREDITATION SYSTEM SANAS Page 11 of 13

TR15-07 Appendix 2: Source OIML R111-1 Edition 2004(E) Table B7 Alloy/Material Assumed Density Uncertainty (k=2) Platinum 21 400 kg/m 3 ± 150 kg/m 3 Nickel silver 8 600 kg/m 3 ± 170 kg/m 3 Brass 8 400 kg/m 3 ± 170 kg/m 3 Stainless Steel 7 950 kg/m 3 ± 140 kg/m 3 Carbon Steel 7 700 kg/m 3 ± 200 kg/m 3 Iron 7 800 kg/m 3 ± 200 kg/m 3 Cast Iron (white) 7 700 kg/m 3 ± 400 kg/m 3 Cast Iron (grey) 7 100 kg/m 3 ± 600 kg/m 3 Aluminium 2 700 kg/m 3 ± 130 kg/m 3 SANAS Page 12 of 13

Addendum 1: Amendment Record Proposed By: Section Change FM 2.1 References updated FM 2.2.3 Definition for Calibration Authority updated FM 2.3 Abbreviations updated FM 5.3.4 Clause added to include eccentricity uncertainty determination FM 5.3.3 & Added references to eccentricity. 5.3.6 FM Appendix 1 Updated to reflect current requirements & added OIML class for weights STC 3.1 Table introduced to simplify environmental requirements STC 5.4.12 Requirement modified STC 5.1.2 & 1 000 kg Make up weight reduced to 500 kg 5.1.3 SANAS Page 13 of 13