Observations Possible Causes Recommended Actions. Turquoise color

Similar documents
Armfield Distillation Column Operation Guidelines

Using the Akta Prime plus October 22, 2012

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax

Read this first. Zetasizer nano series Self installation and Quick start guide MRK825-02

Navigator 600 Silica analyzers

User s Booklet for the Wyatt minidawn Light Scattering Instrumentation

Diver-Office. Getting Started Guide. 2007, Schlumberger Water Services

AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by:

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

BD Cytopeia Fluidic Kit User s Guide

AKTA MC SOP Page 1 9/27/04 AKTA METAL CHELATING SOP

SOP: Anion Exchange Chromatography of Green Fluorescent Protein (GFP) using the AKTA Pure system

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

AKTA pure 25 New Owner s Intro

AKTA 3D SOP. Click on the System Control tab. This screen has 4 windows.

Standard Operational Procedure for Capillary Electrophoresis

Quick Reference Guide

Operator Quick Guide ORBISPHERE 3654

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120

TapeStation Tips & Tricks

Microfluidic Demonstration

6800 Maintenance Instruction System Flush Procedure

SOP: Cation Exchange Chromatography of t-pa using AKTA Pure system

Installation of Your SprayMaster System

VROC (Viscometer/ Rheometer-on-a-Chip) User s Manual 1

PerkinElmer Aanalyst100

QUICK START GUIDE TO DRINKING WATER MONITORING

Mass Spec will not Autotune

POLYMER TEST KIT (CLAPPER METHOD) OFI PART No

Mix 920 ml of 87% glycerol (on shelf) with 1080 ml of Isoton II (in cabinet), stir in a flask until homogenized (the glycerol will sink).

v /07/17 Page 1 of 18

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone:

SOP for use of the SX19 Stopped-Flow Instrument

5890II GC Standard Operating Procedure 9/2/2005

FROG-4000 TM Chemical Analysis System

IVIS Spectrum Whole Animal Imager

Agilent Auxiliary Gas Module AGM 2. User s Guide

SS32L Dissolved O 2 Probe

DXH Troubleshooting FlowCharts

Quick Start Guide. A. Hardware installation B. Software installation C. Start the software for the first time D. Do your first measurement

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

CONSUMER MODEL INSTALLATION GUIDE

HemoCue Hb Procedure Template

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA

6900 Maintenance Instruction System Flush

Technical Procedure for General Laboratory Equipment

Microsoft Windows Software Manual for FITstep Stream Version 4

XC2 Client/Server Installation & Configuration

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

Evaluation copy. Interdependence of Plants and Animals. computer OBJECTIVES MATERIALS

Making OTR Measurements in Bottles using the. BottlePerm Fixture

SOP: Buck Scientific BLC-20P HPLC Operation

The Oxygen Sensor can be connected to all types of einstein Tablets, einstein LabMate, and einstein LabMate +.

Cover Page for Lab Report Group Portion. Head Losses in Pipes

Lipodrop 2.0 device coating

Upgrading Bio-Plex Manager 4.1, 5.0, or 6.0 Software to Bio-Plex Manager 6.1 Software

SOP: Lysine Affinity Chromatography of t-pa on the ÄKTApure Chromatography System

Department of Laboratories St. Louis, MO ISSUE DATE: June 2007 REVISION DATE: July 2010 REVIEWED DATE: May 2017 PRINCIPILE:

no peaks missing peaks too small peaks increasing retention decreasing retention

COULTER LH 500 HEMATOLOGY ANALYZER

Inventory User Guide

Oxygen Sensor with Electrode

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

TITLE: DNA minitube - Red

Diver-Pocket Diver-Pocket Premium

INSTALLATION, CARE & USE MANUAL. EMASMB & LMASMB Surface Mount Bottle Filling Stations

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300

Trion PECVD SOP IMPORTANT: NO PLASTIC, TAPE, RESISTS, OR THERMAL PASTE ARE ALLOWED IN THE CHAMBER

#LZ400 LEAKALYZER. Water Loss Sensor INSTRUCTION MANUAL Country Dr. #190 St. Paul, MN

MASSACHUSETTS STATE POLICE CRIME LABORATORY. Certificate of Calibration Procedure for the Alcotest 9510

For Biogeochemistry at ICEX 2011

Technology. In the My Files [My Files] submenu you can store all the programs that you have made on the NXT or downloaded from your computer.

ChemScan. mini UV254 Analyzer. Installation, Operation and Maintenance Manual. ASA, Inc.

Rhino VE CE Bulk Unloader Operator s Card

UsER manual for Watersens ph -REDOX

Steltronic StelPad User Guide

For running only the scoresheet application without any video features only some very basic hardware / software requirements have to be fulfilled:

Arizona State University Center for Solid State Electronics Research Issue: E Title: Heat Pulse 610 Operating Procedure Page 1 of 7

Instruction Manual. BZ7002 Calibration Software BE

STS ICP-RIE. Scott Munro (2-4826,

operator s manual plus system system nanolc Operator s Manual Revision A1

Hemoglobin Reference Failure Troubleshooting

Figure 1: Plug in ion inlet of mass spec

AQUARIUS. Operating Instructions for the. Type 70 Electrolytic Gas Soldering / Welding Unit. Table of Contents

Requesting Team Activation

FROG-5000TM Chemical Analysis System User s Manual

NMR Service. How to Prepare Samples for NMR

Using the UltraRAE. Firmware 2.35

Manual for continuous distillation

Determination of Zn using Atomic Absorption with Multiple Standard Additions

Agilent ICP-MS Interactive Troubleshooting tool for Low sensitivity

Fiber Optic SMA Z-Flow Cell Manual Version 2.1

Sterisil Ac+ Installation & Operating Manual

PERFORM Operating Document

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

Running the LiCor 6400 Infra Red Gas Analyzer (IRGA) with 4 whole-plant cuvettes attached

Race Screen: Figure 2: Race Screen. Figure 3: Race Screen with Top Bulb Lock

Transcription:

ABI PRISM 3700 Instrument-Hardware-Array View Troubleshooting Module Please contact technical support if you have any questions regarding the document. Text in blue indicates a link to another section or another document. Observations Possible Causes Recommended Actions Turquoise color Go to larger picture Yellow Color Go to larger picture Presence of bubbles in the cuvette. The source of the bubbles will need to be identified, refer to the Checking for Bubbles module. Leak in sheath flow syringe Impaired movement of the syringe pump. Leak at the inline filter Presence of bubbles in the sheath flow syringe. Sheath flow syringe or pump may not be functioning Refer to the Checking the CCD calibration window Module to determine if bubbles have been removed from the cuvette. Verify that within the polymer bottle, the tubing is immersed in polymer. Check for leaking (dried polymer) at the in-line filter or syringes and if necessary replace components. Verify that the fittings on the port above the sheath flow syringe are finger tight. Run the cuvette flush module, this may require several flushes until bubbles are removed. Replace the sheath flow syringe using the change syringe wizard. Check the syringe pump, refer to the Verifying Syringe Pumps module. Clean dried polymer from the area and replace the inline filter. Verify that the fittings on the port above the sheath flow syringe are finger tight. Verify that the tubing in the polymer bottle is immersed in polymer. Verify the syringe pump movement with and without the sheath flow syringe, refer to the Verifying the Syringe Pumps module. Page 1 of 19

Observations Possible Causes Recommended Actions Striations Go to larger picture Green color Polymer mixing of polymer from two bottles. polymer was not at room temp. precipitate in bottle. Presence of foreign fluids in the 1X running buffer or DI water reservoirs. Presence of bubbles Use a new bottle of polymer, we do not recommend the mixing of polymer from different bottles. Bring polymer to room temperature before placing on instrument and gently mix. Verify that the precipitates have gone into solution. When changing polymer, use the change polymer wizard to flush the old polymer out of the system. Thoroughly rinse the running buffer and DI water reservoirs with DI water and replenish with a fresh stock. Run sequencing standards to verify the system. Refer to the Checking the CCD calibration window Module to determine if bubbles have been removed from the cuvette. Verify that within the polymer bottle, the tubing is immersed in polymer. Go to larger picture Contamination in reagents: water, buffer or polymer Check for leaking (dried polymer) at the in-line filter or syringes and if necessary replace components. Verify that the fittings on the port above the sheath flow syringe are finger tight. Run the cuvette flush module, this may require several flushes until bubbles are removed. Verify the reagents on the instrument. Check expiry dates and length of time on the instrument. Replace reagents if necessary. Rinse the buffer and water reservoirs with DI-water and replenish with a fresh stock. Page 2 of 19

Observations Possible Causes Recommended Actions Use another source of DI water. Check the filtration system for the DI water. Leak in sheath flow syringe Sample Contamination Replace the sheath flow syringe using the change syringe wizard. Run a HiDi Formamide only plate to eliminate sample issues. If color is not seen in the HiDi Formamide run please contact Technical Support to troubleshoot sample. Black color entire array is affected Go to larger picture Some lanes are not affected Loss in communication Optics Sample loading issue Bubbles in the sample transfer lines. Bubbles in the sample wells Clogged loading tips Electrophoresis Issues Sample loading issue bubbles in the sample transfer lines Restart the system Check the CCD window, for raman lines. Refer to the Checking CCD Window module. If raman lines are not present, contact service. Check the water reservoir. Replenish the water and verify that the fluid line sinkers for the tubing are at the bottom of the reservoir. Verify that the autoloader tips are screwed in until finger tight. Verify the fittings on the port above the syringes are finger tight. Determine the source of the bubbles. Refer to the Checking for Bubbles module. Replace the sample transfer syringe(s) using the change syringe wizard. Spin down samples before placing on the instrument. Replace the loading tips. Refer to the change tips wizard for instructions Refer to the Electrophoresis Troubleshooting Module Check the water reservoir. Replenish the water and verify that the fluid line sinkers for the tubing are at the bottom of the reservoir. Verify that the autoloader tips are screwed in, they should be finger tight. Page 3 of 19

Observations Possible Causes Recommended Actions Verify the fittings above the syringes are finger tight. Locate the source of the bubbles, refer to the Checking for bubbles module. Replace the sample transfer syringe(s) using the change syringe wizard. Clogged loading tips Capillary specific Replace the autoloader tips. Refer to the change tips wizard for instructions Check the arrayt for any broken capillaries at the loading end or along the length of the array. Block the capillary in the Data Collection software to prevent loading at the capillary or replace the array. The capillary may be clogged. Block the capillary in the Data Collection software to prevent loading at the capillary or replace the array. Sections of the array view are black Sample issue Spatial calibration may have shifted Refer to the Missing Lanes Module Run sequencing standard controls, if these work but samples do not, contact technical support for troubleshooting the sample. Run a spatial calibration. Refer to the Missing Lanes Module Page 4 of 19

Turquoise Array Large bubble in cuvette, affects entire run small bubble in cuvette Page 5 of 19

Yellow Array Page 6 of 19

Striations Page 7 of 19

Green Page 8 of 19

Black Page 9 of 19

Checking fittings on the syringe pumps Array Fill Syringe Sample Transfer Syringes Sheath Flow syringe and pump From water reservoir to syringe Port 8: From polymer bottle to syringe From syringe to robot tip Port 8: From polymer bottle to syringe Port 9: From syringe to array. From water reservoir to syringe Port 9: From syringe to cuvette. From syringe to robot tip Page 10 of 19

Polymer bottle diagram Verifying that the polymer volume is sufficient for the run will prevent the introduction of bubbles into the system. Ensuring that the polymer is at room temperature before placing on the instrument will prevent degassing of the polymer while on the instrument. Verifying that precipitates have been dissolved and gently mixed back into solution will ensure that the polymer is homogenous. Do not combine polymer bottles. Page 11 of 19

Running service modules within Data Collection Step 1: Go to the Instrument menu > Select Utilities > Select Run Service Module > Step 2:Click the Select Module button Page 12 of 19

The service module utility should automatically search the Service Modules Folder, if not the folder can be found in the D://AppliedBio/Support Files/Data Collection Support Files/ Service Modules Double click on the module, in this example CuvetteFlush.mod. Page 13 of 19

After selecting run method, the steps of the service module will be listed. The module selected should be listed. Select Run Method. Page 14 of 19

When this window appears the module has completed, if there is an error during the module, please contact technical support. Array Diagram Check these 2 areas for any capillary breakage. Broken capillaries along the length of the array can be problematic if the array is not changed. Polymer will still flow through the capillary causing buildup of polymer in the area and potentially affecting the performance of neighboring capillaries. Page 15 of 19

Using the wizards Step 1: Go to the Instrument menu > Select Wizards > Select appropriate wizard Replacing the inline filter Step 1: The filter needs to be primed. Step 2: After installing the filter, run the CuvetteFlush module, which will flush lines, filter and cuvette with polymer and purge bubbles from the cuvette. Page 16 of 19

Restarting the 3700 DNA Analyzer IMPORTANT To prevent firmware and software memory problems, we recommend that you restart the instrument and the software once a week. To shut down and restart the instrument: Step 1: The instrument should not be running or extracting data. Step 2: Close the 3700 Data Collection software by selecting Shutdown from the File menu. Note You cannot use the Close button to exit the software. Step 3: Close the OrbixWeb Daemon software by right-clicking on its button in the taskbar and selecting Close from the pop-up menu. If you get a run-time message, click OK to close the message. IMPORTANT Do not shut down the OrbixWeb Daemon until after you have shut down the Data Collection program. Step 4: Restart the computer. After logging in, OrbixWeb Daemon should have automatically launched. Do not start Data Collection at this time. Step 5: a. Turn off the instrument using the On/Off button. b. Wait 30 seconds. c. Turn on the instrument. Step 6: When the green status light is steady, wait 1 minute. Step 7: Restart the 3700 Data Collection software. Page 17 of 19

Fluid Line Sinkers Fluid line sinkers Autoloader tips Tips should be screwed to finger tight and the bottom of the tips should be even. Page 18 of 19

Contacting Technical Support By phone: 1-800-831-6844, option 5 By email: ABTechnicalSupport@appliedbiosystems.com Page 19 of 19