Interlaboratory comparison in the pressure range from 0 to 2 MPa for accredited calibration laboratories

Similar documents
Gas flow calibrations performed at the National Metrology Institute of South Africa (NMISA)

EN ISO/IEC Technical Requirements

Traceable calibration of automatic weighing instruments in dynamic operation

INFLUENCE OF MEASURING PROCESS AUTOMATION ON UNCERTAINTY OF MASS STANDARD AND WEIGHTS CALIBRATION.

Performance Overview calibration Laboratory EP Instruments Messtechnik und Kalibrierung GmbH

THE PRESSURE SIGNAL CALIBRATION TECHNOLOGY OF THE COMPREHENSIVE TEST

Establishing Primary Gas Pressure Standard facility (0.140 to 70) bar Gas Piston Gauge at FCRI and its applications

Bi-lateral Comparison (APMP.L-K ) Calibration of Step Gauge

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Beamex. Calibration White Paper. Weighing scale calibration - How to calibrate weighing instruments

tradition. knowledge. responsibility. CALIBRATION OF ELECTRICAL MEASUREMENT EQUIPMENT CALIBRATION LABORATORY

COOMET TECHNICAL COMMITTEE 1.8 PHYSICAL CHEMISTRY

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

MET-101 Introduction to Metrology

Specific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities

EURAMET P1061 COMPARISON OF AIR TEMPERATURE CALIBRATIONS

FPG8601 Force Balanced Piston Gauge

PRESSURE TRANSDUCERS WITH UNIFIED (4-20) ma OR (10-50) ma OUTPUT SIGNAL

Supersedes: The copy of this document located on Measurement Canada s website is considered to be the controlled copy.

Gas mixtures. Individual solutions specifically for your application

THE USE OF Z FACTORS IN GRAVIMETRIC VOLUME MEASUREMENTS

GUIDANCE FOR THE APPLICATION OF EQUIPMENT CALIBRATION AND TRACEABILITY OF MEASUREMENTS ON INSPECTION BODIES

KENYA ACCREDITATION SERVICE

Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience -

CALIBRATION SERVICES HUMIDITY DEW POINT AIR VELOCITY MASS FLOW TEMPERATURE CO 2 PRESSURE. (+34)

AGILENT TURBOMOLECULAR PUMPS MEASUREMENT METHODOLOGY

Organisation Internationale de Métrologie Légale

Series Ruska Digital Pressure Indicator. GE Sensing & Inspection Technologies. Features

METROLOGICAL ASSURENCE OF O3, CO2, CH4 AND CO CONTROL IN ATMOSPHERE

UKAS calibration laboratory 8809 Temperature, pressure and electrical signals calibration guide

How to evaluate pressure measurement specifications

Pneumatic high-pressure controller Model CPC7000

PG7000 high line differential mode for defining differential pressure at line pressures above atmospheric pressure (PG7000 V2.

Pressure Metrology in Mexico

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

ENVIRONMENTAL AND POLLUTANTS GAS ANALYZERS

Digital Pressure Gauge Main features: Data Sheet: BIT02B.417.R6.EN COVER on request

Light Lab. Reliable Measurements through Regular Calibration. DAkkS

CERTIFICATE OF ACCREDITATION ISO/IEC 17025:2005 ANSI/NCSL Z (R2002)

Vaisala Pressure, RH & Temp. Transmitter. The instrument was operational upon receipt. The instrument was adjusted and calibrated.

NMISA New Gas Flow Calibration Facility

PPC4E Pressure Controller/Calibrator Superior rangeability and reliability

GMC-I Service GmbH Calibration Center

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF PRESSURE METROLOGY

COOMET TECHNICAL COMMITTEE 1.8 PHYSICAL CHEMISTRY. CООМЕТ project 576/RU/12

7250 Sys. Ruska Multi-Range Pressure Calibration System. GE Sensing & Inspection Technologies. Features. Applications

P3000 Deadweight Tester Setup, Part 1: Limited Partial Correction Method

Proficiency Testing for Surface Roughness Standard and Groove Depth Standard

IDL01. Battery Powered Precision Digital Gauge for Leak Testing. Stainless Steel Sensor. class 0.05

Speciality Gases Practice 4. Our quality, your safety. Accredited Test and Calibration Laboratory Speciality Gases Centre in Hörstel.

Withdrawn at December 31, Guideline DKD-R 6-2. Calibration of Measuring Devices for Vacuum DEUTSCHER KALIBRIERDIENST. Part 3

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Pneumatic high-pressure controller Model CPC7000

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Guide No. 2/2009. Design, training and equipment Martin Firlus. Physikalisch Technische Bundesanstalt Braunschweig und Berlin

Calibrating a Micropipette

Certificate of Accreditation

NMISA NEW GAS FLOW CALIBRATION FACILITY. Deona Jonker 7 October 2013

General Accreditation Guidance. User checks and maintenance of laboratory balances

Series 7250 Ruska High-Speed Digital Pressure Controller. GE Sensing. Features

BRML INM. National Metrology Institute. COOMET Comparison COOMET.L-S20 676/MD/15. Calibration of gauge blocks by comparison. Technical protocol

Certificate of Accreditation

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z CALIBRATION

SIKA Ba_ODWT15_en 05/2018. Operating manual for oil deadweight tester 0.015% accuracy ODWT15

Pressure Calibrators. Series PC..., PM...

CALIBRATION CERTIFICATE # 500

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010

Recommended Practices for Calibration Laboratories

CORESTA Guide N 6. February A User Guideline for the use of balances for cigarettes and cigarette related products

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

Test pumps. Pressure sources. Good reasons for proper and reliable calibration

Requirements for the certification of movements and mechanical watches resistant to magnetic fields of 1.5 T ( G)

OPERATING PROCEDURE. Number: V2. Authors. Approvals

Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation

WaterSense Specification for Spray Sprinkler Bodies. Version 1.0

SIM MWG7 Mass and Related Quantities. Report to CCM

Technical Report. 5th Round Robin Test for Multi-Capillary Ventilation Calibration Standards (2016/2017)

Publishable Summary for 14IND06 pres2vac Industrial standards in the intermediate pressure-to-vacuum range

TG GUIDELINES CONCERNING CALIBRATION INTERVALS AND RECALIBRATION

ISO Vacuum gauges Calibration by direct comparison with a reference gauge

Programmable pressure and temperature transmitters PTM/RS485

Comparison APMP.QM-S2.1 Oxygen in nitrogen at atmospheric level

Operational experiences with the EuroLoop Liquid Hydrocarbon Flow Facility

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017 ANSI/NCSL Z

Calibration equipment

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. PRECISION MEASURING CORP Groesbeck Hwy. Fraser, MI Robert Meyer Phone:

Sartorius CCL kg Prototype Mass Comparator

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

Hydraulic pressure balance Model CPB5800

PRESSURE MONITORING ND SAFETY SYSTEMS. Davor ZVIZDIC, Lovorka GRGEC BERMANEC FSB-LPM, Zagreb, Croatia.

High-speed pressure sensor Model CPT6140

ACCURATE PRESSURE MEASUREMENT FOR STEAM TURBINE PERFORMANCE TESTING

Calibration equipment

General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT

Precision pressure sensor Basic version Model CPT6020

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Force Standards Comparison between Mexico and Brazil. Jorge C. Torres Guzmán*, Daniel Ramírez Ahedo*, Jorge P. Cruz**, Jorge M. E.

Programmable pressure and temperature transmitters PTM/RS485

Transcription:

Int. J. Metrol. Qual. Eng. 6, 307 (2015) c EDP Sciences 2015 DOI: 10.1051/ijmqe/2015021 Interlaboratory comparison in the pressure range from 0 to 2 MPa for accredited calibration laboratories Lovorka Grgec Bermanec and Davor Zvizdic Laboratory for Process Measurement, Faculty of Mechanical Engineering and Naval Architecture, University of Zagreb, Croatia Received: 4 September 2015 / Accepted: 6 September 2015 Abstract. This paper describes procedure and results of the interlaboratory comparison that was organised by Croatian national pressure laboratory (HMI/FSB-LPM) between eight accredited pressure calibration laboratories. Comparison was performed in the gauge pressure range from 0 to 2 MPa with pressure transducer as transfer standard and using gas as transmitting medium. Basic aim of this comparison was to improve measurement methods and the measurement uncertainties of each participating laboratory. Analysis of the measurement results are given by calculating the deviation, E n, normalized with respect to the stated measurement uncertainties. Keywords: Interlaboratory comparison, pressure, measurement uncertainty 1 Introduction The interlaboratory comparison (ILC) program described in this paper is designed to help calibration laboratories to achieve confidence in pressure calibration results, validate reliable uncertainty levels and meet ISO 17025 requirements [1]. Participation in interlaboratory comparisons also provides laboratories an opportunity for independent assessment of their measurement performance. Programme is developed in the laboratory for process measurement (LPM) at University of Zagreb, faculty for mechanical engineering and naval architecture (FAMENA), which is part of distributed National Metrology Institute in Croatia (Croatian Metrology Institute-HMI). Calibration in the range from 0 to 2 MPa in gauge pressure range is of great importance for all pressure laboratories because it covers many technical applications and because this range of pressure is mostly used when pressure transmitting medium is gas. The objective of described comparison is to determine the relative agreement between gas pressure standards. 2 Participants, transfer standard and measurement procedure 2.1 Pilot laboratory and participants The reference values for calibration of pressure gauges, described in this paper were provided by the LPM. As pilot laboratory, LPM was responsible for preparing the Correspondence: lovorka.grgec@fsb.hr Table 1. Reference standards used by participating laboratories. Participating Laboratory Pressure standard (Town) 0upto20bar 1. MARUS ATM d.o.o. (Zagreb) Gas Pressure Balance 2. Metron Instruments (Zagreb) Pressure transducer 3. Ravnoteža (Zagreb) Pressure transducer 4. Petrokemija d.d. (Kutina) Gas Pressure Balance 5. Inspekt Metrolab (Zagreb) Pressure transducer 6. BMB (Zagreb) Gas Pressure Balance 7. CEI-IETA (Zagreb) Pressure transducer 8. STSI (Zagreb) Gas Pressure Balance measurement instructions, controlling the stability of the transfer standard, calculating the results and preparing the final report. The activities of the LPM are in the area of measurement of thermal and process values, developing and maintaining national pressure, temperature and humidity standards with published CMC s (Calibration Measurement Capabilities) in the BIPM key comparison database. Laboratory is accredited according to the EN ISO/IEC 17025 for temperature pressure and humidity calibrations by the DAkkS (Deutsche Akkreditierungsstelle registration number DKD-K-35601) from 2002 to 2012 and from 2013 by Croatian Accreditation Agency. Laboratory participated in many EURAMET (European Association of National Metrology Institutes) projects and intercomparisons and start to develop national interlaboratory comparison schemes from 2010 [2]. Other participating laboratories and their pressure standards are shown in Table 1. Article published by EDP Sciences

307-p2 International Journal of Metrology and Quality Engineering 2.2 Transfer standard and calibration procedure Digital pressure transducer (DRUCK DPI 615) as transfer standard of high accuracy was provided by LPM. Measurement range of transfer standard was from 0 up to 20 bar with 0.001 bar resolution. Transfer standard was circulated in eight laboratories from January till June 2010, transported by car. All participants agreed with following protocol: The pressure medium is nitrogen. More than one reference standard can be used. After setting up, there must be enough time for the pressure transducer to reach ambient temperature. The surface of the upper level of the adapters considered as reference level for pressure measurements. Readings of the digital measuring unit must be in bar. The calibration was performed in three measurement series according to, type B procedure [3], as follows: 1. After having opened the reference side of the pressure transducer to atmosphere, the reading is made. 2. The calibrating unit of the digital measuring unit is used to set the reading to zero. The ambient pressure and the ambient temperature shall then be read and recorded. 3. A pressure corresponding to maximum pressure of the device under test, but not more than 25 bar, is summoned up within two minutes. Then, the pressure must be let out to p e = 0 bar, this is repeated twice after two minutes. The zero-point value must be read and recorded. 4. Pressures of 2, 4, 6, 8, 10, 14, 16 and 20 bar must be produced, but not more than 25 bar. 5. Step 4 is repeated in reverse order. 6. Pressure is led off. Reading is made after 5 min. Reading and recording of the ambient pressure and ambient temperature is done. Read the pressure after 10 min for the second time. 7. The participating laboratories are requested to write down their results in the enclosed data sheets and to additionally issue one of their standard calibration certificates. Depending on type of pressure standard, corrections of nominal pressure were applied. Calculations of effective pressure were done by each laboratory according to their own procedure. groups: uncertainties of the pressure standard (usually standard piston/cylinder assembly or standard pressure transducer), uncertainties of the calibration method (including hydrostatic deviations from reference level and current measurements) and uncertainties associated with device under test, DUT (resolution, repeatability and hysteresis). Since we have only three readings on each pressure point, Type A uncertainty can not be calculated from statistics but repeatability and hysteresis, which can be dominant influence quantities, are taken into account as Type B uncertainties. Repeatability was calculated as difference between two ascending measurements and hysteresis as difference between ascending and descending readings, taking into account rectangle probability distribution asdescribedin[3]. Uncertainty of pressure standard was estimated from calibration report as Type B taking into account normal probability distribution. All uncertainty contributions can be expressed in the unit of pressure (bar) and there was no need for determination of sensitivity coefficients. Combined standard uncertainty, u, for each calibration point was calculated after corrections as: where: u PC u 2 = u 2 PC + u 2 Corr + u 2 H + u 2 r + u 2 f 0 + u 2 b + u 2 h, (1) is P/C unit uncertainty estimated from manufacturer s calibration report, u Corr is P/C unit uncertainty under conditions of calibration, u H is uncertainty caused by difference in height, u r is resolution uncertainty caused by DUT, u f0 is uncertainty caused by zero deviation of pressure transducer (DUT), u b is uncertainty caused by repeatability of pressure transducer (DUT), u h is uncertainty caused by hysteresis of pressure transducer (DUT). Expanded uncertainty, U, for each point was calculated as: where coverage factor k =2. U = ku, (2) 3 Measurement uncertainty estimation The uncertainty of laboratory measurement is equally as important as the measurement itself and demonstrates the capability of laboratory to make accurate measurements. For the uncertainty analysis in this intercomparison it was agreed that participants can follow DKD [3] or EURAMET [4] recommendations for calibration of pressure gauges, both based on ISO-GUM method [5]. Evaluation model for one calibration point is described in continuous. Following exactly the references [3, 4] recommendations, sources of uncertainty are divided in three 4 Results Calibration results for 9 measurement points including deviations from standard pressure and measurement uncertainties are reported to LPM and analysed to ensure uniform treatment for all participants. Deviations shown in Figure 1 pertains to all participating laboratories, depicted as LAB 1, LAB 2, LAB 3... without showing the names of the laboratory. Deviations with stated measurement uncertainties for each pressure point are shown individually (Figs. 2 10).

L.G. Bermanec and D. Zvizdic: Interlaboratory comparison for accredited calibration laboratories 307-p3 Fig. 4. Deviations for all participating laboratories respective to associated expanded uncertainty at 4 bar. Fig. 1. Deviations in calibrated range for all laboratories. 6 bar 0 bar Fig. 2. Deviations for all participating laboratories respective to associated expanded uncertainty at 0 bar. Fig. 5. Deviations for all participating laboratories respective to associated expanded uncertainty at 6 bar. Fig. 6. Deviations for all participating laboratories respective to associated expanded uncertainty at 8 bar. Fig. 3. Deviations for all participating laboratories respective to associated expanded uncertainty at 2 bar. 4.1 Stability of the standard Stability of the standard in this case was assessed by re-calibration of transfer standard. LPM has calibrated transfer standard before comparison and six months later after all measurements were performed. No intermediate calibrations were performed. Stability of the standard was within 1 mbar (Fig. 11). 4.2 Reference value and E n numbers The most widely accepted method of data analysis compares each laboratory s measured data and reported measurement uncertainties against established reference value. The result is measure of agreement between participants, E n. When the resultant E n for any measurement is between +1 and 1 no corrective action is usually required [6]. If the resultant E n is greater than +1 or 1this provides a signal that something might be wrong with a laboratory s calibration process, equipment or standards,

307-p4 International Journal of Metrology and Quality Engineering 10 bar Fig. 7. Deviations for all participating laboratories respective to associated expanded uncertainty at 10 bar. - - 14 bar Fig. 8. Deviations for all participating laboratories respective to associated expanded uncertainty at 14 bar. - - 16 bar Fig. 9. Deviations for all participating laboratories respective to associated expanded uncertainty at 16 bar. 0,050 - - 20 bar Fig. 10. Deviations for all participating laboratories respective to associated expanded uncertainty at 20 bar. Fig. 11. Deviations in the calibrated range before and after comparison. Table 2. E n numbers of participating laboratories 1 to 4 compared with LPM. Presssure E nlab1 E nlab2 E nlab3 E nlab4 bar 0 0.33 0.00 0.01 0.49 2 1.00 0.10 0.00 0.21 4 0.93 0.10 0.02 0.85 6 0.93 0.10 0.04 0.98 8 0.91 0.00 0.04 0.58 10 0.92 0.00 0.06 1.08 14 0.92 0.10 0.61 16 0.20 0.78 20 0.10 0.72 competence of personnel, stated uncertainties, etc. E n = M LP M M lab (U 2 lab + U 2 LP M ), (3) where M LP M is measurement result of LPM, M lab is measurement result assigned by participated laboratory, U LP M is uncertainty of M LP M. U lab is the uncertainty of M lab and should include an allowance for the performance of the measurement service. E n is displayed using chart and graph for one real testing. Resulting E n values are given in Tables 2 and 3. 5 Conclusion Eight Croatian accredited laboratories compared their calibration results and estimated measurement uncertainties using the same procedure. Equivalence respective to reference value is fully realized by 4 participants. From the results it can be seen that number of occurrence of E n 1 is 4 but in different pressure points. Those laboratories should analyse their measurements or increase calibration measurement capabilities (CMC values). It is clear that participation in such programme has significant educational element. Basically every part of

L.G. Bermanec and D. Zvizdic: Interlaboratory comparison for accredited calibration laboratories 307-p5 Table 3. E n numbers of participating laboratories 5 to 8 comparedwithlpm. Presssure E nlab5 E nlab6 E nlab7 E nlab8 bar 0 0.05 0.32 0.00 0.02 2 0.15 0.00 0.00 0.18 4 0.25 0.32 0.00 0.35 6 0.34 0.00 0.67 0.47 8 0.53 0.55 0.00 0.58 10 0.62 0.49 0.45 0.61 14 0.78 0.61 0.28 0.80 16 0.86 0.69 0.28 0.90 20 1.02 0.67 0.00 1.11 Laboratory quality system comes into play during participation in a ILC. Participating laboratories can use such intercomparison: For cross train of their personnel. To demonstrate technical competence. To identify potential areas for improvement. Data can be analysed to prevent incorrect results from being reported (ISO 17025: 5.9.2). References 1. International standard ISO/IEC 17025:2005, General requirements for the competence of testing and calibration laboratories 2. D. Zvizdić, L. Grgec Bermanec, Proficiency test in the gauge pressure range for Croatian accredited laboratories, Proceedings of the 3rd International Conference of Metrology (CAFMET 2010), Cairo, Egypt, 2010 3. Guideline DKD-R6-1, Calibration of pressure gauges, 2003 4. EURAMET/cg-17/v.01, Guidelines on the Calibration of Electromechanical Manometers, 2007 5. Evaluation of measurement data, Guide to the expression of uncertainty in measurement, JCGM, September 2008, 100:2008 6. International standard ISO 13528:2005, Statistical Methods for use in proficiency testing by interlaboratory comparisons