Welker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual

Similar documents
Installation, Operation, and Maintenance Manual

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER COMPOSITE GAS SAMPLER

Installation, Operation, and Maintenance Manual

Installation, Operation, and Maintenance Manual. Welker Constant Pressure Cylinder With Solid Indicator (Non-mixer) Model CP-30SI

Installation, Operation, and Maintenance Manual

Installation, Operation, and Maintenance Manual. Welker Constant Pressure Cylinder With Welker Magnetic Indicator (With Gravity Mixer) Model CP2GM-HP

Welker Constant Pressure Cylinders With Welker Solid Indicator, Syringe T-Handle, or Vortex Mixer

Installation, Operation, and Maintenance Manual

PP-1, PP-1W, PP-2, & PP-3

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER RELIEF VALVE

Installation, Operation, and Maintenance Manual

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER FILTER DRYER

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER CONSTANT PRESSURE CYLINDER

Constant Pressure Crude Oil Container Model CPCCP

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER INLOOP ACE CRUDE OIL SAMPLER

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER CONSTANT PRESSURE SAMPLE CONTAINER

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:

RD(H)20/25 Pressure-Reducing Regulator User Manual

LRS(H)4 Pressure-Reducing Regulator User Manual

Anderson Greenwood Series 93 Positive Pressure POSRV Installation and Maintenance Instructions

Discontinued. Powers Controls. Technical Instructions Document No P25 RV Rev. 1, May, RV 201 Pressure Reducing Valves.

Anti-flood device Model B-1

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER TRANSPORTABLE CRUDE OIL CONTAINER

OPERATING AND MAINTENANCE MANUAL

TECHNICAL DATA ANTI-FLOOD DEVICE MODEL B-1 1. DESCRIPTION

DynaPak Gas Sampler S Y S T E M S U P P O R T M A N U A L DP-2000

1805 Series Relief Valves

DynaPak 2000L System Support Manual Version

Product Manual. Description. Specifications. CVS Type 1301F and CVS Type 1301G Regulator. Introduction

Float Operated Level Controllers

P5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

Installation, Operation and Maintenance Manual for Back Pressure Regulator

64 Series Pressure Reducing Regulators

MEGR-1912 Instruction Manual

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

Instruction Manual Updated 7/26/2011 Ver. 2.2

TECHNICAL DATA. Q = C v P S

Model VR6 System. Installation, Operation & Maintenance

Operating Instructions Model and Hydrostatic Test Pump

1200B2 Series Service Regulators. Instruction Manual

PANACEA P100 LOW PRESSURE USER MANUAL

TECHNICAL DATA 3 MODEL G-3000 DRY VALVE RISER ASSEMBLY

Helium-Oxygen Blender

Assembly Drawing: W-311B-A01, or as applicable Parts List: W-311B-A01-1, or as applicable Special Tools: , , &

Pressure Dump Valve Service Kit for Series 2300 Units

Types S100K and S102K Pressure Regulators

RHPS Series RD(H)F40 User Manual. Read the complete manual before installing and using the regulator.

INSTALLATION INSTRUCTIONS. CVS 67CFR Pressure Reducing Instrument Supply Regulator INTRODUCTION

THE HF-300 SERIES. Operating and Service Manual. Series includes all variants of HF-300/301

TECHNICAL DATA. Q= Cv S

INSTRUCTION MANUAL. CVS P37 LP Gas Reducing Regulator

TECHNICAL DATA Q = C. v P S. 2 Model G-2000 Dry valve. Page 1 of 13

THE BP-301 SERIES. Operating and Service Manual. Series includes all variants of BP-301 (LF 0.1Cv / MF 0.5Cv)

MAINTENANCE PROCEDURE FOR X 650

PANACEA P200 STAINLESS STEEL USER MANUAL

Air-Oxygen Blender. Service Manual. Model No. PM5200 Series PM5300 Series (shown)

PANACEA P125 USER MANUAL

Regulator from 1800 B series

4400 TwinHybrid Gas Seal

Types S108K and S109K Pressure Reducing Regulators with Integral Slam-Shut Device

Pressure Dump Valve Service Kit for Series 3000 Units

299H Series. Introduction. P.E.D. Categories. Specifications. Installation. Warning. Installation Guide English September 2012

TECHNICAL DATA. Q = C v P S

RS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.

TECHNICAL DATA. Q = C v P S

Type S301 & S302 Gas Regulators INTRODUCTION INSTALLATION. Scope of Manual. Description. Specifications. Type S301 and S302. Instruction Manual

TECHNICAL DATA. Q = C v P S

RG1200 Service and Repair Manual

1800C and 1800C-HC Series Service Regulators

TECHNICAL DATA. Q = C v P S

Combination Air Valve

TECHNICAL DATA. Q = C v P S

Adjustable. Unloader Valve. O&M Manual. Part #

SUBMITTAL NOTES PROJECT: Ross Model 50RWR-A Pilot Operated Surge Relief Valve with Hydraulic Anticipation. Size: inch / mm

Type HSR Pressure Reducing Regulator for Residential, Commercial, or Industrial Applications

Halsey Taylor Owners Manual STOP!

Model Secure-Gard Pilot Operated Vent Valve SECTION II. Remove all packing material inside and outside of the valve prior to installation.

RIX COMPRESSORS. MICROBOOST Owner s Manual CONFIGURATION A MB-115 MB-230. November 21, 2002 APPLICABILITY [ A]

1305 Series Pressure Reducing Regulators

Halsey Taylor Owners Manual 4410 Freeze Resistant Tubular Fountain STOP!

TECHNICAL DATA CAUTION

Instruction Manual - Diaframless TM Ejector Chlorine, Sulfur Dioxide and Ammonia

WHEATLEY WHEATLEY SERIES 500 SWING CHECK VALVE. Installation, Operation and Maintenance Manual

Operation Manual Piston Sensed Gas Pressure Regulators

ROTATING DISK VALVES INSTALLATION AND MAINTENANCE 1. SCOPE 3 2. INFORMATION ON USAGE 3 3. VALVE TYPES 3 4. OPERATORS 5 5. VALVE CONSTRUCTION 6

Installation Troubleshooting Maintenance Instructions Installation / Start-up

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y

Combination Air Valve Model

Types 749B and R130 Changeover Manifolds

Operating Instructions Models and Hydrostatic Test Pumps

WW-730. Pressure Sustaining/Relief Control Valve

COMPACT TRIPLE WALL OUTLET STATION MODEL INSTALLATION AND OPERATING INSTRUCTIONS

Operating and maintenance manual Filter and reducing station Series / 1.0

Paintball Marker. User s Manual. 530 South Springbrook Road Newberg, OR 97132

LINDGREN-PITMAN, INC. LS-3 / LS-4 Line Setter

INSTALLATION, OPERATION AND SERVICE MANUAL ABS AIR BAG LIFT

Types PT and ETP Dome-Loaded Pressure Regulator Valves

Ball valve HKSF-W100. Ball valve HKSF-W100. RMA Kehl GmbH & Co. KG Oststrasse 17 D Kehl / Germany

Anderson Greenwood Series 400 Diaphragm Pilot Operated Safety Relief Valves. Installation and Maintenance Instructions for

DuraSite 3000 Sample Vessel User's Guide. Version Rev-1

Transcription:

Installation, Operation, and Maintenance Manual Welker Sampler Model GSS-1 The information in this manual has been carefully checked for accuracy and is intended to be used as a guide to operations. Correct operating and/or installation techniques, however, are the responsibility of the end user. Welker reserves the right to make changes to this and all products to improve performance and reliability. 13839 West Bellfort Sugar Land, Texas 77498-1671 U.S.A. Tel.: (800) 776-7267 Tel.: (281) 491-2331 Fax: (281) 491-8344 www.welkereng.com Copyright 2009, Welker Inc. IOM-019

TABLE OF CONTENTS 1. GENERAL...3 1.1 Introduction...3 1.2 Specifications...5 2. INSTALLATION INSTRUCTIONS...6 2.1 Installing the Sampler...6 2.2 Start-Up and Sequence of Operation...9 3. MAINTENANCE INSTRUCTIONS...13 3.1 General Information...13 3.2 Instructions...14 4. INLINE RELIEF INSTRUCTIONS...20 4.1 General Information...20 4.2 For Single Cavity Cylinders...20 4.3 For Constant Pressure Cylinders...21 4.4 Inline Relief Maintenance...22 Welker, Welker Jet, and WelkerScope are Registered Trademarks owned by Welker. IOM-019 Page 2 of 22

1. GENERAL 1.1 Introduction We appreciate your business and your choice of Welker products. The Installation, Operation and Maintenance liability for this product becomes that of the purchaser at the time of receipt. Reading the applicable IO&M Manual prior to installation and operation of this equipment is required so that you have a full understanding of its application and performance prior to commencement of use. If you have any questions, please call 1-800-776-7267 or 1-281-491-2331 in the USA. The Welker GSS-1 Gas Sampler is designed to extract a representative sample of the flowing product from the center one-third of the pipeline with the aid of a sampler probe. The GSS-1, with its Vanishing Chamber collection head, is capable of extracting a representative sample from the flowing stream and pumping the sample into a sample cylinder. IOM-019 Page 3 of 22

FIGURE 1 NOTE: The GSS-1 sampling system does not require an external pneumatic supply when installed as in Figure 1. This system uses the pipeline gas for the instrument pressure. By design, a unique self-purging system through the sample chamber area ensures a true representative sample with each sample cycle. IOM-019 Page 4 of 22

1.2 Specifications Product Sampled Materials of Construction Natural gas, or other compatible gaseous fluids 316 stainless steel, optional aluminum or stainless steel regulator (where applicable) Sample Grab Sizes Grab Rate Line Temperature Limit Maximum Line Pressure Sample Outlet Connection Area Classification 0.22, 0.5, 1.0, 1.5cc Up to 15 grabs per minute -20 F (-28 C) to 250 F (120 C) standard 2,160 psi (148 bar) standard 1/4" FNPT standard Can be used in hazardous areas as applicable IOM-019 Page 5 of 22

2. INSTALLATION INSTRUCTIONS The installation instructions are written from the position that the GSS-1 is part of a complete sampler system. If it is purchased as a sample pump alone, the system should be constructed in a fashion compatible to the following instructions. 2.1 Installing the Sampler To place the sampler into operation, the following steps should be followed: 2.1.1 After unpacking the unit, check it for compliance and any damages that may have occurred during shipment. NOTE: Claims for damages caused during shipment must be initiated by the receiver to the carrier. Welker is not responsible for any damages caused from mishandling by the freight company. NOTE: When sealing fittings with PTFE tape, refer to the proper sealing instructions for the tape used. 2.1.2 A sample probe should be located in the least turbulent area available of the flowing stream; i.e., not in a header or blow-down stack and away from obstructions, elbows or partially closed valves. The sample probe should be installed reaching approximately into the center one-third of the pipeline. NOTE: Typically, the Welker Gas Sampler is installed utilizing a single sample probe. Its unique self-purging feature uses pipeline gas for the instrumentation supply source, thus purging the sample line prior to each sample grab. If instrument air is to be used for the instrument supply, the sampler should be installed with a pitot probe or with 2 single probes. If 2 single probes are used, IOM-019 Page 6 of 22

one should be located upstream and the other one downstream of a moderate pressure drop such as an orifice plate or control valve. This will create a hot loop for the sampler that will allow a real-time sample to be taken with each new actuation. 2.1.3 The sampler should be located as close to the sample point as practical and above the pipeline probe. A location up to three feet from the probe is desirable. 2.1.4 The sample probe valve should be a large ported valve (i.e., fully opening ball valve, block valve, etc.). 2.1.5 Once the sampler is mounted and the probe is installed, hook-up can be completed. The sample cylinder should be located as close to the sampler as is possible. 2.1.6 If using a single-cavity sample cylinder, the inline relief of the pump should be set at 100 psi above maximum line pressure (see Section 4). 2.1.7 If using a constant pressure sample cylinder, the inline relief of the pump should be set at approximately 200 psi and the cylinder must first be pre-charged. 2.1.8 Pre-charging a constant pressure sample cylinder with a pre-charge gas can be done one of three ways: 2.1.8.1 Connect the cylinder pre-charge valve to the pipeline (see Figure 2). 2.1.8.2 Connect the cylinder pre-charge valve to an extra port in the sampler probe as in Figure 1. IOM-019 Page 7 of 22

NOTE: By using the pipeline pressure as the source of pre-charge pressure for the constant pressure sample cylinder the sample will be maintained at actual pipeline pressure conditions. 2.1.8.3 Auxiliary pre-charge gas (i.e., nitrogen bottle). NOTE: If a constant pressure sample cylinder is used, refer to the instructions sent with the cylinder for complete details on pre-charging the cylinder. 2.1.9 Remove the cover or dome for manifold identification where applicable. 2.1.10 Using small diameter stainless steel tubing (1/4" O.D.), tube from the probe to the inlet port of the manifold (labeled "inlet") or the NPT connection directly on the sampler body itself. Where a hot loop or pitot probe is employed, tube to P 2 as previously discussed in 2.1.2. The tubing should always slope downward from the sampler to the probe and be free of sags and loops. NOTE: Nothing should be installed between the probe and the inlet of the sampler, such as filters, drips or regulators. These could have an adverse effect on the integrity of the sample. Also, make sure that all fittings are tight and NPT connections are PTFE taped or doped. 2.1.11 Using 1/8" O.D. stainless steel tubing; connect from the sample outlet port on the manifold if applicable or the standard relief cap to the sample inlet valve on the sample cylinder (see Figure 2). IOM-019 Page 8 of 22

FIGURE 2 2.1.12 Close all valves on the sample cylinder. 2.1.13 If the sampler is to be actuated by an auxiliary air or gas supply as mentioned under 2.1.2, tube the regulated instrument air supply to the port on the rear of the diaphragm pump (80 to 100 psi maximum). 2.1.14 Connect the power supply to the sampler s mode of actuation. 2.1.14.1 If electronics are provided, reference the installation and operating instructions for the particular model supplied 2.1.14.2 For samplers that are controlled from a customer-furnished remote controller, connect the remote controller output voltage signal to the solenoid valve on the GSS-1. 2.1.15 This completes the hook-up procedure. 2.2 Start-Up and Sequence of Operation NOTE: When pressuring the system, always open pipeline valves slowly. All connections must be checked carefully for leaks at full line pressure. No leaks are acceptable within the sample system. IOM-019 Page 9 of 22

2.2.1 All valves are still closed on the sampler and sample cylinder. 2.2.2 For constant pressure cylinders, go to Section 2.2.3. For single-cavity sample cylinders, go to Section 2.2.8. 2.2.3 Constant Pressure Cylinders - Open the product inlet valve on the constant pressure cylinder. Slowly open the pipeline gas supply valve to the pre-charge side of the constant pressure cylinder. If using a nitrogen bottle to pre-charge, open the valve from the bottle and the valve on the cylinder and allow pressure to fill the pre-charge side of the cylinder. This allows pre-charge pressure to be supplied to the cylinder and forces the piston to the product side if it is not already there. Once pressurized to acceptable pressure, leave valves open to allow for gas compression as cylinder is filled with sample. Check for leaks 2.2.4 Close the product inlet valve on the constant pressure cylinder. 2.2.5 Open the pipeline isolation valve on the probe that leads from the pipeline to the sampler. Make sure there are no leaks between the probe, sampler and the cylinder. NOTE: If your sampler was purchased to be used with a constant pressure cylinder, the inline relief is factory set at 200 psi. This can be verified by the gauge on the cylinder once the valve is opened. There will be a 200 psi differential between the gauge reading on the cylinder and the gauged pressure on the pipeline. If this setting is incorrect, see instructions to properly set the inline relief (see Section 4). 2.2.6 After the product inlet valve on the constant pressure cylinder is opened, check for leaks and tighten any fittings that are leaking. IOM-019 Page 10 of 22

NOTE: If your constant pressure cylinder is equipped with a product purge valve, open it to purge the tubing for approximately 3-5 seconds and close. When possible, plug purge valve when not in use. If your cylinder is not equipped with a purge valve, we recommend that a T and valve be used just prior to the inlet valve to provide a purge system. 2.2.7 You are now ready to begin the sampling timing cycles with the control system you chose. 2.2.8 Single Cavity Sample Cylinders - Remember that the inline relief must be set at a minimum of 100 psi above maximum line pressure (see Section 4). Slowly open the pipeline isolation valve on the probe leading to the sampler. NOTE: If your sampler was purchased to be used with a single cavity sample cylinder, the relief is required to be set higher than the maximum pipeline pressure to eliminate bleed by of the relief and pre-filling the cylinder. Mount the standard cylinder in the vertical position. 2.2.9 Now slowly open the bypass valve on the sampler manifold. Using a leak check solution, check for leaks. Tighten any fittings that are leaking. 2.2.10 Purge the tubing between the sampler and cylinder by cracking the tube fitting on the cylinder inlet or by using a by-pass manifold. Follow company procedure for preparing the standard cylinder for use. 2.2.11 Open the product inlet valve on the sample cylinder. 2.2.12 You are now ready to begin the sampling timing cycles with the control system you chose. 2.2.13 The instrument supply air should be set to 65-70 psi. IOM-019 Page 11 of 22

NOTE: In cases where pipeline pressures are in excess of 1,500psi, more instrument supply air may be necessary to take a sample. The instrument supply will have to be increased accordingly. The diaphragm motor can accept 100 psi in these cases. Increased supply should be used only when necessary and a relief valve for overpressure protection should always be used to prevent possible damage and accidents. 2.2.14 To test the sample system, take the following steps: a. If applicable, make sure the bypass valve is closed on the manifold assembly. b. Close the product inlet valve on the sample cylinder. c. Actuate the diaphragm pump. d. Observe gauge on the sampler manifold base if applicable or connect a gauge to the outlet of the pump. Build pressure to above line pressure. Let the unit sit for several minutes and check for a drop in pressure. If it does drop, check for leaks. If the sampler holds pressure, the unit is ready to be placed in operation. 2.2.15 Bleed the test pressure off. Open the product inlet valve on the sample cylinder. 2.2.16 The sampler is ready for operation. IOM-019 Page 12 of 22

3. MAINTENANCE INSTRUCTIONS NOTE: Refer to figure 4. 3.1 General Information Prior to maintenance or disassembly of the unit, it is advisable to have a repair kit handy for the system in case of encountering unexpected wear or faulty seals. Important: Maintenance should be performed on the sampler only after it has been isolated from the pipeline and all pressure has been vented. We recommend that the unit have annual maintenance under normal operating conditions. In the case of severe service, dirty conditions, excessive cycling usage or other unique applications that may subject the equipment to unpredictable circumstances, a more frequent maintenance schedule may be appropriate. Maintenance should be done in as clean a work area as possible. New seals supplied in spare parts kits are not lubricated. They should be lightly coated with lubrication grease (silicone grease or other) before they are installed into the equipment. This helps in the installation of the seals while reducing the risk of damage when positioning them on the parts. After the seals are installed, some additional lubrication can be applied to shafts or cylinder inner diameters to allow smooth transition of parts. All O-rings and seals can be easily cut or damaged and, thus, destroyed. Re-assemble the sampler with care. This is an instrument and should be handled accordingly. Lubricate all polished surfaces and seals with silicone grease or similar lubricant such as Krytox. IOM-019 Page 13 of 22

The following tools will be required for disassembly and maintenance. 1 / 2 " wrench 6" adjustable wrench Hex wrenches 1 / 8 " & 3 / 16 " Adjustable pliers Snap ring pliers FIGURE 4 3.2 Instructions 3.2.1 Close the pipeline isolation valve and vent all pressure. 3.2.2 Disconnect the instrument supply tubing from the sampler body. 3.2.3 Relieve and disconnect all the tubing from the sampler to the sample cylinder. 3.2.4 Unscrew the complete diaphragm housing from the sampler body (see Figure 5). IOM-019 Page 14 of 22

FIGURE 5 3.2.5 To replace the collection head, push the holding pin out (it is held in place by spring tension) and slip the shield off the shield/shaft adapter. Push the collection head out of the shield. The non-extrusion disc will come out first and does not need to be replaced if it is in good condition (see Figure 6). FIGURE 6 IOM-019 Page 15 of 22

NOTE: All collection heads are marked on the back side with the size, compound, and durometer. Take note of what you are replacing (i.e., "V-70 1.0" or V74 is a 70 durometer Viton collection head, 1.0 cc in volume). Consult Welker for other compounds available. 3.2.6 Lightly lubricate the inside surface of the shield and push in the new collection head. Replace the non-extrusion disc. 3.2.7 Slide the shield back onto the shield/shaft adapter and push the hold pin back into place. 3.2.8 To replace the remaining seals, remove the hex head nuts (18 each) and bolts (18 each) that hold the diaphragm case together. Separate the two halves and examine the diaphragm. Replace, if necessary (see Figure 7). Replacement diaphragms are flat and over time will convolute. FIGURE 7 IOM-019 Page 16 of 22

3.2.9 To unscrew the shield/shaft adapter, hold on to the diaphragm plate. With a wrench, unscrew the adapter from the shaft (see Figures 7 and 8). FIGURE 8 3.2.10 Carefully push the shaft through the lower diaphragm case until the diaphragm plate clears the case. Then, pull the plate and shaft out from the top. (See Figure 9). IOM-019 Page 17 of 22

FIGURE 9 3.2.11 Examine the shaft for damage. The shaft is polished and should be free of scratches and pits. If it needs to be replaced, place the diaphragm plate in a vise and remove the shaft with an adjustable wrench on the shaft flats. Replace and tighten securely. 3.2.12 From the body side of the lower diaphragm case, remove the snap ring and the variseal. When replacing the variseal, be careful not to damage it. 3.2.13 The bearing should not need replacing; however, if it is necessary, simply remove the snap ring from the opposite side of the diaphragm case and replace the bearing and/or seals. 3.2.14 Lubricate the shaft. Place the return spring in the center of the diaphragm case. Push the assembly back into the case, carefully guiding it through the seals. Replace the shield/shaft adapter securely. 3.2.15 Replace the diaphragm and install the upper diaphragm case and all nuts (18 each) and bolts (18 each). Cross bolt the case and then tighten all bolts securely. 3.2.16 Replace the collection head assembly, which includes the head shield, collection cup, non-extrusion disc and holding pin. 3.2.17 Re-install the vacuum breaker disc and Kel-f seal into the motor body and replace, if necessary. These are free floating in the body and simply need to be installed with the seal first and then the metal disc. (Note: The collection head will seal against the metal disc.) 3.2.18 Replace the seal around the male threads on the lower diaphragm case assembly and screw the lower diaphragm case back to the body (hand tighten only). IOM-019 Page 18 of 22

3.2.19 Replace instrument tubing. IOM-019 Page 19 of 22

4. INLINE RELIEF INSTRUCTIONS 4.1 General Information The function of the inline relief is to act as a check valve and to assure that the sampler pumps the product into the cylinder and that once the sample is taken, it cannot return to the pipeline even if the pipeline drops in pressure. The relief is located in the standard relief cap at the end of the sampler body. This relief is used with single cavity and constant pressure type cylinders. 4.2 For Single-Cavity Cylinders 4.2.1 See Figure 10. FIGURE 10 4.2.2 Make sure the sample cylinder inlet valve is closed. Slowly open pipeline isolation valve and allow full line pressure to the sampler. 4.2.3 With a 1/8" hex wrench, insert it inside the relief cap to the spring adjusting screw. IOM-019 Page 20 of 22

4.2.4 Adjust the spring tension to where no gas is bleeding through the set screw. Then, turn adjusting screw another full turn clockwise. NOTE: Each full turn on spring adjustment screw increases spring tension approximately 100 psi. The relief needs to be set approximately 100 psi above maximum line pressure. 4.2.5 Replace the tubing fitting and tighten tubing. 4.2.6 If applicable, the gauge on the manifold should show 0 psi. This will assure the relief is holding and the sampler must pump product into the cylinder. 4.2.7 Re-open the cylinder inlet valve. 4.3 For Constant Pressure Cylinders 4.3.1 See Figure 10. 4.3.2 Determine maximum pipeline pressure (i.e., 750 psi). 4.3.3 Make sure the sample cylinder inlet valve is closed. Slowly open pipeline isolation valve and allow full line pressure to the sampler. With a 1/8" hex wrench, insert it inside the relief cap to the spring adjusting screw. 4.3.4 If the pipeline pressure is 200 psi or lower, adjust the spring setting to stop the gas from bleeding past the poppet. Then, replace the seal tubing fitting and re-connect tubing. 4.3.5 If the pipeline pressure is above 200 psi, adjust the spring setting to allow all but approximately 200 psi to flow past the poppet. Replace the seal tubing fitting and re-connect tubing. Look at the gauge. If pipeline pressure is 750 psi, the gauge should read approximately 550 psi. If it reads 650 psi, add more tension by turning the wrench one full turn clockwise. If the gauge reads 450 psi, relieve IOM-019 Page 21 of 22

tension by turning the wrench one full turn counter-clockwise. This may have to be repeated to reach an approximate relief setting of 200 psi. NOTE: Keep in mind that a differential of around only 200 psi is needed as the relief acts only as a check valve in constant pressure applications. 4.4 Inline Relief Maintenance 4.4.1 Close pipeline isolation valve and relieve pressure from sampler. 4.4.2 Disconnect tubing and remove relief cap by unscrewing in a counterclockwise manner. 4.4.3 Remove the spring adjustment screw, spring and poppet. Examine the poppet sealing surface for damage. Replace, if necessary. 4.4.4 Replace the two O-rings (011/003) on the face of the relief cap. 4.4.5 Reassemble the pieces and thread the relief cap back into the body and tighten. 4.4.6 Relief is now ready to be reset. IOM-019 Page 22 of 22