Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS

Similar documents
High speed UHPLC-MS/MS determination of multiple steroids in human serum using the Nexera MX system for multiplex analysis

Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS

Development of a Clinical Research Method for the Measurement of Testosterone. Dihydrotestosterone

The Application of QuEChERS in the Extraction of Anabolic Steroids in Whole Blood

LCMSMS Solutions For Steroid Analysis

Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS

Introduction. Experimental

May 1, By John Heim,Doug Staples

A NOVEL RP-HPLC METHOD FOR THE QUANTIFICATION OF ICATIBANT IN FORMULATIONS

Determination of Selected ASTM Perfluorinated Compounds (PFCs) Using the Shimadzu LCMS No. LCMS-083. No. SSI-LCMS-083

The instrument should have a mass range from 2 to 2000 m/z or better Interface

Integration of amino acid, acylcarnitine and steroids analysis in single FIA/LC-MS/MS platform

IJPAR Vol.6 Issue 1 Jan - Mar Journal Home page:

SOP: Derivatized testosterone_01 Revision: 01 Date Effective: 05/17/15

Analysis of Melamine and Cyanuric Acid in Food Matrices by LC-MS/MS

(2 pts) Draw the line of best fit through the data and estimate the concentration of Fe in your sample solution.

Evaluation of linear and step gradient performance and retention time precision of the Agilent 1200 Series Rapid Resolution LC

Automated Low Background Solid Phase Extraction of Perfluorinated Compounds in Water. Ruud Addink Fluid Management Systems Watertown MA

Extended Application Note

Methylprednisolone detection in urine following local and oral administrations

Rapid Analysis of Metanephrine and Normetanephrine in Urine by Gas Chromatography Mass Spectrometry

GUIDE TO ULTRA-FAST HPLC. Quick Tips for Converting Conventional Reversed-Phase HPLC Separations to Ultra-Fast Separations

Now it is. easy to switch. CHS TM Steroid Profiling kit and software. Brochure not for distribution in the USA and Canada

Sensitive Analysis of Genotoxins by HPLC-ECD Using Boron-Doped Diamond Electrochemical Cell

TRACE LEVEL DETERMINATION OF PER- AND POLYFLUOROALKYL SUBSTANCES (PFASs) IN WATER USING THE AGILENT 6460 LC/MS/MS

Validation of the MRM pesticides from the Working Document SANCO/12745/2013 of November 2017 rev. 9(1)

International Journal of Generic Drugs ANALYTICAL METHOD PROCEDURES THIS SOP IS 'SWITCHED : OFF : ON'

GCMSD-Headspace Analysis SOP

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

1. INTRODUCTION CH 3 CH3

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Rapid and Reliable Detection of Dissolved Gases in Water

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay

Ascentis Express HPLC Columns with Fused-Core Technology

Agilent Dimension Software for ELSD User Manual

Extend the time window for the detection of low level anabolic androgenic steroids and their metabolites

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Fast method for Ginseng Analyses using Agilent Poroshell 120 Columns Scaled from a Traditional Method

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Experiment GC : Analysis of BTEX by GC-FID

Agilent 1290 Infinity LC System Applications requiring the Agilent Ultra-Low Dispersion Kit

More Speed and Resolution for Current Methods

WADA Technical Document TD2014EAAS. Endogenous Anabolic Androgenic Steroids Measurement and Reporting

Analysis of Melamine and Cyanuric Acid in Food Matrices by LC-MS/MS

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

International Journal of Research in Pharmaceutical and Nano Sciences Journal homepage:

ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION

Carrier Gases in Capillary GC

FRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

N Korea Japan China Thailand Taiwan Philipines EP-5 EP km. Source: Google Maps and

Retention Time Locking: Concepts and Applications. Application

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer

MSD 96-Well MULTI-ARRAY CRP Assay

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Acclaim RSLC. (Rapid Separation Liquid Chromatography)

Automated Determination of Dissolved Gases in Water Anne Jurek

biosensis Rat Fibronectin ELISA Kit Protocol

biosensis Human TNFα/Cachectin/TNFSF2 ELISA Kit Protocol

A VALIDATED STABILITY-INDICATING RP-HPLC ASSAY METHOD FOR BOLDENONE UNDECYLENATE AND ITS RELATED SUBSTANCES

biosensis Human Lipocalin-2/NGAL ELISA Kit Protocol

Rev. CO Packed Columns for Normal Phase Chromatography. TSKgel NH m INSTRUCTION MANUAL

SOP: Buck Scientific BLC-20P HPLC Operation

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

biosensis Human IGF-II, Insulin-like growth factor II, Somatomedin-A ELISA Kit Protocol

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

Longterm Detection of Anabolic Steroid Metabolites in Urine

Setting a New Standard in Flash Chromatography Performance

biosensis Mouse Vascular endothelial growth factor A/VEGF-A/VEGF-164/VEGF-1/VEGF- 120/VEGF-2 ELISA Kit Protocol

Carrier Gases in Capillary GC

Beyond Westgard Rules: Quality Control for Mass Spectrometry. Dr Russell Grant VP R&D Laboratory Corporation of America Holdings.

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

The instrument will process the samples one by one till all the samples have been cleaned up.

The Preparation of Low Concentration Hydrogen Sulfide Standards Gulf Coast Conference September 12, 2002 Galveston, TX Paper 050,12:45 pm Moody F

UWPR. LC-MS setup. Before you connect your trap and column. Connecting your trap and column

biosensis Mouse Brain-derived neurotrophic factor (BDNF) ELISA Kit Protocol

HUMAN IL6 KITS PROTOCOL

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

Renzo BAGNATI Mario Negri Institute of Pharmacological Research- IRCCS Milano - Italy

MV5: Automated concentration system

ISO INTERNATIONAL STANDARD

biosensis Rat Interleukin-1 beta, IL-1β ELISA Kit Protocol

Chemical Analysis for Methyltestosterone

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

Newborn Screening for Guanidinoacetate Methyltransferase (GAMT) deficiency. Marzia Pasquali, PhD, FACMG

biosensis Mouse Interleukin-1 beta (IL-1β) ELISA Kit Protocol

Low-Pressure Retention Time Locking with the 7890A GC. Application. Authors. Abstract. Introduction HPI

PFC Analysis Kit for ACQUITY UPLC System Guide

biosensis Rat Glial cell line-derived neurotrophic factor/gdnf total /ATF ELISA Kit Protocol

AKTA pure 25 New Owner s Intro

biosensis Rat IGF-1/Somatomedin/Insulin-like growth factor ELISA Kit Protocol

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation

Transcription:

Abstract Highly sensitive analysis of metanephrines in plasma is critical in the diagnosis and treatment of pheochromocytoma and paraganglioma. Here, a HILIC LC-MS/MS method was developed using a Raptor HILIC-Si column because this approach provides retention of all target compounds in highly organic mobile phases, thus providing the increased sensitivity needed to reach low detection limits. Data from linearity, accuracy, and precision testing demonstrate that accurate results are consistently obtained even at trace levels for these important clinical biomarkers. Introduction Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS By Shun-Hsin Liang and Frances Carroll The metanephrines metanephrine (MN), normetanephrine (NMN), and 3-methoxytyramine (3-MT) are methylated metabolites of epinephrine, norepinephrine, and dopamine, respectively (Figure 1). These catecholamine metabolites are released from the adrenal medulla and sympathetic nervous cells and are normally maintained at very low concentrations in the bloodstream. Elevated levels of these compounds in circulation can indicate the presence of pheochromocytoma or paraganglioma, so highly sensitive measurement of in vivo concentrations is critical for correct diagnosis, treatment, and long-term patient monitoring. Trace-level analysis of metanephrines in plasma using typical reversed-phase LC can be difficult due to limited chromatographic retention and lack of sensitivity. As an alternative, we developed an LC-MS/MS method based on hydrophilic interaction liquid chromatography (HILIC) using a Raptor HILIC-Si column. Combined with a simple solid phase extraction (SPE) procedure, an accurate and precise analysis of these important biomarkers can be achieved in a fast, 5-minute analysis time, making this a beneficial assay for high-throughput clinical labs. Figure 1: Structures of Metanephrine (MN), Normetanephrine (NMN), and 3-Methoxytyramine (3-MT) O OH H N O NH₂ HO HO MN OH 3-MT O NH₂ HO NMN

Experimental Sample Preparation Internal standard (IS) solution was prepared at 4 ng/ml for metanephrine-d3, and at 8 ng/ml for normetanephrine-d3 and 3-methoxytyramine-d4 in methanol. A plasma sample aliquot (200 μl) was mixed with 10 μl of IS solution and 600 μl of 50 mm ammonium acetate. The mixture was loaded onto an EVOLUTE EXPRESS WCX 96-well plate (30 mg) and washed with 1 ml water and 1 ml methanol:acetonitrile (50:50). The elution was performed twice with 0.9 ml of 5% formic acid in methanol:acetonitrile (50:50) and samples were then evaporated to dryness at 55 C under a gentle stream of nitrogen. Dried extracts were reconstituted with 100 μl of 100 mm ammonium formate in water (ph 3.0):acetonitrile (10:90) and injected (10 μl) for analysis. Calibration Standards and Quality Control Samples Charcoal stripped human plasma (BioreclamationIVT) was fortified with metanephrine, normetanephrine, and 3-methoxytyramine to prepare calibration standards and QC samples. The linearity ranges were 0.051-20.28 nmol/l (10-4,000 pg/ml) for metanephrine; 0.14-21.83 nmol/l (24-4,000 pg/ml) for normetanephrine; and 0.060-23.92 nmol/l (10-4,000 pg/ml) for 3-methoxytyramine. Three QC levels were prepared at 40, 400, and 2,500 pg/ml for all three analytes. The fortified standard and QC samples were prepared using the SPE procedure described above. LC-MS/MS analysis of metanephrines in plasma was performed on an ACQUITY UPLC instrument coupled with a Waters Xevo TQ-S mass spectrometer. Instrument conditions were as follows and analyte transitions are provided in Table I. Analytical column: Raptor HILIC-Si (2.7 µm, 50 mm x 2.1 mm; cat.# 9310A52) Mobile phase A: 100 mm ammonium formate in water, ph 3.0 Mobile phase B: Acetonitrile Gradient Time (min) %B 0.00 90 5.00 90 Flow rate: 0.3 ml/min Injection volume: 10 µl Column temp.: 30 C Ion mode: Positive ESI Table I: Analyte Transitions for Plasma Free Metanephrines LC-MS/MS Analysis Analyte Precursor Ion Product Ion Quantifier Metanephrine-d3 183.00 151.15 Product Ion Quantifier Metanephrine 179.94 148.22 165.01 Normetanephrine-d3 169.00 136.96 Normetanephrine 166.00 134.02 121.01 3-Methoxytyramine-d4 155.07 122.93 3-Methoxytyramine 151.00 119.00 91.02 Results and Discussion Chromatographic Performance Using a HILIC approach and Raptor HILIC-Si column, good retention was obtained for all analytes. A fast, 5-minute chromatographic analysis (Figure 2) was achieved with injection of reconstituted sample extract. No chromatographic interferences were observed in the analysis of blank plasma samples (Figure 3), indicating that the simple SPE procedure was effective and specific. In addition, the isocratic elution with HILIC showed consistent results over multiple injections indicating that the method was robust (Figure 4). www.restek.com 2

Figure 2: Human Plasma Fortified with Metanephrine, Normetanephrine, and 3-Methoxytyramine (50 pg/ml Calibration Standard) Conc. Peaks t R (min) (pg/ml) Precursor Ion Product Ion Metanephrine-d3 Metanephrine Normetanephrine-d3 Normetanephrine 3-Methoxytyramine-d4 3-Methoxytyramine 0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 Time (min) LC_CF0689 1. 3-Methoxytyramine-d4 (IS) 1.80 400 155.07 122.93 2. 3-Methoxytyramine 1.80 50 151.00 119.00 3. Metanephrine-d3 (IS) 2.03 200 183.00 151.15 4. Metanephrine 2.03 50 179.94 148.22 5. Normetanephrine-d3 (IS) 2.11 400 169.00 136.96 6. Normetanephrine 2.11 50 166.00 134.02 Column Raptor HILIC-Si (cat.# 9310A52) Dimensions: 50 mm x 2.1 mm ID Particle Size: 2.7 µm Pore Size: 90 Å Temp.: 30 C Sample Diluent: Mobile phase A:mobile phase B (10:90) Inj. Vol.: 10 µl Mobile Phase A: Water, 100 mm ammonium formate, ph 3.0 B: Acetonitrile Detector Ion Mode: Mode: Instrument Notes Time (min) Flow (ml/min) %A %B 0.00 0.3 10 90 5.00 0.3 10 90 MS/MS ESI+ MRM UHPLC Solid Phase Extraction Procedure: Charcoal stripped human plasma was fortified with analytes at 50 pg/ml (0.25, 0.27, and 0.30 nmol/l for metanephrine, normetanephrine, and 3-methoxytyramine, respectively). Internal standard (IS) was prepared at 4 ng/ml for metanephrine-d3 and at 8 ng/ml for normetanephrine-d3 and 3-methoxytyramine-d4 in methanol. The plasma sample (200 μl) was mixed with 10 μl of IS solution and 600 μl of 50 mm ammonium acetate solution. The mixture was loaded in an EVOLUTION EXPRESS WCX 96-well plate (30 mg) and washed with 1 ml water and 1 ml methanol:acetonitrile (50:50). The elution was performed twice with 0.9 ml of 5% formic acid in methanol:acetonitrile (50:50) and evaporated to dryness at 55 C under a gentle stream of nitrogen. Dried extract was reconstituted with 100 μl of diluent and injected (10 μl) for analysis. Figure 3: Blank Human Plasma Metanephrine-d3 Metanephrine Normetanephrine-d3 Normetanephrine 3-Methoxytyramine-d4 3-Methoxytyramine 0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 Time (min) LC_CF0695 Conc. Peaks t R (min) (pg/ml) Precursor Ion Product Ion 1. 3-Methoxytyramine-d4 (IS) 1.80 400 155.07 122.93 2. 3-Methoxytyramine - - 151.00 119.00 3. Metanephrine-d3 (IS) 2.04 200 183.00 151.15 4. Metanephrine - - 179.94 148.22 5. Normetanephrine-d3 (IS) 2.12 400 169.00 136.96 6. Normetanephrine - - 166.00 134.02 Column Raptor HILIC-Si (cat.# 9310A52) Dimensions: 50 mm x 2.1 mm ID Particle Size: 2.7 µm Pore Size: 90 Å Temp.: 30 C Sample Diluent: Mobile phase A:mobile phase B (10:90) Inj. Vol.: 10 µl Mobile Phase A: Water, 100 mm ammonium formate, ph 3.0 B: Acetonitrile Detector Ion Mode: Mode: Instrument Notes Time (min) Flow (ml/min) %A %B 0.00 0.3 10 90 5.00 0.3 10 90 MS/MS ESI+ MRM UHPLC Solid Phase Extraction Procedure: Internal standard solution (IS) was prepared at 4 ng/ml for metanephrine-d3 and at 8 ng/ml for normetanephrine-d3 and 3-methoxytyramine-d4 in methanol. Charcoal stripped plasma (200 μl) was mixed with 10 μl of IS solution and 600 μl of 50 mm ammonium acetate solution.the mixture was loaded to the EVOLUTION EXPRESS WCX 96-well plate (30 mg) and washed with 1 ml water and 1 ml methanol:acetonitrile (50:50). The elution was performed twice with 0.9 ml of 5% formic acid in methanol:acetonitrile (50:50) and evaporated to dryness at 55 C under a gentle stream of nitrogen. Dried extract was reconstituted with 100 μl of diluent and injected (10 μl) for analysis. 3 www.restek.com

Figure 4: Robust Column Performance for the Analysis of Metanephrines in Plasma 1 st Injection 0 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50 2.75 3.00 3.25 3.50 3.75 4.00 135 th Injection 0 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50 2.75 3.00 3.25 3.50 3.75 4.00 Time (min) LC_CF0701 Conc. 1st 135th Peaks (ng/ml) Precursor Product Ion Product Ion Injection (tr) Injection (tr) 1. 3-Methoxytyramine 1 151.00 119.00 91.02 1.80 1.83 2. Metanephrine 1 179.94 148.22 165.01 2.03 2.07 3. Normetanephrine 1 166.00 134.02 121.01 2.11 2.15 Column Raptor HILIC-Si (cat.# 9310A52) Dimensions: 50 mm x 2.1 mm ID Particle Size: 2.7 µm Pore Size: 90 Å Temp.: 30 C Sample Diluent: Mobile phase A:mobile phase B (10:90) Conc.: 1 ng/ml Inj. Vol.: 10 µl Mobile Phase A: Water, 100 mm ammonium formate, ph 3.0 B: Acetonitrile Detector Ion Mode: Mode: Time (min) Flow (ml/min) %A %B 0.00 0.3 10 90 5.00 0.3 10 90 MS/MS ESI+ MRM www.restek.com 4

Linearity Using 1/x weighted linear regression, all three analytes showed acceptable linearity with r 2 values of 0.999 or greater, and deviations of <10% (Figure 5). The established limits of quantitation (LOQ) were 0.051 nmol/l (10 pg/ml), 0.14 nmol/l (24 pg/ml), and 0.060 nmol/l (10 pg/ml) for metanephrine, normetanephrine, and 3-methoxytyramine, respectively. Figure 5: Calibration Curves 22.5 Metanephrine y = 0.00602 x + 0.00289 (r² = 0.9996) 20.0 17.5 15.0 Response 12.5 10.0 7.5 5.0 2.5-0.0 Conc -0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000 3250 3500 3750 4000 18.0 16.0 Normetanephrine y = 0.00486 x + 0.00420 (r² = 0.9997) 14.0 12.0 Response 10.0 8.0 6.0 4.0 2.0-0.0 Conc -0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000 3250 3500 3750 4000 25.0 22.5 3-Methoxytyramine 0.00670 x + -0.00012 (r² = 0.9997) 20.0 17.5 Response 15.0 12.5 10.0 7.5 5.0 2.5-0.0 Conc -0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750 3000 3250 3500 3750 4000 5 www.restek.com

Accuracy and Precision Precision and accuracy testing was performed on three different days. Method accuracy was demonstrated by recovery values that were within 5% of the nominal concentration for all QC levels. The %RSD was 0.2 4.5% and 1.1 4.2% for intraday and interday comparisons, respectively, indicating good method precision (Table II). Table II: Accuracy and Precision of QC Samples for the Analysis of Metanephrines in Plasma Analyte Conc. (pg/ml) QC Level 1 (40 pg/ml) QC Level 2 (400 pg/ml) QC Level 3 (2,500 pg/ml) Accuracy (%) %RSD Conc. (pg/ml) Accuracy (%) %RSD Conc. (pg/ml) Accuracy (%) Metanephrine 39.5 98.7 3.39 411 103 1.10 2,480 99.2 2.89 Normetanephrine 39.3 98.2 4.21 401 100 1.92 2,440 97.3 2.34 3-Methoxytyramine 40 100 1.81 407 102 2.07 2,470 98.9 2.02 %RSD Conclusion As demonstrated here, using HILIC for the analysis of metanephrines in plasma is a viable approach that provides excellent method performance at clinically relevant concentrations. The Raptor HILIC-Si column ensures good retention of metanephrine, normetanephrine, and 3-methoxytyramine, which is generally difficult to achieve at low detection limits using reversed-phase chromatography. Accurate results were consistently obtained without matrix interference even at trace levels. With a fast, simple sample preparation procedure and 5-minute analysis time, the established method provides a reliable high-throughput assay for the clinical diagnosis of pheochromocytoma and paraganglioma. Questions about this or any other Restek product? Contact us or your local Restek representative (www.restek.com/contact-us). Restek patents and trademarks are the property of Restek Corporation. (See www.restek.com/patents-trademarks for full list.) Other trademarks in Restek literature or on its website are the property of their respective owners. Restek registered trademarks are registered in the U.S. and may also be registered in other countries. 2018 Restek Corporation. All rights reserved. Printed in the U.S.A. www.restek.com Lit. Cat.# CFAN2801-UNV