Oxygen measurements: sensors accuracy and scientific needs

Similar documents
In-situ calibration of Oxygen Optodes in the Southeast Pacific Oxygen Minimum Zone

Argo-O2 data Data Management Real-time and Delayed-mode QC Where we are/ Where we go

GLIDERS Aanderaa oxygen optode (models 3830 & 3835)

Dissolved oxygen. Bronte Tilbrook for Craig Neill, Kelly Brown and Erik van Ooijen

Observational approaches to oxygen depletion

Sea-Bird Scientific Carol Janzen, Ph.D. Oceanography David Murphy, M.S.E.E. Dan Quittman, M.S.E.E.

Argo Quality Control Manual For dissolved oxygen concentration

Procedures for correcting in situ CTD data and results obtained during the NEAR-GOOS Cross-Basin Climate Monitoring Section project

TNA PROJECT PRELIMINARY REPORT 1 st Call of Proposals 12 January 3 April, 2012

QC for Hydrographic Data GEOSECS to GO-SHIP. Susan M. Becker Scripps Institution of Oceanography Oceanographic Data Facility

Deep SOLO. Nathalie Zilberman, Dean Roemmich, and SIO float lab. 1. Deep SOLO float characteristics. 2. Deep SOLO float Deployment

Argo Quality Control Manual For Biogeochemical Data

A climatology-based quality control procedure for profiling float oxygen data

6 Evaluation of a life time based optode to measure oxygen in aquatic systems

The 2008 North Atlantic Bloom Experiment. Calibration Report #3

STERILIZABLE DISSOLVED OXYGEN SENSORS

Salinity and Dissolved Oxygen (D.O.) Lecture 2

Sensor Stability. Summarized by Breck Owens David Murphy, Sea-Bird Electronics AST-14 Wellington March 18-21

PERSPECTIVES FOR ACCURACY AND QUALITY ASSURANCE OF CTD & BIOGEOCHEMICAL DATA STREAMS FROM AUTONOMOUS PLATFORMS

Accuracy and Stability of Sea-Bird s Argo CTD and Work Towards a Better CTD for Deep Argo. David Murphy Sea-Bird Electronics

Triton DO8 Dissolved Oxygen Sensors ELECTRO-CHEMICAL DEVICES

E-AIMS. R&D on float technology Synthesis D2.6.1

Vertical in situ profiles of nitrate and oxygen in the northern Japan Sea

Deep-Arvor: a CTD & DO profiling float for Argo

Oxygen in the Ocean. 1/28

SBE61 CTD calibration: results from the June 2014 Tangaroa Voyage

CORIOLIS, A FRENCH PROJECT FOR IN SITU OPERATIONAL OCEANOGRAPHY. S. Pouliquen, A. Billant, Y. Desaubies, G. Loaec, F. Gaillard, G.

Stability of Temperature and Conductivity Sensors of Argo Profiling Floats

MAPCO2 Buoy Metadata Report Project Title:

WEBB RESEARCH CORPORATION

Dissolved Oxygen and measurement possibilities. Berno Lüpkes, 15 th March 2017

Preliminary Cruise Report

How well do CMIP5 climate models reproduce Southern Ocean bottom temperature? Model climatology

High-quality oxygen measurements from profiling floats: A promising new techniques

PRODUCT SHEET. Order probe only as RXPROBE02

Instruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text.

Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter

Water Quality Monitoring Using a Handheld Sonde for in situ Measurements Date Prepared: 11/17/2017

LogIT Dissolved Oxygen Probe

MOTUS Wave Buoys. Powered By the Aanderaa MOTUS Directional Wave Sensor

Minutes of the 6th BGC-Argo meeting 27, 28 November 2017, Hamburg. Catherine Schmechtig

DISSOLVED OXYGEN SENSOR BT34i

(Sea)glider data processing the. sampling and thermal lag

STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE

BI-680 Online Dissolved Oxygen Controller Instruction Manual

Payload CTD. Carol D. Janzen 1 Research and Development 1 Sea-Bird Electronics, Inc. Elizabeth L. Creed 2 MaritimeSystems 2 irobot Corporation

Dissolved Oxygen Guide

Development of the Deep Argo Program

OxiQuant S. Operating Instructions

Argo National Data Management Report Italy (2016)

Engineering Note. Algorithms. Overview. Detailed Algorithm Description. NeoFox Calibration and Measurement. Products Affected: NeoFox

A Depletion Compensated Wet Bath Simulator For Calibrating Evidential Breath Alcohol Analyzers

Glossary Published on LEARNZ (

Oxygen Measurement in Natural Gas

12mm Dissolved Oxygen Sensor Care and Use Instructions

Ocean Observatories Initiative (OOI) Moorings: New Capabilities for Seagoing Science

Quality control of data streams at ANFOG

Bioargo ADMT 16 Meeting report Bermudas 2-3 November 2015

Aerobic reoxidation in marine sediments

Lecture 8 questions and answers The Biological Pump

Systematic Validation of Conductivity and Temperature from Ocean moored buoy data in the northern Indian Ocean with in situ ship based measurements

Water Samplers, Water Sampling with Internal Recording, Cabling, and Deployment

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

CTD Data Documentation

FIU Digital Commons. Florida International University. Henry O. Briceño Florida International University,

Temperature, salinity, density, and the oceanic pressure field

OPERATING AND MAINTENANCE MANUAL SBE 23 DISSOLVED OXYGEN SENSOR

Applications of ELCIRC at LNEC

FAQ to Optimare Precision Salinometer (OPS)

(ml/l), (mg/l) or (ppm)

Cascading of Dense Water along the Peter the Great Bay Slope in the northwestern Japan Sea

Air-sea gas transfer velocity for oxygen derived from float data

Methods Used in Measuring Surface Seawater pco 2 Aboard RVIB Nathaniel B. Palmer

WHOTS Mooring Subsurface Instrumentation

NOTES ON CTD/O 2 DATA ACQUISITION AND PROCESSING USING SEA-BIRD HARDWARE AND SOFTWARE (AS AVAILABLE)

Underwater cryotrap - membrane inlet system (CT-MIS) for improved in situ analysis of gases by mass spectrometry.

NEW INSIGHTS INTO THE SOUTHERN OCEAN'S ROLEINCLIMATE: ANUPDATE

ISE-730 Oxygen Electrode and DO2-100 Currentto-Voltage

2003 Horseshoe Beach Lease Area, Dixie County Quality Assurance/Quality Control (QA/QC) Log

PreSens. Manual OxoPlate. OxoPlate OP96C. PreSens Precision Sensing

Oxygen Measurement Systems High Reliability and Wide Application Coverage

CMORE HOE DYLAN October 2012 Draft cruise plan

Introduction The Smart Q Oxygen Adaptor enables an Oxygen electrode to be connected to an EasySense unit.

Kathleen Dohan. Wind-Driven Surface Currents. Earth and Space Research, Seattle, WA

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

Estuarine, Coastal and Shelf Science

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

Application Note AN-107

Aurora Australis Marine Science Cruise AU Oceanographic Field Measurements and Analysis

Oxygen Sensor with Electrode

FLUORESCENCE DETERMINATION OF OXYGEN

The Ocean is a Geophysical Fluid Like the Atmosphere. The Physical Ocean. Yet Not Like the Atmosphere. ATS 760 Global Carbon Cycle The Physical Ocean

The Oxygen Sensor can be connected to all types of einstein Tablets, einstein LabMate, and einstein LabMate +.

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

Argo Float Pressure Offset Adjustment Recommendations. 1: IORGC/JAMSTEC, 2-15, Natsushima-cho, Yokosuka, , Japan

Temperature, Salinity, Dissolved Oxygen and Water Masses of Vietnamese Waters

For Biogeochemistry at ICEX 2011

Evaluation copy. Interdependence of Plants and Animals. computer OBJECTIVES MATERIALS

Vernier Optical DO Probe (Order Code ODO-BTA)

2017 Float and Sensor Workshop. Conveners: Brian King, Steve Riser, Susan Wijffels Local Host: Steve Riser, U. Washington

Transcription:

Oxygen measurements: sensors accuracy and scientific needs L. Coppola (CNRS-INSU), E.Diamond (CNRS-INSU), L.Delauney (IFREMER), D.Hydes (NOC), M.Haller (HZG), J.Karstensen (IFM-GEOMAR), D.Lefevre (CNRS-INSU), D.Machoczek (BSH), R.Nair (OGS), F.Salvetat (IFREMER), S.Sparnocchia (CNR), V.Thierry (IFREMER) Oceanology International 2014 London, Excel, March 13 th 2014

Why are we measuring oxygen? Physical interests: water mass circulation, new formed deep waters, mixing depth, ventilation age, atmosphere-ocean exchange Very sensitive tracer for studying the ocean biogeochemistry response to climate variability and global climate change Biogeochemical interests: primary production estimates, remineralisation flux (consumption/production) The ocean deoxygenation is one of the most important topic (O 2 <20µmol/kg) Oxygen is one of the first measured oceanographic parameters but with a large spatio-temporal scales range : OMZ are poorly documented

A summary of the manufacturer s stated specifications of present O 2 sensors suitable for use on profiling floats We need accuracy around 1 μmol/kg to do some science!

SBE 43 Dissolved Oxygen Sensor Clark polarographic membrane Mainly used on shipboard CTD unit and moorings Accuracy = 2% of saturation Linear drift due to the biofouling issues Oxygen(ml/l) = Soc x (V + Voffset) x φ Possible to correct the drift with O2 Winkler analysis Correction of the SOC term (linear slope scaling coefficient) SOC new = (DOXY Winkler /DOXY SBE43 )

Experiences and problems with SBE43 sensor: Sources of drift: changes in membrane tension, depletion of electrolyte, impairment of the silver anode, plating of anode metal on the cathode, and the presence of chemical contaminants in the sensor s plastic body. Dynamic errors leading to apparent hysteresis are caused by responsetime mismatch of the compensation temperature sensor Membrane fouling: altering the oxygen diffusion rate through the membrane, thus reducing sensitivity. Biofouling can be particularly troublesome because the living organisms either consume or create oxygen. Mostly adapted for CTD profiler (very fast time response)

Experience on field campaign (Mediterranean Sea) Start of cruise 1 week after MOOSE GE 2012 2 weeks after Raw data Adjusted with O2 Winkler Large drift from SBE43 raw data during 17 days cruise (around 50 umol/kg!!) despite the application of the SBE43 cleaning procedure (Triton and bleach flushing) Correction with O 2 Winkler analysis needed! (1 profile per day)

Experience on time series (Mediterranean Sea) Reference = 4.629 ml/l @ 2000m Example @ 2000m depth Example with 3 years using SBE43 during monthly monitoring at the DYFAMED site Large differences between O 2 reference and O 2 from SBE43 after annual calibration

Optical sensor: Aanderaa optode The sensor is based on the dynamic luminescence quenching of an oxygen-sensitive fluorochrome embedded in the tip Long time stability, no pressure hysteresis, fast response, compact, better accuracy (accuracy <8μmol/kg, precision <1μmol/kg) Adapted for Argo floats, gliders, ferry boxes, moorings, plankton incubators (e.g. IODA, RESPIRE, )

Large interest for the Argo community Oxygen sensors have been deployed on ~ 300 Argo floats: today 200 floats are currently operating

What about the optode accuracy? Sensor comparison to gridded data in WOA 09 for 119 floats Drift No drift in Optodes. Some for SBE. 2.36±3.6 (µm/year) 0.01±3.0 (µm/year) (Avg±Stdev) Accuracy Clear calibration offsets. 10.33±11.6 (µm) 16.72±12.5 (µm) Slides from Yui Takeshita & Todd Martz, SIO

Optode: the multi-point calibration Henry Bittig, Arne Körtzinger (IFM-GEOMAR), Riser (CSIRO), D.Lefevre (CNRS) electrochemical O 2 Generator regulated current, flow and temp. triplicate Winkler samples several optodes in sequence polynomial fit in phase and temp. RMSE of fit: 1.5 µmol/l Henry Bittig, IFM-GEOMAR

Argo-O2 advices and calibration improvements Optodes were stable but gave lower values than expected Main reason : an after-curing appears in the foils during the first 1-3 months after manufacturing which typically leads to 1-4 % lower readings Storage recommendations: keep optode wet and dark at all times With the new optodes (4330/4835/4831), new calibration procedure: Use the Stern-Volmer equation (Uchida et al. 2008) Use a 40 multi-point calibration system (5T x 8O 2 points) Accuracy better than 1.5% (± 2 μm)

ARGO-O2 TEST PROCEDURE BEFORE DEPLOYMENT Mixing bath (homogenization) Seawater with chlorine solution 1 profile per day (max depth 18m) Optode 4330 with new coefficients (CSIRO)

Main results Small vertical T/S gradient in the pool Despite the new calibration and the new storage, all optodes underestimated the O 2 concentrations (average 20 µmol/kg!) Large fluctuations during the drift at the parking depth No clear explanations

QUALITY CHECK DATA AT DEPLOYMENT CTDO 2 profile at t 0 with salinity and oxygen in situ measurements (bottles) Salinity and oxygen data corrected from offset Bittig et al. 2012 and Takeshita et al. 2013 methods (C 0 =offset, C 1 =gain) Offset O 2 conc. 10-15 µmol/kg! Float WMO 6901467

OXYGEN STABILITY OVER 17 MONTHS (600 O 2 profiles) LIW signal WMDW signal

Optical sensor: SBE63 Initial accuracy = ±%2 saturation (± 3µmol/kg) Response time < 6s 24-point calibration O 2 signal measured with Stern-Volmer equation O 2 (ml/l) = [ { (a 0 + a 1 T + a 2 V 2 ) / (b 0 + b 1 V) - 1 } / K sv ] [ S Corr ] [ P Corr ] Mainly used on Argo floats (NAVIS) and moorings (mounted on SBE37)

FIELD EXPERIMENTS WITH SBE43 & SBE63

CONCLUSIONS SBE43: perfect for CTD profilers but need O2 Winkler to correct the membrane drift. O2 correction is quite simple Optode 4330: better calibration. Easier to correct (Bittig & Takeshita methods). Still a drift between lab calibration and first deployment! SBE63: seems more stable than SBE43. Adapted for slow profiles and moorings. O2 underestimated. Need better calibration