Thank You for Attending Our June Webinar When Measuring Trace Oxygen is Critical to Your Process Your Host Mike DeLacluyse President Lesman Instrument Co miked@lesman.com Featured Speaker Ken Soleyn Sr Sales Operation Manager: Measurement and Controls Baker Hughes, a GE Company ken.soleyn@bhge.com Follow the Conversation LIVE @Lesman_Inst #LesmanWebinar Providing Solutions to Detect, Measure, Analyze, and Control Your Process and Your Facility.
GE Measurement & Control Oxygen Measurement Ken Soleyn Senior Sales Operations Manager Measurement & Controls Baker Hughes, a GE Company T: +1 603 953 8649 E: ken.soleyn@bhge.com 1100 Technology Park Drive Billerica, MA 01821 USA www.gemeasurement.com 1
Agenda 1. BHGE Oxygen Measurement Technologies A. Paramagnetic B. Electrochemical 2. Instrumentation 3. Sampling Systems 4. Case Studies 5. Choosing the right technology for your application 2
Oxygen is Drawn to a Magnetic Field Paramagnetic One of the properties of oxygen is that it is paramagnetic or drawn to a magnetic field. The picture above shows liquid Oxygen suspended in a magnetic field. 3
Thermoparamagnetic Oxygen (TPM) Sensor Oxygen s paramagnetic property causes it to be attracted to a magnetic field. Heating the gas cases less attraction of Oxygen to the magnetic field. The sensor has a strong magnet and two pairs of heated thermistors. Gas movement causes magnetic wind which cools the wind generating thermistor. The resistance of the thermistors are proportional their temperature. A bridge circuit provides a voltage proportional to the oxygen concentration W2 W1 V O2 RZ R2 R1 V comp W1, W2 = WIND GENERATING THERMISTORS R1, R2 = WIND RECEIVING THERMISTORS 4
Dumbbell Type Paramagnetic O2 Sensor Two nitrogen filled glass spheres are suspended within an electromagnetic field. Oxygen attracted into the magnetic field displaces the nitrogen filled spheres, causing the suspended spheres to rotate. The mount of current it takes to restore the spheres to a null position is related the oxygen concentration. It is controlled by the amount of light received by a photocell. 5
XMO2 TPM Transmitter 4-wire Transmitter, 24 VDC, 1.2A ±1% of span ±2% of span for 0 to 1% range ±0.2% O2 for 90 to 100% and 80 to 100% ranges Explosion Proof: FM & ATEX Rack Mount Analyzer available May be equipped with TMO2D/XDP Power Supply with programmable relays and two line display. 6
XMO2 Response Compressed Air (21%) Compressed Air (21%) Nitrogen (0%) 7
XMO2 Oxygen Analyzer Applications Inerting of Storage Tanks Olefins (Ethylene) Refineries Oxygen Purity Landfill or Bio Gas Ozone Production Flare Gas & Vents Reactor Feed Gases & Catalyst Recovery Wastewater & Sewage Solvent Recovery 8
Inerting Chemical/Fuel Storage The head space of chemical and fuel storage tanks are often purged with an inert gas to assure that the O2 concentration below the LEL (Lower Explosion Level) Ambient air contains 21% O2 The XMO2 is rugged, explosion proof and has no moving parts. 9
Sampling System For Inerting 10
O2 Measurement for Tanker Inerting Alfa Laval/Aalborg/Smit Gas Inerting utilizes the products of combustion for inerting fuel and chemical tankers. Shipboard inert blanketing gas is produced by combusting fuel utilizing seawater to cool chamber wash the product of combustion which is mainly N2 & CO2. The combustion gas is cooled below the dew point to condense water. The dry clean gas is used to blanket fuel can chemical tankers 11
Oxygen Measurement in Vapor Recovery Systems Aeron located in Louisville, KY is an OEM of vapor recovery units To comply with environmental regulations and to yield more available energy whenever a tank (barge or ship) is loaded vapor recovery units are used and O 2 transmitters are used 12
XMO2 for Plastic Resin Production Background: Sabic produces a variety of engineered plastics including Lexan & Noryl. Application: Oxygen control in polymer reactor vessels. %O2 kept under 9% (LEL). Typically maintaining 4%. System sold: XMO2/TMO2D with Sampling system. Sample system includes filter, eductor (to pull sample) and solenoid for N2 (zero) and Air (21% O2) gases. Past experience with XMO2. No moving parts on TPM technology.reliable for process environment Turnkey sampling system Digital interface for automated zero and span calibration (no manual adjustment needed) 13
Oxygen Analyzers for Floating Platform Stainless Steel body transmitters were used for Kaombo Deep Water Offshore Oil Platform. Two FPSO units, each with a treatment capacity of 115,000 barrels per day, will process, store and offload the production. 14
Oxy.IQ - Galvanic Fuel Cell Electrochemical O 2 Sensor Programmable ranges for ppm and percent oxygen LCD Display + Keys Loop Powered. 9 to 28 VDC. 4-20mA signal Programmable Error Logic 4 to 20 ma Field Programmable Ranges Sealed disposable transducer IS for Div I (Zener Barrier), Div II (w/o Zener) Calibrate directly to ambient air (21 % O2) 15
OXY.IQ Response Air 21% O2 Analog: 4-20mA = 0-25% O2 Air Calibration = 21% O2 Air 21% O2 Nitrogen (0% O2) 16
OXY.IQ Response Analog: 4-20mA = 0-1,000 ppmv O2 Air Calibration = 21% O2 Nitrogen (0% O2) 17
OXY.IQ Response Analog: 4-20mA = 0-100 ppmv O2 Nitrogen (0% O2) 18
OXY.IQ Sample Conditioning Systems Assembled on plate or in enclosure (fiberglass or stainless steel) Pressure Regulator Pressure Gauge Needle valves for flow control Rotameter for flow indication Filter Power Supply Vacuum Pump Customized per the application 19
Delta-F Non-Depleting Electrochmical O 2 Sensor Secondary Electrodes Reservoir MOIS TU RE IM AGE SERIES 1 ON OFF 1.3 PPM O2 1 2 3 4 5 6 7 8 9 0 YES NO 4e 1.3V KOH 4OH O 2 Sample Sample Chamber PA NAME TRICS PPM or PPB applications Measurement Anode Stable long term drift free output No false low readings and infrequent calibration Fast response and fast recovery after oxygen shock. Readings independent of sample flow Measurement Cathode Secondary electrodes prevent build up of contaminants in electrolyte Stab-El version resistant to traces of acid gas in the sample Delta-F cells connect directly to the Moisture.IQ Moisture.IQ s built in safety barrier enables the sensor to be used in a Hazardous area 20
OXY.IQ Galvanic Oxygen Transmitter - Applications Ambient Air Monitor Glove Boxes Food & Pharma Processing Welding Shield Gas Metal Processing, Heat Treating & Annealing 21
Arc Welding Arc welding is process that utilizes electricity to produce the weld. A constant-current welding power supply produces electrical energy, which is conducted across the arc through a column of highly ionized gas and metal vapors known as a plasma. The weld area is protected from atmospheric contamination by an inert shielding gas (Argon or Helium). The shield gas must be devoid of oxygen and H2O TIG (Tungsten Inert Gas) MIG (Metal Inert Gas) MAG (Metal Active Gas) 22
Soft Contact Lens Production Location: Rochester, NY Application: For the production of soft contact lenses the Oxygen concentration must be controlled. The manufacturing plant is under FDA regulations strict quality control. Solution: Oxy.IQ Simple compact loop powered transmitter Easy to calibrate and validate Minimal maintenance Just replace the sensor element every 12 months Sensors may be field replaced and calibrated to ambient air (21% O 2 ) 23
OEM for Oxygen & Trace Moisture Transmitters Location: Salisbury, MA. Manufacturer of glove boxes and gas generators Application: O2 & Trace moisture measurement in Glove Box atmospheres & trace moisture. Requirements: Two-wire 4-20mA transmitters that can connect to their data acquisition and control system. 24
Turnkey, Moisture & Oxygen Analyzer + Sample System Moisture.IQ Location: Columbus Ohio Application: Steel Annealing Furnace (N2/H2 mixture) for sheet steel. Turnkey system including XMO2 (TPM) O2 Analyzer) - % O2 OXY.IQ (Fuel Cell) ppm O2 Aurora TDLAS Laser Hygrometer for trace moisture Filter, valves fittings etc. for calibration gas Moisture.IQ - Multifunction analyzer/data logger Aurora XMO2 - % O2 OXY.IQ ppm O2 25
Annealing Furnace Location: Columbus Ohio Application: H2/O2/H2O concentrations in sheet steel annealing furnace Turnkey analyzer with sample system XMTC, Trace Oxygen & Trace Moisture measurement. Tied to plant wide monitoring system 26
Specifications for Industrial Gases Argon Purity 99.997% Oxygen Less than 0.0005% Dew Point Liquid Argon: -90ºF or lower Helium Purity Combined total of the following impurities Dew Point Hydrogen Liquid 99.995% by volume Argon, carbon dioxide, carbon monoxide, hydrogen, hydrocarbons, oxygen, neon, and Liquid nitrogen Helium: shall -90ºF not or exceed lower 50 parts per million by volume Purity 99.999% Oxygen Less than 0.0002% Dew Point Liquid Hydrogen -105ºF or Hydrogen Gaseous lower Both gas suppliers and gas consumers monitor gas quality Purity 99.95% Oxygen Less than 0.0005% Dew Point -90ºF or lower Nitrogen Nitrogen and trace 99.998% 27 inerts Oxygen Less than 0.001% Dew Point Liquid Nitrogen -90ºF or lower
Industrial Gases Distribution & Use Point of Use Generation & Storage Distribution, Mixing & Metering In the distribution of gas through a facility there are many possible point of O2 or moisture ingress 28
Choosing the Right Technology Technology Range Speed Accuracy Industrial Pkg & XP Trace O2 O2 Purity Cost TPM B B B A D A B- Fuel Cell B+ C B- B- B- D+ A ND Fuel Cell C A B C A- A C+ 1 GENERAL 2 INFORMATION 3 4 PROCESS CONDITIONS 5 6 7 8 9 10 11 12 13 14 15 16 17 24 25 26 27 28 29 30 31 32 Company Name Contact Name Address Phone Component 3 Units Component 4 Units Component 5 Units Measurement Range Min/Nom/Max Units Process Temperature Min/Nom/Max Units Process Pressure Min/Nom/Max Units Sample Return Pressure Min/Nom/Max Units Available Flow Rate Units 18 Ambient Temperature Min/Nom/Max Units 19 Hazardous Area Classification SENSOR / 20 Location Environmental TRANSMITTER 21 Sample System Enclosure Requiremen INSTALLATION 22 Power available 23 Utilities available (Select all that apply) DISPLAY / CONTROLLER INSTALLATION Display Hazardous Area Classification Display Housing Requirement Location Environmental Ambient Temperature Min/Nom/Max Power available Alarm contacts Email Units Required Cable Length to Sensor/Transi Units Utilities available (Select all that apply) Oxygen / Thermal Conductivity Application Data Sheet Highlighed Row = Mandatory Field (Rev 1.3) Tag number Process Name Component to be Measured If Other: Process Composition (include contaminants Units Minimum Nominal Maximum Component 1 Component 2 Units Units Length: Application Gas Compositions Contaminants (Acids Gases) Utilities Power Environmental Accuracy & Precision Maintenance Range of interest Sample system requirements 29
Thank You for your attention Questions? Contact Information Ken Soleyn Senior Sales Operations Manager Measurement & Controls Baker Hughes, a GE Company T: +1 603 953 8649 E: ken.soleyn@bhge.com 1100 Technology Park Drive Billerica, MA 01821 USA www.gemeasurement.com 30