AQUADEX FLEXFLOW SYSTEM QUICK REFERENCE GUIDE

Similar documents
AQUARIUS Continuous Renal Replacement Therapy with Regional Citrate Anticoagulation (RCA) Competency

Contacts. Quick Start Guide

Quiz for Module 2: Aquarius Training (Rev. 1) Lesson 1: Introduction to Aquarius

AQUARIUS Continuous Renal Replacement Therapy Competency

Module 2: Aquarius Training

Indications for Use: Caution: Note:

Troubleshooting Guide Aquarius

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax

Liberty Cycler Handi-Guide

patient education program 4156 South 52 nd Street, Omaha, NE ChildrensOmaha.org/HomeHealthcare

Nimbus II PainPRO Ambulatory Infusion Pump

MANITOBA RENAL PROGRAM

CRRT with Prismaflex LEADS TO More Flexibility, Ease of Use and Safety

Liberty Handi-Guide. A Patient s Reference to the Most Commonly Asked Questions About the Liberty Cycler

HAEMOFILTRATION GUIDELINES

Negative Pressure Wound Therapy

Liberty Handi-Guide A Patient s Reference to the Most Commonly Asked Questions About the Liberty Cycler

Nimbus II PainPRO Ambulatory Infusion Pump

- Repeat calibration until desired results are reached otherwise set SW2 and SW1-8 open.

Blood Pressure Monitoring: Arterial Line

Liberty PDx Cycler Handi-Guide

2008T Hemodialysis Machine with CLiC Device Training Guide

WARNING: All medications should be closely monitored to detect any alterations in the effective concentration due to the ultrafiltration process.

IV IV Training Guide

Introduction. Medley TM Medication Safety System. Alaris Medical Systems

Alaris Pump module FAQs

EASYPUMP II THE FLEXIBLE SOLUTION FOR SHORT AND LONG TERM INFUSION THERAPIES

A few seconds after turning on, the screen changes to Set Up Screen Before adjusting the pressure, calibrate the O 2 sensors

MANITOBA RENAL PROGRAM

Handi-Guide A Patient s Reference to the Most Commonly Asked Questions About the Liberty Select Cycler

INTERNATIONAL MARKETING. 61, avenue Tony Garnier Lyon France Tel.: 33 (0) Fax: 33 (0)

Shared Haemodialysis Care Handbook. Name Hospital number: Shared HD Care Nurse: Date: Machine type: Dialysis Unit:

Chapter 40. Fluid, Electrolyte, and Acid Base Balance. Procedures Checklist INTRAVENOUS THERAPY. Procedure 40.1: Starting an Intravenous Infusion

Xcela Hybrid PICC. Clinician Reference Tool

CADD-Micro. Model 5900 Operator s Manual. Note: Colors are Black and Cyan (as second color) Deltec. Ambulatory Infusion Pump

New. The high security connection

Astral in AirView: Improving patient care through connectivity. ResMed.com

FAULT CODE TROUBLESHOOTING INDEX

Liberty Cycler User s Guide

Lucid System In-Service Training

Making Life Easier Easypump II

MyoSure Setup Simplified

Astral in AirView: Improving patient care through connectivity. ResMed.com

*smith&nephew RENASYS GO Negative Pressure Wound

Heart/Lung Perfusion Packs INSTRUCTIONS FOR USE

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

CircuFlow Quick Setup Guide

NEXUS TKO. Specialty Neonatal/Pediatric. Safety Infusion Sets. Blood Reflux Protection 24/7! Anti-Reflux Technology Provides

PRE-TRANSFUSION GUIDELINES

1.060 Set-Up and Priming of the Bypass Circuit. Perfusion Technology Department. Perfusionists

S/5 Aespire. Quick Reference Guide

STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

Quick Reference Guide

Alfred ECMO Circuit Priming and Storage Guide

Liberty Select Cycler User s Guide

BODYGUARD 2CH TWIN CHANNEL INFUSION PUMP

Liberty PDx Cycler User s Guide

Operation and Maintenance of the EPV200 Portable Ventilator

Nita Newborn Model 1800 User Manual

extricare 3600 Operating Manual

6800 Maintenance Instruction System Flush Procedure

User Manual. For Patients and Clinicians

6900 Maintenance Instruction System Flush

Declining Weight Blending System Quick Start Manual E Main St Marshall, IL Phone: Fax: Website:

HEALTH SERVICES CODE A.6 CATEGORY: RN - SNP

Overview. The user must follow the instructions below.

Hemoglobin Reference Failure Troubleshooting

Assisting with Insertion. Care of Intraspinal Catheters

HemoCue Hb Procedure Template

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

Instruction Manual for Configura Cushionair Portable Pump

Varithena Delivery System Instructions for Use

Technical Support and Customer Service United States: NIOV is a trademark of Breathe Technologies, Inc.

Blood Parameter Monitoring System 550

Cardiac Catheterisation Instrumentation. Lancashire & South Cumbria Cardiac Network

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT

Operation Manual. H & R Healthcare 1750 Oak Street Lakewood, NJ Phone: Fax:


HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

Operation and Maintenance of the MCV100 and MCV100-B Portable Ventilator

Redesign - ATMOS S 201 Thorax. Digital thoracic drainage systems. ATMOS Thorax MOBILITY SAFETY COMFORT FLEXIBILITY

Complex Care Hub Manual: Pressure Ventilation on the LTV 1200

OPERATION MANUAL BFX-2

How does the paramedics work day look like? Have a look at our field research video.

Gas Diluter Portable Gas Dilution System User s Manual

OPERATOR S MANUAL. P/N GB Rev. 008

Bloodlines, Tubing Systems, Accessories

Skills & Scenarios 81

AUTOVENT 4000 VENTILATOR

QUICK REFERENCE GUIDE

My CoughAssist. A patient guide to CoughAssist T70. Please visit

Instructions for Use. OptiLife Nasal Lab Mask

Low-Flow Purging Procedure with MicroPurge basics Equipment. MP10 Bladder Pump Controller MP15 Control. MP30 Drawdown Meter.

Recommendations on Checking Anaesthesia Delivery Systems

OPERATOR S MANUAL HOTLINE 3 Blood and Fluid Warmer

Working version subject to Departmental approval. Drafted by Dr. Olympio and reviewed by clinical members of The Capital Equipment Committee

Recommendations for Standards. Monitoring. Cardiopulmonary Bypass

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

Medical Instruments in the Developing World

WATER CONTROL SYSTEM QUICK START

Transcription:

AQUADEX FLEXFLOW SYSTEM QUICK REFERENCE GUIDE

AQUADEX FLEXFLOW SYSTEM DISCLAIMER The quick reference guide is not intended to replace CHF Solutions, Inc. Aquadex FlexFlow Direction For Use (DFU). Always follow the appropriate and applicable local, institutional, and/or hospital policies, procedures, and requirements. PURPOSE OF QUICK REFERENCE GUIDE Provides an overview on workflow and setup of Aquadex FlexFlow system. ONLY INDICATION: The Aquadex FlexFlow System is indicated for temporary (up to 8 hours) ultrafiltration treatment of patients with fluid overload who have failed diuretic therapy; and extended (longer than 8 hours) ultrafiltration treatment of patients with fluid overload who have failed diuretic therapy and require hospitalization. All treatments must be administered by a healthcare provider, under physician prescription, both of whom having received training in extracorporeal therapies. 1

AQUADEX FLEXFLOW SYSTEM LIST OF SUGGESTED ITEMS Aquadex FlexFlow Console UF 500 Blood Circuit Set 500cc NaCl bag Preloaded NaCl syringes Empty 10 12cc syringe Alcohol preps Extra Luer caps Graduated cylinder KEY CONCEPTS FOR AQUADEX FLEXFLOW SYSTEM THERAPY The following must be achieved, confirmed, and maintained throughout therapy: Catheter patency Recommended blood flow (20-40 ml/min) Therapeutic levels of anticoagulation WARNING It is important for the clinician to balance the risks of anticoagulation with the potential benefit when deciding to use anticoagulation therapy (for example, heparin) in conjunction with Aquadex FlexFlow system therapy. 2 SETUP

PRE-THERAPY PREPARATION 1 Ensure anticoagulation has begun (minimum 30 minutes), per physician order 2 Gather the required items listed 3 Check for venous access patency using 10-in-10 test: Ensure Withdrawal of 10ml over 10 seconds then re-infuse in 10 seconds without resistance (10-in-10 patency test) Aspirate and flush each catheter port If resistance is encountered, consider the following catheter corrections, such as: Re-position catheter, per hospital protocol Contact PICC team or Interventional Radiology as needed WARNING Do not use heparin for anticoagulation in patients with intolerance or allergy to heparin. 3

PRE-THERAPY PREPARATION 4 Load and Prime the Circuit: a. Press PRIME then HELP b. Follow prompts on screen to load the UF 500 Blood Set c. On Prime mode screen, press ACCEPT to enter Prime mode d. Follow prompts on screen during Prime e. To prime access ports, use an empty syringe to withdraw saline from the Withdrawal (blue) access port and use the same syringe to infuse saline into the Infusion (clear) access port WARNING To avoid risk of introducing air into the patient, do not connect the patient to the circuit until priming is complete. 4 OPERATION

INITIATING THERAPY 1 Connect lines to venous access catheter/lumen(s): Disconnect Withdraw (blue) line from priming bag and connect to catheter Disconnect Infusion (clear) line from UF bag and connect to catheter Open all clamps No caps or plugs HUB to HUB connection 2 UF BLOOD Set and Rate, then Press RATE FLOW ACCEPT 3 Press the RUN key 4 5 6 Determine whether automatic Hematocrit monitoring will be enabled on console Monitor Venous Access: Check Pw and Pi pressures and frequency of Occlusions / Disconnects Consider adjusting patient/catheter position in order to maximize blood flow Setting blood flow lower than 20 ml/min is not recommended Monitor the patient per hospital standing orders and per the DFU 5

THERAPY COMPLETION 1 To return blood (35 ml with Hct and 33mL without), connect blue Withdrawal line to saline and press and hold Manual button until blood is returned to patient 2 Close clamps and disconnect blood set from venous access/catheter(s) 3 Remove blood set from device 4 Care for catheter and dispose of set according to institutional policies 6

PRESSURE READINGS: ACCESS Pw WITHDRAWAL PRESSURE: Measures patency of the Withdraw (blue) side of the set Negative pressures -20 to -300 mmhg (greater negative number means greater resistance) If Pw trends in a more negative direction, consider flushing the access ports and/or re-positioning patient Pi INFUSION PRESSURE: Measures patency of the Infusion (clear) side of the circuit Positive pressures +20 to +300 mmhg (greater number means greater resistance) To view graph from the main screen, use the down arrow until the Access status bar is highlighted 7

PRESSURE READINGS: FILTER Pu ULTRAFILTRATE PRESSURE: Measures patency of the filter Pressure range: -250 to +200 mmhg (greater negative number indicates potential clot) FILTER LIFE: If the blood pump has been stopped for 4-5 minutes and the filter resistance is greater than 1.0, consider re-priming Blood Set FILTER STATUS: Measures resistance of the filter If resistance is trending upward towards 1.0, consider re-priming filter, checking that anticoagulation level is appropriate, reducing UF rate Greater than 1.5, may indicate a clotted filter. Consider re-priming or change out filter To view graph from the main screen, use the down arrow until the Filter status bar is highlighted 8

ADDITIONAL TIPS PATIENT SHOWING CLINICAL SIGNS OF HYPOVOLEMIA: These could include: decrease urine output, hypotension, tachycardia Contact prescriber Consider setting UF rate to 0 while awaiting prescriber response REPORTS AND SYSTEM EVENTS HISTORY: Pres the MENU key Choose Alarms and Events for a history of recent events Choose Measurements for a history of settings and pressure readings (logged every 20 minutes) AMBULATING THE PATIENT: Disconnect power cord from outlet Console can operate for no more than 30 minutes on battery power alone Device will alert press Ensure stability of catheter and lines CLEAR Hct measurement may be affected by patient movement WARNING Ignoring and/or indiscriminately pressing the CLEAR key as a response to an alarm may result in serious patient injury or death. Always identify and solve the originating cause of an alarm before pressing the CLEAR key. 9

HEMATOCRIT (HCT) SENSOR OPTIONAL IF HEMATOCRIT SENSING IS NOT DESIRED: Leave the Hematocrit sensor on the dock on the back of the machine When prompted with Confirm Hct System Disabled at start-up, press the ACCEPT key ENABLING HEMATOCRIT SENSING: Move the Hematocrit sensor from the dock to the blood chamber cuvette Set the Hct limit by pressing the Hct button and using the arrows to adjust Upon startup, Hct sensor will automatically establish the patient s baseline Hct Once the automatic baseline process is complete, Hct level will be displayed on the main screen HEMATOCRIT MONITORING: To view graph from the main screen, press the down arrow until the Hct screen is highlighted When Hct limit is reached, the UF pump will stop without stopping the blood pump If Hct limit sustained, system will alert WARNING The setting of a Hct Limit is an adjunct to standard medical practice and is not a replacement. It has the ability to prevent excessive volume depletions and allow the medical practitioner to provide an additional level of safety while treating fluid overloaded patients. 10

COMMON ALARMS AND ALERTS ALARM/ALERT DESCRIPTION POSSIBLE REMEDY The device is beeping ALARM (red LED, audible alarm sounds once per second) ALERT (yellow LED, audible alarm sounds once every two seconds) Press the Alarm Silence key (Silence for 2 minutes) Follow prompts on screen for troubleshooting alarms/alerts Press the CLEAR key when ready to advance If pump was stopped, press run to continue Withdrawal Line Occlusion or Infusion Line Occlusion The withdrawal line or infusion line pressure is higher than expected Consider position of patient, catheter and lines Consider flushing catheters Consider switching the lumens (Infusion and Withdrawal), per institutional policy Consider repositioning catheter, per institutional policy If persistent, reduce blood flow rate in increments of 5 ml/min until consistent blood flow is maintained Withdrawal Line Disconnect or Infusion Line Disconnect The withdrawal line pressure is lower than expected Infusion line is not connected to the ultrafiltrate bag during Prime Mode The console did not see the expected ultrafiltrate bag weight increase during Prime Mode Consider connection to catheter Consider size of catheter resistance may be too low (consider adding resistance) Ensure blood flow rate is at maximum of 40 ml/min Press CLEAR and then RUN to continue UF Bag Full The ultrafiltrate bag is full. If in Run Mode, ultrafiltrate pump stops but blood pump continues operation to prevent clotting The amount of fluid removed is approximate Drain bag completely into graduated cylinder or urinal and discard Ensure the bag drain is fully closed Press CLEAR to continue Ultrafiltrate Bag Weight Mismatch The fluid measured by the weight scale does not match the amount expected to be delivered by the ultrafiltrate pump The amount of fluid removed is approximate If the alarm persists and the ultrafiltrate bag full alarm does not occur at 1.1 liters contact customer service Repeated occurrence of this alert can result in the Excessive Weight Mismatch alert Can occur if fluid collection bag is emptied prematurely Ensure bag is hanging freely on weight scale hook, and the bag drain is fully closed Press CLEAR to continue 11

COMMON ALARMS AND ALERTS ALARM/ALERT DESCRIPTION POSSIBLE REMEDY Air Detected Excessive Withdrawal Pressure or Excessive Infusion Pressure Excessive UF Pressure Unexpected Pressure Difference Air detected in the circuit during Run Mode Blood and ultrafiltration pumps stop, system enters Stop Mode System detected out of range withdrawal or infusion pressure Blood and ultrafiltrate pumps stop, system enters stop mode System detected out of range ultrafiltrate pressure Blood and ultrafiltrate pumps stop, system enters stop mode An unexpectedly high pressure difference over 100 mmhg is detected between the withdrawal and infusion pressures at the beginning of Run or Manual Mode PROMPTLY assess amount of air in the air detector Press the CLEAR key Press and hold the MANUAL key to advance the air past the detector For large air bubbles, continue to advance the air bubble until it reaches the Infusion access port and remove with a syringe Press RUN to continue If obstruction due to compressed vein, re-position patient and catheter position If obstruction due to clot, flush catheter/lumen per institutional policy Disconnect line to relieve pressure and then reconnect Check the line from the filter to the UF bag for any kinks or obstructions Flush catheter/lumen per institutional policy Disconnect line to relieve pressure and then reconnect Ensure ALL clamps are open Consider patient and catheter position Verify anticoagulation is at appropriate level Flush catheter/lumen per institutional policy Manually turn the pumps by hand to achieve desired pressure levels Motion Detected The console is having difficulty measuring pressures Press CLEAR Restrict patient movement Press RUN to continue 12 TROUBLESHOOTING

COMMON ALARMS AND ALERTS ALARM/ALERT DESCRIPTION POSSIBLE REMEDY Circuit Life Exceeded / Expired Circuit Clotted Large increase in Hematocrit (Hct) The indicated life of the circuit has been exceeded Less than 10 minutes remain in current circuit life before therapy using the current circuit will be disabled The system has detected a higher than expected infusion pressure and an inadequate blood flow Blood and ultrafiltration pumps stop, system enters stop mode The current patient Hct reading is significantly above the HCT limit Ultrafiltration has stopped Prepare and replace with a new blood set if treatment is to continue Avoid prolonged or frequent periods in stop mode Verify appropriate anticoagulation levels Flush catheter/lumen per institutional policy Consider re-priming the circuit (see pg 15) Replace the circuit if necessary Ensure that Hct sensor is properly connected to blood chamber cuvette Limit excessive patient movement Clean the Hct sensor of any dust or debris If persistent, consider lowering UF rate (per prescriber s order), raising Hct limit or disabling feature Remove air bubble or saline from blood chamber Hct Limit Not Set Hct No Blood Detected The Hct limit has not been set, ultrafiltration is prevented The HCT sensor clip is not currently detecting blood in the blood chamber Press CLEAR, then set the Hct limit using the Hct key If disabling, place the Hct sensor on the dock Remove Hct sensor from blood chamber and reconnect Remove any air bubble or saline from blood chamber using the MANUAL key Please refer to additional warnings, alarms, and alerts listed in the DFU. 13

RE-PRIMING THE CIRCUIT CONSIDER RE-PRIMING THE CIRCUIT IF: Device is in Stop mode for more than a few minutes Pu is trending negatively or Filter resistance is reaching critical range Circuit Clotted alarm occurs 14

RE-PRIMING STEPS 1 2 3 Empty and document current amount of fluid in UF bag Pause anticoagulation Infusion Disconnect Withdrawal (blue) line and connect to saline bag using spike 4 [OPTIONAL] Press and hold the MANUAL key to return blood to patient 5 6 7 8 9 Disconnect Infusion (clear) line and connect to UF bag Press PRIME, then ACCEPT and follow prompts on screen Press PRIME and ACCEPT for a second prime cycle Reconnect Withdrawal and Infusion line to patient catheters Resume anticoagulation 10 Press RUN 15

CONTACT INFORMATION For more information, refer to the Aquadex FlexFlow User s Guide. For customer and technical support, please call 1-855-786-2778. CHF Solutions, Inc. 12988 Valley View Road Eden Prairie, MN 55344 www.chf-solutions.com 16 CONTACT

2018 CHF Solutions, Inc. All Rights Reserved. 16-05629-D