APPLICATION NOTE. GC Integrated Permeation Device

Similar documents
Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Generating Calibration Gas Standards

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

GCMSD-Headspace Analysis SOP

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER

Development of DMS and Acetonitrile, and Formaldehyde gas standards. Gwi Suk Heo, Yong Doo Kim, Mi-Eon Kim, and Hyunjin Jin

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Application Note AN-107

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Drilling Efficiency Utilizing Coriolis Flow Technology

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

Multiple Gas#5 GC configuration Jan 2016

The Preparation of Low Concentration Hydrogen Sulfide Standards Gulf Coast Conference September 12, 2002 Galveston, TX Paper 050,12:45 pm Moody F

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

Automated Determination of Dissolved Gases in Water Anne Jurek

Specialty Gases & Equipment ISO 9001: 2008 CERTIFIED. A Division of Norco, Inc.

Best Practice for Identifying Leaks in GC and GC/MS Systems

VIC offers a variety of Calibrated Gas Leaks

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

Nonrefillable Cylinders

Thermo Scientific Model 146i Principle of Operation

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module

Agilent 490 Micro GC Natural Gas Analyzer

Elimination of Percent Level H 2 S Calibration Gas from Flare Gas Monitoring Systems Measuring Total Sulfur, H 2 S and BTU.

Gas Clean. Filters. Delivering Clean Gases for GC and GC/MS Operation

Vehicle and Engine Emissions Testing

Validation of Custody Transfer Metering Skid at Site After Laboratory Proving

Experiment GC : Analysis of BTEX by GC-FID

HiQ laboratory gas generators.

Ultimate Performance in Gas Chromatography

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

Recent Advances to Ensure Simple, Leak Free GC Column Connections. Ken Lynam R&D/Application Chemist Agilent Consumables and Supplies Division

CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV

This test shall be carried out on all vehicles equipped with open type traction batteries.

800.Keen.Gas

Gas Chromatographic Applications of the Dielectric Barrier Discharge Detector

White Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology:

Gas mixtures. Individual solutions specifically for your application

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

ASTM UPDATE to COQA. COQA Meeting Feb San Antonio. RJ (Bob) Falkiner SR (Steve) David GJ (Jay) Grills W (Wayne) Kriel

ENVIRONMENTAL AND POLLUTANTS GAS ANALYZERS

Organic Elemental Analysis Series II Nitrogen Analyzer. proven performance. reliable results

Keeping It Simple. Gas Mixtures for Stack Emissions Monitoring.

LABORATORY LINE ON SITE GAS GENERATORS

The analysis of complex multicomponent

HIGH ACCURACY MULTI-GAS MONITORING USING AUTOMATED SELF- CALIBRATION

LUMAZOTE T ECHNICAL BRO CHU RE

Applied Technology and Best Practices in CEE. Conference

Advancements in Gas Chromatography Analyzers - Keeping up with New Technology. Chuck Runkle Gas Phase Product Specialist ASTS June 2013

STD-3-V1M4_1.7.1_AND_ /15/2015 page 1 of 6. TNI Standard. EL-V1M4 Sections and September 2015

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1

Vapor Intrusion. Sampling Solutions VAPOR INTRUSION. Sampling Solutions

Residual Gas Analysis Systems for Industry

Rapid and Reliable Detection of Dissolved Gases in Water

Manual. Online Degasser DDG-75. Manual Degasser DDG-75 April DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE Hockenheim Germany

SAFE AND ENVIRONMENTALLY FRIENDLY

Carrier Gases in Capillary GC

Experiment 18 Properties of Gases

Finding leaks in Main Cryogenic Heat Exchangers without shutdown

System Overview. TCD Detector temperature setpoint Regulator

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

DSC 204 HP Phoenix. High-Pressure Differential Scanning Calorimetry up to 15 MPa Method, Technique and Applications

MV5: Automated concentration system

Real-time Analysis of Industrial Gases with Online Near- Infrared Spectroscopy

Pharmaceutical Grade Nitrogen & Gardner Denver / CompAir PSA Nitrogen Gas Generation Systems

Air Monitoring Directive Chapter 7: Calibration

The Gold Standard in Calibration Mixtures

TECHNIQUES OF COMPOSITE SAMPLING. Kris Kimmel. YZ Systems, Inc Pollok Drive Conroe, Texas 77303

DEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111

Organic Elemental Analysis Series II CHNS/O Elemental Analyzer. proven performance. reliable results

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310


CHE 4115 Chemical Processes Laboratory 2 Experiment 1. Batch Distillation

A Journal of Practical and Useful Vacuum Technology. By Phil Danielson

11/22/ (4) Harmonization: <846> SPECIFIC SURFACE AREA

S on equipment and pipelines along with the constant threat to personnel safety justify major investment in H 2

Background Statement for SEMI Draft document 4657B NEW STANDARD: SPECIFICATIONS FOR TUNGSTEN HEXAFLUORIDE (WF 6 )

no peaks missing peaks too small peaks increasing retention decreasing retention

GPP model GPPT model GPC model CLP model

VICI Metronics Dynacalibrator Model 150 Instruction Manual

Operation of the Perkin Elmer TGA-GC/MS

Both physical and chemical effects come into play either separately or in combination.

May 1, By John Heim,Doug Staples

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

Sorbent Pen Heavy & Labile SVOC s Taints, Off-Flavors, Odors

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay

The Tekran Series 2600-IVS Difference!

Calibration of a GC955

Laser Spectrometers for Online Moisture Measurement in Natural Gas. Ken Soleyn GE M&C

Leaks, contamination and closed loop control how RGAs make coating processes more profitable

During today's webinar we're going to look at the following topics:

Detecting Hydrogen in Helium Streams. Samuel Goodman. Pittsford Mendon High School. Rochester, New York. Advisor: Dr.

Transcription:

GC Integrated Permeation Device GC-integrated Design PPB to PPM Level Calibration No Cylinders or Regulators Required Option for Dual Ovens Cost Effective, Safe, Clean, Flexible, NIST Traceable Keywords: Permeation Device, Sulfur Traces, Low Level Calibration INTRODUCTION Obtaining a reliable and accurate way of calibrating analyzers at low levels is often a challenge, particularly when the components of interest are reactive or tend to cling to contact surfaces (e.g. H 2 S). Most of the times pressurized gas cylinders are used to deliver the required gas compositions but the use of gas cylinders have some distinct disadvantages like: high pressures, unstable/inaccurate composition, use of lab space, difficulties with shipment and delivery. Permeation devices provide an excellent method of producing known gas concentrations in the PPM and PPB level for calibration of analytical instrumentation, overcoming the disadvantages of gas cylinders. Figure 1. Schematic of a Permeation Tube A permeation device is a device based around a permeation tube which is contained within a highly accurate thermally controlled chamber and flushed by a stable accurate flow of inert gas. A permeation tube is a sealed permeable membrane containing a chemical component, typically in liquid form, that permeates through the walls of the membrane at a constant rate (see Figure 1). This constant rate is referred to as the permeation rate. The permeation tube is maintained at a constant temperature inside a permeation chamber to establish constant vapor pressure inside the device. This results in an equilibrium between the liquid and the vapor phase of the chemical component. The vapor escapes through the walls of the permeable membrane at a constant rate as long as the set point temperature is maintained. A stable known flow of an inert gas such as nitrogen, helium, argon etc. is passed through the permeation chamber to mix with the vapor resulting in a known volumetric concentration in ppm/ppb. By varying the dilution flow rate one is able to generate different concentrations using a single tube. Figure 2. Examples of Permeation Tubes

AC PERMEATION DEVICE CONFIGURATION As mentioned the AC permeation device is integrated into AC Gas Chromatography Solutions, offering the following advantages: Small footprint Automated calibration at ppm and ppb level Automated variation of concentration over 1 decade by permeation gas flow variation Permeation gas can be created in different matrices Option to use 2 permeation ovens 3 tubes per permeation oven Optimal Gas composition stability Reduced risk Figure 3. AC Integrated Permeation Device The AC GC-integrated permeation device can be configured with either 1 or 2 permeation ovens. Using 2 permeation ovens provides the possibility to calibrate components that are not chemically compatible or to calibrate components that have certified permeation rates using different temperatures. REPEATABILITY To demonstrate the stability of permeation gas, 3 permeation tubes (H2S, COS and DMS) were inserted into an AC permeation device that was fitted onto an Agilent 7890 GC (see figure 3). The content of the 3 sulfur components in the permeation gas was measured over a period of 4 days (200 analyses), displayed in figure 5. RSD on peak area over this period stayed well within 1.5 %, even for H2S (also see table 1). Figure 4. Chromatogram of Permeation Gas @ 5-7 ppm each on 7890GC with Antek 7090 SCD

Figure 5 with Table 1 included. Permeation Gas Stability over 4 Days Component Conc. (ppm V/V) RSD (%) H2S 7.1 1.2 COS 5.4 1.3 DMS 5 1.5 ACCURACY To validate the accuracy of the GC system equiped with an AC permeation device, a certified reference gas (cylinder) composition was analyzed using the system calibrated with the dual oven permeation device. Certified values on cylinder were around 11 ppm for each component. Permeation Flow was varied between 20 and 200 ml/min to establish bias at both flows. The relative bias between the determined value and the certified target value of this cylinder gas remained within an acceptable window of 10% for all components for both permeation flows (20 ml/min / 200 ml/min, see table 3). Component 20 ml/min 200 ml/min.. (ppm v/v) % (ppm v/v) % H2S 1.1 9.7 0.6 5 COS -0.2 1.4-0.1-0.8 DMS 0.8 7.2 0.5 4.1 MeSH 0.3 2.2 0-0.3 EtSH 0.1 0.6 0.1 0.5 Table 2. Accuracy at 20 and 200 ml/min Permeation Flows. measured at target value of approximately 11 ppm v/v each component, expressed as Measured Value Cylinder Measured Value Calibration Device

SYSTEM INERTNESS One of the main challenges chemists face when working with gas cylinders is creating a decent and realiable calibration curve. Multiple point calibration curves require several cylinders, that often prove unavailable at the desired low concentrations, or a dilution system between a single source calibration gas cylinder and the analyzer, both making rather inflexible solutions. Also the inertness of the sample path will prove extremely critical when creating a calibration curve. Poorly deactivated lines or connections will negatively impact Linear Dynamic Range and lower detection limits. All lines in AC Permeation Device are deactivated specifically for calibrating and analyzing active components at low concetration levels. Furthermore, the flexible permeation flow and dual oven design allow for making a true multipoint calibration curve from up to two sets of 3 of permeation tubes, even at ppb levels of concentration: Below chromatogram represents a CO 2 sample spiked with trace levels sulphur (5 components) at approximately 30 ppb per component level. Peak Elution order is as per table 3. Table 3 summarizes data from 3 runs, mean value measured and repeatability (N=3). Identifying the 5 sulfur component peaks is easy and accuracy/repeatability of quantification at this low level are great. LOD levels are typical. Figure 6. Low concentration Sulfur components in CO 2 : Chromatogram Component Run 1 Run 2 Run 3 Mean (n=3) RSD % (ppb V/V) (ppb V/V) (ppb V/V) (ppb V/V) (n=3) H2S 29 28 28 28 2.6 COS 36 24 31 30 19.2 MeSH 26 28 23 26 9.1 EtSH 29 29 25 28 8.1 DMS 27 24 31 27 13.6 Table 3. Low concentration Sulfur components in CO 2 : Repeatability and accuracy

OTHER PERMEATION DEVICE BENEFITS NIST TRACEABLE, STABLE GAS COMPOSITIONS The permeation rates for permeation tubes can be certified using standards traceable to NIST by the most basic and accurate laboratory procedure measuring the gravimetric weight loss over a known period of time at a known temperature. Ccomponents in gas cylinders may react with each other or adsorb onto the cylinder walls causing the standard to become inaccurate and unreliable. Improper pressure regulators may aggrevate this and impact the calibration gas composition. When using permeation devices the mixture is continuously created which leaves minimal time for reaction and the sample flow path is continuously flushed which deactivates all contact surfaces in the sample flow path. REDUCING RISK Gas cylinders are big and heavy. They take up a lot of space (which often limits the number of standards in inventory), and shipment is slow and expensive. The AC permeation device is fully integrated into the GC and as such doesn t need to be handled and doesn t use any additional lab space. You can also forget about disposal fees and monthly rental charges. Pressurized Gas cylinders can be a potential danger a leak or broken connection can cause a sudden release of a large volume of toxic, corrosive, odorous, or otherwise hazardous gas. Permeation devices contain low amounts of the pure substance, and are permanently sealed, virtually unbreakable, and safe to handle. The permeation device offered by AC is also protected against overheating which further limits risk. HIGH FLEXIBILITY, EASY DELIVERY A Permeation Device may be incorporated in many types of analyzers or even be used to accommodate two different channels of analysis in one GC. Hundreds of components are available in the form of permeation tubes, and rates for new components can easily be certified using NIST-traceable standards. Shipment of gas cylinders to some countries is almost impossible and local supplies may not always provide the desired quality. Because of the limited quantity of chemicals used and its small weight and size, shipment of permeation tubes is much more convenient than shipment of gas cylinders.

CONCLUSION The AC Integrated Permeation Device offers a powerful alternative to the use of certified gas cylinders for calibration in ppb/ppm range. Permeation tubes are easier to ship, transport and store than gas cylinders, thus avoiding the need to deal with high pressure gases that are heavy and take up valuable lab space. The Permeation Device is specifically designed to allow calibrations ensuring full sample integrity, even at the lowest concentrations. The Pereation device can be equiped with 2 permeation ovens, offering the flexibility required when permeation tubes used have different calibration temperatures. Using 2 ovens also allows for calibrating multiple components at a time. Permeation tubes will never fully replace calibration gas cylinders but they are the preferred solution when calibrating at ppm or ppb levels. It offers an easy to use and cost-effective solution that provides better quality calibration even at lower levels, great flexibility towards components, while avoiding known issues commonly seen in Calibration Gas cylinder use. Components Sample Matrix Concentration Range Temperature Range Perm Oven Perm Oven Temp. Accuracy Mass Flow Controller Range Most volatile components that can be determined by Gaschromtagraphy (e.g. H2S, COS, DMS, SO2) See: here for a complete list Gas Low PPB to PPM 35 80 C Figure 3. Better than 0.1 C 20 200 ml/min Mass Flow Control Accuracy +/- 5% Table 4. Specifications AC Integrated Permeation Device AC Analytical Controls has been the recognized leader in chromatography analyzers for gas, naphtha and gasoline streams in crude oil refining since 1981. AC also provides technology for residuals analysis for the hydrocarbon processing industry. Applications cover the entire spectrum of petroleum, petrochemical and refinery, gas and natural gas analysis; ACs Turn-Key Application solutions include the AC Reformulyzer, SimDis, Hi-Speed RGA and Customized instruments. 00.00.231 2013/1 - Copyright 2011 PAC L.P. All rights reserved