Setting up and running a column

Similar documents
7.9. Flash Column Chromatography Guide

Lab Equipment ANALYTICAL BALANCE

LABORATORY TECHNIQUES. Pouring Liquids

Part 1: Distillation and Analysis of an Unknown Alcohol Mixture

CHM 100 / Introductory Laboratory Experiment (r10) 1/11

PREPARING GAS SAMPLES IN LARGE PLASTIC SYRINGES

Chemistry 261 Laboratory Experiment 6: Reduced Pressure Distillation of a Mixture of High Boiling Alcohols

Objective To identify a pure liquid substance using the physical properties of solubility, density, and boiling point.

Packing Procedure Recommendation. For. YMC RP silica. ODS-A, ODS-AQ, Octyl, Butyl, YMCbasic

CHM Introductory Laboratory Experiment (r17sd) 1/13

Using the Akta Prime plus October 22, 2012

CTB3365x Introduction to Water Treatment

DEMONSTRATION 2.1 PROPERTIES OF CO 2. Chapter 2: Gases

CH2250: Techniques in Laboratory Chemistry. Outline Measuring Mass Measuring Volume Significant figures. Mass Measurement

Lab Session #4 AN Physical Properties

Part A: 20 minutes plus 30 minutes setting time Part B: day 1, 20 minutes; then 5 minutes each day for about 5 days

PRE LABORATORY ASSIGNMENT: Lab Section Score: /10 READ THE LAB TEXT BEFORE ATTEMPTING THESE PROBLEMS!

NMR Service. How to Prepare Samples for NMR

Gas Laws. Introduction

SOLUBILITY OF A SOLID IN WATER

1. If grams of the vapor of a volatile liquid can fill a 498 ml flask at o C and 775 mm Hg, what is the molecular mass of the gas?

GRADE 6: Materials 1. UNIT 6M.1 7 hours. Solubility. Resources. About this unit. Previous learning. Expectations. Key vocabulary and technical terms

BOYLE S / CHARLES LAW APPARATUS - 1m long

THE DECOMPOSITION OF POTASSIUM CHLORATE This lab is derived almost entirely from a lab used at the United States Naval Academy

Student s Name: Date : Molar volume of butane

Solvent System Walkthrough

Guidelines on Collection of Water Samples for Phytoplankton Analysis

To derive from experiment the relationships between Pressure (P), Volume (V), Temperature (T), and Water Solubility of gases.

CHE 4115 Chemical Processes Laboratory 2 Experiment 1. Batch Distillation

PARTICLE SPACING PART 1

CHEM254 #4 The Synthesis of a Tertiary Alcohol Using a Pre-Made Grignard Reagent 1

Chromatographic separation procedures: gas chromatography with Cobra4

AKTA pure 25 New Owner s Intro

Eliminating Sources of Error in IV Pump Testing

CHM 2045L Physical Properties

Chemical Spill Clean-up Protocol

Operating Instructions

ExamLearn.ie. The Air & Oxygen

Photolithography. Operating Instructions

MassWWP Standard Operating Procedure R-2 Dissolved Oxygen Rev.0 Page 1 of 10

limbsandthings.com Advanced Venepuncture Arm User Guide For more skills training products visit Limbs & Things Ltd.

Figure. Experimental set up of catalyst, reactants and product syringes.

AQUARIUS. Operating Instructions for the. Type 70 Electrolytic Gas Soldering / Welding Unit. Table of Contents

SOLUBILITY OF A SOLID IN WATER

ASPHALT PLANT LEVEL 1

Experiment #2. Density and Measurements

How to use Solvent Purification System A brief manual

Properties of Gases Observing Atom Pressure of a Gas Measuring Gas Products of Chemical Inferring Molecule Reactions

LESSON Engage. A Model for Liquids. Key Objectives

Deep Water Currents Lab

Measuring Mass and Volume

LEVANT WORMSEED FOR HOMOEOPATHIC PREPARATIONS CINA FOR HOMOEOPATHIC PREPARATIONS

Analysis of Deuterium Enrichment by Fourier Transform Infrared Spectrometry (FTIR): Practice. Christine Slater PhD

VOLUMETRIC TECHNIQUES

Kinetic Model of Matter

Before you start sampling, be sure to read

Experiment 12: MOLAR VOLUME OF AN IDEAL GAS

Marking films - Application guide

Making a Cartesian Diver Toy

. In an elevator accelerating upward (A) both the elevator accelerating upward (B) the first is equations are valid

Students measure the change in pressure by varying the volume of trapped air in a syringe while:

9/28/2017 OBJ: SWBAT determine the solubility of 10 compounds. 1. What makes Magic Sand hydrophobic?

Rev. CO Packed Columns for Normal Phase Chromatography. TSKgel NH m INSTRUCTION MANUAL

The grade 6 English science unit, Gases, meets the academic content standards set in the Korean curriculum, which state students should:

11.1 Dumas Method - Pre-Lab Questions

SRI Model H2-40 Hydrogen Generator Operation and Maintenance Nov 2006

Working correctly with gas

Experiment GC : Analysis of BTEX by GC-FID

Title: Solubility of Gas A Daily Experience. Subject: Chemistry. Grade Level: 10 th 12 th

Standard Operating Procedure

Research Question How does the concentration of catalase affect the speed of the decomposition reaction of Hydrogen Peroxide into oxygen and water?

MV5: Automated concentration system

Experiment 18 Properties of Gases

LABORATORY SAFETY EQUIPMENT Final Grade: /45

Hands-On Experiment Density and Measurement

Standard. Standard. Venepuncture Arm. Light. Part No: Brown. Part No: Black. Part No: 00332

Operating Manual. Manual Blaine Air-Permeability Apparatus for determining the fineness of cement EN 196-6

Emulsiflex C3 Homogenizer Protocol

275 RID Assembly Instructions

Mini-Labs. 7. Movie Fog 8. Jet Power 9. Sauce Pan 10. Magnetic Bubbles 11. Aquarium Magic 12. Hockey Puck 13. Carbon dioxide balloons

29 Pressure, Temperature relationship of a gas

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

BS ml Bottle Sampling Unit

CHEMISTRY FACULTY LABORATORY SAFETY CONTRACT CENTRAL CAMPUS

Elementary Science ACTIVITIES

15 Air Around us. We have learnt in Chapter 9 that. not to be republished NCERT 15.1 IS AIR PRESENT EVERYWHERE AROUND US? Activity 2.

Gases and Pressure SECTION 11.1

SOP: Buck Scientific BLC-20P HPLC Operation

BASIC LAB SKILLS: EIGHING BALANCE AND BUNSEN

METHOD OF TEST FOR THEORETICAL MAXIMUM RELATIVE DENSITY OF BITUMINOUS PAVING MIXTURES LS-264 R27 ASTM D2041/D2041M

Intermolecular Forces

INTRODUCTION TO THE SPECTROPHOTOMETER AND PIPETTING SKILLS

STATIONARY TRUCK INTERNAL HALYARD V-CLEAT FLAGPOLES FOR QUICK AND PROFESSIONAL INSTALLATION READ ALL INSTRUCTIONS BEFORE PROCEEDING

Additional Reading General, Organic and Biological Chemistry, by Timberlake, chapter 8.

Diffusivity in Gases Experiment ver 05.10a

Build your own weather station

A SIMPLE TITRATION METHOD FOR DETERMINING THE SPECIFIC GRAVITY ON ONE DROP OF URINE

of Carbon Dioxide (CO 2

Lipodrop 2.0 device coating

The Safe Use of Pyrophoric Substances

Transcription:

Setting up and running a column PART 1: What you need and theory PART 2: Using a bellow and packing the column PART 3: Loading sample to silica PART 4: Running column PART 5: TLC analysis and combining fractions Chemistry Department

PART 1: What you need and theory

What you need Column Silica Gel (bottle may look different) Funnel (for pouring) Beaker (useful when packing column) Silica in conical (for packing column) Your sample/mixture Spatula Other items not on the picture: - Pasteur pipette/teat - TLC plates and related equipment Long thin pole to help bung column with cotton wool Test tubes and Rack Cotton Wool Eluent Sand (bottle may look different) Round bottomed flask (for sample loading)

What you need Green clip Small piece of rubber tubing Please note: - Some of these items (tubing, connector and bellow) may be preassembled for you Cone Tubing connector Bellow

What is the eluent? In a lot of purification techniques, there is a stationary phase and a mobile phase: Stationary phase: The adsorbant material in which the sample being purified binds to. Mobile phase: The substance used to unbind the sample to produce the analytes. The mobile phase in a column, TLC and GC is referred to as the eluent. It derives from the Latin word for washing out as the eluent pulls the products through the stationary phase: Column/TLC: Eluent is the solvent you are using GC: The eluent is the gas being used.

How does a column relate to a TLC? In a column, you load your sample at the top Silica is the stationary phase in both a TLC and a column Three products in the mixture In a TLC, you load your sample at the bottom The eluent (mobile phase) in a TLC moves upwards, and the products move upwards also. The eluent (mobile phase) in a column moves down, and the products move down also. Therefore a column is essentially an upside down TLC, just on a preparative scale where you can collect the samples as they come through the stopcock at the end

How does a column relate to a TLC? Two products in the mixture NOTE: TLC analysis should always be completed before a column to determine what eluent system you should be using. If this is how the TLC looks in eluent x, then I would happily use eluent x as the starting eluent for the column. This was the case for the real TLC and column system shown here. The middle section of the TLC (called mix ) is the mixture loaded onto the column NOTE: The assumption has been made that if you are attempting a column you are familiar with how to run a TLC

PART 2a: Packing column The chromatography column should be completed inside a fume cupboard. It is preferable to complete the packing, loading and running of a column in one day, where possible,

The column. There are some different variations of column. For example, the diameter can vary, the top of the column may not have a quickfit joint, the stopcock could be glass or teflon, and it may also contain a sinter filter towards the bottom. If the column contains a sinter/filter disc then you can ignore the next few slides. Stopcock

If there is no sinter To the bottom of the empty column, bung the tap by loosely placing cotton/glass wool. Use a long thin pole to help bung column with cotton wool. CARE: If you do this too tightly, the solvent will pass through too slowly meaning the column will take too long.

If there is no sinter Place enough sand to that there is a level base. Add a bit of eluent Take some sand and pour a small portion into the column (using a funnel) After you have added the sand, pour a small amount of eluent to wash the sides of the flask (sand may have stuck). If sand is not level, tap the sides of the column GENTLY and this should help level the sand. Having a small excess of eluent here is good when you pour in the silica, it will not disturb the flat sand layer

Create a slurry. Dissolve 25-35 g of silica gel (that is sufficient for this size column) into your starting eluent. After the next few steps, if you have a silica column path shorter than needed (see in a few slides time), you may need to do this again on a smaller scale to top up your column. CAUTION: Avoid pouring dry silica into the column. Silica has fine dust particles which can cause irritation. The slurry never fully dissolves, but it is now easier, to pour into the column without dry silica dust going everywhere

Carefully pour the slurry into the top of the column at a rate not to disturb the sand layer at the bottom. Wash flask containing slurry residue with additional starting eluent and pour that into the column. Allow silica to settle Pour silica slurry in carefully Although not shown here, wash flask with additional eluent Allow silica to settle. If you can see air bubbles, GENTLY tap the side of the column to help remove the bubbles Note: The level of the silica appears to have decreased a little. This occurs as the silica settles and packs itself

Open the stopcock and allow the solvent to run through. Close the tap as the solvent reaches the top of the silica, ensuring it does not go dry. The aim is to get the silica depth at about 6 inches/15 cm. If you are a long way short, you will need to make up a small bit of slurry and top it up Collect eluent into some sort of flask The sample is ready to be loaded when the solvent level is close to the top, you have the appropriate path length, and the silica is fully packed (it will appear quite firm). Solvent collected upto this point can be reused on the column as it is not contaminated

PART 2b: Using a bellow To speed up the flow rate inside a column, you can apply a slightly positive pressure on into the column reservoir. This is achieved using a bellow. Golden rules and safety when using a bellow: If you seal the column with a green clip, the column tap must be open. If the column tap needs to be closed, the bellow must be removed. Petroleum ether volatility keeps the pressure inside the column quite high, so two squeezes is all that is required.

Using a bellow. Setting up the below is shown below this may already been done for you. Connect the bellow to the tubing connector Connect a piece of tubing to the other side of the tubing connector Add a cone to the other end of the tubing

Using a bellow. Place the bellow onto the top of the column and seal with a green clip. The tap must be open when a green clip is attached Petroleum ether has a high volatility so a natural pressure will build up in the column You are working with pressures greater than atmospheric, using glass to hold the pressure if you over pressurise the column, it could smash.

Using a bellow. With the clip in place, and the tap open, squeeze the top part of the bellow a maximum of 3 times you will see the lower bulb expand. Only once the lower bulb deflates, and the flow rate drops again should you squeeze the top part of the bellow again. The pressure inside the column will increase when the bellow is squeezed and the flow rate (right) will increase until the pressure in the lower bulb drops and the flow rate drops. You are working with pressures greater than atmospheric, using glass to hold the pressure if you over pressurise the column, it could smash. Increased flow rate

Using a bellow. Removing the cone from the top of the column should be done with care. Place one hand over the top of the cone and remove the green clip with the other. Using the same hand that removed the green clip, carefully pull the cone out of the column Remember: If there is still any pressure inside the column, the cone may jump out and smash (trust me, it ll happen!)

PART 3: Loading sample to silica The chromatography column should be completed inside a fume cupboard. It is preferable to complete the packing, loading and running of a column in one day, where possible,

If the sample is not soluble in the starting eluent, it must be pre-adsorbed onto silica to help loading onto the column. Take a small amount of silica in a round bottom flask and add your sample to it (the volume of silica and sample should be roughly the same). Create a slurry using a low boiling & volatile solvent eluent (in a similar fashion to earlier). If the sample is not soluble in the starting eluent, do not use this for pre-loading your sample to silica. You need something the sample dissolves in (ideally a solvent you will use later in the column).

Place your slurry on a rotary evaporator and remove all traces of solvent. It is imperative that the sample is completely dry It is important that the sample is dry for one main reason: If you use a solvent that is more polar than the starting eluent for preloading, any traces of this at the start of the column will ruin your column. The whole point of a column is to gradually increase the solvent polarity to separate materials if there is a more polar solvent present straight away, little separation will be observed Splash guard Take care when removing solvent from a slurry on a rotary evaporator. As it approaches dryness, the silica can have a tendency to bump. Although not shown here, it can be beneficial to place some cotton wool in the end of the splash guard. If the sample does bump, the cotton wool will prevent loss of product into the splash guard.

Scratch the sample/silica mix off the sides of the flask using a spatula.

Carefully pour the mixture onto the top of the column. The aim is that the combined sample/silica mix is not much more than 1-2cm in depth

Using a few millilitres of starting eluent in a pasteur pipette, rinse the sides of the column to ensure that all of your sample/silica mixture is on top of the silica bed. Adjust the amount of starting eluent so that the amount of solvent is approximately level to the top of your sample. The aim is that the combined sample/silica mix is not much more than 1-2cm in depth Add a small amount of staring eluent to wash the sides of the column At this point, you can now add a protective layer of sand

Protect the top of the silica by adding a small quantity of sand (to a depth of about 1-3 cm). Add a few more millilitres of starting eluent from a pasteur pipette to wash sand from the side of the column. Then top up the reservoir and the column is ready to run.

PART 4: Running column The chromatography column should be completed inside a fume cupboard. It is preferable to complete the packing, loading and running of a column in one day, where possible,

After topping up the reservoir with starting eluent, open the tap and start collecting fractions. You can use a bellow (see earlier) to help with this. Note: Do not let the reservoir empty. You can allow the solvent level to reach the top of the sand, but try and avoid the level dropping any further, as this can introduce air bubbles into the column. Remember that the sample can only run through the column at the speed the eluent moves. Whilst it is ok to collect a small amount in the beaker, you should quickly switch to test tubes (next slides)

Collect fractions of about 15 ml (test tubes shown are sufficient). You will need to perform TLC analysis (later) on the fractions to see where your products are. Note: In the example here, we have used a sample which is yellow to the eye, so you can nicely see the progress of the material down the column. This is not usually the case, so TLC analysis is the main way to determine column progress.

Increasing solvent polarity In the example shown, the first product to elute is a white solid, which is colourless in solution (left test tube). The second product to elute is a yellow solid/solution (right test tube). By doing TLC analysis, it was determined when the first product was eluted. After all the first spot was eluted, the eluent polarity was increased to elute the second product. You will probably need to do something similar with your compound. Note: Do not change the polarity too dramatically, too quickly. Instead, do it gradually. In this example, the starting eluent was petroleum ether, and the increased polarity eluent was 10% ethyl acetate in petroleum ether. This level of increase was ok in one go.

PART 5: TLC analysis and combining fractions The chromatography column should be completed inside a fume cupboard. It is preferable to complete the packing, loading and running of a column in one day, where possible,

For the example given here, this is what the TLC looked like under various conditions. This was completed prior to the column to check that the eluent system wass sufficient. Under natural light Under UV Under natural light with products highlighted using pencil The middle section of the TLC (called mix ) is the mixture loaded onto the column

A great result! This is the aim of any column as the top spot has been successfully separated from the bottom spot. The number of empty fractions does not matter Fractions 5 11 could be combined in a round bottom flask and taken to dryness to give you a pure sample of the top spot. Fractions 14 24 could be combined in a round bottom flask and taken to dryness to give a pure sample of the bottom spot The S here is just a reference spot of the starting material. It shows that the spots eluting from the column correlate to what was there at the start Under UV light Note: Every fraction has been TLC ed here for demonstration purposes. It is acceptable to TLC every other fraction until you determine where your sample is. You can then go back and do another TLC, if required Under natural light

A terrible result! Your spots have co-eluted. No separation has been achieved so in this situation you should remove all your solvent and start again Both spots have eluted together so the column has not worked

Nor a good or terrible result! The column has resulted in some pure top spot, some pure bottom spot, but with a mixture in the middle. On this occasion, you should collect 3 fractions: Fraction A: Top spot Fraction B: Mixture Fraction C: Bottom spot Under natural light with products highlighted using pencil Fractions 5 7 could be combined and taken to dryness to give you a pure sample of the top spot. Fractions 8 10 could be combined to give you a mixture. THEN CONSIDER RE-COLUMNING THIS SECTION ONLY Fractions 14 24 could be combined and taken to dryness to give a pure sample of the bottom spot.