Customer Responsibilities. Important Customer Information. Agilent 6400 Series Triple Quad LC/MS Site Preparation Checklist

Similar documents
Customer Responsibilities

Customer Responsibilities. Important Customer Information Series Q-TOF LC/MS Systems Site Preparation Checklist

Customer Responsibilities. Important Customer Information. Agilent Infinity II Preparative HPLC System Site Preparation Checklist

490 Micro GC Site Preparation Checklist

Customer Responsibilities. Important Customer Information. Cary 4000/5000/6000i UV-Vis spectrophotometer Site Preparation Checklist

Customer Responsibilities

490 Micro GC Site Preparation Checklist

Customer Responsibilities

Customer Responsibilities

Customer Responsibilities

For additional information about our solutions, please visit our web site at US/Pages/HomePage.

6890 GC Site Preparation

Varian 325-MS LC/MS Quadrupole Mass Spectrometer Pre-installation Instructions

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 7820MSD Series G3175A (G7020A), G3176A (G7021A)

Customer Responsibilities. Important Customer Information GCMS QQQ Site Preparation Checklist

DD2 Site Preparation Checklist Hardware Site Preparation Specification

Agilent 6100 Series. Single Quad LC/MS Systems. Site Preparation Guide

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 5975C MSD Series G3170A, G3171A, G3172A, G3174A

Pre-installation Checklist

Please refer to the Hydrogen Safety Guide which is shipped with the Instrument. 5977B Series GC/MS Systems Site Preparation Checklist

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Agilent 7890 / 220 Ion Trap GCMS Site Preparation Checklist

AB SCIEX QTRAP 4500 System

7820A GC Site Preparation Checklist

Oxford Instruments Plasma Technology. PlasmaPro NGP80. Installation Data Sheet. Original Instructions.

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Information Sheet. About this information sheet. Floor and wall loadings. Attachment of process module to floor. PlasmaPro Estrelas100

Avio 500 ICP-OES. ICP-Optical Emission Spectroscopy PREPARING YOUR LAB. Preparation Checklist. Suitable Working Area

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Varian 240-MS GC/MS Ion Trap Mass Spectrometer Pre-installation Instructions

Site Planning Guide. API 2000 LC/MS/MS System

Preparing Your Laboratory for the Optima 7300 V and 8300 ICP-OES Spectrometers

DF-310E PROCESS ANALYSERS APPLICATIONS FEATURES

Markes Thermal Desorption Systems Site Preparation Page 1 of 8

Agilent 5975T LTM GC/MSD

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Plasma. Argon Production Helium Liquefaction Truck Loading Pure Gas Bottling

Workhorse 8 Oxygen System Model No 5665, 5673, 5679

Discovery SDT 650. Site Preparation Guide. Revision G Issued December Page 1

Workhorse-15 Oxygen Generator

UltraLo 1800 Alpha Particle Counter

be present throughout these services; otherwise, they will miss important operational, maintenance and safety information.

Columbus Instruments

Agilent 7820A GC. Site Preparation Guide. Agilent Technologies

Technical Information

Series Environmental Chambers

Agilent 5100 and 5110 ICP-OES. Site Preparation Guide

Agilent 7800 ICP-MS. Specifications and Typical Performance. Fast track metals analysis with Solution-Ready ICP-MS

ZEFIRO series NITROGEN GENERATORS. The constant and completely autonomous supply of nitrogen flux DESIGN AND PRODUCTION OF LABORATORY GAS GENERATORS

Nitrogen Supply System

Discovery TGA 55, TGA 550, TGA Site Preparation Guide

NexION 1000/2000 ICP-MS

Installation, Operation, and Maintenance Manual Parker Balston Model N2-04 Nitrogen Generator

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

28 Site Preparation. Supplying valve actuator air

FPG8601 Force Balanced Piston Gauge

TA Instruments Installation Requirements for the Thermal Discovery Series

The instrument should have a mass range from 2 to 2000 m/z or better Interface

TECHNICAL INFORMATION

ECS Protector Nitrogen Generator PGEN-30 (PGEN-30E)

HumiSys HF High Flow RH Generator

Discovery HP-TGA 75/750. Site Preparation Guide

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

Equipment and Accessories. ROMIL LiQuiSure. Designed for High Purity Containment & Dispensing PURE CHEMISTRY

PROCESS ANALYSERS. SERVOPRO PureGas. Gas Purifier. Operator Manual. Part Number: A Revision: 0 Language: UK English

973-SF 6 Analyzer. Precise and Stable SF 6 Gas Analyzer REFLECTING YOUR STANDARDS

Agilent 7700 Series ICP-MS

User Manual. Agilent 1260 Infinity Analytical- and Preparative-scale Fraction Collectors

FAST EVAPORATION AT NORMAL PRESSURE

VETROSON. OXY-GEN SYSTEMS (Oxygen Generating Systems) Question & Answer Brochure

Agilent GC, GC/MS, and ALS

NGN Series nitrogen generator

Preinstallation Checklist and Instructions for the TriStar These Preinstallation Checklist and Instructions were reviewed and approved by:

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

User Manual. Quantos Automated Dosing Liquid Module

DF-550E PROCESS ANALYSERS APPLICATIONS FEATURES

Zebron Gas Management

Agilent 7697A Headspace Sampler

Agilent 1220 Infinity II LC Mobile Upgrade Kit

Digital Vacuum Regulator

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

Analytical Gas Systems Product Catalog

GasSense NDIR User Manual

HiQ laboratory gas generators.

SPECIFICATIONS PARTICLE SENSOR KS-18F Higashimotomachi, Kokubunji, Tokyo , Japan

INTRODUCTION. The Quantum Technology system has the following advantages:, as it does not need plastic gas-bags which are volume-consuming,

Nitrogen generation for modified atmospheric packaging

Mass Spec will not Autotune

Agilent 7667A. Mini Thermal Desorber. Site Preparation

Mooney * Noise Controller Installation, Operation, and Maintenance Manual

WHAT ARE THE SPECIFICATIONS FOR THE ARGON GAS OPTION USED IN JOULE THOMSON REFRIGERATORS?

Oxygen Concentrator. Installation and Operation Manual. ClearWater Tech, LLC. Integrated Ozone Systems

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Superior Result from Analytical Instruments The constant purity of the nitrogen improves system stability and ensures reproducible results

User Manual. Agilent 1260 Infinity Micro-scale Fraction Collector/Spotter

Agilent 5100 and 5110 ICP-OES. Site Preparation Guide

Inert Air (N2) Systems Manual

NGP-250/500 Nitrogen Generator Quick Start Guide

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Transcription:

Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is the key first step in ensuring that your instruments and software systems operate reliably over an extended lifetime. This document is an information guide AND checklist prepared for you that outlines the supplies, consumables, space and utility requirements for your equipment. Customer Responsibilities Make sure your site meets the following specifications before the installation date. For details, see specific sections within this checklist, including: The necessary laboratory or bench space is available The environmental conditions for the lab as well as laboratory gases and plumbing The power requirements related to the product (e.g., number & location of electrical outlets) The required operating supplies necessary for the product and installation Please consult Other Requirements section below for other product-specific information. For more details, please consult the product-specific Site Preparation or Pre-Installation manual. If Agilent is delivering installation and familiarization services, users of the instrument should be present throughout these services; otherwise, they will miss important operational, maintenance and safety information. Important Customer Information 1. If you have questions or problems in providing anything described as a Customer Responsibility above, please contact your local Agilent or partner support/service organization for assistance prior to delivery. In addition, Agilent and/or it's partners reserve the right to reschedule the installation dependent upon the readiness of your laboratory. 2. Should your site not be ready for whatever reasons, please contact Agilent as soon as possible to re-arrange any services that have been purchased. 3. Other optional services such as additional training, operational qualification (OQ) and consultation for user-specific applications may also be provided at the time of installation when ordered with the system, but should be contracted separately. Page 1 of 11

Dimensions and Weight Identify the laboratory bench space before your system arrives based on the table below. Pay special attention to the total height and total weight requirements for all system components you have ordered and avoid bench space with overhanging shelves. 1. The 6400 Series Triple Quad LC/MS dimensions represent the maximum cabinet dimensions with a Spray Chamber installed. 2. At least 30 cm (1 ft) to the left (source end) and right of the instrument must be added to the dimensions to provide adequate instrument access and ventilation. 3. The supporting surface must be relatively vibration free and capable of supporting the combined weight of the Triple Quad system. Instrument/Spray Chamber/Foreline Pump Model Weight Height Depth Width Kg lbs cm in cm in cm in G6410A QqQ LC/MS 107.5 236.5 47 18.5 66 26 111 43.5 G6410B QqQ LC/MS 107.5 236.5 47 18.5 66 26 111 43.5 G6420A QqQ LC/MS 107.5 236.5 47 18.5 66 26 111 43.5 G6430A QqQ LC/MS 115 255 47 18.5 66 26 111 43.5 G6460A QqQ LC/MS 115 255 48 18.8 66 26 111 43.5 G6460C QqQ LC/MS 115 255 48 18.8 66 26 111 43.5 G6490A QqQ LC/MS 115 255 47 18.5 76 30 84 33 G1948B Electrospray Source 1.7 3.8 17 6.8 9.5 3.7 18 7.1 G1947B APCI Source 1.7 3.8 23 9.2 13.0 5.1 18 7.1 G1971B APPI Source 1.7 3.8 23 9.2 13.0 5.1 18 7.1 G1978B Multimode Source 2.29 5.05 23 9.2 13.0 5.1 18 7.1 Agilent Jet Stream 1.7 3.8 23 9.2 11.5 4.5 18 7.1 G4240A HPLC-Chip Cube 14 31 36 14.1 30 11.7 35 13.7 E2M28 Foreline Pump 40.0 88.2 28 11 58.3 23.0 17 7.1 MS40+ Foreline Pump 33.0 72.7 22.8 9.0 41.8 16.5 29.7 11.7 Page 2 of 11

Height Depth Width Figure 1 G6410A/B dimensions Height Height Depth Width Figure 2 G6420A/30A/60A/60C dimensions Depth Width Figure 3 G6490A dimensions Environmental Conditions Operating your instrument within the recommended temperature ranges insures optimum instrument performance and lifetime. 1. Performance can be affected by sources of heat & cold e.g. direct sunlight, heating/cooling from air conditioning outlets, drafts and/or vibrations. 2. The site s ambient temperature conditions must be stable for optimum performance. 3. The Agilent 6400 Series Triple Quad LC/MS is specified for operation under the following conditions: a. Indoor use b. Constant temperature (< +/-3 C from calibration temperature) c. Non-condensing, non-corrosive atmosphere d. Altitude: Not to exceed 3,300 m up to 35 C, not to exceed 3,700 m up to 30 C Instrument Model Operating temp range C (F) Operating humidity range (%) Heat Dissipation (BTU) 6400 Series QqQ LC/MS (all models) 15-35 C (59-95 F) < 85% RH @ 35 C < 4500 BTU/hr Page 3 of 11

Power Consumption 1. If a computer system is supplied with your instrument, be sure to account for those electrical outlets. 2. The LC/MS electrical outlet(s) must have an isolated, noise-free electrical ground that is connected to the main earth ground for the facility. 3. Mains supply voltage tolerances must be between +10% and -5% of nominal line voltage. 4. Models utilizing the E2M28 foreline pump must be ordered specifying installation site voltage. 5. Electrical power for the 6400 Series Triple Quad LC/MS may be delivered in either single-phase or 208-Wye configuration: Configuration Measurement Nominal Voltage Singe Phase Line to neutral 200, 220, 230 or 240 VAC Line to ground 200, 220, 230 or 240 VAC Ground to neutral < 0.5 V rms 208-Wye Line to line (phase A to phase B) 208, 220 VAC Line to ground (phase A to ground) 120, 127 VAC Line to ground (phase B to ground) 120, 127 VAC Instrument Model Line Voltage & Frequency (V, Hz) Supply Circuit Rating (A) Number of Outlets G6410A 200-240 VAC @ 50/60 Hz 15 A 1 2500 G6410B G6420A G6430A G6460A 200-240 VAC @ 50/60 Hz 15 A 1 2700 G6460C 200-240 VAC @ 50/60 Hz 15 A 2 2850 G6490A 200-240 VAC @ 50/60 Hz 15 A 2 2850 Maximum Power Consumption (VA) Page 4 of 11

Main Nitrogen Gas Supply Requirements 1. For information on Agilent consumables, accessories and laboratory operating supplies, please visit http://www.chem.agilent.com/en-us/products/consumables/pages/default.aspx 2. Impurities from LN 2 Dewar being oxygen only. 3. "Hydrocarbon free" means < 0.1 PPM hydrocarbons with the remaining gas being oxygen and trace argon. 4. Nitrogen Pressure as measured at the LC/MS inlet (not the supply side). 5. Minimum Nitrogen Flow required at all times to prevent air from entering the instrument. 6. Main Nitrogen Supply fittings are 1/4" SwageLok. Model Nitrogen Source Nitrogen Purity Pressure Flow G6410A G6410B G6420A G6430A G6460A#100 G6460C#100 LN 2 Dewar Nitrogen Generator 99.5% and 95.0% and 5.5-6.8 bar (80-100 PSI) 18 L/min Maximum ( 1080 L/hour) > 3 L/min Minimum G6460A G6460C LN 2 Dewar Nitrogen Generator 99.5% and 95.0% and 5.5-6.8 bar (80-100 PSI) 30 L/min Maximum ( 1800 L/hour) > 3 L/min Minimum G6490A LN 2 Dewar Nitrogen Generator 99.5% and 95.0% and 5.5-6.8 bar (80-100 PSI) 50 L/min Maximum ( 3000 L/hour) > 9 L/min Minimum Page 5 of 11

Collision Cell Nitrogen Gas Supply Requirements 1. For information on Agilent consumables, accessories and laboratory operating supplies, please visit http://www.chem.agilent.com/en-us/products/consumables/pages/default.aspx 2. Nitrogen is the only supported Collsion Cell gas. 3. Splitting the Main Nitrogen Gas supply for use with the collision cell is not supported due to nitrogen purity requirements. 4. Collision Cell gas supply fittings are 1/8" SwageLok. Instrument Model Nitrogen Source Nitrogen Purity Pressure Flow 6400 Series QqQ LC/MS (all models) High Pressure Cylinder 99.999% and (< 0.1 PPM hydrocarbons) 1-2 bar (15-30 PSI) 0.001 L/min ( 0.006 L/hour) Page 6 of 11

Compressed Air Requirements - HPLC-Chip Cube equipped systems only 1. For information on Agilent consumables, accessories and laboratory operating supplies, please visit http://www.chem.agilent.com/en-us/products/consumables/pages/default.aspx 2. Compressed Air is requried for HPLC-Chip Cube equipped systems only. 3. Air source must be able to deliver air at a constant pressure. 4. Inlet gas pressure for air must be equal to or slightly higher than the pressure used for the Main Nitrogen Gas supply 5. Compressed Air supply fittings are 1/4" SwageLok. Instrument Model Air Source Air Purity Pressure Flow G6410B G6420A G6430A G6460A G6460C High Pressure Air source 99.99% and 6.8-6.9 bar (100-120 PSI) 4 L/min Maximum ( 240 L/hour) G6490A High Pressure Air source 99.99% and 6.8-6.9 bar (100-120 PSI) 10 L/min ( 600 L/hour) Page 7 of 11

Exhaust Venting For information on Agilent consumables, accessories and laboratory operating supplies, please The LC/MS generates exhaust fumes from the foreline pump(s) and drain bottle (from the spray chamber) that must be properly vented for supported instrument operation and compliance with laboratory safety requirements. 1. Exhaust must be vented according to local Environmental Health and Safety regulations. 2. Exhaust gases contain traces of solvent, sample and hydrocarbon pump fluid. 3. Venting Rate is commensurate with Nitrogen comsumption rate. 4. Two independent, negative pressure vents must be available with one for each of the exhaust sources: foreline pump(s) and Spray Chamber. If only 1 vent is available, the exhaust line(s) from the foreline pump(s) must extend beyond the exhaust line from the spray chamber. 5. If a negative pressure vent is not available, the length of the tubing from the foreline pump(s) and the drain bottle to the vent should each not exceed 460 cm (15 ft). 6. Exhaust tubing is 1/2" interior diameter (I.D.). G6410A G6410B G6420A G6430A G6460A#100 G6460C#100 G6460A G6460C G6490A Instrument Model Combined Exhaust Venting Rate (Continuous) 20 L/min Maximum ( 1080 L/hour) > 3 L/min Minimum 30 L/min Maximum ( 1800 L/hour) > 3 L/min Minimum 50 L/min Maximum ( 3000 L/hour) > 9 L/min Minimum Failure to vent the foreline pump and spray chamber separately will void the warranty for the 6400 Series Triple Quad LC/MS. Agilent service representatives will not install an Agilent 6400 Series Triple Quad LC/MS until an adequate exhaust system is present and functioning. Page 8 of 11

Required Operating Supplies by Customer 1. For information on Agilent consumables, accessories and laboratory operating supplies, please visit http://www.chem.agilent.com/en-us/products/consumables/pages/default.aspx 2. For detailed information on Operating Supplies, please refer to the Site Preparation Guide. Item Description, (including dimensions etc) Part Number Recommended Quantity Agilent Triple Quad LC/MS Site Preparation Guide G2571-90010 1 Tuning Calibrant Dependent on LC/MS model and Source Electrospray (ESI) Calibrant G2431A 1 APCI/APPI Calibrant G2432A 1 ESI-Low Calibrant G1969-85000 1 APCI-Low Calibrant G1969-85010 1 Performance Standard Electrospray/APCI Positive Ion G2423A 1 Performance Standard ES Negative Ion (Acid Red 4) G2424A 1 Performance Standard - ES Negative Ion (Chloramphenicol) 5190-0591 1 Performance Standard APCI Negative Ion G2525A 1 Performance Standard Multimode LC Demo Sample G1978-85000 1 Performance Standard Electrospray LC Demo Sample 59987-20033 1 Ammonium Formate G1946-85021 1 Formic Acid Reagent Grade G2453-85060 2 HPLC Flushing Solvent (500 ml) G1969-85026 1 Methanol, High Purity (1 L) 8500-1867 3 Acetonitrile, High Purity (1 L) G2453-85050 2 Water, High Purity (4 L) 8500-2236 1 Pipette, 1 ml 9301-1423 3 Volumetric Flask, 50 ml 9301-1424 1 Volumetric Flask, 100 ml 9301-1433 2 Vials, 2 ml Screw Top, Wide Opening, Amber (100/pk) 5182-0716 1 Vial Caps, Blue, PTFE/red silicone septa (100/pk) 5182-0717 1 Page 9 of 11

Recommended Configurations Agilent recommends 2 standard stacking configurations for your new system depending on the number and type of included modules. Please use these notes and figures as reference for HPLC and LC/MS configurations. 1. Stacking the entire HPLC modules on top of the 6400 Series QqQ LC/MS is not supported. Figure 4 Single HPLC stack configuration Figure 5 Double HPLC stack configuration Page 10 of 11

Important Customer Web Links For additional information about our solutions, please visit our web site at http://www.chem.agilent.com/en-us/pages/homepage.aspx Need to get information on your product? Literature Library - http://www.agilent.com/chem/library Need to know more? Customer Education http://www.agilent.com/chem/education Need technical support, FAQs? http://www.agilent.com/chem/techsupp Need supplies? http://www.agilent.com/chem/supplies Document part number: G2571-90018 Page 11 of 11