OXYTEMP Instructions 1

Similar documents
Operating Instructions

bicycle stand Assembly Instructions

Installation Instructions

PRO SINK -R 1 and 2. Dilution system CONTENTS

Paddle Bar Replacement

5600 ACID FILTER SERVICING INSTRUCTIONS

Operator: Save these instructions for future use!

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

Installation Instructions

Assembly and Disassembly Instructions for Trampoline and Enclosure

HoldUp Plus2. Safety Kit included: See additional instructions for installation. REAR WHEEL TRAY. BASE (1x) lock WASHER (1x) KEY (2x) SAFETY CLIP (1x)

ASSEMBLY & OPERATIONS MANUAL. Straight & Clean Torch Cuts. Every Time.

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

VERSA BIKE RACK INSTRUCTIONS

3. Align ratchet drive with female insert on assembly, and raise ratchet handle until the ratchet drive is firmly seated into the assembly.

Trampoline & Enclosure Assembly Instructions

Unblocking the MicroSprayer Aerosolizer Model IA-1C With the High Pressure Reverse Cleaning Adapter. Instructions for Use

INSTRUCTION MANUAL. January 23, 2003, Revision 0

Operator: Save these instructions for future use!

Oerlikon Sputtering Evaporator SOP

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

LRS(H)4 Pressure-Reducing Regulator User Manual

SAILTEC, INC CONGER COURT, OSHKOSH, WI USA

UNIVERSITY OF ROCHESTER ENVIRONMENTAL HEALTH & SAFETY

ATD LB PRESSURE BLASTER INSTRUCTION MANUAL

Radnoti Liver Perfusion System

SCUBAPRO. Balanced Power Inflator

Installation Instructions

Assembly Instructions

Installation Instructions

Southern Eagles Soaring

Operator: Save these instructions for future use!

12mm Dissolved Oxygen Sensor Care and Use Instructions

Infant Swing ( )

MULTICHANNEL. Labnet International, Inc.

Biochemistry Laboratory I CHEM 4401 Pipetting & Scales

Operating manual for the pressure transducer unit CCS 1000

Patients Guide for Self Administering Intravenous Antibiotics General points

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

Model 400ML Audible Information Device

Trampoline & Enclosure Assembly Instructions

ASSEMBLY INSTRUCTIONES

36H SERIES Combination Gas Valve

How Do I Install My Garden Master Bucket Garden?

CONSTRUCTOPEDIA NXT Kit 9797

OPERATION OF THE DIMPLER


Chapter 2. Pipetting techniques. Guide to Pipetting

Describing a journey made by an object is very boring if you just use words. As with much of science, graphs are more revealing.

FX IMPACT OWNER S MANUAL

Engineering Data Sheet

User s Manual. Digital micro pipette for liquid handling CERTIFIED ISO9001

Operator: Save these instructions for future use!

E. Test, and if needed, adjust tips of antenna. 1. Mark will test antennas with meter at ground end of feedline bundle. Will take 10 minutes.

SUMMITTM 400 & 600. Natural Gas Barbecues. Step-By-Step Guide

Calculating Total Uncertainty of Temperature Calibration with a Dry Block. Calibration White Paper. Beamex.

MAINTENANCE PROCEDURE FOR X 650

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

TECHNICAL BULLETIN. TriControls. Control Materials for Liquid QC and Calibration Verification

GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP

Installation of Your SprayMaster System

RG1200 Service and Repair Manual

Product Information News September 26, 2003

DynaPak Gas Sampler S Y S T E M S U P P O R T M A N U A L DP-2000

BCV31 DN20 - Blowdown Control Valve

PRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator.

Assembly Drawing: W-311B-A01, or as applicable Parts List: W-311B-A01-1, or as applicable Special Tools: , , &

Drive Belt Instructions

TOURNAMENT ICE. 8Ft-Gold Standard Manual Score Home Air Hockey Game Assembly Instructions / Manual

SCUBAPRO Repair Guide. S600-S550 Second Stages

DynaPak 2000L System Support Manual Version

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

Installation Instructions

Unblocking the MicroSprayer Aerosolizer Model IA-1B With the High Pressure Reverse Cleaning Adapter. Instructions for Use

Caliber Sled Wheels Assembly Instructions for PN and 13579

Helium-Oxygen Blender

Blue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION

Regulators repair and maintenance. XS Compact 2nd stage. January 2014 Rev XSC /3 Ed. C /14 1

RS Important Notes. Contact. Bicycle Maintenance Stand instructions manual. Warranty Period : 1 year (from the date of your purchase)

Unisense In Situ UniAmp Connector System

Motion Commotion. KindergarTen-second. Matter and Energy TEKS. Vocabulary

Operator: Save these instructions for future use!

SERIES 2 RAMP OWNER S MANUAL TOOLS REQUIRED: BEFORE YOU BEGIN... Read and understand these instructions before beginning a ramp setup.

Operator: Save these instructions for future use!

New Cortex Software User Guide

NTF-230 OPERATION MANUAL

Hydraulic Punch Drivers

Using a micropipette to transfer exact liquid measurements in scientific experiments By: Nicole Hume

BICYCLE TO MOTORCYCLE BICYCLE RACK

MATCHLESS BULKY PLYER FLYER HEAD

Adjust the Volume Display

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

FT500GF Dat e Code: 2 - FT500GF - - MW Purchase Date: PLEASE RETAIN THIS INSTRUCTION MANUAL FOR FUTURE REFERENCE. All Rights Reserved

Inside Front cover This page will remain blank.

;\- ,.. SAFETY WARNING: Failure to obey a safety warning may result In Injury to you or. T to others. T LOADED AND THE SAFETY IS IN THE

To Purchase This Item, Visit BMI Gaming (800)

Flat Fold Highchair INSTRUCTION MANUAL. Designed and manufactured to BS EN 14988: 2012 IMPORTANT:

Aqua Pilot Suction Pool Cleaner

ThermoVault Max. Extreme Temperature Thermal Barrier. Product User Guide

Shimano Di2 Installation on S5

Transcription:

OXYTEMP Instructions 1 Congratulations on purchasing your new OxyTemp Probe! ASSEMBLY INSTRUCTIONS FOR ECO-OXYGEN PROBES 1. Thread the stem into the bottom of the upper housing. 2. Tighten stem with a wrench. 3. Then, insert the yellow cord into the male connection located on the underside of the upper housing. GENERAL: The REOTEMP OxyTemp Probe allows you to draw a sample of air from your compost pile and then test that same sample to determine the percent oxygen present. All in one simple motion! Samples are only drawn from the tip of the probe, which allows you to control where each one is taken. The unique design of this probe enables you to clean out the tip (if clogging occurs) by forcing ambient air through the tip. For best results, a slow steady flow of air must be pushed across the sensor. See the sampling section for details. Calibration must be completed before use. Below is an overview of the probe components:

YELLOW HANDLE POSITION GUIDE: OXYTEMP Instructions 2 When the yellow handle is in the Horizontal Position or Sampling Position : - Pulling the black plunger will draw air from the tip. - Pushing the black plunger will move the drawn air across the sensor. Note: the pushing stroke should take 6 seconds for best results. When the yellow handle is in the Vertical Position or Cleaning Position : - Pulling the plunger will draw in ambient air. - Pushing the plunger will move air out the tip (cleaning it). Note: the pushing stroke should be fast to clean out tip.

CALIBRATION INSTRUCTIONS: OXYTEMP Instructions 3 1. Turn on probe (allow a few seconds to warm up). 2. Make sure the tip is in clean ambient air. 3. If needed turn yellow handle to sampling position, shown below: 4. Pull & Push 2 samples of fresh air (using black plunger). 5. Pull a 3 rd sample and while pushing back in, adjust calibration knob to read 20.9 (clean ambient air is 20.9% oxygen). Note: pushing stroke should take 6 seconds for best results. 6. Your oxygen probe is now calibrated and ready to measure!

OXYTEMP Instructions 4 SAMPLING: Warning: Avoid sucking liquid into the air chamber. Sediments may collect and cause clogging. 1. Calibrate probe (see instructions above). 2. Once calibrated, insert probe into compost pile. 3. Make sure the handle is in the Horizontal or Sampling position. 4. Pull the black plunger out. Push the black plunger in. Repeat 2 times. (this will purge the probe of any previously trapped gases). 5. Pull the plunger out a 3 rd time. 6. (While watching the display) Slowly Push the plunger in a 3 rd time. This should take 6 seconds for best results. Record the percent oxygen displayed as the plunger was being pushed in. 7. Move to your next sampling location or compost pile and start over with step1. Suggestion - While walking to the next sample location you may want to use the clean out function.

CLEANING INSTRUCTIONS: OXYTEMP Instructions 5 Clogging can be a major issue when sampling oxygen in compost, which is why Reotemp has designed this probe with a cleaning feature. When your probe becomes clogged, perform the steps below. 1. Move the yellow handle to the Vertical or Cleaning Position. 2. Pull the black plunger out (sucking in ambient air). 3. Quickly Push the black plunger in (forcing air out the clogged tip). 4. Repeat as necessary. Note: The cleaning procedure will likely only clean out some of the tip s holes. This is ok. You should be able to re-insert the probe into your next sampling location and successfully pull a new sample.

BATTERY INSTRUCTIONS: OXYTEMP Instructions 6 1. Use your finger to lift the battery holder as shown. 2. Slide battery holder out. 3. Replace battery. 4. Slide battery holder back in until it snaps into place. REPLACEMENT PARTS: Order replacement O2 Sensors and replacements tips at http://reotemp.mybigcommerce.com/