Dissolution Failure Investigation. Ken Boda Dissolution Applications Engineer

Similar documents
Frequently Asked Questions about Mechanical Qualification. Ken Boda Applications Engineer Agilent Technologies

The Transition to Enhanced Mechanical Qualification of the Dissolution Apparatus

Mechanical Qualification (MQ) for Apparatus 1 and 2. Oct 10th, 2008

Food and Drug Administration Division of Pharmaceutical Analysis. Mechanical Qualification of Dissolution Apparatus 1 and 2

02 November, 2017 Moderated by Bryan Crist Agilent Technologies

711 DISSOLUTION. Portions of the present general chapter text that are national USP text, and therefore not part of the

DISSOLUTION TEST FOR SOLID DOSAGE FORMS

BRIEFING 711 DISSOLUTION

Qualification Test Specifications for Dissolution Test Systems

International Journal of Generic Drugs ANALYTICAL METHOD PROCEDURES THIS SOP IS 'SWITCHED : OFF : ON'

USP Dissolution Instrument Calibration or PQ Standard Operating Procedure or SOP

Comparison between Compendial and Enhanced Mechanical Calibration Specifications

Dissolution and Drug Release Testing. DISSOLUTION SYSTEMS SOURCE BOOK Edition

COPLEY TABLET DISSOLUTION TESTERS

Dissolution Toolkit Procedures for Mechanical Calibration and Performance Verification Test Apparatus 1 and Apparatus 2 Version 2.

Effect of the Irregular Inner Shape of a Glass Vessel on Prednisone Dissolution Results

Effects of Dissolved Gases in Surfactant Dissolution Media

ANALYTICAL.

11 May, 2017 Moderated by Bryan Crist Agilent Technologies

Roller Chain and Sprocket Troubleshooting Guide

Hygea Ultrasonic Validation Kit

Chemical Analysis for Methyltestosterone

POOL CARE GUIDE. The E-Z Clor System To Perfect Pool Water.

Quantos Automated Dosing Solution Preparation Precise concentrations Process compliance Minimize out-of-specs

RULES PUBLICATION NO. 52/P UNDERWATER INSPECTION OF MOBILE OFFSHORE DRILLING UNITS IN LIEU OF DRYDOCKING

Peak Pressure Indicator MSI-3

PROCESS CONTROL HANDBOOK Z-6

KOBOLD KSM FLOWMETER. User Instructions

MIL-STD-883H METHOD EXTERNAL VISUAL

Agilent 1220 Infinity II LC Mobile Upgrade Kit

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

Extended Application Note

Reliable System Performance PM valves

Calcimeter Instruction Manual

Best Practice for Calibrating LTH Conductivity Instruments

TECTYL 1422S APPLICATION PROCEDURES AND TROUBLESHOOTING GUIDE

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

Right-Angle Tube Roller Model

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

ON WATER DISPERSIBLE GRANULES (WG)

VITLAB micropipette. Standard Operating Procedure (SOP)

OWNER S MANUAL NMB SERIES

WARNING! DO NOT THROW AWAY THESE INSTRUCTIONS! READ AND UNDERSTAND BEFORE USING EQUIPMENT!

Dissolution test results showing biorelevant media made from SIF Powder Original is the same as media prepared using methylene chloride

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION

CORESTA RECOMMENDED METHOD Nº 67

Transferpette. Single and Multichannel Pipettes

TITLE: DNA minitube - Red

WARNING! DO NOT THROW AWAY THESE INSTRUCTIONS! READ AND UNDERSTAND BEFORE USING EQUIPMENT!

A NOVEL RP-HPLC METHOD FOR THE QUANTIFICATION OF ICATIBANT IN FORMULATIONS

Rubber Dock Hose care, Use & Maintenance

Installation, commissioning and servicing instructions

CLASS D - SENSITIVE LEAK TEST GAS AND BUBBLE METHOD. 1.1 To provide definitive requirements for PNEUMATIC pressure testing of piping systems.

Crusher Bearings: Knowing the Basics Leads to Better Care

How to Achieve Optimal Weighing Performance

ENVIRONMENTAL LABORATORY APPROVAL PROGRAM CERTIFICATION MANUAL

STERILIZABLE DISSOLVED OXYGEN SENSORS

User Instructions 1790 Rail Anchor

130 SSP Soluble Salt Profiler

MODEL 200 KNIFE GATE VALVES INSTALLATION & MAINTENANCE MANUAL

Versatility. Accuracy. Compact. Versa. Automated Headspace Vial Sampler

DETERMINING OPTIMUM RESIDUAL ASPHALT CONTENT (RAC) FOR POLYMER-MODIFIED SLURRY SEAL (MICROSURFACING) MIXTURES

Title: Standard Operating Procedure for Elemental and Organic Carbon (EC and OC) using Non-Dispersive Infrared Detection (NDIR)

J138RSH and J150RSH 3/8" J138RSH 1/2" J150RSH. 3/8" Drive Sealed Head Ratchet 1/2" Drive Sealed Head Ratchet. Operation and Maintenance Manual WARNING

TITAN FLOW CONTROL, INC.

M-06 Nitrogen Generator (Nitrogen Making Machine)

Research Question How does the concentration of catalase affect the speed of the decomposition reaction of Hydrogen Peroxide into oxygen and water?

ISO INTERNATIONAL STANDARD. Pulps Determination of drainability Part 1: Schopper-Riegler method

Transferpette. Lightweight precise rugged reliable!

Installation & Operation Manual Proven Quality since 1892

Air Eliminator Series

Transferpette. Lightweight precise rugged reliable!

ST 18 Aluminium Gas Model

PERFORM Operating Document

MANUFACTURING PROCEDURE SPECIFICATION INSPECTION TEST PLAN

OXYGEN MINERALE ABSOLUTE RANGE OWNER S MANUAL OXY MINERALE 600/1000 SERIES

HEATEC TEC-NOTE Publication No Revised

Eliminating Sources of Pipetting Error in the Forensic Laboratory David M. Epstein Director of Scientific Operations. Abstract.

(2 pts) Draw the line of best fit through the data and estimate the concentration of Fe in your sample solution.

Used with Dissolution Tester for Fraction and Flow Measurements

Series 8500 Expansion Compensators. Catalog 674H

Sterile Visual Flow Indicator

Transferpette -8/-12

PRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator.

AUTOMATIC FLOW CARTRIDGES. Competition Analysis Not All Automatics Are Created Equal

Spa Start Up & Maintenance

Comparison of the Effectiveness of Various Deaeration Techniques

Needle valve. Contents. User s Manual. (1) Be sure to read the following warranty clauses of our product 1. (2) General operating instructions 2

Improve Process Reliability

OD plug gauge. for bores and internal profiles from dia. 2,5 mm. OD plug gauge in open design. Special designs Accessories. OD plug gauge for various

Hi-Temp Spa Care Guide

TECH TIPS: ROPE DEFECTS

v /07/17 Page 1 of 18

ROTATING DISK VALVES INSTALLATION AND MAINTENANCE 1. SCOPE 3 2. INFORMATION ON USAGE 3 3. VALVE TYPES 3 4. OPERATORS 5 5. VALVE CONSTRUCTION 6

HOW TO NOT MEASURE GAS - ORIFICE. Dee Hummel. Targa Resources

28 Day Start Up Instructions July 2018

STANDARD OPERATING PROCEDURES

"HAND, WRIST AND FINGER SAFETY"

Transferpette. Testing Instructions (SOP) 1. Introduction. October 1998

FRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS

Transcription:

Dissolution Failure Investigation Ken Boda Dissolution Applications Engineer

Dissolution Failure Investigation Whenever a dissolution test fails to meet specifications, an investigation should be performed to determine the cause(s) of the failure. Failure can be due to: Man (analyst) Machine (dissolution apparatus) Method (SOPs) Materials (standards, buffers, dosage form, etc.)

Conducting an Investigation Investigations should be performed in a sequential process to ensure all potential sources of error are investigated Review data as a whole Perform investigation in reverse chronological order

Aerial View of the Data The Error itself can help narrow the focus of the investigation: Entire data set trending high/low 1 outlier High %CV Data which makes no sense

Data trending high/low Bad lot Calculation Error Standard Prep Error Improper Filtration System misalignment (overall) Poor evaporation control 120 100 80 60 40 20 0 1 2 3 4 5 6

1 Outlier Misalignment/Bad Component in Single Position Filter fell off Bad sample reading Transcription Error Bad individual dosage form Contamination Wrong volume poured 120 100 80 60 40 20 0 1 2 3 4 5 6

High % CV Sampling Inconsistencies Poor techniques Vibration Improper Degassing System Misalignment (tends to trend) Issues w/ formulation 120 100 80 60 40 20 0 1 2 3 4 5 6

Data that doesn t make sense Filtration Issues Bad Sample Readings Air bubble Contamination 120 100 80 60 40 20 0 1 2 3 4 5 6

Determinant vs. Non-Determinant Errors Determinant Errors are known and controllable Indeterminate errors are unknown, suspected, or beyond control Determinant errors when found will lead to a simpler investigation and require the least re-work Indeterminate errors cannot be proven, and the burden of retesting is often higher

Investigating in Reverse Last thing done, first investigated Limits scope of investigation, if determinant error found can stop Last things done can often allow salvaging of the run All samples/standards/etc. should be kept until results have been checked and approved or fully investigated

Investigation Track Calculations Analysis Sampling and Filtration Run Observations Media Prep and Sample Handling Pre-Run Checks Dissolution Unit Mechanicals SOP and Materials

Checking Calculation Errors Check for transcription errors Are the right values being used for: Standard Conc. Label Claim Dissolution Volume Verify calculations by hand Ensure proper calculations are being used

Calculations without Media Replacement T1 % Dissolved = ((((A1 x V) /As) x C) / W) x 100 T2 % Dissolved = (((((A2 x (V - R)) + (A1 x R)) /As) x C) / W) x 100 T3 % Dissolved = (((((A3 x (V (R x n))) + ((A1 + A2) x R)) /As) x C) / W) x 100 Where, A1 = the absorbance of the first sample pull A2 = the absorbance of the second sample pull An = the absorbance of the nth sample pull As = the average absorbance of the standard V = the initial media volume, ml C = the concentration of the standard, mg/ml W = the weight of the active, mg (label claim) R = the volume of media being withdrawn for a single sample pull, ml n = the number of the sample pull taken

Calculations with Media Replacement T1 % Dissolved = ((((A1 x V) /As) x C) / W) x 100 T2 % Dissolved = (((((A2 x V) + (A1 x R)) /As) x C) / W) x 100 T3 % Dissolved = (((((A3 x V) + ((A1 + A2) x R)) /As) x C) / W) x 100 Where, A1 = the absorbance of the first sample pull A2 = the absorbance of the second sample pull An = the absorbance of the nth sample pull As = the average absorbance of the standard V = the initial media volume, ml C = the concentration of the standard, mg/ml W = the weight of the active, mg (label claim) R = the volume of media being withdrawn for a single sample pull, ml n = the number of the sample pull taken

Check Original UV/HPLC readings Right Wavelength and/or LC conditions used? If sample/standard readings were on UV, check scan if available If sample/standard readings were on HPLC, review chromatogram for: Proper peak shape Non-parent peaks Contaminants

Reanalyze Samples/Standards If available, reanalyze samples in question Reanalyze standard Compare standard against a fresh standard to ensure original standard was prepared correctly Make sure dilutions and purity are properly factored Dilution error possible?

Standard Preparation Create new standard to compare against one used for run Ensure adjustments are made for moisture and purity Right pipet/flask used? Within expiration period?

Storage Were samples within expiration period? Covered? Stored Properly? Labeled?

Sampling and Filtration Sampling and Filtration are one of the most common causes of dissolution failures: USP Position USP Timing Filtration

USP Sampling Area Sampling takes practice and attention! All samples must be withdrawn: ½ way between top of paddle/basket and media No closer than 1cm from vessel wall Recommend not sampling within 1cm of spindle

Imagine A Donut

Sample Timing Must sample (and filter) within 2% of the timepoint or 15 minutes, whichever is less Paddles/baskets must still be stirring while sampling

Filtration Is the filter validated? Efficiency Leaching Adsorption When was the sample filtered? At time of sampling? Immediately after sampling? After time point collected?

Autosampling Autosampling can reduce many of the errors associated with manual sampling: Position constant Timing within 2% Consistent pull force Documentation

Dissolution Observations

Observations Observations of the dosage form are key in order to diagnose potential issues with the dosage form. Observations can tell why something happened, % dissolved issues can only tell you what happened.

% Dissolved When should you take observations? Ideally, observations should be taken at all critical timepoints of the run Wetting/Dose Dumping Brand to Brand Differences in Dissolution Profiles Release Profile Complete Release/Asymptote 110 100 90 80 70 Whenever sampling manually 60 50 40 Brand A Brand B Brand C 30 20 10 0 0 30 60 90 120 150 180 210 240 270 Time (minutes)

What to look for Initial Stages: Floating or Moving? Coning? Dispersion Profile? Dissolved Gasses? Cross-linking? Compare appearance of all dosage forms

Floating/Moving Dosage Forms If a dosage form is floating or moving, it adds additional sources of energy to the dosage form and increases variability Sinkers should be used when floating, spinning, dancing, etc. occurs Use validated sinker

Coning Cones should be loose Cone should be in consistent location (bottom center) Look for consistent behavior across vessels, any trending noticed? In cases of severe coning, method alteration may be needed Higher RPM Peak vessel

Dispersion Profile and Dissolved Gasses Are air bubbles on stirring element, vessel, dosage form? How are particles distributed in vessel (top, middle, and bottom) Are particles floating? Presence of air bubbles may indicate degassing method is not sufficient

Cross-linking Cross-linking of capsule shells can result in hardened and chemically resistant shells. Very common to see in stability testing. Delay opening Trap Drug Product Pellicle Formation If Cross-Linking is seen, testing with pepsin or pancreatin should be performed

Compare Results of all 6-8 Vessels Take observations as a complete set, just like data Write observations down Is there any atypical behavior? Trending of observations?

Observations During Release Profile Size of dosage form/particles Distribution of dosage form Consistency of the dosage forms Anything changed?

Observations at End of Run Any material left? Describe material May be beneficial to assay dosage form post-mortem in formulation development or investigation of strange data or observations

Media Preparation Degassing Procedure validated? Time of degassing vs. time of use Pouring Technique

Sample Handling When exposed to air? Weighed? Inspected? Where placed? How handled? Bare hands Gloves Tweezers

Instrument Checks Pre-run checks of the instrument are always recommended. Pre-run checks are actually required if following the FDA or ASTM Mechanical Qualification procedure. Recheck these same parameters in the event of a run failure

Dissolution Mechanical Qualification Standards Prior to each dissolution test, the guidelines generally agree, requiring documentation for: Vessel examination Paddle examination Baskets and basket shaft examination Vessel medium temperature measurement Vibration evaluation Water bath (USP) Speed measurement on single position (USP)

Dissolution Mechanical Qualification Standards Vessel Examination, must be free from: Scratches Cracks Pits Residue Surface irregularities Centering Device, should also be: Complete Tight to Vessel Plate

Vessel Examination

Dissolution Mechanical Qualification Standards Paddle Examination, must be free of: Defects Rusting Corrosion Peeling or loose coating Knicks, dents or misshapen appearance

Stainless Steel Paddles Minor Rusting

Stainless Steel Paddles Advanced Corrosion

Teflon Paddles Good, Okay, and Terrible

Dissolution Mechanical Qualification Standards Basket and Basket Shaft, must be free from: Defects Rusting Corrosion Loose wires Clogged mesh openings Dented sides or bottom Knicks, dents or misshapen appearance O-ring are not compliant; three clips required

Baskets From New to Dead

Basket Bad Example

Dissolution Mechanical Qualification Standards Other Requirements: Vessel Temperature: The temperature of the medium inside each vessel is measured at the time of use. Media temperature must be ±0.5ºC from target. Vibration: There can be no significant vibration in the dissolution apparatus or medium. Possible sources of vibration are the surrounding environment, the dissolution unit itself or one of its components or an external water bath circulating heater.

Perform Mechanical Checks on System If no issues w/ individual components is noted, investigation of the system s overall alignment and performance should be reviewed MQ tolerances, particularly height and centering can be susceptible to change If height/centering is out, can have major impacts on hydrodynamics

Dissolution Mechanical Qualification Specifications and Tolerances Parameter Basket and Paddle Depth Rotational Speed Shaft Wobble Shaft Verticality ICH Harmonized (USP, JP, EP) FDA DPA-LOP.002 ASTM E2503-07 25 ± 2 mm 25 ± 2 mm 25 ± 2 mm (or <8%) ± 4% of specified rate No significant wobble Not measured Basket Wobble ± 1 mm ± 2 rpm of target 1.0 mm total runout 0.5 from vertical (x and y 90º) 1.0 mm total runout lower rim ± 2 rpm or within 2% of target (larger) 1.0 mm total runout Within Bubble (x and y 90º) 1.0 mm total runout lower rim USP Toolkit Ver 2.0 23-27 mm ± 1 rpm of target 1.0 mm total wobble Not measured 1.0 mm total wobble

Dissolution Mechanical Qualification Specifications and Tolerances Parameter Vessel/Shaft Centering Vessel Verticality Vessel Plate Level Performance Verification Test (PVT) ICH Harmonized (USP, JP, EP) NMT 2 mm from center axis Not Measured FDA DPA-LOP.002 1.0 mm from center line (upper/lower) 1.0 from vertical (2-90 positions) ASTM E2503-07 1.0 mm from center line (upper/lower) 1.0 from vertical (2-90 positions) USP Toolkit Ver 2.0 NMT 2.0 mm difference (4-90 positions) NMT 0.5 from vertical Not Measured Not Measured Not Measured NMT 0.5 from horizontal USP Prednisone Tablets RS Not Measured Not Measured USP Prednisone Tablets RS

SOP investigation In the event a failure can not be diagnosed, and several failures have occurred, SOP/validation review may be needed: Filter choice Chemicals specified Is the method over-discriminatory? Was intermediate precision evaluated for: Chemist to chemist variability Manual vs. autosampling Resident vs. non-resident probes 3-clip vs. O-ring baskets?

SOP Investigation Carryover? Cleaning method established? Dissolution apparatus Sampling equipment

Investigation Conclusion In the event of a determinant error, failing run can be aborted and test re-run In the event of a non-determinant error Multiple runs may be needed Analyst re-training may be advisable

Special Thanks/Reference Dissolution Aberrant Data Investigation, Bryan Crist, Pharmaceutical Canada, March 2004, vol. 4, no. 4., pp. 7-12

Questions? Ken.Boda@Agilent.com 919-677-6797 Dissolution.Hotline@Agilent.com 1-888-826-5351 Dissolution Discussion Group www.dissolution.com

Upcoming Events http://www.chem.agilent.com/en-us/promotions/pages/dissoseminar.aspx