Tips and Tricks of Injector Maintenance. Abby Folk Agilent GC Technical Support August 2009

Similar documents
Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port

Products Designed to Solve Nagging GC Problems

A Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction

Best Practice for Identifying Leaks in GC and GC/MS Systems

no peaks missing peaks too small peaks increasing retention decreasing retention

Agilent Inlet EPC Leak Test

Impact of Air Leaks on the Productivity of GC and GC/MS Systems

6890 GC Site Preparation

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

GC Accessories Inlet Liners Sky Inlet Liners Other Deactivations Inlet Liner Accessories

Carrier Gases in Capillary GC

Instant Connect Helium Saver Injector Module

Agilent G Rebuilt Electron Capture Detector

Low-Pressure Retention Time Locking with the 7890A GC. Application. Authors. Abstract. Introduction HPI

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 7820MSD Series G3175A (G7020A), G3176A (G7021A)

Agilent G3180B Two-Way Splitter Kit

Recent Advances to Ensure Simple, Leak Free GC Column Connections. Ken Lynam R&D/Application Chemist Agilent Consumables and Supplies Division

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer

Carrier Gases in Capillary GC

Gas Clean. Filters. Delivering Clean Gases for GC and GC/MS Operation

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

Converting Helium Carrier Gas GC Methods Nitrogen and Hydrogen

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

A New GC/MS System Designed for Helium Carrier Gas Conservation

3 Flow and Pressure Control

Gas Lines. Technical Manual. Sumitomo (SHI) Cryogenics of America, Inc Vultee Street Allentown, PA U.S.A.

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

Rapid and Reliable Detection of Dissolved Gases in Water

5890II GC Standard Operating Procedure 9/2/2005

Capillary Conversion System CCS-4/NC & CCS-05/NC. Installation and Operating Instructions. Operating Instructions

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

Customer Responsibilities. Important Customer Information GCMS QQQ Site Preparation Checklist

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 5975C MSD Series G3170A, G3171A, G3172A, G3174A

CP-3800 GC Getting Started Manual

Agilent G3185B QuickSwap Accessory

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

MicroJet Cryo-Trap MODEL MJT-1030Ex OPERATION MANUAL

Experiment GC : Analysis of BTEX by GC-FID

Advances in Capillary Column Backflushing

USER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Agilent 7890 Series Gas Chromatograph

instrumentation Pulsed Discharge Detectors Non-Radioactive, Multiple Mode Electron Capture / Helium Photoionization

3-Way Decoking Valve Option - Installation and Operation Instructions Accessory G Background

Agilent 7697A Headspace Sampler

Retention Time Locking: Concepts and Applications. Application

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

STANDARD OPERATING PROCEDURES

Agilent 7890 / 220 Ion Trap GCMS Site Preparation Checklist

Agilent 5975T LTM GC/MSD

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

SRI Model H2-40 Hydrogen Generator Operation and Maintenance Nov 2006

GCMSD-Headspace Analysis SOP

2 Sentry MCL Installation, Operation & Maintenance

The Open-Split Capillary Interface. Part No

7820A GC Site Preparation Checklist

Mass Spec will not Autotune

Water Weir Flow Controller. Introduction. Safety Precautions. Mounting the Hardware

Connecting a SSL/PTV Backflush System

LRS(H)4 Pressure-Reducing Regulator User Manual

Generating Calibration Gas Standards

Tank Blanketing Pressure Regulators RHPS Series

P5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

Super-Clean Gas Filters from Restek

GC Accessories. Agilent Consumables. Agilent Instrument Specific Replacement Parts. SGE Replacement Parts by Instrument Vendor.

System Overview. TCD Detector temperature setpoint Regulator

2 Sentry MVS Installation, Operation & Maintenance

Zebron Gas Management

For additional information about our solutions, please visit our web site at US/Pages/HomePage.

GILMONT ACCUCAL FLOWMETERS

Superior Result from Analytical Instruments The constant purity of the nitrogen improves system stability and ensures reproducible results

Agilent 7667A Mini Thermal Desorber. Installation and Operation

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

Agilent 7820A GC. Site Preparation Guide. Agilent Technologies

Spray Gun 134-A. 1 Liter High- Pressure Gel Coat. Usage Instructions & Parts List

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

INSTRUMENT INSTRUMENT YL INSTRUMENT

Please refer to the Hydrogen Safety Guide which is shipped with the Instrument. 5977B Series GC/MS Systems Site Preparation Checklist

Agilent 6890N, Option 652 Oral Fluids Analyzer

Thermo Scientific Chromatography Well Plates

Thermo Scientific Chromatography Well Plates

Quantitative Analysis of Hydrocarbons by Gas Chromatography

6850 Series Control Module User Information

ms NoVent II Instruction Booklet

RD(H)20/25 Pressure-Reducing Regulator User Manual

Markes Thermal Desorption Systems Site Preparation Page 1 of 8

Running HPLC Methods on an ACQUITY UPLC System

BLAST CABINET FOOT PEDAL & GUN INSTRUCTIONS. Item #21133

6890 GC Not Ready Messages (A16010)

Advancements in Gas Chromatography Analyzers - Keeping up with New Technology. Chuck Runkle Gas Phase Product Specialist ASTS June 2013

Agilent 7697A Headspace Sampler

Installation, Operation, and Maintenance Manual. Welker Constant Pressure Cylinder With Welker Magnetic Indicator (With Gravity Mixer) Model CP2GM-HP

Transcription:

Tips and Tricks of Injector Maintenance Abby Folk Agilent GC Technical Support August 2009

Typical Gas Chromatographic System Mol-Sieve Traps Fixed Restrictors Have to be able to get things to their gas state Key to starting the Chromatographic process right. Regulators Injection Port Detector Electrometer Flow Controller Recorder/ Integrator Air Hydrog gen Carrier Gas Column Cylinders or Generators

Influence of Injection Efficiency Short Concentrated Solute Bands Long Diffuse Same column, same chromatographic conditions

Injectors Split Splitless

Split Injector Flow Path Carrier gas source Septum purge Split vent

Splitless Injector Purge Off At Injection Carrier gas source Septum purge Split vent Flow through injector = Column flow only

Splitless Injector Purge On After Injection Carrier gas source Septum purge Split vent Flow through injector = Column flow + Split Vent Flow

Split/Splitless Injector

So, Why Do I Have To Do Maintenance? 1. Things wear out septa, syringes, nuts, ferrules, o-rings, etc. 2. Things get dirty liners, column, gas lines, traps, etc.

Agilent s Gold Standard GC Autosampler Syringes New syringe Pub #5990-3563EN Improved Quality Improved Packaging More value! BUT They Th Don t Last FOREVER!!

Split/Splitless Injector Parts

Septum and Septum Nut Viton O-Ring P/N 5188-5365

Gold Seal & Washer Column Ferrule & Nut

Redesigned packaging: Agilent Convenience Dial packs Individual packaging Maintains cleanliness Easy access to one at a time Doesn t go flying all over the floor Stores nicely in drawer Inlet O-rings P/N 5188-5365 Inlet O-ring Flip Top 5188-5366

Leak Checking G3388A leak detector Avoids loss of gases, damage to column and detector Portable, handheld unit shown is lightweight only 310g/11oz Simple and easy to use Fast detection 1 second Audible and visual alerts for 12 gases Minimum detection limit of 0.01 ml/minute for hydrogen and helium One year warranty RoHS compliant Rechargeable NiMH battery with over 5 hours of life

Replace Because Most Common Causes of Leaks Gold Seal P/N 5188-53675367 Re-use and mis-installation. Leak from O-ring, Gold Seal, ferrules, column nuts O-rings are elastomer compression fittings designed for one use, not perfectly elastic. Gold seals are designed for one use, knife edge cuts into gold layer giving leak tight seal w/o shrinkage or potential organic contaminants from polyimide out-gassing/degradation. Re-using could result in overlap in seal rings, resulting in a leak. Over-tightening of fittings

Why Are There Different Gold Seals? High Split Flows Limited by flow controller Range usually 1:200 to 1:10001000 Gold plated inlet seal with cross,,p/n 5182-9652

Leaks Due to Septum Nut With repeated use, conical needle guide gets worn, out of round, and needs replacement as septum can begin to bulge out, especially with excessive tightening, Septa fail faster because needle is not guided with as much precision. Under or Over tightening tighten nut until c-clamp on top stops turning, then ½ to ¾ turn more. Non-Agilent septa may be too thin, too thick, or out of round like die-cut septa and may not seal as well. Use Environments that decrease lifetime, like using non-agilent Autosamplers (ours are precisely aligned), manual injection, larger gauge syringes Replace septum nut annually for peace of mind.

Tips to Maximize Septum Life, Minimize Septum Leaks Use Agilent Gold Standard, HP Point, 23-26 gauge taper syringes. The point style cores septa significantly less when used with CenterGuide Septa. Taper minimizes septum coring/wear. HP-Point Style Use Agilent CenterGuide Septa. The molded hole minimizes septa coring, counter-intuitive, but true. Solid Septum CenterGuide Septum

Tips to Maximize Septum Life, Minimize Septum Leaks Use Non-Stick septa, especially Agilent s Centerguide Septa with Proprietary Plasma Treatment Their s Tl Talcum Powder! Our s Stuck septa particles can cause sealing problems on next septum installation. Talc can cause activity/trap plugging problems

Tips to Maximize Septum Life, Minimize Septum Leaks Other Benefits of Agilent s Centerguide Septa Packaging eliminates contamination of septa, first is as good as the last Less Strain on Syringe compared to solid septa Bleed/Temperature Optimized, (to 400C, trace analysis), p/n 5183-4757 Advanced Green, (to 350C, good for general purpose), p/n 5183-4759 Long Life, (to 350C, more injections before failure), p/n 5183-4761 Above are 50 packs, 100 packs also available.

Septa vs GC Column Costs Typical cost of 1 Premium Septum (list), $1.25 Typical cost of 1 GC Column, 30 m x 0.25 mm ID, $450. No accurate leak rate detector at sub 1 ml/min flow rates. Don t step over a dollar to pick up a dime! Proactively change inlet septa.

Or Go Septumless! Merlin Microseal Low bleed, longer life alternative to standard septa for split/splitless injection More than 2000 injections, depending on samples and operating conditions Almost zero downtime for septa changes and injection port liner changes due to septa particulates Double O-ring type seal around the syringe needle Spring assisted duckbill to seal the injection port

So, Why Do I Have To Do Maintenance? 1. Things wear out septa, syringes, nuts, ferrules, o-rings, etc. 2. Things get dirty liners, column, gas lines, traps, etc.

The BIGGEST Problem in GC is There are more things that DON T go through a GC than DO!.therefore, don t inject anything and you ll never have problems. OK, inject, but realize that t everything just got dirty deal d with it!

Where Does it Get Dirty? Here Here Here Here Here

Liner Maintenance Liners become contaminated with use, collecting non-volatiles, salts, excess reagents, etc., or become damaged/cracked. Should inspect and replace liners often. Handle with gloves and forceps. Insert into or remove liners only from cool injection ports. Replacing with a new liner is recommended, to ensure reproducibility

Split Liners What s What? Straight tube Straight tube with glass wool Fixed glass wool Inverted cup Baffle

Split Liner C 10 Packed with Glass Wool Peak Area Ratio n-c 40 /n-c 10 = 0.64 C 40 C 10 Without Glass Wool Packing Peak Area Ratio n-c 40 /n-c 10 = 037 0.37 C 40

Split Liner C 10 Packed with Glass Wool Peak Area Ratio n-c 40 /n-c 10 = 0.64 C 40 C 10 Without Glass Wool Packing Peak Area Ratio n-c 40 /n-c 10 = 037 0.37 C 40

Larger Plug of Glass Wool in the Liner C 8 C 40 10 20 30 10 20 30 Oven: 35 C for4min min, 35320 C 35-320 C at10 /min 10 /min, 320 C for5min Carrier Gas: Helium at 9.5 ml/min

GLASS WOOL Placement in Liner Near top of liner: Wipes syringe needle of sample Can improve injector precision Helps to prevent backflash Near bottom of liner: Helps in volatilization of high MW components Increases mixing

GLASS WOOL Liner Packing Recommendations Amount, size and placement must be consistent for consistent results Can be broken upon installation into the liner, exposing active sites Liner deactivation with glass wool plug in place is ideal

Splitless Injection Liners Liner Part No. Comments Side hole 5181-3316 5062-3587 5181-3315 G1544-80730 G1544-80700 Single taper, deactivated, 900μLvolume. Taper isolatessample sample from metal seal, reducing breakdown of compounds that are active with metals. For trace samples, general application. Single taper, deactivated, with glass wool, 900μL μ volume. Glass wool aides volatilization and protects column. For trace (dirty) samples. Double taper, deactivated, 800μL μ volume. Taper on inlet reduces chance for backflash into carrier gas lines. High efficiency liner for trace, active samples. Direct connect liners, single and dual taper, deactivated. Capillary column press fits into liner end, eliminating sample exposure to inlet. Ultimate protection for trace, active samples. Side hole permits use with EPC.

Do liner types really matter? They do, especially for active compounds like: phenols organic acids pesticides amines drugs of abuse, etc. Phenols for example in a separation of EPA method 8270 Phenols, for example.in a separation of EPA method 8270 compounds

Cool On-Column-FID Injection of 11 Analyte Test Mix From Improvements in the Agilent 6890/5973 GC/MSD System for Use with USEPA Method 8270, Agilent Application Note 5988-3072EN 1 N-Nitrosodimethylamine 7 Pentachlorophenol ISTD 1 Dichlorobenzene-d4 2 Aniline 8 Benzidine ISTD 2 Naphthalene-d8 3 2,4-Dinitrophenol 9 3,3-Dichlorobenzidine ISTD 3 Acenaphthene-d10 4 4-Nitrophenol 10 Benzo(b)fluoranthene ISTD 4 Phenanthrene-d10 5 4,6-Dinitro-2-methylphenol 11 Benzo(k)fluoranthene ISTD 5 Chrysene-d12 6 4-Aminobiphenyl ISTD 6 Perylene-d12 pa 1600 2 ISTD 5, 9 10,11 1400 1200 1000 800 ISTD 2 ISTD 1 1 ISTD 3 600 5 3 4 6 ISTD 4 7 8 ISTD 6 400 200 0 4 6 8 10 12 14 16 18

Splitless Inlet Liners Tested 5062-3587 Single-taper, deactivated, with glass wool 5181-3316 Single-taper, deactivated (open top) 5181-3315 Dual-taper, deactivated (closed top) G1544-80730 Direct Connect, single-taper, deactivated G1544-80700 Direct Connect, Dual-taper, deactivated Vendor X Unknown proprietary deactivation Hole for EPC

Liner Comparison 2,4-Dinitrophenol Response Factors 160 80 20 5 ng injected Experimental: Agilent 6890 with FID Column = HP-5MS 30m x 0.25mm x 0.5μm Compared COC to various liners 0.75 min Splitless time, 3mL/min column flow Oven: Temp programmed per 8270 method Inj. 250 C, Det. 300 C, Sample: 1μL 8270 mix Respo onse Factor 1.400 1.200 1.000 0.800 0.600 0.400 Avg RF (RSD) 0.191 (22%) 0.167 (38%) 0072 0.072 0.170 0.197 0.125 0.214 0.218 0.260 (12%) 0.216 0.253 0.284 0.280 (18%) 0.207 0.280 0.303 0.310 (7%) 0.263 (9%) 0.278 0.229 0.316 0.261 0.318 0.277 0.325 (3%) 0.311 0.332 0.328 0.200 0.228 0.207 0.287 0.328 0.285 0.328 0.331 0.000 5062-3587 5181-3316 5181-3315 Vendor X G1544-80730 G1544-80700 Single-taper Single-taper Dual-taper Direct Connect Direct Connect (with wool) (no wool) Single-taper Dual-taper Cool On- Column PittCon 2006

DIRECT CONNECT Vaporization Liner Side hole for EPC pressure feedback Pulled back 1 mm for flow

Pressure Flow Calculator http://www.chem.agilent.com/en-us/support/downloads/utilities/pages/gcpressureflow.aspx

For Easy Liner Maintenance on 5890/6890/6850 Flip Top for Split/Splitless litl injection ports 30 sec liner change out No more hunting for that funny looking wrench! Saves fingers from getting burned Increases instrument up time For a limited time only Send an email to abby_folk@agilent.com before August 31 st and get your Flip Top for $180 (a 64% discount)

Turn Top Inlet Sealing System on NEW 7890 - Fast/Easy Split/Splitless Inlet Maintenance changing liners has never been easier

Common Care and Maintenance Scheme for GC Columns 1. Bake out the column for no more than 2 hours. 2. Cut off 6-1ft of the inlet end of the column. 3. Cut off more column. (repeat as necessary)

Cleaning the Split/Splitless Injector Caution Carrier gas flow off 38-caliber brass gun barrel brush Disconnect split vent line Replace split vent trap Injector body GC Off Remove column, reducing nut, gold seal, washer and liner MeCl 2 Ace tone

Cleaning the Split/Splitless Injector Caution Carrier gas flow off Disconnect split vent line Replace split vent trap Injector body GC Off Remove column, reducing nut, gold seal, washer and liner MeCl 2 Ace tone

Finding the Split Vent Trap Follow the split vent line back to the EPC

Finding the Split Vent Trap Remove cover at Split Vent

Replacing the Split Vent Trap Finger Tight Knurled Nut G1544-80530

Be Careful When Doing Maintenance You may be the CONTAMINATOR!

Contamination of system by residue on fingers during column installation Column: Carrier: Injector: Detector: Oven: DB-5ms, 30m x 0.25mm, 0.25um H2, 60 cm/sec, constant flow split 1:20, 250 o C FID, 320 o C, N2 makeup gas 40 o C for 0.75 min, 40-325 o C at 20 o C/min, 325 o C for 30 min Red: French Fry Blue: system blank Procedure: (1) Held French fry for 5 seconds. (2) Fingertip was wiped with paper towel to remove as much of the offending material as possible. (3) Lightly touched the part of the column sticking up above the ferrule. (4) Installed column into injector. (5) Set oven temperature to 40 o C. (6) Started oven temperature program as soon as oven reached 40 o C.

Contamination from Liquid Soap Column: Carrier: Injector: Detector: Oven: DB-5ms, 30m x 0.25mm, 0.25um H2, 60 cm/sec, constant flow split 1:20, 250C FID, 320C, N2 makeup gas 40C for 0.75 min, 40-325C at 20C/min, 325C for 30 min Red: Liquid Soap Blue: system blank Procedure: (1) One very small drop of liquid soap placed on one fingertip. (2) Fingertip was wiped with paper towel to remove as much of the offending material as possible. (3) Lightly touched the part of the column sticking up above the ferrule. (4) Installed column into injector. (5) Set oven temperature to 40C. (6) Started oven temperature program as soon as oven reached 40C.

Contamination from Hand Lotion Column: Carrier: Injector: Detector: Oven: DB-5ms, 30m x 0.25mm, 0.25um H2, 60 cm/sec, constant flow split 1:20, 250C FID, 320C, N2 makeup gas 40C for 0.75 min, 40-325C at 20C/min, 325C for 30 min Red: Hand Lotion Blue: system blank Procedure: (1) One very small drop of hand lotion placed on one fingertip. (2) Fingertip was wiped with paper towel to remove as much of the offending material as possible. (3) Lightly touched the part of the column sticking up above the ferrule. (4) Installed column into injector. (5) Set oven temperature to 40C. (6) Started oven temperature program as soon as oven reached 40C.

Contamination from Lubricant Column: Carrier: Injector: Detector: Oven: DB-5ms, 30m x 0.25mm, 0.25um H2, 60 cm/sec, constant flow split 1:20, 250C FID, 320C, N2 makeup gas 40C for 0.75 min, 40-325C at 20C/min, 325C for 30 min Red: WD-40 Blue: system blank Procedure: (1) One very small drop of WD-40 liquid placed on one fingertip. (2) Fingertip was wiped with paper towel to remove as much of the offending material as possible. (3) Lightly touched the part of the column sticking up above the ferrule. (4) Installed column into injector. (5) Set oven temperature to 40C. (6) Started oven temperature program as soon as oven reached 40C.

Root Causes of Inlet Performance Degradation, and Consequences Accumulation of Sample Residues Loss of response, tailing on active analytes, split vent trap fouling and inaccurate EPC flow control Accumulation of consumables wear particles Same as above, plus bleed peaks Leak in Septum Nut, Septum Damage to O 2 sensitive detectors, t irreversible ibl damage to column Non-Optimized Set-up O-ring, Gold Seal, Ferrules, Column Nuts Faster inlet performance degradation between maintenance sessions

Agilent J&W Scientific Technical Support 800-227-9770 (phone: US & Canada) * Select option 3, then 3, then 1. www.agilent.com/chem gc-column-support@agilent.com