Oxygen Dipping Probe PSt3 / PSt6 / PSt9

Similar documents
PreSens. Manual OxoPlate. OxoPlate OP96C. PreSens Precision Sensing

OXYMINI. Fiber-optic Oxygen Meter. INSTRUCTION MANUAL. Serial No. World Precision Instruments

DISSOLVED OXYGEN SENSOR BT34i

PH01 Perfusion Cannula Manual

Dissolved Oxygen Meter. ino2. Instruction Manual. Serial No..

PRODUCT SHEET. Order probe only as RXPROBE02

OXYGEN MICROOPTODE - OPTO SERIES USER MANUAL

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

NEULOG OXYGEN LOGGER SENSOR GUIDE

STERILIZABLE DISSOLVED OXYGEN SENSORS

Instruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text.

Operating manual. TA 197 Oxi. Dissolved Oxygen Depth Armature

STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE

Please read these instructions carefully before use! Please consider the safety instructions! Please keep for future reference!

Oxygen Sensor with Electrode

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

Evaluation copy. Interdependence of Plants and Animals. computer OBJECTIVES MATERIALS

Technical Data Sheet MF010-O-LC

Oxygen Measurement Systems High Reliability and Wide Application Coverage

The Oxygen Sensor can be connected to all types of einstein Tablets, einstein LabMate, and einstein LabMate +.

Membrane covered optical CO 2 sensor

Dissolved Oxygen/Temp Waterproof Tester

INSTRUCTION MANUAL. 840 Transmitter. Loop Powered Dissolved Oxygen Transmitter

Instruction Manual HI Portable Waterproof Dissolved Oxygen Meter.

Oxymax H COS21. Technical Information. Sensor with long-term stability for frequent sterilization and autoclaving

Dextens Thermo Conductivity Sensors. Type 52101, 54101, 55101

Dissolved Oxygen and measurement possibilities. Berno Lüpkes, 15 th March 2017

Piccolo 2 FIBER-OPTIC OXYGEN METER USER MANUAL

Optical Dissolved Oxygen Meter

Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter

HI 2314 HI 2315 HI 23151

BI-680 Online Dissolved Oxygen Controller Instruction Manual

Generating Calibration Gas Standards

Vinmetrica Dissolved Oxygen System User Manual

SS32L Dissolved O 2 Probe

Oxymax H COS21D. Technical Information

deltaflowc deltaflowc Venturi or Probe

12mm Dissolved Oxygen Sensor Care and Use Instructions

Aerobic Respiration. Evaluation copy

Introduction The Smart Q Oxygen Adaptor enables an Oxygen electrode to be connected to an EasySense unit.

Optical Dissolved Oxygen Meter

Gas Sampling Fitting with MK2.1 and L-Probe

Dissolved Oxygen Guide

FTC130 Transmitter. Operating Manual

EC214 EC215 - EC215R Bench Conductivity Meters

Radnoti Liver Perfusion System

Relative Humidity Calibration Kit

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

LogIT Dissolved Oxygen Probe

Best Practice for Calibrating LTH Conductivity Instruments

MULTICHANNEL. Labnet International, Inc.

Dissolved Oxygen OPERATING INSTRUCTIONS

Real-time gas analysis and software for the management of bioprocessing

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

LDO II AQ 2/2014, Edition 1.5

BioAerosol Nebulizing Generator. Operation and Maintenance User Manual

COM280. Technical Information. Oxygen measurement Battery-driven pocket meter for dissolved oxygen measurement

Transmitter CS 21 Operation Manual

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

Transmitter CS 21 Operation Manual

STARLOG. Capacitive Water Depth Probe

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars.

HUMOR 20. High Accuracy Humidity Calibrator. Water T, p1 HUMOR 20 V1.0

METEOR LSR36500 LIGHTING UNDERWATER POOL LIGHT. Operating Instructions

FAST EVAPORATION AT NORMAL PRESSURE

Register your instrument! Eppendorf Xplorer / Eppendorf Xplorer plus. Adjustment

MEASUREMENT OF TOTAL PACKAGE OXYGEN (TPO) AT BOTTLING

VWR UHP j :01 Page 1. ENGLISH 1 22 DEUTSCH FRANÇAIS ESPAπOL ITALIANO

Dr. Thiedig DIGOX 5 CO 2. In-line measurement of CO 2 for constant product quality. Anlagen- und Analysentechnik

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

Improve Process Reliability

QUICK START GUIDE TO DRINKING WATER MONITORING

Simple, Robust, Reliable Measures ph Doses either acid (lower) or alkali (raise) Displays ph, Total Dose Count (DCT) and Dose Count per hour (DC/h)

FTS SUBMERSIBLE PRESSURE TRANSMITTER USER S MANUAL

Dissolved Oxygen Sensor

DTG - LCD Digital Temperature Gauge USER MANUAL

E2K-L. Liquid Level Sensor That Is Unaffected by the Color of the Pipe or Liquid. Liquid Level Sensor. Ordering Information

Gas Pressure Probe. Information Sheet. Total Dissolved Gas

UsER manual for Watersens ph -REDOX

O3 3E 1 F Gas Sensor Module

OPERATING INSTRUCTIONS

dtrans O2 01 Two-wire Transmitter for dissolved oxygen (DO)

Oxymax W COS41. Technical Information. Potentiostatic amperometric two-electrode sensor Long-term stable sensor for water and wastewater applications

PERFORM Operating Document

OxiQuant S. Operating Instructions

OxyProbe II Instruction Manual

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Rev. 1. Cryogenics Thermoflasks. 2122, 2123, 2124, 2129, and 2130

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

LK-SX CO 2 +VOC. Application. Security Advice Caution. Notes on Disposal

Tips for Proper Pipette handling and Maintenance. D. Muruganand, Ph.D ILQA - Bangalore, Feb 2014

TJF-Q180V Cleaning and Disinfection Checklist

Instruction Manual HI Portable Dissolved Oxygen Meter. DO meter HI 8043 STB C OFF O 2 ZERO SLOPE C. ins rum n s

CCS240 and CCS241. Technical Information. Sensors for chlorine dioxide Amperometric, membrane-covered sensors for installation in the CCA250 assembly

OPERATING MANUAL CHEMTRAC 840M MANUAL RETRACTABLE PROCESS HOLDER

Transferpette. Testing Instructions (SOP) 1. Introduction. May 2009

LK-SX VOC. Application. Security Advice Caution. Notes on Disposal. Combined sensor mixed gas. Data sheet

DEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111

Transcription:

Oxygen Dipping Probe PSt3 / PSt6 / PSt9 SENSOR PROBES Instruction Manual

Oxygen Dipping Probe PSt3 / PSt6 / PSt9 Specification: Oxygen dipping probe with high grade stainless steel fitting for normal / trace / ultra-low oxygen range Document filename: IM_DP-PSt3_PSt6_PSt9_dv2 All rights reserved. No parts of this work may be reproduced in any form or by any means - graphic, electronic, or mechanical, including photocopying, recording, taping, or information storage and retrieval systems - without the written permission of the publisher. Products that are referred to in this document may be either trademarks and/or registered trademarks of the respective owners. The publisher and the author make no claim to these trademarks. While every precaution has been taken in the preparation of this document, the publisher and the author assume no responsibility for errors or omissions, or for damages resulting from the use of information contained in this document or from the use of programs and source code that may accompany it. In no event shall the publisher and the author be liable for any loss of profit or any other commercial damage caused or alleged to have been caused directly or indirectly by this document. Specifications may change without prior notice. Manufacturer PreSens Precision Sensing GmbH Am BioPark 11 93053 Regensburg, Germany Phone +49 941 94272100 Fax +49 941 94272111 info@presens.de.de

Table of Contents 1 Preface... 7 2 Safety Notes... 8 3 Description of the Oxygen Dipping Probe... 9 3.1 Scope of Delivery... 10 3.2 Measurement Set-up... 11 3.2.1 Underwater Applications... 11 3.2.2 Gas Phase Applications... 12 4 Operation... 13 4.1 Calibration... 13 4.1.1 Calibration of a DP-PSt3 Preparation of Calibration Standards... 13 4.1.2 Calibration of a DP-PSt6 Preparation of Calibration Standards... 15 4.1.3 Calibration of a DP-PSt9 Preparation of Calibration Standards... 16 4.2 Performance Proof... 17 5 Technical Data... 18 6 Concluding Remarks... 19

7 Preface 1 Preface You have chosen a new, innovative technology for measuring oxygen. Chemical optical sensors (also called optodes) have several important features: They are small. Their signal does not depend on the flow rate of the sample. They can be physically divided from the measuring system which allows a noninvasive measurement. They can be used in disposables. Therefore, they are ideally suited for the examination of small sample volumes, for highly parallelized measurements in disposables, and for biotechnological applications. A set of different minisensors, flow-through cells and integrated sensor systems is available to make sure you have the sensor which matches your application. Please feel free to contact our service team to find the best solution for your application. Your PreSens Team PLEASE READ THE FOLLOWING INSTRUCTIONS CAREFULLY BEFORE WORKING WITH THIS ITEM.

8 Safety Notes 2 Safety Notes! It is the customer s responsibility to validate the sensor under end-user conditions according to safety precautions of the application to ensure that the use of the sensor is safe and suitable for the intended purpose. PreSens is explicitly not liable for direct or indirect losses caused by the application of these sensors. In particular it has to be considered that malfunctions can occur due to the naturally limited lifetime of the sensor depending on the respective application. The set-up of backup measurement stations is recommended when using the sensors in critical applications to avoid consequential losses. It is the customer s responsibility to install a suitable safety system in the event of sensor failure.

9 Description of the Oxygen Dipping Probe 3 Description of the Oxygen Dipping Probe Oxygen dipping probes measure the partial pressure of both gaseous and dissolved oxygen. They consist of a polymer optical fiber (POF) with a polished distal tip which is coated with a planar oxygen sensitive foil. Oxygen dipping probes are available with sensor coatings type PSt3 (detection limit 15 ppb, 0-100 % oxygen), type PSt6 (detection limit 1 ppb, 0-4.2 % oxygen), and PSt9 (detection limit 0.5 ppm, 0-200 ppm). The end of the polymer optical fiber is covered with a high-grade steel tube to protect both the sensor material and the POF. The cable has an outer diameter of 2.8 mm. The inner diameter of the POF is 2.0 mm. The steel tube has an outer diameter of 4 mm. Usually the fiber is coated with an optical isolated sensor material in order to exclude ambient light from the fiber tip and to increase chemical resistance especially against oily samples as well as to reduce biofouling on the sensor membrane. This type of oxygen sensor has excellent long-term stability. Fig. 1 Oxygen dipping probe; insert: enlarged view of sensor coating Fig. 2 Schematic illustration of an oxygen dipping probe

10 Description of the Oxygen Dipping Probe Dipping probe features: No polarization necessary No membrane cleaning and replacement necessary No electrolyte solutions to poison or replenish Long shelf-life Pressure resistant up to 3 bar 3.1 Scope of Delivery The oxygen dipping probe is delivered with a protective rubber cap on the distal end of the steel fitting to protect the sensor coating.! Please remove the protective rubber cap before inserting the dipping probe in your sample. Additionally required equipment (not supplied): Fiber optic oxygen transmitter Fibox 4 / Fibox 4 trace (more oxygen transmitters can be found on www.presens.de/products/o2/meters.html) PC / Notebook

11 Description of the Oxygen Dipping Probe 3.2 Measurement Set-up Fig. 3 Set-up for measurement with an oxygen dipping probe Connect the dipping probe s polymer optical fiber to the sensor connector of the fiber optic oxygen meter (e. g. Fibox 4 trace). Then remove the protective rubber cap from the steel fitting and insert the dipping probe in your sample. 3.2.1 Underwater Applications For underwater applications specially designed weights for dipping probes are available, which keep the probe in a stable, fixed position while it is submerged. For more information please see the Sensor Probe Accessories instruction manual (IM_SP_Acc). Fig. 4 Dipping Probe with attached DP-Weight for under water use.

12 Description of the Oxygen Dipping Probe 3.2.2 Gas Phase Applications For applications in the gas phase, especially when working with gas pipelines the oxygen dipping probe can be integrated in a Swagelok T-cell. The metal T-connector with oxygen probe (FTCM) can easily be integrate in pipes with o. d. 6 mm (or other diameter, when using the respective adapter). Gases (or liquids) can be pumped through the cell and precise oxygen measurements conducted. Fig. 5 Dipping probe (top) integrated in Swagelok T-cell (FTCM)

13 Operation 4 Operation! Please note that oxygen dipping probes (DP) are NOT autoclavable unlike the oxygen probes for inline measurements (OIM) as the polymer optical fiber does not stand high temperatures. Maximum temperature for a standard dipping probe is 80 C. However, a high temperature version is available which stands temperatures up to 110 C. In case the dipping probe must be exposed to temperatures > 80 C please contact our service team. 4.1 Calibration The oxygen dipping probe has to be calibrated before use. A conventional two-point calibration in oxygen-free environment (nitrogen / CO 2, or sodium sulfite), and air saturated (sensor type PSt3), 1 2 % oxygen (sensor type PSt6), or 100 200 ppmv oxygen (sensor type PSt9) environment has to be performed. A recalibration is recommended after 100,000 measurement points or a period of 3 months, whichever is reached earlier. Please see the respective transmitter instruction manual for more detailed information about software settings and calibration procedure. 4.1.1 Calibration of a DP-PSt3 Preparation of Calibration Standards 1st Calibration Point: In Liquid: Oxygen-free water To prepare oxygen-free water dissolve 1 g of sodium sulfite (Na 2SO 3) and 50 µl cobalt nitrate (Co(NO 3) 2) standard solution (ρ(co) = 1000 mg/l; in nitric acid 0.5 mol/l) in 100 ml water. Use a suitable vessel with a tightly fitting screw top and label it cal 0. Make sure there is only little headspace in your vessel. Due to a chemical reaction of oxygen with the Na 2SO 3 the water becomes oxygen-free. Additional oxygen, diffusing from air into the water, is removed by surplus Na 2SO 3. Close the vessel with the screw top and shake it for approximately one minute to dissolve Na 2SO 3 and to ensure that the water is oxygen-free. To prepare oxygen-free water you also can use sodium dithionite (Na 2S 2O 4). Place the oxygen dipping probe in the vessel with cal 0. To minimize the response time, slightly stir the solution. Then follow the instructions in the respective transmitter manual for calibration. After recording the first calibration point remove the dipping

14 Operation probe from the calibration solution cal 0, and rinse it with distilled water. Repeat this procedure at least 5 times to clean the probe from sodium sulfite. For storing the calibration solution cal 0 keep the vessel closed after calibration with a screw top to minimize oxygen contamination. The shelf life of cal 0 is about 24 hours provided that the vessel has been closed with the screw top. In Humid Gases: Nitrogen-saturated atmosphere Use nitrogen-saturated atmosphere as calibration standard cal 0. Use a commercially available test gas N 2(5.0) (suppliers are e. g. Air Liquide, Linde Westfalen AG). Lead the gas into a vessel filled with distilled water before feeding it into the calibration vessel with the dipping probe (see Fig. 6). Then follow the instructions in the respective transmitter manual. In Dry Gases: Nitrogen-saturated atmosphere Use nitrogen-saturated atmosphere as calibration standard cal 0. Use a certified test gas N 2(5.0) (suppliers are e. g. Air Liquide, Linde Westfalen AG). Lead the gas directly into the calibration vessel with the dipping probe in it (see Fig. 7). Then follow the instructions in the respective transmitter manual. 2nd Calibration Point: In Liquid: Air-saturated water Add 100 ml water to a suitable vessel and label it cal 100. To obtain air-saturated water, blow air into the water using an air-pump with a glass-frit (air stone), creating a multitude of small air bubbles, while stirring the solution. After 20 minutes, switch of the air-pump and stir the solution for another 10 minutes to ensure that the water is not supersaturated. Place the oxygen dipping probe in the vessel with cal 100. To minimize the response time, slightly stir the solution. Then follow the instructions in the respective transmitter manual for calibration. In Humid Gases: Water vapor-saturated air or 20.9 % O 2 Use water-vapor saturated air as calibration standard cal 100. Place wet cotton wool in a vessel with a fitting screw top or lid. For inserting dipping probe (and the temperature sensor) into the vessel you might have to drill holes in the lid. Introduce the dipping probe and wait about 2 minutes to ensure that the air is water vapor-saturated. Alternatively, you can use a certified test gas O 2(20.9) (suppliers are e. g. Air Liquide, Linde Westfalen AG). Lead the gas into a vessel filled with distilled water before

15 Operation feeding it into the calibration vessel with the dipping probe (see Fig. 6). Then follow the instructions in the respective transmitter manual for calibration. In Dry Gases: 20.9 % O 2 Leave the dipping probe in ambient air (or certified test gas O 2(20.9); suppliers are e. g. Air Liquide, Linde Westfalen AG). Then follow the instructions in the respective transmitter manual. 4.1.2 Calibration of a DP-PSt6 Preparation of Calibration Standards Calibration of a DP-PSt6 is performed with certified gases (suppliers are e. g. Air Liquide, Linde, Westfalen AG). 1st Calibration Point: In Humid Gases: Nitrogen- or CO 2-saturated atmosphere Use nitrogen or CO 2 (N 2 / CO 2 5.0) as a first calibration standard cal 0. Lead the gas into a vessel filled with distilled water before feeding it into the calibration vessel with the dipping probe (see Fig. 6). Then follow the instructions in the respective transmitter manual. In Dry Gases: Nitrogen- or CO 2-saturated atmosphere Use nitrogen or CO 2 (N 2 / CO 2 5.0) as a first calibration standard cal 0. Lead the gas directly into the calibration vessel with the dipping probe in it (see Fig. 7). Then follow the instructions in the respective transmitter manual. 2nd Calibration Point: In Humid Gases: 1 2 % O 2 (humidified gas) The second calibration value cal 2nd for a PSt6 sensor is ideally in the range between 5 and 10 % air sat. (ca. 1 2 % O 2 % air sat. = % O 2 x 100/20.95). Use a commercially available test gas of 1 2 % O 2 as a second calibration standard cal 2nd. Lead the gas into a vessel filled with distilled water before feeding it into the calibration vessel with the dipping probe (see Fig. 6). Then follow the instructions in the respective transmitter manual. In Dry Gases: 1 2 % O 2 Use a commercially available test gas of 1 2 % O 2 as a second calibration standard cal 2nd. Lead the gas directly into the calibration vessel with the dipping probe in it (see Fig. 7). Then follow the instructions in the respective transmitter manual.

16 Operation If it is not possible to use gases or to build a suitable calibration chamber, a manual calibration can be performed, using calibration values obtained from the final inspection protocol delivered with the dipping probe. You can find more information about manual calibration in the instruction manual of the respective transmitter. Fig. 6 Calibration set-up: Two-point calibration of a dipping probe with humid gases 4.1.3 Calibration of a DP-PSt9 Preparation of Calibration Standards A PSt9 sensor can only be calibrated with dry gases. Calibration is performed with certified gases (suppliers are e. g. Air Liquide, Linde, Westfalen AG). 1st Calibration Point: Use nitrogen-saturated atmosphere as calibration standard cal 0. Use a certified test gas N 2(6.0), and lead the gas directly into the calibration vessel with the dipping probe in it (see Fig. 7). Then follow the instructions in the respective transmitter manual. 2nd Calibration Point: Use a certified test gas of 100 200 ppmv O 2 as calibration standard cal 2nd, and lead the gas directly into the calibration vessel with the dipping probe in. Then follow the instructions in the respective transmitter manual. If it is not possible to use gases or to build a suitable calibration chamber, a manual calibration can be performed, using calibration values obtained from the final inspection

17 Operation protocol delivered with the dipping probe. You can find more information about manual calibration in the instruction manual of the respective transmitter. Fig. 7 Calibration set-up: Two-point calibration of a dipping probe with dry gases 4.2 Performance Proof If you want to prove the sensor performance during the past measurement, please check the calibration values by inserting the sensor in the cal0 and cal2nd calibration standards (see chapter 3.1) when you have finished your measurement. If the device shows 0% air saturation immersing the sensor tip into cal0 and the respective oxygen concentration for the cal2nd standard, the sensor worked perfectly throughout the whole measurement.

18 Technical Data 5 Technical Data Specifications Measurement range Gaseous & Dissolved O 2 0 100 % O 2 0 1000 hpa PSt3 PSt6 PSt9 Dissolved O 2 0 45 mg/l 0 1400 µmol/l Gaseous & Dissolved O 2 0 4.2 % O 2 0 41.4 µmol/l Dissolved O 2 Gaseous O 2 0 1.8 mg/l 0 56.9 µmol/l 0 200 ppm Limit of detection 0.03 % O 2 15 ppb 0.002 % O 2 1 ppb 0.5 ppm Resolution ± 0.01 % O 2 at 0.21 % O 2 ± 0.1 % O 2 at 20.9 % O 2 ± 0.1 hpa at 2 hpa ± 1 hpa at 207 hpa Accuracy ± 0.4 % O 2 at 20.9 % O 2 ± 0.05 % O 2 at 0.2 % O 2 Measurement temperature range Response time (t 90) Properties ± 0.14 µmol/l at 2.83 µmol/l ± 1.4 µmol/l at 238.1 µmol/l ± 0.0007 % O 2 at 0.002 % O 2 ± 0.0015 % O 2 at 0.2 % O 2 ± 0.007 hpa at 0.023 hpa ± 0.015 hpa at 2.0 hpa ± 0.010 µmol/l at 0.03 µmol/l ± 0.020 µmol/l at 2.8 µmol/l 10 ± 0.5 ppm 100 ± 0.8 ppm 200 ± 1.5 ppm ± 1 ppb or ± 3 %; whichever is higher ± 2 ppm or ± 5 %; whichever is higher 0 to +50 C 0 to +50 C 0 to +40 C < 6 sec. < 40 sec. < 6 sec. < 40 sec. < 3 sec. Compatibility Aqueous solutions, ethanol, methanol Gas phase only No crosssensitivity with Cross-sensitivity Sterilization procedure Cleaning procedure Calibration Storage Stability ph 1 14 CO 2, H 2S, SO 2 Ionic species Organic solvents, such as acetone, toluene, chloroform or methylene chloride Chlorine gas Ethylene oxide (EtO) Gamma irradiation 3 % H 2O 2 Acidic agents (HCl, H 2SO 4), max. 4 5 % at room temperature Two-point calibration with oxygen-free environment (nitrogen, sodium sulfite) and air-saturated environment Two-point calibration in oxygen-free environment (nitrogen) and a second calibration value optimally between 1 and 2 % oxygen 2 years provided the sensor material is stored in the dark (-10 to +60 C) CO 2, SO 2 Organic vapor, Chlorine gas - - Two-point calibration in oxygen-free environment (nitrogen 6.0) and a second calibration value optimally between 100 and 200 ppm gaseous oxygen

19 Concluding Remarks 6 Concluding Remarks Dear Customer, With this manual, we hope to provide you with an introduction to work with the oxygen dipping probes type PSt3, PSt6 and PSt9. This manual does not claim to be complete. We are endeavored to improve and supplement this version. We are looking forward to your critical review and to any suggestions you may have. You can find the latest version at.de. With best regards, Your PreSens Team

Manufacturer PreSens Precision Sensing GmbH Am BioPark 11 93053 Regensburg Germany Phone +49 941 94272100 Fax +49 941 94272111 info@presens.de.de