PH01 Perfusion Cannula Manual

Similar documents
User s Guide. CO2 Controller for miniature incubators

Operating Instructions Temperature Measuring System MF420-5T-100

User s Guide. CO2 & O2 Controller for miniature incubators

Warner Instruments, Inc Dixwell Avenue, Hamden, CT (800) / (203) (203) fax

User s Guide. Vacuum Controller for liquid delivery systems

User s Guide. Pressure Controller for liquid delivery systems

Technical Data Sheet MF010-O-LC

Measurement accessories METPOINT OCV for the measurement in systems up to 40 bar

Radnoti Liver Perfusion System

Instruction Manual Contact Pressure Vacuum Gauge

FTC130 Transmitter. Operating Manual

Agilent 1290 Infinity Pump Head Maintenance

Unisense In Situ UniAmp Connector System

URC Voltage Sensor SEN-VOLT for use with MRX units containing sensor ports

The Univentor 1250 Anaesthesia Unit

BPS-2, Rev TABLE OF CONTENTS

E8AA. Pressure Sensor of Stainless Steel Construction Is Ideal for a Wide Range of Applications. Pressure Sensor (Stainless Steel Diaphragm)

E2K-L. Liquid Level Sensor That Is Unaffected by the Color of the Pipe or Liquid. Liquid Level Sensor. Ordering Information

Gas Module for Series GMS800

92831 TEL: (714) FAX:

INSTALLATION MANUAL Matheson Tri-Gas Cabinet Enclosures

THE UNIVENTOR 400 ANAESTHESIA UNIT

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars.

Type BBS-03, BBS-05, BBS-06, BBS-25

STARLOG. Capacitive Water Depth Probe

PRAHER PLC-MP AQUASTAR MANAUL 2009

SERIES 20 CHAMBERS THE RC-25 AND RC-25F CHAMBERS ASSEMBLY

Operator s Manual. The Bullet Blender Gold BB24-AU, BB5E-AU

Manual. Online Degasser DDG-75. Manual Degasser DDG-75 April DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE Hockenheim Germany

E X R A D I N Spherical Ion Chambers A3. A4 REF DOC #

Precision Liquid Settlement Array Manual

MCQ Gas Blender 100 Series 3 Channels Gas Mixer

T EK-SUB 4800C 19 mm Submersible Level Transmitter

Safety instructions and operating manual

BSS 2 Bottle Sampling System. Instruction manual Version 1.8

BioAerosol Nebulizing Generator. Operation and Maintenance User Manual

LC-SPE. Organizer 3.0 User Manual. Innovation with Integrity. Version 001 NMR

03/07/2014 V

QE-1, Rev Warner Instruments, Inc Dixwell Avenue, Hamden, CT (800) / (203) (203) fax

User Manual. Quantos Automated Dosing Liquid Module

InFlow Modular perfusion bioreactor

OPERATING MANUAL CHEMTRAC 840M MANUAL RETRACTABLE PROCESS HOLDER

Safety Manual VEGAVIB series 60

Safety Manual. Process pressure transmitter IPT-1* 4 20 ma/hart. Process pressure transmitter IPT-1*

MASS-STREAM TM Instruction Manual. D-6200 Analog Mass Flow Meters / Controllers

SERIES 20 CHAMBERS THE RC-26, RC-26P AND RC-26G CHAMBERS ASSEMBLY

453 Series Steam Heated Vaporizing Regulator

Basic Flows in a Microfluidic Device

RayPette Plus Autoclavable Pipette. User Manual

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

Operating Instructions in accordance with the Machines Directive 2006/42/EC

MANUAL DIRECT PURGE OPTION. UNION Instruments GmbH CWD2005 PLUS. General information, safety standards and regulations for direct purge option

GasSense NDIR User Manual

PURA. Pure Gas Dewpoint Transmitter. Users Guide

User Manual MFCS TM -EZ

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

PISTOL CALIBER SUPPRESSORS

ROPV R40 E Series User Manual

Instruction Manual EMA 84 Digital Pressure Gauge

Oxygen Dipping Probe PSt3 / PSt6 / PSt9

Mass Flow Controller (MFC) for Gases

Installation and Operation Manual

INSTRUCTION MANUAL. January 23, 2003, Revision 0

RADROVER REAR RACK INSTALLATION MANUAL

The pipettes cover a volume range from 0.1µl to 10ml.

SAFEsystem. Safety Apparatus Filter Enhancement. Filtration Systems. Now you can safely control your filtering process

Linkam Scientific Instruments USER GUIDE DSC600. Temperature Controlled Stage

Model # HC-MIS INSTRUCTION MANUAL

Flow experiment with the ibidi pump system and µ-slide I 0.6 Luer

Preparation and Installation of the Sanitary BDI-FLX Sensor and Connection to the BDI-FLX Interface Cable

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

SPECIFICATIONS APCEPH1

2018 ACCESSORY ATTACHMENTS OWNER S MANUAL

User Manual. ECC-Opto-Gas. Release Electrochemical test cell EL-CELL GmbH

Transferpette. Testing Instructions (SOP) 1. Introduction. May 2009

DK46 - DK800 Supplementary instructions

Product information. Capacitive. Level detection in liquid VEGACAP 62 VEGACAP 63 VEGACAP 64 VEGACAP 66 VEGACAP 69. Document ID: 29983

Digital Melting Point Apparatus

Linkam Scientific Instruments USER GUIDE TST350. Temperature Controlled Tensile Stress Testing Stage

DTG - LCD Digital Temperature Gauge USER MANUAL

CSA Sample Draw Aspirator Adapter Operator s Manual

Pressure Dump Valve Service Kit for Series 3000 Units

BGA244 Binary Gas Analyzer

Installation and Maintenance Instruction Manual

*******WARNING******

Instruction Manual Differential Pressure Transmitter

MODEL NUMBER: AS18-2 USER GUIDE

RGC-IR Remote Gas Calibrator for IR400

Pipettor. User Manual

Operating instructions Safety Rope Emergency Stop Switches ZB0052 / ZB0053 ZB0072 / ZB0073

NovelBright LED Illuminator

Varipette Operating manual

NORMAL OPERATING PROCEDURES Operating Parameter Information

! Warning, refer to accompanying documents.

USER MANUAL. MACMAG magnetic separator MHK MHF

Media-Isolated Mag Plus Probe Low Pressure-ISO Kit

HI 2314 HI 2315 HI 23151

Application of vacuum grease

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION

Electronic gas volume corrector model DGVC-04

Transcription:

PH01 Perfusion Cannula Manual

Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted without the express written permission of Multi Channel Systems MCS GmbH. While every precaution has been taken in the preparation of this document, the publisher and the author assume no responsibility for errors or omissions, or for damages resulting from the use of information contained in this document or from the use of programs and source code that may accompany it. In no event shall the publisher and the author be liable for any loss of profit or any other commercial damage caused or alleged to have been caused directly or indirectly by this document. 2011 Multi Channel Systems MCS GmbH. All rights reserved. Printed: 05.11.2009 Multi Channel Systems MCS GmbH Aspenhaustrasse 21 72770 Reutlingen Germany Fon +49-7121-9 09 25-0 Fax +49-7121-9 09 25-11 info@multichannelsystems.com www.multichannelsystems.com Products that are referred to in this document may be either trademarks and/or registered trademarks of their respective holders and should be noted as such. The publisher and the author make no claim to these trademarks.

Table of Contents 1.1 Operator's Obligations 2 1.2 Guarantee and Liability 2 1.3 Important Safety Advice 3 2.1 Welcome to the PH01 Perfusion Cannula with Heating Element and Sensor 4 2.2 Setting Up and Connecting the PH01 5 2.3 Flow Rate Dependent Temperature Offset 6 2.4 Air Drafts and Ambient Temperature 7 2.5 Pump Types 7 2.6 Flow Stream 8 3.1 Pin Assignment 8 1

1 Important Information and Instructions 1.1 Operator's Obligations The operator is obliged to allow only persons to work on the device, who are familiar with the safety at work and accident prevention regulations and have been instructed how to use the device; are professionally qualified or have specialist knowledge and training and have received instruction in the use of the device; have read and understood the chapter on safety and the warning instructions in this manual and confirmed this with their signature. It must be monitored at regular intervals that the operating personnel are working safely. Personnel still undergoing training may only work on the device under the supervision of an experienced person. 1.2 Guarantee and Liability The General conditions of sale and delivery of Multi Channel System MCS GmbH always apply. The operator will receive these no later than on conclusion of the contract. Guarantee and liability claims in the event of injury or material damage are excluded when they are the result of one of the following. Improper use of the device Improper installation, commissioning, operation or maintenance of the device Operating the device when the safety and protective devices are defective and/or inoperable Non-observance of the instructions in the manual with regard to transport, storage, installation, commissioning, operation or maintenance of the device Unauthorized structural alterations to the device Unauthorized modifications to the system settings Inadequate monitoring of device components subject to wear Improperly executed and unauthorized repairs Unauthorized opening of the device or its components Catastrophic events due to the effect of foreign bodies or Acts of God 2

1.3 Important Safety Advice Warning: Make sure to read the following advice prior to install or to use the device and the software. If you do not fulfill all requirements stated below, this may lead to malfunctions or breakage of connected hardware, or even fatal injuries. Warning: Obey always the rules of local regulations and laws. Only qualified personnel should be allowed to perform laboratory work. Work according to good laboratory practice to obtain best results and to minimize risks. The product has been built to the state of the art and in accordance with recognized safety engineering rules. The device may only be used for its intended purpose; be used when in a perfect condition. Improper use could lead to serious, even fatal injuries to the user or third parties and damage to the device itself or other material damage. Warning: The device and the software are not intended for medical uses and must not be used on humans. Malfunctions which could impair safety should be rectified immediately. Requirements for the installation and operation The heating element produces heat and can get hot during operation. Do not touch the perfusion cannula during operation. Do not use the perfusion cannula for perfusion with flammable or aggressive (corrosive) liquids. Do not store flammable materials nearby during operation. Check in regular intervals that the perfusion cannula does not overheat. 3

2 Installation and Operation 2.1 Welcome to the PH01 Perfusion Cannula with Heating Element and Sensor Cell lines and tissue slices are placed and cultivated directly in the MEA culture chamber, on the microelectrode array. All MEA systems can be upgraded with an enhanced perfusion system (indicated by the code E), featuring a perfusion cannula PH01 with programmable fluid temperature and a two-channel temperature controller TC02 for controlling both the MEA culture chamber temperature and the fluid temperature in parallel. A perfusion system is required especially for recordings from acute slices. The perfusion cannula with heating element and sensor PH01 and a temperature controller TC01/TC02 is also useful for general perfusion purposes, whenever the fluid stream needs to be temperature controlled. The PH01 is to be used only for the temperature-control of non-corrosive liquids up to a temperature of 50 C. The programmable fluid temperature is controlled by a temperature controller TC01/02. The PH01 is equipped with a standard platinum temperature sensor (PT 100). Bath or test solutions can be applied at flow rates from 500 μl to 4.5 ml per minute. The PH01 can be connected to any tubing that fits to the perfusion inlet (luer lock) and outlet (1 mm OD stainless steel cannula). A titanium wire inserted into the cannula results in a turbulent flow that ensures that the temperature of the liquid is constant across the flow stream. For setting up a complete perfusion system, a pump and a perfusion port for aspirating the liquid from the culture chamber are required. Standard setups for MEA recording use a peristaltic pump or syringe pump, and a standard metal cannula with beveled tip for aspiration. 1 PH01 2 Perfusion inlet (luer lock) 3 Perfusion outlet (1 mm OD) 4 TC01/02 connector 5 Heating element cable (connected to TC01/02) 4

2.2 Setting Up and Connecting the PH01 Warning: Do not operate the PH01 unless it is completely filled with liquid. Otherwise, overheating may irreversibly damage the PH01. Warning: Do not start the perfusion until you have double-checked that the perfusion lines are set up properly and that the inflow and outflow rate are matching. Spilled liquid may irreversibly damage electronic instruments. If the setpoint temperature is changed, the controller will need a short time to respond. Please note that there will always be an intrinsic offset between the setpoint and the actual temperature, at the tip of the cannula as well as when the liquid arrives in the culture chamber. The offset at the tip depends mainly on the flow rate, see Flow Rate Dependent Temperature Offset. The temperature in the culture chamber depends on several parameters such as the tubing diameter, length, and material (if a tube is used for guiding the liquid into the chamber), the type of pump used, the flow rate, and so on. The user has to determine empirically the offset of the individual setup once before starting the experiment. The tube that leads into the perfusion chamber should be as short as possible to avoid that the buffer cools down in the line. You can also carefully bend the metal cannula itself, so that the outlet can be positioned directly in the chamber. The flow rate should be constant during the experiment to ensure a stable temperature control. Make sure that the ambient temperature is kept constant and shield the setup from air drafts. You can fix the PH01 directly onto the MEA1060 amplifier (or next to the perfusion chamber), for example with tape. 1. Insert the into the perfusion inlet of the culture chamber (or carefully bend the cannula) so that the liquid can flow into the culture chamber. 2. Connect the (luer lock) perfusion inlet of the to the perfusion tubing (connected to the pump) and tightly secure it. 3. Connect one of the screws on the 's rear side to the ground of the amplifier/setup. Inside the PH01, the cannula is connected to ground via a 10 μf capacitor. This reduces noise to a minimum and avoids DC offset currents. 4. Make sure the temperature controller is switched off. Connect the to one channel of a temperature controller TC01/02. 5. Fix the position of the perfusion cannula (for example, with tape). 6. Set up the aspiration system. 5

7. Set the inflow rate to the optimum operating range (500 μl/min to 4500 μl/min). Make sure the programmed outflow rate of the aspiration system is identical (or slightly higher) than the inflow rate. For example, you can use a double-head peristaltic pump, and use one channel of the pump for the input and the other for the output. An alternative would be a push-pull pump that holds paired syringes, with one syringe for the inflow and one for the outflow. 8. Carefully double-check the setup. Make sure all perfusion lines are tight and set up properly. 9. Switch on the pump. Again, check that the outflow rate matches the inflow rate. 10. Switch on the temperature controller. 11. Check the temperature of the flow stream. Adjust the setpoint temperature accordingly. See also chapter Temperature Control. 2.3 Flow Rate Dependent Temperature Offset The following temperature curve shows the deviation of the setpoint and the actual liquid temperature, depending on the flow rate. This gives you a hint on how to adjust the setpoint temperature. For example, if you have a flow rate higher than 4000 μl/min, you will need to set the setpoint temperature to 37.7 C if you want to achieve an actual temperature of 36.7 C at the cannula tip. Please note that the actual temperature needs to be verified empirically for each perfusion setup. 37 36.5 36 temperature [ C] 35.5 35 34.5 34 33.5 500 1000 1500 2000 2500 3000 3500 4000 4500 flow [µl/min] Fig. 1 Actual temperature of the fluid stream, measured at the cannula tip for different flow rates with a setpoint temperature = 36.7 C and an ambient temperature = 24 C. Note: For flow rates > 1000 μl/min and < 2500 μl/min, the temperature at the outlet is 36.2 C ( 0.1 C) with a given setpoint of 36.7 C. For flow rates lower than 1000 μl/min, the difference between setpoint and actual temperature is due to the temperature loss along the cannula itself, which depends on the ambient temperature. 6

2.4 Air Drafts and Ambient Temperature Ensure that the setup is shielded from air drafts. Even the smallest degree of turbulences in the air such as when a person walks by can affect the PH01's temperature control and thus the outlet liquid temperature as well. Ensure that the ambient temperature remains as constant as possible. 2.5 Pump Types The differences between peristaltic and syringe pumps are shown here for comparison. Both types of pumps deliver the same average outflow temperature (37 C). The offsets are 0.18 C for the peristaltic pump and 0.35 C for the syringe pump (with a flow rate = 1000 μl/min). The two illustrations below show typical long-term properties (20 min) for both pump types at a flow rate of 1000 μl/min. Temperature kinetics using a peristaltic pump 40 temperature [ C] 35 30 25 0 200 400 600 800 1000 1200 time [s] Temperature kinetics using a syringe pump 40 temperature [ C] 35 30 25 0 200 400 600 800 1000 1200 time [s] 7

2.6 Flow Stream A titanium wire inserted into the cannula ensures a turbulent flow. In turbulent flow, there are no discrete layers of flowing liquid. The momentum of the fluid overcomes the viscous shear forces, and there is extensive and continual mixing across the flow stream. This ensures that the temperature of the liquid is constant across the flow stream. Especially flow rates higher than 2500 μl/min would be critical without the wire, because the heat transmission to the inner layers is impaired by the laminar flow. The titanium wire can be removed by the user to increase the inner diameter and thus the maximum flow speed through the cannula. This is not recommended, as removal of the wire results in a laminar fluid flow. In laminar flow, the fluid moves in layers, with one layer sliding smoothly over the other. There is no mixing of fluid from layer to layer, since viscous shear forces damp out relative motions between layers. As a result, the separate layers show different temperatures. The inner layers move much faster than the outer layers. As the temperature sensor is on the outside of the flow stream, the controller will stop the heating if the actual temperature of the outer layer reaches the setpoint temperature. In other words, the inner layers will be considerably colder than the setpoint temperature. 3 Appendix 3.1 Pin Assignment Pin 1, 6 Pin 2, 5 Pin 3, 4 Heating Temperature sensor power supply Temperature sensor probe 8