TANDEM SHOCK WAVES FOR EXTRACORPOREAL SHOCK WAVE LITHOTRIPSY

Similar documents
The Characteristics of Cavitation Bubbles Induced by the Secondary Shock Wave in an HM-3 Lithotripter and Its Effect on Stone Comminution

NIH Public Access Author Manuscript J Endourol. Author manuscript; available in PMC 2008 July 1.

Proceedings of Meetings on Acoustics

PROGRESS IN MODELING AND SIMULATION OF SHOCK WAVE LITHOTRIPSY (SWL)

Radial Response of Individual Bubbles Subjected to Shock Wave Lithotripsy Pulses In Vitro

Shock-induced collapse of a gas bubble in shockwave lithotripsy

Original Contribution

This document is downloaded from DR-NTU, Nanyang Technological University Library, Singapore.

Shock Wave Lithotripsy (ESWL) results aren't improving. What can Radiographers do to improve outcomes with better kidney stone fragmentation

Original Paper. Development of a Cavitation Bubble Observation System and Application to Study Expansion Waves and Shock Waves

Figure 1: Two possible mechanisms of cavitation erosion.

AN ANALYSIS ON HIGH PRESSURE DYNAMIC CALIBRATORS USED IN THE DEFENSE AREAS

5 cm. Flash 20 khz. Camera

Effect of gases on radical production rates during single-bubble cavitation

Analysis of RPC Performance with Different Gas Mixture

Analysis of Pressure Rise During Internal Arc Faults in Switchgear

Detecting sound wave of bubbles produced in a boiling liquid

Acoustical approach to analysis of energy conversions in an oscillating bubble

MEASUREMENTS OF UNDERWATER EXPLOSION PERFORMANCES BY PRESSURE GAUGE USING FLUOROPOLYMER

Offshore platforms survivability to underwater explosions: part I

Study on Intensity of Blast Wave Generated from Vessel Bursting by Gas Explosion

Dynamics of the bubble near a triangular prism array

Cavitation Bubbles in a Starting Submerged Water Jet

Bubble-Based Resonance-Doppler Sensor for Liquid Characterization

An underwater explosion is an explosion where the point of detonation is below the surface of the water.

Waves and Sound. Honors Physics

The Estimation Of Compressor Performance Using A Theoretical Analysis Of The Gas Flow Through the Muffler Combined With Valve Motion

Development of a Shock Loading Simulation Facility

AN EXPERIMENTAL AND ANALYTICAL STUDY OF PLASMA CLOSING SWITCHES FILLED WITH ENVIRONMENTALLY FRIENDLY GASES

DYNAMIC CRUSH TEST ON HYDROGEN PRESSURIZED CYLINDER

Seismic Sources. Seismic sources. Recorders. Requirements; Principles; Onshore, offshore. Digitals recorders; Analog-to-Digital (A/D) converters.

CHAPTER 1 INTRODUCTION

DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS AP PHYSICS

DEVELOPMENTS IN PIPELINE FILLING AND EMPTYING EXPERIMENTATION IN A LABORATORY PIPELINE APPARATUS

EWGAE 2010 Vienna, 8th to 10th September

HIGH-SPEED OBSERVATIONS OF SUBMERGED WATER JETS ISSUING FROM AN ABRASIVE WATER JET NOZZLE

inter.noise 2000 The 29th International Congress and Exhibition on Noise Control Engineering August 2000, Nice, FRANCE

PAPER 2 THEORY QUESTIONS

Fig. 2. M.I. Yaroslavtsev, 2002

Sensor for Air Bubble Detection at Liquid Filled Tubes. SONOCHECK Type ABD06.xx. Operating Manual

ACOUSTIC CAVITATION AND BUBBLE DYNAMICS. Werner LAUTERBORN, Thomas KURZ, Robert METTIN, Philipp KOCH, Dennis KRÖNINGER, Daniel SCHANZ

Waves, Bubbles, Noise, and Underwater Communications

Quantitative detection of bubble dynamics by Doppler ultrasound

Oil-Lubricated Compressors for Regenerative Cryocoolers Using an Elastic Membrane

Acoustic Emission of Bubbly Flow and Its Size Distribution Spectrum

THE EFFECT OF GAS DIFFUSION ON BUBBLE DYNAMICS

Flow in a shock tube

Cooling Characteristics of GM-type Pulse Tube Refrigerator with Neon as Working Gas

PHYSICALLY REALISTIC MODELS OF CATASTROPHIC BUBBLE COLLAPSES

Hunting for the Sweet Spot by a Seesaw Model

Dynamics and acoustics of a cavitating Venturi flow using a homogeneous air-propylene glycol mixture

MODELLING OF FUME EXTRACTORS C. R.

Vibration Analysis and Test of Backup Roll in Temper Mill

The Resistive Plate Chamber detectors at the Large Hadron Collider experiments. Danube School, September 8-13, 2014 Novi Sad

Acoustic Focusing in Shallow Water and Bubble Radiation Effects

High Frequency Acoustical Propagation and Scattering in Coastal Waters

Laser-Induced Bubbles in Glycerol-Water Mixtures

Level MEASUREMENT 1/2016

2 nd Term Final. Revision Sheet. Students Name: Grade: 10 A/B. Subject: Physics. Teacher Signature

Experimental study on the effects of phase change during a bubble collapse

A Depletion Compensated Wet Bath Simulator For Calibrating Evidential Breath Alcohol Analyzers

Air entrainment in Dip coating under vacuum

Development of Fish type Robot based on the Analysis of Swimming Motion of Bluefin Tuna Comparison between Tuna-type Fin and Rectangular Fin -

INTERNATIONAL JOURNAL OF CIVIL AND STRUCTURAL ENGINEERING Volume 1, No 4, 2010

Acoustical Modeling of Reciprocating Compressors With Stepless Valve Unloaders

Application Worksheet

Influence of rounding corners on unsteady flow and heat transfer around a square cylinder

MODELING OF THERMAL BEHAVIOR INSIDE A BUBBLE

IMPROVING PLANT AERATION USING GAS INFUSION TECHNOLOGY

Friction properties of the face of a hand-held tennis racket

1.2 Example 1: A simple hydraulic system

ULTRASOUND-INDUCED POLYMERISATIONS IN HIGH- PRESSURE FLUIDS

Influence of shock-bubble and bubble-bubble interactions on the collapse of a cluster of bubbles

Numerical Simulations of a Train of Air Bubbles Rising Through Stagnant Water

DISSOLVED GAS ANALYSIS OF NATURAL ESTER FLUIDS UNDER ELECTRICAL AND THERMAL STRESS

Influence of the external dynamic wind pressure on the ventilation of double facades

Product Specification

Periodic waves in space and time

Bubble-bubble interactions and wall pressures/temperatures produced by the collapse of a bubble pair near a rigid surface

ABS. Acoustic Bubble Spectrometer. Measurement of Bubble Size, Bubble Number & Void Fraction DYNAFLOW, INC. Research & Development in Applied Sciences

Walking with coffee: when and why coffee spills

Experimental investigation of the check valve behaviour when the flow is reversing

EXPERIMENTAL STUDY ON THE HYDRODYNAMIC BEHAVIORS OF TWO CONCENTRIC CYLINDERS

Chs. 16 and 17 Mechanical Waves

ISOLATION OF NON-HYDROSTATIC REGIONS WITHIN A BASIN

Shock-bubble Interaction Near a Compliant Tissue-like Material

PROPOSAL OF FLUID SELF-EXCITED OSCILLATION PECULIAR TO A FLAT RING TUBE AND ITS APPLICATION

Strand E. Waves. Unit 1. Measuring Waves. Text. Types of Wave 2 Measuring Waves 6 Phase 10

SENSITIVITY ANALYSIS OF THE FIRST CIRCUIT OF COLD CHANNEL PIPELINE RUPTURE SIZE FOR WWER 440/270 REACTOR

ON THE CHARACTERISTICS OF SOUND GENERATED BY BUBBLES INJECTED UNDERWATER HANNAN LOHRASBIPEYDEH, ADAM ZIELINSKI, T. AARON GULLIVER

A CO 2 Waveform Simulator For Evaluation and Testing of Respiratory Gas Analyzers

Vortex Meters for Liquids, Gas, and Steam

A Computational Assessment of Gas Jets in a Bubbly Co-Flow 1

Waves. Name and Surname: Class: L E A R N I N G O U T C O M E. What are waves? Why are waves formed?

4.4 WAVE CHARACTERISTICS 4.5 WAVE PROPERTIES Student Notes

A Study on Roll Damping of Bilge Keels for New Non-Ballast Ship with Rounder Cross Section

Blast Damage Consideratons for Horizontal Pressure Vessel and Potential for Domino Effects

PART EIGHT HIGH FREQUENCY PERCUSSIVE OSCILLATION (HFPOV )

ANALYSIS OF THE POSITIVE FORCES EXHIBITING ON THE MOORING LINE OF COMPOSITE-TYPE SEA CAGE

Similarly to elastic waves, sound and other propagated waves are graphically shown by the graph:

Transcription:

PACS REFERENCE: 43.80.Vj TANDEM SHOCK WAVES FOR EXTRACORPOREAL SHOCK WAVE LITHOTRIPSY Loske, Achim M.; Fernández, Francisco; Prieto, Fernando E. Centro de Física Aplicada y Tecnología Avanzada, UNAM A.P. 1-1010, 76000, Querétaro, Qro., México Tel.: 52 442 2 38 11 64 Fax: 52 442 2 38 11 65 E-mail: loske@fisica.unam.mx ABSTRACT Extracorporeal shock wave lithotripsy (SWL) has become the standard treatment for patients with renal calculi; however, shock waves are not innocuous when focused on the kidney. This paper reports the design of a novel piezoelectric tandem shock wave generator for SWL. The system generates two shock waves with an adjustable time delay (50-950 µsec) to enhance cavitation induced damage to the kidney stone, without increasing tissue trauma. Pressure measurements and fragmentation tests with standardized kidney stone models were compared to that of a conventional generator. Results using tandem shock waves show enhanced stone comminution. INTRODUCTION Disintegration of urinary calculi is possible by extracorporeally induced shock waves. Hundreds of focused shock waves penetrate the body disintegrating the concrements into small enough particles to pass spontaneously through the urinary tract. This technique, known as extracorporeal shock wave lithotripsy (SWL) has been successful for more than twenty years, 1-5 nevertheless, basic research is still necessary to define SWL techniques that minimize tissue damage while improving treatment efficacy. Pancreatic, gallbladder and salivary stones have also been treated with ESWL. 2,4 The objective of our study is to propose improvements to SWL devices (extracorporeal lithotripters) in order to enhance stone comminution without increasing tissue trauma. Lithotripsy shock waves consist of a nonlinear high pressure impulse (up to 150 MPa) with a rapid rise (less than 10 ns) forming the shock front, followed by a negative phase of up to about 20 MPa. They have a wide spectrum of frequencies ranging from a few hundred khz to a few hundred MHz. Due to their nonlinear acoustic behavior, high-pressure impulses may become shock waves on transmission. Their velocity in water and tissue is about 5% faster than sound waves. So far, three main shock wave generation principles have been developed for medical applications: electrohydraulic, piezoelectric and electromagnetic. 3-5 For this study, only piezoelectric shock wave generation was used. Piezoelectric lithotripters produce shock

waves by a fast DC high voltage (5-10 kv) discharge through a set of about 3000 piezoelectric crystals. The crystals change their external dimensions because of the electric discharge, producing a compression wave in water. Tensile or negative pressure results from the return of the crystals to their original shape. Focusing of the energy is obtained by spherical alignment of the piezoelectric crystals. This is necessary to achieve maximum energy at a small volume, ensuing minimal tissue damage. Shock wave coupling is achieved by a water cushion together with coupling ultrasound gel (Fig. 1). Water is degassed to remove dissolved air and microscopic bubbles which absorb shock wave energy due to acoustic mismatch. To localize the stone, X-ray or ultrasound imaging systems are used. Details on piezoelectric lithotripters are given elsewhere. 2-5 Kidney stones fracture mainly due to cavitation, internal crystalline-matrix layer separation, and spalling around the stone. 6-8 A cloud of cavitation bubbles is generated in the vecinity of the focus F by the tensile phase following the positive peak of each shock wave. It also may appear by reflection of the shock wave at the concretion. These bubbles expand in about 50-100 µs, stabilize for 200 500 µs, and finally collapse, creating stone-damaging high energy secondary shock waves and high velocity (up to 300 m/s) Fig. 1. Schematic diagram of piezoelectric shock wave generation and focusing for extracorporeal kidney stone disintegration microjets of fluid. 9-13 The larger the bubbles grow, the more violent their collapse will be. Several authors have shown that if a second shock wave is sent during or shortly after the stable phase of these bubbles, their collapse can be intensified, enhancing stone damage significantly. So far this has been tested using either two electrohydraulic shock wave generators facing each other or using composite reflectors for electrohydraulic lithotripters; 14-21 however, retrofitting a second shock wave generator or composite reflectors on an electrohydraulic lithotripter may be complicated or expensive. We designed and tested a novel piezoelectric tandem shock wave generator capable of producing two shock waves at an adjustable delay between 50 and 950 µs. An important feature of this highly versatile system is the possibility of varying the amplitude of each shock wave independently as well as changing the phase. Inverted pressure waveforms that start with a tensile pulse followed by a compressive one, can be generated independently, switching the system to inverse mode. Pressure measurements and standardized kidney stone model disintegration tests obtained with the tandem system at different settings were compared with that of the conventional lithotripsy system. MATERIAL AND METHODS The electric circuit of a Piezolith 2300 (Wolf GmbH, Knittlingen, Germany) lithotripter was modified. Conventional piezoelectric shock wave generators (Fig. 2) only have one capacitor charging system and one spark gap (SG). Figure 3 shows the tandem system, consisting of two independent high voltage power supplies charging two 0.5 µf capacitors (C1 and C2). Both capacitors remain charged until their spark gap (SG1 or SG2) is fired. This is achieved sending a Fig. 2. Simplified diagram of the conventional piezoelectric generator.

12 kv low current pulse to fire the trigger electrode (TE1 or TE2) of each spark gap, inducing electric breakdown between main electrodes and discharging either C1 or C2 through the piezoelectric array. Fig. 3. The novel tandem shock wave generator The tandem circuit may be operated in manual or repetition mode. Repetition mode offers either conventional or tandem shock wave generation at an adjustable time delay between 50 and 950 µs. Conventional lithotripter pulses, followed by phase inverted pulses, or vice versa, may be selected (Fig. 4). Either two conventional or two inverted shock waves can be generated at the desired delay. Since for experimental purposes the complete lithotripter was not needed, only the shock wave generator was mounted on a bench and a cylindrical 45 cm high Lucite water tank placed on top of it. An X-Y-Z positioner was used to fasten the pressure gauge or a kidney stone model at the focus F. Pressure waveforms were registered at F with polyvinylidene difluoride (PVDF) needle hydrophones (Imotec GmbH, Würselen, Germany) and fed into a digital oscilloscope (Tektronix, Inc., Beaverton, OR, USA, model 2430A). Water temperature, water level and discharge voltage were set to 25 C, 44 cm and 7.5 kv, respectively. Ten pressure waveforms were obtained with the conventional single pulse system and ten records were registered at each time delay with the dual-phase system. All pressure measurements were done using the manual mode. Three standardized (30 x 30 x 14.3 ± 0.1 mm) rectangular kidney stone models (High Medical Technologies, Kreuzlingen, Switzerland) were exposed, one by one, to 600 shock waves at each delay, centering them horizontally at the focus F of the piezoelectric array. The fragmentation efficiency coefficient, defined as F = 100(W i W f )/W i was obtained selecting 17 different delays. W i and W f are the average initial and the average final (shock wave-exposed) model weights. Models were saturated in water for 10 minutes before shock wave application.

(a) (b) Fig. 4. Pressure waveform obtained at the focus of a) the conventional, and b) and c) the novel tandem shock wave generator at a delay of 400 and 500 µs, respectively. SG1 and SG2 correspond to the instant when the first and second spark gap was fired. (c) RESULTS Average amplitudes of the positive and negative peak obtained with the conventional single shock wave system were about 38 and 18 MPa respectively. No statistically significant difference (p 0.01) was obtained between these values and the amplitudes of the first shock wave when using tandem shock wave generation at any delay. The amplitudes of the positive and negative pressure peaks of the second shock wave (about 32 and 15 MPa) were always significantly lower (p 0.01) than those corresponding to the first shock wave. Figure 4a shows a typical lithotripter shock wave obtained at F with the conventional system. Figure 4b and c correspond to tandem shock waves generated at a delay of 400 and 500 µs, respectively. For simplicity, the spark gap discharge signals seen on the screen of the oscilloscope are indicated only as vertical dotted lines. All shock waves reached the hydrophone approximately 230 µs after the spark gaps were fired. This time corresponds to a shock wave velocity of about 1500 m/s. Fragmentation efficiency is shown in Figure 5. Increased efficiency was obtained at a delay of 350 µs. This is about 23% higher than the efficiency observed for the double number (1200) of conventional single lithotripter pulses. DISCUSSION Our results show that tandem shock wave generation at a delay of 350 µs increases stone phantom disintegration efficiency. We believe this to be due to enhanced cavitation. A 350 µs delay is in accordance with bubble collapse times reported by other authors. 22 No

statistically significant difference was obtained for the disintegration efficiency at 350, 400, and 450 µs. At these delays the second positive pressure wave seems to accelerate bubble collapse, intensifying damage to the stone phantoms. If the second positive peak arrives after bubble collapse, no additional damage is produced to the models. At delays between 100 and 250 µs the second shock wave apparently arrives during the growing phase of the bubbles generated by the first shock wave. Because of this, bubbles collapse before reaching their maximum size and damage is reduced, being even lower than using the conventional system. We suppose that since the bubble cloud produced by the passage of the first shock wave is still present as the Fig. 5. Fragmentation efficiency of rectangular kidney stone models exposed to 600 tandem shock waves generated at 7.5 kv. The dotted line corresponds to the fragmentation efficiency obtained with 1200 single (conventional) shock waves. second shock wave arrives, the second positive and negative pressure amplitude is always smaller than the first. More experiments using different kidney stone models as well as in vivo models will be helpful to decide whether tandem shock wave generation should be used for SWL. Due to its versatility, the tandem shock wave generator described here could also be very useful as a research device to increase knowledge of stone fragmentation mechanisms. Since shock wave cytotoxicity is related to cavitation, this system could also be used for research in oncology, shock wave mediated macromolecule delivery, 23 and during exposure of isolated microorganisms to shock waves. 24 A concern about tandem shock waves for SWL could be cavitation-induced tissue damage; however, other authors have shown that bubble collapse in vivo is less violent than in water, because bubble expansion is constrained by the tissue. 25 In contrast to other tandem shock wave systems that have been proposed, our circuit could be installed in clinical piezoelectric lithotripters at relatively low cost. ACKNOWLEDGEMENTS The authors acknowledge technical assistance by Edgar and Arturo Méndez. They would also like to thank Helmuth Busch and Mario Granizo for advice. The Piezolith 2300 shock wave generator was donated to our laboratory by Dewimed S.A. de C.V. REFERENCES 1. Chaussy C, Brendel W, Schmiedt E. Extracorporeally induced destruction of kidney stones by shock waves. Lancet 1980: 2; 1265. 2. Loske AM, Prieto FE. Shock Waves in Medicine: Commented Bibliographical Database. Mexico City. Coordinación de la Investigación Científica, UNAM 1995. ISBN 968-36- 4466-X.

3. Coptcoat MJ, Miller RA, Wickham JEA. The End of the Stone Age. London. BDI Publishing 1987. 4. Loske AM. Applications of shock waves in medicine. In: Handbook of Shock Waves. G Ben-Dor, T Elperin, O Igra, A Lifshitz Eds. New York. Academic Press, 2001, vol 2, pp 415-440. ISBN 0-12-086432-2. 5. Loske AM, Prieto FE. Fundamentos Técnicos de Litotripsia Extracorporal. Mexico City. JGH Editores, 1999. ISBN 970-681-001-3. 6. Steiger E. Full wave modeling of lithotripter fields. In: The Sound of the Future: A Global View of Acoustics in the 21 st Century. PK Kuhl, LA Crum Eds. Seattle, WA. Acoustical Society of America 1998; 2795-2796. 7. Lokhandwalla M, Sturtevant B. Fracture mechanics model of stone comminution in ESWL and implications for tissue damage. Phys. Med. Biol. 2000: 45; 1923-1949. 8. Sass W, Braunlich M, Dreyer HP, Matura E, Folberth W, Preismeyer HG, Seifert J. The mechanisms of stone disintegration by shock waves. Ultrasound Med. Biol. 1991: 17; 239-243. 9. Leighton TG. The Acoustic Bubble. San Diego, CA. Academic Press 1994. ISBN 0-12- 441 920-8. 10. Bailey MR, Blackstock DT, Cleveland RO, Crum LA. Comparison of electrohydraulic lithotripters with rigid and pressure-release ellipsoidal reflectors. II. Cavitation fields J. Acoust. Soc. Am. 1999: 106; 1149-1160. 11. Choi MJ, Coleman AJ, Saunders JE. The influence of fluid properties and pulse amplitude on bubble dynamics in the field of a shock wave lithotripter. Phys. Med. Biol. 1993: 38; 1561-1573. 12. Crum LA. Surface oscillations and jet development in pulsating bubbles. J. of Physique 1979: 40; 285-288. 13. Crum LA. Cavitation microjets as a contributory mechanism for renal calculi disintegration in ESWL. J. Urol. 1988; 140: 1587-1590. 14. Bailey MR. Control of acoustic cavitation with application to lithotripsy. Ph.D. Dissertation, Austin TX; Univ. of Texas at Austin, 1997. 15. Cathignol D. Comparison between the effects of cavitation induced by two different pressure-time shock waveform pulses IEEE Trans. Ultrason. Ferroel. and Freq. Control 1998: 45; 788-799. 16. Loske AM, Prieto FE. Dual-phase reflectors for extracorporeal shock wave lithotripsy. Physica Medica 2001: 17; 141-149. 17. Prieto FE, Loske AM. Multifocal Composite Reflectors to Concentrate Shock Waves. International Patent P.C.T./MX 95/00008 1995. 18. Prieto FE, Loske AM. Zusammengesetzte Multifokalreflektoren zur Konzentration von Stosswellen. German Patent DE 195 81 913.6 1998. 19. Prieto FE, Loske AM. Bifocal reflector for electrohydraulic lithotripters. J. Endourol. 1999: 13; 65-75. 20. Sokolov DL, Bailey MR, Crum LA. Use of two pulses to localize and intensify cavitation in lithotripsy. J. Acoust. Soc. Am. 2001: 110; 1685-1695. 21. Xi X, Zhong P. Improvement of stone fragmentation during shock-wave lithotripsy using a combined EH/PEAA shock-wave generator in vitro experiments. Ultrasound Med. Biol. 2000: 26; 457-467. 22. Zhong P, Cocks FH, Cioanta I, Preminger GM. Controlled, forced collapse of cavitation bubbles for improved stone fragmentation during shock wave lithotripsy. J. Urol. 1997: 158; 2323-2328. 23. Lauer U, Bürgelt E, Squire Z, Messmer K, Hofschneider PH, Gregor M, Delius M. Shock wave permeabilization as a new gene transfer method. Gene Therapy 1997: 4; 710-715. 24. Loske AM, Prieto FE, Zavala ML, Santana AD, Armenta E. Repeated application of shock waves as a possible method for food preservation. Shock Waves 1999: 9; 49-55. 25. Zhong P, Cioanta I, Cocks F H, Preminger GM. Inertial cavitation and associated acoustic emission produced during electrohydraulic shock wave lithotripsy. J. Acoust. Soc. Am. 1997: 101; 2940-2950.