ZIRCONIA OXYGEN DETECTOR SERVICE MANUAL

Similar documents
INFRARED GAS ANALYZER

ZIRCONIA OXYGEN ANALYZER CONVERTER

ZIRCONIA OXYGEN ANALYZER CONVERTER

Columbus Instruments

Specifications of Oxygen Sensor Module V-8344A

Stand-Alone Bubble Detection System

OWNER S MANUAL NMB SERIES

INFRARED GAS ANALYZER

Transmitter CS 21 Operation Manual

IR200 Infrared Gas Analyzer

Model ZX8D Zirconia Oxygen Analyzer

Zirconia Oxygen Sensors. Model 1230 Series

Transmitter CS 21 Operation Manual

MEGAS 2.0. Gas analysis with direct C-Level Calculation and Bus-Connection. Data sheet

USER MANUAL. JET TST and JET TST S.

ACV-10 Automatic Control Valve

BI-680 Online Dissolved Oxygen Controller Instruction Manual

Instruction and Maintenance Manual

MODEL 400 CLD INSTRUCTION MANUAL

Instruction and Maintenance Manual

97C COMPRESSOR KIT 12V PART NO C COMPRESSOR KIT 24V PART NO C COMPRESSOR KIT PART NO

IN-SITU ZIRCONIA OXYGEN ANALYZER < ZIRCOMAT-C >

CCT-7320/ROC-2313 Reverse Osmosis Controller

Model 130M Pneumatic Controller

MODEL GT820 OXYGEN SENSOR

Alphasense 4-20 ma transmitters offer convenience and easy maintenance for toxic sensors:

CONDUCTIVITY POINT/S LEVEL SWITCH JAYCEEAQUA 4000 SERIES

ZIRCONIA OXYGEN ANALYZER CONVERTER

ZIRON Oxygen Measuring Technique. Oxygen Measuring Device ECO1000. Manual

C&G Level 3 Award in the Periodic Inspection, Testing and Certification of Electrical Installations

Manual Oxygen Analyzer A4009T

Oxyfire-LT Modular Oxygen Sensor Installation / Operation Manual

DF-550E PROCESS ANALYSERS APPLICATIONS FEATURES

Inert Air (N2) Systems Manual

INSTRUCTION MANUAL. FLOW CONTROL DRAWERS MANUAL / PLC CONTROL SERIES Model Version Perma Pure LLC Tel:

User's Manual. OX100 General-Purpose Oxygen Analyzer. IM 11M10A01-01E 2nd Edition IM 11M10A01-01E

Gas Sampling Fitting with MK2.1 and L-Probe

Sensoric 4-20 ma Transmitter Board Operation Manual

Ventam 85 Installation & Commissioning Instructions

OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE

DPC-30 DPC-100. Reference Manual

Installation, operating and maintenance Instructions for Seemag bypass level indicator

High-performance submersible pressure transmitter For level measurement Model LH-10

High-performance submersible pressure transmitter For level measurement Model LH-10

100C Air Compressor Kit

200 PSI COMPRESSORS - MODEL NUMBERS

IMPORTANT SAFETY INSTRUCTIONS

Best Practice Guide, Servomex 2700

420C AIR COMPRESSOR KIT PART NO C AIR COMPRESSOR KIT PART NO

STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE

RC 195 Receiver-Controller

Ground Fault Circuit Interrupter(GFCI) Policy

User's Manual. OX100 General-Purpose Oxygen Analyzer. IM 11M10A01-01E 2nd Edition IM 11M10A01-01E

GASGUARD VENT LINE3 Ammonia Sensor OPERATING & INSTALLATION MANUAL

INFRARED GAS ANALYZER

O 2 SENSORS Zirconium Dioxide (ZrO 2 ) Oxygen Sensor Operating Principle Guide

Inoxair. Repair Manual. Content. Edition: 12/2010. Subject to technical changes for reasons of the continous development.

RHPS Series RD(H)F40 User Manual. Read the complete manual before installing and using the regulator.

400H HARDMOUNT AIR COMPRESSOR KIT PART NO H HARDMOUNT AIR COMPRESSOR KIT PART NO

Procedure Troubleshooting the Interconnect Cables

AMS 2710 PCB pressure sensor module with V output

O3 3E 1 F Gas Sensor Module

Overview. Front Panel: Keypad and Display

OXYGEN MONITORING SYSTEM OMS

IR400 Infrared Gas Analyzer

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

Flamco Flamco-Fill STA Flamco-Fill STM

ABB MEASUREMENT & ANALYTICS DATA SHEET. AK100 Series ATEX compliant gas analyzer system for hydrogen-cooled alternators

The Ins and Outs of I/P Transducers

444C DUAL PERFORMANCE VALUE PACK

Oxygen Flue Gas Analyzer BA 2000

Crest Heating Boilers FBN(L) Series &

250C-IG COMPRESSOR KIT 12V PART NO C-IG COMPRESSOR KIT 24V PART NO

DF-310E PROCESS ANALYSERS APPLICATIONS FEATURES

250C-IG COMPRESSOR KIT 12V PART NO C-IG COMPRESSOR KIT 24V PART NO

AUTOMATIC GAS MANIFOLDS

AK100 Series ATEX Compliant Gas Analyzer System for Hydrogen-cooled Alternators

Mass Flow Meter (MFM) for gases

42045 Heavy Duty ADA Base Model Kit: 85/105 PSI (ADA Compressor Only) Heavy Duty ADA Base Model Kit: 110/145 PSI (ADA Compressor Only)

This test shall be carried out on all vehicles equipped with open type traction batteries.

Digital Flow Switch. Series PFM7

RD(H)20/25 Pressure-Reducing Regulator User Manual

OXYGEN MINERALE ABSOLUTE RANGE OWNER S MANUAL OXY MINERALE 600/1000 SERIES

2/ Port G1/4

Commercial Softeners. Installation, Operation & Maintenance Guide. Manual 002.1

INSTRUCTION MANUAL. FLOAT TYPE LEVEL TRANSMITTER HT-100R Series HITROL CO., LTD.

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

400C & 450C DUAL PERFORMANCE VALUE PACKS

TWC Services, Inc. Structured On-The-Job Training for JT- 1, 2, 3

NIV EAU MATIC ST User Manual Made in Canada

EPP4 Pressure Regulator Basic G 1/4"

Engineered Solutions for Gas Detection and Analysis. High Purity Instruments MODEL 3-LPM-N1-SS

Series 6517 Katharometers. The ideal equipment for Process Monitoring and Control. Robust no moving parts. Designed for continuous industrial use

Reverse Osmosis System Installation Guide and Operation Manual.

AHE58/59 AC Servo System

TSE Blood Pressure Monitor Invasive series

Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter

TECHNICAL INFORMATION

Installation, use and maintenance instructions. Gas burner (5)

Navigator 600 Silica analyzers

Transcription:

Service Manual ZIRCONIA OXYGEN DETECTOR SERVICE MANUAL TYPE: ZFK 3, 4, 7 TN5A08a-E

CONTENTS. PREFACE.... MEASUREMENT PRINCIPLE... 3. ADJUSTMENT...3 3. Zero (Air) Adjustment... 3 3. Span Adjustment... 3 4. TROUBLE CHECK SHEET...4 5. REPAIR...5 APPENDIX Wiring Diagram...6 Output characteristic table...7 --

. PREFACE This service manual explains repair and adjustment procedure of the Zirconia oxygen detector (ZFK 3, 4, 7) as a single unit. This manual is intended for use with the instruction manual (INZ-TN4ZFK3).. MEASUREMENT PRINCIPLE When yttria-stabilized zirconia ceramic is heated to hightemperature, only the oxygen ion becomes a movable solid electrolyte. In a zirconia element, the both sides of which are fitted with a platinum electrode, the electrode reactions in the following expressions occur for the oxygen concentrations P and P (P > P ). Electrode of P side O + 4e 0 (negative electrode) Electrode of P side 0 O + 4e (positive electrode) Filter P Measured gas Heater Electrode Insulating material Casing Zirconia element Fig. Detector structure Air P Output That is, the oxygen ion moves from the P side, where oxygen concentration is higher, to P side, where it is lower. This energy for moving the ion generates the electric force (E) indicated in the following expression. RT P E= n (O) (V) 4F P (O) P (O) =50.74 log (mv) (at 800 C) P (O ) E : Electric force P (O ) : Comparative (Air) oxygen concentration P (O ) : Oxygen concentration in the measured gas R : Gas constant 8.344 (J mol - K - ) T : Absolute temperature F : Faraday constant 9.649 0 4 (c mol - ) Coefficient 50.74 is the value when the zircornia element is 750 C. (The sensor unit becomes 750 C with 800 C temperature control.) Therefore, unknown oxygen concentration P (O ) can be calculated from the electric force (E). --

3. ADJUSTMENT 3. Zero (Air) Adjustment ) Remove the 3 screws fixing the case cover. ) Feed the zero gas (Dry Air) through the sample gas inlet (on the lower side of the main unit) at the specified flow rate (0.5L /min). 3) After the voltage indication between the external terminals block (4) and (5) (the terminal block on the front panel 7P) is stabilized (feed zero gas for to 3 minutes), adjust the indication to 0 ± mv by VRZ on the printed board; ZFK 3, 4. For ZFK7, make sure that the indication is within ± 3mV. (No adjustment) 3. Span Adjustment ) Feed the span gas ( to % O /N ) at the specified flow rate. ) After the indication is stabilized, adjust it by VRS; ZFK 3, 4 Find the adjustment value using the following expression. E (Output:V) = 0.38 log /Px Px:Calibration gas concentration Example) In case of.% O /N E=0.38 log /. = 0.488 v Make an adjustment with the VRs control so that 0.488 V is obtained. Span VR VRS Zero VR VRZ 8 C4 0 R4 3 6 R C CN3 JAPAN VRS R7 R3 CN R R8 CN C3 R5 VRZ ZD ZD ZD4 ZD3 C TK4B569R 3 4 3 SSR Fig. Printed board mounting diagram; ZFK 3, 4 For ZFK7, make sure that the indication is within 50 to 70mV when the gas with concentration of to % O /N is flowed. (Refer to Appendix Table.) (No adjustment) -3-

4. TROUBLE CHECK SHEET Phenomena Cause Checking methods (normal value) Protective measures Slow response Sensor element Change over between zero and span Replace deterioration gas and check if 60 seconds or longer Sample gas is not supplied or output is abnormal. Sensor outlet is clogged with crystallized sulfate mist. Output is unstable ZFK 3, 4 Disconnection of connector or poor contact Output is not deflected or remains the same. Output swings over. is needed for 90% response. Visually check for clogging of the outlet of the gas introduction case. Check if the connector is disconnected or move the connector part to check for the indication change. Remove the clogging. Replace with a gas introduction case (provided with gas outlet pot). Repair Disturbance noise Check if output is hunting. Eliminate noise. Sensor element Flow zero/span gas continuously for Replace the sensor deterioration 0 minutes or so, and observe the element. Fuse is blown out or power is not supplied. ZFK 3, 4 Disconnection of connector or poor contact Sensor is defective. Solid relay is defective. Printed board is defective. ZFK 3, 4 Constant voltage power supply is defective. Temperature controller is defective. Measuring range is exceeded. indication status. Check the fuse and the supply voltage specification. Check if the connector is disconnected or move the connector part to check for the indication change. Change over between zero and span gas and check for the indication change. Remove one end of the sensor terminal block (3) (4), and check for continuity. The range should be within to 3 Ω. When 4V DC is applied to the outer end (7) (8) (-), check if AC power between () () on the sensor element is turned on or off. (Check with a tester.) Change over between zero and span gas and check if the voltage value between the sensor outer end (5) (6) (+) indicates as shown below. Air: ± 3mV or lower to % O /N : 50 to 70mV. Check if the CN of the constant voltage power supply indicates ± 0.V. Check if the temperature indication is constant in a range of 800 ± 5 C. Flow measured gas after zero/span calibration. Replace the fuse or check the power supply. Repair Replace the sensor element. Replace the printed board or the solid relay. Replace the printed board or the solid relay or sensor element. Replace the constant voltage power supply. Replace the temperature controller. Review the specifications. -4-

5. REPAIR () Replacement of the sensor element ) Change the measuring gas line to the Air atmosphere and then turn off the power switch. When the device is operating, install an oxygen meter by bypass piping. Leave the device for about 30 minutes till it cools down. ) Remove the case cover (a). 3) Remove the 6 wires connected to the sensor terminal block (b) and the 3 M4 screws fixing the sensor (c). 4) Pull out the sensor (d) from the sensor case (e), and assemble the new sensor in reverse order of disassembly. After wiring is checked, perform energization and running to calibrate zero/span. When replacing the sensor element, also replace the packing with a new one. () Replacement of the sensor case ), ) Follow the procedure in Item (). 3) Remove the 3 M4 screws fixing the sensor (c). 4) Remove the screws fixing the sensor case from the bottom of the main unit, and remove the sensor case from the top. 5) Assemble a new sensor in reverse order of disassembly. (3) Replacement of the printed board (ZFK 3, 4) ), ) Follow the procedure in Item (). 3) Remove the 8 wires connected to the terminal block of the printed board and 3 screws fixing the printed board. 4) Assemble a new printed board in reverse order of disassembly. Perform energization and running to calibrate zero/span. (4) Replacement of the temperature controller ), ) Follow the procedure in Item (). 3) Remove the screws from the top of the temperature controller and the socket from the backside, and pull out the controller forward. 4) Assemble a new temperature controller in reverse order of disassembly. (5) Replacement of the constant voltage power supply (ZFK 3, 4) ), ) Follow the procedure in Item (). 3) Remove the 5 wires and the side fixing screws connected to the upper terminal block of the constant voltage power supply (g), and remove the constant voltage power supply. 4) Assemble a new constant voltage power supply in reverse order of disassembly. (a) Case cover (e) Sensor case (c) Screw (d) Sensor (b) Wiring (h) Printed board: ZFK 3, 4 Solid relay; ZFK 7. (f) Temperature controller (g) Constant voltage power supply: ZFK 3, 4 Not mounted on ZFK 7. Fig.3 Internal diagram of the oxygen analyzer -5-

R T ZFK3,4 APPENDIX Wiring Diagram SW Power supply F. 3A 85 to 65V AC Temperature controller 0 PXR4 7 8 85 to 65V AC 5 4 Approx. 8mV/800 C Sensor mv 85 to 65V AC Constant voltage power supply AC AC V 0 DC POWER SUPPLY 3 4 5 6 7 8 3 4 5 6 IN DCV OUT SSR Heat R OUT IN ZFK CN CN3 Amp. P.C.B O Sensor 00 to 5V AC or 00 to 0V AC 3 4 5 6 7 R T E + - 0 0 03 04 Main power Temperature supply alarm Output 0 0 C Output 03 04 Temperature alarm When the setting value (800 C) is exceeded by ±0 C R T ZFK7 SW *00 to 40V AC for CE compatible types Power supply F. 3A 85 to 65V AC Temperature controller 0 PXR4 7 8 5 4 4 3 SSR 03 04 O Sensor 00 to 5V AC or 00 to 0V AC 3 4 5 6 Heat R OUT ZFK 0 0 C Temperature alarm C Output -6-

APPENDIX TABLE ZFK 3, 4 Zirconia oxygen Analyzer Output characteristic table (ZFK 3, 4 output) %O OUTPUT (mv) %O OUTPUT (mv) 0.05 000.0 9.7 0. 885.3 3.0 79.4 0.5 68.8 4.0 67..0 504.0 5.0 55.7.0 389.3 6.0 45.0 3.0 3. 7.0 35.0 4.0 74.5 8.0 5.5 5.0 37.6 9.0 6.6 6.0 07.4 0.0 8.8 7.0 8.9.0 0.0 8.0 59.8.0-7.7 9.0 40.3 3.0-5. 0.0.8 4.0 -..0 07. 5.0-8.9 E (OUTPUT : V) = 0.38 log /O % APPENDIX TABLE ZFK 3, 4, 7 Zirconia oxygen Analyzer Detector Output characteristic (internal sensor) %O OUTPUT (mv) %O OUTPUT (mv) 0.0 68.57 7.0 4. 0.0 53.3 8.0.7 0.03 44.36 9.0 8.67 0.04 38.0 0.0 6.35 0.05 33..0 4.5 0. 7.83.0.33 0. 0.56 3.0 0.57 0.3 93.6 4.0 8.93 0.4 87.8 5.0 7.4 0.5 8.36 6.0 5.99 0.6 78.35 7.0 4.66 0.7 74.95 8.0 3.40 0.8 7.0 9.0. 0.9 69.4 0.0.08.0 67.09 0.6 0.438.0 5.8.0 0 3.0 4.88.0 -.05 4.0 36.54 3.0 -.005 5.0 3.6 4.0 -.943 6.0 7.6 5.0-3.84 E (OUTPUT : mv) = 50.74 log /O % -7-