Influence of Reynolds number on Cd of Critical flow venturi nozzle and validation with ISO standards

Similar documents
Conversion factors for Thermal Mass flow meters for different operating mediums: An Experimental study

Validation of Custody Transfer Metering Skid at Site After Laboratory Proving

CORESTA RECOMMENDED METHOD N 6

Compiled by: B Beard. Approved by: SH Carstens. Description of requirements and procedures for compact provers to be used as verification standards.

Operational experiences with the EuroLoop Liquid Hydrocarbon Flow Facility

FPG8601 Force Balanced Piston Gauge

THE IMPLEMENTATION OF TOROIDAL THROAT VENTURI NOZZLES TO MAXIMIZE PRECISION IN GAS FLOW TRANSFER STANDARD APPLICATIONS

Establishing Primary Gas Pressure Standard facility (0.140 to 70) bar Gas Piston Gauge at FCRI and its applications

TWO PHASE FLOW METER UTILIZING A SLOTTED PLATE. Acadiana Flow Measurement Society

A centrifugal pump consists of an impeller attached to and rotating with the shaft and a casing that encloses the impeller.

ABSTRACT. flotek.g Innovative Solutions in Flow Measurement and Control - Oil, Water and Gas August 28-30, 2017, FCRI, Palakkad, Kerala, India

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010

HUMOR 20. High Accuracy Humidity Calibrator. Water T, p1 HUMOR 20 V1.0

Sartorius CCL kg Prototype Mass Comparator

Helium Mass Spectrometric Leak Detection In Large Size Process Plants

COMPAFLOW. Compressed Air. Volumetric flow. Gas. Mass flow. Steam. Net volumetric flow. Liquid

Applied Fluid Mechanics

Paper 2.2. Operation of Ultrasonic Flow Meters at Conditions Different Than Their Calibration

Instruction Manual. Pipe Friction Training Panel

ONSITE PROVING OF GAS METERS. Daniel J. Rudroff WFMS Inc West Bellfort Sugar Land, Texas. Introduction

Impact of imperfect sealing on the flow measurement of natural gas by orifice plates

Gas flow calibrations performed at the National Metrology Institute of South Africa (NMISA)

Specific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities

Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow

Pneumatic dead-weight tester Model CPB3500

Performance Overview calibration Laboratory EP Instruments Messtechnik und Kalibrierung GmbH

Unique mounting features for Thermal Mass Flow Meters. By Martin Raab in cooperation with CaTeC b.v.

Radar, Ultrasonic and RF Level Transmitters

Restriction Orifice. Single or Multi Stage Orifice to. Reduce Pressure or. Limit the Flow Rate

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

HUMOR 20 HUMOR 20. High-precision Humidity Calibrator. Water T, p1

DEVELOPMENT OF HIGH ALTITUDE TEST FACILITY FOR COLD JET SIMULATION

PG7000 high line differential mode for defining differential pressure at line pressures above atmospheric pressure (PG7000 V2.

INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE

Performance tests of two insertion type thermal mass flow meters

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF PRESSURE METROLOGY

Effect of Fluid Density and Temperature on Discharge Coefficient of Ogee Spillways Using Physical Models

The Discussion of this exercise covers the following points:

Pressure transmitters EMP 2

Smoke and heat Ventilator Testing

Towards ISO Koos van Helden, B.ASc. International Account Manager Eastern Hemisphere mailto:

Level MEASUREMENT 1/2016

WMO LABORATORY INTERCOMPARISON OF RAIN INTENSITY GAUGES

How to specify a product. Process Sensors and Mechanical Instruments

Third measurement MEASUREMENT OF PRESSURE

Gas Flow Calibration Basics

The M-Series Eletta Flow Meter High accuracy DP Flow Meter with multiple functions

MEMORANDUM. Investigation of Variability of Bourdon Gauge Sets in the Chemical Engineering Transport Laboratory

Vortex Flow Meter Wafer or Flange Connection. - Steam - Liquid - Gas

Calibrating and Monitoring MUSMs in Natural Gas Service

A Depletion Compensated Wet Bath Simulator For Calibrating Evidential Breath Alcohol Analyzers

Hydraulic pressure balance Model CPB5800

for Building facilities Industrial facilities etc.,multipurpose Pilot operated type(high capacity)

Series IRV1000/2000/3000

JOS 15 JOS 25 JOS 40. Manual and semiautomatic test benches from DN 25 to DN 40. Characteristics. Basic characteristics. Test bench using balances

Sizing Pulsation Dampeners Is Critical to Effectiveness

3D drawings available at

Gas mixtures. Individual solutions specifically for your application

ONSITE PROVING OF GAS TURBINE METERS Daniel J. Rudroff Invensys Metering Systems

Effect of Coiled Capillary Tube Pitch on Vapour Compression Refrigeration System Performance

Vehicle-mounted meters, pump supplied

Test Apparatus for Performance Evaluation of Compressed Air End-Use Applications

Integral type Differential pressure flowmeter VNT Series

Experimental Analysis on Vortex Tube Refrigerator Using Different Conical Valve Angles

Electropneumatic Positioner Type 4763 Pneumatic Positioner Type 4765

International Journal of Research in Advent Technology Available Online at:

General Accreditation Guidance. User checks and maintenance of laboratory balances

Vortex Meters for Liquids, Gas, and Steam

Fluid Machinery Introduction to the laboratory measurements

INFLUENCE OF MEASURING PROCESS AUTOMATION ON UNCERTAINTY OF MASS STANDARD AND WEIGHTS CALIBRATION.

THE USE OF Z FACTORS IN GRAVIMETRIC VOLUME MEASUREMENTS

Ultrasonic level transmitter and indicator

Level transmitters and indicators Series LU Ultrasonic level transmitter and indicator for liquids and solids

AGILENT TURBOMOLECULAR PUMPS MEASUREMENT METHODOLOGY

NMISA NEW GAS FLOW CALIBRATION FACILITY. Deona Jonker 7 October 2013

3-way control valves VK 3

HYDRAULICS. H89.8D - Hydraulic Bench

HANDBOOK SAFETY DEVICES. Ed SAFETY DEVICES DS-ED 01/ ENG 1

SINGLE VALVE WITH LOW-FLOW BYPASS

Gas Measurement Fundamentals Certification. Curriculum

RESIDENTIAL WATER DISTRIBUTION

Oil And Gas Office Houston Fax Test Separator / Off-Shore Metering

ENVIRONMENTAL AND POLLUTANTS GAS ANALYZERS

CHARACTERISTICS OF LEAK DETECTION BASED ON DIFERENTIAL PRESSURE MEASUREMENT

Development of Gasdynamic Probe for Total Temperature Measurement in gases

1 Overview. Pressure Measurement Single-range transmitters for general applications. 1/22 Siemens FI US Edition

Tutorial. BOSfluids. Relief valve

Selecting the right pressure sensor for your application

The Estimation Of Compressor Performance Using A Theoretical Analysis Of The Gas Flow Through the Muffler Combined With Valve Motion

Products Solutions Services. Gas Flow Measurement

Study of Thermal Effect on Calibration of an Ultrasonic Flow Meter

H&D Fitzgerald Pressure Pyknometry System Mark 5

MACH ONE MASS FLOW CONTROLLER. MACH ONE SERIES flow control. MASS FLOW CONTROLLERS at the speed of sound.

Chapter 3 EXPERIMENTAL: EQUIPMENT, SET-UP, AND PROCEDURE

KMT. Thermal Mass Flow Meter. for Compressed Air and Gases. Measuring range: m3n/h m3 N

COSMO INSTRUMENTS CO., LTD.

Pneumatic Conveying Principles BSH&E SEMINAR

Technical Report. 5th Round Robin Test for Multi-Capillary Ventilation Calibration Standards (2016/2017)

AFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features

Norval Pressure regulator

Transcription:

flotek.g 2017- Innovative Solutions in Flow Measurement and Control -Oil, Water and Gas August 28-30, 2017, FCRI, Palakkad, Kerala, India Influence of Reynolds number on Cd of Critical flow venturi nozzle and validation with ISO standards Sivaprakash.K.G (*),Unnikrishnan.P.N (**),Dr.Jacob Chandapillai (***) (*)Deputy Director, (**) Junior Research Engineer,(***) Director Fluid Control Research Institute (Under Ministry of Heavy Industries & Public Enterprises) Kanjikode West, Palakkad-678 623, Kerala, India Website: www.fcriindia.com E-mail ID: fcri@fcriindia.com Phone :91-491-2566206,2566120, Fax :91-491- 2566326 ABSTRACT Critical Flow Venturi Nozzles (CFVN) of throat diameter 3.32mm,4.698mm & 6.64mm having Nominal Flow capacities 6.3m 3 /h, 12.5m 3 /h & 25m 3 /h designed at Fluid Control Research Institute (FCRI), India as a transfer standard for undertaking calibration of master air/gas flow meters was tested at different operating pressures using the PVTt (Pressure, Volume,Temperature & time) & Primary standard Gravimetric system (PSGS) facilities of the Institute. The Nozzles fabricated was of Toroidal throat design and was assembled as per the general guidelines of ISO 9300:2005 Measurement of gas flow by means of Critical flow Venturi nozzles. The Air Flow Measurement facilities at FCRI is equipped with Primary Standards PVTt facility of 2m 3 Nominal volume and 90 m³/h maximum flow capacity and a Gravimetric system of 1500 ltr volume and with a maximum flow capacity of 50m³/h. The Coefficient of Discharge (Cd) of the Critical Flow Venturi Nozzle assemblies were established with both PVTt facility and PSGS at different operating pressures so as to achieve a wide range of Reynolds numbers as per the limits specified in ISO 9300 standard for toroidal throat venturi nozzles. Throat Reynolds number range of the nozzles tested were between 4.1E+04 to 1.3E+06. Cd of the nozzles established from Primary Calibration facilities were compared with the values determined using the ISO equation which is a function of Throat Reynolds number and Coefficients as specified in the standard. The uncertainty of ISO specified equation for toroidal throat Venturi nozzle is 0.3% at 95% confidence level and the Uncertainties of Primary Calibration facilities of FCRI are better than 0.1% and hence are at par for validation of experimental results with ISO values. Keywords: Critical Critical flow venturi nozzle, PVTt facility, Primary Gravimetric systam, Coefficent of discharge, Reynolds Number, Validation, Uncertainty 1. ABOUT FCRI The Fluid control Research Institute (FCRI) an autonomous R&D Institute was 1

established in 1987 with active assistance and participation from UNDP and UNIDO, under the Ministry of Industry (Govt. of India). The Institute is accredited by different National/International bodies such as National Accreditation Board for Testing and Calibration Laboratories (NABL, India); Underwriters Laboratory, USA; Chief Controller of Explosives, Nagpur-India; Bureau of Indian Standards (BIS); NMI, Netherlands; Department of weights and measures, India; Central Pollution Control Board(CPCB, India); Department of Science and Technology, India; etc. FCRI regularly participates in International round robin proficiency test programs in association with NIST, USA; CEESI, USA; NEL, UK; DELFT, Holland; DTI, Denmark; and KRISS, Korea etc. Other than flow laboratories with Air, Water and Oil as media, FCRI has got supporting laboratories like metrology section, Noise and Vibration section, and Electronics and Instrumentation section etc. FCRI regularly conducts National/International level Conferences/Training programs. FCRI also develops software in the field of flow meter design and selection valves/pumps, natural gas metering etc. 1.1 Facility at Secondary Air Flow Laboratory (SAFL) The SAFL of FCRI is equipped with critical flow venturi nozzles (sonic nozzles) of capacity 11.25-2880 m 3 /h. This facility is used for model studies, calibration/testing of flow products upto 10,000 m 3 /h at near ambient conditions using three cyclo blowers connected to the test loop. The reference critical flow venturi nozzles are assembled in parallel and are connected to the device under Calibration/test in series. The laboratory is maintained at a temperature of 25±1 C and humidity of 55±5 %. Positive pressure is ensured inside the facility by means of Air Handling units which supplies conditioned air to the laboratory. The schematic of SAFL is given in Fig. 1. Fig 1- Schematic of secondary air flow laboratory 2. DESCRIPTION OF PVTt FACILITY The heart of the facility is the cylindrical stainless steel volume vessel with two spherical dish ends having an approximate internal volume of 2 m 3. It is fitted with necks both at top and bottom. The schematic arrangement of the facility is shown in Fig. 2. The nozzle is connected to the vessel with a quick acting electropneumatically operated high vacuum butterfly valve using ISO KDN100 flanged fittings. The valve operations are controlled manually. Distribution of air temperature inside the vessel is measured using 12 Resistance Temperature Detectors inserted using thermo wells at desired locations spanning the height of the volume vessel. High vacuum couplings (ISO KF, ISO-K) and vacuum valves are provided to ensure sufficient leak tightness in the vessel. A vacuum pump evacuates the vessel to the required initial pressure (<10 Pa) in the vessel. The vacuum inside the vessel is monitored by means of two compact capacitance gauges with digital indicator. The final pressure inside the vessel is measured by means of a high accuracy multifunction digital pressure indicator with resonant sensors. Compact capacitance 2

gauges and RTDs are installed permanently in the vessel This unique primary calibration facility works on the principle of measurement of pressure (P), volume (V ), temperature (T ) and time (t ). It has been established to achieve the following prime objectives: i) Primary calibration of critical flow venturi nozzles having nominal flow capacities less than 90 m 3 /h with uncertainties better than ±0.1% using ambient air. ii) Establishment of traceability to National standards. The following are the salient features of the facility. Method :PVTt method Primary parameter : Volume Medium : Conditioned Air. Temperature : 25 ± 1 o C. Pressure of air in vessel: <10 Pa (a) (Initial) <75000Pa (a) (Final) Volume of vessel : 2 m 3 (Nominal) Collection time : 60 to 250 s Valve opening /closing time : Better than 60 ms Flow rate : 90 m 3 /h (maximum) Uncertainty : Better than ± 0.1% 3. DESCRIPTION OF PRIMARY STANDARD GRAVIMTERIC SYSTEM(PSGS) FACILITY AT FCRI The loop consists of a spherical vessel on a mass comparator, test meter mounted at upstream of a critical flow venturi nozzle and a bypass line opened to atmosphere interconnected with a Tee. The a pilot operated air regulator maintains a preset pressure at upstream of the Critical flow Venturi nozzle. The three pneumatically operated ball valves in the Tee stop or divert the flow from the test meter line to atmosphere or to the spherical vessel. The air passed through the critical flow venturi nozzle and the test meter is collected in a spherical air vessel. The mass of air collected is measured using a mass comparator. The time of collection of air is measured using a universal counter. The pressures are measured using multifunction pressure indicators. The temperatures are measured using digital temperature indicators with RTD. The pulse for a preset time period is measured using universal counters. The relative humidity is measured using pressure dew point measuring instrument. Fig 3 - PSGS facility at FCRI Fig 2- PVTt facility at fcri 3

4. TECHNICAL SPECIFICATION OF CRITICAL FLOW VENTURI NOZZLE (CFVN) PACKAGE Nominal flow capacity : 6.3/12.5/25m 3 /h Type : Toroidal throat Venturi nozzle Design standard : ISO 9300: 2005 Throat diameter : 3.32/4.698/6.64mm Pipe size : 2 NB Fig. 4- Typical assembly of a CFVN Schematic of typical CFVN assembly is shown in Fig.4. Three numbers of CFVN s of throat diameters 3.32, 4.698 and 6.64mm were used in the experimental program. The above Critical flow venturi nozzles were calibrated using PVTt facility in suction method and was further calibrated using the PSGS facility at different operating pressures from 2 to 20 bar abs so as to achieve different Reynolds number range as specified in the ISO standard 9300 for Toroidal throat venture nozzle. 5. CALIBRATION OF CRITICAL FLOW VENTURI NOZZLE WITH PVTT FACILITY Critical Flow Venturi Nozzles are secondary flow measuring devices that operate at the maximum possible flow rate for the existing upstream conditions. The air flow accelerates to the critical velocity at the throat, which is equivalent to the local sound velocity. The nozzle has a throat diameter of 3.32 mm and was assembled in 50 mm NB pipe with a flow straightener at upstream. For this nozzle, to ensure critical flow, a maximum back pressure ratio less than 0.8 must be maintained across the nozzle.the nozzle under calibration is mounted at upstream of the quick acting butterfly valve. With the valve in closed condition, an initial pressure less than 10 Pa (a) is created in the vessel using a vacuum pump. Subsequently the pump is isolated and pressure inside the vessel is monitored to detect leakage if any. Sufficient time is allowed for temperature and pressure stabilization inside the vessel. By opening the quick acting butterfly valve, the ambient conditioned air is drawn into the vessel through the nozzle and simultaneously triggers a time counter. When the pressure of air in vessel reaches about 75 kpa (a), the valve is closed and the signal from proximity sensor stops the timer at the same time. Thus the timer measures the duration of valve opening or time of collection of air in the vessel. More than 2 hours is allowed for stabilization of vessel conditions. The final pressure and temperature of air in the vessel are recorded after stabilization. The actual mass of air is calculated from the difference between the initial and final density of air in the vessel and established volume of PVTt vessel. Coefficient of Discharge (Cd) of the Critical flow venturi nozzle was established by the ratio of actual mass flow rate and theoretical mass flow rate estimated using ISO 9300 specified equations. The entire calibration is repeated a number of times to establish the mean Cd and to assess the random uncertainty associated with calibration. The tests were repeated with CFVN s of throat diameters 4.698 & 6.64 mm. 6. CALIBRATION OF CRITICAL FLOW VENTURI NOZZLE WITH PSGS FACILITY 4

Required operating pressure was maintained at the upstream side of the CFVN using the dome loaded pressure regulator and the actual mass flow rate was determined using the 1.5m 3 volume tank and the Mass comparator weighing system. The mass of air collected in the weighing system varied from 0.5 to 16 kg approximately depending on the operating pressure and nominal flow capacity of the CFVN. Collection time in the vessel was in the range of 60 to 360 seconds depending on flow rate. Cd values were determined for different operating Reynolds number range by this method. Sl. No. Throat Experimental values Coefficient of Nominal Method Reynolds Actual discharge Deviation Coefficient of using from pressure adpted number flow rate ISO 9300 discharge ISO equation values Red ma Cd Cd ISO bar a - kg/h - - % [1] [2] [3] [4] [5] [6] [7] [8] 1 20 1.02000 8.32E+05 147.80 0.9969 0.9929 0.40 2 18 7.52E+05 133.49 0.9972 0.9928 0.44 3 16 6.67E+05 118.18 0.9977 0.9926 0.52 4 14 5.88E+05 103.96 0.9984 0.9924 0.61 5 12 PSGS 5.01E+05 88.51 0.9979 0.9921 0.59 6 10 4.22E+05 74.44 0.9976 0.9917 0.60 7 8 3.43E+05 60.17 0.9974 0.9913 0.62 8 6 2.61E+05 45.79 0.9977 0.9906 0.71 9 4 1.75E+05 30.56 0.9966 0.9894 0.73 10 2 9.39E+04 16.45 0.9943 0.9870 0.74 11 1 PVTt 4.16E+04 7.17 0.9900 0.9826 0.76 Fig 5- Variation of Cd with Reynolds number for 6.3 m3/h CFVN 7. ISO EQUATION FOR CD OF CRITICAL FLOW VENTURI NOZZLE As per ISO 9300: 2005 the equation for Cd of Toroidal throat CFVN is Coefficient of discharge,cd 1.01000 1.00000 0.99000 0.98000 0.97000 Actual Coeffiecent of discharge ISO Coefficent of discharge Cd = a-bre nt -n 0.96000 0.00E+00 1.00E+05 2.00E+05 3.00E+05 4.00E+05 5.00E+05 6.00E+05 7.00E+05 8.00E+05 9.00E+05 Throat Reynolds Number, Red Where a= 0.9959, b= 2.72, n= +0.5 for Throat Reynolds number range of 2.1 E+04 to 3.2E+07. Constants a, b & n depends upon the type of Critical flow venturi nozzle and the corresponding values are given in the ISO 9300 standard and the specified uncertainty of Cd values by the above equation is 0.3% for Toridal throat CFVN. 8. EXPERIMENTAL CD VALUES AND ISO VALUES Cd values, Reynolds number range and the deviations from ISO values for the three tested CFVN s are given in Tables 1 to 3 and Figs 5 to 7. Sl. No. Nomin al pressur e Throat Method Reynolds Actual adopted number flow rate Red ma Coefficient of discharge Cd Coefficient of discharge using Deviation from ISO equation Cd ISO ISO 9300 values bar a - kg/h - - % [1] [2] [3] [4] [5] [6] [7] [8] 1 20 Table 2 - Cd values for 12.5m3/h flow capacity CFVN 1.13E+06 288.5 0.9966 0.9933 0.33 2 18 1.04E+06 266.1 0.9971 0.9932 0.39 3 16 9.25E+05 235.7 0.9976 0.9931 0.46 4 14 8.00E+05 203 0.9982 0.9929 0.54 5 12 6.98E+05 176.3 0.9982 0.9926 0.56 PSGS 6 10 5.92E+05 149.3 0.999 0.9924 0.67 7 8 4.66E+05 117.5 0.9993 0.9919 0.75 8 6 3.56E+05 89.53 0.9986 0.9913 0.74 9 4 2.43E+05 61.1 0.9959 0.9904 0.56 10 2 1.28E+05 32.01 0.9964 0.9883 0.81 11 1 PVTt 5.89E+04 14.37 0.9876 0.9847 0.30 Table 1- Cd values for 6.3m3/h flow capacity CFVN 5

1.02000 1.01000 Experimental Cd values ISO Cd values 9. ESTIMATION OF UNCERTAINTY IN PVTT & PSGS METHODS OF CALIBRATION. Coefficient of discharge,cd 1.00000 0.99000 0.98000 0.97000 0.96000 0.00E+00 2.00E+05 4.00E+05 6.00E+05 8.00E+05 1.00E+06 1.20E+06 Throat Reynolds Number, Red Fig 6 - Variation of Cd with Reynolds number for 12.5m3/h CFVN The overall uncertainty in the calibration of critical flow venturi nozzle using PVTt and PSGS methods are estimated by combining the uncertainties of actual mass flow measurement and the theoretical flow rate from the nozzle. The equation for theoretical flow rate is: 2 * m =π d / 4 C P M /( R T t o e o ) Throat Sl. No. pressure adpted number Coefficient of Nominal Method Reynolds Coefficient of discharge using Deviation from Red bar a - discharge Cd ISO equation ISO 9300 values Cd ISO - - % [1] [2] [3] [4] [5] [6] [7] 1 2 3 4 5 16 10 6 5 PSGS 1.30E+06 0.9924 0.9935-0.11 8.40E+05 0.9929 0.9929 0.00 5.00E+05 0.9923 0.9921 0.02 4.20E+05 0.9904 0.9917-0.13 1.5 BELL PROVER 1.24E+05 0.9867 0.9882-0.15 11 1 PVTt 8.28E+04 0.9826 0.9864-0.39 Table 3- Cd values for 25m3/h flow capacity CFVN 1.02000 1.01000 Actual Coeffiecent of discharge ISO Coefficent of discharge Where P 0 and T 0 are the pressure and temperature measured at the upstream of the nozzle, d is the throat diameter and C * is critical flow factor. The coefficient of discharge of the nozzle defined as the ratio of actual mass flow rate to theoretical flow rate is calculated as: 2. C d = ma mt The systematic uncertainty in the coefficient of discharge E ) is given as the s ( C d combination of individual component uncertainties of the above equations. For PVTt method and PSGS uncertainty in actual flow measurement is estimated and it is combined with uncertainty in theoretical flow rate to estimate the total uncertainty in determination of Cd. Coefficient of discharge,cd 1.00000 0.99000 0.98000 Uncertainty in the Variables of PSGS method and PVTt are given in Tables 4 & 5 0.97000 0.96000 0.00E+00 2.00E+05 4.00E+05 6.00E+05 8.00E+05 1.00E+06 1.20E+06 1.40E+06 Throat Reynolds Number, Red Fig 7- Variation of Cd with Reynolds number for 25m3/h CFVN 6

(P) ±0.03% (V) ±0.1% (T) ±0.28% (t) ±0.007% E ( o ) ±0.03 % s P s( Me E ) ±0.0013 % (R) ±0.01 % ( T o ) ±0.28% Table 4- (PSGS Method) (M) ±0.04% (P) ±0.02% (T) ±0.28% (t) ±0.002% ( P o ) ±0.02% E s( Me) ±0.0013% (R) ±0.01% E ( o ) ±0.28% s T Table 5- (PVTt method) Expanded uncertainties in the Primary standard facilities of FCRI, PVTt and PSGS methods are estimated as better than 0.1% as per NABL 141 guidelines. 10. OBSERVATIONS BASED ON EXPERIMENTAL DATA Deviation between ISO Cd values and Cd values obtained by PVTt method and PSGS method at 1bar abs & 2bar abs results are matching and magnitude of deviation (0.75%) is same as indicated in the previous table for SN5. Generally increasing trend in Cd values is seen with increase of Reynolds number. Same trend is reported in ISO 9300 equation also. Variation of 1% is reported in ISO equation with variations in operating pressures from 1 to 20 bar abs. Maximum variation of 0.9 % is observed in Cd values obtained by experimental method in the operating pressures from 1 to 20 bar abs. At higher operating pressures deviation of Cd from ISO values is only 0.4%. Deviation up to 0.9% is observed for lower operating pressures (From ISO values). There could be an uncertainty component of the order of ±0.3 to 0.4% in Cd values due to variations in pressure from 2 to 20bar and the same has to be accounted in uncertainty estimates. At Higher Reynolds number the experimental Cd values are more closer to ISO values. Closeness to ISO Cd values is more dependent on operating density and hence the Reynolds number rather than the throat diameter of the nozzle. 11. CONCLUSION Experimental Cd values can vary up to 0.9 % from ISO values depending on operating pressures from 1 bar abs to 20bar abs. For 25m3/h CFVN also for higher Reynolds number deviation of Cd from ISO values is lesser when compared to lower Reynolds number values. With increase in Reynolds number Cd values are closer to ISO values. Instead of assuming constant values of Cd, for the operating pressure range from 1 to 20bar abs, 7

interpolated Cd values based on Reynolds number is to be used for flow estimation in precision flow calibration application. Possible causes for Cd variations from ISO values may be due to Machining inaccuracies, surface finish variations, straight length requirement variations, disturbances in flow due to upstream fittings etc of CFVN from ISO recommendations. Recommended Surface roughness for throat and inlet profile of CFVN has to be better than 15 10-6 d as per ISO recommendations. Roughness values of in-service CFVN s at FCRI facilities are on the higher side by 10 to 20 %. ACKNOWLEDGEMENT Authors would like to express their gratitude for the team HPATF (High pressure air test facility) of FCRI comprising of Mr.A.S.Murali; Deputy Director, Mr.C.B.Suresh; SRE, Mr.Sandeep; JE and Miss.Aswathi, JE for their timely technical support and assistance for conducting the experimental program of CFVN at FCRI facilities. REFERENCES [1] ISO 9300:2005 (E) Measurement of gas flow by means of critical flow venturi nozzles [2] Flow Measurement Engineering Handbook, Richard W. Miller [3] Industrial Flow Measurement, David W Spitzer 8