Sensepoint XCD. Calibration Guide

Similar documents
Technical Manual. Sensepoint XCD Gas Detector

CALIBRATION KIT. Instruction Manual

RGC-IR Remote Gas Calibrator for IR400

Operation and Maintenance Manual

Because every life has a purpose. MSA Z GARD S Sensor Calibration Procedure

BI-680 Online Dissolved Oxygen Controller Instruction Manual

Phosgene Sensor. Sensoric COCl2 3E 1

GAS DETECTION: CALIBRATION GAS

Ozone. Sensoric O3 3E 1

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

X-dock Software History

Portable Gas Monitor GX User Maintenance Manual (H4-0050)

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

RK-IR Sample Draw Aspirator Adapter Operator s Manual

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

Bristol Fire Department Standard Operating Procedures. Multi Gas Detector

Columbus Instruments

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Operation and Maintenance Manual

EasyLine Continuous Gas Analyzers. So smart, they re simple

CSA Sample Draw Aspirator Adapter Operator s Manual

G100 Range. Operating Manual

SDM-2012 Docking Station Standalone Configuration Operator s Manual

G100 RANGE G100, G110, AND G150 OPERATING MANUAL

Continuous gas analysis ULTRAMAT 23 Supplement 2 for Operating Instructions C79000-G52xx- C216-03

OPTIMA 7 - BIOGAS Calibrating the gas sensors SENSOR ADJUSTMENT. Diamond Scientific 625 Peachtree St., Cocoa, Florida /Tel:

Instruction Manual. Biogas Analyzer InCa Bio 04

ITEMS REQUIRED TO CALIBRATE A STANDARD GAS DETECTOR USING A 34L, 58L, 78L OR 103L CYLINDER OF GAS:

Advance Optima -- High Performance Measurement Technology Oxygen Analyzer Module Magnos 16/17

Ultima. XI Infrared Gas Monitor

Gas-Detection Instruments

A180 DO/BOD/OUR/SOUR Meter Instruction Manual

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

8.1 Calibration of gas sensors

DF-550E PROCESS ANALYSERS APPLICATIONS FEATURES

2018 National Metal and Nonmetal Mine Rescue Contest. Technician Team Competition Written Test (Dräger BG-4) Good Luck!

Heavy Duty Dissolved Oxygen Meter

Automatic Test and Calibration Station

GASGUARD VENT LINE3 Ammonia Sensor OPERATING & INSTALLATION MANUAL

EX-5150-MOS Sensor/Transmitter For PPM or %LEL Manual

HCL MONITORING. ICAC MARAMA WORKSHOP May 18-19th, Presented by: Ty Smith. CEMTEK

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

CF8-W-Disp-CO. User manual. CO 2 / CO sensor with built-in general purpose controller

Using the UltraRAE. Firmware 2.35

White Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology:

AUTOMATIC TIRE INFLATOR # MW-60, MW-60-4WAY & MW-64HP

Continuous Gas Analysis In situ laser gas analyzers TÜV and MCERTS add-on for LDS 6 operating instructions Compact Operating Instructions

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

An area may be considered a confined space if it meets three criteria.

Dilution Manual Cannonball 3 Multi Gas Detector

TABLE OF CONTENTS INTRODUCTION 3 SAFETY PRECAUTIONS 3 PACKAGE CONTENTS 4 DEVICE OVERVIEW 4 BUTTON OPERATION SUMMARY 5 BASIC OPERATION 6

Spot the Hazards: Review Confined Space Hazards Missed by Most Safety Professionals, and the PPE Solutions that Meet OSHA Requirements

Advanced Calibration Designs, Inc. goacd.com. Certified Calibration Gas Solutions

Instruction and Maintenance Manual

TEL/jlRE" Introduction. Display Features and Modes. Startup Procedure. Power-Up Procedure. Adjustment Modes

Operation Manual. Pro CO O 2 Alarm. Carbon Monoxide and Oxygen Analyzer. Rev

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

GasLab SAN-102 Micro Oxygen Monitor User Manual

GasSense NDIR User Manual

Plasma. Argon Production Helium Liquefaction Truck Loading Pure Gas Bottling

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

FLAMMABLE GAS SENSOR/TRANSMITTER

Generating Calibration Gas Standards

SPECIFICATION. date of manufacture. Intermittent: -20 o C to +50 o C. Pressure Range:

Simrad GD10P Extended On-site Maintenance

BASCOM-TURNER INSTRUMENTS

RAM Operation Manual

This test shall be carried out on all vehicles equipped with open type traction batteries.

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

User Manual. asense miii. / CO sensor with built-in general purpose controller

RK LEL Sample Draw Aspirator Adapter Operator s Manual

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

EVM Series: Frequently Asked Questions

Gas-Detection Instruments

Hot extractive multi-channel gas analyser for continuous process and emissions monitoring.

QRAE II Quick Reference Covers Diffusion & Pump Models with Firmware Version 3.12 and higher

Ultima. X Series Gas Monitor

Calibration Requirements for Direct Reading Confined Space Gas Detectors

Addressing the World Shortage of Helium

Basic Nitriding Sampling System Hydrogen Analyzer with Calculated % DA, % NH 3, and K N Values. Operations Manual

Sensoric 4-20 ma Transmitter Board Operation Manual

Do Not Print This Page.

Model Description

DF-310E PROCESS ANALYSERS APPLICATIONS FEATURES

Verification Of Calibration for Direct-Reading Portable Gas Monitors

Hot extractive multi-channel gas analyser for continuous process and emissions monitoring.

In Vivo Scientific, LLC INSTRUCTION MANUAL

BAPI Pressure Line of Products - FAQs

Dr. Thiedig DIGOX 5 CO 2. In-line measurement of CO 2 for constant product quality. Anlagen- und Analysentechnik

Datasheet: K-30 ASCII Sensor

Model GFC 7000T CO 2 Analyzer (Addendum to GFC 7000TA Manual, PN 07272)

SPECIFICATION. date of manufacture. Intermittent: -20 o C to +50 o C. Pressure Range:

Overview. Front Panel: Keypad and Display

Operation Manual. Pro 4 Warn Analyzer TM Rev 08.17

ABB MEASUREMENT & ANALYTICS DATA SHEET. WaterMaster FEW530 Electromagnetic water meter

G200 Analyzer Range. Operating Manual

HLM-2000-TX Gas Transmitter User Manual

MODEL GENERA PURGE TURBINE INSTR DANGER HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM.

Transcription:

Sensepoint XCD Calibration Guide

Sensepoint XCD Calibration Calibration Handbook It is recommended to periodically carry out a gas response check on the Sensepoint XCD to ensure correct operation. This may be done in two ways; 1. A simple Response Check often referred to as a BUMP TEST is a test using calibration gas applied to the sensor via the nozzle of the Weather Protection or using the Sensepoint XCD Gassing Cap. If a BUMP TEST is done via the Weather Protection nozzle it may be necessary in windy conditions to increase the flow rate of the test gas by a further 1 LPM, OR, to shelter the weather protection from the wind. 2. A full gas calibration of the sensor as described in the following section, using ONLY the Sensepoint XCD Gassing Cap (Part No.: S3KCAL). Zero and Span CAUTION Before initial calibration allow the detector to stabilize for 30 minutes after applying power. When in zeroing and span calibration mode the current output from the detector is inhibited (default 2mA) to avoid false alarms. For Flammable gas calibration use a calibration gas concentration of between 25%LEL and 75%LEL to ensure that the required accuracy can be attained. To calibrate the detector, use an appropriate span gas cylinder, constant flow regulator AND the Sensepoint XCD Gassing Cap. The flow rates used for calibration gas are as follows: Gas Type Flow rate (L / Min) Air or N2 for Zero 0.5 to 1.0 Flammable CAT 1 to 1.5 O2 0.5 to 1.0 H2S 0.5 to 1.0 CO 0.5 to 1.0 H2 0.5 to 1.0 NO2 0.5 to 1.0 Flammable IR 0.4 to 0.6 CO2 IR 0.4 to 0.6 A compressed air cylinder (20.9%Vol oxygen) should be used to perform the zero calibration if the area where the detector is located contains any residual amount of the target gas. If no residual gas is present then the background air can be used to perform the zero calibration. Contact your Honeywell Analytics representative for details of suitable calibration kits. To calibrate the detector follow the procedure below. NOTE Sensepoint XCD 123

Calibration Handbook Zero Calibration 1. If the ambient air is NOT considered reliable to use to set the ZERO, then remove the weather protection and fit the Gassing Cap accessory onto the sensor and apply a clean source of zero gas or compressed air. 2. To access the calibration menu, hold the end of the magnet over the switch located at the top center of the detector display ( ) for at least 3 seconds and then remove. 3. The display will indicate the first configuration mode menu SEt CAL. 4. Put the magnet over the switch again and move to enter the Calibration menu. 5. The display will show the current gas reading, and the icon flashes. 6. When the zero gas reading is stable use to confirm zero calibration. 7. If successful the display shows ZEro PASS (if not successful, the display shows ZEro FAIL and returns to configuration mode). 8. If using zero-air, turn it off. Zeroing is complete and saved. 9. The display shows SPAn with YES flashing. 10. If span calibration is required use proceed to the next step. If span calibration is not required, use to select No and to return to configuration mode. 124 Sensepoint XCD

Span Calibration Calibration Handbook 11. The display shows the current calibration span gas concentration while flashing the icon. Use to change the calibration span gas concentration, and when required span calibration level is set. 12. The display will show the current gas reading, and the icon flashes. 13. Connect the regulator to the span gas cylinder. 14. Apply the span gas to the sensor using the Sensepoint XCD Gassing Cap. The live gas reading is displayed. When the reading is stable, use to confirm span calibration. 15. If the sensor has been replaced the following display may be shown. 16. Use to select YES if the sensor has been replaced or No if it has not been replaced. 17. If the span calibration is successful the instrument will briefly display SPAn PASS (if fails SPAN FAIL displayed and returns to configuration mode). NOTE 18. The display alternates between Purg gas and the gas reading to indicate that the unit is expecting the span gas to be removed from the sensor. 19. Promptly switch off the calibration span gas and remove the Sensepoint XCD Gassing Cap from the sensor to allow the gas to disperse. Sensepoint XCD 125

Calibration Handbook 20. When the reading falls below 50% of the calibration gas level the display indicates a countdown (up to 180 seconds dependant on gas type). 21. When the countdown is finished, the calibration procedure is complete. 22. The instrument returns to the Set CAL menu. Activate the or switch to select another menu or select QuIT to return to normal monitoring mode. NOTE Zero and Span Calibration of Hydrogen Sulfide sensors Hydrogen Sulfide sensors can be affected by extreme humidity changes. A sudden increase in ambient humidity can result in a short-term positive drift in the instrument s reading. A sudden decrease in ambient humidity can result in a short-term negative drift in the instrument s reading. These are most likely to be noticed during calibration with dry or cylinder gas. When calibrating Hydrogen Sulfide cartridges the following should be taken into account while following the Span/Zero procedure: 1. To zero the sensor, use a compressed air cylinder of 20.9%Vol oxygen (not Nitrogen). Do not use background air. 2. Apply the gas to the sensor for three minutes before using to confirm the zero calibration. 3. If a span calibration is to be performed, the span calibration gas should be applied to the sensor immediately after the zeroing procedure. Do not allow the sensor to return to ambient air conditions between steps 2 and 3. 4. Apply the span gas for two minutes before using to confirm the span calibration. Authorized Distributor: GasDetectorsUSA.com Houston, TX USA 832-615-3588 sales@gasdetectorsusa.com 126 Sensepoint XCD