Simrad GD10P Extended On-site Maintenance

Similar documents
Technical Data Sheet MF010-O-LC

GG-VL2-NH3 AMMONIA VENT LINE SENSOR. Installation and Operation Manual

OPERATION MANUAL NTF-15

Coda cx1 Thermal Cycle Test Procedure

BI-680 Online Dissolved Oxygen Controller Instruction Manual

Instruction Manual. PLT Series. 1. General Specification. 2. Installation. 3. Wiring. 4. Function. 5. Setting & Calibration

Transmitter CS 21 Operation Manual

GASGUARD VENT LINE3 Ammonia Sensor OPERATING & INSTALLATION MANUAL

Instruction and Maintenance Manual

Installation and Operation Manual

INSTRUCTION MANUAL. 840 Transmitter. Loop Powered Dissolved Oxygen Transmitter

Gas Pressure Probe. Information Sheet. Total Dissolved Gas

Transmitter CS 21 Operation Manual

Magnetic level switch type MR783 Instruction Manual

Installation, operating and maintenance Instructions for Seemag bypass level indicator

Portable Gas Monitor GX User Maintenance Manual (H4-0050)

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

OPERATION MANUAL NTF-60 Plus

TEL/jlRE" Introduction. Display Features and Modes. Startup Procedure. Power-Up Procedure. Adjustment Modes

PULSAR 5000 SERIES OPERATING & INSTALLATION INSTRUCTIONS SERIES 5000 PLEASE READ CAREFULLY BEFORE INSTALLING

VAPOMATIC ANESTHETIC VAPORIZER

GEMS SENSORS & CONTROLS

INSTRUCTION MANUAL Pressure Relief Device LPT

Operation and Maintenance Manual

Simple, Robust, Reliable Measures ph Doses either acid (lower) or alkali (raise) Displays ph, Total Dose Count (DCT) and Dose Count per hour (DC/h)

GasSense NDIR User Manual

Basic Nitriding Sampling System Hydrogen Analyzer with Calculated % DA, % NH 3, and K N Values. Operations Manual

Maintenance Manual. Liquid Level Transmitter CT801-SI, CT801-TA

97C COMPRESSOR KIT 12V PART NO C COMPRESSOR KIT 24V PART NO C COMPRESSOR KIT PART NO

Operation Manual - PN A MENSOR MODEL 73 SHOP AIR BOOSTER

Datasheet: K-30 ASCII Sensor

ACV-10 Automatic Control Valve

T EK-SUB 4800C 19 mm Submersible Level Transmitter

Nuclear Associates

Instruction and Maintenance Manual

! Warning, refer to accompanying documents.

SAUTER FK Version /2016 GB

Ambient Weather GL150-TH Nautical Thermometer Hygrometer (ComfortMeter) User Manual

Operation and Maintenance Manual

MODEL GT820 OXYGEN SENSOR

WIKA INSTRUMENT CORPORATION

Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL -

FX134, FX1234, FX3030 Problem and Solutions

RGC-IR Remote Gas Calibrator for IR400

Model B Wire Wet Oxidant Gas Transmitter. 6 Iron Bridge Drive Unit 1 & 2 Gatehead Business Park

210 Series Transmitter with External Electrochemical Sensor

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr

RAM 4021 Operation Manual

Installation and operating manual. Pneumatic control station LK product no: PCS 1-10

OFFICINE OROBICHE S.p.A. 1/9

Instruction Manual Contact Pressure Vacuum Gauge

BS12N Piezoresistive OEM Pressure Sensor

Heavy Duty Dissolved Oxygen Meter

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

Safety Manual. Searchpoint Optima Plus

User Instruction Manual

KGW - ISOTHERM. Operating Instructions. Level controlling device for liquid nitrogen LEVEL CONTROL LN2

Instrument Operating Manual

Columbus Instruments

Operating Instructions. Ball valve fitting according to ZB For pressure transmitter VEGABAR 82. Document ID: 50027

FTC130 Transmitter. Operating Manual

STARLOG. Capacitive Water Depth Probe

PURA. Pure Gas Dewpoint Transmitter. Users Guide

Coriolis Mass Flow Meter

PAPR Flow Tester. instructions. part no WARNING

Air Bubbler Depth Gauge DG2200 Installation and Reference Manual

Operating Instructions Part No

Do Not Print This Page

DPC-30 DPC-100. Reference Manual

Budget Range Operators Handbook

User s Guide Temperature Sensor Converter TSC-599

Sensoric 4-20 ma Transmitter Board Operation Manual

KERN EG/EW Version /02

Test and Adjustment Instructions Compressed Air Processing System Air Processing Unit (APU)

Alphasense 4-20 ma transmitters offer convenience and easy maintenance for toxic sensors:

BS12 Piezoresistive OEM Pressure Sensor

Electro-Pneumatic Converter YT-940 SERIES

Peak Pressure Indicator MSI-3

NOVALYNX CORPORATION MODEL BAROMETRIC PRESSURE SENSOR PCB MODEL BAROMETRIC PRESSURE SENSOR IN NEMA ENCLOSURE INSTRUCTION MANUAL

Depth sensor. Product reference : REV 1. USER GUIDE and INSTALLATION GUIDE. nke Sailing competition

Maintenance handbook

Gas density monitor With integrated transmitter Model GDM-100-TI

Specifications and information are subject to change without notice. Up-to-date address information is available on our website.

DL 8 Dome Port. Product Registration. Product Number 75340

Handi+ OPERATING MANUAL & INSTRUCTIONS FOR USE. R218P15 Industrial. R218M15 Rev. D

FTS SUBMERSIBLE PRESSURE TRANSMITTER USER S MANUAL

Model 106 DPI "Micro-switch" Installation and Operating Instructions

BGA244 Binary Gas Analyzer

Pneumatic high-pressure controller Model CPC7000

HLM-2000-TX Gas Transmitter User Manual

SAFETY MANUAL FOR FLAMMABLE PRODUCT TRANSFER

CALIBRATION KIT. Instruction Manual

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

Pressure Automated Calibration Equipment

Target disk flowmeter Series DP. Instructions manual DP65 DP500. R-MI-DP Rev.: 4 English version

Installation Instructions

Installation Instructions

LK-SX VOC. Application. Security Advice Caution. Notes on Disposal. Combined sensor mixed gas. Data sheet

E2K-L. Liquid Level Sensor That Is Unaffected by the Color of the Pipe or Liquid. Liquid Level Sensor. Ordering Information

Transcription:

Simrad Extended On-site Maintenance This procedure covers maintenance of gas detectors with the aid of a handheld terminal (re-zeroing tool). Please also refer to the standard operating manual (P3359E). Reference: Extended on-site maintenance.doc

TERMS FOR USING THE HAND-HELD TERMINAL IN FIRST LINE SERVICE. Preamble The detector is calibrated and temperature tested in a controlled environment at our factory in Oslo, Norway. No further calibration is required during the life time of the product, as the zero point and calibration will stay inside the given tolerances. This implies that slight offsets from the zero point of 4.0mA (0%LEL) is to be expected and values up to 4.5mA at room temperature are within specification. In some environments with excessive heat and/or vibration we have noticed that a higher percentage of the detectors develop an offset outside the specified tolerances. These detectors should normally be returned to the factory for service. An alternative is to do this re-zeroing on site with a dedicated hand held terminal. Training The hand held terminal shall be used by personnel trained specifically by Simrad staff only. Training may be done during a survey on site as on-the-job training (preferred), or at our factory in Oslo. Reporting A reporting scheme is to be agreed upon. All re-zeroing should be reported, either for each detector, or a monthly report. This is needed for the service log for each detector. Re-zeroing We do not recommend re-zeroing of detectors with less than 5.0 ma (6% offset) at room temperature. The actual re-zeroing shall be done in a controlled environment with a temperature of 20 25 degc, following the procedure for field re-zeroing of. Responsibility Simrad Optronics ASA has no responsibility for faults introduced by the on site re-zeroing. Page 2 of 11

TEST OF ZERO POINT WHEN IS CONNECTED TO A SYSTEM 1. Read zero point value on the system display If zero point is too high, clean lens and mirror as follows: 2. Remove the Weather Protection unit by loosening the two screws on the circular front plate. 3. Clean lens and mirror using soap water (e.g. a dishwashing liquid mixed with water. Isopropanol is also recommended). Rinse and dry with a clean, soft paper towel. Take care to clean the lens and mirror circumference. Clean lens and mirror even if they appear to be clean in case there is an invisible layer of oil on the glass surface. 4. Mount the Weather Protection unit. 5. Wait approxx. 20 minutes, letting the reading stabilise. Read zero point value. If the zero point is within specification; OK. If zero point still is too high, the sensor must be decoupled from the system and tested at the instrument shop according to the following approved procedure: Test of zero point at GD10 instrument shop. Page 3 of 11

TEST OF ZERO POINT AT GD10 INSTRUMENT SHOP Required equipment Power supply 24 VDC / 0.3 A Digital voltmeter / amperemeter Sample Flow housing GD10 reg.no. 499-810874.8 Nitrogen gas 1. Remove the sensor from the system. 2. Place the sensor in a clean room where measuring equipment and nitrogen gas are available. Clean instrument air may be used if nitrogen is unavailable. 3. Clean the sensor optics again, both mirror and lens. 4. Hook up the sensor as shown in fig. 1. (Note: Establish whether the sensor has sink or source output). 5. Mount Sample Flow housing with gas hose as shown in fig. 2. 6. Switch on power to the sensor and wait for approx. 2 minutes until the output has stabilized. 7. Apply nitrogen gas to the Sample Flow housing. 8. Read the milliampmeter. 9. If output is outside sensor specification, then a re-zero is applicable. In some cases, you may also have to ventilate the detector housing. Page 4 of 11

VENTING THE HOUSING 1) Turn off power. 2) Remove Sample Flow housing. 3) Clean the sensor around the flame-proof joint marked with an A in fig. 3. This procedure prevents dirt from entering the interior of the sensor during reassembly. 4) Unscrew the 4 screws marked B in fig. 4. NOTE: The red seal on two of the screws must be broken. 5) Place the sensor right in front of you on a bench, place one hand on the housing and one hand on the front nose. Continue by pulling carefully apart. NOTE: The flame-proof joint has very fine tolerances. Consequently, to avoid that the two parts become stuck, it is crucial to pull them straight apart without bending. 6) Check the O-ring for possible damage. Replace if neccesary. 7) Let the sensor stay open for approx. 30 sec. and flutter a sheet of paper for air circulation. 8) Reassemble the parts using the four screws and washers that were removed. Torque: approx. 300 cnm. Note: Ensure that the plastic screen covering the electronics is not squeezed in the flame-proof joint. 9) Point 14 to 16 must be performed quickly within a few minutes (max. 5 minutes) to avoid the desiccator inside the sensor starting to absorb humidity. You may also replace the desiccator (white bag). 10) Check that the lens and mirror surface are clean. If not, clean according to procedure. 11) Mount Sample Flow housing. 12) Connect an ohmmeter betweeen terminal list 2 and the sensor chassis (ref. fig. 5). Ensure that there is no connection. If there is connection, contact Simrad Optronics ASA. 13) Repeat point 5, 6, 7 and 8 above (Test of zero point). 14) If sensor still does not meet specifications, perform a zero calibration using the hand-held terminal. 15) Fill in the Inspection Report data sheet (appendix). It is important that the opening of the detector is reported, as this normally voids the warranty. Page 5 of 11

RE-ZEROING USING THE HAND-HELD TERMINAL Power supply GND +24VDC 1 1 2 2 Hand Held Terminal 4-20mA output 3 4 4 5 5 If zero point is outside specifications after the zero point test procedure, the sensor can be re-zeroed using the hand held terminal. IMPORTANT: The sensor shall only be re-zeroed after performing the complete zero point test procedure. Equipment: Hand held terminal Power supply 24V, 0.3 A Volt/Amp meter Sample flow housing Nitrogen gas Page 6 of 11

Procedure: 1. Connect detector, power and terminal according to illustration. 4. Mount the Sample flow hosing, connect the nitrogen gas bottle. 2. Switch on power. 3. Terminal will display start-up window. Select MENU by pressing ENTER key. 4 The terminal will identify the detector, and the following functions can be performed: -Display gas value. This function displays the optical transmittance( 0 100 % ), and the gas measurement in 0 100 % Full Scale. Optical transmittance should be 100 +- 30 with clean detector optics. ( Fill in values in Inspection Report data sheet ) ( If outside limit; send to Simrad for repair. ) -Fault messages. This function displays fault messages from the sensor. ( If fault message; send to Simrad for repair ) -Zero calibration. Starts zero point calibration routine. 5. Before starting the zero point calibration, verify following: -Ambient temperature must be between 15 C and 25 C. -Power shall be switched on for at least 15 minutes. -Apply a low flow of nitrogen gas to the Sample flow housing. 6. Find Zero calibration function by pressing UP or DOWN key, and select the function by pressing the ENTER key. The terminal will display the zero point calibration conditions. 7. Start the zero point calibration when the terminal displays Start zero cal by pressing the ENTER key. The terminal will display Calculating zero. Please wait app. 3 (6) min. When the zero point calibration is finished ( after app. 3 minutes ), the terminal will display Zero calibration completed. DISPLAY GAS VALUE, and the sensor will switch to normal gas measurement mode. NOTE; s made before summer 1998, have a 6 min. duration zero point calibration routine. 8. Press ENTER, and verify that the detector zero point is OK. 9. Do a calibration check as described in the operating manual to verify operation. 10. Fill in Inspection Report data sheet including serial no. ( appendix ) Page 7 of 11

Power GND +24V Terminal A 1 2 3 4 5 GD10 GD10 Test set-up (source output) Power GND +24V Terminal A 1 2 3 4 5 GD10 GD10 Test set up (sink output) Figure 1 Test set-up source and sink output Page 8 of 11

Figure 2 Set-up calibration test Page 9 of 11

A Figure 3 Flame-proof joint B Figure 4 Screws and washers Page 10 of 11

Ohm Chassi Figure 5 Connection of ohmmeter Terminal 1 +24 VDC Terminal 2 24 V return (0 V) Terminal 3 4-20 ma output Terminal 4 Terminal 5 RS485 RS485 Page 11 of 11