Perfusion Safety Review in Japan

Similar documents
Recommendations for Standards of Monitoring and Alarms During Cardiopulmonary Bypass

Recommendations for Standards. Monitoring. Cardiopulmonary Bypass

Clinical Engineering. Equipment (Heart-lung machines)

S5 TM. Min.I. Mini Maxi Me. The proven safe and reliable, world-leading perfusion system, now optimized for minimally invasive and pediatric surgery

Min.I. Mini Maxi Me. The proven safe and reliable, world-leading perfusion system, now optimized for minimally invasive and pediatric surgery

Point-of-Care Testing: A Cardiovascular Perfusionist s Perspective

Recommendations. For Standards of. Monitoring and Safety during. Cardiopulmonary Bypass

1.060 Set-Up and Priming of the Bypass Circuit. Perfusion Technology Department. Perfusionists

Adult All Integrated Mini Bypass System

Prescriptive Oxygenation

Equipment and monitoring for cardiopulmonary bypass

Patients come in a range of sizes. Now your oxygenator does, too.

CAPIOX FX Advance. Oxygenator with Integrated Arterial Filter and Hardshell Reservoir

PERFUSION SYSTEMS MECC SMART SMART SUCTION SYSTEM

INSPIRE Min.I. Redefining minimally invasive approach to perfusion. Expand your choices with INSPIRE Min.I.

MATERIALS AND METHODS

Optimized Perfusion Management and Patient Outcome

Heart/Lung Perfusion Packs INSTRUCTIONS FOR USE

(12) United States Patent (10) Patent No.: US 6,524,267 B1

INSPIRE TM Inspired choice

Considerations in Circuit Miniaturization

American Society of ExtraCorporeal Technology Standards and Guidelines For Perfusion Practice

CAPIOX SX Family of Oxygenators

I. Subject: Medical Gas Systems Utility Failure Interim Procedure III. Purpose:

CAPIOX SX Family of Oxygenators

CAPIOX SX Family of Oxygenators

Vacuum-Assisted Venous Drainage: A 2014 Safety Survey

Limitations Using the Vacuum-Assist Venous Drainage Technique During Cardiopulmonary Bypass Procedures

A complete family of Pediatric Perfusion Systems

An In-Vitro Study Comparing the GME Handling of Two Contemporary Oxygenators

CDI Blood Parameter Monitoring System 500 A New Tool for the Clinical Perfusionist

INSPIRE Oxygenator Family

(12) United States Patent (10) Patent No.: US 6,918,887 B1

Preventing Massive Gas Embolism During Cardiopulmonary Bypass: A Review of Safety Devices

Equipment and supplies - pediatric intensive care service.

MANITOBA RENAL PROGRAM

An In Vitro Analysis of a One-Way Arterial Check Valve

TSE Blood Pressure Monitor Invasive series

INSPIRE TM C The integrated closed system oxygenator for gentle perfusion

Emboli Production in Hardshell Venous Reservoirs at Low Reservoir Levels Devin Eilers, BS, Willie Glaze, BS, & David Holt, MA, CCT Clinical Perfusion

CRRT with Prismaflex LEADS TO More Flexibility, Ease of Use and Safety

Blood Parameter Monitoring System 550

Using the Lifebox oximeter in the neonatal unit. Tutorial 1 the basics

Mechanical Ventilation

INSPIRE 6 PH.I.S.I.O.

ENT 318/3 Artificial Organ. Artificial Lung. Lecturer Ahmad Nasrul bin Norali

New. The high security connection

Errors in Flow and Pressure Related to the Arterial Filter Purge Line

Gas Cabinets and Valve Manifold Boxes: Applying CGA G13- Rev 2015 to Gas Delivery Equipment

McHENRY WESTERN LAKE COUNTY EMS SYSTEM FALL 2014 CONTINUING EDUCATION MANDATORY FOR ALL PRIMARY AND PROBATIONARY ALS SYSTEM PROVIDERS.

Certified Hyperbaric Specialist Application

NEXUS TKO. Specialty Neonatal/Pediatric. Safety Infusion Sets. Blood Reflux Protection 24/7! Anti-Reflux Technology Provides

Chest Drainage Systems for Cardiac & Thoracic Surgery. Including portable suction with Digital Air Leak Monitoring

Alaris Pump module FAQs

Survey: Retrospective Survey of Monitoring/Safety Devices and Incidents of Cardiopulmonary Bypass for Cardiac Surgery in France

Contacts. Quick Start Guide

INSPIRE TM8 PH.I.S.I.O.

Lifespan Improvements Available in the Industry

Periodic Survey of Fuel Installations on Ships other than Liquefied Gas Carriers utilizing gas or other low flash point fuels

Waste Management System.

EU-Medical Pilots and Cabin Crew

The next generation of Blood and Fluid Warming

INTERNATIONAL MARKETING. 61, avenue Tony Garnier Lyon France Tel.: 33 (0) Fax: 33 (0)

Operating Instructions for Microprocessor Controlled Ventilators

OxygenatOrs QUaDrOx-i adult QUaDrOx-i small adult

Clinical Gaseous Microemboli Assessment of an Oxygenator with Integral Arterial Filter in the Pediatric Population

A Compact Single-Unit Oxygenator,

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR PRESSURE GAUGE. This publication replaces T.O. 33K dated 30 January 2012.

The next generation of Blood and Fluid Warming

Medical Instruments in the Developing World

Medical gases. Safe, reliable supply of high-quality gases and services for your healthcare facility tell me more

Patent Foramen Ovale and Fitness

Transforming Preservation

RULES PUBLICATION NO. 119/P

A portable tool to detect leaks in large diameter pipes AGBAR

Brevard County (Florida) Administrative Procedures

AutoPulse. Mechanical CPR. in the Hospital

Clinical Skills. Administering Oxygen

The safe use of infusion devices

Corporate Overview and Product Summary

Arterial Punctures Blood Gas Collections (ABG)

Periodical surveys of cargo installations on ships carrying liquefied gases in bulk

An In Vitro Study of the Effectiveness of Carbon Dioxide Flushing of Arterial Line Filters

TYPICAL TOXIC GAS SYSTEM INSPECTION VIOLATIONS

Welcome to the Specialized Medical Services Respiratory training webinar series!

Improve Reliability of Turbomachinery Lubrication and Sealing Systems

Indications for Use: Caution: Note:

UNIQUE CHARACTERISTICS OF THE PULMONARY CIRCULATION THE PULMONARY CIRCULATION MUST, AT ALL TIMES, ACCEPT THE ENTIRE CARDIAC OUTPUT

Bulletin TCR-104 & 109 Filling and adding to the Glycol pressure system

Oxygen Therapy. Respiratory Rehabilitation Program. What is oxygen?

Oxygen Saturation Monitors & Pulse Oximetry

GASES FIRE & RESCUE SPECIALTY. Gases & Equipment. Medical Gases Liquid & High Pressure Oxygen U.S.P. Nitrogen N.F. Carbon Dioxide U.S.P.

Thank You for Attending Today s Webinar. Today s Featured Speaker

INTRODUCTION TO HYPERBARIC MEDICINE

American Society of ExtraCorporeal Technology. Standards and Guidelines. for Perfusion Practice

WARNING: All medications should be closely monitored to detect any alterations in the effective concentration due to the ultrafiltration process.

Redesign - ATMOS S 201 Thorax. Digital thoracic drainage systems. ATMOS Thorax MOBILITY SAFETY COMFORT FLEXIBILITY

Should Air Bubble Detectors Be Used to Quantify Microbubble Activity during Cardiopulmonary Bypass?

healthy volunteers were included in the study.

Effect of Hypothermia on Arterial Versus Venous Blood Gases During Cardiopulmonary Bypass in Man

Transcription:

Perfusion Safety Review in Japan Atsushi SEKIGUCHI Department of Medical Engineering, Saitama International Medical Center, Saitama, JAPAN and Secretary General, JaSECT; The Japanese Society of Extra-corporeal Technology in Medicine AmSECT Best Practices in Perfusion 2008 October 16, 2008

Thoracic and cardiovascular surgery in Japan Population: 127 million Number of open heart center: 544(97.1%) Total number of OH Surgery: 53,741 cases

Number of CPB in Japan (2006) MAZE, tumor, VAD, DCM, constrictive pericarditis, Htx, others Congenital 7,386 (17%) Others 4,343 (10%) Vascular 9,376 (21%) CPB 44,231 IHD 7,835 (18%) Valve 15,092 (34%) Conv.- CABG 7,124 (16%) others 910(2%) C-CABG CABG 39% CABG 17,941 OPCAB 61% 11,021

Case load distribution of CPB Average : 71 cases / hospital / year # of Centers # of Cases 200 12,000 180 160 140 120 100 80 60 40 20 0 1~49 50~99 100~149 150~199 200~249 250~299 300~349 350~399 400~449 450~499 500~549 550~599 600~649 650~699 700~749 750~800 CPB Case load range 10,000 8,000 6,000 4,000 2,000 - in 2004 Number of operations performed Number of Centers percentage 1-24 68 12.5% 25-49 104 19.1% 50-99 164 30.1% 100-149 90 16.5% 150-199 39 7.2% >200 79 14.5% in 2006 544

Who is allowed to run CPB? *Certificate system in preparation Pacemaker technician* Bio-medical engineer VAD technician* others Certificates by respective Societies Respiratory therapist Hemodialysis technician Clinical Perfusionist Nat l License grant by the government Clinical Engineer (Legal ground for allowing clinical related practice)

Who is allowed to run CPB? A CLINICAL ENGINEER: Governmental Licensee A Clinical Engineer is the one who officially allowed to perform CPB as well as Hemodialysis, Respiratory therapy and Bio-medical engineering services. 21,500 persons licensed. More than 85% dedicated for HD Approx. 7% dedicated for CPB Requirements 3 to 4 years education after High school education. Pass national examination for permanent license. Limited education and training provided for CPB related subjects

Who is allowed to run CPB? A Certified Clinical Perfusionist Certification grant from the joint board of three physicians societies; JSAO *, JATS *, JSCVS *. Requirement: Licensee of Clinical Engineer Membership of JaSECT and JSAO * More than 3 years clinical experience with >30 CPB cases Complete the 3 year education program (total 6 days) as a continued education seminar provided by JaSECT and JSAO * Pass an examination including individual interview Currently 602 persons certified (1987-2007) Re-certification required every 5 year Minimum 30 CPB cases experience within the period JSAO: The Japanese Society for Artificial Organs JATS; The Japanese Association for Thoracic Surgery JSCVS; The Japanese Society for Cardiovascular Surgery

Perfusion Safety Standard Measures Established by JaSECT April 2007 Level 1 : Minimum requirements <6 items> (Crucial for safety and less barrier for implementation) Level 2 : Strong recommendations <9 items> (Crucial, but some investment required. Expect to be in level 1 in near future) Level 3 : Moderate recommendations <9 items> (Preferred, alternative methods can be acceptable) Based upon consensus at the Council of Professional Perfusionists board of JaSECT

Compliance of the standard measures Survey data as of March 2008 Date collected from 305/454 hosp. (67.3%) Overall Implementation rate (waited average) Level 1 (6 items) 76.5% Level 2 (9 items) 55.5% Level 3 (9 items) 58.0%

Level 1 (minimum requirements) % : Implementation rate Cardioplegia Vent Suction Venous return Arterial (1) Reservoir level sensor 82% 99% (6) Pressure monitor at cardioplegia delivery line Reservoir (2)Continuous pressure monitor prior to Ox. 23% Oxy (3)Pressure monitor line prior to arterial filter 67% Cardioplegia / vent pump Suction pump Arterial pump 3.6 (5) Arterial line filter or bubble trap 99% (4) Flow monitor for centrifugal pump 99%

Level 2 (strong recommendation) % : Implementation rate Cardioplegia Vent Suction Venous return Arterial (8) Bubble detector 38% 225 43 (1) SvO 2 monitor 85% Reservoir (2) Pump auto-regulation system link to reservoir low level detector 35% 138 (9) High pressure alarm at cardioplegia delivery line 80% Oxy (5)High pressure alarm at arterial line 80% (7) Arterial filter (Not bubble trap) (3) bubble detector 79% 31% Cardioplegia / vent pump Suction pump Arterial pump 286 (6) Pump shutdown system link to the high pressure alarm for roller pump (4) Pump shutdown system link to the bubble detect 55% 17%

Level 3 (moderate recommendation) % : Implementation rate Cardioplegia (7) One way valve at vent line 71% Vent Suction Venous 225 43 (1) Arterial blood gas monitor 56% Arterial (4) Pressure monitor line prior to arterial cannula 56% 286 Reservoir (8) One way valve at filter purge line (3) Pressure monitor prior to arterial filter 56% 47% Oxy (9) Pump shutdown system link to high pressure alarm at CP delivery line 90% Cardioplegia / vent pump Suction pump Arterial pump 286 1.6 (5) Low flow alarm for centrifugal pump 60% (2) Pump flow regulation system link to high pressure alarm for centrifugal pump (6) Reverse flow prevention system for centrifugal pump 15% 71%

Conclusion JaSECT has introduced Perfusion Safety Standard Measures in 2007 for Better Practice. The survey conducted one year after the introduction showed better implementation result than the past, however, it looks more time required to establish firmly among the all society members. Faster implementation is crucial especially for pre-oxy arterial line pressure monitor, alarm-linked pump shutdown system and bubble detector at the arterial line. We will continue to educate our members for more higher implementation rate.