How to Design Medical Accelerator Systems for Reliability: IBA PT System

Similar documents
Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription

Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization

Safety Manual VEGAVIB series 60

Safety Manual. Process pressure transmitter IPT-1* 4 20 ma/hart. Process pressure transmitter IPT-1*

Safety Manual VEGAVIB series 60

Electronic Oncology Systems the Management and the Safety Measures. Hansen Chen, Director, TDSI

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.

Review of fundamental photon dosimetry quantities

Safety Manual OPTISWITCH series relay (DPDT)

Safety-critical systems: Basic definitions

Ch.5 Reliability System Modeling.

Zoubir Ouhib Lynn Cancer Institute

SIL Safety Manual. ULTRAMAT 6 Gas Analyzer for the Determination of IR-Absorbing Gases. Supplement to instruction manual ULTRAMAT 6 and OXYMAT 6

NOVEL Hz ACCELERATORS FOR PROTON AND CARBON THERAPY: THE CYCLINACS

Accelerometer mod. TA18-S. SIL Safety Report

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

Proton Therapy QA Tools

Monitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting

An Overview of Types of Isolation Valves

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface

Transactions on the Built Environment vol 7, 1994 WIT Press, ISSN

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009

Relative Dosimetry. Photons

GW DRY PIPE VALVE INSTALLATION & OPERATION 100mm (4 ) 150mm (6 )

YT-3300 / 3301 / 3302 / 3303 / 3350 / 3400 /

THE development of more advanced techniques in radiotherapy,

3. Real-time operation and review of complex circuits, allowing the weighing of alternative design actions.

UNIVERSITY OF EDINBURGH GENERIC RISK ASSESSMENT - RADIATION RP/RA8 - WORK WITH X-RAY FLUORESCENCE ANALYSERS

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers

Adaptability and Fault Tolerance

Bespoke Hydraulic Manifold Assembly

Eutectic Plug Valve. SIL Safety Manual. SIL SM.015 Rev 0. Compiled By : G. Elliott, Date: 19/10/2016. Innovative and Reliable Valve & Pump Solutions

Analysis of Instrumentation Failure Data

SnapShot IMRT with compensators and FFF beams

Reliability Engineering. Module 3. Proactive Techniques - Definitions

Effect of noise in the performance of the transducers in an ultrasonic flow meter of natural gas

PROCEDURE. April 20, TOP dated 11/1/88

Transmitter mod. TR-A/V. SIL Safety Report

Safety Critical Systems

VariSource High Dose Rate Afterloader Procedures. For Performing Breast Brachytherapy with. The SAVI TM Applicator

Math Review. Overview

Variable Depth Bragg Peak Test Method

CORK INSTITUTE OF TECHNOLOGY INSTITIÚID TEICNEOLAÍOCHTA CHORCAÍ. Semester 2 Examinations 2010

OptonPro. High-power laser up to 7000 mw. Laser therapy natural healing with the power of light

Abstract: Introduction:

Distinguishing Features

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas

Systems Theoretic Process Analysis (STPA)

Pneumatic QEV. SIL Safety Manual SIL SM Compiled By : G. Elliott, Date: 8/19/2015. Innovative and Reliable Valve & Pump Solutions

ANZAI AZ-733VI. Respiratory Gating System. since 1976

FP15 Interface Valve. SIL Safety Manual. SIL SM.018 Rev 1. Compiled By : G. Elliott, Date: 30/10/2017. Innovative and Reliable Valve & Pump Solutions

YT-300 / 305 / 310 / 315 / 320 / 325 Series

General Isolation Procedures

SPIROVENT VACUUM DEGASSERS

GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY

Medical Instruments in the Developing World

Special Documentation Proline Promass 80, 83

Customer Responsibilities

Free Surface Flow Simulation with ACUSIM in the Water Industry

Solenoid Valves For Gas Service FP02G & FP05G

Failure Mode and Effect Analysis (FMEA) for a DMLC Tracking System

University of Aberdeen

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry

KOMPRESSOREN COMPRESSORS KAESER AIR SERVICE Our Services An Overview

Combining Speed of Delivery and Quality in Complex Systems. Manuel Pais DevOps & Delivery manuelpais.net

High Speed Direct SAD Radiosurgery Beam Scanner

Good Morning everyone, I will be talking this morning about the review and restructure of Cruise Whitsundays Safety Management Systems.

Electrical Equipment Failures Cause & Liability. Prepared by: Robert Abend, PE on 11 August 2014

Reliability of Safety-Critical Systems Chapter 3. Failures and Failure Analysis

FLUKA Simulations of Radiation Field in Second Accelerator Tunnel

Solenoid Valves used in Safety Instrumented Systems

The latest automotive systems require innovative leak test methods and fixturing.

Yokogawa Systems and PCI Training

Anaesthetic infrastructure and supplies

Reliability predictions in product development. Proof Engineering Co

Radiation Protection Service RADIOISOTOPES : SPECIFIC PROCEDURES AND HAZARD ASSESSMENT

Customer Responsibilities

CT433 - Machine Safety

Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08.

ADDENDUM IMPORTANT DOCUMENT INVITATION TO BID ADDENDUM. OPENING DATE & TIME: January 3, 2018 at 3:00PM

VACUUM REQUIREMENTS FOR LINACS P. Grand Yale University

Understanding the How, Why, and What of a Safety Integrity Level (SIL)

LOCKOUT STATION FOR TRAINING PURPOSES

Surgical Care at the District Hospital. EMERGENCY & ESSENTIAL SURGICAL CARE

New AQF Filter Polymer Filtration System

Hydraulic (Subsea) Shuttle Valves

Overspeed governors (113 OG Моment 250 MR, 114 OG Moment 250 A3, 115 OG Moment 250 MRL, 119 OG Moment 250 MRL/1)

1 in 4 patients will contract an HAI during their hospital stay. were prevented, annual cost savings would be $11.3 billion.

Radiation Safety Program Manual Part 3 Cyclotron - Isotope Production

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

Customer Responsibilities

CONTROL SOLUTIONS DESIGNED TO FIT RIGHT IN. Energy Control Technologies

Questions & Answers About the Operate within Operate within IROLs Standard

Reliability of Safety-Critical Systems Chapter 4. Testing and Maintenance

Operating instructions Electrical switching facility pco

DATA ITEM DESCRIPTION Title: Failure Modes, Effects, and Criticality Analysis Report

Vacuum Simulations of the KATRIN Experiment

Transcription:

How to Design Medical Accelerator Systems for Reliability: IBA PT System Yves Jongen Founder & Chief Research Officer Ion Beam Applications s.a. Belgium 1

Outline A short introduction of a proton therapy facility Why such a high pressure on availability Forecasting the availability Data on MTBF and MTTR Two definitions of availability in radiotherapy systems Designing for availability 2

Why Protons? Relative dose deposited in the tissue 100 Bragg Peak 80 60 Protons 230 MeV 40 20 0 Electrons 20 MeV 0 5 10 15 20 25 30 32 cm 0 = Skin level Depth of the tissue X rays 8 MV 3

Treatment room in PT 4

Proton Therapy Center 5

IBA Proton Therapy System A Proton therapy system is much more than an accelerator It is a complex, multi-room system, filling a Hospital building. The total investment is around 100 M, of which 45 M for the equipment More than 120 people (doctors, therapist, physicists, nurses) work daily in a PT facility A PT facility can treat 3500 patients/year, generating revenues in excess of 60 M /year, or 250,000 /day, or 20,000 /hour! The availability is required as much for financial as for medical reasons 6

Equipment in IBA PT system 230 MeV Proton Accelerator (Isochronous Cyclotron + Energy Selection System) Beam Transport and Switching System Isocentric Gantries (typically 3) and one Fixed Beam Line Nozzles for Scattering and Wobbling (scanning compatibles) Robotic Patient Positioners Control System and Safety System 7

The 230 MeV Cyclotron at MGH 8

The Beam Transport System 9

Isocentric Gantry 10

Beam current regulation Oscillogram of the proton beam current (pink) following the required time function (yellow) 2 1.5 1 signal (V) 0.5 0-0.05-0.04-0.03-0.02-0.01 0 0.01 0.02 0.03 0.04 0.05-0.5-1 t (sec) 11

Calculating the availability Let s concentrate on unscheduled down-time For a single component A = 1- MTTR / MTBF MTBF: mean time between failures MTTR: Mean time to repair Time to identify the problem Time to wait for radiation decay Time to bring the spare part in position Time to repair Time to restart, revalidate 12

Calculating the availability (2) For a chain of components/subsystems in serie, A = product (1 - MTTR(i) / MTBF(i)) The calculation has to take into account the parallel or serie nature of subsystems or components Direct algebric calculation, or Monte Carlo methods (little difference!) The difficulty of the method is not the data processing, it is the collection of the data! (GIGO principle!) The biggest problem is to find the MTBF of the components/ subsystems. Operational experience data of similar accelerators is most useful. Military/Governement reliability databases are useful, but errors are easy and frequent (MTBF of RF amplifier = 500,000 hours) 13

Calculating the availability (3) MTTR is more accessible to prospective calculation based on experience and simulations In the calculation of the MTTR, the composition and location of the spare parts stock is essential Software issues are specially difficult to assess. 14

Two definitions of availability In radiation therapy, not all problems have an equal importance: A problem delaying the delivery of treatments by a couple of hours is annoying, and is a waste of manpower A problem preventing to deliver a treatment fraction on the specified date compromizes the outcome of the treatment and the revenues of the facility As a result, two definitions of reliability have emerged R(a) = 1 (treatment hours lost/ hours scheduled) R(b) = fractions delivered/ fractions scheduled A problem causing a delay of 2 hours 5 times, is more acceptable than a problem causing one loss of 10 hours This is why supraconductivity was eventually discarded for the IBA PT cyclotron 15

Designing for reliability A = Product (i) (1-MTTR(i)/ MTBF(i)) Three methods to improve the availability: 1. Work on simplicity, parallelism 2. Increase MTBF 3. Decrease MTTR Take any incident as an opportunity to improve reliability 16

Simplicity Simplicity of design is one key aspect for reliability (fixed energy cyclotron versus synchrotron) Whenever possible, replace serial chains of components / subsystems by parallel assemblies Robustness and fault tolerance must be at the base of the design Designing fault-tolerant accelerators is still at the level of concepts 17

Increasing MTBF Using high quality, proven components Use conservative designs Well known subcontractors (long term partnership spirit) Well specified, qualified production technologies and methods Components used well below their maximum ratings Implementing and enforcing preventive maintenance Use every problem as an opportunity for improvement: never just fix a problem: go to the root, analyse and take the steps to improve the design or the production quality 18

Decreasing MTTR Fast diagnosis by built-in computer diagnosis, trained personel, written troubleshooting procedures Fast access to repair thanks to low components activation (choice of materials is key) or remote handling Composition, location and accessibility of the stock of spare parts Design of subsystems for repairability (accessibility) Fast exchange of components (pumps, valves, air locks) Design of power supplies (in plane vs. 3D) Choice of vacuum and vacuum technology for fast pump-down Special tools Design for access 19

Cyclotron opens at median plane for easy access 20

21