Is Your Oxygenator Failing? Diagnosis and Suggested Treatment

Similar documents
1.060 Set-Up and Priming of the Bypass Circuit. Perfusion Technology Department. Perfusionists

Heart/Lung Perfusion Packs INSTRUCTIONS FOR USE

Alfred ECMO Circuit Priming and Storage Guide

CDI Blood Parameter Monitoring System 500 A New Tool for the Clinical Perfusionist

Point-of-Care Testing: A Cardiovascular Perfusionist s Perspective

Blood Parameter Monitoring System 550

Recommendations for Standards. Monitoring. Cardiopulmonary Bypass

AQUADEX FLEXFLOW SYSTEM QUICK REFERENCE GUIDE

A complete family of Pediatric Perfusion Systems

TECHNICAL BULLETIN. TriControls. Control Materials for Liquid QC and Calibration Verification

INSPIRE TM Inspired choice

AQUARIUS Continuous Renal Replacement Therapy with Regional Citrate Anticoagulation (RCA) Competency

Blood Pressure Monitoring: Arterial Line

Contacts. Quick Start Guide

EASYPUMP II THE FLEXIBLE SOLUTION FOR SHORT AND LONG TERM INFUSION THERAPIES

VETROSON. OXY-GEN SYSTEMS (Oxygen Generating Systems) Question & Answer Brochure

- Repeat calibration until desired results are reached otherwise set SW2 and SW1-8 open.

INSPIRE Min.I. Redefining minimally invasive approach to perfusion. Expand your choices with INSPIRE Min.I.

CDI System 500. Blood Parameter Monitoring System. Continuous blood parameter monitoring for improved blood gas management

AQUARIUS Continuous Renal Replacement Therapy Competency

Patients come in a range of sizes. Now your oxygenator does, too.

Optimized Perfusion Management and Patient Outcome

Quiz for Module 2: Aquarius Training (Rev. 1) Lesson 1: Introduction to Aquarius

Fail Operational Controls for an Independent Metering Valve

Making Life Easier Easypump II

I. Subject: Medical Gas Systems Utility Failure Interim Procedure III. Purpose:

Reverse Osmosis System Installation Guide and Operation Manual.

(12) United States Patent (10) Patent No.: US 6,524,267 B1

INSPIRE TM C The integrated closed system oxygenator for gentle perfusion

BioAerosol Nebulizing Generator. Operation and Maintenance User Manual

TECHNICAL INFORMATION

Troubleshooting Guide Aquarius

New Generation System M, leading the World in the Non-Invasive Measurement of Critical Real-Time Parameters.

Emboli Production in Hardshell Venous Reservoirs at Low Reservoir Levels Devin Eilers, BS, Willie Glaze, BS, & David Holt, MA, CCT Clinical Perfusion

Oxygen Dialflow Meter. Instructions for Use

IVIG Educational Presentation

SDS -Series 4C-D201/-D202 Supplemental Drying System User s Manual

QUICK REFERENCE GUIDE

Installation Instructions WARNING: WARNING: WARNING: LP AND HIGH ALTITUDE LP GAS CONVERSION KIT FOR UNITED STATES INSTALLATIONS

WARNING: All medications should be closely monitored to detect any alterations in the effective concentration due to the ultrafiltration process.

3100A Competency Exam

CSA Sample Draw Aspirator Adapter Operator s Manual

Summary of Cardiac Anesthesia Technician Responsibilities

Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting

Indications for Use: Caution: Note:

product manual HM-4140, HM-4150, HM-4160 HM-4160A HM-4150 Humboldt FlexPanels

BlenderBuddy 2. Operating Manual & Instructions For Use ENGLISH R233M01 REV. A

QUADROX-i Neonatal & Pediatric Maximum safety for the smallest patients.

InFlow Modular perfusion bioreactor

CORESTA RECOMMENDED METHOD N 6

Corporate Overview and Product Summary

HEALTH SERVICES CODE A.6 CATEGORY: RN - SNP

Flow meter. bellow vaporizer. APL valve. Scavenging system

HYDROVEX TTT Membrane Flow Regulator

CAPIOX SX Family of Oxygenators

Citation for published version (APA): Somer, F. M. J. J. D. (2003). Strategies for optimisation of paediatric cardiopulmonary bypass s.n.

Specifications...Easy CLIA Classification: Sample Type: Blood gas analyzer operation has never been simpler Sample Size: Measured Parameters PCO2

CAPIOX SX Family of Oxygenators

CAPIOX SX Family of Oxygenators

Recommendations. For Standards of. Monitoring and Safety during. Cardiopulmonary Bypass

User Manual. Model 3930 Vacuum Assisted Venous Drainage Controller w/ WAGD Suction Flowmeter

Protocol: i-stat Cardiac Troponin I Performance Verification

Clinical Engineering. Equipment (Heart-lung machines)

Bunnell LifePulse HFV Quick Reference Guide # Bunnell Incorporated

Considerations in Circuit Miniaturization

Digital Vacuum Regulator

DRAFT U S E R M A N U A L CAUTION. Model: 19MFA1001 Series. Federal (USA) law restricts this device to sale by or on the order of a physician.

ACRYLIC FLOW METERS 0-30, 0-70 & LPM

SPIROVENT VACUUM DEGASSERS

INSPIRE Oxygenator Family

An In-Vitro Study Comparing the GME Handling of Two Contemporary Oxygenators

BPS-2, Rev TABLE OF CONTENTS

UsER manual for Watersens ph -REDOX

1.0 PURPOSE 2.0 SCOPE 3.0 DEFINITIONS. ANSI: American National Standards Institute. CCC: Chemical Control Centre

QUICK START GUIDE TO DRINKING WATER MONITORING

AUTOVENT 4000 VENTILATOR

Instructions for Use

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

Cover Page for Lab Report Group Portion. Head Losses in Pipes

User Manual. Model 3930 Vacuum Assisted Venous Drainage Controller w/ WAGD Suction Flowmeter

1020 Industrial Drive, Orlinda, TN fax

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:

Recommendations for Standards of Monitoring and Alarms During Cardiopulmonary Bypass

with O 2 Controller Instruction Manual

Prescriptive Oxygenation

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

Unit 15 Manual Resuscitators

The Proposed Use of Performance Indices to Evaluate and Compare the "Gish Vision" Membrane Oxygenator

OPERATING MANUAL LO-CROSS-FLOW CF M

CAPIOX FX Advance. Oxygenator with Integrated Arterial Filter and Hardshell Reservoir

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Surgical Care at the District Hospital. EMERGENCY & ESSENTIAL SURGICAL CARE

Operators Manual. Supera Floor Stand Anesthesia Machine M1000. Models: M1000 Basic Non-rebreathing M1200 Basic M1200

TESCOM 50-4X Series Safety, Installation & Start-Up Procedures

Operating Guidelines for PermSelect Modules Liquid Contacting

Manual: Biphasic Positive Airway Pressure (BiPAP) Ventilation

Limb Perfusion Device

TSE Blood Pressure Monitor Invasive series

Injection Systems INSTALLATION AND OPERATING GUIDE HI FLO VERTICAL TANK SYSTEMS


Transcription:

Is Your Oxygenator Failing? Diagnosis and Suggested Treatment Michael Higgins, Charles M. Tyndal, Jr., Richard G. Berryessa, and Dan R. Hydrick The University of Colorado Health Sciences Center Denver, CO Division of Cardiothoracic Surgery, Department of Surgery and PSICOR, Inc. San Diego, CA Keywords: complication, oxygenator; equipment, safety; gas transfer, determination; oxygenator, performance; safety, equipment; safety, technique; temperature, variation Abstract (J. Extra-Corpor. Techno/. 19[3] p. 33-337 Fall1987, 3 ref.) Oxygenator failure during open heart surgery can be a serious complication but one with which perfusionists should be prepared to deal. We hypothesize that oxygenators are often replaced unnecessarily and when they are performing within the manufacturers' specifications. This suspicion was confirmed by surveying several oxygenator manufacturers. Discriminating between oxygenator and nonoxygenator problems is critical. In the face of the correct diagnosis of oxygenator failure a simple, safe, expedient, and familiar method of replacement is needed. By using the gas transfer equations one can determine if the oxygenator is transfering adequate gas, after other associated problems are considered and ruled out. After oxygenator failure has been diagnosed, a plan for replacement which has previously been established and practiced should be implemented. We studied 2 techniques for hollow fiber membrane oxygenation (HFMO) replacement and found the risk of micro-air, as a result of the changeout procedure, decreased when the new oxygenator was vacuumed. The purpose of this paper is to help identify when/ if an oxygenator is failing and suggest a planned replacement drill to help minimize confusion, delay, Direct communications to: Michael Higgins, A.A., C.P., PSICOR, Inc., 16818 Via del Campo Ct., San Diego, CA 92127 and possible complications. We will also present a technique for decreasing the risk of micro-air which may result from a quick oxygenator changeout. Introduction The diagnosis and response to oxygenator failure should be planned and well defined. Before changing an oxygenator one must be certain that the unit is at fault and there is not another mechanical or patientrelated problem. This may be accomplished by the use of a checklist and oxygenator replacement drills carried out by the entire perfusion team at regular intervals. The result will be a team of perfusionists who can change an oxygenator rapidly and safely. After surveying 5 major membrane oxygenator manufacturers, we agree that oxygenators are often changed for low Pa 2 s unnecessarily. 1 When manufacturers evaluate returned "failed" oxygenators they often find the problems are not oxygenation but improper anticoagulation, improper Fraction of Inspired Oxygen (Fi 2 ), or small blood to gas leaks. There are many mechanical (equipment) considerations that need to be evaluated prior to the membrane evaluation. Some of these are: 1) Is oxygen being delivered to the membrane? 2) Is the gas path obstructed? 3) Is the gas connected to the correct gas port? 4) Are Fi 2 and gas flow appropriate? All of these equipment-related problems can cause inadequate oxygenation. Certain patient considerations should also be evaluated. Some of these are: 1) Is the hematocrit adequate? 2) What is the temperature? 3) Is blood flow adequate? 33

4) Are anesthesia and muscle relaxant levels adequate? Once these external factors have been evaluated, the gas transfer of the oxygenator should be calculated with use of the 2 Transfer Equation. 2 (Arterial 2 content-venous 2 content) x (1) x (Flow-LPM) Art. Content = [(Art. Sat. %) x (1.34 ml 2) x (Hgb gm%)] + [(Pa2) x (.3)] Ven. Content= [(Ven. Sat.%) (1.34 ml2) (Hgb gm%)] + [(Pv2) (.3)] Maximum Value = 35 cc@ 6 LPM gas flow 12 gm% hemoglobin normothermia sea level Once inadequate gas transfer is diagnosed, the perfusionist should have a replacement strategy that is quick and safe. Our suggestion is to have a rehearsed oxygenator changeout drill. This drill will help identify the safest and most expedient method to deal with oxygenator changeout. If used regularly, this drill will develop and maintain the perfusionist' s proficiency with oxygenator changeout. Our group tested 2 techniques (unprimed and preprimed) for HFMO changeout. Additionally, we measured the effect vacuum had on microbubble activity. The results of this observation suggest that oxygenators are often unnecessarily replaced. We propose a method of streamlining the diagnosis of oxygenator failure and describe a method of changing an HFMO which is quick and has minimal microbubble generation. Materials and Methods The Cobe-St6ckert Heart Lung Machine, J & J Maxima Membrane Oxygenatorb, Terumo 1cc Reservoir Bagc, and Bentley %-inch tubingd with connectors were used for all experiments. The circuit was primed with 2cc Plasmalyte N. Method 1 The following method provides a fully primed oxygenator for changeout. a Cobe Laboratories, Lakewood, CO 8215 b Johnson & Johnson Cardiovascular, King of Prussia, PA 1946 c Terumo Corporation, Piscataway, NJ 8854 d American Bentley, Irvine, CA 92714 e American Travenol, Deerfield, IL 615 Additional Equipment SOOcc Plasmalyte A with pressure bag One Blood Component Infusion Set (Pit. Infusion Set) (Figure 1) One 3 /s" connector Two :%" Leur connectors Two 3-way stopcocks Two 2" pieces :Ys" tubing One 2" piece '14'' tubing One 1 /4" connector Sterile Scissors Tubing Clamps Set-Up Connect to oxygenator blood inlet port two 2- inch pieces of :Ys-inch tubing with 3 /s-inch leur connector and stopcock in middle for priming (Figure 1). Connect to the oxygenator blood outlet port one 2-inch piece of :Ys-inch tubing with a %-inch leur connector and stopcock for debubbling. Connect the 2-inch piece of 1 /4-inch tubing with V4-inch connector to the recirculation port (Figure 1). Priming Place priming solution in pressure bag, clamp infusion line, apply 3 mmhg, and connect infu- 318 Leur Connector PRE-PRIMED OXYGENATOR stopcock ~-~==~~==~~ 2 3/B t~bing I OUTLET MAXIMA HOLLOW FIBER a- 1 I 4 connector!~-2 1/4 tub,ng -Recirculation Port BLOOD COMPONENT INFUSION SET -Spike for priming l}2 3/8 tubing 3/8 Leur Connector ~ To inlet stopcock with - stopcock -2 3/8 tubing -3/8 connector Figure 1: Set-up of membrane to be placed in circuit 331

sion set to meet inlet port stopcock (Figure 1). Position oxygenator vertically and clamp between inlet :YH-inch leur connector and :YH-inch straight connector and between V4-inch connector and recirculation port. Prime through threeway stopcock using the blood component set. Tilt oxygenator to debubble outlet port and clamp line. Turn threeway stopcock off and disconnect infusion set. Replacenzen t 1. Before stopping perfusion double clamp recirculation line; one clamp close to port and the other 2 inches away allowing enough space for reconnecting line to 1 /4-inch connector on new oxygenator. 2. Stop perfusion, turn oxygen off, clamp patient venous and arterial line. 3. Clamp oxygenator inlet blood line close to inlet port and again 2 inches away allowing for reconnection. 4. Clamp oxygenator outlet blood line close to outlet port and again 2 inches away. 5. Sterilize prep tubing and use sterile scissors to cut lines between clamps. 6. Place newly primed oxygenator in holder and connect all lines. 7. Recirculate through recirculation line at 2 LPM. 8. Apply unrestricted vacuum to the gas phase of the oxygenator (all other gas vents occluded). 9. While vacuuming, remove clamp between arterial line connectors and remove air using 6 cc syringe through blood inlet line stopcock. Reclamp arterial line to check for air. 1. Stop pump, clamp recirculation line, remove vacuum, open gas vents, and slowly resume perfusion inspecting for air. 11. Re-establish gas flow. Method 2 The following method may be used for an unprimed oxygenator, to be primed in line. 1. Add 5 ml extra prime to circuit. 2. Clamp recirculation line at V4-inch recirculation port and again 2 inches away allowing for connection. 3. Stop perfusion, clamp patient arterial and venous lines, double clamp oxygenator blood inlet and outlet lines allowing for reconnection. 4. Sterilize prep tubing and cut with sterile scissors between clamps allowing for reconnection. 5. Place new oxygenator in holder and connect all lines. 6. Remove oxygenator blood inlet and recirculation line double clamps and recirculate at 2Limin. 7. Vacuum oxygenator gas path (all gas vents occluded) 3 sec. while tilting and debubbling. Check oxygenator outlet for air under clamps. 8. Inspect entire system. 9. Stop pump, clamp recirculation line, remove vacuum, open gas vents, and resume bypass slowly inspecting for air. 1. Re-establish gas flow. We compared 4 methods of priming and changing an HFMO in the cardiopulmonary bypass circuit. We compared an unprimed oxygenator to a preprimed oxygenator, and compared both with and without vacuum (to decrease micro air). 3 Then we compared the time required for changeout and microbubble removal in all 4 groups. Results A 2-Way ANOV A (analysis of variance) was used to evaluate the results using Statworks TM statistical package for the Macintosh computer. This allowed the simultaneous evaluation of two variables. The changeout times are listed in Figure 2. It took significantly longer to changeout the vacuumed lungs than those not vacuumed (p<.3). This is probably a result of the learning effect due to the order of the groups tested and not of clinical significance. The vacuumed groups were tested first and the nonvacuumed last. Randomizing the order of the tests may have eliminated this difference. 7 6 5!Jl ~ 4 (.) ~ 3 2 1 CHANGEOUT TIMES 64.67 3.23 63. 7.32 Un-primed Figure 2. Membrane changeout times 58.33. 186 55.33. 4.46 Un-primed 332

The 1 micron bubble counts are listed in Figure 3. Use of vacuum had a profound effect on the 1 micron bubble counts (p<.1). Mean bubble counts in the pre-primed group without vacuum were 315.83 ± 9.35* versus 1. ±.89 in the same group with vacuum. The unprimed group also showed substantial decrease in bubble counts. The mean bubble counts for the unprimed lungs without vacuum were 527.33 ± 124.13, versus 14.83±4.45 in the vacuumed lungs. Prepriming of these lungs is significantly, but secondarily, better (p<.5). The 5 micron bubble counts are listed in Figure 4. As with the 1 micron bubbles it is better to vacuum (p<.3). Prepriming the oxygenator did not offer any advantage to debubbling (p<.261). Mean bubble counts in the preprimed without vacuum were 3.8 ± 1. 94 versus.67 ± 1.3 in the same group with vacuum. The mean bubble counts for the unprimed without vac- 1 MICRON BUBBLE COUNTS 6 5 c: 4 ~ '- "' Q) 3 :. :::J aj 2 1 1. '.89 14.83 ' 4.45 Un-primed Figure 3: Ten micron bubbles 9 315.83 ' 9.35 527.33 124.13 Un-primeil 5 MICRON BUBBLE COUNTS uum were 8.67±6.31 versus 2.33±1.75 in the same group with vacuum. There is little ischemic time difference between the preprimed or unprimed methods. The preprimed method offers the convenience of connecting and debubbling rather than connecting, priming, and debubbling in the arterial circuit. For the unpracticed perfusionist this method may also offer less ischemic time. The technique of vacuuming the HFMO for 3 seconds during 2 LPM of recirculation before reinitiating bypass offers the tremendous advantage of a bubblefree membrane. This is especially well demonstrated in the 1 micron range (Figure 3), but is also significant in the 5 micron range (Figure 4). This technique is applicable for all HFMOs, especially those which do not require gas path anatomy modification. If vacuum is not used, there is some advantage to prepriming the HFMO. Although microbubble counts were not significantly different in the 5 micron range (p<.261) in the non-vacuumed groups the 1 micron range exhibited significant reduction in microbubble counts when preprimed (p<.7). Discussion When a HFMO is returned for reported failure to oxygenate, the manufacturer attempts to recreate conditions similar to those that existed at the time of reported failure. If a complete perfusion record is returned with the oxygenator the manufacturer can more efficiently evaluate the complaint. If no records are returned the manufacturer has little to go on, and must impose standard AAMI conditions on the unit. The 1986 Manufacturer Quality Control Inadequate Oxygenation Evaluations for 5 major manufacturers' returned units is summarized in Figure 5. It is important to properly prepare the oxygenator for return to the factory. An improperly prepared oxy- 8.67 ' 6 31 1986 Membrane Failure Statistics Returned Failed %*.67 ' 13 I 2.33 ' 1.75 Un-primed Figure 4: Fifty micron bubbles 3.83 ' 194 Un-pnmed Manufacturer #1 7 1.3 Manufacturer #2 19 1.1 Manufacturer #3 4.5 Manufacturer #4 not given not given.4 Manufacturer #5 4. Figure 5 * All means reported as mean ± standard deviation of the mean (sd) *% of total oxygenators manufactured which failed The numbers are estimates, each manufacturer uses different methods for classifying complaints. 333

genator limits the manufacturer's ability to accurately test it. One manufacturer suggested a method for lung preparation. The lung should first be rinsed with saline until the effluent is clear. Then it should be flushed and recirculated with a 3% solution of H 2 2 Finally, it should be filled with 3% H 2 2, capped, and promptly returned. Testing of oxygenators which may have failed is quite expensive. It often requires an entire day of laboratory time, blood for the circuit, and most importantly, personnel to determine if the lung has in fact failed. Given the exceptionally low failure rate, one should be relatively certain that the lung has failed before returning it for testing. An oxygen analyzer connected to the gas circuit, distal to the vaporizer, may help assure that the proper concentration of oxygen is being delivered from the blender. This blender should be routinely checked for accuracy before each procedure use. No oxygen/air blender should be used without the concomitant use of an oxygen analyzer. A backup oxygen tank with regulator and proper connecting tubing should always be available in the event of loss of wall oxygen supply. Check to be sure that there is a hospital alarm that sounds if the oxygen supply is interrupted. The diagnosis and response to oxygenator failure should be planned and practiced. This may be accomplished through the use of a checklist or algorithm (Figure 6), and oxygenator replacement drills carried out by the perfusion team at regular intervals. The procedure in Figure 6 is provided with the intent of being a convenient flow chart for diagnosing oxygenator failure. It is a helpful tool and is not intended to be all inclusive. Anyone who suspects that an oxygenator is failing may use the algorithm, but should not limit their investigation to it. Any suggestions for revision are welcomed. This algorithm may be enclosed in a plastic cover and placed in a convenient location for referencing. The 2 Transfer formula may be printed on the back of the algorithm for easy reference. We have recommended an algorithm that we think will help in making the diagnosis of oxygenator failure. We have also recommended a technique for replacing the failed oxygenator with one which is preprimed. The replacement technique offers minimal microbubble counts when the lung gas path is vacuumed. On the basis of this study, we strongly recommend that whenever a HFMO is changed, that vacuum be applied to the gas phase to reduce the microbubbles generated by the changeout process. The potential for catastrophic accident during oxygenator changeout may be reduced if the perfusion team will choose an appropriate method for replacing the HFMO, and participate in regular practice drills before a failure occurs. Acknowledgment Special thanks to Anna Baron, Ph.D., for her work in data analysis. Additionally, we would like to thank Marc Voorhees and the Cobe Quality Engineering Department for their help with collecting the microbubble data. References 1. Personal Communication. 2. Reed, C. C. and Clark, O.K. Cardiopulmonary Perfusion, Houston: Texas Medical Press Inc., 1975. 3. Berryessa, R., Tyndal, D.M., and Tornabene, S.: A Technique for the Effective Removal of Air from a Hollow Fiber Membrane Oxygenator Circuit. f. Extra-Corpor. Techno/., 18:156-158, 1986. Questions from the Audience Question: Aaron Hill, Falls Church, VA: Very nice presentation. An observation and a question. In the second step with Fi 2, is Fi 2 maximum? There should be some reference to measuring oxygen content. According to the survey of Kurusz, referring back to accidents and problems, blender failures have been much more on the increase since the advent of the widespread use of oxygenators. So I think it's essential that you establish what's coming out of your blender. Now I think an oxygenator analyzer is essential to be able to do that. And I believe one of the blender manufacturers has that label right on the blender. The other question: during your priming changeout do you have C 2 in that or not? Is it just vacuum? Is it just fluid? Response: No C2. Question: In our experience, we use a different bubble counter, the Hadelon as opposed to the TM8. We were able to establish lower values by CO/ing the oxygenator as well-and decreasing subsequent bubble counts. Would you comment about the Fi 2 thing? Do you believe it's necessary to measure oxygen? Continued, p. 337 334

DIFFERENTIAL DIAGNOSIS OF OXYGENATOR FAILURE NO YES 2. Is gas connected & unobstructed? Is Fi2 maximum? NO Connect Gas Increase Fi2 4. Drawn A & V blood Do so now YES 5. If Pv2 or Sv2 low see 3 Increase flow > 2.4 L/min/m 2 Hematocrit eat 4-6 Calculated 2 Transfer? See formulae YES Do so now 7. Is 2 Transfer Adequate? (For flow, Hgb, altitude BSA, age, sex, temp) YES NO 8. May not need to worry re eat 4-6 CHANGE OUT OXYGENATOR Figure 6: Differential diagnosis algorithm 335

OXYGEN TRANSFER (ARTERIAL 2 CONTENT- VENOUS 2 CONTENT)(1)(FLOW-LPM) Figure 6: continued NORMAL VALUE= 35cc@ 6LPM BLOOD FLOW 12 Q /o HEMOGLOBIN 3]<>C SEA LEVEL 336

Questions, continued Response: Yes. This is a very good point. In the box number 2 there, [onscreen in the figure] it would also be helpful if you had your oxygen analyzer in line. There are also other factors to be considered in there. Is the 2 line obstructed? Is there a hospital alarm system that would go off if there were an oxygen failure in the operating room? So that's a very good point. Question: Dave Hoff, Flint, MI: Very nice presentation. In setting up the Maxima for the switchout, do you modify it or alter it in any way to be able to use an unrestricted vacuum source? And if so, do you recommend, after the oxygenator is in line, making any further alterations? Response: Are you referring to the preprimed or the unprimed? Question: To the oxygenator that is switched out. Because of the various safety vents that are devised in the production of the Maxima, how do you alter them? Response: With the Maxima we bone wax the large vent that goes around the side here. We found no problems. It has created no problem for us. There is also a vent port on the bottom side which we occlude with a small piece of tape. Then once our vacuuming process is completed, we remove the tape. And I found that this offers no deleterious effects for the membrane itself and the bubble-free environment has many advantageous effects. That's how we do it. Comment: Richard Berryessa, Denver, CO: It should be noted that we use the Maxima in this case, because it happened to be one of the few unsterile units that we have. You do not have to modify any of the other oxygenators on the market that I can think of. Most of them have but one gas exit source. If you were to use Maxima, for instance, and wanted to vacuum, it's probably not a good idea to alter the thing, such that you change it differently from the way the manufacturer made it. J&J made this to be foolproof and unfortunately they overlooked the potential for being able to vacuum an oxygenator. And you don't have to worry about that with anybody else. We have talked to them about our desire to be able to vacuum a lung, because there are occasions when you do get air in the circuit for one reason or another and they may or may not address it in the next version of their oxygenator. However, I wouldn't recommend it, and we do not clinically modify the oxygenator to be able to vacuum it. If you were able to do it in a temporary way, that would be reasonable. But you also have to be very careful that you did not leave it modified in the course of the case. You still have one open gas port which is the same as any other oxygenator. But the fact that you modify an oxygenator from the way the manufacturer made it does leave you open for some liability if someone could prove that that caused some problem with the patient. In answer to Aaron's question also, there are a number of things in the text of the paper that Michael did not have time to present in this presentation, including recommendations that blenders be routinely monitored with an oxygen analyzer and that the flow meters in the blenders be routinely calibrated, as well other recommendations. Comment: Frank Hurley, Chicago, IL: Just two comments. One on the blenders: in discussion with our biomedical department, we found that the Bird blenders seem to be especially susceptible to higher failure rates after two years of use. And Aaron's point about checking the output of those units is well taken. The second issue: it might be well for you in the future to look into the comparison of vacuum priming, which is your technique, versus the C 2 flush, which a number of us use. I'd be curious to see which is more effective, and the time it takes to be able to remove the bubbles in the system. Response: I guess I'll do a follow-up study on that when I do the effects of vacuuming. Comment: Richard Berryessa: If I may answer that: in some instances it is difficult and/or time-consuming to C 2 flush. One example is if you decide that you need to changeout an oxygenator, which happens very infrequently, that you really need to. If you take the time to C 2 flush, for instance, you'd probably recirculate, anyway, before you go on. If, however, you do not C 2 flush, and take the oxygenator out of the bag, and stick it in the circuit and recirculate for 3 seconds, with vacuum on the gas phase, we've shown that 3 seconds recirculation time, which you're going to do anyway, is enough to decrease the bubble counts significantly-to the range of a mean of 1. So we do C 2 flush all our clinical units. There are, however, other circumstances, which we reported last year. 337