TGA100A Trace Gas Analyzer Overview

Similar documents
TGA100 TRACE GAS ANALYZER USER AND REFERENCE MANUAL

TGA200 Trace Gas Analyzer Overview 9/11

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Laser Spectrometers for Online Moisture Measurement in Natural Gas. Ken Soleyn GE M&C

HumiSys HF High Flow RH Generator

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

Aurora. Moisture Measurement for Industrial Gases. GE Measurement & Control

Overview. Front Panel: Keypad and Display

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Thermo Scientific Model 146i Principle of Operation

Laser Gas Analyzer LG2000. The Laser Technologies

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

Columbus Instruments

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

HCL MONITORING. ICAC MARAMA WORKSHOP May 18-19th, Presented by: Ty Smith. CEMTEK

BUBBLER CONTROL SYSTEM

Instruction Manual. Biogas Analyzer InCa Bio 04

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

FTIR Users Group Meeting FTIR development

Features. Main Voltage: Vdc. Input Control Signal: 0-5 Vdc or 4-20 madc (Sourcing) Monitor Output Voltage: 0-5 Vdc or 4-20 madc (Sourcing)

TR Electronic Pressure Regulator. User s Manual

FTC130 Transmitter. Operating Manual

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

BUBBLER CONTROL SYSTEM

UltraLo 1800 Alpha Particle Counter

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

MEGAS 2.0. Gas analysis with direct C-Level Calculation and Bus-Connection. Data sheet

Pneumatic high-pressure controller Model CPC7000

EuroFID Total Hydrocarbon Analyzer. Precise Determination of Total Hydrocarbons in Air for Corrosive as well as Condensing Gases

Model GFC 7000T CO 2 Analyzer (Addendum to GFC 7000TA Manual, PN 07272)

FOURIER TRANSFORM INFRARED SPECTROSCOPY

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Pneumatic high-pressure controller Model CPC7000

Application Note 183. Page 1 of Aero Camino, Goleta, CA Tel (805) Fax (805)

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

ACV-10 Automatic Control Valve

Mass Spec will not Autotune

OMNISTAR / THERMOSTAR. The efficient solution for gas analysis. Intelligent software. Wide range of applications.

Mass Flow Controller (MFC) for Gases

X-Flow Mass Flow Controller 授权代理 : 北京品超思瑞科技学院公司

How to order guide. Spectronus FTIR Trace Gas Analyser. Version: 1.0

HUMOR 20. High-precision Humidity Calibrator HUMOR 20. Operation

DOS in Food Science & Nutrition, Manasagangotri, Mysore

Model EXAxt AV550G. Averaging Oxygen Analyzer

8.1 Calibration of gas sensors

Instruction and Maintenance Manual

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300

Agilent Auxiliary Gas Module AGM 2. User s Guide

GasSense NDIR User Manual

Real-time Analysis of Industrial Gases with Online Near- Infrared Spectroscopy

Installation and Operation Manual

Mass Flowmeters. Mass Flowmeters for Gases

HUMOR 20. High Accuracy Humidity Calibrator. Water T, p1 HUMOR 20 V1.0

The M-Series Eletta Flow Meter High accuracy DP Flow Meter with multiple functions

AFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features

The FLUX-METER: implementation of a portable integrated instrumentation for the measurement of CO 2 and CH 4 diffuse flux from landfill soil cover.

Quadruple mass spectrometers (transducer type)

Continuous gas analysis ULTRAMAT 23 Supplement 2 for Operating Instructions C79000-G52xx- C216-03

Series Environmental Chambers

EXPERIMENT XI. Careful!! Improper handling of the vacuum line may result in the release of SO 2 which is an irritating and suffocating gas.

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Mitos Fluika Pressure and Vacuum Pumps Datasheet

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

ACCURATE PRESSURE MEASUREMENT FOR STEAM TURBINE PERFORMANCE TESTING

MATRIX-MG Series. Innovation with Integrity. Automated High-Performance Gas Analyzers FT-IR

DEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111

Technical Data Sheet MF010-O-LC

Operations Manual D , Rev A May Rosemount CT5100 Continuous Gas Analyzer

PHOENIX L300i Best-in-class Helium Leak Detectors

Constant Pressure Inlet (CCN) Operator Manual

TEL/jlRE" Introduction. Display Features and Modes. Startup Procedure. Power-Up Procedure. Adjustment Modes

HUMOR 20 HUMOR 20. High-precision Humidity Calibrator. Water T, p1

Design Features. General Description

O2100C Oxygen Measurement Module Technical Use Notes do not use other wall adapters with the O2100Cmodule. 10% / V 5% / V 2% / V 1% / V 10% / V

21700 Nordhoff St. Chatsworth, CA Phone

Using Vaporized Hydrogen Peroxide

AMT-Ex Dewpoint Transmitter

Parametrable / Programmable Pressure Transmitter

MB3000-CH90 and MBGAS-3000CH FT-IR solutions for laboratory and process gas phase measurements R&D, QA/QC, process optimization and air monitoring

UT 3000 Mercury Ultratracer

FT-IR solutions for laboratory and process gas phase measurements R&D, QA/QC, process optimization and air monitoring

User s Booklet for the Wyatt minidawn Light Scattering Instrumentation

Mass Flow Controller (MFC) for Gases

Discovery HP-TGA 75/750. Site Preparation Guide

Level MEASUREMENT 1/2016

Model PDT Dewpoint Transmitter

AMS 6915 Board mount pressure sensor with digital output (I²C)

Programmable Submersible Level Transmitters

PRC CO ² -LASER PRESENTATION

analytical bulk flow distance FLOW level pressure temperature industrial communication

GCMSD-Headspace Analysis SOP

BAROMETER PRESSURE STANDARD PRESSURE CONTROLLER

Analogue and Digital Mass Flow Meters and Controllers for Gases MASS-STREAM

USING THE CS400/CS405 (KPSI SERIES 169/173) SUBMERSIBLE PRESSURE TRANSDUCER WITH CAMPBELL SCIENTIFIC DATALOGGERS INSTRUCTION MANUAL

Autocalibration Systems For In Situ Oxygen Analyzers

Series 6517 Katharometers. The ideal equipment for Process Monitoring and Control. Robust no moving parts. Designed for continuous industrial use

Sorensen DCS Series. 1 3 kw V A. General Purpose Systems Power Supply

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

Transcription:

TA100A Trace as Analyzer Overview The TA100A Trace as Analyzer measures trace gas concentration in an air sample using tunable diode laser absorption spectroscopy (TDLAS). This technique provides high sensitivity, speed, and selectivity. The TA100A is a rugged, portable instrument designed for use in the field. Common applications include gradient or eddy covariance flux measurements of methane or nitrous oxide and isotope ratio measurements of carbon dioxide or water vapor. The TA100A concentration measurements can be recorded with a Campbell Scientific datalogger, output as analog voltages, or sent to a computer through an Ethernet connection. OV-1

SE 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 1 2 3 4 5 6 7 8 9 10 DIFF H L H L H L H L H L H L H L H L H L H L SE 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 11 12 13 14 15 16 17 18 19 20 DIFF H L H L H L H L H L H L H L H L H L H L VX1 VX2 VX3 VX4 CAO1 CAO2 CONTROL I/O POWER UP Logan, Utah CR5000 MICROLOER IX1 IX2 IX3 IX4 IXR, ' _ Hm SN: P1 CONTROL I/O C5 C6 C7 C8 >2.0V <0.8V 5V 5V SDI-12 12V SDM-C1 SDM-C2 SDM-C3 12V SW-12 SW-12 A B C D E F 1 2 3 H I J K L M N O 4 5 6 P R S T U V W X Y Spc Cap 7 8 9 - + ( * / ) < = > $ Q Z 0 raph/ char P1 C1 C2 C3 C4 PgUp End PgDn Del Ins POWER OUT CURSOR ALPHA SHIFT BKSPC ESC ENTER CAUTION DC ONLY 12V ROUND POWER IN LU pc card 11-16 VDC status 12V MADE IN USA COMPUTER RS-232 (OPTICALLY ISOLATED) CS I/O TA100A Trace as Analyzer Overview OV1. System Components Figure OV1-1 illustrates the main system components needed to operate the TA100A. These system components include: TA100A Analyzer: The analyzer optics and electronics, mounted in an insulated fiberglass enclosure. Computer: A user-supplied computer to display data and set parameters. Optional Datalogger (CR5000 shown): Receives concentration data from the TA100A through the SDM interface cable. Intake (15838 shown): Filters the air sample and sets its flow rate. pump (RB0021-L shown): Pulls the air sample and reference gas through the analyzer at low pressure. Suction hose (7123 shown): Connects the analyzer to the sample pump. Supplied with RB0021 sample pump. Reference gas: tank of reference gas, with pressure regulator (supplied by user). Reference gas connection (15837 shown): Flow meter, needle valve, and tubing to connect the reference gas to the analyzer. Computer Datalogger TA100A Analyzer SDM Cable Ethernet Cable Intake Reference as Connection Suction Hose Reference as Pump FIURE OV1-1. TA100A System Components OV-2

TA100A Trace as Analyzer Overview OV2. Theory of Operation OV2.1 Optical System The TA100A optical system is shown schematically in Figure OV2.1-1. The optical source is a lead-salt tunable diode laser that operates between 80 and 140 K, depending on the individual laser. Two options are available to mount and cool the laser: the TA100A LN2 Laser Dewar and the TA100A Laser Cryocooler System. Both options include a laser mount that can accommodate one or two lasers (up to four lasers can be installed by adding the optional second laser mount). The LN2 Laser Dewar mounts inside the analyzer enclosure. It holds 10.4 liters of liquid nitrogen, and must be refilled twice per week. The Laser Cryocooler System uses a closed-cycle refrigeration system to cool the laser without liquid nitrogen. It includes a vacuum housing mounted inside the analyzer enclosure, an AC-powered compressor mounted outside the enclosure, and 3.1 m (10 ft) flexible gas transfer lines. Reference Detector The laser is simultaneously temperature and current controlled to produce a linear wavelength scan centered on a selected absorption line of the trace gas. The IR radiation from the laser is collimated and passed through a 1.5 m sample cell, where it is absorbed proportional to the concentration of the target gas. A beam splitter directs most of the energy through a focusing lens and short sample cell to the sample detector, and reflects a portion of the beam through a second focusing lens and a short reference cell to the reference detector. A prepared reference gas having a known concentration of the target gas flows through the reference cell. The reference signal provides a template for the spectral shape of the absorption line, allowing the concentration to be derived independent of the temperature or pressure of the sample gas or the spectral positions of the scan samples. The reference signal also provides feedback for a digital control algorithm to maintain the center of the spectral scan at the center of the absorption line. The simple optical design avoids the alignment problems associated with multiple-path absorption cells. The number of reflective surfaces is minimized to reduce errors caused by Fabry- Perot interference. To Pump Reference as In Dewar Detector Cell Laser To Pump In FIURE OV2.1-1. Schematic Diagram of TA100A Optical System OV-3

TA100A Trace as Analyzer Overview OV2.2 Laser Scan Sequence The laser is operated using a scan sequence that includes three phases: the zero current phase, the high current phase, and the modulation phase, as illustrated in Figure OV2.2-1. The modulation phase performs the actual spectral scan. During this phase the laser current is increased linearly over a small range (typically +/- 0.5 to 1 ma). The laser s emission wavenumber depends on its current. Therefore the laser s emission is scanned over a small range of frequencies (typically +/- 0.03 to 0.06 cm -1 ). During the zero current phase, the laser current is set to a value below the laser s emission threshold. Zero signifies the laser emits no optical power; it does not mean the current is zero. The zero current phase is used to measure the detector s dark response, i.e., the response with no laser signal. The reduced current during the zero phase dissipates less heat in the laser, causing it to cool slightly. The laser s emission frequency depends on its temperature as well as its current. Therefore the temperature perturbation caused by reduced current introduces a perturbation in the laser s emission frequency. During the high current phase the laser current is increased above its value during the modulation phase to replace the heat lost during the zero phase. This stabilizes the laser temperature quickly, minimizing the effect of the temperature perturbation. The entire scan sequence is repeated every 2 ms (500 scans per second). High Current Phase (Temperature Stabilization) Modulation Phase (Spectral Scan) Laser Current Zero Current Phase (Laser Off) Omitted Used in Calculation Detector Response 2 ms FIURE OV2.2-1. TA100A Laser Scan Sequence OV-4

TA100A Trace as Analyzer Overview OV2.3 Concentration Calculation The reference and sample detector signals are digitized at 50 khz (100 samples per scan), corrected for detector offset and nonlinearity, and converted to absorbance. A linear regression of sample absorbance vs. reference absorbance gives their ratio. The assumption that temperature and pressure are the same for the sample and reference gases is fundamental to the design of the TA100A. It allows the concentration of the sample, C S, to be calculated by: C s = ( CR )( LR )( D) L + L (1 D) S A where C R = concentration of reference gas, ppm L R = length of the short reference cell, cm L S = length of the short sample cell, cm L A = length of the long sample cell, cm D = ratio of sample to reference absorbance The trace gas concentration is calculated for each scan and then digitally filtered to reduce noise. OV3. Trace as Species Selection The TA100A can measure gases with absorption lines in the 3 to 10 micron range by selecting appropriate lasers, detectors, and reference gas. Lead-salt tunable diode lasers have a very limited tuning range, typically 1 to 3 cm -1 within a continuous tuning mode. In some cases more than one gas can be measured with the same laser, but usually each gas requires its own laser. The laser dewar has two laser positions available (four with an optional second laser mount), allowing selection of up to four different species by rotating the dewar, installing the corresponding cable, performing a simple optical realignment, and switching the reference gas. The standard detectors used in the TA100A are Peltier cooled, and operate at wavelengths up to 5 microns. These detectors are used for most gases of interest, including nitrous oxide (N 2 O), methane (CH 4 ), and carbon dioxide (CO 2 ). Some gases, such as ammonia (NH 3 ), have the strongest absorption lines at longer wavelengths, and require the optional long wavelength, liquid nitrogen-cooled detectors. A prepared reference gas must flow through the reference cell to provide a spectral absorption template for the target gas. The concentration of this reference gas is chosen to give approximately 50% absorption at the center of the absorption line. A 200 ft 3 (5.7 m 3 ) tank of reference gas will last over one year at the recommended flow rate of 10 ml min -1. OV-5

TA100A Trace as Analyzer Overview OV4. Multiple Scan Mode OV5. User Interface Software The TA100A can be configured to measure two or three gases simultaneously by alternating the spectral scan wavelength between two or three absorption lines. This technique requires that the absorption lines be close together (within about 1 cm -1 ), so it can be used only in very specific cases. The multiple scan mode is used to measure δ 13 C and δ 18 O isotope ratios in carbon dioxide or δ 18 O and δd isotope ratios in water vapor by tuning each scan to a different isotopologue. The multiple scan mode may also be used to measure some other pairs of gases, such as carbon monoxide and nitrous oxide, or nitrous oxide and methane, but the measurement noise will be higher than if a single gas is measured. For measurements of a single gas, the laser wavelength is chosen for one of the strongest absorption lines of that gas. Choosing a laser that can measure two gases simultaneously involves a compromise. Weaker absorption lines must be used in order to find a line for each gas within the laser s narrow tuning range. The TA100A user interface software runs on the user's PC. It displays the data in real time and allows the user to modify control parameters. Normally the trace gas concentration data are recorded by a datalogger, and the PC is required only for setting up the TA100A. However, the trace gas concentration data can also be saved to the PC's hard disk, making the datalogger unnecessary in some applications. The real time graphics screen is presented in Figure OV5-1. In the upper left corner is a box which displays the TA software version, the laser and detector temperatures, and the time. Beneath the time and temperature display is a blank area used for information and error message display. The rest of the top of the screen has five menu columns: run mode, dynamic parameters, detector video, special function enable/disable, and graph selections. OV-6

TA100A Trace as Analyzer Overview FIURE OV5-1. Real Time raphics Screen In the middle of the screen are graph 1 and graph 2, used to display userselectable variables. This example shows N 2 O concentration in graph 1 and laser temperature in graph 2. raph 3 is located at the bottom-center of the screen, and is also used to display user-selected variables. In this example graph 3 shows the sample cell pressure. At the bottom left corner of the screen are two high speed graphic windows that show the raw reference (REF) detector signal and the raw sample (SMP) detector signal, scaled to match the analog-to-digital converter (ADC) input range. At the bottom right corner of the screen are two more high speed graphic windows that display processed reference and sample signals. The user may select the type of data to display in these windows using the Detector Video menu or the Quick Keys. The number displayed at the top of these windows is either the transmittance or the absorbance at the center of the spectral scan, depending on the display mode selected. All four of the highspeed graphic windows have three vertical dashed lines. These lines show the center of the spectral scan and the range of data actually used to calculate concentration. Many of the operating parameters can be edited at the real time screen using the Parameters menu column. Other parameters are available at the Parameter Change menu. OV-7

SE 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 1 2 3 4 5 6 7 8 9 10 DIFF H L H L H L H L H L H L H L H L H L H L SE 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 11 12 13 14 15 16 17 18 19 20 DIFF H L H L H L H L H L H L H L H L H L H L CONTROL I/O CONTROL I/O POWER POWER OUT UP CAO1 CAO2 VX1 VX2 VX3 VX4 IX1 IX2 IX3 IX4 IXR P1 P2 C1 C2 C3 C4 C5 C6 C7 C8 >2.0V >0.8V 5V 5V SDI-12 12V SDM-C1 SDM-C2 SDM-C3 12V SW-12 SW-12 Logan, Utah CR5000 M ICROLO ER, _ 1 2 A B C D E F Hm PgUp 3 H I J K L M N O 4 5 6 P R S T U V W X Y Spc Cap End PgDn 7 8 9 - + ( * / ) < = > $ Q Z Del Ins 0 SN : raph/ char CAUTION DC ONLY POW ER IN 11-16 VDC SHIFT BKSPC ESC ENTER 12V ROUND LU CURSOR ALPHA COMPUTER RS-232 (OPTICALLY ISOLATED) M ADE IN USA CS I/O pc card status TA100A Trace as Analyzer Overview OV6. Micrometeorological Applications The TA100A is ideally suited to measure fluxes of trace gases using micrometeorological techniques. Its rugged design allows it to operate reliably in the field with minimal protection from the environment, and it incorporates several hardware and software features to facilitate these measurements. OV6.1 Eddy Covariance The TA100A's frequency response, sensitivity, and selectivity are optimized for measuring trace gas fluxes using the eddy covariance (EC) method. A typical eddy covariance system (illustrated in Figure OV6.1-1) includes a TA100A, a CSAT3 sonic anemometer, and a datalogger. Intake Pump CSAT3 Sonic Anemometer Datalogger PD1000 Dryer SDM Cables Tubing TA100A Analyzer Reference as Dryer Purge Tubing FIURE OV6.1-1. Example Eddy Covariance Flux Application The sonic anemometer and air sample intake are mounted on the measurement mast. Tubing connects the air sample intake to the inlet of a PD1000 high flow sample air dryer, which filters and dries the air sample. A needle valve at the outlet of the PD1000 sets the sample flow rate, typically to approximately 15 slpm. The TA100A analyzer is located near the base of the measurement mast to minimize the length of sample tubing. This minimizes the attenuation of high frequencies in the concentration data as the air sample flows through the tube. The sample pump requires minimal shelter and can be located up to 90 m (300 ft) away from the analyzer, connected by 1 ID suction hose. A Campbell Scientific datalogger collects synchronized data from the sonic anemometer and the TA100A through SDM interface cables. Analog outputs are available for use with other data loggers. The datalogger records the raw time series data as well as the results of an online EC flux calculation. OV-8

TA100A Trace as Analyzer Overview OV6.2 Flux radient The TA100A also supports the measurement of trace gas fluxes by the gradient method. A gradient valve switches the sample between the two intakes, the valid samples from each intake are averaged and the difference is computed. The valve control and the calculations can be performed either by a datalogger or by a computer and I/O module. The computer and I/O module provide a convenient user interface, but require shelter from the environment. The datalogger is generally preferred when an instrumentation trailer or other shelter is not available. Figure OV6.2-1 illustrates a typical gradient application. Two intake assemblies are mounted at different heights on the measurement mast. Tubing connects each intake assembly to a gradient valve assembly that selects one of the intakes at a time. The air sample from the selected intake flows through the PD1000 high flow sample air dryer, which filters and dries the air sample. A needle valve at the outlet of the PD1000 sets the sample flow rate, typically 5 to 10 slpm. Tubing connects the outlet of the dryer to the TA100A analyzer, which may be located 200 m (650 ft) or more away. The sample pump requires minimal shelter and can be located up to 90 m (300 ft) away from the analyzer, connected by 1 ID suction hose. This example shows a gradient flux measurement at a single site. However, the TA100A can also support flux gradient measurements at multiple sites by installing intake assemblies, a gradient valve assembly, and a sample dryer at each site, and a site selection system near the analyzer. The site selection system connects one site at a time to the analyzer. Normally each site is measured for 15 to 60 min before switching to the next site. Intake Assemblies radient Valve Assembly TA100A PC Surge Protector TA100A Analyzer Reference as PD1000 Dryer Digital Signal Cable Dryer Purge Pump Suction Hose FIURE OV6.2-1. Example radient Flux Application OV-9

TA100A Trace as Analyzer Overview OV6.3 Site Means Sampling Mode The TA100A s site means mode is similar to the flux gradient mode in that it controls switching valves and calculates mean concentrations for each intake. The difference between the two sampling modes is that the gradient mode considers the sample intakes in pairs, switching several times between an upper and lower intake before moving to another site, but the site means mode considers all of the intakes as one group. It cycles through all of the intakes in sequence. Applications for the site means mode include concentration profile measurements and trace gas flux measurements using the mass balance technique. Similar to the gradient mode, the valve switching and calculations can be performed either by a datalogger or by the TA100A computer and I/O module. Figure OV6.3-1 illustrates an eight-level vertical profile. The eight intake assemblies are arranged vertically on a single measurement tower. These intake assemblies include a filter to remove particulates and a critical flow orifice to set the sample flow (typically less than 1 slpm). A separate tube connects each intake assembly to the site selection system, which selects one of the intakes at a time. All of the unselected intakes are connected through the bypass tube to the 1 suction hose leading to the sample pump, keeping air flow at all times in all intake tubes. The flow from the selected intake goes through a PD625 low flow sample dryer to the TA100A analyzer. A second PD625 is used to provide dry air to purge the sample dryer. TA100A PC TA100A Analyzer Reference as Intakes Digital Control Cable Site Selection Sampling System Purge Dryer Dryer Dryer Purge Bypass Pump FIURE OV6.3-1. Example Profile Application OV-10

TA100A Trace as Analyzer Overview OV6.4 Absolute Concentration / Isotope Ratio Measurements The TA100A can be configured for highly accurate measurements of trace gas concentrations by performing frequent calibration. The TA100A has a small offset error caused by optical interference. This offset error changes slowly over time, with a standard deviation roughly equal to the short-term noise. Offset errors have little effect on flux measurements by either the gradient or eddy covariance technique, but may be important in other applications. For measurements of absolute trace gas concentration, the offset error can be removed by switching between a nonabsorbing gas (e.g. zero air) and the sample, using the gradient mode of operation. Applications such as isotopic ratio measurements require the highest possible accuracy. This is achieved using a frequent two-point calibration to correct for drift in the instrument gain and offset. High accuracy requires the flow rate for the calibration gases to be the same as for the sample air. Even though the sampling system can be designed so that calibration gases flow only when they are used, frequent calibration (every few minutes) consumes a large amount of calibration gas if high flow rates are used. The site means sampling mode is normally used because it works well at low flow rates. Figure OV6.4-1 illustrates a typical CO 2 isotope application. It is similar to the site means example above, but it includes two intakes connected to calibration tanks. A tank of nitrogen or CO 2 -free air is also shown connected to the analyzer to purge the air gap between the laser dewar and sample cell. This purge is required for CO 2 isotope measurements because of the high ambient concentration of CO 2 and the need for high accuracy. TA100A PC TA100A Analyzer Reference as Intakes Digital Control Cable Purge Dryer Dryer Purge as Site Selection Sampling System Dryer Purge Bypass Calibration ases Pump FIURE OV6.4-1. Example CO 2 Isotope Application OV-11

TA100A Trace as Analyzer Overview OV7. Specifications OV7.1 Measurement Specifications The preliminary frequency response and noise specifications given in this section are based on performance of the TA100, which has a 10 Hz update rate. The TA100A has a faster update rate, which should give improved performance. The frequency response is determined by the time required to flush the sample cell (480 ml volume). The frequency response was measured at 14.4 slpm flow rate and 50 mbar sample pressure (4.8 actual l/s) by injecting approximately 1 μl of N 2 O into the sample stream. The resulting time series and frequency response graphs are shown below (TA100 data are shown, with a 10 Hz update rate). 4 1 Concentration (ppmv) 3 2 1 Frequency Response 0.1 0 0.0 0.5 1.0 1.5 2.0 Time (sec) 0.01 0.1 1 5 Frequency (Hz) FIURE OV7.1-1. Time Series (left) and Frequency Response (right) The typical 10 Hz concentration measurement noise is given as the square root of the Allan variance with no averaging (i.e. the two-sample standard deviation. This is comparable to the standard deviation of the 10 Hz samples calculated over a relatively short time (10 s). The typical 30-minute average gradient resolution is given as the standard deviation of the difference between two intakes, averaged over 30 minutes. Values shown in Table OV7.7-1 are for ambient concentrations. TABLE OV7.1-1. Typical Concentration Measurement Noise as Wave number (cm -1 ) 10 Hz Noise (ppbv) 30-min radient Resolution (pptv) Nitrous Oxide N 2 O 2208.575 1.5 30 Methane CH 4 3017.711 7 140 Ammonia NH 3 1065.56 6 200 Carbon Monoxide CO 2176.284 3 60 Nitric Oxide NO 1900.08 13 260 Nitrogen Dioxide NO 2 1630.33 3 60 Sulfur Dioxide SO 2 1366.60 25 500 OV-12

TA100A Trace as Analyzer Overview The TA100A multiple-scan mode can be used to measure suitable pairs of gases. Typical performance for some examples is given in Table OV7.1-2. TABLE OV7.1-2. Typical Concentration Measurement Noise as Wave number (cm -1 ) 10 Hz Noise (ppbv) 30-min radient Resolution (pptv) Nitrous Oxide N 2 O 1271.077 7 140 and Methane CH 4 1270.785 18 360 N 2 O 2190.350 5 100 Nitrous Oxide and CO 2190.018 5 100 Carbon Monoxide N 2 O 2203.733 1.8 35 CO 2203.161 10 200 Nitrous Oxide and N 2 O 2243.110 2 40 Carbon Dioxide CO 2 2243.585 400 8000 The multiple-scan mode may also be used to measure different isotopologues of the same gas. Isotope ratio measurements are typically given in delta notation: δ = R R m std 1 where R m is the ratio of the isotopologue concentrations measured by the TA100A and R std is the standard isotope ratio. δ is reported in parts per thousand (per mil or ). Typical isotope ratio performance is given in Table OV7.1-3. The calibrated noise assumes a typical sampling scenario: two air sample intakes and two calibration samples measured in a 1 minute cycle. It is given as the standard deviation of the calibrated air sample measurements. TABLE OV7.1-3. Typical Isotope Ratio Measurement Noise as Isotope Ratio Wavenumber (cm -1 ) 10 Hz Noise Calibrated Noise Carbon Dioxide, CO 2 2293.881 0.2 ppm 0.05 ppm δ 13 C only δ 13 C 2294.481 0.5 0.1 CO 2 2308.225 0.6 ppm 0.15 ppm Carbon Dioxide, δ 13 C 2308.171 2.0 0.4 δ 13 C and δ 18 O δ 18 O 2308.416 2.0 0.4 Water, H 2 O 1501.846 10 ppm 2 ppm δd only δd 1501.813 8 2 H 2 O 1500.546 10 ppm 2 ppm Water, δ 18 O 1501.188 2 0.5 δ 18 O and δd δd 1501.116 20 5 OV-13

TA100A Trace as Analyzer Overview OV7.2 Physical Specifications Analyzer Length: 211 cm (83 in) Width: 47 cm (18.5 in) Height: 55 cm (21.5 in) Weight: 74.5 kg (164 lb) Optional Cryocooler Compressor Length: 31 cm (12 in) Width: 45 cm (18 in) Height: 38 cm (15 in) Weight: 32 kg (71 lb) Power Requirements Analyzer: Optional Heater: Optional Cryocooler Compressor: RB0021-L sample pump: This is a blank page. 90-264 Vac, 47-63 Hz, 50 W (max) 30 W (typ) 90-264 Vac, 47-63 Hz, 150 W (max) 100, 120, 220, or 240 Vac, 50/60 Hz, 500 W 115 Vac, 60 Hz, 950 W (other power options are available) OV-14