Development of high pressure-high vacuum-high conductance piston valve for gas-filled radiation detectors

Similar documents
Development of a high vacuum sample preparation system for helium mass spectrometer

Revision 2013 Vacuum Technology 1-3 day Good Vacuum Practice 1 Day Course Outline

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

PNEUMATIC PRESSURE CONTROLLERS

Quantitative study of sniffer leak rate and pressure drop leak rate of liquid nitrogen panels of SST-1 tokamak

Chapter 6 Leak detection

Air Operated Hydraulic Pumping Systems to 50,000 psi

Description Advanced Features Typical Applications

Smooth Vent Valve. Series XVD XL XL Q D- XVD XGT CYV

SPECIFICATIONS. Approximate Weight: 10 oz. Surface Finish: Ra micro inch or less MATERIALS OF CONSTRUCTION

Helium Mass Spectrometric Leak Detection In Large Size Process Plants

Self-operated Regulators Accessories Differential Pressure and Flow Regulators

PTF4 Pivotrol Pump (patented) version Dual Mechanism - Pressure Powered Pump

Technical Specifications of Hydrogen Isotope Handling and Recovery System

Introduction of Vacuum Science & Technology. Diffusion pumps used on the Calutron mass spectrometers during the Manhattan Project.

TECHNICAL DATA. Q = C v P S

Smooth Vent Valve XVD. Series

The Helium Leak Detector

TECHNICAL DATA. Q= Cv S

The Compact, Portable, Efficient and Economical Solution to boost Shop Compressed Air Pressure by 2 or 3 times

Smooth Vent Valve. XVD Series

Flow control valves Series RFU - RFO 2/

Series IRV1000/2000/3000

This educational seminar discusses creating, measuring, and troubleshooting Rough Vacuum.

Lesson 6: Flow Control Valves

Development of Helium Leak Testing System and Procedure for testing Welds of Steam Generator

Helium Leak Detection by Detector Probe Technique : Is it a Leak Detection or Concentration Detection Method?

TECHNICAL DATA. Q = C v P S

CSA Sample Draw Aspirator Adapter Operator s Manual

Fig. 1: Mechanical vacuum gauge (left) and capacitive vacuum gauge (right)

SERIES 30. Spring Operated Tank Blanketing Valve PROTECTOSEAL.

TECHNICAL DATA 3 MODEL G-3000 DRY VALVE RISER ASSEMBLY

Design of Pressure vessel for the Hydrocarbon release from the Vent Header Lines

VACUUM REGULATORS CONTENTS

Operation Manual Piston Sensed Gas Pressure Regulators

TECHNICAL DATA Q = C. v P S. 2 Model G-2000 Dry valve. Page 1 of 13

Calibrated or locked versions available. TREATMENT > Series MD regulators GENERAL DATA. New

Pressure Measurement

SEMASPEC Test Method for Determination of Cycle Life of Automatic Valves for Gas Distribution System Components

Please refer F 90 of the present manual for the identification of the valves used in the detector.

DIAPHRAGM VACUUM PUMPS AND COMPRESSORS

COMPACT TRIPLE WALL OUTLET STATION MODEL INSTALLATION AND OPERATING INSTRUCTIONS

Specialty Products. Specialty Products Index

TECHNICAL DATA OBSOLETE

Welker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual

Results of mathematical modelling the kinetics of gaseous exchange through small channels in micro dischargers

TECHNICAL DATA. Q = C v P S

TECHNICAL DATA. Q = C v P S

TECHNICAL DATA. Q = C v P S

JR Series. Gas Pressure Reducing Valve GAS PRESSURE REDUCING VALVE

Model 7989T Steel Pipe Squeezer Sch. 40 & Sch. 80. Operations Manual

WATER MADE EASY MARINE ENERGY MUNICIPAL INDUSTRIAL

Residual Gas Analysis Systems for Industry

Model 4000 Pressure Controller

Instruction Manual Updated 7/26/2011 Ver. 2.2

RK-IR Sample Draw Aspirator Adapter Operator s Manual

Vacuum System at IUAC

GA / 3.02 TL 4 TL 6. Cat. No Calibrated leaks GA / 3.02 TL 4 TL 6. Cat. No

Vacuum Science Techniques and Applications Dan Dessau Adv. Lab 2007

Description Pressure range Connection thread Device Page

Manual Solvent Purification System. Innovative Technology, Inc.

Level MEASUREMENT 1/2016

TECHNICAL DATA. Q = C v P S

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

SIHI LPH-X - Liquid Ring Compressors

RELIEF VALVES. General Description

Torque Tube TB300 Digital Transmitters

CVI Valve Line. Exceeds the industry s highest standards for reliability and performance

Chlorinator MODEL MK-I

BACK PRESSURE / SUSTAINING

RHPS Series RD(H)F40 User Manual. Read the complete manual before installing and using the regulator.

Model 130M Pneumatic Controller

Test pumps. Pressure sources. Good reasons for proper and reliable calibration

Mounting and operating instructions EB 2530 EN. Self-operated Pressure Regulator. Pressure Reducing Valve Type M 44-2

FLAMMABLE GASES AND FLAMMABLE CRYOGENIC FLUIDS

BACK PRESSURE / SUSTAINING

Vacuum Systems and Cryogenics for Integrated Circuit Fabrication Technology 01

Variable Temperature Storage Dewar Mount Inserts

Super-Clean Gas Filters from Restek

Basic & Controlled Atmosphere Glove Boxes

Vacuum. Dry Turbo Pumping Stations. Roughing Pumps for 1 K Helium Pots and Lambda Refrigerators. Hermetically Sealed Pumps for He3 Recirculation

World Area Differences Technical Information

Figure 1: Hydrostatic Pressure Forces Are Perpendicular to the Surface

HP/UHP Spring-Loaded Regulator

6.6 Relief Devices. Introduction

INSTRUCTIONS FOR MODELS SG3897 AND SG3898 CROSS PURGE ASSEMBLIES

ASME Boiler & Pressure Vessel Code Analysis of the 1497 MHz High-Current Cryomodule Helium Vessel

SERIES 10. Pilot Operated Tank Blanketing Valve, 2" PROTECTOSEAL.

How to specify a product. Process Sensors and Mechanical Instruments

Basic Pneumatics. Module 8: Pressure control valves. Academic Services PREPARED BY. April 2012

VACUUM & AIRLESS. Select from Kimble vacuum traps, manifolds, Schlenk tubes, adapters, and bubblers for your vacuum and airless glassware needs.

MEDICAL CYLINDER VALVES

CLASS DH - PNEUMATIC TESTING - GREATER THAN 1 BAR (15 PSIG)

Major Design Features

The Principles of Vacuum Technology

A Journal of Practical and Useful Vacuum Technology. By Phil Danielson

Flange Bolt Torquing. for Resistoflex Plastic-Lined Piping Products. Torquing. Retorquing. Hydrotesting. Annual retorquing

Differential pressure

[Instruments for vacuum measurement, checking and adjustment] 3

Inflatable Packer Single & Double. Single & Double Packer Dimension. Wireline Packer. Water Testing Packer (WTP) Packer

Transcription:

Journal of Physics: Conference Series Development of high pressure-high vacuum-high conductance piston valve for gas-filled radiation detectors To cite this article: D N Prasad et al 2008 J. Phys.: Conf. Ser. 114 012055 Related content - Kinetic Energy of Trapped Ions Cooled by Buffer Gas Chen Liang, She Lei, Li Jiao-Mei et al. - O-ring vacuum valve for use with a visible flow industrial glass apparatus G S Martin and A Lench - Simple half-turn valve A Lasbley and J Rosenblatt View the article online for updates and enhancements. This content was downloaded from IP address 148.251.232.83 on 03/11/2018 at 10:15

Development of High Pressure-High Vacuum-High Conductance Piston Valve for Gas-filled Radiation Detectors D N Prasad 1, R Ayyappan, Lata P Kamble, J P Singh, L V Muralikrishna, Mary Alex, V Balagi and P K Mukhopadhyay Electronics Division, Bhabha Atomic Research Centre, Mumbai, India dnprasad@barc.gov.in, dnprasadbarc@yahoo.co.in, jpsingh@barc.gov.in, maryalex@barc.gov.in, balagiv@barc.gov.in Abstract. Gas-filled radiation detectors need gas filling at pressures that range from few cms of mercury to as high as 25kg/cm 2 at room temperature. Before gas-filling these detectors require evacuation to a vacuum of the order of ~1 x 10-5 mbar. For these operations of evacuation and gas filling a system consisting of a vacuum pump with a high vacuum gauge, gas cylinder with a pressure gauge and a valve is used. The valve has to meet the three requirements of compatibility with high-pressure and high vacuum and high conductance. A piston valve suitable for the evacuation and gas filling of radiation detectors has been designed and fabricated to meet the above requirements. The stainless steel body (80mmx160mm overall dimensions) valve with a piston arrangement has a ½ inch inlet/outlet opening, neoprene/viton O-ring at piston face & diameter for sealing and a knob for opening and closing the valve. The piston movement mechanism is designed to have minimum wear of sealing O-rings. The valve has been hydrostatic pressure tested up to 75bars and has Helium leak rate of less than 9.6x10-9 m bar ltr/sec in vacuum mode and 2x10-7 mbar ltr/sec in pressure mode. As compared to a commercial diaphragm valve, which needed 3 hours to evacuate a 7litre chamber to 2.5x 10-5 mbar, the new valve achieved vacuum 7.4x10-6 mbar in the same time under the same conditions. 1.Introduction Gas-filled radiation detectors 1 (such as ionisation chambers and proportional counters for reactor control instrumentation and area monitoring applications) typically consist of two or more coaxial metallic electrodes mounted on high purity alumina spacers and housed in an enclosure (Fig.1). 1 Corresponding Author Fig. 1 Schematic Diagram of a Gas-Filled Detector c 2008 Ltd 1

One or more of these electrodes serve as the HT electrode while the remaining is used as signal electrode. The volume between these electrodes forms the sensitive volume of the chamber and is filled with a suitable gas at adequate pressure. Radiation falling on this volume generates ion pairs. The drift of these charges in opposite direction under the influence of an applied electric field constitutes the signal output from the chamber. The assembly is welded to a pair of end plates at the ends of the enclosure and the chambers are filled with a gas or a gas mixture. Prior to gas filling at pressures that range from few cm of mercury to as high as 25kg/cm 2 at room temperature, these detectors require evacuation to a vacuum of the order of ~1 x 10-5 mbar. For these operations of evacuation and gas filling, a system consisting of a vacuum pump with a high vacuum gauge, gas cylinder with a pressure regulator gauge, gas manifold and a high vacuum-high pressure-high conductance valve is used (Fig. 2). Since the volume of the detectors can be as high as a few litres, the valves used in these operations need to have high conductance as well. Fig. 2 Schematic diagram of evacuation-cum-gas-filling system Disadvantages of existing valves and need for indigenous development of a special valve: Conventional valves like bellow sealed valves; diaphragm valves, gate valve and butterfly valves are compatible with high vacuum but cannot withstand high pressures. These valves cannot be used in applications where the pressure exceeds few bars. Certain high-pressure cylinder gas regulators and needle valves for precious rare gases are compatible with both high pressure and high vacuum but have a low conductance. Hence the detectors need to be evacuated for longer periods of time to ensure complete evacuation of occluded gases. Moreover since the three requirements such as high pressurehigh vacuum and high conductance are rarely encountered in industry; commercial suppliers do not commonly manufacture such valves. On account of the difficulties encountered in the existing valves, there is a need to develop valves having high vacuum, high pressure and high conductance. Electronics Division, BARC has indigenously developed a piston valve suitable for the evacuation and gas filling of radiation detectors. 2

1.1 Design of the valve Figure 3 shows the construction details of the valve. Fig. 4&5 show the photographs of the valve and its components. The valve consists of a stainless steel body with a cylindrical hole for piston. A ½ inch inlet/outlet opening is provided for connecting the valve to the gas-filling system. Neoprene/viton O-rings are used at the piston face & diameter for sealing. A circular knob is provided for opening and closing the valve. The valve has an overall diameter of 80mm and an overall length of 160mm. One important feature of the valve is that the piston movement mechanism is designed to have minimum wear and tear of sealing O-rings. This has been achieved by the use of a screw mechanism, which facilitates the axial movement of the piston and prevents the rotational movement thereby minimizing the wear and tear of the O-rings. In addition to this, replacement of O-ring is facilitated by the easily demountable flange design. O-rings are well supported in grooves and have about 20 % compression between cylinder and piston. On account of the narrow surface of O-ring, the sealing material encounters least force to cause deformation/deflection when exposed to high pressure and ensures leak tightness at high pressure and high vacuum. The inlet / outlet opening has been designed to be ½ inch, compatible with a ½ vacuum angle valve and also since the gas filling tube in the filling assembly for any detector has a diameter of less than 10 mm. The valve body and components which need to withstand high pressure of 50 bar (750 psi) have been designed as per ASME pressure vessel code. Fig. 3. Construction and details of the valve 3

1.2 Two types of piston design were test: In the first design facial-sealing method for closing & opening of the valve was adopted by providing an O-Ring on piston face (Fig.3). This design is suitable only in uses where the flow is unidirectional. If the flow direction is reversed and valve is in the closed position there are possibilities of the O-Ring getting displaced from the groove during opening of the valve. This is an undesirable situation. This phenomenon is more pronounced at pressures in excess of a few bars. In the second design dynamic gland method for closing & opening of the valve was employed. In this design the O-Ring is mounted on the step diameter grooves provided in the piston (Fig. 3) Here, flow in both directions is possible since the O-ring is held is such a manner as to prevent it from getting displaced due to high pressures. This design therefore permits interchanging inlet and outlet. Fig. 4. Photograph of the piston valve Fig. 5. Photograph of the components of the Piston valve 4

2.0 Tests and results 2.1 Pressure and Leak Tests The valve was leak and pressure tested at CDM, BARC as per ASME pressure vessel code Section V and VIII. The hydrostatic pressure testing was carried out at 75 kg/cm 2. The valve did not show any pressure drop for half-hour period. Also the test ensured the mechanical ruggedness of the design. In the vacuum mode the valve had helium leak rate 9.6x10-9 m bar ltr/sec and in the pressure mode the Helium leak rate was 2x10-7 mbar ltr/sec. 2.2 Comparison of performance of valves To compare the performance of the indigenously developed valve with the conventional valves, an experiment was conducted to determine the vacuum achieved, the time taken to achieve the vacuum and the maximum pressure that the valves could withstand. Three valves which included a diaphragm valve (1/2 ) and a needle valve (1/2 ) obtained from commercial suppliers and the in-house developed piston valve were tested. Each of the valves was connected to an evacuation and gas filling system (Fig. 6), which consisted of a turbo molecular pump, high vacuum gauge, pressure gauge, and a 7ltr spherical ionisation chamber. Fig. 6. Photograph of the evacuation-cum-gas-filling system With the diaphragm valve in the circuit the system was able to evacuate the chamber to a vacuum of 3.0 x 10-4 mbar in 30 minutes. The ultimate vacuum achieved after 6 hours was 1.2x 10-5 mbar. It was also observed that the diaphragm valve started leaking at pressure above 3 bars. With the needle valve in the circuit the system was able to evacuate the same chamber to a vacuum of 2.2 x 10-4 mbar in 30 minutes. The ultimate vacuum achieved after 6 hours was 8.6 x 10-6 mbar. 5

Although needle valves can withstand high pressures of the order of few tens of bars, the flow rate is low as seen from the above result. On the other hand the present piston valve provided 8.6x 10-5 mbar vacuum in 30 minutes. The ultimate vacuum achieved in 6 hours was 4.6 x 10-6 mbar. Fig. 7 shows the comparison of the performance of the three valves. Fig. 7. Comparison of Performance of Different Valves 3.0 Conclusion The present design of the valve is capable of meeting the three-fold requirement of ability to withstand high-pressure, compatibility with high vacuum and high conductance. The valve is compatible with vacuums of the order of 2.5x10-6 mbar. 4.0 Acknowledgement The authors wish to acknowledge the staff of Electronics Division Workshop for their prompt and whole-hearted support to this work. References G F. Knoll, Radiation Detection and Measurement ASME B&P Vessel Code Section V&VIII V.V. Rao, T.B. Ghose and K.L. Chopra Vacuum Science and Technology, Allied Publishers Ltd., New Delhi (2001) 6