TITLE: DNA minitube - Red

Similar documents
PIPETMAN F ENGLISH. Other languages available on our website:

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

TapeStation Tips & Tricks

Chapter 2. Pipetting techniques. Guide to Pipetting

Adjust the Volume Display

Standard Operational Procedure for Capillary Electrophoresis

1020 Industrial Drive, Orlinda, TN fax

Lab 1. Instrumentation Familiarity: Using Micropipetters and Serological Pipettes

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay

A N F G J I C D D L A B C D E F

Fisherbrand. Finnpipette INSTRUCTIONS FOR USE PARK LANE Pittsburgh PA

UNCC Biotechnology and Bioinformatics Camp. Dr. Jennifer Weller Summer 2010

Pipetting Small Volumes

MULTICHANNEL. Labnet International, Inc.

Login to ilab Kiosk. Revised 05/22/2018. Load your sample:

Improving Pipetting Techniques

Gel Loading Practice MiniLab

RayBio Human vwf ELISA Kit

Version BA D

Varipette Operating manual

Lab Skills Practice: Pipetting Small Volumes. B3 Summer Science Camp at Olympic High School 2016

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

II. MATERIALS REQUIRED FOR PERFORMANCE TESTING

DISSOLUTION TEST FOR SOLID DOSAGE FORMS

Aerobic Respiration. Evaluation copy

VIAFLO 96. Electronic 96 Channel Hand Held Pipettor

Multiplate Platelet Function Analyzer: An Overview

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010

biosensis Mouse CXCL10/IP-10 ELISA Kit Protocol

Agilent 1290 Infinity Pump Head Maintenance

MOLEBIO LAB #1: Microquantity Measurement

25/04/56. LTC Carmen A. Bell Chief, Laboratory Operations USAMC-AFRIMS

Experimental Procedure

Improve pipetting results in pharmaceutical formulation by using MICROMAN E

CHM250 Calibration and Measurement Lab. Balance Calibration

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Transferpette. Testing Instructions (SOP) 1. Introduction. May 2009

711 DISSOLUTION. Portions of the present general chapter text that are national USP text, and therefore not part of the

Packing Procedure Recommendation. For. YMC RP silica. ODS-A, ODS-AQ, Octyl, Butyl, YMCbasic

RayBio Human TNF-alpha ELISA Kit

Cell Disruption System

Register your instrument! Eppendorf Xplorer / Eppendorf Xplorer plus. Adjustment

UWPR. LC-MS setup. Before you connect your trap and column. Connecting your trap and column

Observations Possible Causes Recommended Actions. Turquoise color

Operating Instructions. Rainin Classic Manual Pipette

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

Armfield Distillation Column Operation Guidelines

USER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve

PROTOCOLS. Coomassie Brilliant Blue G (Sigma 98%) ,4 mg. Ethanol 95% ml. Phosphoric acid 85%.(w/v) ml

S-series USER MANUAL. a single tube system developed for high performance sample preparation using adaptive focused acoustic energy

QUICK START GUIDE TO DRINKING WATER MONITORING

Microprojectile Bombardment Protocol

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

GCMSD-Headspace Analysis SOP

Introduction IP codes are used in electrical product catalogs, etc., to indicate waterproofness and so on.

ab IgG1 Human ELISA Kit

TEST PROTOCOL VERIFICATION OF THE BEHAVIOUR OF A MECHANICAL COMPRESSION COUPLING DURING SETTLEMENT

Laboratory #2 Pipetting Technique and Micropipette Calibration Skills=40 pts

Competency Record for Sonic Cleaning

Transferpette -8/-12

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION

VERTICAL BLADDER TANK

General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

A N B P F G J I C D D L A B C D E F

Exhibit 4. Determination of Static Pressure Performance of the Healy Clean Air Separator (Executive Orders VR-201-N and VR-202-N)

R I T. Title: Amray 1830 SEM Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 09/29/03 1 SCOPE 2 REFERENCE DOCUMENTS

Lipodrop 2.0 device coating

Eliminating Sources of Pipetting Error in the Forensic Laboratory David M. Epstein Director of Scientific Operations. Abstract.

User s Manual. Digital micro pipette for liquid handling CERTIFIED ISO9001

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

TECHNICAL DATA. Q = C v P S

Thermo Scientific Chromatography Well Plates

Thermo Scientific Chromatography Well Plates

Laser-Induced Bubbles in Glycerol-Water Mixtures

Integrity Testing LifeASSURE BNA045 and BNA065 Series Filter Cartridges

Contacts. Quick Start Guide

Pipettor. User Manual

ANALOG ULTRASOUND WATTMETER USP-30

EVOLUTION OF RUPTURE DISC MANUFACTURING TECHNOLOGY AND PERFORMANCE

Final Assembly Instructions Bikes with Quill Stems

Technical Procedure for General Laboratory Equipment

COMBINATION AIR RELEASE DEGASSING (CARD) VALVES INSTALLATION AND MAINTENANCE MANUAL

VWR UHP j :01 Page 1. ENGLISH 1 22 DEUTSCH FRANÇAIS ESPAπOL ITALIANO

Institute For Thermal Processing Specialists

HORIZONTAL BLADDER TANK

THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY

Radnoti Liver Perfusion System

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:

Preparation and Installation of the Sanitary BDI-FLX Sensor and Connection to the BDI-FLX Interface Cable

AKTA pure 25 New Owner s Intro

Angstrom E-Beam Instructions. PROCESS CHECKS: the tooling factors of each metal; Ti/Au layer for wire bonding pull test.

RayPette Plus Autoclavable Pipette. User Manual

Calibration of Volumetric Glassware

Final Assembly Instructions Bikes with Threaded Headsets

[USING THE VITROBOT April 9, Using the Vitrobot

SOP: Buck Scientific BLC-20P HPLC Operation

Mass Spec will not Autotune

Transcription:

TITLE: DNA minitube - Red Summary of Operating Conditions: Target Base Pair (Peak) 5 kb (nominal) Please see specific recommendation chapter regarding optimization of operating conditions Temperature (bath) 19 21 C Duty Cycle 20% Intensity 1 Cycles per Burst 1,000 Total Treatment Time 600 seconds Power mode Frequency Sweeping Degassing mode Continuous Volume 200µl Buffer Tris EDTA, ph 8.0 DNA mass 2-20µg Starting material > 50kb Water level (FILL/RUN) S2 level 16 E210 level 13 AFA Intensifier No Intensifier (remove from E transducer) Supplies Part Number Sample vessel minitube Red polyethylene tube (Red) with Cap 520066 Preparation station minitube loading and unloading station 500207 Holder for S2 minitube holder (single tube) 500206 Holders for E210 minitube rack (24 tubes) 500205 Recommended settings are subject to change without notice. See following link www.covarisinc.com/pdf/pn_010301.pdf for updates to this document. Part number: 010301 Rev A 1 P a g e

Recommendations specific for minitubes Red The Covaris AFA process is highly reproducible, however, in the extremely low power application of fragmenting DNA molecules to large mean fragment sizes, additional attention must be paid to the following treatment attributes to ensure best results: DNA Sample size: Especially for the shearing into large fragments, it is very important to keep at least a 10 fold ratio as a minimum between the size of the starting material and the targeted mean of the fragment size distribution. For example, this means a starting material size should be > 50kb for the minitube red to achieve a target fragment distribution with a mean of 5kb. Treatment: As the DNA fragmenting process is rate limited, fragment size generation (defined by mean peak base pair size) is affected by treatment duration and parameters: 1. Duration: Minor adjustments in treatment duration may be made to optimize results for various sample types and concentrations. Duration listed in this document is recommended guidelines. Actual results may vary depending on the amount and type of starting material, concentration, and/or viscosity. Covaris recommends setting up a time dose response experiment (e.g., 10, 15, 20 minutes) for determining appropriate treatment times. Larger length starting material (e.g., 100kb) and larger mass (above. 20 µg) may require a longer dose to ensure a homogeneous shearing result. A longer duration will produce smaller fragments. 2. Parameters: Control should be maintained over editing of Method files, so that acoustic parameters (intensity, duty cycle, cycles per burst), and well plate definitions, once proven effective, are preserved. Water: The bath water is employed to couple acoustic energy to the sample vessel: 1. Purity: When applying low power acoustics in rate limited applications, foreign materials such as algae and particulates may scatter the high frequency focused acoustic beam, resulting in a shift to larger mean fragment size. Bath water should be pure distilled or DI water, changed daily or cleansed by a Covaris Water Conditioning System. 2. Degas Level: Similarly, insufficient degas levels within the bath may result in poor acoustic coupling and thereby shift the mean fragment size. System degas pumps should be run in advance of and during AFA treatments, as detailed in instrument User Manuals. 3. Temperature: Warmer temperatures promote less forceful collapse of acoustic cavities within the sample fluid, causing a shift toward larger mean fragment size. Bath temperature (as reported by SonoLAB software) should therefore be closely controlled and matched runto run and day to day. Employ the temperature alert feature in SonoLAB to warn of a failure to maintain control of bath temperature. 4. Level: The water level is a critical parameter in the red minitube process; attention should be paid to maintaining a consistent water level, according to published protocols. If using a Covaris Water Conditioning System, check levels daily to restore water lost to evaporation. Part number: 010301 Rev A 2 P a g e

In summary, when employing the Covaris AFA for generation of large kilobase fragments, control and verification of treatment attributes and water quality will reduce variance and promote consistent, satisfactory shearing results. Materials 1. Sample Vessel: Covaris minitube Red. The minitube is a high density polyethylene tube (with a vented, separate polypropylene cap) with an AFA disc at the bottom. Warning: store the plugs in a cool, dry place before use. Avoid contacting or touching the bottom of the MiniTUBE 2. Buffer: Tris EDTA, ph 8.0. 3. Sample Concentration: 2 20μg DNA in 200 μl 4. Rack or holder: E210 minitube rack (p/n: 500205) or S2 minitube holder (p/n: 500206) Operating Conditions 1. Fill the tank with fresh de ionized water to proper fill line. The S2 or E210 tank should be equipped with a graduated water level label. If the tank lacks this label, please contact Covaris. During treatments, the minitube should be slightly immersed in the water to ensure a good acoustic path from the AFA transducer. a. For S2 system equipped with a graduated fill line label, level = 16 b. For E210 system equipped with a graduated fill line label, level = 13 2. Degas water for recommended time period. To maintain degassed water, keep the pump continuously on during operation and sample processing. Do not turn the pump off. a. For S2 system: at least 30 minutes b. For E210 system: at least 60 minutes 3. Set the chiller to the right temperature. a. When set at 18.5 C, the S2 and E210 temperature software display should settle near 20 C. b. Depending on environmental conditions (for example, high relative humidity) the chiller may have to be set a little lower (or higher) to maintain the bath temperature between 19 21 C to offset the thermal transfer loss between the chiller and the apparatus. Method for Using the S2 Part number: 010301 Rev A 3 P a g e

1. Set up the Covaris S2 at the appropriate temperature following the operating conditions above. 2. Slowly pipette 200 µl of DNA solution into the minitube. As the bottom is hydrophobic, the sample sometimes doesn t cover it completely. To break the surface tension, gently stir the sample with your pipette tip until the sample fully covers the bottom of the minitube. Ensure that all of the solution is at the interior bottom of the tube and avoid droplets adhered to the interior sides of the tube. Note that this is a low power AFA process; the sample will not splash during the acoustic energy delivery. CAUTION: keep the tube vertical at all times during the fragmentation process 3. Gently press the cap into place until it seats in the top of the tube. The caps provided in the minitube kit are vented to relieve pressure within the tube. 4. Place the capped minitube into the holder, 1) Depress the steel piston of the S2 holder. 2) Gently insert the tube into the S2 holder from the side. The holder is not symmetrical insert or remove the tube from the open side. NOTE: Do not contact the bottom of the tube. 3) Pull tube to the bottom of the holder. Attempt to center the tube on the axis of the holder. 4) Gently release the steel piston, taking care that the tube is centered and aligned to the axis of the holder. 5. Place the holder in the S2. Take care to slightly tilt the holder inserting it, in order to avoid trapping bubbles at the bottom of the minitube. If a bubble is trapped at the bottom, remove the holder from the S2 and repeat this step. CAUTION: Do not run treatments with a bubble trapped at the outside bottom of a minitube Part number: 010301 Rev A 4 P a g e

6. Process the samples according to the settings provided in the Summary of Operating Conditions on page 1. 7. At the completion of this treatment remove the minitube from the holder. For maximum recovery of the sample, tilt the minitube and slowly remove the sample with a pipette. Alternately, you may return the tube to the preparation station tilted approximately 30 degrees on its stand. Typically, you should recover more than 190 μl. The sample should be transferred to another tube for long term storage. Method for Using the E210 When operating an E210, please verify that any 500141 Intensifier has been removed from the transducer prior to running a method. The Intensifier may be removed by following the instruction in Covaris document 010111. 1. Set up the Covaris E210 at appropriate temperature following the operating conditions above. 2. Load up to 24 minitubes in the E210 rack. 3. Slowly pipette 200 µl of DNA solution into the minitubes. As the bottom is hydrophobic, the sample sometimes doesn t cover it completely. To break the surface tension, gently stir the sample with your pipette tip until the sample fully covers the bottom of each minitube. Ensure that all of the solution is at the interior bottom of the tube and avoid droplets adhered to the interior sides of the tube. Note that this is a low power AFA process; the sample will not splash during the acoustic energy delivery. 4. Gently press the caps into place until they seat in the top of the tube. The caps provided in the minitube kit are vented to relieve pressure within the tube. 5. Press the rack cover into place until it rests on top of the caps. Tighten the center screw to secure the rack cover. Part number: 010301 Rev A 5 P a g e

6. Set up your treatment in SonoLAB according to the settings provided in the Summary of Operating Conditions on page 1. Use plate definition 500205: 24 position MiniTube +15mm Z offset.wpl. 7. Before loading your rack in the holder, press the Check Alignment button in SonoLAB to bring the holder into position at the water surface. Then, place the rack into the holder, tipping the rack slightly as the tubes enter the water, so that any bubbles forming under the tubes can escape and are not trapped in the acoustic path. Observe the underside of the tubes if a bubble is trapped at the bottom of a minitube, remove the rack from the holder and repeat this step. With a correct water level, you shouldn t see any air trap below the rack Warning: Do not run treatments with a bubble trapped at the bottom of a minitube 8. Run the treatment and process your samples. 9. At the completion of this treatment, slowly remove the rack from the holder keeping the samples in a vertical orientation. For maximum recovery of the sample, slightly tilt the rack and slowly remove the sample with a pipette. Alternately, you may return the tubes to the preparation station tilted approximately 30 degrees on its stand. Typically, you should recover more than 190 μl. The sample should be transferred to another tube for long term storage. Part number: 010301 Rev A 6 P a g e

Supplementary Data: Please note that there are two sources of variation in both peak value and distribution: 1) the physical process of DNA fragmentation is random and 2) the analytical process to determine fragment size has inherent variances (for example, gel electrophoresis and electropherograms). Therefore, fragment distributions and peak values, even from technical replicates, may not appear identical. If the sheared DNA sample will be column purified or concentrated prior to analysis on an agarose gel or Agilent Bioanalyzer, please remember to take out an aliquot for use as control prior to that step. Column purification and concentration of the sheared DNA will generate a biased fragment distribution profile due to the inherent greater loss of the smaller DNA fragments. Also please note that for the large fragment size, sizing by Bioanalyser and Agarose gel don t give exactly the same value. Typical DNA Recovery Volume recovery (for 200 μl initial sample) Concentration recovery (as measured with an Invitrogen Qubit fluorometer) Overall recovery > 95 % > 95 % > 90 % Part number: 010301 Rev A 7 P a g e

Sample to sample reproducibility Figure 1 Agilent 2100 Bioanalyzer electropherogram. 12 replicates using Covaris MiniTUBE - Red (10ug of E. Coli DNA in 200 μl TE) Inset: 0.7% Agarose gel. 14 replicates using Covaris MiniTUBE Red Part number: 010301 Rev A 8 P a g e

Concentration independent 5kb DNA shearing Figure 2 Different mass of E. Coli DNA (in 200 μl TE buffer) sheared in minitube red following the same operating conditions on page 1. 0.7% Agarose gel and Bioanalyzer 2100 electropherogram. Part number: 010301 Rev A 9 P a g e