A NEW INFANT OSCILLATORY VENTILATOR
|
|
- Clifton Potter
- 5 years ago
- Views:
Transcription
1 British Journal of Anaesthesia1990; 64: A NEW INFANT OSCILLATORY VENTILATOR M. K. CHAKRABARTI, A. HOLDCROFT, S. SAPSED-BYRNE AND J. G. WHITWAM SUMMARY A new, simple and inexpeive oscillatory ventilator is described in which a rotating jet mounted in the breathing duct generates cyclically positive and negative pressures in the airway with a sinusoidal flow waveform. Unlike conventional oscillatory ventilators it is free from restrictio to ipirator/ or expiratory gas flows and open to atmosphere at all times, making it intriically a safe system for ventilation. A prototype rotating jet oscillatory ventilator designed for application in infants was evaluated in rabbits (mean weight 3.8 kg). The positive peak and mean airway pressures were significantly less during oscillatory ventilation at 300 and 420 b.p.m. compared with normal and high frequency positive pressure ventilation at 30 and 300 b.p.m., respectively, while maintaining blood-gas teio within the normal range. An increase in the oscillatory frequency from 300 to 420 b.p.m. provided no further benefit in terms of airway pressure, tidal volume or blood-gas teio. KEY WORDS Ventilation: high frequency, oscillation, airway pressure. High frequency oscillatory ventilation (HFOV) has been used successfull y in the treatment of infants with respiratory failure from several causes [1-6]. Conventional oscillatory ventilators contain a piston or bellows pump which provides cyclically positive and negative pressures in the airway with a sinusoidal flow waveform. The respiratory fresh gas (RFG) is supplied as a continuous bias flow, usually at the proximal end of the tracheal tube. A narrow-bore outlet tube is used to discharge the expired gas from the patients along with the bias flow gas. It acts as a low pa s filter to impede the discharge of the high frequency ipiratory gas flow so that it is delivered to the lung. The tidal volume is adjusted normally by changing the excursion of the piston or bellows or by introducing a leak [5,6]. In all such systems, to avoid the accumulation of carbon dioxide from expired gas and to meet the peak ipiratory flow demand during spontaneous ipiration, the RFG must be high, which in turn may generate undesirably high continuous positive pressure in the airways (PEEP) because of the high resistance of the expiratory outlet. If the expiratory resistance is decreased to obviate the PEEP effect, most of the ipiratory tidal volume is lost. Some workers have used very high RFG with balanced suction at the expiratory outlet to counter this problem [7]. A new, simple and inexpeive ventilator is described which provides oscillatory ventilation especially in infants. METHODS Oscillatory ventilator The new oscillatory ventilator is essentiall y a rotating gas jet (J r ) placed dowtream in the widebore open limb of a T-piece breathing system and alternately directed upstream and dowtream (fig. 1). The warmed and humidified, low pressure RFG is supplied to the proximal end of the tracheal tube, which is connected to a standard three-way connector, at a flow rate of not less than 300 ml kg" 1 with a minimum of 2 litre min" 1. The rotating jet, the orifice of which has an internal diameter of 1.2 mm, is coupled to a variable speed electric motor and the oscillation frequency is set by controlling the speed of the motor. The driving M. K. CHAKRABARTI, B.SC, M.PHIL. ; A. HOLDCROFT, M.B., CH.B., F.F.A.R.C.S., M.D.J S. SAPSED-BYRNE, A.I.BIOL.J J. G. WHITWAM, M.B., CH.B., PH.D., F.R.c.p., F.F.A.R.C.S. ; Department of Anaesthetics, Royal Postgraduat e Medical School, Hammersmith Hospital, DuCane Road, London W12 ONN. Accepted for Publication: September 4, 1989.
2 NEW INFANT VENTILATOR 375 RFG- Air-oxygen or anaesthetics Gas Humidifier Outlet FIG. 1. Schemati c diagram of oscillatory ventilator. RFG = Respiratory fresh gas; Fl and F2 = bacterial niters; J r = rotating jet; J p = PEEP jet; PR1 and PR2 = pressure regulators for the driving pressure for J r and J p. gas to the rotating jet is supplied from a standard hospital pipe line (400 kpa, 60 lbf in 2 ) through a pressure or flow regulator (PR1). The outlet of the pressure regulator is connected to the rotating jet assembly. The gas flows into a cavity surrounding the jet shaft, which in turn connects to the jet nozzle through a 5-mm channel in the shaft. The outer body in which the jet shaft rotates is made of PTFE material so that no gas leaks between the shaft and the body; a high pressure gas seal could be used. However, as the gas flows continuously through a comparativel y low resistance path, gas leakage has not been a problem. The driving gas may be any dry gas, either the same air-oxygen mixture as the RFG, or air. The open limb of the T-piece system is a standard disposable infant ventilator tube (i.d. 10 mm; not less than 50 cm long). The outlet tube of the jet assembl y is 8-10 mm i.d. and cm long. When driving gas is supplied to the rotating jet it cyclically generates sinusoidal flow in the breathing duct, confirmed by Fourier analysis using a "waterfall" programme (Cambridge Electronic Design, CED 1401). A sinusoidal positivenegative pressure wave-form is generated with an ipiratory to expiratory (I:E) ratio of 1:1, which acts as a pneumatic piston to drive the RFG in and out of the lungs. The magnitude of the pressure which is generated and applied to the airways depends on the driving pressure, the gas flow, and the design of the jet assembly. The nominal driving gas coumption varies between 10 and 15 litre min" 1, depending on the airway pressure requirement. PEEP is created by a fixed jet (J p ) with an internal diameter of 0.8 mm placed dowtream of the rotating jet (fig. 1). A pressure or a flow regulator (PR2) controls the magnitude of externally applied PEEP, in a manner identical to that of a new infant ventilator described elsewhere [8]. At the outlet of the jet manifold (fig. 1), a wide bore tube (standard disposable tube of 22 mm i.d.) and a bacterial filter (F2) of negligible resistance to gas flows ( < 0.1 kpa at 1 litre s" 1 ) are used as a silencer to suppres jet noise. In clinical application, another similar filter (Fl) may be used between the breathing system and jet manifold to avoid bacterial contamination of the jet assembl y [8]. The jet assembl y can be sterilized in situ by nebulizing a chemical agent through the assembl y [8]. Alternatively, it may be removed and autoclaved or gas-sterilized. Although this particular oscillatory ventilator is designed for infants, the same principle could be used for adults, with suitable adjustment of the jet assembly. However, the gas coumptio are much greater.
3 376 Evaluationin rabbits The new ventilator was used to apply HFOV to seven rabbits (mean (SD) weight 3.8 (0.4) kg) with normal lungs. Comparison was made with high frequency ventilation (HFPPV) at 300 b.p.m. using a similar valvele s infant ventilator designed for both normal and high frequency ventilation [8,9]. Anaesthesi a was induced with thiopentone mg kg" 1 into an ear vein and maintained with a continuous infusion of 1 % thiopentone 6-9 ml h" 1 (20 mg kg" 1 h" 1 ). The femoral artery and vein were cannulated for arterial blood sampling and injection of drugs and fluids, respectively. The animals were paralysed with suxamethonium 10 mg i.v. in repeated bolus doses after tracheotomy had been performed with an uncuffed paediatric tracheal tube (4.5 mm i.d., length 10 cm). A snare was placed around the trachea to prevent gas leak and, initially, ventilation was controlled (IPPV) at 30 b.p.m. with an I: E ratio of 1:2 using the infant ventilator [8,9]. The pressure applied to the driving jet was adjusted to maintain the airway pressure at cmhjo and Pa COt at kpa. The arterial and airway pressures were measured by strain gauge traducers and displayed and recorded (M19 Devices, Welwyn Garden City, U.K.). Calibration was performed agait mercury and water colum. Airway pressure (Paw) was measured at the proximal end of the tracheal tube at the three-way connector. After the control samples and recordings were taken either HFPPV at 300 b.p.m. or HFOV at 300 and 420 b.p.m. were applied. The driving pressure of the rotating jet was adjusted to maintain a similar Pa co during HFOV as during HFPPV. At least 20 min was allowed after each change in mode or frequency of ventilation before bloodgas samples were taken and pressure measurements recorded. Arterial blood-gas samples were analysed immediately with a Radiometer bloodgas analyser. Tidal volumes (KT) were measured at each frequency of ventilation or oscillation with a pneumotachograp h system (Gould Medical) by placing the flow traducer between the tracheal tube and the three-way connector of the breathing duct. After the measuremen t of VT the flow traducer head was removed to avoid an increase in ventilation deadspace. The pneumotachograp h recording system used has a frequency respoe flat to 80 Hz and was calibrated before each measuremen t with a gas syringe. BRITISH JOURNAL OF ANAESTHESIA Statistical analysis was performed using twoway analysis of variance and paired t tests where appropriate. P < 0.05 was regarded as significant. RESULTS The flow waveform during HFOV at 300 b.p.m. in rabbits had a sinusoidal pattern creating sinusoidal positive and negative airway pressures as ipiration and expiration, respectivel y (fig. 2). The driving gas coumption was litre min" 1. Airway pressure waveform also was sinusoidal (fig. 2). The meanpeak airway pressure(paw) decreased significantly during HFPPV at 300 b.p.m. and HFOV at 300 and 420 b.p.m. compared with the control IPPV (table I), but differences in ipired peak Paw between HFOV and HFPPV were not significant. During HFOV, but not HFPPV, negative airway pressures were generated during the expiratory phase (table I). The mean airway pressuresduring HFOV decreased significantly compared with HFPPV and control IPPV (30 b.p.m.). A low PEEP of 0.2 kpa was observed during HFPPV, but not during HFOV (table I). Tidal volumewaveform produced by the ventilator is shown in figure 2. FT decreased significantly from a mean value of about 10 ml kg" 1 to 2 ml kg" 1 during HFPPV and HFOV compared ~ , 0 20J M M FIG. 2. Record of flow, airway pressure and volume characteristics of the new oscillatory ventilator. Oscillatory frequency = 420 b.p.m. Upward deflectio = ipiratory phase. 1s
4 NEW INFANT VENTILATOR 377 TABLE I. Ventilatory and blood-gasdata {mean {SEM)) urith different modesof ventilation in rabbits (n = 7). IPPV = Intermittent positive pressureventilation {control); HFPPV = high frequency positive pressureventilation; HFOV = high frequency oscillatory ventilation. *P < 0.05, paired t test comparedwith IPPV; ^analysis of variance IPPV 30 b.p.m. HFPPV 300b.p.m. HFOV 300 b.p.m. HFOV 420 b.p.m. Peak airway pressure (kpa) lpiratory Expiratory Mean airway pressure (kpa) Tidal volume (ml) /X-o, (kpa) Pao.CkPa) ph (arterial) Mean arterial pressure (mm Hg) 1.3(0.17) (0.07) 41 (5.6) 4.2 (0.27) 54.3 (3.3) (0.052) 113(8) 0.9 (0.20)* + 0.2(0.05) 0.50 (0.09) 8(1.4)* 4.0 (0.41) 57.3 (3.6) (0.019) 114(12) 0.6(0.16)* -0.4(1.0)* 0.17(0.04)* 7.5(1.2)* 4.1 (0.51) 55.2 (3.6) (0.019) 113(10) 0.5(0.10)* -0.4(0.09)* 0.16(0.04)* 7.1 (1.8)* 4.5 (0.81) 51.4(3.8) (0.025) 110(9) with control IPPV, without a change in Pa COi (table I). There was no significant difference in VT between HFPPV and HFOV at 300 and 420 b.p.m. Ventilation was maintained adequatel y throughout, as reflected in Pa COf values (table I). However, there was a smal increase in Pa^ during HFOV at 420 b.p.m. compared with 300 b.p.m., although the driving pressure was kept cotant. There were no significant differences in ph, Pao or mean arterial pressure (MAP) during the study period (table I). DISCUSSION The phenomenon of gas trapping in the lungs during high frequency ventilation (HFV) because of short expiratory time compared with the time cotant of the pulmonary system is well known [8]. An active expiratory phase during HFOV may overcome or minimize the problem of an undesirable increase in lung volume during HFV, and hence maintain a lesser mean airway and alveolar pressure. Oscillatory ventilators normally comprise either a reciprocating piston or bellows and the ventilation system is almost closed, with high impedance to expiratory flow [5 7]. The new oscillatory ventilator described here, although providing very similar ipiratory and expiratory sinusoidal flow patter (fig. 2) is open to the atmosphere at all times, with a negligible resistance to both ipiratory and expiratory flow; therefore, spontaneous respiration is unimpeded, even when the RFG is only 2 litre min" 1, which is less than the peak ipiratory flow rate of an infant. The rotating jet could also be applied directly to the airway, in which case the driving gas would become the respiratory fresh gas; however, this configuration would increase the apparatus deadspace because of the connecting tubes. In this study the frequency of oscillatio was maintained at 300 and 420 b.p.m. Rossing and colleagues [7] demotrated that, at frequencies greater than 300 b.p.m., no further improvement in elimination of carbon dioxide could be achieved, even with a cotant tidal volume. The present study confirmed this observation. A desirable feature of this oscillatory ventilator is the absence of resistance during expiration, thereby preventing uncontrolled PEEP and gas trapping in the lungs. During HFPPV at a similar frequency of ventilation, 0.2 kpa PEEP was observed. A trapped gas volume of 2 litre has been observed in man after application of an oscillatory ventilator at frequencies less than 10 Hz for only 30 s [10]. If PEEP is required in this new system, it may be applied by using the second fixed jet. Although the blood-gas teio during HFPPV and IPPV were very similar to HFOV at 300 b.p.m., the mean airway pressures were significantly lower during HFOV, which has been shown to be beneficial to infants with persistent pulmonary hyperteion [6]. A multicentre trial [11] of oscillatory ventilation in critically ill pre-term infants has not shown any benefit compared with conventional positive pressure ventilation. However, some of the earlier studies [1-6], which led to the multicentre trial, were more encouraging. The differences between the studies may be explained by lack of uniformity of lung pathology, the ventilation frequency used and the characteristics of the ventilators used. This new jet driven oscillatory ventilation
5 378 BRITISH JOURNAL OF ANAESTHESIA system using a relatively low respiratory fresh gas flow upstream in a wide bore breathing tube free of restrictio and valves prevents an increase in airway pressures. The preliminary results suggest that it may overcome some problems of earlier ventilators and is worthy of further investigation. REFERENCES 1. Bohn DJ, Miyasaka K, Marchak BE, Thompson WK, Froese AB, Bryan AC. Ventilation by high-frequency oscillation. Journal of Applied Physiology1980; 48: Frantz ID in, Werthammer J, Stark AR. High frequency ventilation in premature infants with lung disease: adequate gas exchange at low trachcal pressure. Pediatrics 1983; 71: 483-^ Marchak BE, Thompson WK, Duffty P, Miyaki T, Bryan MH, Bryan AC, Froese AB. Treatment of RDS by high frequency oscillatory ventilation: A preliminary report. Journal of Pediatrics 1981;99: Boynton BR, Mannino FL, Davis RF, Koptic RJ, Friederichsen G. Combined high frequency oscillatory ventilation and intermittent mandatory ventilation in critically ill neonatcs. Journal of Pediatrics1984; 105: Froese AB, Butler PO, Fletcher WA, Byford LJ. Highfrequency oscillatory ventilation in premature infants with respiratory failure: a preliminary report. Anesthesiaand Analgesia 1987; 66: Kohelet D, Perlman M, Kripalani I, Hanna G, Koren G. High-frequency oscillation in the rescue of infants with persistent pulmonary hyperteion. Critical Care Medicine 1988; 16: Rossing TH, Slutsky AS, Lehr JL, Drinker PA, Kamm R, Drazen JM. Tidal volume and frequency dependence of carbon dioxide elimination by high frequency ventilation. New England Journal of Mediane 1981; 305: Chakrabart i MK, Whitwam JG. A new infant ventilator for normal and high-frequency ventilation: influence of tracheal tube on distal airway pressure during high frequency ventilation. Critical Care Medicine1988; 16: Chan KN, Chakrabart i MK, Whitwam JG, Silverman M. Assessmen t of new valveles infant ventilator. Archivesof Diseasesof Childhood 1988; 63: Saari AF, Rossing TH, Solway J, Drazen JM. Lung inflation during high frequency ventilation. American Reviewof RespiratoryDiseases1984; 129: The HIFI Study Group. High-frequency oscillatory ventilation compared with conventional mechanical ventilation in the treatment of respiratory failure in preterm infants. New England Journal of Medicine1989; 320:
http://www.priory.com/cmol/hfov.htm INTRODUCTION The vast majority of patients who are admitted to an Intensive Care Unit (ICU) will need artificial ventilation (Jones et al 1998). The usual means through
More informationHow does HFOV work? John F Mills MBBS, FRACP, M Med Sc, PhD Neonatologist Royal Children s Hospital. Synopsis
How does HFOV work? John F Mills MBBS, FRACP, M Med Sc, PhD Neonatologist Royal Children s Hospital Synopsis Definition of an oscillator Historical perspective Differences between HFOV and CMV Determinants
More informationCARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS
Brit. J. Anaesth. (1956), 28, 196 CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS BY RUSSELL M. DAVIES, I. R. VERNER Queen Victoria Hospital, East Grinstead AND A. BRACKEN Research and Development Centre,
More informationTest Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 6th Edition by Cairo
Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 6th Edition by Cairo Link full download: http://testbankair.com/download/test-bank-for-pilbeams-mechanicalventilation-physiological-and-clinical-applications-6th-edition-by-cairo/
More information3100A Competency Exam
NAME DATE (Circle the appropriate answer) 3100A Competency Exam 1. Of the following, which best describes the mechanics of ventilation used by the 3100A? a. Active inspiration with passive exhalation b.
More informationPART EIGHT HIGH FREQUENCY PERCUSSIVE OSCILLATION (HFPOV )
PART EIGHT HIGH FREQUENCY PERCUSSIVE OSCILLATION (HFPOV ) Note: For maximal comparative understanding, FIRST read PART SEVEN which defines the concept of High Frequency Oscillatory Ventilation (HFOV).
More informationSLE5000 Infant Ventilator with HFO
SLE5000 Infant Ventilator with HFO When the smallest thing matters SLE5000 - The Total Solution for Infant Ventilation Shown on optional roll stand. Ventilator may be mounted either way round on stand.
More informationEXCESSIVE WORK OF BREATHING DURING INTERMITTENT MANDATORY VENTILATION
Br. J. Anaesth. (1986), 58, 1048-1054 EXCESSIVE WORK OF BREATHING DURING INTERMITTENT MANDATORY VENTILATION J. S. MECKLENBURGH, I. P. LATTO, T. A. A. AL-OBAIDI, E. A. SWAI AND W. W. MAPLESON Ventilators
More informationChapter 4: Ventilation Test Bank MULTIPLE CHOICE
Instant download and all chapters Test Bank Respiratory Care Anatomy and Physiology Foundations for Clinical Practice 3rd Edition Will Beachey https://testbanklab.com/download/test-bank-respiratory-care-anatomy-physiologyfoundations-clinical-practice-3rd-edition-will-beachey/
More informationHONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS
RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS 1. INTRODUCTION An anaesthesia delivery system includes any machine, equipment or apparatus which supplies gases, vapours, local anaesthesia and/or
More informationCARBON DIOXIDE CLEARANCE DURING HIGH FREQUENCY JET VENTILATION
Br. J. Anaesth. (86), 58, 404-43 CARBON DIOXIDE CLEARANCE DURING HIGH FREQUENCY JET VENTILATION Effect of Deadspace in a Lung Model A. J. MORTIMER, J.-L. BOURGAIN, J. UPPINGTON AND M. K. SYKES During high
More informationSLE4000. Infant Ventilator with touch-screen operation. When the smallest thing matters
SLE4000 Infant Ventilator with touch-screen operation When the smallest thing matters SLE4000 - The Total Solution for Conventional Infant Ventilation SLE is a world leader in the design and manufacture
More informationRESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE
Course n : Course 3 Title: RESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE Sub-category: Intensive Care for Respiratory Distress Topic: Pulmonary Function and
More informationPART SEVEN THE HISTORY AND APPLICATION OF HIGH FREQUENCY OSCILLATORY VENTILATION (HFOV)
PART SEVEN THE HISTORY AND APPLICATION OF HIGH FREQUENCY OSCILLATORY VENTILATION (HFOV) Reciprocating pistons with an eccentric travel speed, moving to and fro within a cylinder (with a common inlet/outlet),
More informationIntroduction to Conventional Ventilation
Introduction to Conventional Ventilation Dr Julian Eason Consultant Neonatologist Derriford Hospital Mechanics Inspiration diaphragm lowers and thorax expands Negative intrathoracic/intrapleural pressure
More informationTechnical Data and Specifications
Technical Data and Specifications INTENDED USE Ventilator designed to provide Invasive and Non-invasive ventilation for the critical care management of adult, pediatric and neonate-infant (including premature)
More informationRecommendations on Checking Anaesthesia Delivery Systems
Page 1 of 11 Recommendations on Checking Anaesthesia Delivery Version Effective Date 1 Oct 1992 (reviewed Feb 07, Feb 02) 2 2004 3 Nov 2011 4 Dec 2016 Document No. HKCA T1 v4 Prepared by College Guidelines
More informationHow Ventilators Work. Chapter 3
How Ventilators Work Chapter 3 To care for a ventilator patient, you need to know: The various functions of the ventilator used How the ventilator interacts with the patient How changes in lung condition
More informationPrinciples of Mechanical Ventilation: A Graphics-Based Approach
Principles of Mechanical Ventilation: A Graphics-Based Approach Steven M. Donn, MD, FAAP Professor of Pediatrics Neonatal-Perinatal Medicine C.S. Mott Children s Hospital University of Michigan Health
More informationMechanical Ventilation. Mechanical Ventilation is a Drug!!! is a drug. MV: Indications for use. MV as a Drug: Outline. MV: Indications for use
Mechanical Ventilation is a Drug!!! Mechanical Ventilation is a drug I am an employee of Philips Healthcare Hospital Respiratory Care Group and they help me pay for my kids education Jim Laging, RRT, RCP
More informationPrinciples of mechanical ventilation. Anton van Kaam, MD, PhD Emma Children s Hospital AMC Amsterdam, The Netherlands
Principles of mechanical ventilation Anton van Kaam, MD, PhD Emma Children s Hospital AMC Amsterdam, The Netherlands Disclosure Research grant Chiesi Pharmaceuticals Research grant CareFusion GA: 27 weeks,
More informationMEDICAL EQUIPMENT IV MECHANICAL VENTILATORS. Prof. Yasser Mostafa Kadah
MEDICAL EQUIPMENT IV - 2013 MECHANICAL VENTILATORS Prof. Yasser Mostafa Kadah Mechanical Ventilator A ventilator is a machine, a system of related elements designed to alter, transmit, and direct energy
More informationVT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator)
VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator) VT PLUS HF provides a special mode for evaluating the performance of high frequency ventilators while connected
More informationMechanical Ventilation
Mechanical Ventilation Chapter 4 Mechanical Ventilation Equipment When providing mechanical ventilation for pediatric casualties, it is important to select the appropriately sized bag-valve mask or endotracheal
More informationIntroduction. Respiration. Chapter 10. Objectives. Objectives. The Respiratory System
Introduction Respiration Chapter 10 The Respiratory System Provides a means of gas exchange between the environment and the body Plays a role in the regulation of acidbase balance during exercise Objectives
More informationDisclosures. The Pediatric Challenge. Topics for Discussion. Traditional Anesthesia Machine. Tidal Volume = mls/kg 2/13/14
2/13/14 Disclosures Optimal Ventilation of the Pediatric Patient in the OR Consulting Draeger Medical Jeffrey M. Feldman, MD, MSE Division Chief, General Anesthesia Dept. of Anesthesiology and Critical
More informationCompleted downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo
Completed downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo Link full download: http://testbankcollection.com/download/pilbeams-mechanicalventilation-physiological-and-clinical-applications-5th-edition-test-bank-cairo
More informationVentilators. Dr Simon Walton Consultant Anaesthetist Eastbourne DGH KSS Basic Science Course
Ventilators Dr Simon Walton Consultant Anaesthetist Eastbourne DGH KSS Basic Science Course Objectives Discuss Classification/ terminology Look at Modes of ventilation How some specific ventilators work
More informationRESPIRATORY PHYSIOLOGY RELEVANT TO HFOV
RESPIRATORY PHYSIOLOGY RELEVANT TO Physiology of CMV 1 What is? Ventilation using a relatively high continuous distending pressure at the airway opening, around which an oscillatory wave is generated to
More informationV8800 Ventilator System. Your healthcare, we care.
V8800 Ventilator System Your healthcare, we care. V8800 Ventilator System V8800 is a critical care ventilator for infants, children and adults. It has comprehensive functionality, a user-friendly design,
More informationSLE5000 Neonatal Ventilator with High Frequency Oscillation
SLE5000 Neonatal Ventilator with High Frequency Oscillation The SLE5000 is superb: the best High Frequency Ventilator I ve used When the smallest thing matters SLE5000 - The Total Solution for Infant Ventilation
More informationMechanical ven3la3on. Neonatal Mechanical Ven3la3on. Mechanical ven3la3on. Mechanical ven3la3on. Mechanical ven3la3on 8/25/11. What we need to do"
8/25/11 Mechanical ven3la3on Neonatal Mechanical Ven3la3on Support oxygen delivery, CO2 elimination" Prevent added injury, decrease ongoing injury" Enhance normal development" Mark C Mammel, MD University
More informationThe Crossvent 2i+ 2. Ventilator Concept (brief theory of operation and features)
The Crossvent 2i+ 1. How is this ventilator classified 2. Ventilator Concept (brief theory of operation and features) -Your Two Choices with this Ventilator 3. An overview of the device (an in-service)
More informationBunnell LifePulse HFV Quick Reference Guide # Bunnell Incorporated
Bunnell Incorporated n www.bunl.com n 800-800-4358 (HFJV) n info@bunl.com 436 Lawndale Drive n Salt Lake City, Utah 84115 n intl 801-467-0800 n f 801-467-0867 Bunnell LifePulse HFV Quick Reference Guide
More informationPERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR:
PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: 1. **Identify and name the filters on the 7200ae. 2. **Explain how each filter is sterilized. 3. **Trace the gas flow through the ventilator
More informationFlow meter. bellow vaporizer. APL valve. Scavenging system
Introductory Lecture Series: The Anesthesia Machine PORNSIRI WANNADILOK Objectives Anesthesia Machine Ventilators Scavenging Systems System Checkout 1 Flow meter ventilator bellow vaporizer Corrugated
More informationEMS INTER-FACILITY TRANSPORT WITH MECHANICAL VENTILATOR COURSE OBJECTIVES
GENERAL PROVISIONS: EMS INTER-FACILITY TRANSPORT WITH MECHANICAL VENTILATOR COURSE OBJECTIVES Individuals providing Inter-facility transport with Mechanical Ventilator must have successfully completed
More informationNeonatal Assisted Ventilation. Haresh Modi, M.D. Aspirus Wausau Hospital, Wausau, WI.
Neonatal Assisted Ventilation Haresh Modi, M.D. Aspirus Wausau Hospital, Wausau, WI. History of Assisted Ventilation Negative pressure : Spirophore developed in 1876 with manual device to create negative
More informationAdvanced Ventilator Modes. Shekhar T. Venkataraman M.D. Professor Critical Care Medicine and Pediatrics University of Pittsburgh School of Medicine
Advanced Ventilator Modes Shekhar T. Venkataraman M.D. Shekhar T. Venkataraman M.D. Professor Critical Care Medicine and Pediatrics University of Pittsburgh School of Medicine Advanced modes Pressure-Regulated
More informationHigh Frequency Ventilation. Neil MacIntyre MD Duke University Medical Center Durham NC USA
High Frequency Ventilation Neil MacIntyre MD Duke University Medical Center Durham NC USA High frequency ventilation Concept of ventilator induced lung injury and lung protective ventilatory strategies
More informationEmergency Transport and Ventilation
Emergency Transport and Ventilation When you get a call and the patient is not breathing, are you and your equipment ready? Can you make the difference? Pneupac portable gas powered ventilators (PGPVs)
More informationCapnography in the Veterinary Technician Toolbox. Katie Pinner BS, LVT Bush Advanced Veterinary Imaging Richmond, VA
Capnography in the Veterinary Technician Toolbox Katie Pinner BS, LVT Bush Advanced Veterinary Imaging Richmond, VA What are Respiration and Ventilation? Respiration includes all those chemical and physical
More informationSafe Accurate Reliable. CWM-301 Anesthesia System
Safe Accurate Reliable CWM-301 Anesthesia System CWM-301 Anesthesia System Extraordinary performance points 8.4 inch colour LCD screen Compact Anesthesia ventilation system with the standard of ISO 9001:13485
More informationየIትዮዽያ የደረጃዎች ኤጀንሲ ETHIOPIAN STANDARDS AGENCY ES DRAFT UPLOAD REQUEST FORM FOR PUBLIC COMMENTS
OF/ESA/SDD/04 EHS This form is used to maintain uniformity and consistency in uploading the list of draft documents for the development of Ethiopian standards for public comment. 1. The draft documents
More information(Received 16 January 1946)
186 J. Physiol. (I946) I05, I86-I90 6I2.2I5.9 THE ABSORPTION OF FLUIDS FROM THE LUNGS BY F. C. COURTICE AND P. J. PHIPPS From the Experimental Station, Porton and the Laboratory of Physiology, Oxford (Received
More informationThe Compact, Portable, Efficient and Economical Solution to boost Shop Compressed Air Pressure by 2 or 3 times
Tseries Series AB AB Air Air Boosters Boosters The Compact, Portable, Efficient and Economical Solution to boost Shop Compressed Air Pressure by 2 or 3 times.01/1 MANIFOLD MOUNTED VALVES FOR EASY SERVICING
More informationWhat can we learn from high-frequency ventilation?
What can we learn from high-frequency ventilation? Dipartimento di Medicina Perioperatoria, Terapia Intensiva ed Emergenza Azienda Sanitaria Universitaria Integrata di Trieste Università degli Studi di
More informationHIGH FREQUENCY JET VENTILATION (HFJV): EQUIPMENT PREPRATION
POLICY The physician orders High Frequency Jet Ventilation (HFJV). The Respiratory Therapist in discussion with the physician will determine blood gas targets and ventilation settings for the treatment
More informationA9600 Anesthesia Workstation. Your healthcare, we care.
A9600 Anesthesia Workstation Your healthcare, we care. A9600 is an advanced yet easy to use anesthesia workstation that provides accurate, pneumatically driven and electronically controlled ventilation.
More informationActivity 2: Examining the Effect of Changing Airway Resistance on Respiratory Volumes
1 BGYC34 PhysioEx Lab 7 Respiratory Systems Mechanics Marking Scheme Part 1 Complete PhysioEx lab #7. Hand-in all of the pages associated with the lab. Note that there are 5 activities to be completed.
More informationInitiation and Management of Airway Pressure Release Ventilation (APRV)
Initiation and Management of Airway Pressure Release Ventilation (APRV) Eric Kriner RRT Pulmonary Critical Care Clinical Specialist Pulmonary Services Department Medstar Washington Hospital Center Disclosures
More informationOperation of Oxylog 3000 ventilator
Operation of Oxylog 3000 ventilator Overview Suitable for patients from approximately 7Kg body weight upwards (smallest tidal volume 50mls). Entirely power dependant- will not ventilate if no available
More informationQUICK REFERENCE GUIDE
cm H O 2 cm H O 2 cm HO 2 PSI cm H O 2 ON OFF UPPER LIMIT LOWER LIMIT UPPER LIMIT LOWER LIMIT LIFE PULSE HIGH-FREQUENCY VENTILATOR QUICK REFERENCE GUIDE 01388-08.11 MONITOR PIP JET VALVE ALARMS READY SILENCE
More informationVENTILATORS PURPOSE OBJECTIVES
VENTILATORS PURPOSE To familiarize and acquaint the transfer Paramedic with the skills and knowledge necessary to adequately maintain a ventilator in the interfacility transfer environment. COGNITIVE OBJECTIVES
More informationMechanical Ventilation
PROCEDURE - Page 1 of 5 Purpose Scope Physician's Order Indications Procedure Mechanical Artificial Ventilation refers to any methods to deliver volumes of gas into a patient's lungs over an extended period
More informationRESPIRATORY PHYSIOLOGY, PHYSICS AND
Course n : Course 3 Title: RESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE Sub-category: Techniques Topic: Respiratory monitoring Date: May 05-07, 2016 Language:
More informationHospital and Transport for Controlled Breathing
Hospital and Transport for led Breathing When transporting a critically ill patient you need a ventilator that can go anywhere in any situation. Smiths Medical Pneupac small portable gas powered ventilators
More informationRESISTANCE TO AIRFLOW IN ANAESTHETIC BREATHING SYSTEMS
Br. J. Anaesth. (1989), 62, 456-461 RESISTANCE TO AIRFLOW IN ANAESTHETIC BREATHING SYSTEMS D. G. MARTIN, K. L. KONG AND G. T. R. LEWIS The resistance to airflow of the components of breathing systems have
More informationAespire 7100 Essential performance Compact design
GE Healthcare Aespire 7100 Essential performance Compact design Features Enhanced monitor integration capabilities with our Datex-Ohmeda Anesthesia Monitor and Compact Anesthesia monitor Lightweight and
More informationCOMPARISON OF PORTABLE EMERGENCY VENTILATORS USING A LUNG MODEL
British Journal of Anaesthesia 1993; 70: 2-7 APPARATUS COMPARISON OF PORTABLE EMERGENCY VENTILATORS USING A LUNG MODEL L. ATTEBO, M. BENGTSSON AND A. JOHNSON SUMMARY A lung model was used to test the performance
More informationLung Volumes and Capacities
Lung Volumes and Capacities Normally the volume of air entering the lungs during a single inspiration is approximately equal to the volume leaving on the subsequent expiration and is called the tidal volume.
More informationMedical Instruments in the Developing World
2.2 Ventilators 2.2.1 Clinical Use and Principles of Operation Many patients in an intensive care and the operating room require the mechanical ventilation of their lungs. All thoracic surgery patients,
More informationONLINE DATA SUPPLEMENT. First 24 hours: All patients with ARDS criteria were ventilated during 24 hours with low V T (6-8 ml/kg
APPENDIX 1 Appendix 1. Complete respiratory protocol. First 24 hours: All patients with ARDS criteria were ventilated during 24 hours with low V T (6-8 ml/kg predicted body weight (PBW)) (NEJM 2000; 342
More informationUNDERSTANDING NEONATAL WAVEFORM GRAPHICS. Brandon Kuehne, MBA, RRT-NPS, RPFT Director- Neonatal Respiratory Services
UNDERSTANDING NEONATAL WAVEFORM GRAPHICS Brandon Kuehne, MBA, RRT-NPS, RPFT Director- Neonatal Respiratory Services Disclosures Purpose: To enhance bedside staff s knowledge of ventilation and oxygenation
More informationBreathing Circuits. Product training
Breathing Circuits Product training Agenda Introduction to Breathing Circuits Anaesthesia Circuits Intensive Care Circuits Accessories 2 Covidien Introduction 3 Covidien What is a breathing circuit? In
More informationCHAPTER 3: The respiratory system
CHAPTER 3: The respiratory system Practice questions - text book pages 56-58 1) When the inspiratory muscles contract, which one of the following statements is true? a. the size of the thoracic cavity
More informationQuick Response Pneupac VR1 Emergency Ventilator AIRWAYMANAGEMENT
Quick Response Pneupac VR1 Emergency Ventilator AIRWAYMANAGEMENT Responding to your needs One of the best selling resuscitators... Smiths Medical is bringing you the next generation of emergency ventilators.
More informationNeonatal Ventilators. Touch-screen Neonatal Ventilator with High Frequency Oscillation SLE5000
Neonatal Ventilators Touch-screen Neonatal Ventilator with High Frequency Oscillation SLE5000 SLE5000: Designed for Neonates Neonatal ventilation poses real challenges to neonatologists and caregivers.
More informationUnit 15 Manual Resuscitators
15-1 Unit 15 Manual Resuscitators GOAL On completion of this unit, the student should comprehend the proper operation of self-inflating resuscitation bags, flow-inflating resuscitation bags and gas-powered
More informationFlight Medical presents the F60
Flight Medical presents the F60 Reliable Ventilation Across the Spectrum of Care Adult & Pediatric Pressure/Volume Control Basic/Advanced Modes Invasive/NIV High Pressure/Low Flow O2 Up to 12 hours batteries
More informationMaking Life Easier Easypump II
Making Life Easier Easypump II The elastomeric pump system for short- and long-term infusion therapy Elastomeric Infusion Systems Technical Guide Easypump II Disposable Elastomeric Infusion Pump System
More informationHomecare Pneumology NEONATOLOGY ANAESTHESIA INTENSIVE CARE VENTILATION Step beyond the future
Homecare Pneumology NEONATOLOGY ANAESTHESIA INTENSIVE CARE VENTILATION Step beyond the future Sleep Diagnostics Service Patient Support The future starts now! For over 30 years, Heinen + Löwenstein has
More informationOperating Instructions for Microprocessor Controlled Ventilators
Page 1 of 5 Operating Instructions for Microprocessor Controlled Ventilators Purpose Audience Scope Physician's Order To provide guidelines for the procedure for the use of microprocessor controlled ventilators.
More informationDPA 02 TM Diamedica Portable Anaesthesia System INSTRUCTIONS AND USER GUIDE
DPA 02 TM Diamedica Portable Anaesthesia System INSTRUCTIONS AND USER GUIDE INSTRUCTIONS FOR PORTABLE ANAESTHETIC MACHINE DPA 02 TM The Diamedica Portable Anaesthetic machine DPA 2 TM has three principal
More informationRespiratory Physiology Gaseous Exchange
Respiratory Physiology Gaseous Exchange Session Objectives. What you will cover Basic anatomy of the lung including airways Breathing movements Lung volumes and capacities Compliance and Resistance in
More informationWhat is an Optimal Paw Strategy?
What is an Optimal Paw Strategy? A Physiological Rationale Anastasia Pellicano Neonatologist Royal Children s Hospital, Melbourne Acute injury sequence Barotrauma Volutrauma Atelectotrauma Biotrauma Oxidative
More informationHuman gas exchange. Question Paper. Save My Exams! The Home of Revision. Cambridge International Examinations. 56 minutes. Time Allowed: Score: /46
Human gas exchange Question Paper Level Subject Exam oard Topic Sub Topic ooklet O Level iology ambridge International Examinations Respiration Human gas exchange Question Paper Time llowed: 56 minutes
More informationState-of-the-art anesthesia technology
State-of-the-art anesthesia technology anesthesia system Get closer to the MRI scanner Superb certified performance in close proximity to strong magnetic fields. Our innovative machine provides benefits
More informationBASIC PHYSICS APPLIED TO ANAESTHESIOLOGY
BASIC PHYSICS APPLIED TO ANAESTHESIOLOGY Dr.R.Selvakumar.M.D.D.A.DNB Professor of Anaesthesiology, K.A.P.Viswanatham Govt medical college, Trichy. The current practice of Anaesthesiology demands knowledge
More informationCollin County Community College. Lung Physiology
Collin County Community College BIOL. 2402 Anatomy & Physiology WEEK 9 Respiratory System 1 Lung Physiology Factors affecting Ventillation 1. Airway resistance Flow = Δ P / R Most resistance is encountered
More informationImprovement of gas exchange during high frequency intermittent oscillation in rabbits
Original Contribution Kitasato Med J 2014; 44: 56-68 Improvement of gas exchange during high frequency intermittent oscillation in rabbits Shingo Kasahara, 1 Kagami Miyaji 2 1 Department of Cardiovascular
More informationKey words: intrahospital transport; manual ventilation; patient-triggered ventilation; respiratory failure
Intrahospital Transport of Critically Ill Patients Using Ventilator With Patient- Triggering Function* Toshiaki Nakamura, MD; Yuji Fujino, MD; Akinori Uchiyama, MD; Takashi Mashimo, MD; and Masaji Nishimura,
More informationElectronics Respironics Lifecare PLV-102 / PLV-102 Ventilator To t a l S o l u t i o n This lightweight ventilator is well suited for transport and portable applications. Controlled by a microprocessor,
More informationAUGUST ANESTHESIA MACHINES MEDICAL PRODUCTS ANESTHESIA MACHINES. Quality System Standard applied: ISO 9001, ISO 13485
ANESTHESIA MACHINES AUGUST - 2008 MEDICAL PRODUCTS ANESTHESIA MACHINES Quality System Standard applied: ISO 9001, ISO 13485 CE Certification 93/42/EEC. Arsiconsult 38, Av. Guy de Maupassant 78400 Chatou
More informationVentilator Training Module
Ventilator Training Module LTV (Lap Top Ventilator) Teaching Script LTV User guides: http://www.carefusion.com/documents/guides/user-guides/rc_ltv-1000_ug_en.pdf http://www.carefusion.com/documents/guides/user-guides/rc_ltv-1100_ug_en.pdf
More informationThe physiological functions of respiration and circulation. Mechanics. exercise 7. Respiratory Volumes. Objectives
exercise 7 Respiratory System Mechanics Objectives 1. To explain how the respiratory and circulatory systems work together to enable gas exchange among the lungs, blood, and body tissues 2. To define respiration,
More informationHypoxia Following Rapid Decompression to 18,288 m (60,000 ft) Attributable to Alveolar Hypoventilation
Hypoxia Following Rapid Decompression to 18,288 m (60,000 ft) Attributable to Alveolar Hypoventilation Desmond M Connolly PhD QinetiQ Aircrew Systems Senior Medical Officer Timothy J D Oyly BSc Amanda
More informationRESPIRATORY CARE POLICY AND PROCEDURE MANUAL. a) Persistent hypoxemia despite improved ventilatory pattern and elevated Fl02
The University of Mississippi AND PROCEDURE MANUAL Effective Date: June 30, 1990 Revised Date: December 2009 MANUAL CODE Page 1 of 5 PREPARED BY: Respiratory Care Policy and Procedure Review Committee
More informationPART ONE CHAPTER ONE PRIMARY CONSIDERATIONS RELATING TO THE PHYSIOLOGICAL AND PHYSICAL ASPECTS OF THE MECHANICAL VENTILATION OF THE LUNG
PART ONE CHAPTER ONE PRIMARY CONSIDERATIONS RELATING TO THE PHYSIOLOGICAL AND PHYSICAL ASPECTS OF THE MECHANICAL VENTILATION OF THE LUNG POSSIBLE ORIGIN OF THE MECHANICAL VENTILATION OF THE LUNG- The first
More informationNuffield 200 Ventilator User Instruction Manual
Nuffield 200 Ventilator User Instruction Manual Quality and Assurance in Anaesthesia THE IMPORTANCE OF PATIENT MONITORING WARNING Anaesthetic systems have the capability to deliver mixtures of gases and
More informationDATA SHEET. VENTILATION SERVO-s
DATA SHEET VENTILATION The Gold Standard Critical Care 3 PROVEN SERVO SIMPLICITY, RELIABILITY AND SENSITIVITY MAQUET THE GOLD STANDARD Leading the way: MAQUET is a premier international provider of medical
More informationNeonatal tidal volume targeted ventilation
Neonatal tidal volume targeted ventilation Colin Morley Retired Professor of Neonatal Medicine, Royal Women s Hospital, Melbourne, Australia. Honorary Visiting Fellow, Dept Obstetrics and Gynaecology,
More informationClassification of Mechanical Ventilators
Classification of Mechanical Ventilators Kacmarek s 12 Point Classification Positive/Negative Pressure Powering Mechanism Driving Mechanism Single or Double Circuited Modes of Ventilation Cycling Parameter
More informationSATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products
Saturn Evo SATURN EVO COLOR A N A E S T H E S I A S Y S T E M i n t e l l i g e n t s i m p l i c i t y Manufacturer of high quality anaesthesia products SATURN EVO COLOR PATIENT SAFETY AND USER PATIENT
More informationVirginia Beach EMS. Oxylator EMX. Debra H. Brennaman, RN, MPA, NREMT-P
Virginia Beach EMS Oxylator EMX Debra H. Brennaman, RN, MPA, NREMT-P Oxylator EMX Overview Patient responsive oxygen powered resuscitation / ventilation device intended to provide emergency ventilatory
More informationCHAPTER 3: The cardio-respiratory system
: The cardio-respiratory system Exam style questions - text book pages 44-45 1) Describe the structures involved in gaseous exchange in the lungs and explain how gaseous exchange occurs within this tissue.
More informationSelecting and Connecting Breathing Systems
Selecting and Connecting Breathing Year Group: BVSc3 + Document number: CSL_A03 Equipment for this station: Equipment list: Pen Paper Calculator T-piece (in CSL a strip of white tape is around this system)
More informationLAB 7 HUMAN RESPIRATORY LAB. Complete the charts on pgs. 67 and 68 and read directions for using BIOPAC
66 LAB 7 HUMAN RESPIRATORY LAB Assignments: Due before lab: Quiz: Three Respiratory Interactive Physiology Animations pages 69 73. Complete the charts on pgs. 67 and 68 and read directions for using BIOPAC
More informationAPUS Anesthesia Machines
APUS Anesthesia Machines APUS x1 Anesthesia Machine Breathing circuit applicable for complete range Autoclavable metal and plastic materials. Suitable for adult, pediatric and neonatal patients. Featured
More informationOperation and Maintenance of the EPV200 Portable Ventilator
Operation and Maintenance of the EPV200 Portable Ventilator 1 Applications of the EPV200 The EPV200 Portable Ventilator is a gas powered electronically controlled mechanical ventilator, designed to provide
More information