Name: Oxygen Care Ltd Address: 2 Holfeld Business Park, Kilmacanogue, Co. Wicklow, Ireland URGENT FIELD SAFETY NOTICE - UPDATE

Size: px
Start display at page:

Download "Name: Oxygen Care Ltd Address: 2 Holfeld Business Park, Kilmacanogue, Co. Wicklow, Ireland URGENT FIELD SAFETY NOTICE - UPDATE"

Transcription

1 Carefusion UK Jays Close The Crescent Basingstoke RG22 4BS United Kingdom Name: Oxygen Care Ltd Address: 2 Holfeld Business Park, Kilmacanogue, Co. Wicklow, Ireland URGENT FIELD SAFETY NOTICE - UPDATE Product Name: AVEA ventilators Product Reference: All AVEA models Serial Numbers: All (See attached list in Appendix 2) FSCA Identifier: RES2013-AVEA-01 Date: 7 th October 2013 Type of Action: Field Safety Notice Dear Mr/Mrs I write following my letter dated 18 th September 2013 to inform you that the Field Safety Notice issued by CareFusion has been updated. Please find below the changes highlighted in green. Product Name: AVEA ventilator all models CareFusion has identified a potential risk associated with AVEA ventilators when used at higher altitudes. A proactive complaint review identified an error in the barometric pressure compensation. CareFusion is voluntarily initiating a field correction of the affected devices to correct the error. The published volume accuracy specification of the neonatal hotwire flow sensor is +/- 1ml + 10%. PROBLEM AND AFFECTED DEVICES ISSUE: AVEA ventilators may experience the underreporting of tidal volume if used in conjunction with the neonatal hotwire flow sensor, CareFusion part number This error is due to the barometric pressure sensor compensation. Clinically significant changes in barometric pressure are generally seen at altitudes greater than 5,000 feet above sea level (ASL). While all AVEA ventilators fall under this voluntary Field Safety Corrective Action, hospitals in locations at high elevations (higher than 1524 meters or 5000 feet above sea level) that are using the AVEA ventilator in the neonatal patient care setting are at greatest risk of this error. Single patient use VarFlex sensors are not affected by this corrective action and will compensate for changes to atmospheric barometric pressure. The adult and pediatric patient setting are not affected by this issue.

2 POTENTIAL RISK: In cases where a patient is ventilated at an altitude of equal or greater than 1,524 meters ASL the ventilator may under report the delivery of tidal volume. The degree of which is based upon the altitude at the location of use. For reference: The measurement error at 1,220 meters ASL (approximately 681 mmhg) would be less than 10% and the error at 1,829 meters ASL would be approximately 16-19%. ACTIONS TO BE TAKEN BY CAREFUSION: CareFusion has initiated a project to develop a new version of software that will resolve this issue. ACTION TO BE TAKEN BY THE CUSTOMER CareFusion does not require the return of your devices. Once this software is available it must be installed on all AVEA ventilators. If the location of your AVEA ventilator(s) is above 1,524 meters ASL, and you are using the AVEA in the neonatal patient setting, customers are asked to take one of the following actions. o AVEA Comprehensive: Continue the use of your AVEA Comprehensive ventilator with the use of a VarFlex single patient use flow sensor, CareFusion part number (10 pack). o AVEA Standard: Discontinue the use of your AVEA Standard ventilator in the neonatal patient setting. If the location of your hospital is located at altitude above 1,524 meters ASL a customer support representative will contact you to arrange for a suitable alternative if necessary. If the location of your hospital is below 1,524 meters, or the ventilator is exclusively used in the adult and/or pediatric patient setting, no action is necessary other than to perform the necessary software update when made available. Please promptly return the enclosed response card (Appendix 2) to expedite the correction process and acknowledge receipt of this notification. The US Food and Drug Administration (FDA) has been notified of this action. Please use the chart provided below for questions and support. CareFusion Contact Contact Information Areas of Support CareFusion Technical Recall Related Support Support.CC.EU@carefusion.com Questions CareFusion Business Partner Oxygen Care Ltd 2 Holfeld Business Park, Kilmacanogue, Co. Wicklow, Ireland Recall Related Questions Please promptly return the enclosed Response Card to expedite the correction process and acknowledge receipt of this Notification. We appreciate your cooperation and we recognize the inconvenience this may cause your facility. Thank you for your support on this important matter.

3 Sincerely, Mirela Boureanu CareFusion QRA Manager UK and Ireland Enclosed: List of affected units, customer response card

4 Appendix 1: List of affected part and serial numbers PART NUMBER CONFIGURATION BASE MODEL SHIP DATE TO OXYGEN CARE LTD CLEAN SN PN SN AVEA COMP CONF 10/09/2003 ADV ADV AVEA COMP 27/11/2002 ACV01721 ACECACV AVEA STD 22/04/2003 ADV01605 ASHOADV AVEA STD 22/04/2003 ADV01624 ASHOADV AVEA STD 22/04/2003 ADV01648 ASHOADV AVEA STD 22/04/2003 ADV01653 ASHOADV AVEA STD 22/04/2003 ADV01669 ASHOADV AVEA COMP 23/04/2003 ADV01718 ACECADV AVEA COMP 16/12/2003 ADV ADV AVEA COMP 16/12/2003 ADV ADV AVEA COMP 16/12/2003 ADV ADV06113 R AVEA STD W/COMP CONFIG 20/05/2005 ADV06218 R172ADV AVEA COMP 15/10/2004 AEV AEV AVEA COMP 08/10/2004 AEV AEV AVEA COMP 08/10/2004 AEV AEV AVEA COMP 08/10/2004 AEV AEV AVEA COMP 15/10/2004 AEV AEV AVEA COMP 08/10/2004 AEV AEV AVEA COMP 08/10/2004 AEV AEV AVEA COMP 08/10/2004 AEV AEV AVEA COMP 08/10/2004 AEV AEV AVEA PLUS 08/12/2005 AFV AFV AVEA PLUS 08/12/2005 AFV AFV AVEA PLUS 08/12/2005 AFV AFV AVEA PLUS 08/12/2005 AFV AFV AVEA STD 25/11/2009 AKV AKV AVEA COMP 25/11/2009 AKV AKV AVEA COMP 25/11/2009 AKV AKV AVEA COMP 25/11/2009 AKV AKV AVEA COMP 25/11/2009 AKV AKV AVEA COMP 25/11/2009 AKV AKV AVEA COMP 25/11/2009 AKV AKV AVEA COMP 25/11/2009 AKV AKV AVEA STD 25/11/2009 AKV AKV AVEA STD 25/11/2009 AKV AKV04423 Page 4 of 7

5 AVEA STD 25/11/2009 AKV AKV AVEA STD 25/11/2009 AKV AKV AVEA STD 25/11/2009 AKV AKV AVEA STD 25/11/2009 AKV AKV AVEA STD 25/11/2009 AKV AKV AVEA STD 01/04/2010 BAV BAV AVEA STD 01/04/2010 BAV BAV AVEA STD 01/04/2010 BAV BAV04250 Page 5 of 7

6 Appendix 2 Business Partner Response Card UPDATE - FIELD SAFETY NOTICE AVEA Ventilators Acknowledgment and Verification Form Product Name: Product Reference: FSCA Identifier: AVEA ventilators All AVEA models RES2013-AVEA-01 Name of Business Partner Organization Business Partner Address Name of Business Partner Correspondent Telephone number address Signature Date I have read and understand the contents of this Field Safety Notice and confirm that all affected medical equipment within my area of responsibility has been checked, and Software will be upgraded as required. I have completed the enclosed table to reflect those devices, that cannot be upgraded (not found or scrapped). Page 6 of 7

7 Equipment Country Distribution Table: Model Number Serial Number Affected and will be done Other (not found or scrapped) Note: If more space is needed for the above table please attach a separate document. Please return this form to: Miss Mirela Boureanu Address: Carefusion UK, Jays Close, The Crescent, Basingstoke, RG22 4BS or By mirela.boureanu@carefusion.com (Return of your product is not necessary) For further questions please call CareFusion Technical Support Page 7 of 7

Alaris / Asena GS, GH, CC, TIVA, PK, Enteral Syringe Pump

Alaris / Asena GS, GH, CC, TIVA, PK, Enteral Syringe Pump Name Address Product Name: URGENT FIELD SAFETY NOTICE Alaris / Asena GS, GH, CC, TIVA, PK, Enteral Syringe Pump Product codes with prefix (all variants): 8001, 8002, 8003, 8004, 8005, 8007 FSCA Identifier:

More information

Airox Supportair Ventilator

Airox Supportair Ventilator Airox Supportair Ventilator A versatile hospital ventilator with a titration platform that enables a smooth transition to the patient s home care ventilator Titration Platform The Supportair ventilator

More information

URGENT MEDICAL DEVICE RECALL NOTICE

URGENT MEDICAL DEVICE RECALL NOTICE January 31, 2015 URGENT MEDICAL DEVICE RECALL NOTICE ATTN: Regulatory Affairs Manager, Risk Manager, Hospital Administrator, OxyTOTE Distributor, OxyTOTE Wholesaler, Oxygen Service Provider, Health Care

More information

B. Diagnostic Equipment Information 01 Number of Manometers Used to Measure Home Pressurization Manometer Make. Manometer Serial Number

B. Diagnostic Equipment Information 01 Number of Manometers Used to Measure Home Pressurization Manometer Make. Manometer Serial Number (Page 1 of 3) A. Building Air Leakage General Information 01 Test Procedure Used: 02 Indoor temperature during test (degreef) 03 Outdoor temperature during test (degreef) 04 Blower door location 05 Building

More information

B. Diagnostic Equipment Information 01 Number of Manometers Used to Measure Home Pressurization Manometer Make. Manometer Serial Number

B. Diagnostic Equipment Information 01 Number of Manometers Used to Measure Home Pressurization Manometer Make. Manometer Serial Number STATE OF CALIFORNIA BUILDING LEAKAGE WORKSHEET CEC-CF2R-MCH-24-H (Revised 06/14) CERTIFICATE OF INSTALLATION Building Leakage Worksheet CALIFORNIA ENERGY COMMISSION Project Name: Enforcement Agency: Permit

More information

MasterScreen PFT. Powered by SentrySuite

MasterScreen PFT. Powered by SentrySuite MasterScreen PFT Powered by SentrySuite MasterScreen PFT powered by SentrySuite, is an effective combination of robust hardware technology and innovative software platform. A complete, stabile and durable

More information

Urgent Medical Device Recall Medtronic StrataMR Adjustable Valves & Shunts

Urgent Medical Device Recall Medtronic StrataMR Adjustable Valves & Shunts Urgent Medical Device Recall Medtronic StrataMR Adjustable Valves & Shunts 125 Cremona Drive Goleta, CA 93117 Catalog Number 42955, STRATAMR VALVE, SMALL E11656, E12556, E12753, E13482, E13483, E13484,

More information

URGENT Medical Device Recall

URGENT Medical Device Recall PH18022R/FSN86100191A Product: Date: August 16, 2018 Dear HeartStart HS1/OnSite/Home AED Owner, Our records show that you purchased one or more Philips HeartStart Infant/Child SMART Pads Cartridges (M5072A)

More information

Kimal plc: URGENT FIELD SAFETY NOTICE Field Safety Corrective Action IMMEDIATE ACTION REQUIRED

Kimal plc: URGENT FIELD SAFETY NOTICE Field Safety Corrective Action IMMEDIATE ACTION REQUIRED (Customer Name) (Address Line 1) (Address Line 2) (Address Line 3) (Address Line 4) (Post Code) Kimal plc: URGENT FIELD SAFETY NOTICE Field Safety Corrective Action IMMEDIATE ACTION REQUIRED Adult Altius

More information

Vivo Ventilator Manual READ ONLINE

Vivo Ventilator Manual READ ONLINE Vivo Ventilator Manual READ ONLINE If you are searched for the book Vivo ventilator manual in pdf form, then you have come on to the faithful site. We presented utter release of this book in epub, DjVu,

More information

March 2018 FSN Medical Device Recall/Notification HeartStart FRx, HeartStart Home, and Heartstart OnSite AEDs

March 2018 FSN Medical Device Recall/Notification HeartStart FRx, HeartStart Home, and Heartstart OnSite AEDs March 2018 FSN86100186 Medical Device Recall/Notification HeartStart FRx, HeartStart Home, and Heartstart OnSite AEDs Dear HeartStart AED Owner, We are contacting you because our records show you are the

More information

Blease900 Ventilator PATIENT CENTERED VENTILATION

Blease900 Ventilator PATIENT CENTERED VENTILATION Blease900 Ventilator PATIENT CENTERED VENTILATIN PATIENT CENTERED VENTILATIN From the first Blease/Manley ventilator sixty years ago, Spacelabs Healthcare has built a heritage of innovative and intuitive

More information

CDI Blood Parameter Monitoring System 500 A New Tool for the Clinical Perfusionist

CDI Blood Parameter Monitoring System 500 A New Tool for the Clinical Perfusionist Original Article Blood Parameter Monitoring System 500 A New Tool for the Clinical Perfusionist David W. Fried, MS Ed, CCP; Joseph J. Leo, BS, CCP; Gabriel J. Mattioni, BS, CCP; Hasratt Mohamed, CCP; Raymond

More information

Urgent Field Safety Notice (Field Safety Corrective Action) for CORAIL AMT Neck Trials Surgical Instruments

Urgent Field Safety Notice (Field Safety Corrective Action) for CORAIL AMT Neck Trials Surgical Instruments Urgent Field Safety Notice (Field Safety Corrective Action) for CORAIL AMT Neck Trials Surgical Instruments Product Name CORAIL AMT NECK SEG 125D STD CORAIL AMT NECK SEG 125D KLA CORAIL AMT NECK SEG 135D

More information

Rad-57 Pulse CO-Oximeter

Rad-57 Pulse CO-Oximeter Pulse co-oximeter Rad-57 Pulse CO-Oximeter Helps detect methemoglobinemia and carbon monoxide poisioning on-site in seconds PeRFoRMance Specifications measurement range Oxygen Saturation (SpO2)........................................

More information

SLE4000. Infant Ventilator with touch-screen operation. When the smallest thing matters

SLE4000. Infant Ventilator with touch-screen operation. When the smallest thing matters SLE4000 Infant Ventilator with touch-screen operation When the smallest thing matters SLE4000 - The Total Solution for Conventional Infant Ventilation SLE is a world leader in the design and manufacture

More information

RESPIRATORY CARE POLICY AND PROCEDURE MANUAL. a) Persistent hypoxemia despite improved ventilatory pattern and elevated Fl02

RESPIRATORY CARE POLICY AND PROCEDURE MANUAL. a) Persistent hypoxemia despite improved ventilatory pattern and elevated Fl02 The University of Mississippi AND PROCEDURE MANUAL Effective Date: June 30, 1990 Revised Date: December 2009 MANUAL CODE Page 1 of 5 PREPARED BY: Respiratory Care Policy and Procedure Review Committee

More information

Technical Data and Specifications

Technical Data and Specifications Technical Data and Specifications INTENDED USE Ventilator designed to provide Invasive and Non-invasive ventilation for the critical care management of adult, pediatric and neonate-infant (including premature)

More information

Milestones of Aeonmed Ventilators

Milestones of Aeonmed Ventilators Aeonmed Development History Milestones of Aeonmed s 2003 More than 1000 Aeonmed ventilators contributed to save lives during SARS 2008 Aeonmed nominated emergency ventilator supplier for 2008 Beijing Olympic

More information

BODYGUARD 2CH TWIN CHANNEL INFUSION PUMP

BODYGUARD 2CH TWIN CHANNEL INFUSION PUMP BODYGUARD 2CH TWIN CHANNEL INFUSION PUMP The BodyGuard 2CH Twin Channel Infusion Pump is an ambulatory multi-therapy pump suitable for hospital and homecare environments with five infusion programs (continuous,

More information

* William A. Sala, Jr. John A. Roscher. Columbia Gas Transmission, LLC * Sorana Linder 700 Louisiana Street, Suite 700 Director, Regulated Services

* William A. Sala, Jr. John A. Roscher. Columbia Gas Transmission, LLC * Sorana Linder 700 Louisiana Street, Suite 700 Director, Regulated Services December 1, 2016 Ms. Kimberly D. Bose, Secretary Federal Energy Regulatory Commission 888 First Street, NE Washington, DC 20426 Columbia Gas Transmission, LLC 700 Louisiana Street, Suite 700 Houston, TX

More information

1811G/H PITOT-STATIC TEST SET

1811G/H PITOT-STATIC TEST SET 1811G/H PITOT-STATIC TEST SET USER INSTRUCTION MANUAL M/N: 1811G/H, P/Ns: 101-00165 101-00168 101-00169 Doc. P/N: 56-101-00165_00168_00169 Revision C December 9, 2014 BARFIELD, INC. Corporate Headquarters

More information

SLE5000 Infant Ventilator with HFO

SLE5000 Infant Ventilator with HFO SLE5000 Infant Ventilator with HFO When the smallest thing matters SLE5000 - The Total Solution for Infant Ventilation Shown on optional roll stand. Ventilator may be mounted either way round on stand.

More information

The centerpiece of a comprehensive anaesthesia workstation

The centerpiece of a comprehensive anaesthesia workstation The centerpiece of a comprehensive anaesthesia workstation IntelliSave AX700 anaesthesia machine A slim design with rich functionality The IntelliSave AX700 is a comprehensive anaesthesia machine from

More information

fabian HFO: Modular, Powerful, Versatile.

fabian HFO: Modular, Powerful, Versatile. fabian HFO: Modular, Powerful, Versatile. Introducing the flagship of the ACUTRONIC FABIAN FAMILY. The FABIAN HFO, a ventilator developed specifically for the NICU and PICU needs, packed with many great

More information

PROTOCOL Oxygen Delivery Testing of the POM Panoramic Oxygen Mask Under Simulated Patient Conditions. April 8, 2014

PROTOCOL Oxygen Delivery Testing of the POM Panoramic Oxygen Mask Under Simulated Patient Conditions. April 8, 2014 PROTOCOL 14-0022 Mask Under Simulated Patient Conditions April 8, 2014 Dave Piper, PE Piper Medical Products 4807 El Camino Ave, Suite C Carmichael, CA 95608 Tel 916-483-3687 Dave@PiperMedical.com www.pipermedical.com

More information

INSTRUCTION MANUAL. 840 Transmitter. Loop Powered Dissolved Oxygen Transmitter

INSTRUCTION MANUAL. 840 Transmitter. Loop Powered Dissolved Oxygen Transmitter INSTRUCTION MANUAL Loop Powered Dissolved Oxygen Transmitter Contents Contents...2 Introduction...3 Range and Display Settings...4 Installation...5 Electrical Installation...6 Calibration...7 EasyCal...8

More information

NMISA NEW GAS FLOW CALIBRATION FACILITY. Deona Jonker 7 October 2013

NMISA NEW GAS FLOW CALIBRATION FACILITY. Deona Jonker 7 October 2013 NMISA NEW GAS FLOW CALIBRATION FACILITY Deona Jonker 7 October 2013 Introduction Flow the total volume of a fluid that flows past a fixed point in a given time Flow measurements used in industry as a means

More information

CRICKET AUSTRALIA. Schedule for the Cricket Australia 9 to Thrive Business Card Draw Promotion

CRICKET AUSTRALIA. Schedule for the Cricket Australia 9 to Thrive Business Card Draw Promotion CRICKET AUSTRALIA Schedule for the Cricket Australia 9 to Thrive Business Card Draw Promotion Promotion Name Cricket Australia 9 to Thrive Business Card Draw Promoter Cricket Australia ABN 53 006 089 130

More information

IMPORTANT SAFETY RECALL NOTICE. Voluntary Recall (Field Retrofit) Program for Scotsman Commercial Cubers

IMPORTANT SAFETY RECALL NOTICE. Voluntary Recall (Field Retrofit) Program for Scotsman Commercial Cubers IMPORTANT SAFETY RECALL NOTICE February 25, 2010 Re: Voluntary Recall (Field Retrofit) Program for Scotsman Commercial Cubers Dear Scotsman Ice Machine Owner: We understand from product distribution records

More information

D Transport ventilation without compromise DRÄGER OXYLOG 3000 PLUS

D Transport ventilation without compromise DRÄGER OXYLOG 3000 PLUS D-33257-2011 Transport ventilation without compromise DRÄGER OXYLOG 3000 PLUS 02 Rising up to the challenge D-33317-2011 D-33494-2011 D-33299-2011 ICU ventilation standard at all times Whether you re in

More information

Study Management SM STANDARD OPERATING PROCEDURE FOR Interactions with the Institutional Review Board (IRB)

Study Management SM STANDARD OPERATING PROCEDURE FOR Interactions with the Institutional Review Board (IRB) Study Management SM 302.00 STANDARD OPERATING PROCEDURE FOR Interactions with the Institutional Review Board (IRB) Approval: Nancy Paris, MS, FACHE President and CEO 24 May 2017 (Signature and Date) Approval:

More information

7.0 DEVIATION and EXCEPTION of a PREVIOUSLY APPROVED PROTOCOL

7.0 DEVIATION and EXCEPTION of a PREVIOUSLY APPROVED PROTOCOL 7.0 DEVIATION and EXCEPTION of a PREVIOUSLY APPROVED PROTOCOL 7.1 OBJECTIVE To describe the policies and procedures for reviewing a modification or a deviation/exception to a previously approved protocol.

More information

BleaseDatum Anesthesia Vaporizers PEACE OF MIND. Serving The HealthCare Industry

BleaseDatum Anesthesia Vaporizers PEACE OF MIND. Serving The HealthCare Industry BleaseDatum Anesthesia Vaporizers PEACE OF MIND Serving The HealthCare Industry PEACE OF MIND While ventilation ensures that patients receive adequate oxygenation, reliable vaporization of inhaled anesthesia

More information

Via Federal Express Rockville, MD 20850

Via Federal Express Rockville, MD 20850 DEPARTMENT OF HEAKTH & HUMAN SERVICES JUL 8 -- WARNING LETTER Food and Drug Administration Center for Devices and Radiological Health 2098 Gaither Road Via Federal Express Rockville, MD 20850 Gordon Binder

More information

United Kingdom of Great Britain and Northern Ireland

United Kingdom of Great Britain and Northern Ireland (UK/0126/0074) MI-005 United Kingdom of Great Britain and Northern Ireland Certificate of EC type-examination of a measuring instrument Number: UK/0126/0074 issued by the Secretary of State for Business,

More information

Patient Setup Information

Patient Setup Information VOCSN Out of Box Tips For temporary use before entering patient settings, VOCSN can be configured as follows to help ease the setup process: 1 Remove VOCSN from its shipping packaging. 2 Attach the circuit.

More information

ali-q TM Gravimetric Verification Procedure

ali-q TM Gravimetric Verification Procedure ali-q TM Gravimetric Verification Procedure Introduction This document describes the recommended procedure for verifying performance of the ali-q aliquoting pipet controller using the industry standard-

More information

TECHNICAL BULLETIN. TriControls. Control Materials for Liquid QC and Calibration Verification

TECHNICAL BULLETIN. TriControls. Control Materials for Liquid QC and Calibration Verification i-stat TECHNICAL BULLETIN Tris Materials for Liquid QC and OVERVIEW As part of the READi initiative (Responds, Enhances, And Delivers innovations), Abbott Point of Care (APOC) has released a new set of

More information

Evita 2 dura. First-class ventilation in your patient s best interests. Because you care. Emergency Care OR/Anesthesia Critical Care Home Care

Evita 2 dura. First-class ventilation in your patient s best interests. Because you care. Emergency Care OR/Anesthesia Critical Care Home Care M E D I C A L Evita 2 dura First-class ventilation in your patient s best interests Emergency Care OR/Anesthesia Critical Care Home Care Because you care A workhorse in everyday clinical routine For more

More information

To assure knowledge and compliance by documenting the annual administrative review and continuing IRB review of non-exempt projects for the IRB.

To assure knowledge and compliance by documenting the annual administrative review and continuing IRB review of non-exempt projects for the IRB. Institutional Review Board.... University of Missouri-Columbia.. Standard Operating Procedure Annual Review of Research Annual Review of Research Effective Date: January 21, 2019 Original Approval Date:

More information

Warranty The device shall have a 6-year warranty at minimum

Warranty The device shall have a 6-year warranty at minimum Bid Specifications Defibrillator The AED must have a high-resolution liquid crystal display with capacitive touch panel. The AED must have an ON/OFF button. The AED must have a SHOCK button that illuminates

More information

VENTILATION SERVO-i INSPIRATION IN EVERY BREATH

VENTILATION SERVO-i INSPIRATION IN EVERY BREATH VENTILATION SERVO-i INSPIRATION IN EVERY BREATH The Gold Standard Critical Care SERVO-i 3 PROVIDING THE HIGHEST LEVEL OF CLINICAL CARE MAQUET THE GOLD STANDARD Leading the way: MAQUET is a premier international

More information

SEP-1 Additional Notes for Abstraction for Version 5.0b

SEP-1 Additional Notes for Abstraction for Version 5.0b SEP-1 Additional Notes for Abstraction for Version 5.0b Data Element Administrative Contraindication to Care Blood Culture Collection Date Blood Culture Collection Time Broad Spectrum or Other Antibiotic

More information

CRICKET AUSTRALIA. Schedule for the ASICS Superstore Promotion

CRICKET AUSTRALIA. Schedule for the ASICS Superstore Promotion CRICKET AUSTRALIA Schedule for the ASICS Superstore Promotion Promotion Name ASICS Superstore Promotion Promoter Cricket Australia ABN 53 006 089 130 of 60 Jolimont Street, Jolimont, VIC, 3002. Permit

More information

DNA-OC GENOMIC BREEDING VALUE CONTRACT

DNA-OC GENOMIC BREEDING VALUE CONTRACT DNA-OC GENOMIC BREEDING VALUE CONTRACT The undersigned assigns the Warmblood Studbook of Ireland (WSI) to engage the KWPN to perform a genomic profile, conduct the DNA-OC test, and estimate the OC breeding

More information

Transport ventilation without compromise DRÄGER OXYLOG 3000 PLUS

Transport ventilation without compromise DRÄGER OXYLOG 3000 PLUS D-33257-2011 Transport ventilation without compromise DRÄGER OXYLOG 3000 PLUS 2 Rising up to the challenge D-33317-2011 D-33494-2011 D-33299-2011 ICU ventilation standard at all times Whether you re in

More information

VT900. Gas Flow Analyzer. Technical data. Key benefits and features:

VT900. Gas Flow Analyzer. Technical data. Key benefits and features: VT900 Gas Flow Analyzer Technical data The Fluke Biomedical VT900 is designed to accurately and reliably test all types of medical gas flow equipment ventilators, insufflators, oxygen meters especially

More information

Optical Dissolved Oxygen Meter

Optical Dissolved Oxygen Meter Optical Dissolved Oxygen Meter HI764113 Rugged Optical Dissolved Oxygen Probe for Fresh and Saltwater Applications Digital, weighted probe No membranes No electrolytes No oxygen consumption No flow dependence

More information

Push-To-Set Intermittent Suction Unit (PTS-ISU) Instructions for Use

Push-To-Set Intermittent Suction Unit (PTS-ISU) Instructions for Use Push-To-Set Intermittent Suction Unit (PTS-ISU) Instructions for Use F H M L F H M L MEDICAL VACUUM Analog Digital 8700-0000-000 (Rev.7) 05/2010 Regulator Identification Push-To-Set TM Intermittent Suction

More information

D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA

D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA Software version 2.6 Table of contents 1 Introduction...3 1.1 Optical Sensor Technology...3 1.2 Installation...4 1.3 Operation...5

More information

Effective Date: January 16, 2012 Policy Number: MHC_RP0107. Revised Date: November 2, 2015 Oversight Level: Corporate

Effective Date: January 16, 2012 Policy Number: MHC_RP0107. Revised Date: November 2, 2015 Oversight Level: Corporate Policy Title: Initial Review of Human Subject Research Effective Date: January 16, 2012 Policy Number: Review Date: November 20, 2015 Section: Revised Date: November 2, 2015 Oversight Level: Corporate

More information

SOP 801: Investigator Qualifications and Responsibilities

SOP 801: Investigator Qualifications and Responsibilities 1. POLICY University of Oklahoma : Investigator Qualifications and Responsibilities The purpose of this policy is to outline the qualifications and responsibilities of the principal investigator and key

More information

Narkomed MRI-2. A perfect fit for the MRI environment. Because you care. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care

Narkomed MRI-2. A perfect fit for the MRI environment. Because you care. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care Narkomed MRI-2 A perfect fit for the MRI environment Emergency Care Perioperative Care Critical Care Perinatal Care Home Care Because you care The complete anesthesia solution for the MRI environment A

More information

Standards and guidelines for care and management of patients requiring oxygen therapy.

Standards and guidelines for care and management of patients requiring oxygen therapy. PURPOSE Standards and guidelines for care and management of patients requiring oxygen therapy. STANDARDS Ongoing management of oxygen therapy requires a prescriber s order. The order must specify oxygen

More information

Oxylog 3000 plus Emergency & Transport Ventilation

Oxylog 3000 plus Emergency & Transport Ventilation Oxylog 3000 plus Emergency & Transport Ventilation Offering high ventilation performance with features such as AutoFlow, integrated capnography and non-invasive Ventilation, the compact and robust Oxylog

More information

The vital hand-held held tool that replaces an entire belt weather kit.

The vital hand-held held tool that replaces an entire belt weather kit. The vital hand-held held tool that replaces an entire belt weather kit. Kestrel Pocket Weather Meters measure the following: Relative Humidity Temperature Wind Speed Wet Bulb Density Altitude Barometric

More information

State-of-the-art anesthesia technology

State-of-the-art anesthesia technology State-of-the-art anesthesia technology anesthesia system Get closer to the MRI scanner Superb certified performance in close proximity to strong magnetic fields. Our innovative machine provides benefits

More information

Astral in AirView: Improving patient care through connectivity. ResMed.com

Astral in AirView: Improving patient care through connectivity. ResMed.com Astral in AirView: Improving patient care through connectivity ResMed.com This guide will assist you with: Setting up the ResMed Connectivity Module for Astral 2 Troubleshooting the ResMed Connectivity

More information

VT650. Gas Flow Analyzer. Technical data. Key benefits and features:

VT650. Gas Flow Analyzer. Technical data. Key benefits and features: VT650 Gas Flow Analyzer Technical data Accurately test gas flow equipment including ventilators with the all-in-one, portable Fluke Biomedical VT650 Gas Flow Analyzer. Accurate The VT650 Gas Flow Analyzer

More information

Corporate Overview and Product Summary

Corporate Overview and Product Summary 2008 Corporate Overview and Product Summary Vapotherm Summary: Market leader and originator of High Flow Therapy Founded in 1999 Headquartered in Stevensville, MD Clinical and education leadership in High

More information

Expanding versatility. The upgraded Trilogy family of ventilators continues to meet the changing needs of your patients

Expanding versatility. The upgraded Trilogy family of ventilators continues to meet the changing needs of your patients Expanding versatility The upgraded Trilogy family of ventilators continues to meet the changing needs of your patients Trilogy Series is a family of remarkably versatile life support ventilators that simplifies

More information

NMISA New Gas Flow Calibration Facility

NMISA New Gas Flow Calibration Facility NMISA New Gas Flow Calibration Facility Speaker / Author: D. Jonker Co-author: E.P. Tarnow National Metrology Institute of South Africa (NMISA) Private Bag X34, Lynnwood Ridge, Pretoria, 0040, South Africa

More information

Misuse of Combustible Gas Meters

Misuse of Combustible Gas Meters Misuse of Combustible Gas Meters An incident investigation raised concerns that users of combustible gas meters may be greatly underestimating fire and explosion hazards. Errors result when meters calibrated

More information

VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator)

VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator) VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator) VT PLUS HF provides a special mode for evaluating the performance of high frequency ventilators while connected

More information

Solaris Nitrogen. Installation Guide. Solaris Nitrogen Installation Guide Rev 1 PN IG /09/16

Solaris Nitrogen. Installation Guide. Solaris Nitrogen Installation Guide Rev 1 PN IG /09/16 Solaris Nitrogen Installation Guide Solaris Nitrogen Installation Guide Rev 1 PN IG-0014 01/09/16 Change History Rev Comment Name Date How to use this Manual This installation guide is intended to be used

More information

Standard Operating Procedure Measuring & Testing Equipment

Standard Operating Procedure Measuring & Testing Equipment Standard Operating Procedure Measuring & Testing Equipment Purpose: 7.6 Responsibility: 7.6d Scope To have a procedure to ensure that all is controlled, maintained and calibrated to ensure that Product

More information

MicroTim XB. User Manual. Precision Digital Barometric Altimeter / Barometer / VSI. Document Revision 1.0 Firmware Version 3.0

MicroTim XB. User Manual. Precision Digital Barometric Altimeter / Barometer / VSI. Document Revision 1.0 Firmware Version 3.0 MicroTim XB Precision Digital Barometric Altimeter / Barometer / VSI User Manual Document Revision 1.0 Firmware Version 3.0 Table of Contents Table of Contents...2 1 General Operation...5 1.1 Altitude

More information

NCCP position paper on the use of Tall Man Lettering

NCCP position paper on the use of Tall Man Lettering NCCP position paper on the use of Tall Man Lettering Version Date published Amendment Approved By 1 August 2017 Version 1 NCCP All comments and feedback are welcome at oncologydrugs@cancercontrol.ie 2

More information

GNX 20/21 Owner s Manual

GNX 20/21 Owner s Manual GNX 20/21 Owner s Manual February 2014 190-01659-00_0B Printed in Taiwan All rights reserved. Under the copyright laws, this manual may not be copied, in whole or in part, without the written consent of

More information

GE Healthcare. Aestiva/5. More than superior ventilation

GE Healthcare. Aestiva/5. More than superior ventilation GE Healthcare Aestiva/5 More than superior ventilation A balance of superior ventilation and award-winning design SmartVent the shape of superior ventilation At the heart of the Datex-Ohmeda Aestiva /5

More information

D-Opto. Dissolved Oxygen Sensor. Operation Manual (Software Version 1.0)

D-Opto. Dissolved Oxygen Sensor. Operation Manual (Software Version 1.0) D-Opto Dissolved Oxygen Sensor Operation Manual (Software Version 1.0) Table of contents 1 Introduction...3 1.1 Optical Sensor Technology...3 2 Software...4 2.1 Installation...4 2.2 Operation...4 3 Operation...9

More information

Astral in AirView: Improving patient care through connectivity. ResMed.com

Astral in AirView: Improving patient care through connectivity. ResMed.com Astral in AirView: Improving patient care through connectivity ResMed.com Using Astral in AirView via the ResMed Connectivity Module (RCM) Astral is ResMed s portable, invasive and non-invasive life support

More information

United Kingdom of Great Britain and Northern Ireland

United Kingdom of Great Britain and Northern Ireland (UK/0126/0026) MI-005 United Kingdom of Great Britain and Northern Ireland Certificate of EC type-examination of a measuring instrument Number: UK/0126/0065 issued by the Secretary of State for Business,

More information

Industrial Hygiene Sampling. Husqvarna Construction Products

Industrial Hygiene Sampling. Husqvarna Construction Products ! Report of: Industrial Hygiene Sampling Husqvarna Construction Products Olathe, KS Palmetto EHS Report No: 2017039 Rev. 1 Submitted Prepared for: Husqvarna Construction Products Prepared By: Palmetto

More information

Continuous Gas Analysis In situ laser gas analyzers TÜV and MCERTS add-on for LDS 6 operating instructions Compact Operating Instructions

Continuous Gas Analysis In situ laser gas analyzers TÜV and MCERTS add-on for LDS 6 operating instructions Compact Operating Instructions Continuous Gas Analysis In situ laser gas analyzers Compact Operating Instructions 1 LDS 6 TÜV/MCERTS certified analyzers for NH3/H2O, NH3 and H2O 1.1 Operation requirements Please observe the following:

More information

Solutions for transcutaneous monitoring

Solutions for transcutaneous monitoring Solutions for transcutaneous monitoring About Radiometer Radiometer is a leading provider of technologically advanced acute care solutions that simplify and automate all phases of acute care testing. By

More information

Panther 5 Acute Care Ventilator

Panther 5 Acute Care Ventilator 1 HIGHLIGHTS High performance and advanced features target the ICU environment Utilizes an internal blower with a specially designed flow control valve removing the need for using compressed air without

More information

CITREX H4 Mobile gas flow analyser. The ideal mobile device for flow and pressure analysing.»

CITREX H4 Mobile gas flow analyser. The ideal mobile device for flow and pressure analysing.» CITREX H4 Mobile gas flow analyser «The ideal mobile device for flow and pressure analysing.» Ventilator tester for mobile use CITREX is designed for mobile and easy use while still meeting all requirements

More information

Diffusing Capacity: 2017 ATS/ERS Standards for single-breath carbon uptake in the lung. Susan Blonshine RRT, RPFT, FAARC, AE-C

Diffusing Capacity: 2017 ATS/ERS Standards for single-breath carbon uptake in the lung. Susan Blonshine RRT, RPFT, FAARC, AE-C Diffusing Capacity: 2017 ATS/ERS Standards for single-breath carbon uptake in the lung Susan Blonshine RRT, RPFT, FAARC, AE-C Joint ATS/ERS Taskforce Recent literature reviewed Surveyed current technical

More information

EMERGENCY USE 03/02/2016

EMERGENCY USE 03/02/2016 DUKE UNIVERSITY HEALTH SYSTEM Human Research Protection Program EMERGENCY USE 03/02/2016 Emergency Exemption for an Investigational Drug or Biologic or Unapproved Device Emergency use of an investigational

More information

PROTOCOL January 25, 2008

PROTOCOL January 25, 2008 PROTOCOL 08-0002 SouthMedic Oxymask and a Non-Rebreather Mask with 80-20 Heliox as Measured at the Mouth January 25, 2008 Dave Piper, PE Piper Medical Products PO Box 993 Carmichael, CA 95609 Tel 916-834-3283

More information

An intuitive and innovative answer to your anaesthesia challenges

An intuitive and innovative answer to your anaesthesia challenges An intuitive and innovative answer to your anaesthesia challenges Dameca AX500 anaesthesia machine Meeting your needs today and tomorrow As an expert in anaesthesia care, you are looking to continuously

More information

Meet the changing needs of your patients. Trilogy Series, with advanced ventilator technology

Meet the changing needs of your patients. Trilogy Series, with advanced ventilator technology Meet the changing needs of your patients Trilogy Series, with advanced ventilator technology Trilogy100 and Trilogy200, unique features for unique patients The Trilogy Series, versatile life support ventilators

More information

UNIVERSITY OF TENNESSEE HEALTH SCIENCE CENTER INSTITUTIONAL REVIEW BOARD CONTINUING REVIEW OF RESEARCH

UNIVERSITY OF TENNESSEE HEALTH SCIENCE CENTER INSTITUTIONAL REVIEW BOARD CONTINUING REVIEW OF RESEARCH UNIVERSITY OF TENNESSEE HEALTH SCIENCE CENTER INSTITUTIONAL REVIEW BOARD CONTINUING REVIEW OF RESEARCH I. PURPOSE This document outlines the University of Tennessee Health Science Center Institutional

More information

GNX 20/21. Owner s Manual

GNX 20/21. Owner s Manual GNX 20/21 Owner s Manual March 2016 190-01659-00_0C All rights reserved. Under the copyright laws, this manual may not be copied, in whole or in part, without the written consent of Garmin. Garmin reserves

More information

Introducing Astral. Greater freedom. Confident care. Designed efficiency

Introducing Astral. Greater freedom. Confident care. Designed efficiency Enrich life Introducing Astral From initial setup to everyday use, the Astral 100 and Astral 150 life support ventilators offer greater freedom, confident care and designed efficiency to enrich life for

More information

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

Bante821 Portable Dissolved Oxygen Meter Instruction Manual Bante821 Portable Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante821 Portable Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante821 portable dissolved oxygen

More information

UNIVERSAL MASK TEST SYSTEM

UNIVERSAL MASK TEST SYSTEM f UNIVERSAL MASK TEST SYSTEM Quotation #110811 Automated Motion Inc. 225 NW Victoria Drive Lee s Summit, Missouri 64086-4709 www.automatedmotion.com SYSTEM OVERVIEW AMI s Universal Mask Test System is

More information

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

Bante820 Portable Dissolved Oxygen Meter Instruction Manual Bante820 Portable Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante820 Portable Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante820 portable dissolved oxygen

More information

DESIGNED FOR PROFESSIONAL RESCUERS

DESIGNED FOR PROFESSIONAL RESCUERS DESIGNED FOR PROFESSIONAL RESCUERS SMART TECHNOLOGY. UNRIVALED SUPPORT. Designed for professional rescuers, the ZOLL AED 3 BLS defibrillator provides in-depth rescue support for both adult and child victims

More information

NEONATAL VOLUME VENTILATION CLINICAL GUIDELINE V2.0

NEONATAL VOLUME VENTILATION CLINICAL GUIDELINE V2.0 NEONATAL VOLUME VENTILATION CLINICAL GUIDELINE V2.0 Summary Commence ventilation Set up ventilator; PTV rate 40, Ti 0.36, TTV 5ml/kg, Trigger sensitivity 0.2, PiP limit 30 mbar Consider SIMV if risk of

More information

Procedure Department: HUMAN RESEARCH PROTECTIONS PROGRAM Policy Number: II.B.1

Procedure Department: HUMAN RESEARCH PROTECTIONS PROGRAM Policy Number: II.B.1 Procedure Department: HUMAN RESEARCH PROTECTIONS PROGRAM Policy Number: II.B.1 Section: HRPP Compliance Review Responsibility: HRPP Policy and Procedure Committee Original Creation Dare: November 1, 2003

More information

Welcome to Flogas. Metered Estates. You would never know you were off the grid. Metered Estates Welcome Pack.indd 1 18/01/ :48

Welcome to Flogas. Metered Estates. You would never know you were off the grid. Metered Estates Welcome Pack.indd 1 18/01/ :48 Welcome to Flogas Metered Estates You would never know you were off the grid. Metered Estates Welcome Pack.indd 1 18/01/2016 16:48 Welcome to Flogas Metered Estates Why Flogas is the right choice A leading

More information

2011 HOUSTON ROCKETS MVP REWARDS SWEEPSTAKES ENTRY FORM. State: Zip: Home Phone:

2011 HOUSTON ROCKETS MVP REWARDS SWEEPSTAKES ENTRY FORM. State: Zip: Home Phone: 2011 HOUSTON ROCKETS MVP REWARDS SWEEPSTAKES ENTRY FORM Name: Age: Address: City: State: Zip: Home Phone: Please indentify for which Drawing this Entry is being submitted (e.g., Week 1, Week 2): Email:

More information

UNDERSTANDING, ACCELERATED

UNDERSTANDING, ACCELERATED UNDERSTANDING, ACCELERATED MEASURE FLOW, PRESSURE, AND TEMPERATURE ALL IN ONE INSTRUMENT! Designed for Performance TSI thermal mass flowmeters incorporate a proprietary platinum film sensor design for

More information

Distribution Connection Forms

Distribution Connection Forms Distribution Connection Forms Form No. DR01 DR02 DR03 DR04 DR05 DR06 DR07 DR08 DR09 DR10 DR11 GDP101 GDP102 GDP105 GDP106 GDP107 GDP108 GDP111 KG-D01 KG-D02 KG-D03 KG-D04 Description Application For Gas

More information

Held under sanction of Pacific Northwest Swimming, Inc., and USA Swimming, Inc. SCHEDULES SESSION 1 SESSIONS 2 & 3 SESSIONS 4 & 5

Held under sanction of Pacific Northwest Swimming, Inc., and USA Swimming, Inc. SCHEDULES SESSION 1 SESSIONS 2 & 3 SESSIONS 4 & 5 SCHEDULES SESSION 1 SESSIONS 2 & 3 SESSIONS 4 & 5 Dates Coaches' Meeting Officials' Meetings Timers Meetings Warm-ups Timed Finals Positive Check-in Deadlines: Friday, May 4, 2018 Saturday, May 5, 2018

More information