Standard Operating Procedure Measuring & Testing Equipment

Size: px
Start display at page:

Download "Standard Operating Procedure Measuring & Testing Equipment"

Transcription

1 Standard Operating Procedure Measuring & Testing Equipment

2 Purpose: 7.6 Responsibility: 7.6d Scope To have a procedure to ensure that all is controlled, maintained and calibrated to ensure that Product conforms to the specified company and regulatory standards. The overall responsibility is with the Technical Department Management and the Inspector, who will safeguard equipment from adjustments that would invalidate any measurement. All in-house Measuring and Testing Equipment used for product release. 1. General c,e c a a,b d a c All Measuring and Testing Equipment used throughout all processing activities will be identified. The description, serial number, location and the intervals for calibration will recorded and documented on the Calibration Record Card MLD Details of all equipment will also be entered on the Calibration Records spreadsheet on the computer system. All calibration equipment will be adequately protected from damage and deterioration during handling, maintenance and storage. All the Measuring and Testing Equipment will be individually identified with a Serial Number and the valid status of calibration. All the Measuring and Testing Equipment will be calibrated on a scheduled basis or prior to use. Equipment may be calibrated prior to use only if there is adequate resources or where the use of Measuring and Testing Equipment is intermittent. The Calibration Record Card will specify if Measuring and Testing Equipment, is to be calibrated in-house by the Inspection Department or by a Sub- Contractor/Test House. Calibration activities will be performed in accordance to a recognised International Standard. It is the responsibility of the Technical Department Management to define the accuracy requirement for the calibration of in-house Testing & Inspection Equipment. The acceptance criteria will be recorded on the Calibration Record Card MLD The Technical Department Management or a Deputy will ensure that all the Measuring and Testing Equipment used has a valid calibration status prior to use. The status of the Measuring and Testing, which includes the calibration validation dates, tolerances and the relative calibration certificate number (where applicable), will be recorded on the Calibration Record Card MLD Calibration labels (from the testing house) will be attached to the equipment as appropriate. 1.8 Equipment that has Failed a calibration test will be placed into quarantine and the relevant details recorded on the Calibration Record spreadsheet b If the equipment can be repaired and the costs not exceed a replacement, this will be sanctioned accordingly. An indication of the fault and corrective action will be requested from the repairer and the item will be recalibrated and certificated If a piece of equipment has Failed calibration and cannot or will not be repaired, it will be permanently withdrawn from use. This will be recorded both on the Calibration Records spreadsheet and on its Calibration Record Card MLD All calibration records and certificates (where applicable) will be retained and filed accordingly.

3 2. In-House Calibration 2.1 It is the responsibility of the Technical Department Management to define the accuracy requirement for the calibration of in-house Testing and Inspection Equipment. The acceptance criteria will be recorded on the Calibration Record Card MLD Measuring and Testing Equipment that requires in-house calibration will be performed recorded as stated in section 1.0. The identification of the person who performs calibration activities will be recorded. 2.3 Measuring and Testing Equipment that is calibrated In House and the calibrating methods are as follows: Ring Type Thread Gauges. These are used for testing external threads on components. The gauge has a clearly marked GO side. Method: Apply GO side to thread. This side must fit fully into the thread with no play to give conformance Plug Type Thread Gauges. This gauge is used for testing internal threads of components. The gauge has two test positions clearly marked GO and NO GO. Method: Screw the GO position into the thread to be tested. The entry MUST be smooth and there MUST be no play. Secondly try to enter the NO GO position. For conformance the GO must go as above and the NO GO must not fit Six-Inch Digital Vernier. This gauge is used to measure internal and external diameters and distances across and inside faces. It can be used on certain types of components for measuring depths. Method: Firstly the gauge must be turned on, secondly the jaws must be closed and thirdly the readout indicator must be zeroed. Apply the gauge pointers to the component ensuring by observation of the digital readout that the plane of measurement goes through the centre of the component. Ensure that the measurement conforms to the stated tolerance Analogue/Digital Depth Gauge. This is used for measuring the depth of holes and recesses in components Method: Apply the gauge to the hole/recess ensuring that it reaches the base. Take the reading from the digital readout (or analogue scale if battery fails). Ensure that the measurement conforms to the stated tolerance Master High Pressure Gauge. This is a calibrated gauge used to calibrate the high-pressure testing gauge in the Inspection Department. This must have a current calibration certificate. Method: Firstly, ensure that the pressurised cylinder is turned off and the highpressure regulator Testing Rig has been drained. Fit the High Pressure Master Test Gauge into the valve adjacent the test gauge that is being calibrated. Turn on the cylinder valve and then turn on the valves under each gauge. Compare the readings on each gauge and record the results.

4 2.3.6 Master Low Pressure Gauge. This is a calibrated pressure gauge and is used to calibrate pressure-testing gauges in the factory. This gauge MUST have a current calibration certificate. Method: The Master Pressure Gauge is coupled to an air cylinder via a T-piece gas tight connection. The gauge to be calibrated is similarly coupled to the opposite T-piece. The indicator on the gauge to be calibrated is adjusted to be the same as the Master Pressure Gauge. The date of calibration is attached to the face of the now calibrated gauge Master Vacuum Gauge. This is a calibrated vacuum gauge and is used to calibrate vacuum gauges within the factory area. This gauge MUST have a current calibration certificate. Method: The gauges are coupled up to a vacuum source and are proceeded with as in above Digital Flowmeter. This is a calibrated Flowmeter used to calibrate Flowmeters in use within the Factory and Repair areas in the workshop. The Flowmeter may be used in absence or as an addition of other Flowmeters. This unit must have a current calibration certificate. Method: Using a standard Flowmeter for a supply, connect the calibrated Master Flowmeter to the supply Flowmeter with a 1mt length of Oxygen Delivery Tube. Open the Supply Flowmeter Valve (starting from the lowest calibration point) until the first full flow reads on the calibrated Flowmeter. Remove the tube from the Calibrated Flowmeter without disturbing the flow from the supply Flowmeter then connect it to the Flowmeter being calibrated. Observe the reading and compare with the Master Flowmeter. Continue throughout the scale of the Flowmeter. 2.4 All details of the calibrations performed will be recorded on the Calibration Record Card MLD Incorrectly Calibrated Products 3.1 Where it has been found that previous production has been produced using testing equipment that was out of calibration, the Technical Department Management will review the effects on the quality of the product. 3.2 Where it has been found that the effect of the product may be hazardous which may cause injury to the user, the patient and/or effect the environment, the Technical Department Management will refer to the Recall & Advisory Notices Procedure (SOP 25) for further action paragraph 2 The relevant competent authorities and the Notifying Bodies will be informed accordingly as stated in SOP 25 Recall & Advisory Notices.

5 Clause(sub) BS EN ISO 9001/13485 Details referenced 7.6 Control of monitoring and measuring devices paragraph 2 The organization shall determine the monitoring and measurement to be undertaken and the monitoring and measuring equipment needed to provide evidence of conformity of product to determined requirements. The organization shall establish documented procedures (processes) to ensure that monitoring and measurement can be carried out and are carried out in a manner that is consistent with the monitoring and measurement requirements. Where necessary ensure valid results, measuring equipment shall a) be calibrated or verified, or both, at specified intervals, or prior to use, against measurement standards traceable to international or national measurement standards; where no such standards exist, the basis used for calibration or verification shall be recorded; b) be adjusted or re-adjusted as necessary; c) have identification in order to determine its calibration status; d) be safeguarded from adjustments that would invalidate the measurement result; e) be protected from damage and deterioration during handling, maintenance and storage. In addition, the organization shall assess and record the validity of the previous measuring results when the equipment is found not to conform to requirements. The organization shall take appropriate action on the equipment and any product affected. Records of the results of calibration and verification shall be maintained (see 4.2.4). NOTE: Conformation of the ability of computer software to satisfy the intended application would typically include its verification and configuration management to maintain its suitability for use. The organization shall establish documented procedures for the issue and implementation of advisory notices. These procedures shall be capable of being implemented at any time.

6 History Page DCN Author: Date: Details Red references ISO 9001 clauses; blue ISO clauses.

MEDICAL REGULATOR RANGE (MEDIREG RANGE)

MEDICAL REGULATOR RANGE (MEDIREG RANGE) MEDICAL REGULATOR RANGE (MEDIREG RANGE) Medireg Dialreg Flowreg INSPECTION Remove the Medireg from the packaging and inspect for damage. If there is any damage, DO NOT USE, and contact Therapy Equipment

More information

Title: INSTRUMENT AND EQUIPMENT DOCUMENTATION AND RECORDS

Title: INSTRUMENT AND EQUIPMENT DOCUMENTATION AND RECORDS Page 1 of 7 Sections Included in this Document and Document History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References/(Deleted date) 6. Procedure 7. Definitions 8. Records 9. Supporting

More information

Pressure Equipment Directive PED 2014/68/EU Commission's Working Group "Pressure"

Pressure Equipment Directive PED 2014/68/EU Commission's Working Group Pressure F. INTERPRETATION OF THE ESSENTIAL SAFETY REQUIREMENTS ON MANUFACTURING Guideline F-01 Guideline related to: Annex 1 Section 3.1.2 According to section 3.1.2 (permanent joining) of Annex I, the third party

More information

INSTALLATION AND OPERATING INSTRUCTIONS

INSTALLATION AND OPERATING INSTRUCTIONS INSTALLATION AND OPERATING INSTRUCTIONS For 3D Instruments Direct Drive Pressure Gauges 3D Instruments Inc., 15542 Chemical Lane, Huntington Beach, CA, 92649, U.S.A. Tel. (714) 894-5351, Fax. (714) 895-4309

More information

Oxygen Dialflow Meter. Instructions for Use

Oxygen Dialflow Meter. Instructions for Use Oxygen Dialflow Meter Instructions for Use 702-0031.12 December 2017 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in injury to the patient,

More information

WIKA INSTRUMENT CORPORATION

WIKA INSTRUMENT CORPORATION WIKA INSTRUMENT CORPORATION Instruction Manual DIFFERENTIAL PRESSURE GAUGE Series 1500 Series 1000 Series 300 WIKA Instrument Corporation 1000 Wiegand Boulevard Lawrenceville, GA 30043 1-888-945-2872 http://www.wika.com

More information

Compiled by: B Beard. Approved by: SH Carstens. Description of requirements and procedures for compact provers to be used as verification standards.

Compiled by: B Beard. Approved by: SH Carstens. Description of requirements and procedures for compact provers to be used as verification standards. 1. Scope Description of requirements and procedures for compact provers to be used as verification standards. 2. Reference documents Trade Metrology Act SANS1698 3. Policy A. BASIC REQUIREMENTS Compact

More information

TEST BENCH SAFETY VALVES ¼ - 5 DN10 DN125

TEST BENCH SAFETY VALVES ¼ - 5 DN10 DN125 TEST BENCH SAFETY VALVES ¼ - 5 DN10 DN125 Model: VC-40-VYC Table of contents 1. - Installing the test bench 1.1.1- Connecting the compressed air / nitrogen source 1.1.2- Maximum test pressure according

More information

M-06 Nitrogen Generator (Nitrogen Making Machine)

M-06 Nitrogen Generator (Nitrogen Making Machine) Guideline No.M-06 (201510) M-06 Nitrogen Generator (Nitrogen Making Machine) Issued date: 20 th October, 2015 China Classification Society Foreword This Guideline is a part of CCS Rules, which contains

More information

SEMEM3-73 Maintaining medical gas pipeline systems and equipment

SEMEM3-73 Maintaining medical gas pipeline systems and equipment Maintaining medical gas pipeline systems and equipment Overview This unit identifies the competences you need to carry out maintenance activities on medical gas pipeline systems and equipment, in accordance

More information

Accreditation for In-Service Inspection of Transportable Pressure Receptacles (TPRs)

Accreditation for In-Service Inspection of Transportable Pressure Receptacles (TPRs) RG 3 Edition 4 December 2018 Accreditation for In-Service Inspection of Transportable Pressure Receptacles (TPRs) RG 3 / Edition: 4 Page 1 of 8 Contents 1. Introduction 2 2. Scope - Inspection Services

More information

Oxygen Dialflow Meter. Instructions for Use

Oxygen Dialflow Meter. Instructions for Use Oxygen Dialflow Meter Instructions for Use 702-0031.9 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

Compact Triple Cabinet Outlet Station Model Installation and Operating Instructions

Compact Triple Cabinet Outlet Station Model Installation and Operating Instructions Compact Triple Cabinet Outlet Station Model 6258-1 Installation and Operating Instructions The Porter Compact Triple Outlet Station (6258-1) provides a quick, safe, and reliable method of connection to

More information

INSPECTION FOCUS ON:

INSPECTION FOCUS ON: Newsletter of the South African National Accreditation System Issue 28, October 2016 INSPECTION FOCUS ON: INSPECTION OF GAS CYLINDERS AND OCCUPATIONAL HYGIENE REGULATIONS BY LINDA GRUNDLINGH 1. Occupational

More information

SFHGEN77 - SQA Code HD48 04 Perform first line calibration on clinical equipment to ensure it is fit for use

SFHGEN77 - SQA Code HD48 04 Perform first line calibration on clinical equipment to ensure it is fit for use Perform first line calibration on clinical equipment to ensure it is fit for Overview This standard relates to first line calibration of specific equipment prior to clinical. Individuals will be able to

More information

Operating Manual. Models AQT-02 & AQT-COO2 Manual No. MON004 (Rev 2 April 2000)

Operating Manual. Models AQT-02 & AQT-COO2 Manual No. MON004 (Rev 2 April 2000) Models AQT-02 & AQT-COO2 Manual No. MON004 (Rev 2 April 2000) Operating Manual AIR SYSTEMS INTERNATIONAL, INC. 829 Juniper Crescent, Chesapeake, Va., 23320 Telephone (757) 424-3967 Toll Free 1-800-866-8100

More information

Limerock Bearing Ratio Technician Training Course

Limerock Bearing Ratio Technician Training Course Limerock Bearing Ratio Technician Training Course Module 2: Equipment LBR Technician Release 10, Module 2 1 - The equipment section of this training will cover equipment referenced in AASHTO T 180, FM

More information

CalCheck. Instrument User Manual V1.8

CalCheck. Instrument User Manual V1.8 CalCheck Instrument User Manual V1.8 Unrivalled Part number: 21518 Detection. Part number: 21518 Declaration of conformity Manufacturer:, The Way, Fowlmere, Cambridge, England. SG8 7UJ Product: CalCheck

More information

TECHNICAL DATA CAUTION

TECHNICAL DATA CAUTION Page 1 of 6 1. DESCRIPTION The Viking Model D-2 Accelerator is a quick-opening device, with an integral anti-flood assembly, used to increase the operating speed of a differential type dry pipe valve.

More information

PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL Company Registered Office: Crown House, Stockport, Cheshire, SK13RB No. 850 0700 67 Registered in England and Wales No. 05058855

More information

TG GUIDELINES CONCERNING CALIBRATION INTERVALS AND RECALIBRATION

TG GUIDELINES CONCERNING CALIBRATION INTERVALS AND RECALIBRATION GUIDELINES CONCERNING CALIBRATION INTERVALS AND RECALIBRATION Approved By: Senior Manager: Mpho Phaloane Revised By: Field Manager: Neville Tayler Date of Approval: 2015-08-26 Date of Implementation: 2015-08-26

More information

Dialflow Regulator. Instructions for Use

Dialflow Regulator. Instructions for Use Dialflow Regulator Instructions for Use 702-0030.11 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

HPICAL Operation & Data Logging Procedures. Click spacebar to advance through slides 1

HPICAL Operation & Data Logging Procedures. Click spacebar to advance through slides 1 HPICAL-15000 Operation & Data Logging Procedures Click spacebar to advance through slides 1 WARNING Always wear proper safety equipment when using high pressure equipment. Do not exceed 125 psi air pressure.

More information

159 mm 154 mm. Subject to changes 10/2012

159 mm 154 mm. Subject to changes 10/2012 Low-pressure calibrator LPX calibrator medium: air / up to 10 bar KELLER pressure calibrators are highly accurate calibration instruments which are available for the following pressure ranges: -0,85 10

More information

Compact Triple Cabinet Outlet Station Model B Installation and Operating Instructions

Compact Triple Cabinet Outlet Station Model B Installation and Operating Instructions PORTER Parker Hannifin Corporation Porter Instrument Division 245 Township Line Rd. P.O. Box 907 Hatfield, PA 19440-0907 USA (215) 723-4000 / fax (215) 723-5106 Compact Triple Cabinet Outlet Station Model

More information

Micro Dial-Flowmeter. Instructions for Use

Micro Dial-Flowmeter. Instructions for Use Micro Dial-Flowmeter Instructions for Use 702-0082.8 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

300 Held Drive Northampton Pa

300 Held Drive Northampton Pa 300 Held Drive Northampton Pa 18067 610.262.6090 June 10, 2015 Recommended testing for verifying device performance. Intervals of testing are to be set by the Healthcare Provider / Dealer. Table of Contents

More information

Pressure Calibrators. Series PC..., PM...

Pressure Calibrators. Series PC..., PM... Pressure Calibrators Series PC..., PM... SIKA-Precision Pressure Calibrators, Series PC..., PM... We help you, if you are under pressure! Pneumatic pressure calibrators PM 25 A, PM 25 D, PM 110 PIC pressure

More information

Model: AT-1000 L.P.G. FREEZE VALVE

Model: AT-1000 L.P.G. FREEZE VALVE Model: AT-1000 P/N: 2156006200 3829 Forest Parkway Suite 500 Wheatfield, NY 14120-3763 Phone: 716-629-3802 Fax: 716-693-9162 solutions@viatran.com www.viatran.com 1 OPERATION AND MAINTENANCE INSTRUCTIONS

More information

SEMPEO SQA Unit Code FP2P 04 Maintaining fluid power equipment

SEMPEO SQA Unit Code FP2P 04 Maintaining fluid power equipment Overview This standard covers a broad range of basic fluid power maintenance competences which will prepare you for entry into the engineering or manufacturing sectors, creating a progression between education

More information

Installation and operating manual. Pneumatic control station LK product no: PCS 1-10

Installation and operating manual. Pneumatic control station LK product no: PCS 1-10 LK product no: PCS 1-10 Article no: 74503 Revision:8 Article no: 74503 Revision: 8 2 (23) Contents 1. General information... 5 2. Safety precautions... 5 2.1 Significance of symbols... 5 2.2 Explanatory

More information

SKSTEX36(SQA Unit Code-H7C9 04) Calibrate textile production equipment

SKSTEX36(SQA Unit Code-H7C9 04) Calibrate textile production equipment Overview This standard is for those who carry out routine calibrations at prescribed intervals in order to keep equipment in calibration. The job role will involve: 1. Performing calibration 2. Identifying

More information

GUIDANCE IN-SERVICE INSPECTION PROCEDURES

GUIDANCE IN-SERVICE INSPECTION PROCEDURES GUIDANCE IN-SERVICE INSPECTION PROCEDURES Number: PEC 13 Prepared by: Pressure Equipment Committee (TC1) Status: Issue: 07 Date: 16 th January 2018 Reference: Guidance for safety valve examination. SITUATION:

More information

20. Technical Equipment. Archived. This section applies to all equipment used in forensic examinations, such as

20. Technical Equipment. Archived. This section applies to all equipment used in forensic examinations, such as Instruments and Equipment Introduction 5.5. Authorization to use technical equipment 5.5.3 Use of instruments and equipment Instruction and maintenance manuals 5.4. 5.5.3 Use of nonlaboratory equipment

More information

SAPAG. Safety valves, type 5700 Storage, Use, Operation and Maintenance Instructions. IMPORTANT NOTICE

SAPAG. Safety valves, type 5700 Storage, Use, Operation and Maintenance Instructions.  IMPORTANT NOTICE SAPAG IMPORTANT NOTICE Contents Important notice 1 0 Valve identification 2 1 Storage 2 2 Installation 2 3 Operation 2 4 Maintenance 3 4.1 Dismantling 3 4.2 Inspection 3 4.3 Repair 3 4.4 Assembly 4 4.5

More information

Supersedes: The copy of this document located on Measurement Canada s website is considered to be the controlled copy.

Supersedes: The copy of this document located on Measurement Canada s website is considered to be the controlled copy. Bulletin Category: GAS Bulletin: G-23 Page: 1 of 7 The copy of this document located on Measurement Canada s website is considered to be the controlled copy. Recognition of data from specialty gas suppliers

More information

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER 1250 METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER 1.0 Scope and Application. 1.1 Analytes. Analyte CAS No. Sensitivity Total Organic Compounds N/A

More information

Operating manual for the pressure transducer unit CCS 1000

Operating manual for the pressure transducer unit CCS 1000 Operating manual for the pressure transducer unit CCS 1000 Please read the operating manual before using the device. 1 Content: 1. General description 2. Technical data 3. Operation 4. Scope of supply

More information

Requirements for the certification of movements and mechanical watches resistant to magnetic fields of 1.5 T ( G)

Requirements for the certification of movements and mechanical watches resistant to magnetic fields of 1.5 T ( G) Institut fédéral de métrologie METAS METAS-N001/EN Requirements for the certification of movements and mechanical watches resistant to magnetic fields of 1.5 T (15 000 G) This document is a translation

More information

1811G/H PITOT-STATIC TEST SET

1811G/H PITOT-STATIC TEST SET 1811G/H PITOT-STATIC TEST SET USER INSTRUCTION MANUAL M/N: 1811G/H, P/Ns: 101-00165 101-00168 101-00169 Doc. P/N: 56-101-00165_00168_00169 Revision C December 9, 2014 BARFIELD, INC. Corporate Headquarters

More information

GUIDELINES. Systems. Pressure. Guidelines Acceptance of equipment, establishment of safe operating limits and fitness for service

GUIDELINES. Systems. Pressure. Guidelines Acceptance of equipment, establishment of safe operating limits and fitness for service GUIDELINES Pressure Systems Guidelines Acceptance of equipment, establishment of safe operating limits and fitness for service Document Reference: PSG 16 Issue 04 Dated 13/09/16 The Safety Assessment Federation

More information

Accessories. Differential pressure gauges for air and gas MAG-A & MAG-G (E5275 rev /04/2010)

Accessories. Differential pressure gauges for air and gas MAG-A & MAG-G (E5275 rev /04/2010) Accessories Differential pressure gauges for air and gas MAG-A & MAG-G (E5275 rev. 02-12/04/2010) GENERAL WARNINGS: DISPOSAL: ¾ All installation, maintenance, ignition and setting must be performed by

More information

PROCEDURES FOR REPAIRS TO ASME NV STAMPED PRESSURE RELIEF DEVICES OF NUCLEAR SAFETY RELATED PRESSURE RELIEF VALVES

PROCEDURES FOR REPAIRS TO ASME NV STAMPED PRESSURE RELIEF DEVICES OF NUCLEAR SAFETY RELATED PRESSURE RELIEF VALVES NB16 0603 NR Task Group 1 9 16 1. One update by NR task group page 3: 7/17/19 2. Two updates to S6.1 and S6.3 on page 1: 7/19/17 SUPPLEMENT 6 PROCEDURES FOR REPAIRS TO ASME NV STAMPED PRESSURE RELIEF DEVICES

More information

ACCREDITATION OF ON-LINE LEAK SEAL ORGANIZATIONS

ACCREDITATION OF ON-LINE LEAK SEAL ORGANIZATIONS ACCREDITATION OF ON-LINE LEAK SEAL ORGANIZATIONS TSSA GUIDE FOR SURVEY TEAMS The Technical Standards and Safety Authority Boilers and Pressure Vessels Safety Division 3300 Bloor Street West 14th Floor,

More information

User Manual Pressure Calibration System C and Hydraulic Pressure Pump Models T-620/T-620H Copyright 2008 AMETEK Denmark A/S

User Manual Pressure Calibration System C and Hydraulic Pressure Pump Models T-620/T-620H Copyright 2008 AMETEK Denmark A/S User Manual Pressure Calibration System C and Hydraulic Pressure Pump Models T-620/T-620H Copyright 2008 AMETEK Denmark A/S About this manual. The structure of the manual This user manual is aimed at users

More information

EXPIRATION DATE: (FOR RENEWAL, SEE 49 CFR ) 1. GRANTEE: Structural Composites Industries (SCI) Pomona, CA

EXPIRATION DATE: (FOR RENEWAL, SEE 49 CFR ) 1. GRANTEE: Structural Composites Industries (SCI) Pomona, CA East Building, PHH-30 1200 New Jersey Avenue S.E. U.S. Department of Transportation Washington, D.C. 20590 Pipeline and Hazardous Materials Safety Administration DOT-SP 13583 (NINTH REVISION) EXPIRATION

More information

ST 18 Aluminium Gas Model

ST 18 Aluminium Gas Model ST 18 Aluminium Gas Model Operators Manual Issue 5 For correct operation of this autoclave it is essential that the chamber be filled with water up to the height of the studs before every use. DIXONS SURGICAL

More information

CALIBRATION SYSTEM REQUIREMENTS. ESCC Basic Specification No

CALIBRATION SYSTEM REQUIREMENTS. ESCC Basic Specification No Page 1 of 8 CALIBRATION SYSTEM REQUIREMENTS ESCC Basic Specification Issue 4 February 2014 Document Custodian: European Space Agency see https://escies.org PAGE 2 LEGAL DISCLAIMER AND COPYRIGHT European

More information

PRESSURE REGULATORS AND FLOWMETERS, ALL MEDICAL GASES

PRESSURE REGULATORS AND FLOWMETERS, ALL MEDICAL GASES PRESSURE REGULATORS ALL MEDICAL GASES In this section we propose our pressure regulators (FM series, MU and MERCURY ) and our Rs flowmeters for applications other than oxygen therapy (presented in the

More information

Specific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities

Specific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities Specific Accreditation Criteria Calibration ISO IEC 17025 Annex Mass and related quantities January 2018 Copyright National Association of Testing Authorities, Australia 2014 This publication is protected

More information

Technical Specification

Technical Specification Medical Gas Terminal Units (Gas Outlet Points) Technical Specification EN 9170-1 Terminal Unit Product Description CPX Terminal Units comprise a first fix, second fix and check valve. The first fix assembly

More information

DPC-30 DPC-100. Reference Manual

DPC-30 DPC-100. Reference Manual DPC-30 DPC-100 Reference Manual 1. Introduction 1.1 Description The Martel DPC Digital Pneumatic Calibrator improves upon traditional dial gauge pneumatic calibrators. The Martel DPC improves accuracy,

More information

INSTALLATION, OPERATION AND MAINTENANCE MANUAL. Model Sanitary Vent SECTION I

INSTALLATION, OPERATION AND MAINTENANCE MANUAL. Model Sanitary Vent SECTION I INSTALLATION, OPERATION AND MAINTENANCE MANUAL Model 1100 Sanitary Vent IOM-1100 03-17 ISO Registered Company SECTION I I. DESCRIPTION AND SCOPE The Model 1100 is a stainless steel sanitary vent designed

More information

IMPLEMENTATION OF EIGA CARBON DIOXIDE STANDARDS

IMPLEMENTATION OF EIGA CARBON DIOXIDE STANDARDS CODE OF PRACTICE 42 IMPLEMENTATION OF EIGA CARBON DIOXIDE STANDARDS REVISION 1 : 2014 CODE OF PRACTICE 42 IMPLEMENTATION OF EIGA CARBON DIOXIDE STANDARDS REVISION 1: 2014 Copyright 2014 by British Compressed

More information

Pressure Equipment Directive (PED) 97/23/EC Page 033 of 124

Pressure Equipment Directive (PED) 97/23/EC Page 033 of 124 Pressure Equipment Directive (PED) 97/23/EC Page 033 of 124 13.7 Pressure Equipment Directive (PED) 97/23/EC 1 The Pressure Equipment Directive (PED) 97/23/EC applies to the design, manufacturing and conformity

More information

SEMEM3-65 Servicing anaesthetic and ventilation equipment

SEMEM3-65 Servicing anaesthetic and ventilation equipment Overview This unit identifies the competences you need to carry out servicing activities on anaesthetic and ventilation equipment, in accordance with approved procedures. You will be required to service

More information

Pneumatic high-pressure controller Model CPC7000

Pneumatic high-pressure controller Model CPC7000 Calibration technology Pneumatic high-pressure controller Model CPC7000 WIKA data sheet CT 27.63 Applications Healthcare and avionics industry Industry (laboratory, workshop and production) Transmitter

More information

Technical Procedure for Instrument Calibration and Maintenance

Technical Procedure for Instrument Calibration and Maintenance Technical Procedure for Instrument Calibration and Maintenance 1.0 Purpose To outline the schedules and procedures involved in the maintenance and verification of critical instrument calibration in the

More information

STAD. Balancing valves DN (MD60, MD61, MD62)

STAD. Balancing valves DN (MD60, MD61, MD62) STAD Balancing valves DN 15-50 (MD60, MD61, MD62) IMI TA / Balancing valves / STAD STAD The STAD balancing valve delivers accurate hydronic performance in an impressive range of applications. Ideally suited

More information

Flexible hoses - Non-metallic materials

Flexible hoses - Non-metallic materials CLASS PROGRAMME Type approval DNVGL-CP-0183 Edition December 2015 The electronic pdf version of this document, available free of charge from http://www.dnvgl.com, is the officially binding version. FOREWORD

More information

SAMPLE WRITTEN PROGRAM for Control of Hazardous Energy LOCKOUT TAGOUT

SAMPLE WRITTEN PROGRAM for Control of Hazardous Energy LOCKOUT TAGOUT SAMPLE WRITTEN PROGRAM for Control of Hazardous Energy LOCKOUT TAGOUT SAMPLE WRITTEN PROGRAM The objective of this procedure is to establish a means of positive control to prevent the accidental starting

More information

Pressure Regulator. Instructions for Use

Pressure Regulator. Instructions for Use Pressure Regulator Instructions for Use 702-0083.9 December 2017 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in injury to the patient, the

More information

Pressure is our Business

Pressure is our Business Pressure is our Business Count on gauges and calibrators that perform Wherever you work Reduce Measurement Error Protect against diminished accuracy from environmental conditions, accidental damage, and

More information

BDT20 Safety process pressure gauge 100 & 160mm

BDT20 Safety process pressure gauge 100 & 160mm BDT20 Safety process pressure gauge 100 & 160mm Product description Badotherm pressure gauge model BDT20 is the solid front, safety pattern gauge according the highest class of the EN 837-1 / 9.7.2 and

More information

BCGA GUIDANCE NOTE 17

BCGA GUIDANCE NOTE 17 BCGA GUIDANCE NOTE 17 BCGA POLICY AND GUIDANCE FOR THE SAFE FILLING OF THIRD-PARTY OWNED AND / OR MAINTAINED TANKS Revision 2: 2013 BCGA GUIDANCE NOTE 17 BCGA POLICY AND GUIDANCE FOR THE SAFE FILLING OF

More information

Smart Water Application Technologies (SWAT) TM

Smart Water Application Technologies (SWAT) TM Smart Water Application Technologies (SWAT) TM Equipment Functionality Test Protocol Version 2.3 June 2014 SWAT Committee 8280 Willow Oaks Corporate Drive Suite 400 Fairfax, VA 22031 www.irrigation.org

More information

Technical Procedure for General Laboratory Equipment

Technical Procedure for General Laboratory Equipment Technical Procedure for General Laboratory Equipment 1.0 Purpose - This procedure specifies the required elements for the use of general laboratory equipment. 2.0 Scope - This procedure applies to all

More information

ISO INTERNATIONAL STANDARD. Flow-metering devices for connection to terminal units of medical gas pipeline systems

ISO INTERNATIONAL STANDARD. Flow-metering devices for connection to terminal units of medical gas pipeline systems INTERNATIONAL STANDARD ISO 15002 Second edition 2008-07-01 Flow-metering devices for connection to terminal units of medical gas pipeline systems Dispositifs de mesure de débit pour raccordement aux prises

More information

Differential Pressure Regulator Type Type 45-6 (0.1 to 1 bar, DN 15) Mounting and Operating Instructions EB 3226 EN

Differential Pressure Regulator Type Type 45-6 (0.1 to 1 bar, DN 15) Mounting and Operating Instructions EB 3226 EN Differential Pressure Regulator Type 45-6 Type 45-6 (0.1 to 1 bar, DN 15) Mounting and Operating Instructions EB 3226 EN Edition March 2008 Contents Contents Page 1 Design and principle of operation...................

More information

NOVALYNX CORPORATION MODEL RAIN GAUGE CALIBRATOR INSTRUCTION MANUAL

NOVALYNX CORPORATION MODEL RAIN GAUGE CALIBRATOR INSTRUCTION MANUAL NOVALYNX CORPORATION MODEL 260-2595 RAIN GAUGE CALIBRATOR INSTRUCTION MANUAL REVISION DATE: August 2018 Receiving and Unpacking Carefully unpack all components and compare to the packing list. Notify NovaLynx

More information

Installation and Maintenance Instruction Manual

Installation and Maintenance Instruction Manual Installation and Maintenance Instruction Manual Bourdon tube pressure gauge, model T5400, T5500 and T6500 for industrial application In the following configuration: ###T5400### bourdon tube pressure gauge

More information

Approved Installation Method (AIM)

Approved Installation Method (AIM) AIM-08-01 Issue 2 (DRAFT) 2015 Water Undertakers Approved Installation Method (AIM) Type BA Device Verifiable Backflow Preventer with Reduced Pressure Zone (RPZ Valve) Requirements for installation, commissioning

More information

Model: AT-1000 L.P.G. FREEZE VALVE

Model: AT-1000 L.P.G. FREEZE VALVE Model: AT-1000 P/N: 2156006200 3829 Forest Parkway Suite 500 Wheatfield, NY 14120-3763 Phone: 716-629-3802 Fax: 716-693-9162 solutions@viatran.com www.viatran.com 1 OPERATION AND MAINTENANCE INSTRUCTIONS

More information

deltaflowc deltaflowc Venturi or Probe

deltaflowc deltaflowc Venturi or Probe deltaflowc Mass Flowmeter for Gases - Multivariable with ultra fast dp, p and T-sensors - Compact, accurate and user-friendly - Ideal for OEMs deltaflowc Venturi or Probe Precise mass flow metering deltaflowc

More information

SCOPE OF WORK. The Repair of the Mobile Harbour Crane leaking Luffing Cylinder

SCOPE OF WORK. The Repair of the Mobile Harbour Crane leaking Luffing Cylinder EAST LONDON MULTI-PURPOSE TERMINAL Document Title: SCOPE OF WORK Scope Of Work Title: The Repair of the Mobile Harbour Crane leaking Luffing Cylinder TPT MHC luffing cylinder repair 1 1. BACKGROUND East

More information

Control of Energy - Isolation Process

Control of Energy - Isolation Process Issue Date: 24/03/17 Last Reviewed: 24/03/17 Next Review Date: 24/03/18 Authorised By: Mike Muir Safe Operating Procedure Control of Energy - Isolation Process Purpose: This SOP is based on a Generic 12-step

More information

ANNEX A. (Normative) Tank Calibration Frequency and Re-computation of Calibration Tables

ANNEX A. (Normative) Tank Calibration Frequency and Re-computation of Calibration Tables ANNEX A (Normative) Tank Calibration Frequency and Re-computation of Calibration Tables Terms used in this document: Complete tank calibration External tank calibration Class A tanks Class B tanks Uniform

More information

Anti-flood device Model B-1

Anti-flood device Model B-1 December 4, 2009 Dry Systems 123a 1. DESCRIPTION The Anti-flood Device is required when Viking accelerators are installed on dry systems according to Viking Model E-1 Accelerator Trim Charts. In the SET

More information

Australian Standard. Bourdon tube pressure and vacuum gauges AS

Australian Standard. Bourdon tube pressure and vacuum gauges AS AS 1349 1986 Australian Standard Bourdon tube pressure and vacuum gauges [Title allocated by Defence Cataloguing Authority: GAUGES, PRESSURE AND VACUUM (Bourdon Tube Type)... NSC Gp 66] This Australian

More information

DTG - LCD Digital Temperature Gauge USER MANUAL

DTG - LCD Digital Temperature Gauge USER MANUAL DTG - LCD USER MANUAL Page 1 of 7 USERS GUIDE page 1. Description 3 2. Function 3 3. Safety Instruction 3 3.1. Safety Conventions 3 3.2. Proper Use 3 4. Installation 4 4.1. Unpacking 4 4.2. Storage 4 4.3.

More information

INSPECTORATE FOR DIVING EQUIPMENT SERVICING & TESTING CODE OF PRACTICE CP11:2011

INSPECTORATE FOR DIVING EQUIPMENT SERVICING & TESTING CODE OF PRACTICE CP11:2011 D003 INSPECTORATE FOR DIVING EQUIPMENT SERVICING & TESTING CODE OF PRACTICE CP11:2011 For INSPECTING and TESTING Transportable Pressure Receptacles (Cylinders) for Breathing Gases manufactured from Steel

More information

ST 18 Aluminium Gas Model

ST 18 Aluminium Gas Model DIXONS SURGICAL INSTRUMENTS LTD, 1 ROMAN COURT HURRICANE WAY, WICKFORD BUSINESS PARK WICKFORD, ESSEX, SS11 8YB, UK TEL: +44 (0)1268 764614 - FAX: +44 (0)1268 764615 EMAIL: info@dixons-uk.com - WEB: http://www.dixons-uk.com

More information

Calibration Procedure

Calibration Procedure Management Procedure 2535 Revision: E Date Issued: October 23, 1998 Date Revised: March 6, 2012 Calibration Procedure DeFelsko Corporation PosiTector 6000 N PosiTector 6000 NS PosiTector 6000 NRS Coating

More information

Appendix C Field Testing of Pipelines

Appendix C Field Testing of Pipelines 1 (Normative) C1 Scope Appendix C is based on some of the test methods in AS/NZS 2566.2, section 6, and associated appendices. This appendix specifies suggested methods of test and their application to

More information

AND OPERATION INSTRUCTIONS

AND OPERATION INSTRUCTIONS ADI 2111-D Certified ISO 9001 MEDICAL FLOWMETER (Back Pressure Compensated Thorpe Tube) INSTALLATION AND OPERATION INSTRUCTIONS Before Installing or Operating, Read and Comply with These Instructions Controls

More information

STD-3-V1M4_1.7.1_AND_ /15/2015 page 1 of 6. TNI Standard. EL-V1M4 Sections and September 2015

STD-3-V1M4_1.7.1_AND_ /15/2015 page 1 of 6. TNI Standard. EL-V1M4 Sections and September 2015 page 1 of 6 TNI Standard EL-V1M4 Sections 1.7.1 and 1.7.2 September 2015 Description This TNI Standard has been taken through all of the voting stages and has received consensus approval by the TNI membership.

More information

Installation, operating and maintenance Instructions for Seemag bypass level indicator

Installation, operating and maintenance Instructions for Seemag bypass level indicator Issue: S Date: 05-09-14 Type G35 General information The Seetru bypass magnetic level indicator, abbreviate SEEMAG, serves to show the filling level of fluids in tanks, basins, tubes etc. The Seemag operates

More information

COSASCO 6000 PSI SINGLE ISOLATION SERVICE VALVE (FR) MAINTENANCE

COSASCO 6000 PSI SINGLE ISOLATION SERVICE VALVE (FR) MAINTENANCE COSASCO 6000 PSI SINGLE ISOLATION SERVICE VALVE (FR) MAINTENANCE 11841 Smith Avenue Santa Fe Springs, CA 90670 Tel: (562) 949-0123 (800) 635-6898 Fax: (562) 949-3065 www.cosasco.com Page: 2 of 26 Revision

More information

VACUUM TESTING PRECAST CONCRETE MANHOLES

VACUUM TESTING PRECAST CONCRETE MANHOLES 1 OF 5 testing is a quick, safe and practical way to validate manhole system integrity. Manhole sections can be tested at the precast concrete plant prior to delivery or on site prior to backfilling. Here

More information

Pressure Regulators. Instructions for Use

Pressure Regulators. Instructions for Use Pressure Regulators Instructions for Use 702-0032.11 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

973-SF 6 Analyzer. Precise and Stable SF 6 Gas Analyzer REFLECTING YOUR STANDARDS

973-SF 6 Analyzer. Precise and Stable SF 6 Gas Analyzer REFLECTING YOUR STANDARDS Precise and Stable SF 6 Gas Analyzer Simultaneous measurement of humidity, SF 6 purity and SO 2 concentration Integrated gas recovery system with automatic pump back Fully automated SF 6 gas testing Fundamental

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS Air Entrainment Meter ELE International Chartmoor Road, Chartwell Business Park Leighton Buzzard, Bedfordshire, LU7 4WG England phone: +44 (0) 1525 249200 fax: +44 (0) 1525 249249

More information

Mr Roger Wood Measurement & Process Group (M&P) Network Strategy National Grid Gas plc Brick Kiln Street Hinckley Leicestershire LE10 0NA

Mr Roger Wood Measurement & Process Group (M&P) Network Strategy National Grid Gas plc Brick Kiln Street Hinckley Leicestershire LE10 0NA Mr Roger Wood Measurement & Process Group (M&P) Network Strategy National Grid Gas plc Brick Kiln Street Hinckley Leicestershire LE10 0NA Dear Roger, Your Ref: Our Ref: Direct Dial: 020 7901 7468 Email:

More information

San Antonio Water System Standard Specifications for Construction ITEM NO. 841 HYDROSTATIC TESTING OPERATIONS

San Antonio Water System Standard Specifications for Construction ITEM NO. 841 HYDROSTATIC TESTING OPERATIONS ITEM NO. 841 HYDROSTATIC TESTING OPERATIONS 841.1 DESCRIPTION: This item shall consist of hydrostatic testing operations, of water mains in accordance with these specifications. 841.2 MATERIALS: The materials

More information

CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL Company Registered Office: Crown House, Stockport, Cheshire, SK13RB No. 850 0700 67 Registered in England and Wales No. 05058855

More information

Fluid Sealing Association

Fluid Sealing Association Fluid Sealing Association STANDARD FSA-PSJ-701-06 PIPING SYSTEMS NON-METALLIC EXPANSION JOINT HYDROTESTING AND VACUUM TESTING 994 Old Eagle School Road, Suite 1019 Wayne, Pennsylvania 19087-1866 Phone:

More information

Pressure Equipment Directive PED 2014/68/EU Commission's Working Group "Pressure"

Pressure Equipment Directive PED 2014/68/EU Commission's Working Group Pressure H. INTERPRETATION OF OTHER ESSENTIAL SAFETY REQUIREMENTS Guideline H-02 Guideline related to: Annex I Section 3.2.2 and 7.4 Final assessment (Annex I Section 3.2.2) of pressure equipment must include a

More information

TECHNICAL DATA ANTI-FLOOD DEVICE MODEL B-1 1. DESCRIPTION

TECHNICAL DATA ANTI-FLOOD DEVICE MODEL B-1 1. DESCRIPTION Page 1 of 6 1. DESCRIPTION The Model B-1 Anti-flood Device is required when Viking accelerators are installed on dry systems according to Viking Model E-1 Accelerator Trim Charts. In the SET condition,

More information

[Reserved] Packagings ICC 3 1 3AA 3AL 3AX 3A480X 3AAX

[Reserved] Packagings ICC 3 1 3AA 3AL 3AX 3A480X 3AAX 173.300 [Reserved] 173.301 General requirements for shipment of compressed gases and other hazardous materials in cylinders, UN pressure receptacles and spherical pressure vessels. (a) General qualifications

More information

IN-PLACE DENSITY AND MOISTURE CONTENT OF SOIL AND SOIL- AGGREGATE BY NUCLEAR METHODS (SHALLOW DEPTH) FOP FOR AASHTO T 310

IN-PLACE DENSITY AND MOISTURE CONTENT OF SOIL AND SOIL- AGGREGATE BY NUCLEAR METHODS (SHALLOW DEPTH) FOP FOR AASHTO T 310 AND MOISTURE CONTENT OF SOIL AND SOIL- AGGREGATE BY NUCLEAR METHODS (SHALLOW DEPTH) FOP FOR AASHTO T 310 Scope This procedure covers the determination of density, moisture content, and relative compaction

More information