F4504 MULTI-AIR TESTER OPERATING MANUAL

Size: px
Start display at page:

Download "F4504 MULTI-AIR TESTER OPERATING MANUAL"

Transcription

1 F4504 MULTI-AIR TESTER OPERATING MANUAL FAC QAM 173 Issue B 07/17 0

2 1

3 Contents Page No Introduction 3 Calibration & Warranty 4 Equipment List Pre Use Start-Up Selecting the test standard and configuring the instrument. Breathing-Air Tests - All Completing an Airline Test Completing a High Pressure Test Medical Air Tests Pharmacopeia and HTM Medical Gas Tests Pharmacopeia Continuous Oxygen Reading Exporting Test Results and PC Software 23 Appendix 1 - Helpful Tips 25 Appendix 2 - Draeger Tube Tip Cutter 26 Appendix 3 - Reading the Draeger Oil Impactor and Chemical 27 Reagent Tubes Appendix 4 Breathing Air Standard 32 Appendix 5 Breathing Air Quality Testing Why? 33 Frequency of BA Tests / Recording results Recording Breathing Air Quality Test Results A.4.5 Compressed Air for Breathing Apparatus from EN COSHH L5 Sixth Edition 37 Respiratory Protective Equipment at Work 39 Appendix 6 Medical Air Standards 41 Conversion chart mg/m3 to ppm v/v 42 2

4 Introduction Multi-Air Testers are designed to provide a quick and easy method for carrying out accurate periodic testing of breathing air, medical-air and medical gas supplies. The instrument can test against a range of standards, for breathing air the European Standard is EN12021, which should be referred to in conjunction with any overriding national standards. For medical air, surgical air and medical gas the F4504 is programmed to test against the requirements of European Pharmacopoeia and in the UK, HTM The F4504, is designed primarily for use on airline systems up to 10bar when used in conjunction with the F3005 regulator can also test systems up to 20 bar. For testing high pressure cylinders or charging systems the F3002 High Pressure Regulator is available which has a 200/300 bar 5/8 DIN thread. Additionally the F4504 will test the following medical gases - CO 2, N 2 and N 2 O For some of the medical tests the F4504 uses alternative test methods to those described in the narrative of the standard. Standards which the F4504 can test against are: Breathing Air: EN12021:2014 Medical: HTM02-01 Medical and Surgical Air HTM02-01 Dental Air European Phamacopeia Medical and Surgical Air European Phamacopeia CO 2 European Phamacopeia N 2 European Phamacopeia N 2 O The F4504 has a colour touch screen display and is supplied in a hard weatherproof case with storage provision for a wide range of accessories and equipment. It is primarily designed to be used from within the case; however the tester can also be removed and operated separately where space is at a premium. As well as test ports for chemical reagent tubes/impactors the F4504 incorporates an electronic dewpoint meter, electrochemical oxygen cell, digital flowmeter and pressure transducer. Depending on the standard selected tests are carried out using the Draeger Impactor for oil and chemical reagent tubes for carbon monoxide, carbon dioxide, sulphur dioxide, nitrous fumes, hydrogen sulphide and polytest. These are supplied in packs of 10 per type and the chemical reagent tubes show the degree of contaminant present as a colour change to the crystals that they contain. The extent of this is read against the scale on the tube. For the Impactor the degree of oil contamination is displayed on a screen. 3

5 Results from the F4504 can be downloaded to PC software which can be downloaded from Factair s website ( The software allows you a convenient way to easily store, retrieve and print previous test results. Calibration and Warranty Multi-Air Testers leave our factory with a 12-month warranty and calibration certificate. Our standard turnaround on annual calibration is working days providing there is no major damage that requires an extensive rebuild. Note:- Please download all stored data before returning the Tester to Factair. Temperature Parameters Storage: -20/+50 C Operating Range: -10/+45 C This tester is calibrated for the following approved reagent tubes/impactor. Test Draeger ref. Factair Part Sensitivity No: Carbon Monoxide CO TUB ppm Carbon Dioxide CO TUB ppm Oil (Impactor) TUB mg/m 3 Oil (chemical reagent tube)* TUB mg/m 3 Nitrous Fumes NO x TUB ppm Sulphur Dioxide SO TUB ppm Polytest CH28401 TUB0017 Sensitivity dependant on compound present Hydrogen Sulphide H 2 S TUB ppm **Note:- The tester can also be used with the old type Draeger reagent tube for oil, ref (Refer to Appendix 2 at the rear of the manual for instructions.) IMPORTANT IT IS RECOMMENDED THAT YOUR MULTI-AIR TESTER IS RETURNED FOR RECALIBRATION AND SERVICING WITHIN 12 MONTHS FROM THE ISSUE DATE OF ITS CALIBRATION CERTIFICATE Equipment List: F4504 Multi-Air Tester Hard Case Mains Power Supply USB Memory Cable 6 x AA Alkaline Batteries Travel Container for used detector tubes Tube Tip Cutter F1946 Connection ¼ BSP inlet adaptor Stylus Pen 4

6 Optional Extras F Bar Regulator F /300 Bar 5/8 DIN thread High-Pressure Regulator F2193 Connection adaptor terminating in a CEJN Plug 1. Pre-Use F4504 Multi-Air Testers are fitted with internal lithium rechargeable batteries and in addition also have a battery tray which accommodates 6 off size AA alkaline type batteries packed separately and stored underneath the tester within the case. In the event that the rechargeable batteries are not sufficiently recharged then you can continue to use the instrument with either the Alkaline AA batteries or mains power supply, To install the alkaline AA batteries lift the tester out of its case, open the battery drawer at the rear of the unit and place the batteries inside. Note: - It is important to strictly observe polarity when fitting batteries and ensure that they are correctly located. During use the battery condition is displayed on the screen. If the Tester is to be used adjacent to a power socket, the mains adaptor may be used. Please note the unit should be switched off prior to connection or disconnection of the mains supply. When connected to a mains power supply the internal lithium batteries will be recharged. 5

7 2. Start Up To carry out a test first switch on the machine by pressing and holding the green on button, located on the right-hand side of the unit, for approximately 3 seconds. A splash screen will be displayed which shows the instruments hardware and firmware versions. Periodically the firmware is upgraded as part of the recalibration process as improvements are identified. A Stylus pen is provided for use with the touch screen. The display will show the, date last calibrated, serial number and a tick or cross symbol will indicate whether it is within the recommended calibration period. The F4500 series has 2 battery level displays, the left hand one is for the alkaline AA batteries, if fitted. The right hand one is the rechargeable internal lithium battery. Unless the instrument is connected to a mains power supply the instrument will use the highest of the 2 battery power sources. Please note if alkaline AAs are inserted whilst the unit is on their battery level will not be displayed until the unit is turned off and back on again. Press the arrow key to progress to the main menu. The main menu has 3 options available from left to right these are: Begin test, Review Previous Tests and Set-Up Menu. Each of these sub-menus are described in further detail below The instrument can also be turned off from this menu by pressing the red power symbol in the top right hand corner. 6

8 Note If the begin test icon is in grey this will be because the instrument s memory is full (20 tests maximum) and you need to remove at least 1 test before you can proceed. See Section 4 Reviewing Test Results for further details. 3. Selecting the test standard and configuring the instrument From the main menu press the configure icon. In the configuration sub menu there are 3 icons available, from left to right these are: Set Date Set Time (the instrument uses a 24 hour clock) Set units of measure and test standard The date and time are set prior to the unit being supplied but they can be adjusted as necessary. To select the test standard or the units of measure press the right hand icon. In this menu you can then choose date format; pressure and temperature units of measure and the breathing-air test standard. Once you have made your selection press the arrow in the top right corner to exit. To then exit from the configuration sub menu to the main menu press the arrow at the top left of the screen. 7

9 4. Breathing-Air Tests- All Having selected the breathing-air standard EN12021 in the configuration screen, to carry out a test press the begin test icon on the main menu The unit will automatically calibrate the oxygen cell against the surrounding atmosphere; the unit therefore needs to be in a well ventilated area. Note:- If the temperature is greater than 40 C inside the instrument the calibration will operate for an extended period to stabilise the Oxygen cell. You now have the option of entering a description for the test location, if you wish to use numbers press the function button. Push the right arrow button when you are ready to proceed. Choose between using a Draeger chemical reagent tube or Draeger oil Impactor in the test port. For further details on alternative chemical reagent tubes see appendices 2 and 3. For breathing-air tests the standard option is to select a Draeger oil Impactor. 8

10 The unit has 4 breathing-air test programmes available: Airline Systems up to 40 bar (top left); High Pressure Charging Compressors (top right) High Pressure Systems up to and including 200bar and High Pressure Cylinders above 200bar. All of the high-pressure tests require the use of the F3002 regulator which has a DIN connector that can accept input pressures up to 300bar. 4.1 Completing an Airline Test Select (Airline < 40 bar) note the maximum pressure the instrument is rated to is 10 bar. With the flow control valve in the closed position connect to the air supply to be tested. Ideally this should be done by using the hose which normally supplies the breathing apparatus. Fit the appropriate adaptor to the Tester first and then connect the supply hose to the adaptor. A test can be carried out at any pressure between 2 and 10 bar. Note:- If the unit is over-pressurised it will automatically shut down and display an overpressure warning; it will then need to be returned to the manufacturer for checking and resetting. Input pressure will be displayed on the measured pressure bar graph and the adjacent display. System pressure will automatically default to this reading. Note:- The pressure dewpoint at the end of the test will be calculated on the recorded system pressure. For the majority of applications the system pressure can be left at this default setting, however if the breathing-air system is distributed as breathing air at a different pressure to that being displayed this should be entered at this stage. 9

11 To enter press the right hand side (Set) icon use the keypad to enter the required value, press the right arrow to return. The (Reset) button will automatically reset the system pressure back to the measured pressure. Press the arrow key to proceed. We can now confirm the flow and pressure is adequate for your respiratory protective device. Adjust the system s flow control valve so that the flow meter registers the required level*, check the pressure is sufficient for the breathing-air equipment being used. It is recommended that this flow is left running for the duration of the test to simulate the conditions of use. Press the arrow key to proceed if the flow and pressure meet the minimum requirements for your RPD. *Note:- At high flow rates hearing protection should be considered. Carry out the odour test and smell the air exhausted from the sintered outlet. If there is a serious odour problem, fail the sample at this point of the test by pressing the fail button. Initiating flows of seriously contaminated air could result in damage to the instrument. If the odour test is satisfactory, press the pass button. The instrument will now commence a 5-minute purge sequence, during which time an airflow will be passed through the test port and across the electronic cells. This clears contaminants from previous tests and ensures that a representative sample is being tested. 10 Once the purge has finished, the CO, CO 2 and Draeger Impactor/oil tube should be inserted into the corresponding test port and tightened firmly; ensure the Impactor has an intact protective seal. This seal must remain in place for the duration of the test. Press the button in top right hand corner

12 to start the test. The test will now commence and the duration will be displayed as a countdown, together with pressure, flowrate, ambient temperature, carbon dioxide, carbon monoxide and oxygen content. During the test the flowmeter should be set to simulate the requirements of the respiratory protective device RPD. The unit can be left until the test time has elapsed. If the instrument detects that the dewpoint meter s rate of change is still too high (>0.2 dec C per minute) it will automatically continue the test until this stabilises (maximum test time 15 minutes). Once the test is complete close the flow control valve, the readings from the Impactor can now be entered. To enter the results press the corresponding icon on the screen. Note:- Instructions on reading the Impactor can be found in Appendix 3 at the back of the manual. 11

13 The tube readings should also then be entered, push the button corresponding to each chemical reagent tube / Impactor. Input the reading from the tube / Impactor and press the arrow icon in the top right hand corner of the screen. The test complete screen displays all the measurements recorded during the test. And automatically determines whether the test was a pass or fail against the standard displayed in the top right corner of the screen. The water content on low pressure tests is displayed as a pressure dewpoint. The test result is recorded in the instrument s memory, which can store up to 20 previous tests. If the test was a fail an orange exclamation mark would be indicating those criteria which did not meet the standard. The photograph opposite is an example of a failed test where carbon dioxide failed to meet the requirements of the selected standard. 12

14 For airline systems less than 40 bar the pressure dewpoint, i.e. the point at which water will begin to condense into a liquid form in the airline, must be at least 5 degrees centigrade below the lowest temperature at which the system will be used. The F4504 automatically calculates the pressure dewpoint from the measured water content and system pressure, it then compares this to the ambient temperature to determine whether it was a pass or fail. This test result is automatically recorded in the instrument s memory, which can store up to 20 previous tests. To view these tests return to the main menu and press the Review Tests button. Test results are stored accordingly to the date and time they were completed. From this menu previous tests can be viewed or deleted as required. To disconnect the Tester from the system make sure you leave the inlet adaptor on the instrument and only disconnect at the end of the supply hose. The adaptor can then be disconnected afterwards if required. To turn off the Tester use the return arrows to return to the main menu screen and press the X in the top right hand corner of the screen. Note: - The unit will automatically turn off after a few minutes if left unused at the main menu screen. 4.2 Completing a High Pressure Test For high pressure tests there are 3 different test options available. All of the high-pressure tests require the use of the F3002 regulator which has a DIN connector which can accept input pressures of up to 300 bar. High-pressure charging compressors High-pressure cylinder less than 200bar High-pressure cylinder greater than 200bar Note: For cylinder tests please ensure the test is carried out on a recently filled, fully charged cylinder. Note: For the high-pressure charging compressor test a DIN to DIN adaptor ref: F2158 is required to connect the F3002 high-pressure regulator to a charging hose. To carry out a test press the appropriate button. 13

15 Ensure that the flow control valve is fully closed and then connect the F3002 high-pressure regulator assembly to the cylinder or charging hose to be tested, ensuring that the regulator has the correct type of connector for the cylinder. Connect the regulator hose coupling to both the regulator and the inlet of the Tester. Open the cylinder valve and check that the cylinder is fully charged by reading the contents gauge on the regulator. (A partially discharged cylinder can give an incorrect water result.) The pressure displayed on the tester will be that of the regulated supply from the cylinder and must not exceed 10 bar. Press the button in the top right hand corner to proceed. Note:- if the unit is over-pressurised it will automatically shut down and display an overpressure warning; it will then need to be returned to the manufacturer for checking and resetting. Carry out the odour test by gently opening the flow control valve and smelling the air exhausted from the sintered outlet, then close the valve. If there is a serious odour problem, fail the sample at this point of the test. Initiating flows of seriously contaminated air through the tube ports could result in damage to the unit. If the odour test is satisfactory, press the pass icon. The instrument will now commence a 5-minute purge sequence, during which time an airflow will be passed through the test port and across the electronic cells. This clears contaminants from previous tests and ensures that a representative sample is being tested. 14

16 Once the purge has finished, the Draeger tube/impactor should be inserted into the corresponding test port and tightened firmly; ensure the Impactor has an intact protective seal. This seal must remain in place for the duration of the test. Press the button in top right hand corner to start the test. The test will now commence and the duration will be displayed as a countdown, together with pressure, ambient temperature, carbon dioxide, carbon monoxide and oxygen content. The unit can be left until the test time has elapsed. If the instrument detects that the dewpoint meter s rate of change is still too high (>0.2 dec C per minute) it will automatically continue the test until this stabilises (maximum test time 15 minutes). The Impactor can now be removed and evaluated. To enter the results press the corresponding icon on the screen. Note:- Instructions on reading the Impactor can be found in Appendix 3 at the back of the manual. 15

17 The test complete screen displays all the measurements recorded during the test. And automatically determines whether the test was a pass or fail against the standard displayed in the top right corner of the screen. For water content on high pressure tests both the measured value in mg/m 3 and the atmospheric dewpoint are provided. If the test was a fail an orange exclamation mark would be indicating those criteria which did not meet the standard. The photograph opposite is an example of a failed test where carbon dioxide failed to meet the requirements of the selected standard. The test result is automatically recorded in the instrument s memory, which can store up to 20 previous tests. To disconnect the Tester from the cylinder firstly turn off the cylinder valve, open the flow control valve until all the air is exhausted then close the valve. It is now safe to disconnect the connection hose and cylinder regulator. 16

18 5 Medical Air Tests Pharmacopeia and HTM02-01 Having selected the appropriate medical-air standard in the configuration screen, to carry out a test press the begin test icon on the main menu The unit will automatically calibrate the oxygen cell against the surrounding atmosphere; the unit therefore needs to be in a well ventilated area. Note:- If the temperature is greater than 40 C inside the instrument the calibration will operate for an extended period to stabilise the Oxygen cell. You now have the option of entering a description for the test location, if you wish to use numbers press the function button. Push the right arrow button when you are ready to proceed. For EU Phamacopeia Medical Air and HTM02-01 Medical and Surgical Air and Dental Air only Choose between using a Draeger oil Impactor or oil tube. For further details on alternative chemical reagent tubes see appendix 3. The standard option is to select a Draeger oil Impactor. 17

19 With the flow control valve in the closed position connect to the air supply to be tested. Ensure the supply hose has non hydroscopic properties to ensure it will not retain excessive moisture and effect the result. Fit the appropriate adaptor to the Tester first and then connect the supply hose to the adaptor. A test can be carried out at any pressure between 2 and 10 bar. Note:- If the unit is over-pressurised it will automatically shut down and display an overpressure warning; it will then need to be returned to the manufacturer for checking and resetting. Input pressure will be displayed on the measured pressure bar graph and the adjacent display. System pressure will automatically default to this reading. Note:- The pressure dewpoint at the end of the test will be calculated on the recorded system pressure. For the majority of applications the system pressure can be left at this default setting, however if the medical-air system is distributed at a different pressure to that being displayed this should be entered at this stage. To enter press the right hand side (Set) icon use the keypad to enter the required value, press the right arrow to return. The (Reset) button will automatically reset the system pressure back to the measured pressure. Press the arrow key to proceed. The instrument will now commence a 5-minute purge sequence, during which time an airflow will be passed through the test port and across the electronic cells. This clears contaminants from previous tests and ensures that a representative sample is being tested. 18

20 Once the purge has completed the chemical reagent tubes / Impactor should be inserted into the corresponding test port and tightened firmly. The tubes/impactors to be used are shown at the top of the screen and will be dependent on the test standard selected earlier. If using an Impactor ensure it has an intact protective seal. This seal must remain in place for the duration of the test. Press the button in top right hand corner to start the test. The test will now commence and the duration will be displayed as a countdown, together with test standard. The pressure, ambient temperature, and oxygen content will also be displayed. During the test the flowmeter should be kept closed. Test port volumes are automatically adjusted for different gas mediums. The unit can be left until the test time has elapsed. Depending on the standard and therefore the chemical reagent tubes / Impactor being used some will complete at different test times and the instrument will automatically stop the flow to these tubes. Each complete test port will be shown with a green tick. If the instrument detects that the dewpoint meter s rate of change is still too high (>0.2 dec C per minute) it will automatically continue the test until this stabilises (maximum test time 15 minutes). 19

21 Once the test is complete push the button corresponding to each chemical reagent tube / Impactor. Input the reading from the tube / Impactor and press the arrow icon in the top right hand corner of the screen. For the oil and polytest tubes select the appropriate pass or fail icons. After you have entered each result that box will change from red to green. Once you have completed entering all the results press the arrow icon in the top right hand corner of the screen. you enter each result the Note:- Instructions on reading the Impactor can be found in Appendix 4 at the back of the manual. 20

22 The test complete screen displays all the measurements recorded during the test. And automatically determines whether the test was a pass or fail against the standard displayed in the top right corner of the screen. The water content is automatically also displayed, together with oxygen content, ambient temperature and test pressure. The test result is recorded in the instrument s memory, which can store up to 20 previous tests. To disconnect the Tester from the system make sure you leave the inlet adaptor on the instrument and only disconnect at the end of the supply hose. The adaptor can then be disconnected afterwards if required. To turn off the Tester use the return arrows to return to the main menu screen and press the X in the top right hand corner of the screen. Note: - The unit will automatically turn off after a few minutes if left unused at the main menu screen. 6.0 Medical Gas Tests Pharmacopeia Having selected the appropriate medical-gas standard in the configuration screen, to carry out a test press the begin test icon on the main menu then follow the procedures as detailed in 5.0 Medical_Air Tests, with the following exception. When testing CO 2 to EU Pharmacopeia the H 2 S (Hydrogen Sulphide) tube should be inserted in the CO 2 port. 21

23 7.0 Continuous Oxygen Reading The F4504 can also provide a continuous oxygen reading. When you start a test press and hold the O2 cylinder for a few seconds which it is in its self calibration mode. Note:- If the temperature is greater than 40 C inside the instrument the calibration will operate for an extended period to stabilise the Oxygen cell. The instrument will then change operating modes to continuously display the Oxygen reading, connect your air source and the instrument will continue to provide a reading until you press the arrow key in the top corner. 22

24 8.0 Exporting test results and PC software To transfer test results from the Safe-Air Tester to the PC software you must first install the software on your PC. This is available in the download section of Factair s website: Then click on the File drop down menu and select import test results from F4500 via USB connection. You will then be able to choose the location where you would like test results to be stored. Individual test records can be opened from the file drop down menu by selecting Open an F4500 test results file. 23

25 Once loaded further information can be added about the following:- Test location. Date next test due. Test engineer and relevant address. Note:- These details can be saved as default. For each Airline test you will need to confirm whether the recorded test volume and pressure was sufficient for the RPD by selecting pass or fail as appropriate. Note:- In addition there is an option of entering N/A against these categories. In this instance the resultant test certificate will only validate air quality and exclude validation of the pressure and flow. When complete you can save the test result file and if required print off a certificate. In the UK, for breathing-air tests the Health and Safety Executive guideline HSG53 recommends test results should be retained for a minimum of 5 years. 24

26 With detector tubes Appendix 1 - HELPFUL TIPS Ensure that there are no fragments of glass in the tube glands prior to fitting detector tubes, clean if required. Always remove the ends of the detector tubes cleanly using the correct tube tip cutter. Do not use pincers or other devices. Always remove both ends of the detector tubes prior to fitting. Always fit detector tubes with the arrows facing outwards. Never break the glass ampoule in the oil tube before the test. Only use the detector tubes that the Tester is calibrated for. Once used, be aware that the tube tip cutter contains ground glass and glass fragments. Take appropriate precautions for the disposal of these. Dispose of as sharps. When using the tube tip cutter ensure only a light pressure is applied to the tube whilst rotating for scoring. General Make sure the system flow control valve is closed after each test. Connecting an air supply to the Tester with the system flow regulator in the open position can damage the unit. Never connect to a non-regulated supply from an HP cylinder or compressor. Maximum inlet pressure to the Tester is 10 bar. If exposed to an overpressure the Tester will display an overpressure warning necessitating its return to the manufacturer for checking and resetting. If the Tester has been used on an excessively wet supply it can be purged dry by running a full test on a known dry source without any detector tubes fitted. When using the mains power supply, ensure that the Tester is switched off prior to connection and disconnection. For any additional advice and information please contact Factair on: +44 (0)

27 Appendix 2 Draeger Tube Tip Cutter The F2187 is especially designed to prevent glass from falling out of the opener by accident. The reservoir for the broken-off tips is easy to empty. Instructions 1. Place the end of the tube between the 3 blades, and turn to score the end. 2. Push the tube at an angle to break the tip. 3. Repeat steps 1 and 2 with the other end of the tube 26

28 Appendix 3 Reading the Draeger Oil Impactor and Chemical Reagent Tubes Oil Impactor The Oil Impactor s protective seal must be kept in place for the duration of the test and only removed after the test has been completed. With a standard measurement range of 0.1 to 1.0 mg/m 3 the Impactor has a limit of detection of 0.05 mg/m 3. The Impactor can detect all mineral and synthetic oils, it features a series of 3 horizontal lines, each formed by a series of precision manufactured nozzles. These horizontal lines of nozzles are calibrated to a different concentration of oil. When air is passed through these nozzles any oil present in the air is deposited on the glass plate prior to the air being exhausted by vents around its circumference, this allows the user to easily and quickly identify minute quantities of oil. For tests that pass with an oil concentration of less than 0.05mg/m 3, the screen will remain blank. For tests above that read as follows: The bottom line represents an oil concentration of 0.1 mg/m 3. When the oil deposited forms a continuous line then the concentration is in excess of 0.1 mg/m 3. The middle line represents a concentration of 0.5 mg/m 3, again when the oil deposited forms a continuous line the concentration is in excess of 0.5 mg/m 3. The top line represents a concentration of 1.0 mg/m 3, again when this forms a continuous line the concentration is in excess of 1.0 mg/m 3. Pass - greater than 0.05mg/m 3 but less than 0.1mg/m 3 Fail / (Pass for EN12021 breathingair only) - greater than 0.1mg/m 3 but less than 0.5mg/m 3 Fail - greater than 0.5mg/m 3 but less than 1.0mg/m 3 Fail - greater than 1.0mg/m 3 27

29 Each pack of tubes has its own instruction leaflet but the following notes should help you take readings after the tests have been completed. Each tube has an expiry date which is located on the back of its storage box. Oil ( ) REQUIREMENT - THE AIR SHOULD HAVE A MAXIMUM OIL CONTENT OF 0.5 MG/M³ AND SHOULD BE WITHOUT SIGNIFICANT ODOUR OR TASTE. 1. Using the Tube Tip Cutter, place the ampoule section in the platform at the bottom. Note: Make sure the tube is against the back wall of the cutter. Line up the black dot nearest the end of the tube with dot on the cutter. 2. With one hand holding the cutter, the other holding the tube and with your thumb against the base of the cutter, apply pressure to the tube, this should break the inner tube. 3. Place your used detector tubes in the F2154 Travel Container, until they can be disposed of properly as sharps or glass. Ensure the oil tube is kept vertical throughout this process. Satisfactory test: - The white crystals will turn translucent and show at worst a slight discoloration. Failed test: - (Mineral oil) - The white crystals will show a light brown or darker discoloration. - (Synthetic oil) - The white crystals will show a yellow discoloration [Note: best seen by comparing with an unused tube]. 28

30 Carbon Dioxide (CO 2 ) ( ) REQUIREMENT - MAXIMUM READING OF 500 PPM. (300 PPM EU Pharmacopoeia N2/N 2 O) The media in the detector tube will discolour to show the presence of carbon dioxide. The total length of the discoloration read from the printed scale at that point is a measure of the concentration in parts per million. Carbon Monoxide (CO) ( ) REQUIREMENT - MAXIMUM READING OF 5 PPM. The media will discolour to show the presence of carbon monoxide in the air sample. The total length of the discoloration is the measure of concentration read directly from the scale in parts per million. Nitrous Fumes (NO x ) ( ) REQUIREMENT - MAXIMUM CONCENTRATION OF 2 PPM. The media in the detector tube will discolour to show the presence of nitrous fumes. The total length of the discoloration is the measure of concentration read directly from the scale in parts per million. 29

31 Sulphur Dioxide (SO 2 ) ( ) REQUIREMENT - MAXIMUM READING OF 1 PPM. ( 2PPM EU Pharmacopoeia in CO2) The media in the detector tube will change from grey blue to white to show the presence of sulphur dioxide. The total length of the white section is the measure of concentration read directly from the 1-25ppm scale in parts per million. Polytest (CH 28401) REQUIREMENT - NO DISCOLORATION. The media in the detector tube will discolour to provide qualitative determination of easily oxidizable substances in air. Refer to Polytest data sheets for more specific information. If discoloration occurs further investigation will be required to determine the source of the contamination. Hydrogen Sulphide (H 2 S) ( ) REQUIREMENT - MAXIMUM READING OF 1 PPM. (EU Pharmacopoeia in CO2) The media in the detector tube will change from white to pale brown to show the presence of Hydrogen Sulphide. The total length of the pale brown section is the measure of concentration read directly from the 1-15ppm scale in parts per million. 30

32 IMPORTANT IT IS RECOMMENDED THAT YOUR MULTI-AIR TESTER IS RETURNED FOR RECALIBRATION AND SERVICING WITHIN 12 MONTHS FROM THE ISSUE DATE OF ITS CALIBRATION CERTIFICATE Disposal of Draeger-Tubes When Draeger tubes have been used, or unopened tubes have exceeded their expiry date, they should be disposed of using one of the following methods: Used Tubes Submerge the tube(s) in a beaker or metal container filled with water and allow to soak for 24 hours. Treat the residual water in accordance with local authority waste regulations (some tube aqueous waste may require neutralisation prior to disposal). Place the tubes in a sharps or glass bin wearing protective gloves and safety spectacles. Dispose of the bin via the company s normal industrial waste disposal method(s) i.e. landfill or incineration. OR Place the tubes in a sharps or glass bin wearing protective gloves and safety spectacles. Dispose of the bin via the company s normal hazardous waste disposal method(s) i.e. landfill or incineration. OR Place the tubes in a sharps or glass bin wearing protective gloves and safety spectacles. Dispose of the container via incineration. Unused Tubes Open the Draeger tube at both ends using the special tube opener or the cutter on the hand pump. Break any ampoules where applicable. Dispose of the tubes as stated in Methods 1, 2 or 3. NOTE: As an alternative a local authority approved waste disposal contractor can be employed to collect used and unused tubes from site and dispose of them in a safe manner. 31

33 Appendix 4 Breathing-Air Standard BS EN12021 & EN12021:2014 Odour The gas shall be free from unsatisfactory odour or taste. Oxygen (21 ± 1)% Carbon 500 PPM Dioxide Carbon 5 PPM Monoxide Oil 0.5 mg/m 3 Water Airline <40Bar Water High Pressure Where the apparatus is used and stored at a known temperature the pressure dewpoint shall be at least 5 C below likely lowest temperature. Where the conditions of usage and storage of any compressed air supply is not known the pressure dewpoint shall not exceed -11 C. 40 to 200 bar 50 mg/m 3 >200 bar 35 mg/m 3 HP Charging Comp 25 mg/m 3 Above is an extract only for full details refer to the individual standard. 32

34 Appendix 5 - Air Quality Testing Why? The required quality of breathing air is stated in EN12021 is to provide information on the safe limits of potential contaminant gases within breathing air and to ensure that the life support gas of oxygen is of an adequate level. Compressed air for breathing normally originates from a compressor system installed or operating at the place of use and there are various factors that can affect the quality and safety of this air. The air intake to the compressor can ingest airborne contamination from local processes and vehicle exhaust fumes which are not removed by standard breathing air filtration. Such air borne contamination may not be continuous but the pollution of the air supply may persist for hours or days. Malfunctioning compressors, especially reciprocating type, can produce unsafe levels of both carbon monoxide and carbon dioxide. Breathing air filtration has a finite life and can fail causing high levels of oil and water contamination to be present in the air. The performance of desiccant filters is dramatically affected by operating temperature. Infrequent validation may result in poor quality air being supplied for an extended period Failure of the compressed air aftercooling will result in air entering the filtration at too high a temperature, this will cause the filtration to prematurely fail and pass excess levels of oil and water. Malfunctioning dryers can disturb the oxygen concentration to outside safe levels within the breathing air. High levels of water in breathing air can freeze within RPD demand valves causing the air supply to fail. Insufficient air flow or pressure to the RPD will reduce the protection factor of the RPE and potentially expose the user to ingress of external contaminants. The effects of contaminants when breathed at elevated pressure can have a much greater effect on users than it would at normal pressure. Changes in the performance of compressor and filtration equipment are usually rapid in nature. Any failure affecting outlet air quality may injure users for an extended period if quality validation is infrequent. Odour alone is a poor indicator of air quality, toxic as asphyxiant gasses are often odourless, the limits for oil pollution are lower than the threshold detection level that most people will notice. All employers have a duty of care to their employees to ensure that the breathing air they are supplied with is adequate for the RPD they are using and safe to breathe. The points raised above may form the basis of the risk assessment called for in the European guidance document for the selection and use of respiratory protective devices EN

35 Frequency of Breathing-Air Tests Low Pressure Systems The purpose of periodically testing air quality is to make sure that the control measures you have put in place are delivering the air quality required. In the UK national forward to EN12021 advises that samples should be taken and analysed at least every three months or more frequently if there has been a change in, or concerns relating to, the production process. In the HSE guideline document Respiratory Equipment at Work (HSG53) it states you should base the frequency of such tests on a risk assessment, but again they should take place at least every three months, and more often when the quality of air cannot be assured to these levels. For mobile breathing-air compressors, in the UK, COSHH stipulates that, the employer should ensure that wherever a compressor is located, the quality of air it supplies is not compromised by nearby contaminants. We strongly therefore recommend that for mobile compressors the air quality is tested whenever it is first moved into a new position or prevailing wind conditions change. The final decision on frequency of test is the responsibility of the Employer and needs to not only reflect local legislation but also the task and frequency of use. It should be incorporated into their risk assessment and updated regularly to reflect results from ongoing breathing-air tests to maintain a robust control system. High Pressure Systems Whilst up to a 3 month periodicity for testing may be suitable for low pressure breathing-air systems, further consideration should be given to high pressure compressors where the life of filter elements are normally much shorter, typically 50 hours for a HP filter cartridge and this is reduced further in high ambient temperatures. Accordingly for these systems we recommend that your risk assessment for testing HP systems should be based on the expected filter life and an air quality test should first be completed when new filter elements are fitted and then again when they reach 50% of their life based on the hours run usage. Subsequent tests would then be dependent on usage with a maximum interval between tests of 3 months. 34

36 Recording Breathing-Air Quality Test Results When undertaking breathing-air quality tests, results should be retained. In the UK the regulations COSHH stipulates the information retained should include: the name and address of the employer responsible for the RPE; particulars of the equipment and of the distinguishing number or mark, together with a description sufficient to identify it, and the name of the maker; the date of examination and the name and signature or other acceptable means of identifying the person carrying out the examination and test; the condition of the equipment and details of any defect found, including for canister or filter respirators, the state of the canister and the condition of the filter; for self-contained compressed air/gas breathing apparatus, the pressure of air/gas in the supply cylinder; and for powered/power-assisted respirators and breathing apparatus, the volume flow rate to ensure that they can deliver at least the manufacturer's minimum recommended flow rate. Records can be in paper or electronic format but should be kept readily accessible and retrievable at any reasonable time for examination by safety representatives or inspectors etc. In the guideline document Respiratory Protective Equipment, a practical guide (HSG53) it advises results should be kept for 5 years. 35

37 A.4.5 Compressed air for breathing apparatus from EN529 A.4.5 Compressed air for breathing apparatus (EN12021) A General A compressor system will have produced the compressed air supplied to a breathing apparatus. The compressor system may be used for filling individual high-pressure pressure vessels or those on a mobile trolley or to supply air direct to breathing apparatus and other air-tools used in the workplace. Contaminants can mix in compressed air at various stages of its production and supply. Any presence of contaminants in acceptable quantities will render the air unsuitable as breathable air and can threaten the health and safety of the respiratory protective device wearer. For this reason quality assured compressed air should be supplied to a breathing apparatus. EN12021 stipulates the minimum quality standards for breathable compressed air and includes the levels for oxygen, carbon monoxide, carbon dioxide, lubricants, water and other types of contaminant and odour. A Compressor system A General A competent person should be consulted when planning or installing a compressed air system for producing breathable air. This will help to minimise problems associated with compressors and the down stream effects on the quality of the air supplied. Table A.2 provides a summary of the main elements associated with a compressor system for producing breathable air. In addition to the careful and installation of the system it should be maintained by a competent person to ensure the safe operation of the system. The compressor should be installed in an area providing sufficient space on all side to ensure good ventilation. The area should be cool as possible but avoid place where freezing is possible. The air intake point should be located in open air and away from potential contaminants (e.g. not close to ventilation outlets or in down stream of the outlets or near vehicle exhaust emission points). A Air purification elements The air purification elements should be placed in the correct sequence to ensure the delivery of acceptable quality breathing air. These purification elements should be replaced in accordance with the advice provided by the competent person and the manufacturers of these elements. A Testing and inspection The volume flow and quality of the supplied air should be thoroughly tested as specified by a competent person after risk assessment. Permission to reproduce extracts of EN529 is granted by BSI. British Standards can be obtained from BSI Customer Services, 389 Chiswick High Road, London W4 4AL. Tel: +44 (0) cservices@bsi-global.com Contains public sector information licensed under the Open Government Licence v3.0 36

38 COSHH L5 (Sixth Edition 2013) EXTRACTS FROM CODE OF PRACTICE RELATING TO RESPIRATORY PROTECTIVE EQUIPMENT (RPE) 178. The maintenance, examination and tests should be in accordance with the manufacturer's instructions. Examinations should comprise a thorough visual examination of all parts of the respirator or breathing apparatus, to ensure that all parts are present, correctly fltted, and the equipment is in good working order. In particular, the examination should ensure that the straps, facepieces, filters and valves are sound and in good working condition. For powered and power-assisted respirators, tests should: be made on the condition and efficiency of those parts; ensure that the battery pack is in good condition; ensure that the respirator delivers at least the manufacturer's recommended minimum volume flow rate For RPE incorporating compressed gas cylinders, tests should include the condition and efficiency of all parts, the pressure in the cylinders and the volume flow rate. Frequency of examination and tests 180. The quality of the air supplied to a breathing apparatus should be tested at suitable intervals, depending on the task and the frequency of use. When the air supply is from mobile compressors, the employer should ensure that wherever a compressor is located, the quality of air it supplies is not compromised by nearby contaminants. In every case, the air supplied to a breathing apparatus should meet the relevant quality standard. As it is not reasonably practicable to test for all contaminants, the risk assessment made under regulation 6 should guide what other contaminants will require testing Thorough maintenance examinations and, where appropriate, tests of items of RPE, other than one-shift disposable respirators, should be made at suitable intervals. The frequency should increase where the health risks and conditions of exposure are particularly severe In situations where respirators are used only occasionally, an examination and test should be made before their next use and maintenance carried out as appropriate. The person who is responsible for managing the maintenance of RPE should determine suitable intervals between examinations. Emergency escape-type RPE should be examined and tested in accordance with the manufacturer's instructions Suitable arrangements should be made to ensure that no employee uses RPE which has previously been used by another person, unless it has been thoroughly washed and cleaned in accordance with the manufacturer's instructions. 37

39 Suitable records 184. The record of each thorough examination and test of RPE carried out should include: the name and address of the employer responsible for the RPE; particulars of the equipment and of the distinguishing number or mark, together with a description sufficient to identify it, and the name of the maker; the date of examination and the name and signature or other acceptable means of identifying the person carrying out the examination and test; the condition of the equipment and details of any defect found, including for canister or filter respirators, the state of the canister and the condition of the filter; for self-contained compressed air/gas breathing apparatus, the pressure of air/gas in the supply cylinder; and for powered/power-assisted respirators and breathing apparatus, the volume flow rate to ensure that they can deliver at least the manufacturer's minimum recommended flow rate. Keeping records 185. Employers may keep records in any format, eg on paper or electronically. Records should be kept readily accessible and retrievable at any reasonable time for examination by safety representatives or inspectors etc. Accommodation for, and checking of, PPE 186. Employers should ensure that accommodation is provided for PPE so that it can be safely stored or kept when it is not in use. The adequacy of the accommodation will vary according to the quantity, type and its use, eg pegs, (labelled) lockers, shelves or containers etc. The storage should be adequate to protect the PPE from contamination, loss or damage by, for example, harmful substances, damp or sunlight. Where quantities of PPE are stored, equipment which is ready for use should be clearly segregated from that which is awaiting repair or maintenance. Where PPE becomes contaminated during use, and especially by biological agents, the accommodation should be separate from any the employer provides for ordinary clothing and equipment. Employers may also have duties under the Workplace (Health, Safety and Welfare) Regulations 1992 to provide accommodation for PPE All PPE should be checked regularly to ensure that it continues to function and provide protection. The types of checks should be suited to that item of PPE and be able to detect significant deterioration. The more likely the performance of a particular item of PPE is to deteriorate, the more often it needs checking. Whoever does this work should be sufficiently knowledgeable and trained to identify deterioration and significant faults. Equipment that has deteriorated significantly or is faulty should be effectively repaired or disposed of safely. Reproduced from COSHH Approved Codes of Practice L5 by Factair Ltd with the permission of the Controller of HMSO. Contains public sector information licensed under the Open Government Licence v3.0 38

40 Dos and don ts Respiratory Protective Equipment at Work A practical guide HSG53 (Fourth edition, published 2013) Do s Always ensure the breathing apparatus is in good working order before putting it on, even when new. Always look after your supply hose during use your life may depend on it. Always use all the straps provided, making sure they are correctly positioned and adjusted. Follow the manufacturer s instructions. Ensure that an adequate clean air supply is available for all users. Don t s Never place the hose inlet near to potential sources of contamination, eg vehicle exhausts. Never use the equipment without the waist belt. Never use a light-duty airline hose where there is any potential for crushing by vehicles or passers-by etc. Never keep working if the airflow rate drops or any warning devices are activated. Leave the work area immediately. Ensure that the compressed air quality meets the minimum requirements of BS EN Always plan your exit from the contaminated area so you don't run out of air. Ensure the other PPE you need to wear is compatible with the BA. Appendix 3 Quality of air for breathing apparatus Air quality 1. Air supplied to breathing apparatus (BA) should be clean and safe to breathe, whether it is supplied via a fresh air hose or a source of compressed air. Fresh air hose You should securely anchor the inlet for fresh air hose BA in an area that is free of contaminant. This can usually be achieved by siting the inlet well away from the work area (eg in free air outside the building), and upwind of any local sources of airborne contamination (eg vehicle exhaust). 39

41 Compressed air 2. Compressed air for BA normally originates from a compressor system. The maintenance, examination and testing of compressors should be carried out according to the manufacturer s instructions. The siting of air inlets to compressors should follow the same principles as for fresh air hose. However, because compressors themselves can generate and concentrate a wide range of contaminants, you should take extra care in assuring air quality. 3. As the BA wearer s life and health depend on the air supplied by the compressor, you should ensure that the air supplied meets the quality requirements in British Standard BS EN Respiratory protective devices. Compressed air for breathing apparatus,* in addition to the pressure and airflow rate requirements of the BA manufacturer. 4. Compressors which are moved from site to site, such as those used by the emergency services or on construction sites, will require a higher standard of maintenance and should be sited so that the quality of air they provide is not compromised by nearby contaminants. * BS EN states: Compressed air for breathing apparatus shall not contain any contaminants at a concentration which can cause toxic or harmful effects. In any event all contaminants shall be kept to as low a level as possible and shall be far below the national exposure limit. Combination effects of more than one contaminant shall be taken into account. (1999)Respiratory protective equipment at work Page 48 of 59 Periodic testing of air quality 5. The purpose of periodically testing air quality is to make sure that the control measures you have put in place are delivering the air quality required by BS EN You should base the frequency of such tests on a risk assessment, but they should take place at least every three months, and more often when the quality of air cannot be assured to these levels. 6. As part of the risk assessment, if a mobile compressor is being used consideration should be given as to how often the air supply should be checked when the compressor is moved. Testing for these components may be carried out using any appropriate method, eg: simple colour change tubes; on-line gas testers; sample collection for laboratory analysis elsewhere. 7. The supplier of your compressor or BA should be able to advise you on the best method for you. You should keep records of air quality tests for five years. Crown copyright material is reproduced with the permission of the Controller of HMSO and Queen s Printer for Scotland. Source Acknowledgement: HSG53 Respiratory Protective Equipment at Work ISBN Health and Safety Executive 2013 Contains public sector information licensed under the Open Government Licence v3.0 40

42 Appendix 6 - Medical Air Standards HTM02-01 Medical & Surgical Air Dental Air Synthetic Air Oil 0.1 mg/m mg/m 3 Water 67 vpm ( 0.05 mg/l, 1020 vpm ( 0.78 mg/l, 67 vpm ( 0.05 mg/l, 50mg/m 3, atmospheric 78mg/m 3, atmospheric 50mg/m 3, atmospheric dewpoint -46 C) dewpoint -20 C) dewpoint -46 C) Carbon 5 mg/m 3 5 PPM v/v 5 mg/m 3 5 PPM v/v Monoxide Carbon 900 mg/m PPM 900 mg/m PPM v/v Dioxide v/v Carbon 5 PPM 10 PPM Monoxide NO and 2 PPM v/v 2 PPM v/v NO 2 SO 2 1 PPM v/v 1 PPM v/v Polytest No discoloration No discoloration No discoloration (Optional) Odour None None Note There is also a requirement to conduct a visual particulate test and this should be free from visible particles in a 75 L sample (175 L for surgical air) Above is an extract only for full details refer to the HTM02-01 standard. European Pharmacopoeia 5.0 Air, Medicinal N 2 O CO 2 N 2 Oxygen 20.4% v/v to 21.4% v/v Oil 0.1 mg/m 3 Water 67 vpm ( 0.05 mg/l, 60 vpm ( vpm ( vpm ( mg/m 3, atmospheric mg/m 3 ) mg/m 3 ) mg/m 3 ) dewpoint -46 C) Carbon Monoxide 5 PPM v/v 5 PPM v/v 5 PPM v/v Carbon 500 PPM v/v 300 PPM v/v 300 PPM v/v Dioxide Carbon Monoxide 5 PPM v/v 5 PPM v/v 5 PPM v/v NO and 2 PPM v/v 2 PPM v/v 2 PPM v/v NO 2 SO 2 1 PPM v/v 2 PPM v/v H 2 S 1 PPM v/v Odour None 41

43 Water mg/m3 to PPM v/v Conversion Charts 42

44 Factair Ltd 49 Boss Hall Road Ipswich Suffolk IP1 5BN UK Tel Sales: +44 (0) Tel Hire: +44 (0) Fax: +44 (0)

F4500 SAFE-AIR TESTER OPERATING MANUAL

F4500 SAFE-AIR TESTER OPERATING MANUAL F4500 SAFE-AIR TESTER OPERATING MANUAL FAC QAM 181 Issue A 07/16 1 Contents Page No Introduction 3 Calibration & Warranty 4 Equipment List 4 1 - Pre Use 5 2 - Start-Up 6 3 - Selecting the test standard

More information

F4501 & F4501ED SAFE-AIR TESTER OPERATING MANUAL

F4501 & F4501ED SAFE-AIR TESTER OPERATING MANUAL F4501 & F4501ED SAFE-AIR TESTER OPERATING MANUAL FAC QAM 190 Issue D 09/18 1 FAC QAM 190 Issue D 09/18 Contents Page No Introduction 3 Calibration & Warranty 3 Equipment List 4 1 - Pre Use 5 2 - Start-Up

More information

F4000 & F4000ED SAFE-AIR TESTER OPERATING MANUAL

F4000 & F4000ED SAFE-AIR TESTER OPERATING MANUAL F4000 & F4000ED SAFE-AIR TESTER OPERATING MANUAL FAC QAM 093 Issue B 03/15 0 1 Contents Page No Introduction 3 Calibration & Warranty 3 Diagram and Equipment List 4 1 - Pre Use 5 2 - Testing- Airline up

More information

F6000 SAFE-AIR TESTER OPERATING MANUAL

F6000 SAFE-AIR TESTER OPERATING MANUAL F6000 SAFE-AIR TESTER OPERATING MANUAL FAC QAM 094 Issue A 09/13 0 1 Contents Page No Introduction 3 Calibration & Warranty 3 Diagram and Equipment List 4 1 - Pre Use 5 2 - Start Up 6 3 - Beginning a new

More information

F4001 & F4001ED SAFE-AIR TESTER OPERATING MANUAL

F4001 & F4001ED SAFE-AIR TESTER OPERATING MANUAL F4001 & F4001ED SAFE-AIR TESTER OPERATING MANUAL FAC QAM 094 Issue B 03/15 0 1 Contents Page No Introduction 3 Calibration & Warranty 3 Diagram and Equipment List 4 1 - Pre Use 5 2 - Testing- High Pressure

More information

F4001 & F4001ED SAFE-AIR TESTER OPERATING MANUAL

F4001 & F4001ED SAFE-AIR TESTER OPERATING MANUAL F4001 & F4001ED SAFE-AIR TESTER OPERATING MANUAL 0 1 Contents Page No Introduction 3 Calibration & Warranty 3 Diagram and Equipment List 4 1 - Pre Use 5 2 - Testing- High Pressure systems 200 & 300 bar

More information

AN INTRODUCTION TO COMPRESSED AIR QUALITY for BREATHABLE PURPOSES

AN INTRODUCTION TO COMPRESSED AIR QUALITY for BREATHABLE PURPOSES AN INTRODUCTION TO COMPRESSED AIR QUALITY for BREATHABLE PURPOSES 1 of 1 Introduction Supplied air breathing apparatus systems require a breathable compressed air source to feed the equipment. The quality

More information

A. Evaluation of the written program

A. Evaluation of the written program A. Evaluation of the written program Begin your audit with an evaluation of the written respiratory protection program. In addition to the operating procedures describing respirator selection and use in

More information

BGC HOLDINGS LTD RESPIRATORY PROTECTIVE EQUIPMENT (RPE) POLICY

BGC HOLDINGS LTD RESPIRATORY PROTECTIVE EQUIPMENT (RPE) POLICY BGC HOLDINGS LTD RESPIRATORY PROTECTIVE EQUIPMENT (RPE) POLICY Introduction This document sets out the Health & Safety arrangements we have established for BGC Holdings Ltd, and all subsidiary and associated

More information

CONTROL OF SUBSTANCES HAZARDOUS TO HEALTH PROCEDURE

CONTROL OF SUBSTANCES HAZARDOUS TO HEALTH PROCEDURE CONTROL OF SUBSTANCES HAZARDOUS TO HEALTH PROCEDURE 1. INTRODUCTION The University uses substances hazardous to health in many of its operations, and accepts that it has a duty of care to reduce exposure

More information

PremAire INSTRUCTIONS FOR VORTEX TUBE MODE OF OPERATION

PremAire INSTRUCTIONS FOR VORTEX TUBE MODE OF OPERATION DUAL SUPPLY REGULATOR ESCAPE CYLINDER VORTEX MAIN INLET PremAire INSTRUCTIONS FOR VORTEX TUBE MODE OF OPERATION WARNING THIS MANUAL MUST BE READ CAREFULLY BY ALL PERSONS WHO HAVE OR WILL HAVE THE RESPONSIBILITY

More information

Niltox Breathing Air Purifier Range

Niltox Breathing Air Purifier Range Breathing Air Requirements The main standards affecting breathing air are BSEN529, EN12021 and COSHH. The requirements for a breathing air system are shown in the table opposite and the requirements for

More information

Pressure Automated Calibration Equipment

Pressure Automated Calibration Equipment GE Measurement & control Pressure Automated Calibration Equipment Safety Instructions and User Guide - K0447 PACE5000 PACE6000 K0447 Issue No. 9 1 10 1 PACE5000 1 2 3 4 5 PACE6000 2 6 7 8 3 4 5 6 7 8 9

More information

GSA POLICY ON LOCAL EXHAUST AND OTHER VENTILATION (FUME CUPBOARDS)

GSA POLICY ON LOCAL EXHAUST AND OTHER VENTILATION (FUME CUPBOARDS) GSA POLICY ON LOCAL EXHAUST AND OTHER VENTILATION (FUME CUPBOARDS) Local Exhaust Ventilation Inspection & Maintenance 1 LOCAL EXHAUST AND OTHER VENTILATION Contents 1. Introduction 2. Design specifications

More information

COSHH Hazardous Substances Health & Safety Management Standard Issue 2 (September, 2010)

COSHH Hazardous Substances Health & Safety Management Standard Issue 2 (September, 2010) COSHH Hazardous Substances Health & Safety Management Standard Issue 2 (September, 2010) Introduction This document contains corporate standards and guidance on the management of risks arising from the

More information

TOXIC GAS DETECTORS IN THE WORKPLACE UK REGULATIONS AND STANDARDS

TOXIC GAS DETECTORS IN THE WORKPLACE UK REGULATIONS AND STANDARDS SUMMARY TOXIC GAS DETECTORS IN THE WORKPLACE UK REGULATIONS AND STANDARDS In recent articles supplied by the Council of Gas Detection and Environmental Monitoring (CoGDEM) for inclusion in International

More information

Section1. Knowledge of RPE Sets out the minimum knowledge requirements in respect of respiratory protective equipment. [Mandatory] 1.

Section1. Knowledge of RPE Sets out the minimum knowledge requirements in respect of respiratory protective equipment. [Mandatory] 1. Subject / Content Section1. Knowledge of RPE Sets out the minimum knowledge requirements in respect of respiratory protective equipment. [Mandatory] 1.1 A general understanding of RPE use in the workplace

More information

Laboratory Self Inspection Checklist LAB01

Laboratory Self Inspection Checklist LAB01 Laboratory Self Inspection Checklist LAB01 General Chemical/Biological Laboratories Location Assessor Date This core checklist is provided by the Health and Safety Department as a template, which provides

More information

SAFETY TRAINING LEAFLET 06 CARBON DIOXIDE

SAFETY TRAINING LEAFLET 06 CARBON DIOXIDE SAFETY TRAINING LEAFLET 06 CARBON DIOXIDE Doc 23.06/18 EUROPEAN INDUSTRIAL GASES ASSOCIATION AISBL AVENUE DES ARTS 3-5 B 1210 BRUSSELS Tel: +32 2 217 70 98 Fax: +32 2 219 85 14 E-mail: info@eiga.eu Internet:

More information

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4 Calibration....

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS euro-gas@btconnect.com www.euro-gasman.com AIRCHECK OPERATING MANUAL By carefully following these instructions, the AIRCHECK with KITAGAWA gas detection tubes will enable you to

More information

Respiratory Protective Equipment

Respiratory Protective Equipment Page 1 of 5 Health and Safety Fact File - Protective Clothing Respiratory Protective Equipment Respiratory protective equipment (RPE) selected for compliance with the COSHH regulations must conform to

More information

RPE with a UK Standard Assigned Protection Factor 10 (APF10)

RPE with a UK Standard Assigned Protection Factor 10 (APF10) Health and Safety Executive R COSHH essentials: Supplementary advice RPE with a UK Standard Assigned Protection Factor 10 (APF10) Control approach R Respiratory protective equipment (RPE) The Control of

More information

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet: OXY Integral INTERCON ENTERPRISES INC Tel: 800 665 6655 Fax: 604 946 5340 E-Mail: sales@intercononline.com Internet: www.intercononline.com Manual Integral 2006 1 INDEX 2-3 PREFACE 4 INTRODUCTION 5 Principle

More information

Breathable Compressed Air

Breathable Compressed Air aerospace climate control electromechanical filtration fluid & gas handling hydraulics pneumatics process control sealing & shielding Breathable Compressed Air Typical Hazardous Substances Biological agents

More information

Respiratory Protection Self Inspection Checklist

Respiratory Protection Self Inspection Checklist Optional Information Name of School: Date of Inspection: Vocational Program/Course/Room: Signature of Inspector: Respiratory Protection Self Inspection Checklist Guidelines: This checklist covers some

More information

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021-PR Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021-PR For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4

More information

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments Pegas 4000 MF Gas Mixer InstructionManual Contents I Table of Contents Foreword Part I Introduction 1 2 1 System overview... 2 2 Specifications... 3 Part II Installation 4 1 Rear panel connections...

More information

Respiratory Protection

Respiratory Protection Respiratory Protection Purpose The purpose is to establish a set of guidelines for the selection of respiratory protection equipment, situations for its use, and training in the use of the respirator required.

More information

Peninsula Dental Social Enterprise (PDSE)

Peninsula Dental Social Enterprise (PDSE) Peninsula Dental Social Enterprise (PDSE) Medical Gas Policy Version 3.0 Date approved: August 2018 Approved by: The Board Review due: August 2021 Policy will be updated as required in response to a change

More information

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4 Calibration....

More information

RAM 4021 Operation Manual

RAM 4021 Operation Manual RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For your safety...3 Description...3 Set-up mode...4 Annunciator lights/alarms...4 Operation...5 Calibration...6

More information

RAM 4021-DPX Operation Manual

RAM 4021-DPX Operation Manual RAM 4021-DPX Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS ABL 4021-DPX / RAM 4021-DPX For Your Safety... 3 Description... 3 Setup Mode... 4 Lights/Alarms... 4 Operation...

More information

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008 UNITY 2 TM Air Server Series 2 Operators Manual Version 1.0 February 2008 1. Introduction to the Air Server Accessory for UNITY 2...2 1.1. Summary of Operation...2 2. Developing a UNITY 2-Air Server method

More information

VERTICAL AIR COMPRESSORS

VERTICAL AIR COMPRESSORS VERTICAL AIR COMPRESSORS MODEL NO: VE11C150, VE15C150, VE18C150 PART NO: 2226005, 2226000, 2226015 OPERATION & MAINTENANCE INSTRUCTIONS LS0615 INTRODUCTION Thank you for purchasing this CLARKE Vertical

More information

Key Technology, Inc. Confined Space Entry Program. July, 2017

Key Technology, Inc. Confined Space Entry Program. July, 2017 Key Technology, Inc. Confined Space Entry Program July, 2017 Page 1 of 11 Key Technology, Inc. CONFINED SPACE ENTRY PROGRAM OVERVIEW Purpose: The purpose of this program is to ensure the protection of

More information

Working Environment Safety Assessment

Working Environment Safety Assessment For your own safety and/or safety of your workforce, and to improve the efficiency of your business practices, you should be aware of, and be up to date with, the key safety issues and good practices relevant

More information

Respiratory Protection Self-Inspection Checklist

Respiratory Protection Self-Inspection Checklist Respiratory Protection Self-Inspection Checklist Guidelines: This checklist applies to the use of either atmosphere-supplying or air-purifying respirators being worn, voluntarily or otherwise, for comfort

More information

Vertex VP Operation and Maintenance

Vertex VP Operation and Maintenance AIRBENCH LTD Vertex VP Operation and Maintenance This Edition: 2 October 2018 AirBench Ltd, 14 Grange Farm Road, Colchester CO2 8JW www.airbench.com +44 (0) 1206 791191 Copyright AirBench Ltd 2016 Page

More information

Texas Department of Insurance. Respiratory Protection. Provided by. Workers Health & Safety HS01-027B (1-06)

Texas Department of Insurance. Respiratory Protection. Provided by. Workers Health & Safety HS01-027B (1-06) Texas Department of Insurance Respiratory Protection Provided by Workers Health & Safety HS01-027B (1-06) Contents Introduction... 3 Definitions... 3 Approved Respirators... 3-4 New Respirator Selection

More information

ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS

ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS Annex 3, page 2 ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS The text of existing chapter 15 is replaced by the following: "1 Application This

More information

PERSONAL AIR BREATHING UNIT

PERSONAL AIR BREATHING UNIT PERSONAL AIR BREATHING UNIT Operation & Maintenance Manual MARTECH SERVICES C O M P A N Y 1-800-831-1525 Table of Contents INTRODUCTION: Page 2 COMPONENTS DRAWING: Page 3 START UP: Page 4 OPERATION: Page

More information

RAM Operation Manual

RAM Operation Manual RAM 4021-1 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021-1 For Your Safety... 2 Description... 2 Setup Mode... 3 Lights/Alarms... 3 Operation... 4 Calibration...

More information

VERTICAL AIR COMPRESSORS

VERTICAL AIR COMPRESSORS VERTICAL AIR COMPRESSORS MODEL NO: VE15C150, VE18C150, VE25C150 PART NO: 2226010, 2226020, 2226025 OPERATION & MAINTENANCE INSTRUCTIONS LS0715 INTRODUCTION Thank you for purchasing this CLARKE Vertical

More information

Oxygen Dialflow Meter. Instructions for Use

Oxygen Dialflow Meter. Instructions for Use Oxygen Dialflow Meter Instructions for Use 702-0031.12 December 2017 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in injury to the patient,

More information

OPERATION MANUAL NTF-15

OPERATION MANUAL NTF-15 OPERATION MANUAL NTF-15 Nitrogen Tire Filling Valve Stem Caps (Qty=200) Order P/N 436075 RTI Technologies, Inc 10 Innovation Drive York, PA 17402 800-468-2321 www.rtitech.com 035-81235-00 (Rev B) TABLE

More information

Instruction Manual. Model DSP-Ex Portable Dewpoint Meter. Alpha Moisture Systems Alpha House 96 City Road Bradford BD8 8ES England

Instruction Manual. Model DSP-Ex Portable Dewpoint Meter. Alpha Moisture Systems Alpha House 96 City Road Bradford BD8 8ES England Model DSP-Ex Portable Dewpoint Meter Instruction Manual Alpha Moisture Systems Alpha House 96 City Road Bradford BD8 8ES England Tel: +44 1274 733100 Fax: +44 1274 733200 Email: info@amsytems.co.uk Web:

More information

SUBSTATION MAINTENANCE ELECTRICAL OPERATING PROCEDURE

SUBSTATION MAINTENANCE ELECTRICAL OPERATING PROCEDURE Page 1 of 6 A. PURPOSE AND SCOPE This procedure establishes safe working practices for handling sulfur hexafluoride (SF 6 ) gas and release minimization practices. This document also specifies methods

More information

Pressure Systems Safety Regulation

Pressure Systems Safety Regulation Pressure Systems Safety Regulation Introduction This document informs Faculty of the key requirements of the UK and Chinese Pressure Systems Safety regulations. The aim of these regulations is to prevent

More information

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer July 2011 OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer WARNING! Before operating this product, read and understand this Operator s Manual. Become familiar with the potential hazards of this unit. Contact

More information

Oxygen Dialflow Meter. Instructions for Use

Oxygen Dialflow Meter. Instructions for Use Oxygen Dialflow Meter Instructions for Use 702-0031.9 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

Fire Escape Hood OPERATION AND INSTRUCTIONS

Fire Escape Hood OPERATION AND INSTRUCTIONS TM Fire Escape Hood OPERATION AND INSTRUCTIONS THIS MANUAL MUST BE CAREFULLY READ AND FOLLOWED BY ALL PERSONS WHO HAVE, OR WILL HAVE, THE RESPONSIBILITY FOR USING THE S-CAP FIRE ESCAPE HOOD. THESE ESCAPE

More information

Manual Actuated Boiler Blowdown Valves

Manual Actuated Boiler Blowdown Valves Manual Actuated Boiler Blowdown Valves Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation 4. Operation 5. Maintenance 6. Spare parts p.1 1. Safety

More information

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor INSTALLATION and INSTRUCTION MANUAL for QUALITY AIR BREATHING SYSTEMS Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor M A R T E C H S E R V I C E S C O M P A N Y P.O. Box 7079 OFFICE: (507)

More information

SEMEM3-73 Maintaining medical gas pipeline systems and equipment

SEMEM3-73 Maintaining medical gas pipeline systems and equipment Maintaining medical gas pipeline systems and equipment Overview This unit identifies the competences you need to carry out maintenance activities on medical gas pipeline systems and equipment, in accordance

More information

Training. Testor Training Manual

Training. Testor Training Manual Training Testor Training Manual Index Section 1 Introduction and Safety Warnings Section 2 Test Procedures Section 3 Test Hoses Section 4 Fault Location 1:1 1.1 Introduction The Dräger Testor test equipment

More information

CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL Company Registered Office: Crown House, Stockport, Cheshire, SK13RB No. 850 0700 67 Registered in England and Wales No. 05058855

More information

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems Corporation Issued: 5/09 P/N: 600104 Rev. 2 Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems The Gas Addition Accessory permits the

More information

ST53SD. Self Contained Breathing Apparatus. User Manual

ST53SD. Self Contained Breathing Apparatus. User Manual Self Contained Breathing Apparatus User Manual User Manual for the (Short Duration) Self Contained Breathing Apparatus WARNING Disassembly of the components beyond the procedures described herein shall

More information

Working in Gas Hazard Areas

Working in Gas Hazard Areas / Health & Safety OPERATIONS STANDARDS Ref.: Aperam Safety ST 012 Working in Gas Hazard Areas Issue: 15.05.2012 Vers.: v0 Rev.: AM Safety ST 012 v1.2 08.05.2012 Working in Gas Hazard Areas Controlled Updated

More information

High Pressure Chem-SCAN Operating Manual

High Pressure Chem-SCAN Operating Manual GAS INLET VALVES REACTIVE GAS PRESSURE RELIEF VALVE INERT GAS VENT VENT VALVE REACTOR INLET VALVES PRESSURE TRANSDUCERS REACTORS STIRRER & THERMOWELL HEATING JACKET STIRRER MOTORS High Pressure Chem-SCAN

More information

TExT Thorough Examination & Test of LEV Local Exhaust Ventilation Systems

TExT Thorough Examination & Test of LEV Local Exhaust Ventilation Systems ILEVE -Institute of LEV Engineers CPD & Information Day TExT Thorough Examination & Test of LEV Local Exhaust Ventilation Systems Jane Bastow CMIOSH Chartered Safety and Health Practitioner OSCHR Registered

More information

The Airfield Volunteer Fire Service

The Airfield Volunteer Fire Service The Airfield Volunteer Fire Service Breathing Apparatus Deployment Policy Policy Number: P23 Origin: KP Date: 25/06/2015 1 Contents Breathing Apparatus Deployment Policy Page Contents 3 Introduction 3

More information

Installation Instructions and User Guide 15mm & 22mm Thermostatic Mixing Valve

Installation Instructions and User Guide 15mm & 22mm Thermostatic Mixing Valve Installation Instructions and User Guide 15mm & 22mm Thermostatic Mixing Valve TMV3 / TMV2 Combined Valve C85079 C85081 C85080 C85082 It is important that these guidance notes are read and fully understood

More information

PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL Company Registered Office: Crown House, Stockport, Cheshire, SK13RB No. 850 0700 67 Registered in England and Wales No. 05058855

More information

SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL

SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL www.modsafe.com SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL WARNING: Do not attempt to operate this equipment without first reading and understanding the service manual enclosed with this device.

More information

Micro Dial-Flowmeter. Instructions for Use

Micro Dial-Flowmeter. Instructions for Use Micro Dial-Flowmeter Instructions for Use 702-0082.8 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

SERVICE MANUAL MODEL BA100BMST 100 SCFM PORTABLE AIR PANEL

SERVICE MANUAL MODEL BA100BMST 100 SCFM PORTABLE AIR PANEL Product Manual MST Portable Air Filtration System SERVICE MANUAL MODEL BA100BMST 100 SCFM PORTABLE AIR PANEL WARNING: Do not attempt to operate this equipment without first reading and understanding the

More information

Hydrogen Cyanide OC 275/3. Target Audience:

Hydrogen Cyanide OC 275/3. Target Audience: Hydrogen Cyanide OC 275/3 Target Audience: Agricultural and Factory Inspectors FCG Specialist Inspectors (Occ Hyg) Senior/Employment Medical Advisers Employment Nursing Advisers Date issued: 1990-11-22

More information

UsER manual for Watersens ph -REDOX

UsER manual for Watersens ph -REDOX UsER manual for Watersens -REDOX Cl 8 1 2 6 3 3 7 7 4 4 4 4 Parts List 1 Redox Probe 1 x 2 PH Probe 1 x 5 Tube Weight 2 x 6 Connection Valve 1 x chlorine 3 Chlorine and Pumps 2 x 7 Dosing Valve 2 x 5 5

More information

SA121, SA122, SA123, SA128 and SA1219 Self-acting Temperature Control Systems (Dial Adjustment)

SA121, SA122, SA123, SA128 and SA1219 Self-acting Temperature Control Systems (Dial Adjustment) 3820050/4 IM-P382-01 CH Issue 4 SA121, SA122, SA123, SA128 and SA1219 Self-acting Temperature Control Systems (Dial Adjustment) Installation and Maintenance Instructions 1. Safety information 2. Use 3.

More information

PSSI 36 General Confined Spaces, Tunnels, Culverts and Similar Spaces

PSSI 36 General Confined Spaces, Tunnels, Culverts and Similar Spaces 1. SCOPE This document now includes provision for safe access to cable tunnels and culverts previously covered in PSSI 37 which is now withdrawn. The term Confined Space is defined in the Confined Spaces

More information

Theatre Control Panel. Instructions for Use. January 2017 ASF-29 (1)

Theatre Control Panel. Instructions for Use. January 2017 ASF-29 (1) Theatre Control Panel Instructions for Use January 2017 0197 ASF-29 (1) Contents 1. Instructions For Use 1.1. Introduction 1.2. Explanation of Symbols 1.3. General Safety Instructions Intended Use Operating

More information

Model PDT Dewpoint Transmitter

Model PDT Dewpoint Transmitter Model PDT Dewpoint Transmitter Instruction Manual Alpha Moisture Systems Alpha House 96 City Road Bradford BD8 8ES England Tel: +44 1274 733100 Fax: +44 1274 733200 email: mail@amsytems.co.uk web: www.amsystems.co.uk

More information

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc. Advanced Calibration Designs, Inc. Calibration Gas Instrument INSTRUCTION MANUAL Release I www.goacd.com Instruction Manual Gas Generator Release I TABLE OF CONTENTS I. General Description Page 2 II. Start-Up

More information

How to calibrate the formaldehyde sensor within a PPMonitor Wireless Unit:

How to calibrate the formaldehyde sensor within a PPMonitor Wireless Unit: How to calibrate the formaldehyde sensor within a PPMonitor Wireless Unit: Please read this document thoroughly before attempting to check or adjust the calibration of the formaldehyde sensor within the

More information

OPERATION MANUAL NTF-60 Plus

OPERATION MANUAL NTF-60 Plus OPERATION MANUAL NTF-60 Plus Nitrogen Tire Filling Valve Stem Caps (Qty=200) Order P/N 436075 RTI Technologies, Inc 10 Innovation Drive York, PA 17402 800-468-2321 www.rtitech.com 035-81264-00 (Rev A)

More information

Breathing Air Purifiers. Comply with European Pharmacopoeia and International Breathing Air Standards

Breathing Air Purifiers. Comply with European Pharmacopoeia and International Breathing Air Standards Breathing Air Purifiers Comply with European Pharmacopoeia and International Breathing Air Standards 1 Breathing Air Purifiers domnick hunter Breathing Air Purifiers provide air 1,000,000 times cleaner

More information

973-SF 6 Analyzer. Precise and Stable SF 6 Gas Analyzer REFLECTING YOUR STANDARDS

973-SF 6 Analyzer. Precise and Stable SF 6 Gas Analyzer REFLECTING YOUR STANDARDS Precise and Stable SF 6 Gas Analyzer Simultaneous measurement of humidity, SF 6 purity and SO 2 concentration Integrated gas recovery system with automatic pump back Fully automated SF 6 gas testing Fundamental

More information

Airtester HP for compressed air according to EN 12021

Airtester HP for compressed air according to EN 12021 Airtester HP for compressed air according to EN 12021 Instruction for use Part Number D3188001 (Edition Dec. 2003) TABLE OF CONTENTS Page 1. Use 3 2. Measuring Principle 3 3. Design 3 4. Technical Data

More information

RESPIRATORY PROTECTION PROGRAM

RESPIRATORY PROTECTION PROGRAM A P P E N D I X 14 RESPIRATORY PROTECTION PROGRAM AP14.1. Purpose. AP14.2. Policy. To establish a written Respiratory Protection Program for CBP. The purpose of an effective respirator program is to provide

More information

QCF Unit and Assessment Specification

QCF Unit and Assessment Specification QCF Unit and Assessment Specification Unit title Ofqual Unit code Working Safely in an Engineering Environment L/600/5781 SQA Unit code FT2R 60 SSC Ref Unit 78 History of changes Publication date: March

More information

SAPAG. Safety valves, type 5700 Storage, Use, Operation and Maintenance Instructions. IMPORTANT NOTICE

SAPAG. Safety valves, type 5700 Storage, Use, Operation and Maintenance Instructions.  IMPORTANT NOTICE SAPAG IMPORTANT NOTICE Contents Important notice 1 0 Valve identification 2 1 Storage 2 2 Installation 2 3 Operation 2 4 Maintenance 3 4.1 Dismantling 3 4.2 Inspection 3 4.3 Repair 3 4.4 Assembly 4 4.5

More information

Hi-Force Limited Prospect Way Daventry Northants NN11 8PL United Kingdom Tel: +44(0) : Fax: +44(0) : Website:

Hi-Force Limited Prospect Way Daventry Northants NN11 8PL United Kingdom Tel: +44(0) : Fax: +44(0) : Website: 1.0 Inspection of the product upon receipt: On receipt of the product, visually inspect the item for any evidence of shipping damage. Please note shipping damage is not covered by warranty. If shipping

More information

COLUMBUS STATE COMMUNITY COLLEGE EMPLOYEE SAFETY MANUAL

COLUMBUS STATE COMMUNITY COLLEGE EMPLOYEE SAFETY MANUAL COLUMBUS STATE COMMUNITY COLLEGE EMPLOYEE SAFETY MANUAL (Hand, Eye, Face, Head, Foot, and Respiratory) Effective October 15, 1995 Page 1 of 15 Updated September 2010 A. At the request of the College Safety

More information

Operation manual Level sensor DC-LS-50 Operation Manual Level Sensor DC-LS-50

Operation manual Level sensor DC-LS-50 Operation Manual Level Sensor DC-LS-50 Operation Manual Level Sensor DC-LS-50 Rev.02 (2014.06) OL Page 1 of 8 Table of contents I. STRUCTURE OF THE MANUAL / CLARIFICATION... 3 II. SAFETY AND HEALTH CONCERNS... 4 1 INTRODUCTION... 5 1.1 Use

More information

West Clinton Fire District

West Clinton Fire District 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 West Clinton Fire District Best Practice Respiratory Protection Program Best

More information

PRESSURISED PAINT CONTAINER

PRESSURISED PAINT CONTAINER PRESSURISED PAINT CONTAINER MODEL NO: CPP2B PART NO: 3082115 OPERATION & MAINTENANCE INSTRUCTIONS GC0913 INTRODUCTION Thank you for purchasing this CLARKE Pressurised Paint Container. Before attempting

More information

ADVANCED AIRLINE FILTER INSTRUCTION MANUAL

ADVANCED AIRLINE FILTER INSTRUCTION MANUAL ADVANCED AIRLINE FILTER INSTRUCTION MANUAL CONTENTS Read all instructions and warnings before using this product. Keep this manual for future reference. INTRODUCTION 3!WARNINGS! 5 BASEMOUNT ASSEMBLY 6

More information

Hampstead Hill School Health and Safety Policy: Confined Spaces

Hampstead Hill School Health and Safety Policy: Confined Spaces All Hampstead Hill School Policies are always to be read and considered in conjunction with Equal Opportunities, Race Equality and Inclusion Policies. This Policy of Hampstead Hill School applies to all

More information

Instruction Manual Dräger MSI P7 and MSI P7 plus

Instruction Manual Dräger MSI P7 and MSI P7 plus Dräger MSI GmbH Rohrstraße 32 58093 Hagen Tel.: +49-2331 / 9584-0 Fax: +49-2331 / 9584-29 e-mail: info@draeger-msi.de D 923; Edition 2011-01-01 Content 1. General Hints Page 4 2. The Instrument 2.1 Front

More information

CONFINED SPACE ENTRY PROGRAM FOR GOUCHER COLLEGE

CONFINED SPACE ENTRY PROGRAM FOR GOUCHER COLLEGE I. INTRODUCTION CONFINED SPACE ENTRY PROGRAM FOR GOUCHER COLLEGE The Goucher College campus has confined spaces that, due to various chemical and physical properties, may cause death or serious injury

More information

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120 MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120 QUICK START GUIDE P/N 6006154, REVISION C MAY 2013 Model 8120 Mask Integrity Tester is patented under U.S. Patent No. 8,312,761. Additional patents are pending.

More information

MODULE 2 BRAZING CYLINDER TRANSPORT, STORAGE AND EQUIPMENT

MODULE 2 BRAZING CYLINDER TRANSPORT, STORAGE AND EQUIPMENT MODULE 2 BRAZING CYLINDER TRANSPORT, STORAGE AND EQUIPMENT ISSUED BY REVISION NUMBER REVISION DATE DOCUMENT NUMBER S. GINBEY 0 7 th Nov 2016 QP74 Module 2 Cylinder Transport, Storage & Equipment ADR Regulations

More information

PRODUCT OPERATING MANUAL

PRODUCT OPERATING MANUAL PRODUCT OPERATING MANUAL PANBLAST TM VISIFLO RESPIRATOR AIRLINE FILTERS Manual Number: ZVP PC 0084 01 Pressure Equipment Directive 97/23/EC SECTION 1. GENERAL INFORMATION 2. INTRODUCTION 3. PREPARATION

More information

TEMPLE UNIVERSITY ENVIRONMENTAL HEALTH AND RADIATION SAFETY

TEMPLE UNIVERSITY ENVIRONMENTAL HEALTH AND RADIATION SAFETY Page 1 of 9 ISSUED: 5/00 REVISED: 08/06 Introduction Purpose: In accordance with applicable regulations and Temple University, this policy was developed to minimize exposure to Ethylene Oxide. Applicability

More information

BCGA GUIDANCE NOTE 17

BCGA GUIDANCE NOTE 17 BCGA GUIDANCE NOTE 17 BCGA POLICY AND GUIDANCE FOR THE SAFE FILLING OF THIRD-PARTY OWNED AND / OR MAINTAINED TANKS Revision 2: 2013 BCGA GUIDANCE NOTE 17 BCGA POLICY AND GUIDANCE FOR THE SAFE FILLING OF

More information

PERSONAL PROTECTIVE EQUIPMENT GUIDANCE

PERSONAL PROTECTIVE EQUIPMENT GUIDANCE Legislation requires a workplace to identify and assess significant hazards (via a risk assessment) associated with workplace activities. The purpose of the assessment is to enable a valid decision about

More information

SEMEM3-65 Servicing anaesthetic and ventilation equipment

SEMEM3-65 Servicing anaesthetic and ventilation equipment Overview This unit identifies the competences you need to carry out servicing activities on anaesthetic and ventilation equipment, in accordance with approved procedures. You will be required to service

More information

SA121, SA122, SA123, SA128 and SA1219 Self-acting Temperature Control Systems (Knob Adjustment)

SA121, SA122, SA123, SA128 and SA1219 Self-acting Temperature Control Systems (Knob Adjustment) 3810050/5 IM-P381-01 CH Issue 5 SA121, SA122, SA123, SA128 and SA1219 Self-acting Temperature Control Systems (Knob Adjustment) Installation and Maintenance Instructions 1. Safety information 2. Use 3.

More information