02 November, 2017 Moderated by Bryan Crist Agilent Technologies

Size: px
Start display at page:

Download "02 November, 2017 Moderated by Bryan Crist Agilent Technologies"

Transcription

1 02 November, 2017 Moderated by Bryan Crist Agilent Technologies 1

2 USP<1058> Analytical Instrument Qualification In this session we will discuss the revised USP chapter <1058> Analytical Instrument Qualification which became effective August 2017 in USP 40 First Supplement. Although the bulk of the AIQ chapter remains unchanged we will review some of the expanded components in terms of processes, phases and risk management. We will also discuss a re-definition of instrument performance qualification relative to its life cycle and its impact on the dissolution apparatus in the GMP laboratory. 2

3 Welcome to the DDG Online Meeting This is our 28 th session Fourth and final meeting in 2017; now in our 7 th year of consecutive meetings Worldwide users group for dissolution chemists Free, on-line, interactive bulletin board Provides annual and regional meetings around the world Place for practical answers to everyday questions on dissolution 3

4 DDG Mission The mission of the Dissolution Discussion Group (DDG) is to provide an independent forum to freely discuss the practical issues which challenge the pharmaceutical industry and affect the day-to-day task of developing, validating, and performing dissolution tests and related chemical analyses. 4

5 5

6 Previous DDG Online Meetings: Aug 10, 2017 Dissolution Testing of Chewable Tablets May 11, 2017 USP Proposed Chapter <1236> Solubility Measurements Feb 09, 2017: Concepts for Qualifying Non-Compendial Dissolution Equipment Nov 10, 2016: USP Performance Testing of Ointments, Creams and Gels Aug 11, 2016: EMEA Draft Reflection Paper on Dissolution Specifications for Generic Oral Immediate Release Products May 12, 2016: USP <1092> and its Impact on Dissolution Automation Feb 25, 2016: Dissolution SOP s; Do What You Say and Say What You Do Nov 12, 2015: FDA Draft Guidance; Dissolution Testing and Specifications for BCS Class 1 and 3 Drugs Aug 13, 2015: Breaking Bad Dissolution Habits May 14, 2015: Crosslinking of Gelatin Capsules and Dissolution Feb 12, 2015: Dissolution Method Development Considerations: Part 6 Intrinsic Dissolution Nov 13, 2014: Dissolution Method Development Considerations: Part 5 Alcohol Induced Dose Dumping Aug 14, 2014: Dissolution Method Development Considerations: Part 4 Time Point Estimation May 15, 2014: Dissolution Method Development Considerations: Part 3 Challenges for Poorly Soluble Compounds Feb 06, 2014: Dissolution Method Development Considerations; Part 2 Media Volume and Sink Conditions for App 1&2 Nov 08, 2013: Dissolution Method Development Considerations; Part 1 Filtration Requirements Aug 15, 2013: Operational Checks; Enhanced Mechanical Qualification Requirement - At Time of Use May 16, 2013: Deaeration of Dissolution Media; What s All The Hot Air About?. Feb 14, 2013: Thinking Small - Challenges with Small Volume Dissolution Nov 15, 2012: Suggestions for the ASTM E Dissolution Apparatus Mechanical Qualification; 5-Year Review Aug 16, 2012: Use of Enzymes for Dissolution Testing of Gelatin Capsules May 24, 2012: The Role of Dissolution in Quality by Design (QbD) Feb 16, 2012: The Dissolution Method Transfer Nov 17, 2011: The Dissolution Laboratory Audit Aug 18, 2011: Do f1 and f2 Tell the Whole Story? May 19, 2011: Best Practices for Success with the USP Dissolution Performance Verification Test (PVT) Mar 15, 2011: Mechanical Calibration: ASTM vs USP; Impact of ICH on Dissolution; QBD and PAT 6

7 Today s Panelists: Bryan Crist Agilent Technologies, Inc. Ken Boda Agilent Technologies, Inc. Greg Martin - Complectors 7

8 11/2/2017 Agilent How USP <1058> Has Evolved 1. AAPS White Paper 2004 [From AAPS March 2003] 4. USP Stimulus Paper Jan Proposed Draft <1058> 6. Draft USP <1058> Aug Bob McDowall & Chris Burgess 2. USP <1058> Aug PF 41(3) June PF 42(3) May AAPS Round Table Nov New USP<1058> [Effective Aug. 2017] AAPS = American Association of Pharmaceutical Scientists 8

9 11/2/2017 Agilent Effective August 2017: New USP <1058> USP 40-NF 35 - First Supplement There are many ways of determining that an instrument is qualified and under control, and these can include, qualification, calibration, validation, and maintenance. In order to ensure fitness for purpose, an integrated approach, based upon a risk assessment, is recommended The USP is the only Pharmacopeia with a general chapter specific to Analytical Instrument Qualification (AIQ) Changes are highlighted 9

10 11/2/2017 Agilent New USP <1058> Summary Explanation Roles & Responsibilities Software Validation Firmware Change Control Documentation AIQ Process Qualification Phases Risk Management by Categorisation: A / B / C Observe Calibrate Components of <1058> Qualify Re-Definition of the Life Cycle Process Components of Data Quality Principles Apply to All Labs. A B C 10

11 New USP <1058> Components Analytical Instrument Qualification User Requirement Specification (URS) The first activity is the generation of a URS, which defines the laboratories specific needs along with technical and operational requirements that are to be met. Key Messages: Define laboratory needs Requirements that must be met 11

12 New USP <1058> Components Analytical Instrument Qualification Design Qualification (DQ) The DQ is the documented collection of activities that define the functional and operational specifications and intended purpose of the instrument. The DQ states what the laboratory wants the instrument to do. Key Messages: Defined functional and operational use. Instrument must do what the lab wants 12

13 New USP <1058> Components Analytical Instrument Qualification Instrument Qualification (IQ) The IQ is the documented collection of activities necessary to establish that an instrument is delivered as designed and specified, is properly installed in the selected environment, and that this environment is suitable for the instrument. Key messages: Delivered as designed and specified Properly installed Environment is suitable 13

14 New USP <1058> Components Analytical Instrument Qualification Operational Qualification (OQ) The OQ is the documented collection of activities necessary to demonstrate that an instrument will function according to its operational specification testing in the selected environment. The OQ demonstrates fitness for selected use, and should reflect the URS*. Key Messages: Instrument Operational Specification performed in lab Demonstrate fitness for purpose Reflects the URS *Note: change to URS from DQ in previous <1058> 14

15 New USP <1058> Components Analytical Instrument Qualification Performance Qualification (PQ) The PQ is the documented collection of activities necessary to demonstrate that an instrument consistently performs according to the specifications defined by the user, and is appropriate for its intended use. After IQ and OQ have been successfully performed, the instruments continued suitability for its intended use is demonstrated through continued performance qualification. Key Messages: Specifications defined by the User Document that instrument is appropriate for intended use Demonstrated through continued and ongoing PQ 15

16 PQ Requirements PQ tests: are based on typical on-site applications are based on good science reflect the intended use of the instrument may be modular or holistic (dissolution emq or PVT) may consist of analyzing known components or stds. 16

17 PQ Frequency The testing frequency: depends on the ruggedness of the instrument depends on the criticality of the analytical method may be scheduled at regular intervals, or unscheduled with repeated test to compile performance history 17

18 Preventive Maintenance Preventive maintenance: is required at scheduled intervals may include calibration (as found, as left and after PM) plans and procedures must be documented when PM fails; the cause must be investigated and documented the instrument must be maintained or repaired the relevant OQ and PQ tests should be repeated 18

19 Practices for PQ, change control and review Document each PQ, maintenance and calibration activity. Change control procedures required with changes in: Instrument configuration Firmware and software Critical instruments should have periodic review to ensure that the system is still under control. Instrument owner and management is responsible for PQ; although portions can be carried out by internal or external suppliers or service providers. 19

20 Questions or comments about <1058>? NEXT: DISSOLUTION APPARATUS PERFORMANCE QUALIFICATION 20

21 The Dissolution Environment In vitro dissolution data will be of great importance when assessing changes in production site, manufacturing process, or formulation, and assist in decisions concerning the need for bioavailability studies. The proper qualification of the apparatus is critical to accurate and precise evaluation of dosage form performance

22 An Overview of the Dissolution Performance Test The dissolution apparatus allows the testing of six dosage forms. Each position has: An inert hemispheric vessel A dissolution solvent (medium) A rotating spindle which provides the hydrodynamic flow of the solvent across the surface of the dosage form The dissolution apparatus must maintain these three components in terms of alignment, stability and isolation from the environment

23 Background of Performance Qualification USP Performance Verification Test (PVT) In existence since 1978 As of March 1 st, 2010, a new set of acceptance criteria based on both geometric mean and %CV was adopted. The current lot is now lot R In addition to the changes in acceptance criteria, some minor changes to the dissolution test recommendations have occurred

24 Background of Performance Qualification Note, this is not the current USP PVT lot.

25 Background of Performance Qualification FDA Concerns with USP PVT Prednisone Tablets: Perceived variability with PVT Prednisone tablets not suitable for evaluating variability in the apparatus. 6-month qualification interval too long. High risk for failure which leads to Out-of-Specification (OOS) investigations. Not keeping the apparatus in tune may lead to a false sense of security regarding calibration. Difficulty meeting %CV Acceptance Criteria and values occasionally drift over time. Stability issues?

26 USP Performance Verification Test (PVT) Other concerns: Expense of Tablets and RS. Down time. When it fails, what was the cause? Poor vessel quality and condition Poor paddle and basket quality and condition Vessel retaining systems failing Unvalidated deaeration techniques Unvalidated automated methods Unvalidated filters None of the above? 26

27 Analytical Instrument Qualification USP <1058> Analytical Instrument Qualification Definition of Performance Qualification (PQ): Usually based on the instrument s application and may consist of analyzing known components or standards PQ tests may be modular or holistic PQ tests should be based on good science and reflect the general intended use of the instrument When an instrument fails to meet PQ test specifications, it requires maintenance or repair

28 Dissolution Mechanical Qualification Guidance The Use of Mechanical Calibration of Dissolution Apparatus 1 and 2 Current Good Manufacturing Practice (CGMP) FDA Guidance for Industry January anceregulatoryinformation/guidances/ucm pdf CGMP requires laboratory apparatus be calibrated at suitable intervals in accordance with established written specifications (21 CFR (b)(4))

29 Dissolution Mechanical Qualification Guidance The Use of Mechanical Calibration of Dissolution Apparatus 1 and 2 Current Good Manufacturing Practice (CGMP) - Claims: Enhanced Mechanical Calibration (MC or MQ) of dissolution apparatus may be used as an alternative to the current Apparatus Suitability procedure for Dissolution Apparatus 1 and 2 described in USP <711> The MC procedure should specify the frequency at which each calibration step is performed Either the USP procedure or an appropriate MC method executed according to a written procedure will satisfy the CGMP requirement for calibration of the laboratory apparatus Regardless of which procedure is used, appropriate measures must be taken to control the following sources of significant variability in dissolution testing: dissolved gases, vibration, and vessel dimensions.

30 Dissolution Enhanced Mechanical Qualification Standards Mechanical Qualification of Dissolution Apparatus 1 and 2 US Food and Drug Administration (FDA), DPA-LOP.002, Version 2.0, Effective Date 02 June, Standard Practice for Qualification of Basket and Paddle Dissolution Apparatus American Society for Testing and Materials, International (ASTM), Designation E , Effective

31 The 5 Steps for Implementation of Enhanced Mechanical Qualification of the Dissolution Apparatus 1. Check vessel, basket and paddle dimensions on receipt or obtain COC 2. Perform maintenance procedures recommended by manufacturer 3. Perform mechanical qualification: At documented intervals (6-month, 3-month, 1-month, weekly) based on risk After apparatus is moved After apparatus is repaired 4. Perform Operational Checks at each time of use 5. FDA requires implementation of laboratory controls for significant sources of vibration affecting dissolution testing Vessel quality Deaeration Vibration

32 Dissolution MQ Standard Requirements #1 Check Dimensions Upon Receipt: Paddle, Basket/Shaft, and Vessel apparatus components must be verified that they conform with the harmonized standard USP <711> Dissolution (USP, JP, EP) Individual measurements for each dimension of each component must be documented Certificates of Conformance (COC) may be obtained from Agilent to document conformance. Otherwise, measurements must be documented by the end user.

33 Parameter ICH Harmonized (USP, JP, EP) FDA DPA-LOP.002 ASTM E #3 Mechanical 25 ± 2 mmqualification 25 ± 2 Specifications 25 ± 2 mm & (or <8%) Tolerances Basket and Paddle Depth Rotational Speed Shaft Wobble ± 4% of specified rate No significant wobble ± 2 rpm of target ± 2 rpm or within 2% of target (larger) 1.0 mm total runout Shaft Verticality Not measured 0.5 from vertical Basket Wobble ± 1 mm 1.0 mm total runout Vessel/Shaft Centering Vessel Verticality Vessel Plate Level NMT 2 mm from center axis Not Measured 1.0 mm from center line,2 pts upper and lower 1.0 from vertical (2-90 positions) 1.0 mm total runout Within Bubble 1.0 mm total runout 1.0 mm from center line,2 pts upper and lower 1.0 from vertical (2-90 positions) USP Toolkit Ver mm ± 1 rpm of target 1.0 mm total wobble Not measured 1.0 mm total wobble NMT 2.0 mm difference (4-90 positions) NMT 0.5 from vertical Not Measured Not Measured Not Measured NMT 0.5 from horizontal

34 Dissolution Mechanical Qualification Standard Requirements #4 Operational Checks (document each time of use): Basket/shaft examination Paddle examination Vessel examination Vessel Temperature Vibration

35 Dissolution Mechanical Qualification Standard Requirements #5 FDA Control of significant sources of variability: Dissolved gases - to be removed through a validated deaeration procedure Vibration - sources to be eliminated: Shakers, mixers, fume hoods, pumps, centrifuge, radio, etc Malfunctioning internal components Heater/circulator problems Vessel dimensions - conform with cylindrical, circular sides and hemispherical bottom

36 The Advantage for MQ Time Savings (35-min vs 2-days with PVT) Ease of use allows shorter qualification intervals Data that helps you pinpoint possible problems Now have the ability to trend parameter variation over time No guesswork associated with manual and analog gauges No standards or tablets to purchase Provides instant failure investigation information Reports exact information required by Enhanced MQ Standards

37 General Discussion

38 November Confidentiality 38 2, 2017 Label References United States Pharmacopoeial Convention, Inc. United States Pharmacopeia (USP 40-NF 35). Rockville(MD); 2017: <1058> Analytical Instrument Qualification <711> Dissolution FDA Guidance for Industry; The Use of Mechanical Calibration of Dissolution Apparatus 1 and 2 Current Good Manufacturing Practice, January 2010 FDA Division of Pharmaceutical Analysis, DPA-LOP.002 Mechanical Qualification of Dissolution Apparatus 1 and 2, Ver. 2.0, June 2, 2006 ASTM E Standard Practice for Qualification of Basket and Paddle Dissolution Apparatus.

39 Mark your calendars for the next Online DDG meeting: Thursday, 08 February, Topic to be announced. We look for your suggestions for current hot topics in the field of dissolution such as drug release from novel dosage forms, FDA Guidance, Apparatus Qualification, etc. We do want to hear from you so please continue to suggest topics at the DDG site Meetings are held quarterly on Tuesdays from 10:30-11:30 Eastern US Time. Please consult the DDG site for recorded sessions, future topics and updated information. 39

40 40

The Transition to Enhanced Mechanical Qualification of the Dissolution Apparatus

The Transition to Enhanced Mechanical Qualification of the Dissolution Apparatus The Transition to Enhanced Mechanical Qualification of the Dissolution Apparatus Bryan Crist Scientific Affairs Manager Dissolution Systems Agilent Technologies, Inc. Disso Europe 2016 Bucharest, Romania

More information

Mechanical Qualification (MQ) for Apparatus 1 and 2. Oct 10th, 2008

Mechanical Qualification (MQ) for Apparatus 1 and 2. Oct 10th, 2008 Mechanical Qualification (MQ) for Apparatus 1 and 2 Oct 10th, 2008 Overview of Present Qualifications Installation Qualification (IQ) Operational Qualification (OQ) Performance Qualification (PQ) 2 IQ

More information

Frequently Asked Questions about Mechanical Qualification. Ken Boda Applications Engineer Agilent Technologies

Frequently Asked Questions about Mechanical Qualification. Ken Boda Applications Engineer Agilent Technologies Frequently Asked Questions about Mechanical Qualification Ken Boda Applications Engineer Agilent Technologies Is MQ Acceptable to the FDA? Yes Either FDA or ASTM approach is acceptable Many people (including

More information

11 May, 2017 Moderated by Bryan Crist Agilent Technologies

11 May, 2017 Moderated by Bryan Crist Agilent Technologies 11 May, 2017 Moderated by Bryan Crist Agilent Technologies 1 Welcome to the DDG Online Meeting This is our 26 th session Second meeting in 2017; now in our 7 th year of consecutive meetings Worldwide users

More information

Dissolution Failure Investigation. Ken Boda Dissolution Applications Engineer

Dissolution Failure Investigation. Ken Boda Dissolution Applications Engineer Dissolution Failure Investigation Ken Boda Dissolution Applications Engineer Dissolution Failure Investigation Whenever a dissolution test fails to meet specifications, an investigation should be performed

More information

Food and Drug Administration Division of Pharmaceutical Analysis. Mechanical Qualification of Dissolution Apparatus 1 and 2

Food and Drug Administration Division of Pharmaceutical Analysis. Mechanical Qualification of Dissolution Apparatus 1 and 2 Page 1 of 10 Sections included in this document Table of Contents Document history 1. Purpose 2. Scope/Policy 3. Responsibilities 4. Background 5. References 6. Procedure 7. Records 8. Glossary - Not applicable

More information

Qualification Test Specifications for Dissolution Test Systems

Qualification Test Specifications for Dissolution Test Systems 35 for Vantage Dissolution Point Drive, Rochester NY Test 14624 Systems Phone: 800-828- 1470 Fax: 800-395- 0543 Transcat.com Test Set Points/Range Acceptance Criteria Rotational Speed Runout (Wobble) Shaft

More information

711 DISSOLUTION. Portions of the present general chapter text that are national USP text, and therefore not part of the

711 DISSOLUTION. Portions of the present general chapter text that are national USP text, and therefore not part of the Change to read: 711 DISSOLUTION This general chapter is harmonized with the corresponding texts of the European Pharmacopoeia and/or the Japanese Pharmacopoeia. The texts of these pharmacopeias are therefore

More information

BRIEFING 711 DISSOLUTION

BRIEFING 711 DISSOLUTION BRIEFING 711 Dissolution, USP 28 page 2412 and page 234 of PF 30(1) [Jan. Feb. 2004]. The United States Pharmacopeia is the coordinating pharmacopeia in the efforts toward interntional harmonization of

More information

DISSOLUTION TEST FOR SOLID DOSAGE FORMS

DISSOLUTION TEST FOR SOLID DOSAGE FORMS Seite 1 von 15 01/2010:20903 corrected 6.8 2.9.3. DISSOLUTION TEST FOR SOLID DOSAGE FORMS This test is provided to determine compliance with the dissolution requirements for solid dosage forms administered

More information

Dissolution Toolkit Procedures for Mechanical Calibration and Performance Verification Test Apparatus 1 and Apparatus 2 Version 2.

Dissolution Toolkit Procedures for Mechanical Calibration and Performance Verification Test Apparatus 1 and Apparatus 2 Version 2. Dissolution Toolkit Procedures for Mechanical Calibration and Performance Verification Test Apparatus 1 and Apparatus 2 Version 2.0 March 22, 2010 Scope: The dissolution toolkit provides a description

More information

Title: INSTRUMENT AND EQUIPMENT DOCUMENTATION AND RECORDS

Title: INSTRUMENT AND EQUIPMENT DOCUMENTATION AND RECORDS Page 1 of 7 Sections Included in this Document and Document History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References/(Deleted date) 6. Procedure 7. Definitions 8. Records 9. Supporting

More information

Comparison between Compendial and Enhanced Mechanical Calibration Specifications

Comparison between Compendial and Enhanced Mechanical Calibration Specifications Comparison between Compendial and Enhanced Mechanical Calibration Specifications Calibration Parameter Current Pharmacopoeia Enhanced USP Specification Enhanced FDA Specification Bench Horizontality 1

More information

Quantos Automated Dosing Solution Preparation Precise concentrations Process compliance Minimize out-of-specs

Quantos Automated Dosing Solution Preparation Precise concentrations Process compliance Minimize out-of-specs Quantos Automated Dosing Solution Preparation Precise concentrations Process compliance Minimize out-of-specs Gravimetric Sample Preparation The Alternative to Volumetric Flasks From Volumetric to Gravimetric

More information

Dissolution and Drug Release Testing. DISSOLUTION SYSTEMS SOURCE BOOK Edition

Dissolution and Drug Release Testing. DISSOLUTION SYSTEMS SOURCE BOOK Edition Dissolution and Drug Release Testing DISSOLUTION SYSTEMS SOURCE BOOK 2011-2012 Edition Welcome Agilent Technologies, Inc. is well known within the pharmaceutical community for its broad portfolio of instruments

More information

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries. Informal Commentary Limit of Diethylene Glycol (DEG) and Ethylene Glycol (EG) in Sorbitol Solution, Sorbitol Sorbitan Solution and Noncrystallizing Sorbitol Solution December 2009 Monograph/Section(s):

More information

Effect of the Irregular Inner Shape of a Glass Vessel on Prednisone Dissolution Results

Effect of the Irregular Inner Shape of a Glass Vessel on Prednisone Dissolution Results dx.doi.org/10.14227/dt120405p15 Effect of the Irregular Inner Shape of a Glass Vessel on Prednisone Dissolution Results Makoto Tanaka 1, 2, Hiroyasu Fujiwara 1, and Morimasa Fujiwara 1 e-mail: takao-mt@mbj.nifty.com

More information

COPLEY TABLET DISSOLUTION TESTERS

COPLEY TABLET DISSOLUTION TESTERS Apparatus 1 (Basket) Apparatus 2 (Paddle) COPLEY TABLET DISSOLUTION TESTERS In the majority of cases, the effectiveness of tablets or capsules administered orally relies on the drug dissolving in the fluids

More information

Improve Process Reliability

Improve Process Reliability Convenient Conductivity Calibration to Meet Pharmacopeia Regulations Pharmacopeias require that conductivity measurement systems be periodically calibrated to ensure their precision. The range of accessories

More information

Title: Standard Operating Procedure for Elemental and Organic Carbon (EC and OC) using Non-Dispersive Infrared Detection (NDIR)

Title: Standard Operating Procedure for Elemental and Organic Carbon (EC and OC) using Non-Dispersive Infrared Detection (NDIR) Procedure No: SOP-025 Revision No: 1.0 (January 21, 2011) Page No.: 1 of 8 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting and to

More information

Effects of Dissolved Gases in Surfactant Dissolution Media

Effects of Dissolved Gases in Surfactant Dissolution Media Effects of Dissolved Gases in Surfactant Dissolution Media Kyle A. Fliszar*, Richard J. Forsyth, Zhong Li, Gregory P. Martin Pharmaceutical Analysis & Control, PR&D, Merck & Co. West Point, PA, USA. email:

More information

Int r oduct ion t o Int egr it y Tes t ing Theor y

Int r oduct ion t o Int egr it y Tes t ing Theor y Int r oduct ion t o Int egr it y Tes t ing Theor y G. Somasundaram (Som) Associate Director Technology Management South East Asia & Oceania Singapore Somasundaram.g@merckgroup.com 16 th May 2017 Merck

More information

INSPECTION OF MEDICINAL GASES

INSPECTION OF MEDICINAL GASES PHARMACEUTICAL INSPECTION CONVENTION PHARMACEUTICAL INSPECTION CO-OPERATION SCHEME PI 025-2 25 September 2007 AIDE-MEMOIRE INSPECTION OF MEDICINAL GASES PIC/S September 2007 Reproduction prohibited for

More information

Characterization and formulation of cocrystals

Characterization and formulation of cocrystals Characterization and formulation of cocrystals Cocrystals can deliver unique physical properties, but taking full advantage of these improved properties requires a new approach to cocrystal formulation

More information

ROTARY VISCOMETER. Rotary Viscometers MYR V0/V1/V2 acc. to ISO 2555/ASTM (Brookfield method) SERIE VR 3000

ROTARY VISCOMETER. Rotary Viscometers MYR V0/V1/V2 acc. to ISO 2555/ASTM (Brookfield method) SERIE VR 3000 ROTARY VISCOMETER Rotary Viscometers MYR V0/V1/V2 acc. to ISO 2555/ASTM (Brookfield method) SERIE VR 3000 VISCOMETERS VR 3000 The MYR viscometers, models V0, V1 and V2, are rotational viscometers for the

More information

ANNEX 4 ALTERNATIVE TEXT FOR OPERATIONAL GUIDELINES FOR VERIFICATION OF DAMAGE STABILITY REQUIREMENTS FOR TANKERS

ANNEX 4 ALTERNATIVE TEXT FOR OPERATIONAL GUIDELINES FOR VERIFICATION OF DAMAGE STABILITY REQUIREMENTS FOR TANKERS Annex 4, page 1 ANNEX 4 ALTERNATIVE TEXT FOR OPERATIONAL GUIDELINES FOR VERIFICATION OF DAMAGE STABILITY REQUIREMENTS FOR TANKERS GUIDELINES FOR VERIFICATION OF DAMAGE STABILITY FOR TANKERS PART 2 OPERATIONAL

More information

Significant Change to Dairy Heat Treatment Equipment and Systems

Significant Change to Dairy Heat Treatment Equipment and Systems Significant to Dairy Heat Treatment September 2008 Page 1 Significant to Dairy Heat Treatment Equipment and Systems September 2008 1 Background Requirements for the assessment of dairy heat treatment equipment

More information

USP Dissolution Instrument Calibration or PQ Standard Operating Procedure or SOP

USP Dissolution Instrument Calibration or PQ Standard Operating Procedure or SOP USP Dissolution Instrument Calibration or PQ Standard Operating Procedure or SOP Part 2: Medium Preparation and Calibrator Tablet Tests Methods, Calculations and Typical Results PHARMA TEST APPARATEBAU

More information

STD-3-V1M4_1.7.1_AND_ /15/2015 page 1 of 6. TNI Standard. EL-V1M4 Sections and September 2015

STD-3-V1M4_1.7.1_AND_ /15/2015 page 1 of 6. TNI Standard. EL-V1M4 Sections and September 2015 page 1 of 6 TNI Standard EL-V1M4 Sections 1.7.1 and 1.7.2 September 2015 Description This TNI Standard has been taken through all of the voting stages and has received consensus approval by the TNI membership.

More information

Calibration Requirements for Direct Reading Confined Space Gas Detectors

Calibration Requirements for Direct Reading Confined Space Gas Detectors : Calibration Requirements for Direct Reading Confined Space Gas Detectors However, the definition of bump test has always been a little slippery. Some manufacturers differentiate between a bump test that

More information

Beamex. Calibration White Paper. Weighing scale calibration - How to calibrate weighing instruments

Beamex. Calibration White Paper.  Weighing scale calibration - How to calibrate weighing instruments Beamex Calibration White Paper info@beamex.com Weighing scale calibration - How to calibrate weighing instruments Weighing scale calibration - How to calibrate weighing instruments Weighing scales, weighing

More information

AGILENT TURBOMOLECULAR PUMPS MEASUREMENT METHODOLOGY

AGILENT TURBOMOLECULAR PUMPS MEASUREMENT METHODOLOGY AGILENT TURBOMOLECULAR PUMPS MEASUREMENT METHODOLOGY Agilent Vacuum Products: Our Committment to Quality Agilent s commitment to quality and Measurement s philosophy Reference Standards (PNEUROP, ISO)

More information

Engineering: Measurement Technology Pressure/Level (SCQF level 6)

Engineering: Measurement Technology Pressure/Level (SCQF level 6) National Unit Specification: general information CODE F5KT 12 SUMMARY This Unit can be delivered as part of a National Qualification Group Award but can also be taken as a free-standing Unit. This Unit

More information

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator Procedure No: SOP-029 Revision No: 1.1 (December 29, 2010) Page No.: 1 of 7 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting, air quality

More information

CTG Certification Protocol -- Version 3.0 SUMMARY

CTG Certification Protocol -- Version 3.0 SUMMARY Date of Test: Test Report Number: SUMMARY Date of Prior Certification: Certified by: Re-certification Due Date: Unit Description: Make : Model: Serial Number: Location: CERTIFIER INFORMATION Name: Client

More information

OPERATING PROCEDURE. Number: V2. Authors. Approvals

OPERATING PROCEDURE. Number: V2. Authors. Approvals OPERATING PROCEDURE Title: Calibration of the Denver XL5K Top-Loading Balance (Note: This SOP is for demonstration purposes, and has numerous problems) Effective Date: April 30, 2003 Author Date: October

More information

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION Application Note Doc. G8108-001 Rev. A - 23-Jul-02 ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION For more information regarding Digiquartz products contact: Paroscientific,

More information

JetVent SAFE, AUTOMATIC PRESSURIZED CONTAINMENT.

JetVent SAFE, AUTOMATIC PRESSURIZED CONTAINMENT. JetVent SAFE, AUTOMATIC PRESSURIZED CONTAINMENT. PROTECTING PEOPLE AND THEIR ENVIRONMENTS FROM EXPOSURE. Handling today s higher-potency active pharmaceutical ingredients (HPAPIs) presents ever-increasing

More information

3M Steri-Vac Sterilizer/Aerator GSX Series. For Life Science Applications. Product. Profile

3M Steri-Vac Sterilizer/Aerator GSX Series. For Life Science Applications. Product. Profile M Steri-Vac Sterilizer/Aerator GSX Series Product Profile Product Profile M Steri-Vac Sterilizer/Aerator GSX Series ethylene oxide (EO) sterilizer/aerators are designed to be used with M Steri-Gas EO Gas

More information

Safety Services Guidance. Guidance on: The Provision and Use of Work Equipment Regulations 1998 (PUWER)

Safety Services Guidance. Guidance on: The Provision and Use of Work Equipment Regulations 1998 (PUWER) Guidance on: The Provision and Use of Work Equipment Regulations 1998 (PUWER) Key word(s) : Work equipment. Using, maintaining and repairing. Information, instruction and supervision. Risk Assessment.

More information

SKSTEX36(SQA Unit Code-H7C9 04) Calibrate textile production equipment

SKSTEX36(SQA Unit Code-H7C9 04) Calibrate textile production equipment Overview This standard is for those who carry out routine calibrations at prescribed intervals in order to keep equipment in calibration. The job role will involve: 1. Performing calibration 2. Identifying

More information

IMCA DP Station Keeping Bulletin 04/18 November 2018

IMCA DP Station Keeping Bulletin 04/18 November 2018 DP STATION KEEPING BULLETIN IMCA DP Station Keeping Bulletin 04/18 November 2018 The following event trees have been compiled from recent reports received by IMCA. The originators granted IMCA permission

More information

Design Qualification. DQ Regulatory Requirements EU-GMP, Annex 15

Design Qualification. DQ Regulatory Requirements EU-GMP, Annex 15 Design Qualification by Dr. Ingrid Walther May 2015, page 1 DQ Regulatory Requirements EU-, Annex 15 Design qualification (DQ) 3.3. The next element* in the qualification of equipment, facilities, utilities,

More information

Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience -

Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience - Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience - Dr. Wolfgang Jockel, Martin Schneider, TÜV Rheinland Group, D-51105 Cologne / Germany 1. Introduction A new basis for the certification

More information

Proposed Uniform Standards for Equipment Leaks

Proposed Uniform Standards for Equipment Leaks 12 th LDAR Symposium May 15-16, 2012 New Orleans, LA Proposed Uniform Standards for Equipment Leaks Standards Certification Education & Training Publishing Conferences & Exhibits 1 Proposed LDAR Uniform

More information

EuroBioConcept s.a.s Hospital Pharmacy Isolator Systems

EuroBioConcept s.a.s Hospital Pharmacy Isolator Systems EuroBioConcept s.a.s Hospital Pharmacy Isolator Systems Presented by: Joe Sacca Dynamic Design Pharma, Inc. Mission Viejo, California USA June 2014 INTRODUCTION EBC PRODUCTS TYPICAL INSTALLATIONS ISOLATION

More information

Requirements for the certification of movements and mechanical watches resistant to magnetic fields of 1.5 T ( G)

Requirements for the certification of movements and mechanical watches resistant to magnetic fields of 1.5 T ( G) Institut fédéral de métrologie METAS METAS-N001/EN Requirements for the certification of movements and mechanical watches resistant to magnetic fields of 1.5 T (15 000 G) This document is a translation

More information

SILICA EXPOSURE MEASUREMENT IN REAL-TIME FOR CONSTRUCTION

SILICA EXPOSURE MEASUREMENT IN REAL-TIME FOR CONSTRUCTION SILICA EXPOSURE MEASUREMENT IN REAL-TIME FOR CONSTRUCTION APPLICATION NOTE EXPMN-023(A4) Frequently Asked Questions What is crystalline silica and why is it important to measure exposure? Silica is found

More information

ANSI / ASSE Z88.2 American National Standard Practices for Respiratory Protection (1992 rev. versus 2015 rev.)

ANSI / ASSE Z88.2 American National Standard Practices for Respiratory Protection (1992 rev. versus 2015 rev.) ANSI / ASSE Z88.2 BACKGROUND: A substantial body of research has been published since the 1992 revision of Z88.2. This information related to the proper use and performance of respiratory protection in

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs. Laboratory Hardware Custom Gas Chromatography Solutions Custom solutions for your analytical needs. Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions for advanced chromatography

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results. Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions

More information

Installation Procedure Esco Miri Multi-Room Incubator

Installation Procedure Esco Miri Multi-Room Incubator Esco Miri Multi-Room Incubator Version 1.1 ESCO MEDICAL Esco Micro Pte Ltd 21 Changi South Street 1 Singapore 486777 Tel. : +65 6542 0833 Fax : +65 6542 5732 / 6546 2913 Mail : csis-medical@escoglobal.com

More information

Chemical Analysis for Methyltestosterone

Chemical Analysis for Methyltestosterone APPLIED CONSUMER SERVICES, INC. Chemical Analysis for Methyltestosterone Report #: 27584 Report to: Mr. Mitch Ross 24 Dockside Lane # 131 Key Largo, Florida 33017 2016 1 1 8 9 0 N W 8 7 t h C o u r t,

More information

HumiSys HF High Flow RH Generator

HumiSys HF High Flow RH Generator HumiSys HF High Flow RH Generator Designed, built, and supported by InstruQuest Inc. Versatile Relative Humidity Generation and Multi-Sensor System The HumiSys HF is a high flow version of the previously

More information

1. Scope: This document describes the procedure for verifying the performance of vaccine vial monitors.

1. Scope: This document describes the procedure for verifying the performance of vaccine vial monitors. PQS Independent type-testing protocol WHO/PQS/E06/IN05.VP.2 Original: English Distribution: General TITLE: Vaccine Vial Monitor Product verification protocol: E06/IN05.VP.2 Applies to specification ref(s):

More information

VS Series Leak Detectors for Vacuum Furnaces LEAK DETECTION FOR VACUUM FURNACES

VS Series Leak Detectors for Vacuum Furnaces LEAK DETECTION FOR VACUUM FURNACES VS Series Leak Detectors for Vacuum Furnaces LEAK DETECTION FOR VACUUM FURNACES VS Series Leak Detectors for Vacuum Furnaces indicate the presence of a leak. If the vacuum level is slow to reach the original

More information

Finding Leak Testing Success Through Calibration and Validation. Paul Chamberlain President, CEO

Finding Leak Testing Success Through Calibration and Validation. Paul Chamberlain President, CEO Finding Leak Testing Success Through Calibration and Validation Paul Chamberlain President, CEO 2 Overview Confidence in Your Leak Test Results Sources of Uncertainty Pressure decay leak testing Helium

More information

Top Tips to Mitigate. Food Fraud COPYRIGHT ALL RIGHTS RESERVED.

Top Tips to Mitigate. Food Fraud COPYRIGHT ALL RIGHTS RESERVED. Top Tips to Mitigate Food Fraud COPYRIGHT 2017. ALL RIGHTS RESERVED. Today s Speakers Peter Claise Marketing Director Foods Programs Jeff Chilton Vice President of Professional Services Today s Discussion

More information

Distributed Control Systems

Distributed Control Systems Unit 41: Unit code Distributed Control Systems M/615/1509 Unit level 5 Credit value 15 Introduction With increased complexity and greater emphasis on cost control and environmental issues, the efficient

More information

Index. Batch processing 56 Branching xiii 52 62

Index. Batch processing 56 Branching xiii 52 62 Index A Accessibility electronic 40 105 of emergency operating procedures 85 of procedures 15 32 Accidental chemical release xiii Administrative procedures xiii 27 Alignment 75 American Petroleum Institute

More information

ATTACHMENT O WIPP MINE VENTILATION RATE MONITORING PLAN

ATTACHMENT O WIPP MINE VENTILATION RATE MONITORING PLAN ATTACHMENT O WIPP MINE VENTILATION RATE MONITORING PLAN (This page intentionally blank) ATTACHMENT O WIPP MINE VENTILATION RATE MONITORING PLAN TABLE OF CONTENTS O- Definitions... O- Objective... O- Design

More information

WIKA INSTRUMENT CORPORATION

WIKA INSTRUMENT CORPORATION WIKA INSTRUMENT CORPORATION Instruction Manual DIFFERENTIAL PRESSURE GAUGE Series 1500 Series 1000 Series 300 WIKA Instrument Corporation 1000 Wiegand Boulevard Lawrenceville, GA 30043 1-888-945-2872 http://www.wika.com

More information

TG GUIDELINES CONCERNING CALIBRATION INTERVALS AND RECALIBRATION

TG GUIDELINES CONCERNING CALIBRATION INTERVALS AND RECALIBRATION GUIDELINES CONCERNING CALIBRATION INTERVALS AND RECALIBRATION Approved By: Senior Manager: Mpho Phaloane Revised By: Field Manager: Neville Tayler Date of Approval: 2015-08-26 Date of Implementation: 2015-08-26

More information

Ultima. X Series Gas Monitor

Ultima. X Series Gas Monitor Ultima X Series Gas Monitor Safety Manual SIL 2 Certified " The Ultima X Series Gas Monitor is qualified as an SIL 2 device under IEC 61508 and must be installed, used, and maintained in accordance with

More information

SIL Safety Manual. ULTRAMAT 6 Gas Analyzer for the Determination of IR-Absorbing Gases. Supplement to instruction manual ULTRAMAT 6 and OXYMAT 6

SIL Safety Manual. ULTRAMAT 6 Gas Analyzer for the Determination of IR-Absorbing Gases. Supplement to instruction manual ULTRAMAT 6 and OXYMAT 6 ULTRAMAT 6 Gas Analyzer for the Determination of IR-Absorbing Gases SIL Safety Manual Supplement to instruction manual ULTRAMAT 6 and OXYMAT 6 ULTRAMAT 6F 7MB2111, 7MB2117, 7MB2112, 7MB2118 ULTRAMAT 6E

More information

7 th International Conference on Wind Turbine Noise Rotterdam 2 nd to 5 th May 2017

7 th International Conference on Wind Turbine Noise Rotterdam 2 nd to 5 th May 2017 7 th International Conference on Wind Turbine Noise Rotterdam 2 nd to 5 th May 2017 Sound power level measurements 3.0 ir. L.M. Eilders, Peutz bv: l.eilders@peutz.nl ing. E.H.A. de Beer, Peutz bv: e.debeer@peutz.nl

More information

Understanding safety life cycles

Understanding safety life cycles Understanding safety life cycles IEC/EN 61508 is the basis for the specification, design, and operation of safety instrumented systems (SIS) Fast Forward: IEC/EN 61508 standards need to be implemented

More information

Dissolution test results showing biorelevant media made from SIF Powder Original is the same as media prepared using methylene chloride

Dissolution test results showing biorelevant media made from SIF Powder Original is the same as media prepared using methylene chloride Dissolution test results showing biorelevant media made from SIF Powder Original is the same as media prepared using methylene chloride Michael Schaich, Bastian Kloefer, Mathew Leigh : info@biorelevant.com

More information

Air Handling Unit (AHU) Air handling unit which serves to condition the air and provide the required air movement within a facility.

Air Handling Unit (AHU) Air handling unit which serves to condition the air and provide the required air movement within a facility. IOSR Journal of Mechanical and Civil Engineering (IOSR-JMCE) e-issn: 2278-1684,p-ISSN: 2320-334X, Volume 13, Issue 5 Ver. I (Sep. - Oct. 2016), PP 28-35 www.iosrjournals.org. Heating, Ventilation and Air

More information

Solenoid Valves used in Safety Instrumented Systems

Solenoid Valves used in Safety Instrumented Systems I&M V9629R1 Solenoid Valves used in Safety Instrumented Systems Operating Manual in accordance with IEC 61508 ASCO Valves Page 1 of 7 Table of Contents 1 Introduction...3 1.1 Terms and Abbreviations...3

More information

Type Testing Procedure for Crankcase Explosion Relief Valves

Type Testing Procedure for Crankcase Explosion Relief Valves (Jan 2005) (Corr.1 Nov 2005) (Rev.1 Oct 2006) (Corr.1 Mar 2007) (Rev.2 Sept 2007) (Corr.1 Oct 2007) (Rev.3 Jan 2008) Type Testing Procedure for Crankcase Explosion Relief Valves 1. Scope 1.1 To specify

More information

Extended Application Note

Extended Application Note Extended Application Note Harmful Substances in Dietary Supplements HPLC Columns with APP A-337 www.mtc-usa.com 1-732-578-1777 INTRODUCTION A dietary supplement is a substance to be consumed for the purpose

More information

This test shall be carried out on all vehicles equipped with open type traction batteries.

This test shall be carried out on all vehicles equipped with open type traction batteries. 5.4. Determination of hydrogen emissions page 1 RESS-6-15 5.4.1. This test shall be carried out on all vehicles equipped with open type traction batteries. 5.4.2. The test shall be conducted following

More information

2600T Series Pressure Transmitters Plugged Impulse Line Detection Diagnostic. Pressure Measurement Engineered solutions for all applications

2600T Series Pressure Transmitters Plugged Impulse Line Detection Diagnostic. Pressure Measurement Engineered solutions for all applications Application Description AG/266PILD-EN Rev. C 2600T Series Pressure Transmitters Plugged Impulse Line Detection Diagnostic Pressure Measurement Engineered solutions for all applications Increase plant productivity

More information

Children s Hospital of Philadelphia Committees for the Protection of Human Subjects Policies and Procedures. IRB Review Processes

Children s Hospital of Philadelphia Committees for the Protection of Human Subjects Policies and Procedures. IRB Review Processes Page: 1 of 8 I. PURPOSE II. III. IV. The purpose of this standard operating procedure is to describe IRB review processes, including pre-review, review, and post-review procedures. POLICY STATEMENT The

More information

Quality Assurance Charting for QC Data

Quality Assurance Charting for QC Data Quality Assurance Charting for QC Data September 2018 Iowa s Environmental & Public Health Laboratory Copyright the State Hygienic Laboratory at the University of Iowa 2017. All rights reserved. Images

More information

ANALYTICAL.

ANALYTICAL. ANALYTICAL P R O D U C T P R O G R A M 2 0 0 8 / 0 9 R PHARMA TEST PAGE 3 > EDITORIAL PAGE 4 > DISSOLUTION TESTER - PT-DT70 PAGE 5 > DISSOLUTION TESTER - PT-DT8S PAGE 6 > DISSLOUTION TESTER - PTWS 310

More information

ESTATES OFFICE OPERATIONS WORKS INSTRUCTIONS WI 6. Fume Cupboards, Ventilated Cabinets, Local Exhaust Ventilation

ESTATES OFFICE OPERATIONS WORKS INSTRUCTIONS WI 6. Fume Cupboards, Ventilated Cabinets, Local Exhaust Ventilation ESTATES OFFICE OPERATIONS WORKS INSTRUCTIONS WI 6 ISSUE RECORD Revision Description Updated Date Compiled by Review date 01 Modified Mar 2010 J Cummins Mar 2011 02 Reviewed Aug 2011 J Cummins Aug 2012

More information

1811G/H PITOT-STATIC TEST SET

1811G/H PITOT-STATIC TEST SET 1811G/H PITOT-STATIC TEST SET USER INSTRUCTION MANUAL M/N: 1811G/H, P/Ns: 101-00165 101-00168 101-00169 Doc. P/N: 56-101-00165_00168_00169 Revision C December 9, 2014 BARFIELD, INC. Corporate Headquarters

More information

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc. The HumiSys RH Generator Designed, built, and supported by InstruQuest Inc. Versatile Relative Humidity Generation and Multi-Sensor System The new HumiSys with single or dual RH probes capabilities is

More information

THE CANDU 9 DISTRffiUTED CONTROL SYSTEM DESIGN PROCESS

THE CANDU 9 DISTRffiUTED CONTROL SYSTEM DESIGN PROCESS THE CANDU 9 DISTRffiUTED CONTROL SYSTEM DESIGN PROCESS J.E. HARBER, M.K. KATTAN Atomic Energy of Canada Limited 2251 Speakman Drive, Mississauga, Ont., L5K 1B2 CA9900006 and M.J. MACBETH Institute for

More information

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service? Pipette specifications according to ISO 8655 Chapter 5 5.1 The ISO 8655 standard gives the accuracy and precision limits as both absolute and relative values. Specifications will depend on the technique

More information

PHASE 1 WIND STUDIES REPORT

PHASE 1 WIND STUDIES REPORT PHASE 1 WIND STUDIES REPORT ENVIRONMENTAL STUDIES AND PRELIMINARY DESIGN FOR A SUICIDE DETERRENT SYSTEM Contract 2006-B-17 24 MAY 2007 Golden Gate Bridge Highway and Transportation District Introduction

More information

SKAC25 Operate plant to maintain the quality of pool water

SKAC25 Operate plant to maintain the quality of pool water Overview This unit is about operating swimming pool plant. The unit covers start-up and shut-down, routine maintenance and ensuring that the water quality, clarity and temperature are safe and within the

More information

General Accreditation Guidance. User checks and maintenance of laboratory balances

General Accreditation Guidance. User checks and maintenance of laboratory balances General Accreditation Guidance User checks and maintenance of laboratory balances January 2018 Copyright National Association of Testing Authorities, Australia 2010 All intellectual property rights in

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 8655-4 First edition 2002-09-15 Piston-operated volumetric apparatus Part 4: Dilutors Appareils volumétriques à piston Partie 4: Diluteurs Reference number ISO 8655-4:2002(E)

More information

CODE OF PRACTICE 42 IMPLEMENTATION OF EIGA CARBON DIOXIDE STANDARDS REVISION 2: 2018

CODE OF PRACTICE 42 IMPLEMENTATION OF EIGA CARBON DIOXIDE STANDARDS REVISION 2: 2018 CODE OF PRACTICE 42 IMPLEMENTATION OF EIGA CARBON DIOXIDE STANDARDS REVISION 2: 2018 CODE OF PRACTICE 42 IMPLEMENTATION OF EIGA CARBON DIOXIDE STANDARDS REVISION 2: 2018 Copyright 2018 by British Compressed

More information

Experiment Instructions. Circulating Pumps Training Panel

Experiment Instructions. Circulating Pumps Training Panel Experiment Instructions Circulating Pumps Training Panel Experiment Instructions This manual must be kept by the unit. Before operating the unit: - Read this manual. - All participants must be instructed

More information

IMPLEMENTATION OF EIGA CARBON DIOXIDE STANDARDS

IMPLEMENTATION OF EIGA CARBON DIOXIDE STANDARDS CODE OF PRACTICE 42 IMPLEMENTATION OF EIGA CARBON DIOXIDE STANDARDS REVISION 1 : 2014 CODE OF PRACTICE 42 IMPLEMENTATION OF EIGA CARBON DIOXIDE STANDARDS REVISION 1: 2014 Copyright 2014 by British Compressed

More information

MET-101 Introduction to Metrology

MET-101 Introduction to Metrology Pasig city Tel: 654-13-42 MET-101 Introduction to Metrology This 2-day course offers a basic introduction into Metrology, meant for people who are working in calibration and testing labs, maintenance services

More information

Navy Guidance and Tips for Using DOEHRS-IH Ventilation NAVY & MARINE CORPS PUBLIC HEALTH CENTER

Navy Guidance and Tips for Using DOEHRS-IH Ventilation NAVY & MARINE CORPS PUBLIC HEALTH CENTER Navy Guidance and Tips for Using DOEHRS-IH Ventilation NAVY & MARINE CORPS PUBLIC HEALTH CENTER October 2010 Purpose This document is a supplemental Navy guide to the DOEHRS Student Guide/User Manual Version

More information

The Complete Stability Testing for a Successful Application. Strategies

The Complete Stability Testing for a Successful Application. Strategies The Complete Staility Testing for a Successful Application, Requirements, Basic Principles Performance, Documents 1 Introduction Ojectives To secure the registration application and the successful marketing

More information

Customer Responsibilities

Customer Responsibilities Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Gas mixtures. Individual solutions specifically for your application

Gas mixtures. Individual solutions specifically for your application Gas mixtures Individual solutions specifically for your application For many applications besides pure gases also defined gas mixtures are required. For many routine applications, from operating gases

More information

A Depletion Compensated Wet Bath Simulator For Calibrating Evidential Breath Alcohol Analyzers

A Depletion Compensated Wet Bath Simulator For Calibrating Evidential Breath Alcohol Analyzers A Depletion Compensated Wet Bath Simulator For Calibrating Evidential Breath Alcohol Analyzers Slemeyer, A. University of Applied Sciences of Giessen, Dept. of Electrical Engineering 1 Wiesenstr. 14, D-35390

More information

Fall Protection. ANSI-What is it, Why is it important? Craig Firl. March 5 th, Company Private

Fall Protection. ANSI-What is it, Why is it important? Craig Firl. March 5 th, Company Private Fall Protection ANSI-What is it, Why is it important? New ANSI Standards Craig Firl March 5 th, 2013 ANSI Fall Protection Standards What is ANSI? American National Standards Institute Voluntary Consensus

More information

G.B. Collins Engineering, P.A. Comments to Proposed Code Modification SW7074-A1

G.B. Collins Engineering, P.A. Comments to Proposed Code Modification SW7074-A1 G.B. Collins Engineering, P.A. Comments to Proposed Code Modification SW7074-A1 I. REFERENCES: CURRENT CODE PROVISIONS Florida Building Code, 5th Edition (2014) Chapter 4 Special Detailed Requirements

More information

LABORATORY FUME HOOD SPECIFICATION AND MAINTENANCE PROGRAM

LABORATORY FUME HOOD SPECIFICATION AND MAINTENANCE PROGRAM LABORATORY FUME HOOD SPECIFICATION AND MAINTENANCE PROGRAM Last Updated: Tuesday, January 31, 2006 1. INTRODUCTION AND SCOPE 2. RESPONSIBILITIES 3. USER GUIDELINES 4. SYSTEM DESIGN SPECIFICATIONS o The

More information

DEVELOPING YOUR PREVENTIVE MAINTENANCE PROGRAM

DEVELOPING YOUR PREVENTIVE MAINTENANCE PROGRAM DEVELOPING YOUR PREVENTIVE MAINTENANCE PROGRAM By Owe Forsberg, Senior Consultant You are the new Reliability Engineer in a plant that currently lacks documented procedures to maintain the plant. You have

More information

New Highly Productive Phased Array Ultrasonic Testing Machine for Aluminium Plates for Aircraft Applications

New Highly Productive Phased Array Ultrasonic Testing Machine for Aluminium Plates for Aircraft Applications 19 th World Conference on Non-Destructive Testing 2016 New Highly Productive Phased Array Ultrasonic Testing Machine for Aluminium Plates for Aircraft Applications Christoph HENKEL 1, Markus SPERL 1, Walter

More information