VENTI-O 2 plus. O 2 switching valve for: VENTImotion from serial number 4000 VENTIlogic from serial number 4000 VENTImotion 2

Size: px
Start display at page:

Download "VENTI-O 2 plus. O 2 switching valve for: VENTImotion from serial number 4000 VENTIlogic from serial number 4000 VENTImotion 2"

Transcription

1 VENTI-O 2 plus O 2 switching valve for: VENTImotion from serial number 4000 VENTIlogic from serial number 4000 VENTImotion 2 Device description and operating instructions

2 Overview VENTI-O 2 plus with therapy device 2 Therapy device 3 Tube system 1 VENTI-O 2 plus 4 Test bellows VENTI-O 2 plus 6 Status indicator 7 O 2 inlet 5 O 2 outlet 8 Plug-in contact 9 Locking pins 2 Overview

3 Legend 1 VENTI-O 2 plus VENTI-O 2 plus is an O 2 switching valve. Using VENTI-O 2 plus, up to 15 l/min oxygen can be delivered to the respiratory air by the therapy device during ventilation. 2 Therapy device VENTI-O 2 plus can be used with the following therapy devices: VENTImotion from serial no.: 4000 VENTIlogic from serial no.: 4000 VENTImotion 2 3 Tube system The respiratory air enriched with oxygen flows through the tube system into the mask. 4 Test bellows The test bellows are used to perform the functional check on the VENTI-O 2 plus. 5 O 2 outlet Oxygen flows out of the O 2 outlet into the therapy device, where the oxygen is mixed with the respiratory air. 6 Status indicator Used for the functional check of the VENTI-O 2 plus. The status indicator comes on when the VENTI-O 2 plus delivers oxygen to the therapy device. 7 O 2 inlet For connecting an oxygen system (concentrator, liquid oxygen system, oxygen cylinder with pressure reducer, central gas supply system with pressure reducer). 8 Plug-in contact For supplying energy to and controlling the VENTI-O 2 plus. 9 Locking pins For positioning the VENTI-O 2 plus securely on the casing of the therapy device. Overview 3

4 Markings on the device VENTI-O 2 plus 1 2 Rating plate Symbol VENTI-O 2 plus 1 2 O 2 inlet O 2 outlet Significance Rating plate Symbol Significance Do not dispose of the device with domestic waste Year of manufacture Degree of protection against electric shock: device Type B Oxygen supply, max. 15 l / min O 2 Type of protection against electric shock: device of protection class II 4 Markings on the device

5 Markings on the pack Symbol Significance Permitted storage temperature: - 40 C to + 70 C Permitted humidity for storage: max. 95% relative humidity Protect pack from wet Do not drop pack or let it fall Safety information in these instructions Safety information in these operating instructions is marked as follows: Warning! Warns of risk of injury and potential material damage. Caution! Warns of material damage and potentially false therapy results. Note Contains useful tips. Markings on the pack 5

6 Contents 1. Description of the device Intended use Description of function Safety information Safety instructions Setting up the device Connecting to a therapy device Connecting to an oxygen system Using humidifiers Operation Switching on the device Switching off the device Hygienic preparation Intervals Cleaning/disinfecting Change of patient Functional check Intervals Performing the check Troubleshooting Servicing Intervals Disposal Product and accessories Standard product Spare parts Technical data Specifications Pneumatic plan Warranty Declaration of Conformity Index Contents

7 1. Description of the device 1.1 Intended use VENTI-O 2 plus can be used with following Weinmann devices: VENTImotion from serial number 4000 VENTIlogic from serial number 4000 VENTImotion 2 The VENTI-O 2 plus can be used to deliver up to 15 l/min oxygen (O 2 ) into respiratory air during ventilation. The VENTI-O 2 plus can be used in combination with an oxygen concentrator, an oxygen cylinder with pressure reducer, a continuous-flow liquid oxygen system or a central gas supply system with pressure reducer. The oxygen system must have its own flow control device. Warning! The VENTI-O 2 plus may only be used for patients who are not at risk from an undetected failure of the oxygen supply as the VENTI-O 2 plus - like the therapy device - does not have an alarm to indicate oxygen delivery faults. Only use the device for the purpose described here. 1.2 Description of function The VENTI-O 2 plus is clicked onto the side of the therapy device. It is controlled and supplied with energy by the therapy device. If the therapy device and the oxygen system are switched on and working properly, the VENTI-O 2 plus automatically starts delivering oxygen. The green status indicator of the VENTI-O 2 plus comes on. If the therapy device is switched off or if a fault (e.g. power failure) occurs, the VENTI-O 2 plus stops delivering oxygen to the therapy device. The green status indicator of the VENTI-O 2 plus goes out. Description of the device 7

8 2. Safety information 2.1 Safety instructions Read these operating instructions carefully. They are an integral part of the device and must be available at all times. Only use the device for the purpose described (see 1.1 Intended use on page 7). For your own safety and the safety of your patients and in accordance with the requirements of Directive 93/42/EEC, please note the following points. Oxygen delivery Warning! Smoking and naked flame are prohibited during delivery of oxygen into the respiratory flow. Risk of fire. Oxygen can accumulate in clothing, carpets, bedding or hair. Ventilate thoroughly after use to reduce the accumulation of oxygen. Keep all parts free of oil and grease. Hydrocarbon compounds such as oils, greases, cleaning alcohols, hand cream and adhesive plasters can lead to explosive reactions in combination with oxygen. If a fault occurs, the VENTI-O 2 plus stops delivering oxygen. It is therefore essential in the event of a fault that you shut off your oxygen system to ensure that the tube connection between the oxygen system and the VENTI-O 2 plus does not come loose. If the tube connection should have come loose, shut off your oxygen system. Ventilate the room thoroughly after use to reduce the accumulation of oxygen. The amount delivered must be set using a flow control device (not included in the scope of supply). The flow rate may not exceed 15 l/min for this. Higher delivery rates can lead to accumulation of oxygen in the therapy device. Risk of fire. Never operate the VENTI-O 2 plus in a bag. Oxygen may accumulate in the bag. Risk of fire. Caution! Oxygen therapy can have adverse effects if used/dosed incorrectly. For this reason you must only use the device as prescribed by a doctor. Make sure that it always delivers the flow rate prescribed by the doctor. 8 Safety information

9 Always follow the safety information in the operating instructions for your oxygen system! Use only oxygen systems with a maximum static pressure not exceeding 1000 hpa. At higher static pressures, a tube connection in the oxygen supply system may come loose. Make sure that all components used are designed for the static pressure of your oxygen system, otherwise there is a risk of damage to your device. Operating the device Caution! Only use the device for the purpose described (see 1. Description of the device on page 7). To prevent infection or bacterial contamination, observe section 5. Hygienic preparation on page 13. Do not allow liquids to enter the therapy device or the casing of the VENTI-O 2 plus. This could damage the electronic system. Do not use any antistatic or electrically conductive tubes. Maintain a safe distance between the therapy device and equipment which emits HF radiation (e.g. cell phones). The device may otherwise malfunction (see operating instructions for the therapy device). Accessories/Spare parts/repairs Caution! Malfunctions and bioincompatibility may result if third-party articles are used. Please note that in such cases, all warranty entitlement and liability claims shall be void where items other than the accessories or original spare parts recommended in the operating instructions are used. Have repairs carried out only by the manufacturer - Weinmann - or consult your specialist dealer. Safety information 9

10 3. Setting up the device 3.1 Connecting to a therapy device 1. Set up your therapy device as described in the relevant operating instructions. Then assemble the VENTI-O 2 plus as follows. 2. Clip the VENTI-O 2 plus onto the side of the therapy device. Ensure that it engages fully. 3.2 Connecting to an oxygen system To connect to the oxygen system, you need an oxygen tube with an internal diameter of 4 mm. To ensure proper functioning of the VENTI-O 2 plus, the tube should be no longer than 20 m. Connecting to an oxygen concentrator 1. Set up your oxygen concentrator as described in the relevant operating instructions. Do not switch on the concentrator yet. 2. Fit the oxygen tube of your concentrator onto the O 2 inlet of the VENTI-O 2 plus. Connecting to an oxygen cylinder with pressure reducer 1. Set up your oxygen cylinder with pressure reducer as described in the relevant operating instructions. Do not open the cylinder valve yet. 2. Fit the oxygen tube of your cylinder system to the O 2 inlet of the VENTI-O 2 plus. 10 Setting up the device

11 Connecting to a continuous-flow liquid oxygen system Connecting to a central gas supply system 1. Set up your liquid oxygen system as described in the relevant operating instructions. Leave the flow control in position Fit the oxygen tube of your liquid oxygen system to the O 2 inlet of the VENTI-O 2 plus. 1. Insert the flow control with pressure reducer into the appropriate wall socket. Do not open the flow control yet. 2. Connect the flow control and the O 2 inlet of the VENTI-O 2 plus to the oxygen tube. 3.3 Using humidifiers You can use the VENTIclick humidifier to humidify respiratory air. Alternatively, an HME filter can be used close to the patient. Additional humidification of the oxygen using bubble or sterile water humidifiers is permitted, but not necessary. A humidifier needs to be protected from excessive pressure. On humidifiers with an excess pressure valve, it should be ensured that no oxygen escapes at the excess pressure valve during operation, otherwise too little oxygen is mixed into the respiratory air. Follow the operating instructions for your humidifier. The use of nebulizers is not permitted. Caution Make sure that the bubble/sterile water humidifier used is designed for the maximum static pressure of your oxygen system. Follow the operating instructions for the VENTIclick and for the bubble humidifier, otherwise there is a risk of damage to your device. Setting up the device 11

12 4. Operation 4.1 Switching on the device VENTI-O 2 plus is ready for operation once the device has been connected as described in 3. Setting up the device on page 10. To start therapy, first switch on your therapy device and then your oxygen system (at the prescribed flow rate, max. 15 l/min). The VENTI-O 2 plus automatically enables oxygen delivery. The green status indicator comes on. There is no further need to control the device during operation. 4.2 Switching off the device Please observe the following sequence when switching off the devices. 1. Switch off the O 2 supply on your oxygen system; depending on the system: switch off oxygen concentrator, close cylinder valve (cylinder oxygen), close flow control (liquid oxygen) or disconnect therapy device from the central gas supply system. 2. Switch off your therapy device. This also switches off the VENTI-O 2 plus. The green status indicator goes out. 12 Operation

13 5. Hygienic preparation 5.1 Intervals The casing should be wiped down regularly. 5.2 Cleaning/disinfecting Cleaning the casing 1. Detach the tube from the VENTI-O 2 plus. 2. Detach the VENTI-O 2 plus from the therapy device. 3. Wipe the casing, the plug-in contact, the locking pins and the O 2 inlet with a soft damp cloth. The plug-in contact and the O 2 inlet must be completely dry before you start up the device. Disinfecting the casing Clean the casing, the plug-in contact, the locking pins, the O 2 inlet, the O 2 outlet and the connecting line to the O 2 system simply by wiping with disinfectant. We recommend TERRALIN for this purpose. We recommend that you wear suitable gloves (e.g. household or disposable gloves) for disinfecting. Refer to the operating instructions for the disinfectant used. Sterilizing Sterilizing is not permitted. 5.3 Change of patient If there is a change of patient, replace the tube system and clean the outside of the VENTI-O 2 plus by wiping with disinfectant. Hygienic preparation 13

14 6. Functional check 6.1 Intervals Perform a functional check every day. 6.2 Performing the check Use the following test to see whether the valve of the VENTI-O 2 plus is opening and closing correctly. Proceed as follows for the test. Connect the VENTI-O 2 plus to the therapy device. Do not switch on the therapy device. The status indicator on top of the VENTI-O 2 plus does not come on. 1. Slip the opening of the test bellows over the O 2 inlet of the VENTI-O 2 plus. 2. Squeeze the test bellows. The device is working correctly if you are unable to force air through the valve. The test bellows can only be squeezed with difficulty. You feel considerable resistance. The test bellows remain firm and retain their round shape. 3. Plug the drying adapter onto the outlet of the therapy device. 4. Switch on your therapy device. The therapy device starts working. The status display of the VENTI-O 2 plus comes on and the valve opens with a soft click. 14 Functional check

15 5. Squeeze the test bellows. The device is working correctly if you are able to force air through the valve. The test bellows are easy to squeeze. The test bellows remain in the squeezed position. 6. Switch off your therapy device. The test bellows remain squeezed. Take the test bellows off the O 2 inlet of the VENTI-O 2 plus. If your VENTI-O 2 plus does not work as described above, please contact your specialist dealer or the manufacturer - Weinmann - to have the device repaired. Functional check 15

16 7. Troubleshooting Fault Cause Remedy Status indicator does not No power supply come on or comes on only from time to time, although the therapy Leak in breathing mask device is working properly. Check plug-in contact for firm fit Adjust cap so that mask is a firm fit. If necessary, replace faulty mask. Flow rate does not reach prescribed level. O 2 supply tubes kinked Outlet pressure of oxygen system not high enough Check tube connections. If necessary use a different oxygen system (e.g. OXYMAT 3). Passage in therapy device blocked Please contact your specialist dealer. Warning! If there are faults which cannot be eliminated immediately, switch off the device and the oxygen supply and do not use any more. Otherwise there is a risk of fire. Contact the manufacturer - Weinmann - or your specialist dealer to have the device repaired. 16 Troubleshooting

17 8. Servicing 8.1 Intervals The VENTI-O 2 plus requires no servicing. 9. Disposal Disposal Do not dispose of the device as domestic waste. To dispose of the valve properly, contact a licensed, certified electronic scrap recovery firm. You can find out their address from your environment officer or from your municipal authorities. Servicing 17

18 10. Product and accessories 10.1 Standard product Oxygen switching valve VENTI-O 2 pluswm Description of component Order no. Oxygen switching valve VENTI-O 2 plus, basic device WM Test bellows WM Operating instructions VENTI-O 2 plus WM Spare parts Description of component Order no. Test bellows WM Operating instructions VENTI-O 2 plus WM Product and accessories

19 11. Technical data 11.1 Specifications VENTI-O 2 plus Product class under 93/42/EEC II a Dimensions W x H x D in cm 8.0 x 9.3 x 3.0 Weight approx. 190 g Temperature range Operation Storage Air pressure range Pneumatic connection Inlet Outlet Permitted humidity for operation and storage Electrical connection Current consumption during operation Classification to EN Type of protection against electric shock - Degree of protection against electric shock Electromagnetic compatibility (EMC) to EN Radio interference suppression Radio interference immunity +5 C to +35 C 40 C to +70 C hpa only tubes with 4 mm internal Ø O 2 outlet into therapy device 95 % rh (no condensation) 12 V DC 135 ma Class II Type B EN EN to 3 EN Max. permitted flow rate 15 l/min. O 2 Min. required O 2 supply pressure at valve inlet for 4 l/min O 2 for 15 l/min O 2 Max. permitted static pressure of pressure source Permitted therapy pressures 100 hpa 300 hpa 1000 hpa 4 to 35 hpa (1 hpa = 1 mbar 1 cm H 2 O) We reserve the right to make design changes. Technical data 19

20 11.2 Pneumatic plan VENTI-O 2 plus External oxygen supply Control by therapy device Delivery of O 2 to the therapy device 20 Technical data

21 12. Warranty Weinmann offers a warranty that the product, when used in accordance with requirements, will remain free from defects for a period of two years from date of purchase. For products whose durability is clearly indicated as less than two years, the warranty expires on the expiration date indicated on the packaging or in the user s manual. Claims against the warranty can be made only when accompanied by the sales receipt, which must show salesperson and date of purchase. We offer no warranty in the case of: Disregard of usage instructions Operating errors Improper use or improper handling Third-party intervention by non-authorized persons for the purpose of device repair Acts of God, e.g., lightning strikes, etc. Transport damage as a result of improper packaging of returned items Lack of maintenance Operational and normal wear and tear, which includes, for example, the following components. Filter Batteries / recheargable batteries Articles for one-time usage, etc. Failure to use original spare parts. Weinmann is not liable for consequential harm caused by a defect if it is not based on intention or gross negligence. Weinmann is also not liable for minor physical injury to life or limb resulting from negligence. Weinmann reserves the right to decide whether to eliminate defects, to deliver a defectfree item or to reduce the purchase price by a reasonable amount. If Weinmann rejects a claim against the warranty, it assumes no expense for transport between customer and manufacturer. Implied warranty claims remain unaffected by these changes. Warranty 21

22 13. Declaration of Conformity Weinmann Geräte für Medizin GmbH + Co. KG declares herewith that the product complies fully with the respective regulations of the Medical Device Directive 93/42/EEC. The unabridged text of the Declaration of Conformity can be found on our website at 22 Declaration of Conformity

23 14. Index C Change of patient 13 Cleaning 13 D Disinfecting 13 F Functional check 14 O Overview 2 P Pneumatic plan 20 S Safety instructions 8 Servicing 17 Spare parts 9, 18 T Technical data 19 Troubleshooting 16 W Warranty 21 Index 23

24 Weinmann Geräte für Medizin GmbH + Co. KG P.O. Box D Hamburg Kronsaalsweg 40 D Hamburg T: +49-(0) F: +49-(0) E: info@weinmann.de Center for Production, Logistics, Service Weinmann Geräte für Medizin GmbH + Co. KG Siebenstücken 14 D Henstedt-Ulzburg WM 66831b 12/2012 EN

USER MANUAL SAVE THESE INSTRUCTIONS. For the most current manual revision, please visit our Website:

USER MANUAL SAVE THESE INSTRUCTIONS. For the most current manual revision, please visit our Website: USER MANUAL Model: PM4300 Series SAVE THESE INSTRUCTIONS For the most current manual revision, please visit our Website: www.precisionmedical.com 300 Held Drive Tel: (+001) 610-262-6090 Northampton, PA

More information

Instrument Description and Operating Instructions OXYMAT 2. Its own oxygen source

Instrument Description and Operating Instructions OXYMAT 2. Its own oxygen source Instrument Description and Operating Instructions OXYMAT 2 Its own oxygen source OXYMAT 2 Fig. 1 5 6 1 15 2 3 4 8 15 7 9 12 10 13 11 14 Legends 1 OXYMAT 2 oxygen concentrator 2 Oxygen outlet 3 Main switch

More information

Pressure Regulator. Instructions for Use

Pressure Regulator. Instructions for Use Pressure Regulator Instructions for Use 702-0083.9 December 2017 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in injury to the patient, the

More information

Portable oxygen concentrator

Portable oxygen concentrator Table of contents G3 1. Introduction 2 2. Important Safety Rules 3 3. Unpack your Lovego 4 4. User Controls & System Status Indicators 5 5. Quick Start 6 6. How to operate 7 7. How to operate with battery

More information

EasyFit Lite Nasal Mask, Silicone. Nasal Mask. EasyFit Lite Nasal Mask. Device Description and Instructions for Use. Caution:

EasyFit Lite Nasal Mask, Silicone. Nasal Mask. EasyFit Lite Nasal Mask. Device Description and Instructions for Use. Caution: EasyFit Lite Nasal Mask EasyFit Lite Nasal Mask, Silicone Nasal Mask Device Description and Instructions for Use Caution: Federal Law (U.S.A.) restricts this device to sale by or on the order of a physician.

More information

Oxygen Dialflow Meter. Instructions for Use

Oxygen Dialflow Meter. Instructions for Use Oxygen Dialflow Meter Instructions for Use 702-0031.9 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

Oxygen Dialflow Meter. Instructions for Use

Oxygen Dialflow Meter. Instructions for Use Oxygen Dialflow Meter Instructions for Use 702-0031.12 December 2017 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in injury to the patient,

More information

U S E R M A N U A L. Model: PM4300 Series SAVE THESE INSTRUCTIONS

U S E R M A N U A L. Model: PM4300 Series SAVE THESE INSTRUCTIONS U S E R M A N U A L Model: PM4300 Series (PM4351 Shown) SAVE THESE INSTRUCTIONS CAUTION Federal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive Tel: (+001) 610-262-6090

More information

U S E R M A N U A L AIR FLOWMETER. MODELS: 1MFA2001 (shown) 1MFA9001 CAUTION. ISO Certified

U S E R M A N U A L AIR FLOWMETER. MODELS: 1MFA2001 (shown) 1MFA9001 CAUTION. ISO Certified U S E R M A N U A L AIR FLOWMETER MODELS: 1MFA2001 (shown) 1MFA9001 ISO 13485 Certified Authorized Your EU local Representative: distributor: EMERGO EUROPE, INC. Molenstraat 15 2513866 BH -624 The -3952

More information

Instruction Manual for Configura Cushionair Portable Pump

Instruction Manual for Configura Cushionair Portable Pump Instruction Manual for Configura Cushionair Portable Pump Fitted with battery powered pump, suitable for Configura Portable chairs V E R S I O N O N E M A Y 2 0 1 6 Contents Introduction 3 Set up of Cushionair

More information

BOC: Living healthcare. Manual. LIV IQ BOC Integrated Valve with digital display portable delivery system for Medical Oxygen. BOC: Living healthcare

BOC: Living healthcare. Manual. LIV IQ BOC Integrated Valve with digital display portable delivery system for Medical Oxygen. BOC: Living healthcare BOC: Living healthcare Manual LIV IQ BOC Integrated Valve with digital display portable delivery system for Medical Oxygen. BOC: Living healthcare 02 Manual LIV IQ Oxygen Manual LIV IQ Oxygen 03 Contents.

More information

Pressure Injection Cell Operator s Manual

Pressure Injection Cell Operator s Manual Pressure Injection Cell Operator s Manual Operator s Manual Pressure Injection Cell CONGRATULATIONS! Congratulations on your purchase of a Next Advance Pressure Injection Cell. Please read this operator

More information

Med Aire Alternating Pressure Pump and Pad System

Med Aire Alternating Pressure Pump and Pad System User Manual Med Aire Alternating Pressure Pump and Pad System 14002E 14001E Symbols & Statements NOTE Indicates some tips or some information users should be aware of. CAUTION Indicates correct operating

More information

USER MANUAL SAVE THESE INSTRUCTIONS. 300 Held Drive Tel: (+001) Northampton, PA USA Fax: (+001) ISO Certified

USER MANUAL SAVE THESE INSTRUCTIONS. 300 Held Drive Tel: (+001) Northampton, PA USA Fax: (+001) ISO Certified USER MANUAL Model: PM4300 Series (PM4351 Shown) SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive Tel: (+001) 610-262-6090 Northampton,

More information

453 Series Steam Heated Vaporizing Regulator

453 Series Steam Heated Vaporizing Regulator ADI 0453A Certified ISO 9001:2000 453 Series Steam Heated Vaporizing Regulator INSTALLATION AND OPERATION INSTRUCTIONS Before Installing or Operating, Read and Comply with These Instructions Controls Corporation

More information

Micro Dial-Flowmeter. Instructions for Use

Micro Dial-Flowmeter. Instructions for Use Micro Dial-Flowmeter Instructions for Use 702-0082.8 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

VENTIlogic LS VENTIlogic plus. 100 % Mobility and Reliability in IV and NIV

VENTIlogic LS VENTIlogic plus. 100 % Mobility and Reliability in IV and NIV VENTIlogic LS VENTIlogic plus 100 % Mobility and Reliability in IV and NIV VENTIlogic VENTIlog LS Your Requirements for Reliability and Mobility are Our Benchmark. VENTIlogic LS and VENTIlogic plus are

More information

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator Operating Instructions Table of Contents Resource v.2 Conserving Regulator Safety Information Device Precautions Introduction Product Features Product Specifications Feature Illustrations Set Up Usage

More information

Safety. Operating instructions Solenoid valve for gas VG 6 VG 15/10 DANGER. Contents WARNING CAUTION. Changes to edition 09.14

Safety. Operating instructions Solenoid valve for gas VG 6 VG 15/10 DANGER. Contents WARNING CAUTION. Changes to edition 09.14 15 Elster GmbH Edition 7.15 Translation from the German 519 D F NL I E DK S N P GR TR CZ PL RUS H www.docuthek.com Operating instructions Solenoid valve for gas VG VG 15/1 Contents Solenoid valve for gas

More information

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician. U S E R M A N U A L 1600 SERIES OXYGEN REGULATOR 168715G (Shown) SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive Tel: (+001) 610-262-6090

More information

AND OPERATION INSTRUCTIONS

AND OPERATION INSTRUCTIONS ADI 2111-D Certified ISO 9001 MEDICAL FLOWMETER (Back Pressure Compensated Thorpe Tube) INSTALLATION AND OPERATION INSTRUCTIONS Before Installing or Operating, Read and Comply with These Instructions Controls

More information

AIR HAMMERS MODEL NO: CAT138/CAT139 OPERATING & MAINTENANCE INSTRUCTIONS PART NO: / GC064

AIR HAMMERS MODEL NO: CAT138/CAT139 OPERATING & MAINTENANCE INSTRUCTIONS PART NO: / GC064 AIR HAMMERS MODEL NO: CAT138/CAT139 PART NO: 3120152 /3120153 OPERATING & MAINTENANCE INSTRUCTIONS GC064 INTRODUCTION Thank you for purchasing this CLARKE Air Hammer. Before attempting to use this product,

More information

VENTIlogic LS VENTIlogic plus. 100 % Mobility and Reliability in IV and NIV

VENTIlogic LS VENTIlogic plus. 100 % Mobility and Reliability in IV and NIV VENTIlogic LS VENTIlogic plus 100 % Mobility and Reliability in IV and NIV VENTIlogic LS VENTIlogic plus Your requirements for reliability and mobility are our benchmark. VENTIlogic LS and VENTIlogic plus

More information

BiLevel ST 22. Ventilation and Sleep Therapy Device. BiLevel ST 22. Description of unit and instructions for use

BiLevel ST 22. Ventilation and Sleep Therapy Device. BiLevel ST 22. Description of unit and instructions for use BiLevel ST 22 Ventilation and Sleep Therapy Device BiLevel ST 22 Description of unit and instructions for use Overview BiLevel ST 22 17 Hood 16 Nasal mask 15 Exhalation system 14 Hose system 11 Drying

More information

Dialflow Regulator. Instructions for Use

Dialflow Regulator. Instructions for Use Dialflow Regulator Instructions for Use 702-0030.11 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

User Instruction Manual

User Instruction Manual User Instruction Manual 4500 psi Air Compressor Ver 2, 1.18 Contents Parts Included...3 Assembly Instructions...3-5 Operation Instructions...6-7 Oil Change Intervals...8 Air Filter Replacement...9 Setting

More information

CUT OFF TOOL MODEL: CAT113

CUT OFF TOOL MODEL: CAT113 CUT OFF TOOL MODEL: CAT113 Part No: 3120135 ASSEMBLY & INSTRUCTION MANUAL LS0309 INTRODUCTION Thank you for purchasing this CLARKE product Before attempting to use the product, it is essential that you

More information

CLEMENTS. ANA Oxygen Regulator with Yoke Connector. User Manual. Manual No. ANA Issue 1

CLEMENTS. ANA Oxygen Regulator with Yoke Connector. User Manual. Manual No. ANA Issue 1 ANA 74300 Oxygen Regulator with Yoke Connector User Manual Manual No. ANA 74300 001 Issue 1 Safety Thank you for purchasing this Clements Oxygen Regulator. For your safety it is imperative that this unit

More information

VERTICAL AIR COMPRESSORS

VERTICAL AIR COMPRESSORS VERTICAL AIR COMPRESSORS MODEL NO: VE15C150, VE18C150, VE25C150 PART NO: 2226010, 2226020, 2226025 OPERATION & MAINTENANCE INSTRUCTIONS LS0715 INTRODUCTION Thank you for purchasing this CLARKE Vertical

More information

Rev. B. Operating Instructions

Rev. B. Operating Instructions 780745-00 Rev. B Operating Instructions 780745-00 Rev. B TABLE OF CONTENTS PAGE Warnings, Cautions, and Notes..............................4-7 Definition of Symbols......................................

More information

Operation Manual. O2 Quickstick. Oxygen Analyzer 08.17

Operation Manual. O2 Quickstick. Oxygen Analyzer 08.17 Operation Manual O2 Quickstick Oxygen Analyzer 08.17 If you have any questions on this equipment please contact Technical Support at: Nuvair 1600 Beacon Place Oxnard, CA 93033 Phone: 805-815-4044 FAX:

More information

USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS. Federal (USA) law restricts this device to sale by or on the order of a physician.

USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS. Federal (USA) law restricts this device to sale by or on the order of a physician. USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS CAUTION Federal (USA) law restricts this device to sale by or on the order of a physician. RECEIVING / INSPECTION Remove the Precision Medical,

More information

VERTICAL AIR COMPRESSORS

VERTICAL AIR COMPRESSORS VERTICAL AIR COMPRESSORS MODEL NO: VE11C150, VE15C150, VE18C150 PART NO: 2226005, 2226000, 2226015 OPERATION & MAINTENANCE INSTRUCTIONS LS0615 INTRODUCTION Thank you for purchasing this CLARKE Vertical

More information

Operating Instructions in accordance with the Machines Directive 2006/42/EC

Operating Instructions in accordance with the Machines Directive 2006/42/EC Lubricating Equipment Metering Technology for Grease and Oil High Pressure Hydraulic Customised Developments Operating Instructions in accordance with the Machines Directive 2006/42/EC Compressed air filling

More information

OPERATOR S MANUAL. Pediatric/Infant cap-one Mask YG-232T/YG-242T A. General. Safety Information WARNING WARNING WARNING. Components WARNING

OPERATOR S MANUAL. Pediatric/Infant cap-one Mask YG-232T/YG-242T A. General. Safety Information WARNING WARNING WARNING. Components WARNING OPERATOR S MANUAL 0614-905493A First Edition: 12 Dec 2011 Second Edition: 12 Feb 2015 Printed: Pediatric/Infant cap-one Mask YG-232T/YG-242T Symbol Description Attention, consult operator s manual Date

More information

Safety Powder Spray Systems

Safety Powder Spray Systems Instruction Sheet P/N 107 952C Safety Powder Spray Systems 1. Introduction This section contains general safety instructions for using your Nordson equipment. Task- and equipment-specific warnings are

More information

ACRYLIC FLOW METERS 0-30, 0-70 & LPM

ACRYLIC FLOW METERS 0-30, 0-70 & LPM ACRYLIC FLOW METERS 0-30, 0-70 & 0-120 LPM ASSEMBLY INSTRUCTIONS & INSTRUCTIONS FOR USE R219P86 (0-120LPM) R219P87 (0-70LPM) R219P88 (0-30 LPM) R138P11 Rev. D TABLE OF CONTENTS: 1.0 PRODUCT OVERVIEW...1

More information

testo Leakage detector for refrigerants Instruction manual

testo Leakage detector for refrigerants Instruction manual testo 316-3 Leakage detector for refrigerants Instruction manual 2 1 Contents 1 Contents 1 Contents... 3 2 Safety and the environment... 4 2.1. About this document... 4 2.2. Ensure safety... 4 2.3. Protecting

More information

Model ASSEMBLY and OPERATING INSTRUCTIONS

Model ASSEMBLY and OPERATING INSTRUCTIONS QUICK CHANGE AIR BRUSH KIT Model 93506 ASSEMBLY and OPERATING INSTRUCTIONS Due to continuing improvements, actual product may differ slightly from the product described herein. 3491 Mission Oaks Blvd.,

More information

Operating and maintenance manual Filter and reducing station Series / 1.0

Operating and maintenance manual Filter and reducing station Series / 1.0 Operating and maintenance manual Filter and reducing station Series 961 04.2017 / 1.0 Original instructions ARCA Regler GmbH. All rights reserved. Cover picture background: Freepik.com ARCA Regler GmbH

More information

Solenoid gas valve. The solenoid gas valve is also applicable as a main gas valve Pilot gas valve for forced draft burners

Solenoid gas valve. The solenoid gas valve is also applicable as a main gas valve Pilot gas valve for forced draft burners 7 634 VGS1... with AGA67 connector VGS2... with AGA67 connector Solenoid gas valve VGS The solenoid gas valve is also applicable as a main gas valve Pilot gas valve for forced draft burners The VGS and

More information

MODEL Maverick Heavy Duty Aerosol Compressor Operator s Manual

MODEL Maverick Heavy Duty Aerosol Compressor Operator s Manual MODEL 50012 Maverick Heavy Duty Aerosol Compressor Operator s Manual This page left blank intentionally CONTENTS Page No. INSPECTION... 1 GENERAL INFORMATION... 1 SAFETY PRECAUTIONS... 2 BASIC OPERATING

More information

U S E R M A N U A L. (CGA 870 Connection)

U S E R M A N U A L. (CGA 870 Connection) U S E R M A N U A L 1900 Series (CGA 870 Connection) SAVE THESE INSTRUCTIONS CAUTION Federal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive Tel: (+001) 610-262-6090

More information

Instructions for use Compressed air injector with water trap

Instructions for use Compressed air injector with water trap Instructions for use Compressed air injector with water trap Table of Contents Introduction 1. Description of Functions 2. Putting into Operation 3. Instructions for Cleaning and Care 3.1 Disassembling

More information

Oxygen Concentrator Instruction

Oxygen Concentrator Instruction WARNING-Read instruction before operating this equipment Oxygen Concentrator Instruction K5BW WARNING Oxygen therapy can be hazardous in certain conditions. Seeking medical advice before using an oxygen

More information

OWNER S MANUAL. Page: 1 of 8

OWNER S MANUAL. Page: 1 of 8 Air Needle Scaler OWNER S MANUAL WARNING: Read carefully and understand all INSTRUCTIONS before operating. Failure to follow the safety rules and other basic safety precautions may result in serious personal

More information

Operating Instructions Temperature Measuring System MF420-5T-100

Operating Instructions Temperature Measuring System MF420-5T-100 Operating Instructions Temperature Measuring System Dnhqdhnjah MF420-5T-100 Dmsklakla Read before use! Observe all safety instructions! Keep for future reference! Fon +49/7221/64103 Fax +49/7221/17103

More information

Innovative Vacuum for Automation. Operating Instructions Dust Filter STF P / STF-D P.

Innovative Vacuum for Automation. Operating Instructions Dust Filter STF P / STF-D P. Innovative Vacuum for Automation EN Operating Instructions Dust Filter STF P / STF-D P 30.30.01.00059/02 11.2014 www.schmalz.com Note These operating instructions were written in the German language. This

More information

PLEASE READ CAREFULLY BEFORE INSTALLING OR USING MEGA POOL SAVER MPS 1100

PLEASE READ CAREFULLY BEFORE INSTALLING OR USING MEGA POOL SAVER MPS 1100 MPS-1100 User Manual Mega Pool Saver Ltd PLEASE READ CAREFULLY BEFORE INSTALLING OR USING MEGA POOL SAVER MPS 1100 For further up to date instructions on how to install Mega Pool Saver MPS 1100, please

More information

SOMNOvent CR. CS therapy device with autotrilevel principle. SOMNOvent CR with SOMNOclick 300 SOMNOvent CR 230 V

SOMNOvent CR. CS therapy device with autotrilevel principle. SOMNOvent CR with SOMNOclick 300 SOMNOvent CR 230 V SOMNOvent CR CS therapy device with autotrilevel principle SOMNOvent CR SOMNOvent CR with SOMNOclick 300 SOMNOvent CR 230 V Device description and instructions for use with devices as of serial number

More information

EvolutionTM OM-900 PRODUCT INFORMATION AND INSTRUCTIONS OXYGEN CONSERVER

EvolutionTM OM-900 PRODUCT INFORMATION AND INSTRUCTIONS OXYGEN CONSERVER EvolutionTM OXYGEN CONSERVER PRODUCT INFORMATION AND INSTRUCTIONS OM-900 TABLE OF CONTENTS General Information....3 Important Safety Rules and Precautions Operating the Unit... 4-6 Accessories/Repairs....6

More information

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc. Advanced Calibration Designs, Inc. Calibration Gas Instrument INSTRUCTION MANUAL Release I www.goacd.com Instruction Manual Gas Generator Release I TABLE OF CONTENTS I. General Description Page 2 II. Start-Up

More information

MODEL NUMBER: PSI AIR SOURCE KIT 200 PSI Compressor on 2.0 Gallon 200 PSI Air Tank

MODEL NUMBER: PSI AIR SOURCE KIT 200 PSI Compressor on 2.0 Gallon 200 PSI Air Tank IMPORTANT SAFETY INSTRUCTIONS CAUTION - To reduce risk of electrical shock or Electrocution: MODEL NUMBER: 20008 200 PSI AIR SOURCE KIT 200 PSI Compressor on 2.0 Gallon 200 PSI Air Tank IMPORTANT: It is

More information

- GAS REGULATOR - COMPACT STYLE

- GAS REGULATOR - COMPACT STYLE INSTRUCTIONS FOR USE - GAS REGULATOR - COMPACT STYLE Preset at 50 or 60 psi Rx ONLY GR - XXX - [(TT) (UU) - (XX)(Z)] GR XXX (TT) (UU) (XX) (Z) Gas Connection Outlet Flowrate Addition Outlet Additional

More information

MODEL NUMBER: M20005 AIR SOURCE KIT. 30% Duty Compressor on. 2.0 Gallon Air Tank SAVE THIS MANUAL FOR FUTURE REFERENCE

MODEL NUMBER: M20005 AIR SOURCE KIT. 30% Duty Compressor on. 2.0 Gallon Air Tank SAVE THIS MANUAL FOR FUTURE REFERENCE MODEL NUMBER: M20005 AIR SOURCE KIT 30% Duty Compressor on 2.0 Gallon Air Tank SAVE THIS MANUAL FOR FUTURE REFERENCE USER MANUAL IMPORTANT SAFETY INSTRUCTIONS CAUTION - To reduce risk of electrical shock

More information

User manual. ZX trolley.

User manual. ZX trolley. User manual. ZX trolley. Date July 2011 02 User manual. ZX trolley. ZX trolley. Safe transportation for medical gas cylinders. Important Please read this manual carefully before attempting to use the BOC

More information

CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR RESUSCITATOR

CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR RESUSCITATOR CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR OXYLATOR FR-300 RESUSCITATOR Symbols used in the Operating Manual for the Oxylator FR-300 and/or on the Oxylator FR-300 unit: WARNING! Do not use grease or

More information

The Univentor 1250 Anaesthesia Unit

The Univentor 1250 Anaesthesia Unit THE UNIVENTOR 1200/1250 ANAESTHESIA UNIT The Univentor 1250 Anaesthesia Unit TABLE OF CONTENTS EDITION 1 Section 1 - WARRANTY & SERVICE 1.1. WARRANTY 2 1.2. DAMAGED SHIPMENTS 2 1.3. SERVICE 2 Section 2

More information

AUTOVENT 4000 VENTILATOR

AUTOVENT 4000 VENTILATOR OVERVIEW AUTOVENT 4000 Only properly trained and approved Escambia County Bureau of Public Safety Paramedics are to use the AutoVent 4000 ventilator manufactured by LSP to transport patients already on

More information

Vapotherm Q50 Compressor. Instructions for Use

Vapotherm Q50 Compressor. Instructions for Use Vapotherm Q50 Compressor Instructions for Use Contents Quick Start Guide...2 Intended Use...3 Indications, Warnings and Cautions...4 Primary Indications...4 Contraindications...4 Warnings and Cautions...4

More information

Operation Manual. O2 Quickstick. Oxygen Analyzer

Operation Manual. O2 Quickstick. Oxygen Analyzer Operation Manual O2 Quickstick Oxygen Analyzer 01.11 www.nuvair.com If you have any questions on this equipment please contact Technical Support at: Nuvair 2949 West 5 th St. Oxnard, CA 93030 Phone: 805-815-4044

More information

4-in-1 Professional Inflator Kit

4-in-1 Professional Inflator Kit 4-in-1 Professional Inflator Kit Owner s Manual WARNING: Read carefully and understand all ASSEMBLY AND OPERATION INSTRUCTIONS before operating. Failure to follow the safety rules and other basic safety

More information

AC1810 / AC1810-A TECHNICAL SPECIFICATIONS. Operating Pressure psi ( kgs/cm²) [AC1810] Displacement. Net Weight

AC1810 / AC1810-A TECHNICAL SPECIFICATIONS. Operating Pressure psi ( kgs/cm²) [AC1810] Displacement. Net Weight Technical Specifications Operating Instructions Maintenance Information Troubleshooting Guide Parts Diagrams AC1810 / AC1810-A THE EVOLUTION OF PERFECTION CAUTION: Before attempting to use or service this

More information

Home Oxygen Therapy in NHS Scotland. Level 1 Education for Patients & Carers

Home Oxygen Therapy in NHS Scotland. Level 1 Education for Patients & Carers Home Oxygen Therapy in NHS Scotland Level 1 Education for Patients & Carers National Home Oxygen Service Guidance/Best Practice Recommendations Changes to the National Oxygen Service in Scotland in 2012-2013

More information

RARS5000 AIR BODY SAW OWNER S OPERATING MANUAL

RARS5000 AIR BODY SAW OWNER S OPERATING MANUAL RARS5000 AIR BODY SAW OWNER S OPERATING MANUAL DESCRIPTION 1. No mar 2. No mar tip 3. Housing grip 4. Trigger 5. Air inlet 6. Air inlet plug 7. Plastic board Important! It is essential that you read the

More information

OXYWAY. Service and Repair Instructions

OXYWAY. Service and Repair Instructions OXYWAY Single-stage pressure reducer, fixed: Fix I; Fix I side outlet; Fix II; Fix III; Fix III left Single-stage pressure reducer, variable: Fine I; Fine II; Fine III Single-stage pressure reducer, indexed:

More information

AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST

AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST OIL-LESS TYPE IMPORTANT: PLEASE READ CAREFULLY BEFORE STARTING OPERATIONS. THE CONTENTS ARE FOR GENERAL INFORMATION OF ALL THE SIMILAR MODELS. Record

More information

English USER GUIDE User guide

English USER GUIDE User guide USER GUIDE English ENGLISH A Elbow E Cushion B Elbow ring 1 Vent (in cushion and elbow) C Headgear 2 Side buttons D Frame 3 Sleeve Intended use The AirFit N30i mask is intended to be used by patients weighing

More information

Instructions for use Evacuation system compressed air injector SL 20 / SL 20 Max SL 30 / SL 30 Max

Instructions for use Evacuation system compressed air injector SL 20 / SL 20 Max SL 30 / SL 30 Max Instructions for use Evacuation system compressed air injector SL 20 / SL 20 Max SL 30 / SL 30 Max Table of Contents Introduction 1. Description of Functions 2. Putting into Operation 3. Instructions for

More information

AIR INLINE METAL SHEAR

AIR INLINE METAL SHEAR AIR INLINE METAL SHEAR ASSEMBLY and OPERATING INSTRUCTIONS 3491 Mission Oaks Blvd. / Camarillo, CA 93011 Copyright 1997 by Harbor Freight Tools. All rights reserved. No portion of this manual or any artwork

More information

Domiciliary Oxygen Switch. Instructions for Use

Domiciliary Oxygen Switch. Instructions for Use Domiciliary Oxygen Switch Instructions for Use 702-0041.11 July 2017 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the

More information

For use with select HPC Fire Pits ONLY- refer to product catalog or website. 100K BTU Maximum WARNING: FOR OUTDOOR USE ONLY

For use with select HPC Fire Pits ONLY- refer to product catalog or website. 100K BTU Maximum WARNING: FOR OUTDOOR USE ONLY For use with select HPC Fire Pits ONLY- refer to product catalog or website. 100K BTU Maximum : FOR OUTDOOR USE ONLY Small Tank LP Kit (STLPK) Installation & Operation Instructions Installation We suggest

More information

444C DUAL PERFORMANCE VALUE PACK

444C DUAL PERFORMANCE VALUE PACK (Chrome) PART NO. 44432 IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of this manual before installing and using this product. SAVE THIS MANUAL

More information

97C COMPRESSOR KIT 12V PART NO C COMPRESSOR KIT 24V PART NO C COMPRESSOR KIT PART NO

97C COMPRESSOR KIT 12V PART NO C COMPRESSOR KIT 24V PART NO C COMPRESSOR KIT PART NO 97C COMPRESSOR KIT 12V PART NO. 00097 97C COMPRESSOR KIT 24V PART NO. 02497 98C COMPRESSOR KIT PART NO. 00098 97C 98C IMPORTANT: It is essential that you and any other operator of this product read and

More information

Mounting and Operating Instructions EB 2558 EN. Self-operated Pressure Regulators. Type Pressure Build-up Regulator

Mounting and Operating Instructions EB 2558 EN. Self-operated Pressure Regulators. Type Pressure Build-up Regulator Self-operated Pressure Regulators Type 2357-31 Pressure Build-up Regulator with safety function and integrated excess pressure valve Type 2357-31 with non-return unit at port C Ports A and B with soldering

More information

CJMJ.DPNQBDU. Instruction for use. Neonatal phototherapy device. BILI-COMPACT WY 1816/ Year of manufacture D2017_00

CJMJ.DPNQBDU. Instruction for use. Neonatal phototherapy device. BILI-COMPACT WY 1816/ Year of manufacture D2017_00 Instruction for use Neonatal phototherapy device CJMJ.DPNQBDU D2017_00 GmbH > BILI-COMPACT WY 1816/ Year of manufacture Herrenhöhe 4 D-51515 Kürten/Germany 02207-9698-0 02207-4750 E-mail: info@weyermed.com

More information

MEDUMAT Standard a MEDUMAT Standard. The Ventilation Standard for Professionals

MEDUMAT Standard a MEDUMAT Standard. The Ventilation Standard for Professionals a The Ventilation Standard for Professionals a The Ventilation Standard for Professionals a offers you everything you need for reliable emergency ventilation. Users around the world are impressed by its

More information

CPR MEDICAL DEVICES INC. OPERATING MANUAL

CPR MEDICAL DEVICES INC. OPERATING MANUAL CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR OXYLATOR EMX RESUSCITATION AND INHALATION SYSTEM Symbols used in the Operating Manual for the OXYLATOR EMX and/or on the OXYLATOR EMX unit: WARNING! Do not

More information

Yoke Block Instruction Manual

Yoke Block Instruction Manual Yoke Block Instruction Manual ! WARNING IMPORTANT: READ MANUAL COMPLETELY BEFORE OPERATING THIS DEVICE This manual contains instructions on periodically required checks to be performed by the user. These

More information

Operating Manual Body fat balance

Operating Manual Body fat balance KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen email: info@kern-sohn.com Phone: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com Operating Manual Body fat balance KERN MFB Version

More information

FLOWMETER MODELS: 1MFA, 4MFA, 6MFA and 8MFA Series SAVE THESE INSTRUCTIONS

FLOWMETER MODELS: 1MFA, 4MFA, 6MFA and 8MFA Series SAVE THESE INSTRUCTIONS USER MANUAL FLOWMETER MODELS: 1MFA, 4MFA, 6MFA and 8MFA Series 1MFA3001 (Shown) 8MFA1001 (Shown) SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

More information

200 PSI FAST-FILL AIR SOURCE KIT

200 PSI FAST-FILL AIR SOURCE KIT 200 PSI FAST-FILL AIR SOURCE KIT 55% Duty Compressor on 2.0 Gallon Air Tank PART NO. 20007 IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of

More information

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE INSTALLATION, OPERATION & MAINTENANCE MANUAL INTERFACE DEVICES, INC. 230 Depot Road, Milford, CT 06460 Ph: (203) 878-4648, Fx: (203) 882-0885, E-mail: info@interfacedevices.com

More information

PPS PORTABLE POWER SUPPLY

PPS PORTABLE POWER SUPPLY PPS User Manual Rev 1-2015 ENG MASTER PPS PORTABLE POWER SUPPLY Disclaimer About ActSafe Safety messages and warnings Product safety System description Pre-use inspection and connection General safety

More information

Instruction Manual LIMITED 1 YEAR WARRANTY. Hydraulic Punch Driver Read this material before using this product.

Instruction Manual LIMITED 1 YEAR WARRANTY. Hydraulic Punch Driver Read this material before using this product. Instruction Manual Hydraulic Punch Driver 902-483 LIMITED 1 YEAR WARRANTY We make every effort to assure that its products meet high quality and durability standards, and warrant to the original purchaser

More information

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% ADMINISTRATION SYSTEM (01EQ7000E) DEMAND VALVE (01EQ7001E) O-TWO MEDICAL TECHNOLOGIES INC.

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% ADMINISTRATION SYSTEM (01EQ7000E) DEMAND VALVE (01EQ7001E) O-TWO MEDICAL TECHNOLOGIES INC. MANUFACTURERS AND DISTRIBUTORS OF: O-Two eseries Electronic Transport Ventilators O-Two Single Use Open CPAP Systems O-Two CAREvent Range of Automatic Transport Ventilators O-Two CAREvent Range of Automatic

More information

Type 3709 Pneumatic Lock-up Valve. Translation of original instructions. Mounting and Operating Instructions EB 8391 EN

Type 3709 Pneumatic Lock-up Valve. Translation of original instructions. Mounting and Operating Instructions EB 8391 EN Type 3709 Pneumatic Lock-up Valve Translation of original instructions Mounting and Operating Instructions EB 8391 EN Edition July 2017 Note on these mounting and operating instructions These mounting

More information

Oxygen Demand Valve. Instructions for Use

Oxygen Demand Valve. Instructions for Use Oxygen Demand Valve Instructions for Use 702-0059.11 May 2017 Contents 1. Symbols... 2 2. Warnings and Cautions... 2 2.1. Warnings!... 2 2.2. Cautions!... 3 2.3. Notes... 4 3. Functional Description...

More information

VS18/26 with PROFINET and EtherNet/IP Interface. ATEX Installation Instructions

VS18/26 with PROFINET and EtherNet/IP Interface. ATEX Installation Instructions VS18/26 with PROFINET and EtherNet/IP Interface ATEX Installation Instructions INDEX 1. INTENDED USAGE 3 2. OPERATING MANUAL ATEX 4 2.1 General conditions 4 2.2 Installation 5 2.3 Operating 5 2.4 Failures

More information

User Manual PTW-DensiX. D / Di/Zi

User Manual PTW-DensiX. D / Di/Zi User Manual PTW-DensiX D148.131.0/3 2005-05 Di/Zi General Information General Information The product bears the CE-mark "CE-0124" in accordance with the Council Directive 93/42/EEC about Medical Devices

More information

Nitrogen Supply System

Nitrogen Supply System ATMOSCOPE Nitrogen Supply System NS500 Users Guide Copyright 2001 EDSYN, INC. ATMOSCOPE Nitrogen Supply Systems TABLE OF CONTENTS Introduction............................. 3 General Information.......................

More information

COMBIBAG Resuscitator. Operating Instructions

COMBIBAG Resuscitator. Operating Instructions COMBIBAG Resuscitator Operating Instructions Overview COMBIBAG Patient valve Safety valve Squeeze bag (compressable unit) Intake valve Accessories PEEP valve Oxygen reservoir DEMAND valve 2 EN Overview

More information

Welcome to the Specialized Medical Services Respiratory training webinar series!

Welcome to the Specialized Medical Services Respiratory training webinar series! Welcome to the Specialized Medical Services Respiratory training webinar series! SMS is your Long Term Care (LTC) facility single source for oxygen, medical equipment, respiratory care services and supplies

More information

English. ivolve Full Face Mask. User Manual

English. ivolve Full Face Mask. User Manual ivolve Full Face Mask User Manual 0123 Table of Contents ivolve Full Face Mask... 1 Intended Use... 1 Before Using the Mask... 2 Parts of the ivolve Full Face Mask... 4 Getting the Right Cushion Size...

More information

80 Litre Suction Oil Drainer

80 Litre Suction Oil Drainer 80 Litre Suction Oil Drainer Please dispose of packaging for the product in a responsible manner. It is suitable for recycling. Help to protect the environment, take the packaging to the local amenity

More information

200 PSI COMPRESSORS - MODEL NUMBERS

200 PSI COMPRESSORS - MODEL NUMBERS 200 PSI COMPRESSORS - MODEL NUMBERS 380C AIR COMPRESSOR KIT PART NO. 38033 480C AIR COMPRESSOR KIT PART NO. 48043 380C 480C IMPORTANT: It is essential that you and any other operator of this product read

More information

Operating Instruction English. Revision A Date author NH This operating instruction is not subject to updating. Airgun

Operating Instruction English. Revision A Date author NH This operating instruction is not subject to updating. Airgun Revision A Date 23.11.2012 author NH This operating instruction is not subject to updating Airgun 95400 1 This airgun is a quality product with special focus on high functionality, easy handling, safety

More information

PRESSURE TEST PUMP OPERATING & MAINTENANCE INSTRUCTIONS. Model No: PTP100. Part No: GC01/09

PRESSURE TEST PUMP OPERATING & MAINTENANCE INSTRUCTIONS. Model No: PTP100. Part No: GC01/09 PRESSURE TEST PUMP Model No: PTP00 Part No: 06020 OPERATING & MAINTENANCE INSTRUCTIONS GC0/09 INTRODUCTION Thank you for purchasing this CLARKE Pressure Testing Pump. Before attempting to use the product,

More information

APC Pressure Reducer with Sensor

APC Pressure Reducer with Sensor OBJ_DOKU-194490-001.fm Seite 1 Donnerstag, 16. Februar 2017 4:22 16 NOTES ON USE USA APC Pressure Reducer with Sensor 20134-006 OBJ_DOKU-194490-001.fm Seite 2 Donnerstag, 16. Februar 2017 4:22 16 Table

More information

Interface Devices, Inc. Hydraulic Mini Mule

Interface Devices, Inc. Hydraulic Mini Mule Interface Devices, Inc. Hydraulic Mini Mule INSTALLATION, OPERATION & MAINTENANCE MANUAL IMPORTANT! FILE THIS MANUAL IN A SAFE PLACE FOR FUTURE SERVICE & PARTS NEEDS ALWAYS REFERENCE THE SERIAL NUMBER

More information