Fluidized Bed Unit (FBU) Operating Manual. Fall 2008
|
|
- Rhoda Sullivan
- 6 years ago
- Views:
Transcription
1 Fluidized Bed Unit (FBU) Operating Manual Fall 2008
2 Background Fluidized bed technology is based on the fluid like behavior of a bed of solid particles when subjected to the buoyant forces exerted by a gas or liquid. Though composed of an inhomogeneous mixture of fluid and solids, the bed fluidized by the gas or liquid behaves like a fluid (i.e., exhibiting hydrostatic surface properties and an effective bulk density lower than the original solids). The outcrop of this behavior is a host of attractive benefits including enhancement of heat and mass transport, high contacting efficiency for reactants, and improved flow and transport options for the solids. Figure 1: Bed of glass beads fluidized by air Understanding the behavior of fluidized beds is important to realizing these and other benefits. This behavior can be seen in the short video captured in Figure 1, where a bed of small glass beads moves from the slumped condition through bubbling fluidization and back again as the air flow rate is first increased then decreased. Description of Facilities The FBU is formally a part of the Permeameter Unit skid (see Figure 2) but is separable from that unit from both operations and data collection points of view. The unit is controlled and operated entirely in manual fashion using valves, a regulator, and rotameters with the exception of observation and data collection which rely on the Honeywell Experion system. Figure 2: Schematic diagram of the FBU portion of the PERM unit skid
3 Using the Honeywell Experion Process Knowledge System (PKS) Honeywell s Experion PKS is a distributed control system (DCS) widely used in industry and on many lab experiments here at LSU. In Experion, each process variable is represented by an entity called a Control Module (CM). Each CM is a collection of Function Blocks (FB) and each FB consists of many values called parameters. Within a CM (and sometimes between CMs), the FBs are wired together in various ways to monitor and control the process. Desired values of many parameters may be entered via the computer keyboard. The purpose of the next few sections is to explain how to use Experion to observe and collect data from the Fluidized Bed Unit (FBU). Logon to a Honeywell Experion Equipped PC On a PC loaded with Honeywell Station software, logon using your normal ID, password and domain. Open the Honeywell Station software by going to Start/Programs/Honeywell Experion PKS/Client Software/Station. If your ID is authorized for use of the Honeywell PKS system, the Station program full menu will appear. From the Unit item on the menu, select PERM. The Permeameter/FBU P&ID schematic will appear. You will be using this schematic to follow the operation of the FBU. This schematic has NOT been updated to reflect recent modifications to the unit on the FBU portion of the skid. Rely on the included drawing in this manual (Figure 2) to provide you some of the missing details. Controlling from the PID Schematic (At present, there are no controllers on the PERM schematic for use in controlling the FBU. The FBU is controlled using manual devices at the unit skid itself.) Every controller is represented by a small colored circle containing the tagname of the CM, with the values of the setpoint (backlit in green) and the process value (backlit in cyan) near the circle. Any standalone transmitter (those without controllers) is represented by a small white circle containing the tagname of the transmitter, with the process value (backlit in cyan) near the circle. The first letter indicates the type of measurement F for flow rate, P for pressure, or T for temperature. To change any analog value from a schematic, you must click on it (if it can be changed, its backlighting will change), and then enter the new desired value. The changeable objects in this display are the values near each controller for the Permeameter, and there are additional objects on the faceplates, all of which are explained below. The main value associated with a controller or transmitter is the measured input, or process value (PV). In addition to the PV, controllers have several additional values, the most important of which is the setpoint, or SP. This is just like the speed setting on a cruise control the controller will manipulate its output (the throttle position in this case) to move the PV to the SP and hold it there. The SP (in green) and the PV (in cyan) are shown immediately to the side of the circle representing the controller. Clicking on any of the small colored circles the controllers will bring up the respective controller s faceplate, on which you can see the tagname, description, engineering units, and several of the most important parameters on the controller. You can also change many of these parameters. Near the bottom of the faceplate is a combo box labeled MD, which can be used to select the mode of the controller. Immediately above the mode are the OP (changeable), the PV (not changeable) and the SP
4 (changeable). To change the OP or SP, single click the value, type in the new value and press ENTER. The SPs and tuning constants on the schematic can be changed the same way. There are also special buttons on the schematic to accomplish various functions for the Permeameter unit. These are not needed to run the FBU. Controller Modes (At present, there are no controllers on the PERM schematic for use in controlling the FBU. The FBU is controlled using manual devices at the unit skid itself.) As mentioned above, a controller has an SP and an OP. The OP is always given in percent (0 100%) and the SP has the same engineering units as the PV (in this experiment, the flow rates are in ml/min or SCFM, the temperatures are in C, and pressures are in inches of water column). When the controller mode is MANual, the OP is held until the operator changes it. When you want to change it, simply click the OP in the faceplate, type in the new value, and press ENTER. The new OP will be held until you change it again. Note that you may enter an OP only while a controller is in MAN. An OP value of 6.9% is known as tight shutoff in the Experion system. Notice the small bar under the control valve on the schematic its length is proportional to the output. When the mode is not MAN, the controller uses the PV, SP and tuning constants to calculate the OP. When the mode is AUTOmatic, you may enter a new SP to be used for control. Note that changing an SP affects the OP only while a controller is in AUTO. The circle representing a controller is filled with a color which indicates the current mode of the controller. Yellow means the mode is MAN, and white means the mode is AUTO. Sampled Data Control Unlike dedicated analog instrumentation for process control, the Honeywell Experion uses microprocessor based digital computers to perform measurement and regulatory control tasks. It uses workstation and server PCs to configure the system, build control strategies and schematics, gather and display data, etc. The calculations in such systems are typically performed at regular intervals of time. The time between two consecutive data readings is referred to as the sampling time and such systems are referred to as sampled data systems. The sampling time used in Experion is one (1) second. All data are collected each second, control algorithms processed, and all outputs resent each second. Display Navigation When you first opened the Station software, you used an item from the menu bar to call up the main schematic. There are several additional ways to go from one display to another. For example, you can enter the tagname of a controller in the Command field at the top of the screen and press F12 to call up the detail display. For a controller, the detail display has 7 tabs. The one labeled LOOP TUNE is useful for tuning controllers.
5 Most of the toolbar buttons are used for navigation some require a name or number to be entered, and some go directly to the display. Most of the same functions are on the function keys. For example, to return to the previous display, click or press F8. To return to the display before that, do it again. From most displays (both system displays and custom schematics such as PERM), double clicking any value associated with a CM will take you to its detail display. From a detail display, click or press F2 to return to the main PERM schematic. On most custom schematics there may also be buttons to quickly get you from one display to another. Using Trends There is one button labeled Trend 50 on the main PERM schematic that is used to call up a trend. This trend button displays the PVs of all the controllers and transmitters on the Permeameter/FBU. The only variable on this display of real interest to FBU users is P501, the Fluidized Bed Pressure Drop. At the bottom of the trend is the legend with all the tag.block.parameters associated with the traces. The checkboxes in the Pen column indicate which traces are currently on the trend. Click on the chart area of the trend and a white hairline cursor appears on the chart and the values at the hairline cursor appear in the Reference Value column of the legend. Along the bottom of the chart area is a horizontal scroll bar which allows you to scroll the chart area back and forth. Along the left axis are listed the low and high range of the selected trace. These allow you to change the range of the trace for the selected parameter. Practice by changing the range of the P501 to 0 to 20. Immediately above the left side of the chart area is a combo box which allows you to select one of the traces (you may also click anywhere on the line for this trace in the legend area). When you select an active trace, it is highlighted (thicker) in the chart area. Above the right side of the chart area is the Period combo box which allows you to select how much data, on a time wise basis, is displayed in the chart area. To the right of that is the Interval combo box which allows you to select the interval between points in the chart area. Practice changing to a different period and interval. Leave the period set to 1 day and the interval at 1 minute for now. For practice, scroll back until some variation in some of the traces appears. Notice that the timestamps below the chart area change as you scroll. Find some local max or min in one of the traces and click or drag the hairline to it. Now change the period back to one hour and notice that the cursor is centered on (or at least near) the local maximum or minimum. If necessary, move the hairline so it is exactly on the peak or valley and notice that the values, as well as the date and time, are shown in the Reference Value column in the legend. Now return the trend to the current time by clicking. All changes you make to the trend can be saved by clicking the familiar Windows Save icon just above the right end of the chart area next to the word (Modified). Saving Process Variable Values into Excel Normal Speed Data Collection Successful completion of most experiments on the FBU will require the analysis of a good deal of data. To collect these data, an Excel workbook containing a Visual Basic Add in is provided.
6 Open Excel 2003 and navigate to: C:\Program Files\Honeywell\Experion PKS\Client\PermRecorder.xls This file is used for recording normal speed (i.e., no faster than 1Hz) process data from the Perm and the included FBU subsystem while the experiment is running. Double click it and enable macros. The workbook will open with a Start button, the experiment name, a collection frequency Combo Box, and a Stop button on the top line. Click on the Start button, and the workbook will start collecting the relevant data at the specified collection frequency. While the workbook is collecting data, it may be scrolled, but you should not attempt to do anything else in this instance of Excel until after you click on the Stop button. If you do, the collector may stop and you may lose valuable data. When you finish a run, click on the Stop button and cut or copy whatever data you need into your daily workbook in a separate instance of Excel. Fast Data Collection Some experiments using the FBU will require the collection of data at faster than normal speeds. To collect these data, an additional Excel workbook containing a Visual Basic Add in is provided. Open Excel 2003 and navigate to: C:\Program Files\Honeywell\Experion PKS\Client\FluidBedRecorder.xls This file is used for recording fast data i.e., 20 Hz or 50 ms data from the FBU differential pressure transmitter while the experiment is running. Double click it and enable macros. The workbook will open with a Start button, the experiment name, a non functional collection frequency Combo Box, and a Stop button on the top line. Click on the Start button, and the workbook will start collecting the relevant data at a fixed frequency of 20 Hz. While the workbook is collecting data, it may be scrolled, but you should not attempt to do anything else in this instance of Excel until after you click on the Stop button. If you do, the collector may stop and you may lose valuable data. When you finish a run, click on the Stop button and cut or copy whatever data you need into your daily workbook in a separate instance of Excel. Startup Procedure To establish a desired flow of air through the FBU, execute the following procedure: 1. Ensure that the air regulator is closed (off) by turning the round black handle knob counterclockwise until the regulator gage reads 0 psig. 2. Ensure that both rotameter variable flow valves are closed. Doing so ensures that there is no air flow UNTIL desired by operations personnel. Doing so guards against having both rotameters open simultaneously and thereby getting a misleading total air flow rate through the FBU.
7 3. Ensure that the block valve just before the dry test meter is closed. The dry test meter in not in service at this time. Additionally, the maximum air flow deliverable by the large rotameter exceeds the capacity of the dry test meter. Exceeding the capacity of this meter will result in erroneous readings and can seriously damage the meter. Opening these block valves (BV12, BV13, BV14, and BV15) lines up potential air flow to the FBU. Doing so establishes an initial pressure for air flow. 4. Open block valves BV12, BV13, BV14, and BV Adjust the air regulator by turning the large black handle knob clockwise until the gage reads 5 psig. 6. Adjust the desired rotameter variable flow Doing so establishes an initial air flow value. valve until the desired reading is obtained. 7. Readjust the air regulator to 5 psig. You will notice that the regulator pressure had changed after performing Step Repeat Steps 6 and 7 until both regulator pressure and air flow reading are at the desired values. Changing the setting on either valve changes the reading at the other device, requiring that these two steps are repeated until both objectives are achieved. Shutdown Procedure To place the FBU is a safe shutdown mode, execute the following procedure: 1. Ensure that the air regulator is closed (off) by turning the round black handle knob counterclockwise until the regulator gage reads 0 psig. 2. Ensure that both rotameter variable flow valves are closed. 3. Ensure that the block valve just before the dry test meter is closed. 4. Close block valves BV12, BV13, BV14, and BV15. Doing so ensures that there is no air flow to the FBU. Doing so guards against having both rotameters open simultaneously and thereby getting a misleading total air flow rate through the FBU. The dry test meter in not in service at this time. Additionally, the maximum air flow deliverable by the large rotameter exceeds the capacity of the dry test meter. Exceeding the capacity of this meter will result in erroneous readings and can seriously damage the meter. Closing these block valves (BV12, BV13, BV14, and BV15) isolates FBU from potential air flow. Special Procedures Calibrating Rotameters There are two rotameters available to meter air into the FBU a larger one and a smaller one. If there is any uncertainty regarding the rated performance of either of these two instruments, calibration (or at least a check of the calibration) might be warranted. The accuracy of an off the shelf rotameter is typically ± 5% of the full scale reading. Improving on that accuracy would require calibration using
8 system components with even better accuracy. An independently calibrated dry test meter, for example, with ± 1% accuracy would be a good candidate for rotameter calibration. However, without a recent certified calibration certificate for the dry test meter itself or at least some knowledge that the dry test meter is measuring properly it s unclear that this method offers value. Another possibility for performing at least a single point flow check for a rotameter would be to pass a known volume of gas through the rotameter at a constant rate. A standard cylinder (i.e., known total volume) of compressed air can be used to provide such a flow and the change in pressure over a fixed time period can be used to estimate the volumetric flow rate that passes through the rotameter. Comparing this to that indicated by the rotameter would provide a single calibration check point. The FBU can be equipped to perform this test if necessary. Failing that, the rotameter faceplate calibrations are given on the meters themselves and can be found in this manual in the Instrumentation Specs section. Determining Distributor Pressure Drop The fritted metal plate used to distribute air uniformly to the FBU bed adds pressure drop to the observed total differential pressure across the unit. Increasing the air flow through the fritted metal plate increases this added pressure drop, which then becomes an increasingly larger percentage of the total observed differential pressure. If one needs to determine the fluidized bed differential pressure ex the added pressure drop due to the distributor, a pressure drop versus air flow measurement study should be conducted prior to introducing solids to determine needed correction(s). Determining Bed Void Fraction at Minimum Fluidization εmf Determining bed void fraction at minimum fluidization velocity requires some involved testing, measurement and calculations. A thorough discussion of this is best left to a literature reference (Subramani, Balaiyya and Miranda 2007). Laboratory Procedures Determining Average Particle Diameter using Sieves Fluidization behavior and virtually every correlation attempting to explain that behavior is greatly dependent on a representative average particle size of the fluidized solids. This average is determined in some manner from the distribution of particle sizes present in the admixture of solids. There are a variety of methods for determining particle size distribution but the most common method is a sieve analysis in which a known quantity of solids is shaken through a series of sieves of ever decreasing aperture. The distribution of solids retained on each screen is used to estimate an average particle size. A representative diameter can be determined as follows (Baker and Herrman 2002): d gw = log n 1 i= 1 ( W log d ) n i i= 1 W i i (1.1)
9 Where: dgw is the geometric mean, mass weighted diameter, W i d i is the mass of particles retained on the is the diameter of the th i th i sieve, and sieve in the stack, calculated from d = ( d d ) i u o 0.5 (1.2) d u th Where: = diameter opening through which particles will pass (sieve preceding i ), and do = diameter opening through which particles will not pass (i sieve). Determining Average Particle Sphericity Depending on the particular solids used in FBU, the particles may be shaped more like spheres or they may be quite angular. If the particles are large enough, a micrometer can be used to make multiple length measurements at different particle orientations. If the particles are too small for measurement by micrometer, a microscope and quantitative reticule can be used instead. Specifically, if one defines the smallest dimension of a particle to be r and the remaining two dimensions to be s and t respectively, then the sphericity of the particle φ s can be calculated as follows (Siwiec 2007): th φ = s r st (1.3) The average of a sufficient number of replicate measurements can produce a reasonable value which should compare favorably to that asserted by the manufacturer of the material. Determining Bed Void Fraction ε The bed void fractionε absent fluidization also described as the bed porosity is basically a property of the bulk solids and can be estimated ex situ by volume and mass measurements and a simple implicit calculation (Chase n.d.): ρ = ε ρ+ (1 ε) ρ (1.4) o Where: ρo is the bed bulk density as determined by volume and mass measurements, ρ is the fluid density (normally air, unless bulk density is determined in another other), ρ p is the solids particle density. p
10 Instrumentation Specs Rotameters Device Flow Specific Gravity Temperature Pressure Large rotameter SCFM = Scale x F 0 psig Small rotameter Scale = L/min C 760 mm Hg abs. Safety Considerations See fluidized bed solids Materials Safety Data Sheet for hazard information. Works Cited Baker, Scott, and Tim Herrman. "Evaluating Particle Size." Kansas State University (accessed October 22, 2008). Chase, George G. "Solids Notes 4." University of Akron. (accessed October 22, 2008). Siwiec, Tadeusz. "The Sphericity of Grains of Filtration Beds Applied for Water Treatment on Examples of Selected Minerals." Electronic Journal of Polish Agricultural Universities 10, no. 1 (2007). Subramani, Hariprasad J., M.B. Mothivel Balaiyya, and Lima Rose Miranda. "Minimum fluidization velocity at elevated temperatures for Geldart's group B powders." Experimental Thermal and Fluid Science 32 (2007):
Vertical Tube Heat Exchanger (VTHX) or The Heat Exchanger Network (HEN) Control Process. Operating Manual
Vertical Tube Heat Exchanger (VTHX) or The Heat Exchanger Network (HEN) Control Process Operating Manual 1/17 9/4/2013 1. Background Heat exchangers are ubiquitous in the industrial world with millions
More informationCHEMICAL ENGINEERING LABORATORY CHEG 239W. Control of a Steam-Heated Mixing Tank with a Pneumatic Process Controller
CHEMICAL ENGINEERING LABORATORY CHEG 239W Control of a Steam-Heated Mixing Tank with a Pneumatic Process Controller Objective The experiment involves tuning a commercial process controller for temperature
More informationPegas 4000 MF Gas Mixer InstructionManual Columbus Instruments
Pegas 4000 MF Gas Mixer InstructionManual Contents I Table of Contents Foreword Part I Introduction 1 2 1 System overview... 2 2 Specifications... 3 Part II Installation 4 1 Rear panel connections...
More informationCover Page for Lab Report Group Portion. Head Losses in Pipes
Cover Page for Lab Report Group Portion Head Losses in Pipes Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 February 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section
More informationExercise 2-2. Second-Order Interacting Processes EXERCISE OBJECTIVE DISCUSSION OUTLINE. The actual setup DISCUSSION
Exercise 2-2 Second-Order Interacting Processes EXERCISE OBJECTIVE Familiarize yourself with second-order interacting processes and experiment with the finer points of controller tuning to gain a deeper
More informationCover Page for Lab Report Group Portion. Drag on Spheres
Cover Page for Lab Report Group Portion Drag on Spheres Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 29 September 2017 Name 1: Name 2: Name 3: [Name 4: ] Date: Section number:
More informationAdsorption Unit Operating Procedures: CO2 Adsorption on to Activated Carbon
Adsorption Unit Operating Procedures: CO2 Adsorption on to Activated Carbon Background Safety Concerns Description of the Unit Switching the Unit On Preparation of the Unit Configuration of the Unit for
More information5890II GC Standard Operating Procedure 9/2/2005
5890II GC Standard Operating Procedure 9/2/2005 This procedure is for the analysis of natural gas using the HP 5890II GC with the Restek ShinCarbon column. Detector Type: TCD Column Type: Restek Micropacked,
More informationUNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008
UNITY 2 TM Air Server Series 2 Operators Manual Version 1.0 February 2008 1. Introduction to the Air Server Accessory for UNITY 2...2 1.1. Summary of Operation...2 2. Developing a UNITY 2-Air Server method
More informationLOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary
ADH 1/7/014 LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary This experiment will introduce you to the kinetic properties of low-pressure gases. You will make observations on the
More informationArmfield Distillation Column Operation Guidelines
Armfield Distillation Column Operation Guidelines 11-2016 R.Cox Safety SAFETY GLASSES ARE REQUIRED WHEN OPERATING THE DISTILLATION COLUMN Wear gloves when mixing alcohol feedstock The column will become
More informationLOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12
LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/ This experiment will introduce you to the kinetic properties of low-pressure gases. You will make observations on the rates with which selected
More information( ) ( ) *( A ) APPLICATION DATA. Procidia Control Solutions Coarse/Fine Control. Split-Range Control. AD Rev 2 April 2012
APPLICATION DATA Procidia Control Solutions / Control AD353-118 Rev 2 April 2012 This application data sheet describes implementing a coarse/fine control strategy in a Siemens Procidia 353 controller.
More informationExercise 8. Closed-Loop Pressure Control, Proportional-Plus-Integral Mode EXERCISE OBJECTIVE
Exercise 8 Closed-Loop Pressure Control, EXERCISE OBJECTIVE To understand open and closed-loop pressure control; To learn how to sense the pressure in a pneumatic circuit; To control the pressure in a
More informationLABORATORY EXERCISE 1 CONTROL VALVE CHARACTERISTICS
Date: Name: LABORATORY EXERCISE 1 CONTROL VALVE CHARACTERISTICS OBJECTIVE: To demonstrate the relation between valve stem position and the fluid flow through a control valve, for both linear and equal
More informationL 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions
L 100 Bubble-Tube Level System Installation, Operation and Maintenance Instructions Figure 1 Contents Section Description Page 1.0 Introduction 2 2.0 Specifications 3 3.0 Installation 3 4.0 Warranty 6
More informationRM-80 respiration monitor
RM-80 respiration monitor User Manual September 18, 2015 0025-003M 950 North Hague Avenue Columbus, Ohio 43204-2121 USA Sales: sales@colinst.com Service: service@colinst.com Phone: (614) 276-0861 Fax:
More informationThe Discussion of this exercise covers the following points:
Exercise 3-2 Orifice Plates EXERCISE OBJECTIVE In this exercise, you will study how differential pressure flowmeters operate. You will describe the relationship between the flow rate and the pressure drop
More informationExperiment 11: The Ideal Gas Law
Experiment 11: The Ideal Gas Law The behavior of an ideal gas is described by its equation of state, PV = nrt. You will look at two special cases of this. Part 1: Determination of Absolute Zero. You will
More informationCover Page for Lab Report Group Portion. Pump Performance
Cover Page for Lab Report Group Portion Pump Performance Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 March 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section number:
More informationACV-10 Automatic Control Valve
ACV-10 Automatic Control Valve Installation, Operation & Maintenance General: The Archer Instruments ACV-10 is a precision automatic feed rate control valve for use in vacuum systems feeding Chlorine,
More informationPressure Control. where: p is the pressure F is the normal component of the force A is the area
Pressure Control First of all, what is pressure, the property we want to control? From Wikipedia, the free encyclopedia. Pressure is the application of force to a surface, and the concentration of that
More informationFireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS
FireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS WARNING THE WARRANTIES MADE BY MSA WITH RESPECT TO THE PRODUCT ARE VOIDED IF THE PRODUCT IS NOT USED AND MAINTAINED IN ACCORDANCE
More informationHPICAL Operation & Data Logging Procedures. Click spacebar to advance through slides 1
HPICAL-15000 Operation & Data Logging Procedures Click spacebar to advance through slides 1 WARNING Always wear proper safety equipment when using high pressure equipment. Do not exceed 125 psi air pressure.
More informationThe HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator
The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator Designed, built, and supported by InstruQuest Inc. Universal pycnometer, no
More informationLab 4: Root Locus Based Control Design
Lab 4: Root Locus Based Control Design References: Franklin, Powell and Emami-Naeini. Feedback Control of Dynamic Systems, 3 rd ed. Addison-Wesley, Massachusetts: 1994. Ogata, Katsuhiko. Modern Control
More informationDrilling Efficiency Utilizing Coriolis Flow Technology
Session 12: Drilling Efficiency Utilizing Coriolis Flow Technology Clement Cabanayan Emerson Process Management Abstract Continuous, accurate and reliable measurement of drilling fluid volumes and densities
More informationCover Page for Lab Report Group Portion. Lift on a Wing
Cover Page for Lab Report Group Portion Lift on a Wing Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 17 January 2017 Name 1: Name 2: Name 3: [Name 4: ] Date: Section number:
More informationInprocess Operator Training Programme
2016 Inprocess Operator Training Programme ABSORPTION COLUMN These exercises are intended to provide an understanding of absorption columns and the fundamental principles they use to eliminate pollutants
More informationMoLE Gas Laws Activities
MoLE Gas Laws Activities To begin this assignment you must be able to log on to the Internet using Internet Explorer (Microsoft) 4.5 or higher. If you do not have the current version of the browser, go
More informationVISIMIX LAMINAR. BLENDING OF NON-NEWTONIAN LIQUID. POWER. CIRCULATION. MIXING TIME.
VISIMIX LAMINAR. BLENDING OF NON-NEWTONIAN LIQUID. POWER. CIRCULATION. MIXING TIME. Subject of the example application of program VisiMix Laminar for analysis and improvement of process characteristics
More informationExercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler.
Exercise 5-2 Bubblers EXERCISE OBJECTIVE Learn to measure the level in a vessel using a bubbler. DISCUSSION OUTLINE The Discussion of this exercise covers the following points: Bubblers How to measure
More informationThe Discussion of this exercise covers the following points: Range with an elevated or suppressed zero Suppressed-zero range Elevated-zero range
Exercise 4-3 Zero Suppression and Zero Elevation EXERCISE OBJECTIVE In this exercise, you will learn the effect that mounting a pressure transmitter above or below the reference level has on the hydrostatic
More informationWorkshop 1: Bubbly Flow in a Rectangular Bubble Column. Multiphase Flow Modeling In ANSYS CFX Release ANSYS, Inc. WS1-1 Release 14.
Workshop 1: Bubbly Flow in a Rectangular Bubble Column 14. 5 Release Multiphase Flow Modeling In ANSYS CFX 2013 ANSYS, Inc. WS1-1 Release 14.5 Introduction This workshop models the dispersion of air bubbles
More informationDynamic Simulation for T-9 Storage Tank (Holding Case)
Dynamic Simulation for T-9 Storage Tank (Holding Case) CASE 1: 19,642 Kg/Hr (Holding: 52 o C), No Liquid Draw Workshop Description Estimation of vapor flow rate coming out from the T-9 tank for holding
More informationThe HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator
The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator Designed, built, and supported by InstruQuest Inc. Temperature controlled, multi-technique volumetric
More informationDynamics and Control of Chemical Processes Solution to Lab #8 Shutdown of C4-C5 separation section
Prof. Davide Manca Politecnico di Milano Dynamics and Control of Chemical Processes Solution to Lab #8 Shutdown of C4-C5 separation section RUN A DISTILLATION COLUMN USING HYSYS/UNISIM IN DYNAMIC MODE
More informationControlling the prefeeder
Controlling the prefeeder A prefeeder is a modulating device of some kind, controlling the material flow into the belt feeder infeed. You need it for three reasons: 1. The material can not be sheared at
More informationInjection Controller Program User Manual (for FloBoss 107 and ROC800-Series)
Part D301757X012 June 2016 Injection Controller Program User Manual (for FloBoss 107 and ROC800-Series) Remote Automation Solutions Revision Tracking Sheet June 2016 This manual may be revised periodically
More informationModel 130M Pneumatic Controller
Instruction MI 017-450 May 1978 Model 130M Pneumatic Controller Installation and Operation Manual Control Unit Controller Model 130M Controller is a pneumatic, shelf-mounted instrument with a separate
More informationSHIMADZU LC-10/20 PUMP
SHIMADZU LC-10/20 PUMP Clarity Control Module ENG Code/Rev.: M091/70C Date: 24.10.2017 Phone: +420 251 013 400 DataApex Ltd. Fax: +420 251 013 401 Petrzilkova 2583/13 clarity@dataapex.com 158 00 Prague
More informationWave Load Pattern Definition
COMPUTERS AND STRUCTURES, INC., AUGUST 2010 AUTOMATIC WAVE LOADS TECHNICAL NOTE DEFINING WAVE LOADS This section describes how to define automatic wave loads. The automatic wave load is a special type
More informationThis portion of the piping tutorial covers control valve sizing, control valves, and the use of nodes.
Piping Tutorial A piping network represents the flow of fluids through several pieces of equipment. If sufficient variables (flow rate and pressure) are specified on the piping network, CHEMCAD calculates
More informationSSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA
Page 1 of 9 SSI Solaris 150 RTA The Solaris 150 RTA is a rapid thermal annealing system capable of handling sample sizes up to 100mm (4 diameter) or smaller. The system can anneal in N 2 and Forming gas
More informationCode Basic module and level control complete with optionals code
AUTOMATION AND PROCESS CONTROL MPCT - Modular Process Control Trainer Code 916960 Basic module and level control complete with optionals code 916961-916962-916963-916964-916965 1. General Didacta proposes
More informationExp. 5 Ideal gas law. Introduction
Exp. 5 Ideal gas law Introduction We think of a gas as a collection of tiny particles in random, thermal motion. When they collide with the sides of a container, they exert a force on the container walls.
More informationHydrostatics Physics Lab XI
Hydrostatics Physics Lab XI Objective Students will discover the basic principles of buoyancy in a fluid. Students will also quantitatively demonstrate the variance of pressure with immersion depth in
More informationQuantitative Analysis of Hydrocarbons by Gas Chromatography
Quantitative Analysis of Hydrocarbons by Gas Chromatography Introduction Gas-liquid chromatography (GLC) accomplishes a separation by partitioning solutes between a mobile gas phase and a stationary liquid
More informationTR Electronic Pressure Regulator. User s Manual
TR Electronic Pressure Regulator Page 2 of 13 Table of Contents Warnings, Cautions & Notices... 3 Factory Default Setting... 4 Quick Start Procedure... 5 Configuration Tab... 8 Setup Tab... 9 Internal
More informationLab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow
Georgia Institute of Technology School of Earth and Atmospheric Sciences EAS 4641, Spring 2008 Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow Purpose of Lab 3: 1) To gain a
More information800 Series Mass Flow Meters and Controllers. High Performance Mass Flow Meters and Controllers
800 Series Mass Flow Meters and Controllers High Performance Mass Flow Meters and Controllers WHY MASS FLOW? In most processes it is mass, not volume, which is the critical variable. Volumetric flow measurements
More informationEXPERIMENT XI. Careful!! Improper handling of the vacuum line may result in the release of SO 2 which is an irritating and suffocating gas.
Chem 366-3 Page XI - 1 EXPERIMENT XI INFRARED SPECTRUM OF SO2 (S&G, 5th ed. Expt 36, 6th ed. Expt. 35) 1. Pre-Lab preparation. The description of this experiment has disappeared from the more recent editions
More informationIdeal gas law. Introduction
Ideal gas law Introduction We think of a gas as a collection of tiny particles in random, thermal motion. When they collide with the sides of a container, they exert a force on the container walls. The
More informationP3000 Deadweight Tester Setup, Part 1: Limited Partial Correction Method
P3000 Deadweight Tester Setup, Part 1: Limited Partial Correction Method 19 May 2014 M. Daniels This tutorial is for configuring a P3000 series (Pressurements) deadweight tester to be used with COMPASS
More informationLong term stability tests of INO RPC prototypes
Long term stability tests of INO RPC prototypes While the problem of sudden aging in the glass RPC prototypes is being investigated, a few RPCs of dimension 40 cm 30 cm were fabricated, using glass procured
More informationAgilent ICP-MS Interactive Troubleshooting tool for Low sensitivity
Agilent ICP-MS Interactive Troubleshooting tool for Low sensitivity Rev. B 2017,Janyary,20 Next > Preconditions 7900, 7800, 7700, or 8800 ICP-MS is used. Sample introduction type is PeriPump. If you are
More informationONSITE PROVING OF GAS METERS. Daniel J. Rudroff WFMS Inc West Bellfort Sugar Land, Texas. Introduction
ONSITE PROVING OF GAS METERS Daniel J. Rudroff WFMS Inc. 13901 West Bellfort Sugar Land, Texas Introduction With the increased use of Natural Gas as a fuel, and higher natural gas prices buyers and sellers
More informationCHAPTER 2 EXPERIMENTAL SETUP AND PROCEDURE
22 CHAPTER 2 EXPERIMENTAL SETUP AND PROCEDURE 2.1 EXPERIMENTAL COLUMN All the experiments were carried out in an internal loop airlift fluidized bed and combined loop fluidized bed (an external down comer
More informationRon Gibson, Senior Engineer Gary McCargar, Senior Engineer ONEOK Partners
New Developments to Improve Natural Gas Custody Transfer Applications with Coriolis Meters Including Application of Multi-Point Piecewise Linear Interpolation (PWL) Marc Buttler, Midstream O&G Marketing
More informationTechnical Procedure for Instrument Calibration and Maintenance
Technical Procedure for Instrument Calibration and Maintenance 1.0 Purpose To outline the schedules and procedures involved in the maintenance and verification of critical instrument calibration in the
More informationTutorial. BOSfluids. Relief valve
Tutorial Relief valve The Relief valve tutorial describes the theory and modeling process of a pressure relief valve or safety valve. It covers the algorithm BOSfluids uses to model the valve and a worked
More informationTechnical Data Sheet High Performance Digital Gas Mass Flow Meters and Controllers Features Description
Smart-Trak Model 100 High Performance Digital Gas Mass Flow Meters and Controllers Features Measure & control gas mass flow rates up to 1000 slpm True digital performance provides high accuracy and great
More informationUNIVERSITY OF WATERLOO
UNIVERSITY OF WATERLOO Department of Chemical Engineering ChE 524 Process Control Laboratory Instruction Manual January, 2001 Revised: May, 2009 1 Experiment # 2 - Double Pipe Heat Exchanger Experimental
More informationAVA Purpose AVA Background spacing factor specific surface AVA Principle AVA AVA
Purpose The AVA (Air Void Analyzer) is used to measure the air-void parameters (spacing factor and specific surface) of samples of fresh air-entrained concrete. Samples are taken after concrete has been
More information780S Ultra High Purity Series
Smart Ultra High Purity Thermal Gas Mass Flow Meter Features Measures mass flow directly, no seperate temperature or pressure inputs required Field adjustment of critical flow meter settings via on-board
More informationExperiment P18: Buoyant Force (Force Sensor)
PASCO scientific Physics Lab Manual: P18-1 Experiment P18: (Force Sensor) Concept Time SW Interface Macintosh file Windows file Newton's Laws 45 m 300/500/700 P18 P18_BUOY.SWS EQUIPMENT NEEDED CONSUMABLES
More information3 1 PRESSURE. This is illustrated in Fig. 3 3.
P = 3 psi 66 FLUID MECHANICS 150 pounds A feet = 50 in P = 6 psi P = s W 150 lbf n = = 50 in = 3 psi A feet FIGURE 3 1 The normal stress (or pressure ) on the feet of a chubby person is much greater than
More informationLab #1 Pressure: Bubblers and Water Balloons CEE 331 Fall 2003
CEE 331 Lab 1 Page 1 of 9 SAFETY Lab #1 Pressure: Bubblers and Water Balloons CEE 331 Fall 2003 Laboratory exercise based on an exercise developed by Dr. Monroe Weber-Shirk The major safety hazard in this
More informationThis addendum describes the back pressure control implementation if FuelCell Version 4.1a and later.
FuelCell Addendum - Back Pressure Controls D. Johnson, Scribner Associates, Inc. 7/23/2015, Ver. 6 Introduction This addendum describes the back pressure control implementation if FuelCell Version 4.1a
More informationEXPERIMENT 1 AIR PRESSURE CONTROL SYSTEM
EXPERIMENT 1 AIR PRESSURE CONTROL SYSTEM 1.0 OBJECTIVE To study the response of gas pressure control using PID controller. 2.0 INTRODUCTION TO THE APPARATUS a) The process plant consists of two air vessels
More informationMoLE Gas Laws Activities
MoLE Gas Laws Activities To begin this assignment you must be able to log on to the Internet using Internet Explorer (Microsoft) 4.5 or higher. If you do not have the current version of the browser, go
More informationMass Flowmeters. Mass Flowmeters for Gases
Mass Flowmeters Mass Flowmeters for Gases Fast, Accurate, Low Pressure Drop for critical measurement applications! Flow (Std l/min) Response to a step change in flowrate 300 200 100 4 Milliseconds 63%
More informationXactix XeF2 OPERATION MANUAL
General Information The Xactix e-1 is a xenon difluoride (XeF 2) isotropic silicon etcher. XeF 2 is a vapor phase etch, which exhibits very high selectivity of silicon to photo-resist, silicon dioxide,
More informationTECHNICAL INFORMATION
TECHNICAL INFORMATION Installation, Operation, and Maintenance Balston Model 75-77 Nitrogen Generator Figure 1-75-77 Overall Dimensions These instructions must be thoroughly read and understood before
More informationBoyle s Law: Pressure-Volume Relationship in Gases
Boyle s Law: Pressure-Volume Relationship in Gases The primary objective of this experiment is to determine the relationship between the pressure and volume of a confined gas. The gas we will use is air,
More informationInstructions for SMV 3000 Multivariable Configuration (MC) Data Sheets
Instructions for SMV 3000 Multivariable Configuration (MC) Data Sheets Similar to the TC option for ST 3000 transmitters, the MC option for the SMV 3000 provides a service to our customers which results
More informationSwing Labs Training Guide
Swing Labs Training Guide How to perform a fitting using FlightScope and Swing Labs Upload Manager 3 v0 20080116 ii Swing labs Table of Contents 1 Installing & Set-up of Upload Manager 3 (UM3) 1 Installation.................................
More informationAC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW
AC 2010-2145: MEASUREMENT OF HYDROGEN IN HELIUM FLOW Randy Buchanan, University of Southern Mississippi Christopher Winstead, University of Southern Mississippi Anton Netchaev, University of Southern Mississippi
More informationProcess Dynamics, Operations, and Control Lecture Notes - 20
Lesson 0. Control valves 0.0 Context Controller output is a signal that varies between 0 and 100%. Putting this signal to use requires a final control element, a device that responds to the controller
More informationNumerical Simulations of a Train of Air Bubbles Rising Through Stagnant Water
Numerical Simulations of a Train of Air Bubbles Rising Through Stagnant Water Hong Xu, Chokri Guetari ANSYS INC. Abstract Transient numerical simulations of the rise of a train of gas bubbles in a liquid
More informationDeploying the TCM-1 Tilt Current Meter in an Inverted (Hanging) Orientation By: Nick Lowell, Founder & President
Lowell Instruments Application Note #TCMA Deploying the TCM-1 Tilt Current Meter in an Inverted (Hanging) Orientation By: Nick Lowell, Founder & President 1 Introduction The TCM-1 Tilt Current Meter (TCM)
More informationLab 1: Pressure and surface tension. Bubblers, gravity and the mighty paper clip.
Lab 1: Pressure and surface tension. Bubblers, gravity and the mighty paper clip. CEE 3310 - Summer 2012 SAFETY The major safety hazard in this laboratory is a shock hazard. Given that you will be working
More informationNavy Guidance and Tips for Using DOEHRS-IH Ventilation NAVY & MARINE CORPS PUBLIC HEALTH CENTER
Navy Guidance and Tips for Using DOEHRS-IH Ventilation NAVY & MARINE CORPS PUBLIC HEALTH CENTER October 2010 Purpose This document is a supplemental Navy guide to the DOEHRS Student Guide/User Manual Version
More informationFor Multi-Parameter Meters see mvx
Design Features VORTEX IN-LINE FLOW METERS For Multi-Parameter Meters see m Principles of Operation n No moving parts to wear or fail. n Electronics can be remotely mounted up to 30.5 m (0 ft). n No fluid
More informationLaboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.
Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions
More informationElectronic gas volume corrector model DGVC-04
Electronic gas volume corrector model DGVC-04 1. Introduction The following user s manual gives information about the installation, configuration, usage and storage of the electronic gas volume corrector
More informationAFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features
ANALOG MASS FLOW CONTROLLERS Model AF mass fl ow controllers are designed to indicate fl ow rates and control set fl ow rates of gases. Each of these units incorporates an advanced straight tube sensor
More informationAppendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation
Page 1 of 19 : SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor Description and Principle of Operation The photoacoustic multi-gas monitor (INNOVA 1412, Innova AirTech Instruments, Denmark) is a highly
More informationTechnical Information
Technical Information Installation, Operation, Installation, Operation and Maintenance and Maintenance Manual Manual Balston Model 75700-K728 Nitrogen Generator Figure 1-75700-K728 Overall Dimensions These
More informationDigital Level Control One and Two Loops Proportional and Integral Control Single-Loop and Cascade Control
Digital Level Control One and Two Loops Proportional and Integral Control Single-Loop and Cascade Control Introduction This experiment offers a look into the broad field of process control. This area of
More informationGolf Genius Software
CARSON VALLEY WOMEN S GOLF CLUB Golf Genius Software On-Line User Manual Kathy Belvel 6/3/2018 User step by step instructions for accessing and using the full array of capabilities available in the Golf
More informationExperiment AMe-1: Small Animal Respiratory Exchange Ratio (RER)
Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER) Background There are two main sources of energy available for animal metabolism: carbohydrates (CHO) and fats. These molecules are broken
More informationFor Multi-Parameter Meters see mvx
VORTEX IN-LINE FLOW METERS Design Features For Multi-Parameter Meters see mvx Principles of Operation VX No moving parts to wear or fail. Electronics can be remotely mounted up to 30.5 m (0 ft). No fluid
More informationFeatures. Main Voltage: Vdc. Input Control Signal: 0-5 Vdc or 4-20 madc (Sourcing) Monitor Output Voltage: 0-5 Vdc or 4-20 madc (Sourcing)
1 X-Flow Mass Flow Controller Flow Range from 0.8 ml/min to 20 l/min Target Markets Laboratory and Process Instrumentation Pharma / Bio-Pharma Equipment Air Quality Monitoring Systems Furnace and Coatings
More informationIntegral type Differential pressure flowmeter VNT Series
Integral type Differential pressure flowmeter VNT Series OUTLINE VH series Wafer-Cone differential pressure flowmeter and high precision differential pressure transmitter are integrated into one flowmeter.
More informationMETHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER
1250 METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER 1.0 Scope and Application. 1.1 Analytes. Analyte CAS No. Sensitivity Total Organic Compounds N/A
More informationHUMOR 20. High-precision Humidity Calibrator HUMOR 20. Operation
High-precision Humidity Calibrator The role of humidity calibrations that are accurate, reproducible, and documentable is becoming more and more important. ISO quality guidelines and regulations according
More informationHigh Pressure Chem-SCAN Operating Manual
GAS INLET VALVES REACTIVE GAS PRESSURE RELIEF VALVE INERT GAS VENT VENT VALVE REACTOR INLET VALVES PRESSURE TRANSDUCERS REACTORS STIRRER & THERMOWELL HEATING JACKET STIRRER MOTORS High Pressure Chem-SCAN
More informationFlow Vision I MX Gas Blending Station
Flow Vision I MX Gas Blending Station Alicat Scientific, Inc. 7641 N Business Park Drive Tucson, Arizona 85743 USA alicat.com 1 Notice: Alicat Scientific, Inc. reserves the right to make any changes and
More informationTransPort PT878GC Panametrics Portable Gas Ultrasonic Flowmeter. GE Sensing. Applications. Features
Applications The TransPort PT878GC clamp-on gas flowmeter is a complete ultrasonic flow metering system for measurement of most gases, including: Natural gas Compressed air Fuel gases Erosive gases Corrosive
More information