Hygea Ultrasonic Validation Kit
|
|
- Theodora Hampton
- 6 years ago
- Views:
Transcription
1 Hygea Ultrasonic Validation Kit Ultrawave s Hygea Ultrasonic Validation kit contains the tools and accessories required to perform Validation and Periodic testing of your ultrasonic cleaning bath. Designed to meet the testing guidelines for measuring the ultrasonic activity and cleaning efficacy of an ultrasonic bath as stipulated in HTM guidelines, the Hygea Ultrasonic Validation kit assures the operator that the ultrasonic bath is performing effectively. Why measure? The need to confirm the performance of all processing equipment in healthcare and industrial manufacturing environments is all important. HTM01-05 recommends that a measurement of ultrasonic activity and cleaning efficacy for dental ultrasonic baths is taken on a regular basis. Ultrasonic activity is measured using the Hygea Ultrasonic Activity Meter and aluminium foil test, while STF Load Check indicators are used to measure cleaning efficacy. All results can then be recorded in the Validation logbook provided. Kit Contents The Hygea Ultrasonic Validation Kit consists of the following components which will allow you to periodically test your ultrasonic cleaning system to ensure optimum performance: 1 x Hygea Ultrasonic Activity Meter Aluminium foil roll 2 x STF Load Check Holders STF Load Check Strips Ultrasonic Testing log book all packaged in a handy carry case.
2 Ultrasonic Activity Detection Foil Ablation Test The activity of an ultrasonic cleaner may be investigated by the erosion pattern which is created on aluminium foil exposed in the bath for a short period. The activity will not be uniform throughout the ultrasonic bath. Tests carried out during commissioning are intended to establish the variation in activity at different positions within the bath and the time required to obtain a characteristic erosion pattern. Tests should be carried out quarterly to ensure the ultrasonic activity remains unchanged. Equipment and materials The following equipment and materials are required: 1. Aluminium foil of nominal thickness mm to mm (A 20mm wide, 240m reel can be ordered from Ultrawave: Part number C , aluminium foil). 2. Waterproof adhesive tape (autoclave indicator tape is ideal). 3. A stopwatch, graduated in 0.2s or better and with an accuracy over a period of 15 min of ± 0.5 s, or better. 4. A rule or tape measure graduated in mm. Method 1. Measure the depth of the bath from the level of the lid to the bottom of the bath. Let the depth be D mm. 2. Cut nine strips of aluminium foil 15 mm to 20 mm wide and {D + 120} mm in length. 3. Roll 120mm of prepared foil strip into a cylindrical shape and secure with a paper-clip. Repeat for all nine strips.
3 4. Ensure that the water in the tank is at the required level, that the required amount of any chemical additive has been added and that the water in the tank is at the specified operating temperature. 5. Start the ultrasonic cycle and carry out the degas procedure to eliminate dissolved gases from the liquid in the bath. 6. Using strips of autoclave indicator tape across the top of the bath suspend nine strips of the prepared foil in the bath in a 3 x 3 grid. 7. The rolled end of each foil strip acts as a sinker weight to maintain the foil in an approximately vertical position. The sinker weight should be not more than 10 mm above, but not touching, the bottom of the bath 8. Run an ultrasonic cycle for a minimum of 30 seconds. Foil strips in position Inadequate or uneven erosion patterns 9. Remove the strips from the bath, blot dry and examine. 10. Drain the bath and clean to remove debris of eroded aluminium foil. Results Positive or satisfactory erosion patterns The zones of maximum erosion should be at similar positions on all nine foils and each should be eroded to a similar extent (by visual inspection). On re-testing the extent of erosion and the erosion pattern should have remained consistent with those originally determined during commissioning. Notes. The strips may be filed conveniently by sticking them to an A4 sheet of plain paper using a transparent adhesive tape. Laminating will provide a permanent record. The erosion pattern will depend on the thickness of the foil, the hardness of the foil, and the temperature of the liquid in the ultrasonic bath.
4 Ultrasonic Load Check Indicator Cleaning Efficacy Test The ultrasonic load check indicator is a consistent, repeatable way to monitor the cleaning efficacy of your ultrasonic bath. The indicator strips have been designed to perform in an equivalent manner to the test soils described in documents such as HTM2030 and HTM They represent typical soiling that would occur during normal use of reusable surgical instruments. These ultrasonic indicator strips contain two sources of protein - lipids and polysaccharides. Safe to handle and made with no blood products, the indicator strips allow easy to interpret results to be obtained, time after time. Test strips can be used with no risk of adding blood borne contaminants to the bath. The test strips can be used in the mesh design holder and should be placed in a basket within the ultrasonic bath. If the strips are being used in a hygea 2 benchtop bath, the strips should be placed on the middle instrument tray and the two other trays removed. As the ultrasonic activity occurs, the red test soil will be removed from the strips and different levels of soil removal will indicate different levels of ultrasonic activity. This is a bench marked performance test which is suitable for validation. The indicator strips mimic occluded surfaces to present a realistic ultrasonic cleaning challenge. In order to obtain meaningful results, test strips should be stored in a dry environment, preferably in a sealed container.
5 Instructions for use: 1. Ensure the ultrasonic bath has sufficient Water. 2. Ensure the detergent has been thoroughly mixed in the water and that the correct dosage has been applied (50ml of Sonozyme). 3. Ensure that the hygea 2 minimum temperature is set to 20 o C. 4. Ensure that the water has been degassed. This can be done by allowing the heating cycle to complete then running at least 1 x 10 minutes ultrasonic cycle. Allow the water to rest for no less than 3 minutes. 5. Ensure the strip holders and your hands are clean and dry. 6. Place one indicator strip in each holder, ensuring it is centrally placed and not protruding from either side. Place the holders in the basket as shown in Picture A. 7. After running a complete cycle (at least six minutes in the hygea 2), remove the holders from the basket and carefully take out the indicator strips. Caution: the device may be hot and any residual soil from the strips may stain. Picture A 8. Inspect the indicator strips for evidence of remaining soil by placing the strips against a white background. 9. If more than 2% of the soil remains, cleaning of the load should be considered inadequate and departmental procedures should be followed in respect of failed cleaning efficacy testing. (see picture B below) Note: It is important to check the expiry date of the strips and that the holders must be placed horizontally as shown above. Not doing so will affect the cleaning efficacy results. Fail Fail Fail Pass Pass Picture B
6 Hygea Ultrasonic Activity Meter Specifications Probe Assembly in Stainless Steel mm long Handle in ABS Plastic 105mm long Instrument Case in ABS Plastic mm Weight (probe and cable) 350 grams Weight (instrument including battery) 200 grams Indicator 1: Frequency 5-50 KHz Indicator 2: Power % Battery 9V PP3 Cable Connector BNC 2P Warranty 12 months Calibration Period 12 months Operating Instructions 1. The Hygea Ultrasonic Activity Meter should be used to measure the level of ultrasonic activity throughout the tank. It is recommended that the tank be divided into 9 sections and measurements taken at just below the surface and at least 3 cm from the base of the bath, where the depth of the tank allows. 2. Insert the probe into the cleaning fluid inside the tank. To take a reading press and hold the button on the meter. Record the results displayed on both the Frequency and Power displays. 3. Repeat this in the 9 specified places at the two different depths throughout the bath, including each corner. 4. Record the results so that future tests can be compared. If there is a reduction in the readings over time, your ultrasonic cleaning system may not be working at full efficiency and may need validation and servicing. Note: It should not be expected that the Meter will give readings of 100% every time. The Hygea Ultrasonic Activity Meter is designed to be used as a comparative measurement device over a period of time. If the readings change significantly, it is recommended that you contact Ultrawave for information and advice.
Ultrasonic baths. XUB range of digital ultrasonic baths. XUBA range of analogue ultrasonic baths. Digitally controlled range with a choice of 5 baths
7 Ultrasonic baths XUB range of digital ultrasonic baths Digitally controlled range with a choice of 5 baths XUBA range of analogue ultrasonic baths Entry level baths with analogue controls Ultrasonic
More informationSUGGESTED POLICY FOR THE MONITORING OF THE CLEANING EFFICIENCY OF AN ULTRASONIC CLEANERS( SONIC & TOSI) WEEKLY
SUGGESTED POLICY FOR THE MONITORING OF THE CLEANING EFFICIENCY OF AN ULTRASONIC CLEANERS( SONIC & TOSI) WEEKLY SUBJECT: ULTRASONIC CLEANER MONITORING WEEKLY DEPARTMENT: Central Service APPROVED BY: EFFECTIVE:
More informationCompetency Record for Sonic Cleaning
Competency Record for Sonic Cleaning Name: Competency Achieved: (Date) Evaluator: Learner: Comments: Background information on sonic testing ST 79; 2009A section 7.5.3.3 states this on weekly testing Mechanical
More informationUsER manual for Watersens ph -REDOX
UsER manual for Watersens -REDOX Cl 8 1 2 6 3 3 7 7 4 4 4 4 Parts List 1 Redox Probe 1 x 2 PH Probe 1 x 5 Tube Weight 2 x 6 Connection Valve 1 x chlorine 3 Chlorine and Pumps 2 x 7 Dosing Valve 2 x 5 5
More informationQi Litre Ultrasonic Cleaner
Qi-300 17Litre Ultrasonic Cleaner Ultrawave s Qi-Series are robust and reliable, delivering outstanding cleaning performance in a wide range of industries. From precision, pharmaceutical and medical device
More informationEurosonic 4D - 3D - Micro ENGLISH
Eurosonic 4D - 3D - Micro ENGLISH INDEX The state of the art ultrasonic cleaners. 05 Eurosonic 4D. For those who don t settle for second best. 07 HOW TO USE THE COVER 15 DIRECT METHOD AND INDIRECT METHOD
More informationInstruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text.
Hach LDO Sensor Instruction Sheet 00745589 Safety Precautions Please read this entire instruction sheet before operating this sensor. Pay particular attention to all danger and caution statements. Failure
More informationBioAerosol Nebulizing Generator. Operation and Maintenance User Manual
BioAerosol Nebulizing Generator Operation and Maintenance User Manual INTRODUCTION The BANG or BioAerosol Nebulizing Generator is a unique nebulizer for the generation of aqueous aerosols at a low air
More informationthe annulus between the second casing string and the surface casing [must be] open to the atmosphere in the manner described in subsection (2).
Bulletin 2011-35 December 20, 2011 Surface Casing Vent Requirements for Wells This bulletin clarifies Section 6.100(4) of the Oil and Gas Conservation Regulations (OGCR) and provide additional surface
More informationENVIRONMENTAL LABORATORY APPROVAL PROGRAM CERTIFICATION MANUAL
1 of 7 The quality assurance plan as presented in the Quality Manual should assure that constant and consistent test conditions are met and verified and should be tailored to the laboratory s activities.
More informationANALOG ULTRASOUND WATTMETER USP-30
ANALOG ULTRASOUND WATTMETER USP-30 USER MANUAL BC BIOMEDICAL USP-30 ULTRASOUND WATTMETER TABLE OF CONTENTS WARNINGS, CAUTIONS, NOTICES... ii DESCRIPTION... 1 TEST SETUP... 2 TESTING... 5 MANUAL REVISIONS...
More informationTQC Bresle Chloride test 2008
TQC Bresle Chloride test 2008 SP7310 MANUAL See also : ISO8502-6 ISO8502-9 1 DETERMINATION OF SOLUBLE CONTAMINANTS ON SURFACES IN CONFORMITY WITH ISO 1.1 PREPARATION With this system very low concentrations
More informationSECTION SHOWER AND BATH ENCLOSURES. Display hidden notes to specifier by using Tools / Options / View / Hidden Text.
SECTION 10185 SHOWER AND BATH ENCLOSURES Display hidden notes to specifier by using Tools / Options / View / Hidden Text. PART 1 GENERAL 1.1 SECTION INCLUDES A. Bathtub wall enclosures. B. Shower wall
More informationCP10 Sensor Installation and Maintenance Instructions
4030150/9 IM-P403-26 EMM Issue 9 CP10 Sensor Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation 4. Maintenance 5. Spare parts Copyright 2017
More informationDISSOLUTION TEST FOR SOLID DOSAGE FORMS
Seite 1 von 15 01/2010:20903 corrected 6.8 2.9.3. DISSOLUTION TEST FOR SOLID DOSAGE FORMS This test is provided to determine compliance with the dissolution requirements for solid dosage forms administered
More informationSDX Submersible Depth Transmitter User Manual
SDX Submersible Depth Transmitter User Manual July 2017 USER INFORMATION Stevens makes no warranty as to the information furnished in these instructions and the reader assumes all risk in the use thereof.
More informationGas Pressure Probe. Information Sheet. Total Dissolved Gas
Total Dissolved Gas Information Sheet T900 Total Dissolved Gas Pressure Probe Description The In-Situ Inc. T900 probe is a specially adapted submersible pressure transducer designed to monitor Total Dissolved
More informationWater Weir Flow Controller. Introduction. Safety Precautions. Mounting the Hardware
57007-88 Introduction Safety Precautions This instruction sheet describes how to set up and use the Hach (Figure 1). A water weir is a device that raises or diverts water to regulate the flow. Hach s water
More informationBest Practice for Calibrating LTH Conductivity Instruments
Application Note Best Practice for Calibrating LTH Conductivity Instruments As accurate process measurement becomes an everyday requirement it is vital to be able to calibrate conductivity instruments
More informationOPERATING INSTRUCTIONS MANUAL FOR QACV PNEUMATIC DOSING PUMP
This operating instructions contains safety information that if ignored can endanger life or result in serious injury. They are indicated by this icon. Use of this pump with radioactive chemicals is forbidden!
More informationDTG - LCD Digital Temperature Gauge USER MANUAL
DTG - LCD USER MANUAL Page 1 of 7 USERS GUIDE page 1. Description 3 2. Function 3 3. Safety Instruction 3 3.1. Safety Conventions 3 3.2. Proper Use 3 4. Installation 4 4.1. Unpacking 4 4.2. Storage 4 4.3.
More informationTourniquet Systems Pressure Infusion
Tourniquet Systems Pressure Infusion VBM Medizintechnik GmbH VBM 1 VBM 2 Summary Tourniquets and Accessories Digital Tourniquets... Compressed Air Tourniquets... Electronic Tourniquets... Application Tourniquets...
More information711 DISSOLUTION. Portions of the present general chapter text that are national USP text, and therefore not part of the
Change to read: 711 DISSOLUTION This general chapter is harmonized with the corresponding texts of the European Pharmacopoeia and/or the Japanese Pharmacopoeia. The texts of these pharmacopeias are therefore
More informationElcometer 138. Bresle Kit and Patches. Operating Instructions
Elcometer 138 English Bresle Kit and Patches Operating Instructions R English This product meets the Electromagnetic Compatibility Directive. The product is Class B, Group 1 ISM equipment according to
More informationStandard. Standard. Venepuncture Arm. Light. Part No: Brown. Part No: Black. Part No: 00332
Setup Guide Standard Standard Light. Part No: 000 Brown. Part No: 00 Black. Part No: 00 Part No: 065-964 Issue, June 0 0 Limbs & Things For more skills training products visit limbsandthings.com Limbs
More informationAqua-DP. Aqua-DP Waterproof Dissolved Oxygen - ph -mv Temperature Meter. TPS Australia - Quality hand-made instruments since 1968.
TPS Australia - Quality hand-made instruments since 1968. Waterproof Dissolved Oxygen - ph -mv Temperature Meter Dual readout of Dissolved Oxygen and ph/mv Waterproof to IP65 Easy to use One-touch calibration
More informationBP48-1 AND SSP48-1 IN-GROUND POST INSTALLATION INSTRUCTIONS
BP48-1 AND SSP48-1 IN-GROUND POST INSTALLATION INSTRUCTIONS WARNING: THIS IN-GROUND POST IS NOT DESIGNED FOR USE WITH AN LP GAS CYLINDER. WARNING: SEE YOUR GRILL OWNER S MANUAL FOR PROPER LOCATION, MINIMUM
More informationUser Guide. Pall Laboratory Manifold. For laboratory use. Not for use in a manner other than indicated. Introduction. Regulatory References
User Guide Pall Laboratory Manifold For laboratory use. Not for use in a manner other than indicated. Introduction Application and Intended Use The membrane filtration (MF) technique is a method of analyzing
More informationSDX Submersible Depth Transmitter User Manual
SDX Submersible Depth Transmitter User Manual October 2007 USER INFORMATION Stevens makes no warranty as to the information furnished in these instructions and the reader assumes all risk in the use thereof.
More informationElcometer 138. Bresle Kit and Patches. Operating Instructions
Elcometer 138 English Bresle Kit and Patches Operating Instructions R English The Conductivity Meter model B-173 has been tested in accordance with EU regulations governing Electro-magnetic compliance
More informationSTERILIZABLE DISSOLVED OXYGEN SENSORS
STERILIZABLE DISSOLVED OXYGEN SENSORS USER MANUAL Table of Contents 1. Introduction 2. Operation 3. Calibration 4. Electrode Verification 5. Mounting 6. Maintenance 7. Storage 8. Instructions for Exchange
More informationPP-1, PP-1W, PP-2, & PP-3
Welker Pitot Probe Model PP-1, PP-1W, PP-2, & PP-3 The information in this manual has been carefully checked for accuracy and is intended to be used as a guide for the installation, operation, and maintenance
More informationStandard. Bag & Stand Advanced Venepuncture Arm. Light. Part No: Brown. Part No: Black. Part No: 00442
Setup Guide Bag & Stand Advanced Standard Light. Part No: 00440 Brown. Part No: 0044 Black. Part No: 0044 Part No: 065-0 Issue, May 0 0 Limbs & Things For more skills training products visit limbsandthings.com
More informationQE-1, Rev Warner Instruments, Inc Dixwell Avenue, Hamden, CT (800) / (203) (203) fax
QE-1, Rev. 03.27.03 35 mm Quick Exchange Platform Model QE-1 1125 Dixwell Avenue, Hamden, CT 06514 (800) 599-4203 / (203) 776-0664 (203) 776-1278 - fax QE-1, Rev. 03.27.03 Table of Contents DESCRIPTION...3
More informationWIND CLIPPER KTS ILLUM SCALE INC DEC CLIPPER WIND SYSTEM
CLIPPER WIND KTS ILLUM SCALE DEC INC CLIPPER WIND SYSTEM TABLE OF CONTENTS INTRODUCTION PRE-TEST OF INSTRUMENT INSTALLING THE MASTHEAD SENSOR UNIT INSTALLING THE DISPLAY NORMAL OPERATION CHANGING THE
More information9408 Series Dissolved Oxygen Systems. Dissolved oxygen systems designed for reliable operation with minimal operating costs
Data sheet DS/9408 EN Rev. M 9408 Series Dissolved oxygen systems designed for reliable operation with minimal operating costs Reduced sensor fouling design minimizes operational costs Low maintenance,
More informationField Manual. Simple by Design
Simple by Design US Patents No. 6,481,300; No. 6,837,120; others pending Field Manual The HydraSleeve is a simple tool. In keeping with the Simple by Design motto, these are the basic instructions. Please
More informationRoyce Interface Level Analyzer and Sensors
Royce Interface Level Analyzer and Sensors Model 2511A Interface Level Analyzer The Model 2511A is capable of having the ultrasonic speed of sound signal from its transducers changed by the user in applications
More informationGILMONT ACCUCAL FLOWMETERS
OPERATING MANUAL GILMONT ACCUCAL FLOWMETERS 28W092 Commercial Ave. Barrington, IL U.S.A. 60010-2392 (847) 381-4888 (847) 381-7053 (Fax) 800-962-7142 www.barnant.com e-mail: barnant@barnant.com A-1299-0766
More informationHOW TO DIG A TEST PIT
HOW TO DIG A TEST PIT An archaeological test pit is a small 1m by 1m trench dug in a series of layers, or spits, to a depth of approximately 1m. The archaeological materials collected from these layers
More informationABB MEASUREMENT & ANALYTICS DATA SHEET series Dissolved oxygen systems
ABB MEASUREMENT & ANALYTICS DATA SHEET 9408 series Dissolved oxygen systems Measurement made easy Dissolved oxygen systems designed for reliable operation with minimal operating costs Reduced sensor fouling
More informationTECHNOLOGIES. M Series. Positive return metering pumps [API675]
M Series Positive return metering pumps [API675] Hydraulic double diaphragm metering pump in AISI 316L stainless steel A line of plunger and hydraulic double diaphragm metering pumps designed according
More informationACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION
Application Note Doc. G8108-001 Rev. A - 23-Jul-02 ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION For more information regarding Digiquartz products contact: Paroscientific,
More informationSooriyan Pool and Spa Ionizer. by Hilltop Solar
Sooriyan Pool and Spa Ionizer by Hilltop Solar If you re reading this, then you ve made the right choice to save money on pool chemicals and electricity. Thank you for purchasing the premium solar ionizer
More informationSUMMITTM 400 & 600. Natural Gas Barbecues. Step-By-Step Guide
SUMMITTM 400 & 600 Natural Gas Barbecues Step-By-Step Guide W E B E R W E B E R W E B E R W E B E R Summit 400 NG Summit 600 NG CANADIAN GAS ASSOCIATION R A P P R O V E D WARNING: Follow all leak check
More informationOptical Dissolved Oxygen Meter
Optical Dissolved Oxygen Meter HI764113 Rugged Optical Dissolved Oxygen Probe for Fresh and Saltwater Applications Digital, weighted probe No membranes No electrolytes No oxygen consumption No flow dependence
More informationUser Manual V14.1 IP68 LED FLEX NEON
User Manual V14.1 IP68 LED FLEX NEON PLEASE READ THESE INSTRUCTIONS CAREFULLY BEFORE INSTALLATION LEAVE A COPY FOR THE END USER/MAINTENANCE ENGINEER FOR FUTURE REFERENCE 1/11 PRODUCT SPECIFICATION 1. Operating
More informationMAKE YOUR SAMPLE COUNT! A PRIMER ON COLLECTING QUALITY, REPRESENTATIVE SAMPLES
TURF MATTERS I BY SAM FERRO President Turf Diagnostics & Design DIAGNOSTIC SPECIAL FOCUS SERVICES REPORT MAKE YOUR SAMPLE COUNT! A PRIMER ON COLLECTING QUALITY, REPRESENTATIVE SAMPLES "Follow the guidelines
More informationModel PDT Dewpoint Transmitter
Model PDT Dewpoint Transmitter Instruction Manual Alpha Moisture Systems Alpha House 96 City Road Bradford BD8 8ES England Tel: +44 1274 733100 Fax: +44 1274 733200 email: mail@amsytems.co.uk web: www.amsystems.co.uk
More informationRAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions
RAM 4021-PR Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021-PR For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4
More informationBRIEFING 711 DISSOLUTION
BRIEFING 711 Dissolution, USP 28 page 2412 and page 234 of PF 30(1) [Jan. Feb. 2004]. The United States Pharmacopeia is the coordinating pharmacopeia in the efforts toward interntional harmonization of
More informationEquipment Accuracy / Selection
Revision 5.0: Included the PPP program for clarification. Revision 6.0: Update references for CTL PDSH 251B to CTL DSH 251E" Added leakage current tolerances for values above 5KHz For Client Labs Purpose
More informationSupratec AERATION TECHNOLOGY. von-drais-straße 7 D Simmern / Hunsrück. tel.: / fax: /
PERMOX OM120 Oval Membrane Tu b e D i ffuser w i th stainless steel b o d y Supratec for fine bubble or coarse bubble diffusion of liquids Supratec UND GESELLSCHAFT FÜR UMWELT- VERFAHRENSTECHNIK MBH AERATION
More informationCLASS D - SENSITIVE LEAK TEST GAS AND BUBBLE METHOD. 1.1 To provide definitive requirements for PNEUMATIC pressure testing of piping systems.
Page 1 of 7 CLASS D - SENSITIVE LEAK TEST GAS AND BUBBLE METHOD 1. SCOPE 1.1 To provide definitive requirements for PNEUMATIC pressure testing of piping systems. 1.2 The piping system as used herein is
More informationList of Contents vma. Compressed Air Conditioning 2. System description 2. Individual Units Pre-Filters 3. Combinations 6
Úprava stlačeného vzduchu EWO VMA List of Contents System description Individual Units Pre-Filters 3 Micro-Filters 4 Activated-Charcoal-Filters 5 Combinations 6 Accessories and Main spare parts 6 Service
More informationUltraLo 1800 Alpha Particle Counter
XIA LLC UltraLo 1800 Alpha Particle Counter Site Requirements & Planning Version: 0.4 Friday, February 15, 2013 Table of Contents I. Site Preparation Tasks... 3 A. Instrument Overview... 3 B. Installation
More informationOptical Dissolved Oxygen Meter
Optical Dissolved Oxygen Meter HI764113 Rugged Optical Dissolved Oxygen Probe for Fresh and Saltwater Applications Digital probe Fast and stable readings No membranes Not affected by sunlight No electrolytes
More informationInternal Sealing of Masts
595-814-E 2007-02-15 Internal Sealing of Masts (Replaces the sealing instructions 595-551-SE) Vertical high line Kicker bracket Sec. sealing hole Drain hole Conventional masts This instruction is enclosed
More informationEcoCell Vertical Electrophoresis Unit
Instructions for EcoCell Vertical Electrophoresis Unit Cat. No.: AN 12005 Ringstr. 4 64401 Gross-Bieberau Tel. ++49-6162-809840 Fax ++49-6162-8098420 www.anamed-gele.com Warning These units are capable
More informationUltima Salt System Overview
When only the BEST will do for your salt pool! Ultima Salt System Overview The reasons we started a comprehensive Salt System program: Although salt chlorine generators automate the daily addition of your
More informationBante810 Benchtop Dissolved Oxygen Meter Instruction Manual
Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante810 Benchtop Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante810 benchtop dissolved oxygen
More informationCompatible with a Wide Range of Gases or Liquids for Laboratory or Plant Applications
Riteflow Flowmeters Compatible with a Wide Range of Gases or Liquids for Laboratory or Plant Applications Used for flow measurements in the lab, process flows in pilot plants, metering carrier gases in
More informationRisk Analysis and Hazard Reduction Report V1.0
Risk Analysis and Hazard Reduction Report V1.0 2 Contents 1. Definitions and Risk Classifications 3 2. Risk and Hazard reduction associated with Personnel 4-5 3. Risk and Hazard reduction associated with
More informationSTEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE
STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE GENERAL INFORMATION This family of dissolved oxygen electrodes is of the steam sterilizable polarographic type, designed to be interchangeable with NBS
More informationSEMEM3-73 Maintaining medical gas pipeline systems and equipment
Maintaining medical gas pipeline systems and equipment Overview This unit identifies the competences you need to carry out maintenance activities on medical gas pipeline systems and equipment, in accordance
More informationCanNeed-DSST-100 Digital Secure Seal Tester --- To evaluate the secure seal quality of PET bottles, glass bottles and cans
CanNeed-DSST-100 Digital Secure Seal Tester --- To evaluate the secure seal quality of PET bottles, glass bottles and cans New piercing mode Traditional Piercing Mode The CanNeed -DSST-100 Digital Secure
More informationDepth sensor. Product reference : REV 1. USER GUIDE and INSTALLATION GUIDE. nke Sailing competition
Depth sensor Product reference : 90-60-456 REV 1 USER GUIDE and INSTALLATION GUIDE nke Sailing competition Z.I. Kerandré Rue Gutenberg 56700 HENNEBONT- FRANCE http://www.nke.fr After sale service n 33
More informationTS1. Ultrasonic Tank Sender. Installation and Operating Instructions. For TS1 Firmware v3.8. Page 1 INST-TS1-V13 18/11/10
TS1 Ultrasonic Tank Sender Installation and Operating Instructions For TS1 Firmware v3.8 Page 1 Table of Contents 1. FEATURES... 3 2. SPECIFICATIONS... 3 3. DIMENSIONS... 4 4. MOUNTING AND INSTALLATION...
More informationExperiment 18 Properties of Gases
Experiment 18 Properties of Gases E18-1 E18-2 The Task In this experiment you will investigate some of the properties of gases, i.e. how gases flow, their phase changes and chemical reactivity. Skills
More informationExpert Hydrostatic Level Transmitters
Expert Hydrostatic s General Features MJK Expert hydrostatic level transmitters are designed for level measurement by submerging the transmitter in open channels, drains and tanks. Expert hydrostatic level
More informationRuskinn Hypoxia and Anoxia Workstations. innovation in incubation
Ruskinn Hypoxia and Anoxia Workstations The Ruskinn range of hypoxia workstations has been designed to replicate low oxygen in vivo physiology providing the ideal research platform for cell biologists
More informationRAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions
RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4 Calibration....
More informationDigiquartz Water-Balanced Pressure Sensors for AUV, ROV, and other Moving Underwater Applications
Digiquartz Water-Balanced Pressure Sensors for AUV, ROV, and other Moving Underwater Applications Dr. Theo Schaad Principal Scientist Paroscientific, Inc. 2002 Paroscientific, Inc. Page 1 of 6 Digiquartz
More informationMaintenance and Calibration
Maintenance and Calibration PeriFlux 6000 peripheral pressure made intelligent 44-00315-01 Disclaimer The information contained in this document is intended to provide general information only. It is not
More informationSCOUT DECK ANCHOR 3 1/4" BONDING SCREW
SCOUT DECK ANCHOR 16" 16" 1" X 4" ANCHOR BOLT (4) 10" 3 1/8" 5" 6" MIN COPPER WIRE TO ATTACH TO BONDING GRID 3 1/4" BONDING SCREW CONCRETE DECK SPECIFICATIONS: THE AQUA CREEK PRODUCTS' SCOUT DECK ANCHOR
More informationPOLYMER TEST KIT (CLAPPER METHOD) OFI PART No
OFI Testing Equipment Polymer Test Kit (Clapper Method) Part # 295-00 Instructions Page 1 of 4 POLYMER TEST KIT (CLAPPER METHOD) OFI PART No. 295-00 This test determines the Partially Hydrolyzed Polyacrylamide
More information11. Aquasalt. Aquasalt Aquasalt 440. Salt electrolysis demonstration 442. Spare parts and Accessories 443. Chemical Products
. 434. 440. Salt electrolysis demonstration 442. Spare parts and Accessories 443. Chemical Products 432 433 Pleasure and benefits for our health removes the unpleasant smell from chlorine in your swimming
More informationFood and Drug Administration Division of Pharmaceutical Analysis. Mechanical Qualification of Dissolution Apparatus 1 and 2
Page 1 of 10 Sections included in this document Table of Contents Document history 1. Purpose 2. Scope/Policy 3. Responsibilities 4. Background 5. References 6. Procedure 7. Records 8. Glossary - Not applicable
More informationGilson Aqua-Check Moisture Tester MA-26X
OPERATING MANUAL Gilson Aqua-Check Moisture Tester MA-26X Rev: 01/2019 PHONE: 800-444-1508 P.O. Box 200, Lewis Center, Ohio 43035-0200 740-548-7298 E-mail: customerservice@gilsonco.com Product Web Page:
More informationINDEX THE FINITO UNIVERSE BLANKET PLATE COATING PRESS WRAPPING
The Finito Universe INDEX 03 THE FINITO UNIVERSE 04 UNIVERSAL UNDERPACKING 06 BLANKET 08 PLATE COATING 10 PRESS WRAPPING The Finito Universe 3 Printgraph started to manufacture in 2005 and is one of the
More informationOperation manual Level sensor DC-LS-50 Operation Manual Level Sensor DC-LS-50
Operation Manual Level Sensor DC-LS-50 Rev.02 (2014.06) OL Page 1 of 8 Table of contents I. STRUCTURE OF THE MANUAL / CLARIFICATION... 3 II. SAFETY AND HEALTH CONCERNS... 4 1 INTRODUCTION... 5 1.1 Use
More informationST 18 Aluminium Gas Model
ST 18 Aluminium Gas Model Operators Manual Issue 5 For correct operation of this autoclave it is essential that the chamber be filled with water up to the height of the studs before every use. DIXONS SURGICAL
More informationDecontamination Procedures for Gilson Pipettes
Standard Operating Procedure for Pipettes Decontamination Procedures for Gilson Pipettes (defined for Users to Incorporate in SOPs) in Accordance with ISO8655 Standard Procedure LT802288/E - 2008 Gilson
More informationMULTICHANNEL. Labnet International, Inc.
Labnet International, Inc. A B A1 A2 C F D H E 4A 4B 4C 4D G CONTENTS 1 - INTRODUCTION 2 - VOLUME SETTING 3 - METHOD OF PIPETTING 4 - RECOMMENDATIONS - RECALIBRATION 6 - CLEANING AND STERYLIZATION 7 -
More informationBS Series Basket Strainer
BS Series Basket Strainer Operating, Installation, & Maintenance Manual Corrosion Resistant Fluid and Air Handling Systems. Dated 04-26-12 PRESSURE DROP SIMTECH strainers are engineered to offer the lowest
More informationGenerating Calibration Gas Standards
Technical Note 1001 Metronics Inc. Generating Calibration Gas Standards with Dynacal Permeation Devices Permeation devices provide an excellent method of producing known gas concentrations in the PPM and
More informationInstallation Operation Maintenance
682 Seal Cooler New generation seal cooler to meet and exceed the seal cooler requirements stated in the 4th Edition of API Standard 682 Installation Operation Maintenance Experience In Motion Description
More informationNuclear Associates , &
Nuclear Associates 76-801, 76-801-1926 76-804 & 76-807 Flood Phantoms Users Manual March 2005 Manual No. 76-801-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product
More informationInstructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems
Corporation Issued: 5/09 P/N: 600104 Rev. 2 Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems The Gas Addition Accessory permits the
More informationOperating Instructions for BAIR22-6 AIR-POWERED CRIMPING TOOL
Operating Instructions for BAIR22-6 AIR-POWERED CRIMPING TOOL Read and understand all of the instructions and safety information in this manual before operating or servicing this tool. Table of Contents
More informationModel PSI Compressor with 3-Gallon Air Tank 12VDC
Model 6350 150 PSI Compressor with 3-Gallon Air Tank 12VDC IMPORTANT: It is essential that you and any other operator of this product read and understandd the contents of this manual before installing
More informationSEMMME3-073 Checking that completed assemblies comply with specification
Checking that completed assemblies comply with specification Overview This unit identifies the competences you need to carry out checks on mechanical assemblies, in accordance with approved procedures.
More informationSS32L Dissolved O 2 Probe
BIOPAC WWW.biopac.com Systems, Inc. Application Note PH-185 SS32L Dissolved O 2 Probe SS32L Dissolved O 2 Probe BIOPAC Software SS32L Specifications BSL PRO v. 3.6.6 BIOPAC Hardware SS32L Dissolved O 2
More informationAdvanced Test Equipment Rentals ATEC (2832)
WARRANTY CLAUSE METROSONICS warrants each new cl-304 manufactured and sold to be free from defects in material, workmanship and construction, except for batteries which may be contained therein, and that
More informationRayPette Plus Autoclavable Pipette. User Manual
RayPette Plus Autoclavable Pipette User Manual CONTENTS 1. YOUR NEW PIPETTE... 1 1.1. Adjustable volume pipettes... 1 1.3 Fully autoclavable... 2 2. UNPACKING... 3 3. INSTALLING THE PIPETTE HOLDER... 4
More informationWork Method Statement. Prepared for: The Bolton NHS Foundation Trust Prepared by: Mark Fentiman Date:
Work Method Statement Prepared for: The Bolton NHS Foundation Trust Prepared by: Mark Fentiman Date: 06.03.2014 Detailed Work Plan The Bolton NHS Foundation Trust Date: 06.03.2014 1 Contents: 1. The Deprox
More informationOPERATOR S MANUAL Ar-Gone Weld Gas Analyzer
July 2011 OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer WARNING! Before operating this product, read and understand this Operator s Manual. Become familiar with the potential hazards of this unit. Contact
More informationCCPaSEC. Procedure for the Quality Control Yearly Equipment Check. July 2018 CCPaSEC Quality Control Team
CCPaSEC Procedure for the Quality Control Yearly Equipment Check July 2018 CCPaSEC Quality Control Team 1 ph and Conductivity Equipment Check. Battery check: replace the batteries. Inspect the conductivity
More informationProduct Manual B-Safety ClassicLine & PremiumLine Emergency Eyewash and Eye/Face Wash Equipment
Product Manual B-Safety ClassicLine & PremiumLine Emergency Eyewash and Eye/Face Wash Equipment 1. Application Emergency eyewash and eye/face wash equipment are prescribed first aid installations for workplaces
More informationModel VR6 System. Installation, Operation & Maintenance
Model VR6 System Installation, Operation & Maintenance General: All Archer Instruments chlorination systems are carefully designed and tested for years of safe, accurate field service. All Archer Instruments
More information