Total Pressure Measurement Using Residual Gas Analyzers

Size: px
Start display at page:

Download "Total Pressure Measurement Using Residual Gas Analyzers"

Transcription

1 Total Pressure Measurement Using Residual Gas Analyzers M. Maskell, Old Dominion University; G. Myneni, P. Adderley, Jefferson Lab; G. Brucker, Stanford Research Systems; C. Day, Forschungszentrum, Karlsruhe BACKGROUND The aim of this work is to improve measurement techniques using Residual Gas Analyzers (RGAs) through better understanding of the effect of gas species on pressure measurement. The behavior of ionization gauges (IGs), such as the extractor gauge, is quite linear into the 10^-11 Torr range; however the accuracy of these instruments suffers for gases other than the calibration standard, Nitrogen. Different ionization cross sections may cause greater or lesser sensitivity, and fragmentation may cause ion currents to become multiplied in the detector, giving total pressures that are too high. Some instruments, such as Ulvac s Axtran, include correction factors for various gases in their documentation. This information improves the accuracy of these devices for single gas species; however they are often calculated rather than found experimentally with the device itself, and can only be applied one at a time, thus limiting their usefulness for gases containing more than one species. Residual Gas Analyzers (RGAs) contain a quadrupole mass filter, which makes their calibration much more complicated than ion gauges with relatively static configurations, but also allows them to determine the relative concentrations of various species of gas and apply the appropriate corrections before summing to find the total pressure. This method of using an RGA as a total pressure gauge is cumbersome due to the need for calibration specific to its settings, but in situations where a high degree of accuracy is needed in a mixed gas species environment, the benefit may merit the effort. RGAs typically include a total pressure measurement display in their software; however, responsible manufacturers admonish the user that this total pressure reading is inherently inaccurate for any gas other than that which was used to obtain the total pressure calibration factor. The sensitivity of an RGA is dependent on its ion source configuration and its operating parameters, including ionization energy (electron energy), focus voltage, resolution setting, and ion energy. When used with an electron multiplier (as is advised at UHV pressures), the sensitivity also varies with the mass of the gas being detected and the age of the detector. Sensitivity is gas dependent, and for a gas g it is defined as (Use S g instead of S), where I and I 0 are the RGA reading at a given partial pressure and the background reading for that mass peak, respectively, and P and P 0 are the reference gauge readings for the pressure and background pressure. The sensitivity values presented in this paper are specific to the devices and parameters that were used. Thus the preferred method for using an RGA as a total pressure gauge is to operate in conjunction with a reference, such as the SRG and extractor gauge used in this experiment, so that the sensitivity for each gas can be determined for the particular device and parameters being used. Once the sensitivity has been found, the extractor can be removed and the RGA used as a stand alone total pressure gauge so long as the operating

2 parameters remain the same as when the sensitivity was determined. The gain of the electron multiplier must remain fixed for sensitivity calibrations to remain valid. It is therefore recommended that the electron multiplier voltage be routinely tuned to maintain a constant gain. This procedure is easily accomplished through RGA operating software. It should be noted that the removal of the extractor changes the system, and thus can create a source of error for future measurement. A SRG does not alter the vacuum system, so it is the preferred reference gauge for determining sensitivity. The SRG measures pressure directly and passively via molecular drag, it does not emit heat or ionizing electrons into the system. The following section describes in more detail the experimental setup and procedure. The basic premise is that the RGA sensitivity is normalized to 1 at ~4E-7 Torr by correcting both the RGA and the other ion gauges to match the SRG, and then observing the change in sensitivity of the RGA relative to the other ion gauges as pressure is decreased. EXPERIMENTAL PROCEDURE The experiment is performed in a cylindrical stainless steel vacuum chamber with gauge ports situated symmetrically around the circumference. The gas inlet is designed such that the gas is introduced at the top of the chamber and equidistant from all gauges. The chamber is pumped by a Pfeiffer Balzers TPU-180H turbo pump, and backed by a Varian scroll pump. The gas inlet is connected by a leak valve to a system of inlets pumped by its own turbo pump, which can be valved off, and backed by a diaphragm pump. Each of the inlets in this system has a leak valve leading to the gas samples to be studied. Stable pressures are achieved by establishing a flow through the vacuum chamber. The gauges installed on the vacuum chamber include three SRS 100 AMU RGAs, an MKS Spinning Rotor Gauge (SRG), an Ionivac IM 520 extractor gauge, and an Ulvac Axtran gauge. The RGAs are labeled,, and. is a unit that has been used extensively in systems at Jlab, with no factory reconditioning. is a used RGA that has undergone factory reconditioning. is a new RGA. The RGA resolutions are set to the standard 1 AMU peak width at 10% peak height. All RGA settings are kept at factory defaults or optimized for the particular device, unless otherwise noted. The characteristics of the RGAs are shown in Table 1. The same number of ionization gauges are active in each test (one RGA, the extractor, and the AxTRAN), so that addition or removal of ion gauges does not cause a temperature difference between tests. The SRG does not rely on ionization for its operation, and its mass dependence is completely calculable, making it easily corrected for by entering the molecular mass into the SRG controller; therefore, calibrating the extractor with the SRG eliminates these factors from the extractor reading, making the extractor an acceptable transfer standard to lower pressures. The RGAs are calibrated using the SRG as the reference gauge under ~10^-6 Torr of nitrogen, and except where otherwise stated, they remain calibrated for nitrogen. At the beginning of each trial, the extractor and Axtran are corrected to match the reading of the SRG at approximately 4*10^-7 Torr. Since these gauges are known to be quite linear into the UHV range, the correction factor applied to accomplish this is used throughout the trial. The RGA being tested is given a correction from the SRG at

3 this pressure as well. The sensitivities calculated from the data corrected by the SRG are normalized sensitivities, all starting at 1 at the highest pressure measured; this allows the percent change in sensitivity to be easily read from the graphs. The actual sensitivity of the RGA (when calibrated as above to have a sensitivity of 1 for nitrogen) at the highest pressure is equal to the inverse of the correction factor determined from the SRG. Nitrogen, Carbon Dioxide, Methane, Hydrogen and Helium are individually introduced into the vacuum chamber in order to obtain normalized sensitivity plots as a function of pressure. RESULTS LINEARITY OF The results of the relative sensitivity measurements are shown in Figure 1. While all of the RGAs show some variations in relative sensitivity for each gas species, these variations are small for the most part, indicating linear behavior. FRAGMENTATION EFFECTS Fragmentation of mass spectra by dissociation or multiple ionization of molecules complicates pressure measurement for both total pressure gauges and partial pressure gauges such as the RGA. Gauges such as the extractor detect all ions produced; thus a single CH4 molecule, for example, may dissociate into two ions, resulting in an ion current disproportionate to the number of molecules present in the sample. RGAs have an advantage in that they sort these fragmentary ions into different mass channels, allowing the user to deal with them. A general quantitative description of the relation between fragmentary peak heights and pressure is difficult, and this is one reason why RGAs are considered by some to be solely qualitative devices as explained in the SRS RGA documentation. If a constant fragmentary pattern can be assumed, however, the previously discussed calibration procedure can be performed for any desired peak in a particular gas species cracking pattern. The choice of the peak to use requires intuition on the part of the user, and some prior knowledge of the likely gas species present. By correcting the chosen peak based on this procedure and ignoring other peaks from that species, the RGA can provide a more precise measurement of total pressure than gauges without a mass filter. This method also provides the user with a means of dealing with overlapping peaks from different gases. For example, when both CO2 and CH4 are present, the primary peak of CH4 at mass 16 will have a contribution from the secondary peak of CO2; in this case, the problem can be circumvented by using the sensitivity correction for the mass 15 peak for CH4 (which can be assumed to be about 85% of the sensitivity for mass 16, or can be directly measured), then adding the corrected mass 15 and 44 peaks to get the total pressure. The above discussion is meaningful only if the fragmentation pattern remains constant with pressure. Fortunately, for the pressure range measured here at least, this is the case. Figure 2 shows the abundance of the secondary mass 15 peak relative to the primary mass 16 peak for CH4 as a function of pressure. Fragmentation percentage may be altered by adjustment of electron emission current, but should remain constant at least

4 down to 10^-10 Torr. The linearity of ionization total pressure gauges is further evidence of constant fragmentary patterns, as a nonlinear fragmentation would result in a nonlinear relationship between ion current and pressure. SPACE CHARGE EFFECTS AND RGA EMISSION CURRENT Space charge effects affect the RGA in two places: the ionizer and the quadrupole. In the ionizer, the electronic space charge can decrease the sensitivity of the detector by altering the trajectory of ions that otherwise would have entered the detector. This effect has been shown to effectively decrease the ionization volume 2. In the quadrupole, the ionic space charge causes the ion beam to become defocused. This effect depends on the number of ions in the quadrupole, and the time an individual ion spends in the quadrupole. Because higher velocity ions exit the quadrupole sooner, lower mass ions experience less defocusing, resulting in higher sensitivities for lower mass ions. Both the electronic and ionic space charge effects can be reduced by lowering the emission current. This decreases the number of electrons in the ionizer, and the number of ions in the quadrupole. Although lowering the emission current also lowers the sensitivity of the detector, it improves the linearity of the sensitivity, especially at pressures above 10^-7 Torr. The loss of sensitivity due to lowering the emission current can be corrected by recalibrating the gauge from the standard once the emission current has been changed. Figure 3 shows the effect of changing emission current on the sensitivity for Hydrogen; this figure shows that reducing the emission current by a factor of 10 does not necessarily change the sensitivity by the same factor, thus recalibration is necessary. This is because while lowering the emission current decreases the ion production rate, it also increases the sensitivity by lowering space charge effects, and the exact net effect can only be experimentally determined. Operating with a lower emission current can allow an RGA to be used at higher pressures without loss of linearity. MASS FILTER THROUGHPUT RGAs can be viewed as containing a window that is only open to ions of a certain charge to mass ratio at a given time, and this window slides over a range of masses at a certain scanning rate. Because the ions collected are produced to possess a certain uniform energy, lower mass ions will travel through the filter faster than higher mass ions; therefore, more mass 2 ions will make it through the window before it closes to them than will mass 28 ions. In order to maintain a constant resolution as mass increases, RGAs a relationship between the d.c. voltage U and the a.c. max voltage V such that U=KV-Uoffset, where K is a dimensionless constant and Uoffset is an offset voltage. This method of maintaining constant resolution (important for enabling the device to resolve very close mass peaks) has the drawback of increasing mass dependence of sensitivity 3. The degree of this mass dependence is determined by K and Uoffset, which are set by the manufacturing technician when initially tuning the instrument, and so this particular portion of the mass dependence must be determined by measurement, since K and Uoffset are not typically available to the end user for alteration. Table 1 shows the Faraday Cup sensitivities for the three RGAs to Hydrogen and Helium, relative to Nitrogen. A total pressure ion gauge, whose sensitivity is determined primarily by ionization cross section, would have a typical sensitivity around 0.5 for Hydrogen relative to Nitrogen. Table 2 shows that RGAs are capable of having

5 Hydrogen sensitivities exceeding 1 relative to Nitrogen, requiring that mass peak to be actually multiplied rather than divided (as would be the case for a total pressure gauge with no mass spectrometer). CONCLUSIONS AND SUMMARY There are factors other than pressure and emission current that can affect RGA sensitivity, and overall sensitivity changes with time; however, as shown here, it can be demonstrated that RGAs can be calibrated to provide accurate and consistent measurements over a wide pressure range for most experimental time scales. The throughput of the quadrupole is related to its resolution setting and the ion energies used. While these were both kept constant for this experiment, their effect should be noted so that users will check and maintain consistent resolution as sensor age causes the value to drift, and so that users will redetermine calibration values if a resolution change becomes necessary. Additional work would be useful to establish the minimum pressure RGAs are capable of measuring, and the linearity of the gauges as the pressure approaches this minimum. While many RGAs, especially high-end units, are very sensitive (some claiming miminum measureable pressures down to as low as 10^-13 Torr), the total pressure gauges that were used here to determine linearity have difficulty measuring pressures that low. In addition, while the chamber used in this experiment provides very good vacuum (as low as around 10^-11 Torr), specially treated chambers to reach even lower pressures and ion gauges capable of measuring those pressures are needed if a true lower bound of RGA total pressure measurement capability and sensitivity is to be determined. We have seen here that RGAs behave quite linearly into the UHV pressure range for the gases studied, and that the linearity can be improved by adjusting the emission current of the gauge. In typical UHV applications, generally only a few gas species are present; in the ideal case only Hydrogen is present. RGAs can therefore serve as a good total pressure gauge for UHV applications once the sensitivities specific to the gauge and gases present have been found. Residual Gas Analyzers are complex gauges, both in their construction and their use. Accurate measurements require the user to consider many different physical factors involved in the operation of the gauge. FIGURES Figure 1: Sensitivity data for all RGAs and gases studied Generally there is little more than 10% deviation in sensitivity over the pressure range studied. The factory reconditioned RGA () seems to provide the greatest degree of linearity. 1a: Nitrogen:

6 RGA N2 VS EXT READING 1 MA EMIS CURRENT E E E E E-06 EXT PRESSURE 1b: Methane: H4 MASS 16 VS EXTRACTOR READING E E E E E-06 EXT PRESSURE (TORR) 1c: Carbon Dioxide:

7 O2 VS EXTRACTOR READING 1 MA EMIS CURRENT E E E E E-06 EXT PRESSURE 1d: Helium: RGA He VS AXTRAN READING 1 MA EMIS CURRENT E E E E E-06 AXT PRESSURE (TORR) 0 1e: Hydrogen:

8 RGA H2 NORMALIZED VS PRESSURE, 1 MA EMISSION CURRENT E E E E E-06 PRESSURE (TORR) Figure 2: Fragmentation percent of Methane It can be seen here that the fragmentation percentage for Methane remains nearly constant as pressure is decreased.

9 Fragmentation Percent vs. Pressure Fragmentation E E E E E-006 Pressure Figure 3: RGA sensitivity to background H2 as a function of emission current sensitivity can be seen to decrease almost linearly with emission current RGA NORMALIZED H2 VS EMISSION CURRENT 1.20E E E E E E E CURRENT (MA) Table 1: Properties of the SRS 100 AMU RGAs Filament Type ThO 2 /Ir Electron Energy ev Ion Energy 8-12 ev Focus Voltage V Standard 90 V setting used

10 Emission Current ma -- 1 ma used except where noted Table 2: RGA Faraday Cup H2 sensitivities relative to N2 1 ma ma REFERENCES: 1. M.G Rao and C. Dong Evaluation of Low Cost Residual Gas Analyzers for Ultrahigh Vacuum Applications. J. Vac. Sci. Tech. A, May 1997, Vol. 15, Issue 3, pp M.C. Cowen, W. Allison, and J.H. Batey Electron Space Charge Effects in Ion Sources for Residual Gas Analysis. J. Meas. Sci. Tech., January 1993, Vol 4, Issue 1, pp Robert Ellefson Partial Pressure Analysis in Foundations of Vacuum Science and Technology J.M Lafferty, ed. John Wiley & Sons 1998, pp Peter Dawson Quadrupole Mass Spectrometry and its Applications American Institute of Physics P.A. Redhead, J.P. Hobson, and E.V. Kornelson The Physical Basis of Ultrahigh Vacuum American Institute of Physics 1993.

Gun Vacuum Performance Measurements February 2003

Gun Vacuum Performance Measurements February 2003 JLAB-TN-03-013 Gun Vacuum Performance Measurements February 2003 Marcy Stutzman Abstract. The vacuum in the polarized photoguns at Jefferson Lab is crucial to long lifetime operation of the polarized source.

More information

Practical approach and problems in in-situ RGA calibration

Practical approach and problems in in-situ RGA calibration Practical approach and problems in in-situ RGA calibration Oleg Malyshev and Keith Middleman Vacuum Science Group, ASTeC Accelerator Science and Technology Centre STFC Daresbury Laboratory UK Workshop

More information

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION BEST KNOWN METHODS Transpector XPR3 Gas Analysis System DESCRIPTION The Transpector XPR3 is a third-generation, quadrupole-based residual gas analyzer that operates at PVD process pressures and is the

More information

The Helium Leak Detector

The Helium Leak Detector The Helium Leak Detector Helium Leak Detector Main Components The main components of a helium leak detector are: 1. The analyzed, which enables to separate the tracer gas from other gases inside leak detector.

More information

Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples

Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples Outline of Presentation Introduction Process Gas Analysis Magnetic Sector Mass Spectrometer Standard

More information

Leak Checking Large Vacuum Chambers

Leak Checking Large Vacuum Chambers Leak Checking Large Vacuum Chambers Technical Overview Vacuum Technologies Introduction Understanding the pump-down characteristics of a large vacuum vessel is critical for determining whether the vacuum

More information

Residual Gas Analysis Systems for Industry

Residual Gas Analysis Systems for Industry HIDEN RC SYSTEMS QUADRUPOLE MASS SPECTROMETERS FOR RGA, GAS ANALYSIS AND PROCESS MONITORING The HAL RC systems are designed for RGA, gas analysis and process monitoring applications including leak detection,

More information

Choosing a Quadrupole Gas Analyzer Application Note #9

Choosing a Quadrupole Gas Analyzer Application Note #9 Choosing a Quadrupole Gas Analyzer Application Note #9 Modern-day contamination control requirements for gas phase processes are constantly pushing the limits of performance of quadrupole gas analyzers.

More information

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Oneida Research Services, Inc. 8282 Halsey Road Whitesboro, NY 13492 Phone: (315) 736-5480 FAX: (315) 736-9321 1.0 Purpose The

More information

LCLS Heavy Met Outgassing Tests * K. I. Kishiyama, D. M. Behne Lawrence Livermore National Laboratory

LCLS Heavy Met Outgassing Tests * K. I. Kishiyama, D. M. Behne Lawrence Livermore National Laboratory UCRL-TR-223160 LCLS-TN-06-12 July 12, 2006 LCLS Heavy Met Outgassing Tests * K. I. Kishiyama, D. M. Behne Lawrence Livermore National Laboratory Abstract A Heavy Met that is 95% tungsten, 3% nickel and

More information

ISO/TC 112/WG 2 N 49. ISO TS QMS WD v6. Date of document Expected action Comment Due Date Background

ISO/TC 112/WG 2 N 49. ISO TS QMS WD v6. Date of document Expected action Comment Due Date Background ISO/TC 112/WG 2 N 49 ISO/TC 112/WG 2 Vacuum instrumentation E-mail of Secretary: juergen.eisenreich_vdma.org@livelink.din.de Secretariat: DIN ISO TS 20175 QMS WD v6 Date of document 2016-05-02 Expected

More information

This educational seminar discusses creating, measuring, and troubleshooting Rough Vacuum.

This educational seminar discusses creating, measuring, and troubleshooting Rough Vacuum. This educational seminar discusses creating, measuring, and troubleshooting Rough Vacuum. Specifically, today s talk will cover: Brief review of Vacuum Fundamentals Applications Using Rough Vacuum Rough

More information

Resist round robin ELETTRA

Resist round robin ELETTRA Resist round robin outgassing @ ELETTRA Trieste - Italy M. Bertolo, P. Lacovig, L.Rumiz, D. Giuressi, R. Sergo, A. Stolfa Experimental setup (1) 100 mm linear translator Sample manipulator Mask manipulator

More information

Hiden QIC Series Gas Analysers Real-time gas analysis systems

Hiden QIC Series Gas Analysers Real-time gas analysis systems Hiden QIC Series Gas Analysers Real-time gas analysis systems vacuum analysis surface science plasma diagnostics gas analysis Application Focussed Options Multistream valves 20, 40 and 80 way versions

More information

OMNISTAR / THERMOSTAR. The efficient solution for gas analysis. Intelligent software. Wide range of applications.

OMNISTAR / THERMOSTAR. The efficient solution for gas analysis. Intelligent software. Wide range of applications. OMNISTAR / THERMOSTAR The efficient solution for gas analysis. Intelligent software. Wide range of applications. OMNISTAR /THERMOSTAR The efficient solution for gas analysis. Intelligent software. Wide

More information

Quadruple mass spectrometers (transducer type)

Quadruple mass spectrometers (transducer type) Quadruple mass spectrometers that meet the need for sensitive analysis and low outgassing which were specifically designed to be controlled by PCs Quadruple mass spectrometers (transducer type) M-070QA-TDF,

More information

The Principles of Vacuum Technology

The Principles of Vacuum Technology The Principles of Vacuum Technology Vacuum Terminology Vacuum units Vacuum regimes How to measure vacuum. Gauge designs. How to create vacuum Pump classifications and designs UHV compatibility considerations

More information

METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL PRESSURE AND COMPOSITIONAL ANALYSIS

METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL PRESSURE AND COMPOSITIONAL ANALYSIS METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL PRESSURE AND COMPOSITIONAL ANALYSIS Robert E. Ellefson REVac Consulting Dayton, OH 45459 USA robert.ellefson@sbcglobal.net ERMP IND12 QMS Workshop Bled,

More information

The Experts in Vacuum Solutions

The Experts in Vacuum Solutions By Woodrow Farrow - Reprinted with permission from Specialty Gas Report Vacuum: a space that is relatively empty of matter, especially when that void has been created through artificial means. The earth

More information

Leaks, contamination and closed loop control how RGAs make coating processes more profitable

Leaks, contamination and closed loop control how RGAs make coating processes more profitable Leaks, contamination and closed loop control how RGAs make coating processes more profitable Backup: What s an RGA Residual Gas Analyzer An instrument that is measuring the gas composition in a chamber

More information

Vacuum Science Techniques and Applications Dan Dessau Adv. Lab 2007

Vacuum Science Techniques and Applications Dan Dessau Adv. Lab 2007 Vacuum Science Techniques and Applications Dan Dessau Adv. Lab 2007 Vacuum increases the mean-free-path of gas molecules. Vacuum prevents chemical reaction. Vacuum removes contaminants from surfaces. Vacuum

More information

High Performance Mass Spectrometers

High Performance Mass Spectrometers High Performance Mass Spectrometers For Vacuum and Semiconductor Process Monitoring Detailed product information / introduction MOCVD semiconductor processing Applications: XXCVD/MOCVD XXPECVD XXALD XXSputtering

More information

1)! DO NOT PROCEED BEYOND THIS MARK

1)! DO NOT PROCEED BEYOND THIS MARK Operating Instructions for X-ray Photoelectron Spectrometer: Physical Electronics Model 555 XPS/AES (John H. Thomas, III, Ph.D., Electron Spectroscopy) Sample Insertion: figure 1. Sample insertion rod

More information

VS Series Leak Detectors for Vacuum Furnaces LEAK DETECTION FOR VACUUM FURNACES

VS Series Leak Detectors for Vacuum Furnaces LEAK DETECTION FOR VACUUM FURNACES VS Series Leak Detectors for Vacuum Furnaces LEAK DETECTION FOR VACUUM FURNACES VS Series Leak Detectors for Vacuum Furnaces indicate the presence of a leak. If the vacuum level is slow to reach the original

More information

Detecting Hydrogen in Helium Streams. Samuel Goodman. Pittsford Mendon High School. Rochester, New York. Advisor: Dr.

Detecting Hydrogen in Helium Streams. Samuel Goodman. Pittsford Mendon High School. Rochester, New York. Advisor: Dr. 1 Detecting Hydrogen in Helium Streams Samuel Goodman Pittsford Mendon High School Rochester, New York Advisor: Dr. Walter Shmayda Laboratory for Laser Energetics University of Rochester Rochester, New

More information

Technical Specifications of Hydrogen Isotope Handling and Recovery System

Technical Specifications of Hydrogen Isotope Handling and Recovery System SECTION - C TECHNICAL SPECIFICATIONS OF STORES AND DRAWINGS. Technical Specifications of Hydrogen Isotope Handling and Recovery System INSTITUTE FOR PLASMA RESEARCH GANDHINAGAR, GUJARAT 382428 ANNEXURE-I

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310 www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Varian 310 310 LC/MS Manual System INCLUDES: 310-MS LC/MS triple quadrople

More information

Go/no-go test for the acceptance of components for unbaked vacuum sectors

Go/no-go test for the acceptance of components for unbaked vacuum sectors Go/no-go test for the acceptance of components for unbaked vacuum sectors Paolo Chiggiato (Thanks to Sophie Meunier for the preparation of the presentation) Introduction CERN policy: All vacuum components

More information

Low-Power Atmospheric Gas Sampling System Based on ART MS Sensor and NEG-Ion Pump

Low-Power Atmospheric Gas Sampling System Based on ART MS Sensor and NEG-Ion Pump 8 th Harsh Environment Mass Spectrometry Wokshop Low-Power Atmospheric Gas Sampling System Based on ART MS Sensor and NEG-Ion Pump Sept 22, 2011 Gerardo A. Brucker and G. Jeffery Rathbone Granville-Phillips

More information

AGILENT TURBOMOLECULAR PUMPS MEASUREMENT METHODOLOGY

AGILENT TURBOMOLECULAR PUMPS MEASUREMENT METHODOLOGY AGILENT TURBOMOLECULAR PUMPS MEASUREMENT METHODOLOGY Agilent Vacuum Products: Our Committment to Quality Agilent s commitment to quality and Measurement s philosophy Reference Standards (PNEUROP, ISO)

More information

Vacuum. Methods for in situ QMS calibration for partial pressure and composition analysis. Robert E. Ellefson * abstract

Vacuum. Methods for in situ QMS calibration for partial pressure and composition analysis. Robert E. Ellefson * abstract Vacuum 101 (2014) 42e42 Contents lists available at ScienceDirect Vacuum journal homepage: www.elsevier.com/locate/vacuum Methods for in situ QMS calibration for partial pressure and composition analysis

More information

at NIST: ultra-low outgassing rates

at NIST: ultra-low outgassing rates Traceable outgassing measurements at NIST: ultra-low outgassing rates and new materials James Fedchak Thermodynamic Metrology Group Sensor Science Division Physical Measurement Laboratory Vacuum Project:

More information

EuroFID Total Hydrocarbon Analyzer. Precise Determination of Total Hydrocarbons in Air for Corrosive as well as Condensing Gases

EuroFID Total Hydrocarbon Analyzer. Precise Determination of Total Hydrocarbons in Air for Corrosive as well as Condensing Gases P r o d u c t i n f o r m at i o n EuroFID Total Hydrocarbon Analyzer Precise Determination of Total Hydrocarbons in Air for Corrosive as well as Condensing Gases Proven Analyzer Technology Measuring total

More information

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies IMPROVED CALIBRATION AND OPERATION OF A STABLE-GAS QUANTIFICATION MANIFOLD AS PART OF A RADIOXENON COLLECTION SYSTEM ABSTRACT Richard M. Williams, James C. Hayes, and Randy R. Kirkham Pacific Northwest

More information

Cover Page for Lab Report Group Portion. Pump Performance

Cover Page for Lab Report Group Portion. Pump Performance Cover Page for Lab Report Group Portion Pump Performance Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 March 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section number:

More information

Major Design Features

Major Design Features Major Design Features 3Flex Features Design Benefits Discussion 316 SS fittings and pneumatically actuated, hard-seal valves Provides virtually leak-free gas management with the lowest outgassing rate

More information

Improving the Measurement Accuracy of Water Partial Pressure Using the Major Constituent Analyzer

Improving the Measurement Accuracy of Water Partial Pressure Using the Major Constituent Analyzer Improving the Measurement Accuracy of Water Partial Pressure Using the Major Constituent Analyzer Ben D. Gardner, Phillip M. Erwin, Wai Tak Lee, Amber M. Tissandier, Souzan M. Thoresen Hamilton Sundstrand

More information

Mass Spec will not Autotune

Mass Spec will not Autotune Mass Spec will not Autotune Applies to 5973A/N MSD What could be the problem? There could be several things that would cause your Mass Spec not to Autotune. The most common, easily corrected Autotune problems

More information

Measuring mass flows in hermetically sealed MEMs & MOEMs to ensure device reliability

Measuring mass flows in hermetically sealed MEMs & MOEMs to ensure device reliability Measuring mass flows in hermetically sealed MEMs & MOEMs to ensure device reliability R. C. Kullberg*, D. J. Rossiter** Oneida Research Services, Inc., 8282 Halsey Road, Whitesboro, NY13492 Abstract Many

More information

The Broadest Gauging Line in the Industry

The Broadest Gauging Line in the Industry The Broadest Gauging Line in the Industry Vacuum Measurement Varian s broad line of vacuum gauge controllers and gauge tubes are the most reliable, accurate, and economical means of measuring, monitoring,

More information

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW AC 2010-2145: MEASUREMENT OF HYDROGEN IN HELIUM FLOW Randy Buchanan, University of Southern Mississippi Christopher Winstead, University of Southern Mississippi Anton Netchaev, University of Southern Mississippi

More information

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER Michael T Tang, Ph.D. Grace Feng Greg Merideth Rui Huang Matheson Gas Applied Lab

More information

Helium Mass Spectrometric Leak Detection In Large Size Process Plants

Helium Mass Spectrometric Leak Detection In Large Size Process Plants National Seminar & Exhibition on Non-Destructive Evaluation, NDE 2014, Pune, December 4-6, 2014 (NDE-India 2014) Vol.20 No.6 (June 2015) - The e-journal of Nondestructive Testing - ISSN 1435-4934 www.ndt.net/?id=17898

More information

Real-time Analysis of Industrial Gases with Online Near- Infrared Spectroscopy

Real-time Analysis of Industrial Gases with Online Near- Infrared Spectroscopy Real-time Analysis of Industrial Gases with Online Near- Infrared Branch 5 Petrochemical, Biofuels Keywords Process, NIR, Gases, Gas Cell, Ethylene, Carbon Dioxide, Propane, Butane, Acetylene Summary Near-infrared

More information

A NOVEL SENSOR USING REMOTE PLASMA EMISSION SPECTROSCOPY FOR MONITORING AND CONTROL OF VACUUM WEB COATING PROCESSES

A NOVEL SENSOR USING REMOTE PLASMA EMISSION SPECTROSCOPY FOR MONITORING AND CONTROL OF VACUUM WEB COATING PROCESSES A NOVEL SENSOR USING REMOTE PLASMA EMISSION SPECTROSCOPY FOR MONITORING AND CONTROL OF VACUUM WEB COATING PROCESSES F. Papa 1, J. Brindley 2, T. Williams 2, B. Daniel 2, V. Bellido-Gonzalez 2, Dermot Monaghan

More information

Vacuum Simulations of the KATRIN Experiment

Vacuum Simulations of the KATRIN Experiment Vacuum Simulations of the KATRIN Experiment Marcel Krause (KATRIN collaboration) 31.03.2014 KIT University of the State of Baden-Wuerttemberg and National Research Center of the Helmholtz Association www.kit.edu

More information

Revision 2013 Vacuum Technology 1-3 day Good Vacuum Practice 1 Day Course Outline

Revision 2013 Vacuum Technology 1-3 day Good Vacuum Practice 1 Day Course Outline Revision 2013 Vacuum Technology 1-3 day Good Vacuum Practice 1 Day Course Outline This training course outline is intended to cover the following: Introduction to vacuum Measurement Lubricated rotary pumps

More information

Procedure 1: Volume vs. Pressure 1.) Using the lap tops, go to the Physics Education Technology from the University of Colorado at:

Procedure 1: Volume vs. Pressure 1.) Using the lap tops, go to the Physics Education Technology from the University of Colorado at: Deriving the Gas Laws Background The gaseous state of matter consists of particles (gas molecules like oxygen, nitrogen, and carbon dioxide) which, according to the kinetic theory of gases, are in constant

More information

Introduction of Vacuum Science & Technology. Diffusion pumps used on the Calutron mass spectrometers during the Manhattan Project.

Introduction of Vacuum Science & Technology. Diffusion pumps used on the Calutron mass spectrometers during the Manhattan Project. Introduction of Vacuum Science & Technology Diffusion pumps used on the Calutron mass spectrometers during the Manhattan Project. 1 What is a vacuum? 760 mm Hg Vacuum ATM A vacuum is defined as less than

More information

Canadian Light Source Inc. Vacuum Component Leak Test Technical Procedure

Canadian Light Source Inc. Vacuum Component Leak Test Technical Procedure Canadian Light Source Inc. Vacuum Component Leak Test Technical Procedure CLSI Document 8.7.33.2 Rev. 2 Date: 2006-03-08 Copyright 2006, Canadian Light Source Inc. This document is the property of Canadian

More information

INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE

INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE TELEDYNE HASTINGS TECHNICAL PAPERS INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE Proceedings of FEDSM 98 1998 ASME Fluids Engineering

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results. Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions

More information

LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12

LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12 LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/ This experiment will introduce you to the kinetic properties of low-pressure gases. You will make observations on the rates with which selected

More information

Matching Vacuum Pump to Process

Matching Vacuum Pump to Process A Journal of Practical and Useful Vacuum Technology From By Phil Danielson Matching Vacuum Pump to Process What s the best vacuum pump? is one of those maddening questions that can only be answered with

More information

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial Supporting Information for Micro-Collection of Gases in a Capillary Tube: Preservation of Spatial and Temporal Resolution AUTHOR INFORMATION Kristin A. Herrmann Favela, 1 * Pieter Tans, 2 Thomas Jaeckle,

More information

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator Procedure No: SOP-034 Revision No: 1.0 Revised Dec. 29, 2010 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE This procedure is intended to describe the operations of the Dasibi model 5008 calibrator. The Dasibi

More information

LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary

LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary ADH 1/7/014 LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary This experiment will introduce you to the kinetic properties of low-pressure gases. You will make observations on the

More information

Chapter 13 Temperature, Kinetic Theory, and the Gas Laws 497

Chapter 13 Temperature, Kinetic Theory, and the Gas Laws 497 Chapter 13 Temperature, Kinetic Theory, and the Gas Laws 497 Figure 13.25 This photograph of Apollo 17 Commander Eugene Cernan driving the lunar rover on the Moon in 1972 looks as though it was taken at

More information

Recent steps toward traceability for partial pressure and outgassing measurements Karl Jousten, PTB, Berlin

Recent steps toward traceability for partial pressure and outgassing measurements Karl Jousten, PTB, Berlin Recent steps toward traceability for partial pressure and outgassing measurements Karl Jousten, PTB, Berlin 1. Problems of calibrating QMS 3. Toward traceability in EMRP IND12 2. Toward standardization

More information

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES 1. Applicability and Principle 1.1 Applicability. This method applies to the analysis of carbon dioxide

More information

Pressure & Vacuum Measurement. Series 370. Solutions. Stabil-Ion Vacuum gauge and Controller

Pressure & Vacuum Measurement. Series 370. Solutions. Stabil-Ion Vacuum gauge and Controller w w w. m k s i n s t. c o m Pressure & Vacuum Measurement Solutions Series 370 Stabil-Ion Vacuum gauge and Controller Features & Benefits All-metal, rack-mount controller for Stabil-Ion and Convectron

More information

FUGITIVE EMISSIONS EXPERIMENTAL MEASUREMENTS AND EQUIVALENCY

FUGITIVE EMISSIONS EXPERIMENTAL MEASUREMENTS AND EQUIVALENCY FUGITIVE EMISSIONS EXPERIMENTAL MEASUREMENTS AND EQUIVALENCY ABSTRACT Jay Abrahimzadeh Joe Steinke Control Components Inc. Fugitive Emissions requirements, compliance, and use have gained momentum in the

More information

Equation 1: F spring = kx. Where F is the force of the spring, k is the spring constant and x is the displacement of the spring. Equation 2: F = mg

Equation 1: F spring = kx. Where F is the force of the spring, k is the spring constant and x is the displacement of the spring. Equation 2: F = mg 1 Introduction Relationship between Spring Constant and Length of Bungee Cord In this experiment, we aimed to model the behavior of the bungee cord that will be used in the Bungee Challenge. Specifically,

More information

Introductory Lab: Vacuum Methods

Introductory Lab: Vacuum Methods Introductory Lab: Vacuum Methods Experiments in Modern Physics (P451) In this lab you will become familiar with the various components of the lab vacuum system. There are many books on this topic one of

More information

Extreme High Vacuum: The Need, Production, and Measurement

Extreme High Vacuum: The Need, Production, and Measurement Extreme High Vacuum: The Need, Production, and Measurement, Philip Adderley, Matt Poelker (Jefferson Lab) Polarized Electron Gun Group Newport News, Virginia Run by JSA for the US DOE What is XHV Extreme

More information

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs. Laboratory Hardware Custom Gas Chromatography Solutions Custom solutions for your analytical needs. Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions for advanced chromatography

More information

Application Note AN-107

Application Note AN-107 SPEC Sensor TM Characterization & Calibration Considerations Scope This document is provided to describe the considerations needed to characterize, calibrate, verify and validate the measurement performance

More information

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE. Gases&Technology FEATURE Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. B Y M A T T H E W M O N A G L E Abstract Bulk gases are often delivered to

More information

EFFECTS OF BAFFLE SIZE ON PRESSURE DISTRIBUTION IN VACUUM CHAMBER DURING DYNAMIC GAS FLOW

EFFECTS OF BAFFLE SIZE ON PRESSURE DISTRIBUTION IN VACUUM CHAMBER DURING DYNAMIC GAS FLOW XIX IMEKO World Congress Fundamental and Applied Metrology September 6 11, 2009, Lisbon, Portugal EFFECTS OF BAFFLE SIZE ON PRESSURE DISTRIBUTION IN VACUUM CHAMBER DURING DYNAMIC GAS FLOW Wakil Khan 1,

More information

DF-310E PROCESS ANALYSERS APPLICATIONS FEATURES

DF-310E PROCESS ANALYSERS APPLICATIONS FEATURES PROCESS ANALYSERS DF-310E The DF-310E is a Coulometric sensor based oxygen analyzer designed to measure trace and percent level oxygen in pure and multi-gas backgrounds for process and quality control

More information

Advantages of Carrier Gas Leak Detection using Novel Helium or Hydrogen Leak Detectors with Specific Sensor Types

Advantages of Carrier Gas Leak Detection using Novel Helium or Hydrogen Leak Detectors with Specific Sensor Types 19 th World Conference on Non-Destructive Testing 2016 Advantages of Carrier Gas Leak Detection using Novel Helium or Hydrogen Leak Detectors with Specific Sensor Types Klaus HERRMANN 1, Daniel WETZIG

More information

The instrument should have a mass range from 2 to 2000 m/z or better Interface

The instrument should have a mass range from 2 to 2000 m/z or better Interface Revised Specifications for Liquid Chromatography Mass Spectrometry (LC-MS/MS ) Triple Quadrupole Specification for Ultra-Fast Liquid Chromatography system Solvent Delivery 1) A quaternary Pump system for

More information

Installation and Operation Manual

Installation and Operation Manual Manual Static pressure transducer with controller Differential static pressure transducer with analog output and optional PI control mode Large diaphragm element with differential transformer Transducer

More information

Phase Changes * OpenStax

Phase Changes * OpenStax OpenStax-CNX module: m42218 1 Phase Changes * OpenStax This work is produced by OpenStax-CNX and licensed under the Creative Commons Attribution License 3.0 Abstract Interpret a phase diagram. State Dalton's

More information

835 VQM. Features & Benefits

835 VQM. Features & Benefits Pressure & Vacuum Measurement Solutions w w w. m k s i n s t. c o m 835 VQM VACuum Quality Monitor System The Vacuum Quality Monitor (VQM) combines the highest performance gas analysis technology with

More information

Gas Pressure. Pressure is the force exerted per unit area by gas molecules as they strike the surfaces around them.

Gas Pressure. Pressure is the force exerted per unit area by gas molecules as they strike the surfaces around them. Chapter 5 Gases Gas Gases are composed of particles that are moving around very fast in their container(s). These particles moves in straight lines until they collides with either the container wall or

More information

Lab #1 Pressure: Bubblers and Water Balloons CEE 331 Fall 2003

Lab #1 Pressure: Bubblers and Water Balloons CEE 331 Fall 2003 CEE 331 Lab 1 Page 1 of 9 SAFETY Lab #1 Pressure: Bubblers and Water Balloons CEE 331 Fall 2003 Laboratory exercise based on an exercise developed by Dr. Monroe Weber-Shirk The major safety hazard in this

More information

Development of a high vacuum sample preparation system for helium mass spectrometer

Development of a high vacuum sample preparation system for helium mass spectrometer Journal of Physics: Conference Series Development of a high vacuum sample preparation system for helium mass spectrometer To cite this article: P Kumar et al 2012 J. Phys.: Conf. Ser. 390 012056 View the

More information

Another convenient term is gauge pressure, which is a pressure measured above barometric pressure.

Another convenient term is gauge pressure, which is a pressure measured above barometric pressure. VACUUM Theory and Applications Vacuum may be defined as the complete emptiness of a given volume. It is impossible to obtain a perfect vacuum, but it is possible to obtain a level of vacuum, defined as

More information

(2 pts) Draw the line of best fit through the data and estimate the concentration of Fe in your sample solution.

(2 pts) Draw the line of best fit through the data and estimate the concentration of Fe in your sample solution. Name: 1. You are asked to measure the percent iron in a steel sample you believe to be about 50% Fe by mass. You dissolve four replicate samples in nitric acid and dilute them by mass to a concentration

More information

Retention Time Locking: Concepts and Applications. Application

Retention Time Locking: Concepts and Applications. Application Retention Time Locking: Concepts and Applications Application Gas Chromatography December 1997 Authors Vince Giarrocco Bruce Quimby Matthew Klee Agilent Technologies, Inc. 2850 Centerville Road Wilmington,

More information

Basic Nitriding Sampling System Hydrogen Analyzer with Calculated % DA, % NH 3, and K N Values. Operations Manual

Basic Nitriding Sampling System Hydrogen Analyzer with Calculated % DA, % NH 3, and K N Values. Operations Manual Basic Nitriding Sampling System Hydrogen Analyzer with Calculated % DA, % NH 3, and K N Values Operations Manual Please read, understand, and follow these instructions before operating this equipment.

More information

A Journal of Practical and Useful Vacuum Technology. By Phil Danielson

A Journal of Practical and Useful Vacuum Technology. By Phil Danielson A Journal of Practical and Useful Vacuum Technology From By Phil Danielson Thermal Conductivity Gauges Thermal conductivity pressure gauges are extremely common in vacuum technology, but an understanding

More information

TOTAL AND PARTIAL PRESSURE MEASUREMENT IN VACUUM SYSTEMS

TOTAL AND PARTIAL PRESSURE MEASUREMENT IN VACUUM SYSTEMS TOTAL AND PARTIAL PRESSURE MEASUREMENT IN VACUUM SYSTEMS J.H. LEeK Emeritus Professor Department of Electrical Engineering and Electronics University of Liverpool Blackie Glasgow and London B1ackie & Son

More information

Outgas of Methane from NEG Coating

Outgas of Methane from NEG Coating WS63: SURFACE PHENOMENA LIMITING PRESSURE IN VACUUM SYSTEM Outgas of Methane from NEG Coating Chia-Mu Cheng ( 鄭家沐 ) Vacuum Group NSRRC, Taiwan 18 /09/2010 e-mail : chiamucheng@nsrrc.org.tw 1 Application

More information

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1 Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost HERMAN HOLM SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 4C CONFERENCE APRIL 2018 SAN ANTONIO, TX SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC

More information

GP1 & GP2. Electropneumatic Regulators FOR PRESSURE CONTROL TO 1,000 PSI

GP1 & GP2. Electropneumatic Regulators FOR PRESSURE CONTROL TO 1,000 PSI GP1 & GP2 Electropneumatic Regulators FOR PRESSURE CONTROL TO 1, PSI GP1 & GP2 Functional Description The GP series control valve is an electronic pressure regulator designed to precisely control the pressure

More information

Chapter 6 Leak detection

Chapter 6 Leak detection Chapter 6 Leak detection 1 2 1. Leak rate q L = V p t mbar l s -1 A leak rate of 1 mbar l s -1 exists in a closed vessel having a volume of 1 liter when the pressure increases by 1 mbar within one second,

More information

ACV-10 Automatic Control Valve

ACV-10 Automatic Control Valve ACV-10 Automatic Control Valve Installation, Operation & Maintenance General: The Archer Instruments ACV-10 is a precision automatic feed rate control valve for use in vacuum systems feeding Chlorine,

More information

Multiple Gas#5 GC configuration Jan 2016

Multiple Gas#5 GC configuration Jan 2016 History: Unfortunately there is no single column that can separate: Hydrogen Oxygen Nitrogen Methane CO CO2 Ethane Water Propane Butane Pentane Over the years SRI Instruments has devised several solutions

More information

Extend the time window for the detection of low level anabolic androgenic steroids and their metabolites

Extend the time window for the detection of low level anabolic androgenic steroids and their metabolites Application Note Doping Control Extend the time window for the detection of low level anabolic androgenic steroids and their metabolites Steroid profiling, screening, and confirmatory analysis using the

More information

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet: OXY Integral INTERCON ENTERPRISES INC Tel: 800 665 6655 Fax: 604 946 5340 E-Mail: sales@intercononline.com Internet: www.intercononline.com Manual Integral 2006 1 INDEX 2-3 PREFACE 4 INTRODUCTION 5 Principle

More information

RADIATION PROCEDURES MANUAL Procedure Cover Sheet

RADIATION PROCEDURES MANUAL Procedure Cover Sheet RADIATION PROCEDURES MANUAL Procedure Cover Sheet Procedure Title: Calibration of Eberline Portable Particulate Noble Gas Monitor Procedure Number: TSO-08-13-REV 1 Effective Date: July 1, 2008 Approved

More information

PHOENIX L300i Best-in-class Helium Leak Detectors

PHOENIX L300i Best-in-class Helium Leak Detectors 180.85.02 Best-in-class Helium Leak Detectors The perfect Helium leak detector matching your demands The helium leak detector opens up new dimensions of productivity and reliability for your applications.

More information

Lab 1. Adiabatic and reversible compression of a gas

Lab 1. Adiabatic and reversible compression of a gas Lab 1. Adiabatic and reversible compression of a gas Introduction The initial and final states of an adiabatic and reversible volume change of an ideal gas can be determined by the First Law of Thermodynamics

More information

DISCLAIMER. Government. Neither the United States Government nor any agency thereof, nor any of their

DISCLAIMER. Government. Neither the United States Government nor any agency thereof, nor any of their S-76,491 03 r3 0 m 0 X u a r3 I (v 0 2 I W CI +,.rl k 3. W E c 6) X A MEANS FOR POSITIVELY SEATING A.PIEZOCERAMIC ELEMENT IN A PIEZOELECTRIC VALVE DURING INLET GAS INJECTION Kenneth E. Wright DISCLAIMER

More information

BAPI Pressure Line of Products - FAQs

BAPI Pressure Line of Products - FAQs Table of Contents 1. Several manufacturers produce pressure transmitters, why should I purchase from BAPI?... p. 2 2. BAPI makes several styles of pressure transmitters. What are the features of each?...

More information

Smart Water Application Technologies (SWAT)

Smart Water Application Technologies (SWAT) Smart Water Application Technologies (SWAT) Turf and Landscape Irrigation Equipment PRESSURE REGULATING SPRAY HEAD SPRINKLERS Equipment Functionality Test Testing Protocol Version 3.0 (May 2012) Developed

More information

AUTOMATIC FLOW CARTRIDGES. Competition Analysis Not All Automatics Are Created Equal

AUTOMATIC FLOW CARTRIDGES. Competition Analysis Not All Automatics Are Created Equal - AUTOMATIC FLOW CARTRIDGES Competition Analysis Not All Automatics Are Created Equal IMI FLOW DESIGN / AUTOMATIC FLOW CARTRIDGES / F345.0 COMPETITION ANALYSIS Improve the of your Premium System Components...

More information

White Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology:

White Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology: Title: Comparison of Chemical Sensor and NDIR Technologies TSN Number: 25 File:\\MII- SRV1\Metron\Bridge_Analyzers\Customer_Service_Documentation\White_Papers\25_G en EGA NDIR vs Chemical Sensor.docx Created

More information