Dissolution Failure Investigation. Ken Boda Dissolution Applications Engineer

Size: px
Start display at page:

Download "Dissolution Failure Investigation. Ken Boda Dissolution Applications Engineer"

Transcription

1 Dissolution Failure Investigation Ken Boda Dissolution Applications Engineer

2 Dissolution Failure Investigation Whenever a dissolution test fails to meet specifications, an investigation should be performed to determine the cause(s) of the failure. Failure can be due to: Man (analyst) Machine (dissolution apparatus) Method (SOPs) Materials (standards, buffers, dosage form, etc.)

3 Conducting an Investigation Investigations should be performed in a sequential process to ensure all potential sources of error are investigated Review data as a whole Perform investigation in reverse chronological order

4 Aerial View of the Data The Error itself can help narrow the focus of the investigation: Entire data set trending high/low 1 outlier High %CV Data which makes no sense

5 Data trending high/low Bad lot Calculation Error Standard Prep Error Improper Filtration System misalignment (overall) Poor evaporation control

6 1 Outlier Misalignment/Bad Component in Single Position Filter fell off Bad sample reading Transcription Error Bad individual dosage form Contamination Wrong volume poured

7 High % CV Sampling Inconsistencies Poor techniques Vibration Improper Degassing System Misalignment (tends to trend) Issues w/ formulation

8 Data that doesn t make sense Filtration Issues Bad Sample Readings Air bubble Contamination

9 Determinant vs. Non-Determinant Errors Determinant Errors are known and controllable Indeterminate errors are unknown, suspected, or beyond control Determinant errors when found will lead to a simpler investigation and require the least re-work Indeterminate errors cannot be proven, and the burden of retesting is often higher

10 Investigating in Reverse Last thing done, first investigated Limits scope of investigation, if determinant error found can stop Last things done can often allow salvaging of the run All samples/standards/etc. should be kept until results have been checked and approved or fully investigated

11 Investigation Track Calculations Analysis Sampling and Filtration Run Observations Media Prep and Sample Handling Pre-Run Checks Dissolution Unit Mechanicals SOP and Materials

12 Checking Calculation Errors Check for transcription errors Are the right values being used for: Standard Conc. Label Claim Dissolution Volume Verify calculations by hand Ensure proper calculations are being used

13 Calculations without Media Replacement T1 % Dissolved = ((((A1 x V) /As) x C) / W) x 100 T2 % Dissolved = (((((A2 x (V - R)) + (A1 x R)) /As) x C) / W) x 100 T3 % Dissolved = (((((A3 x (V (R x n))) + ((A1 + A2) x R)) /As) x C) / W) x 100 Where, A1 = the absorbance of the first sample pull A2 = the absorbance of the second sample pull An = the absorbance of the nth sample pull As = the average absorbance of the standard V = the initial media volume, ml C = the concentration of the standard, mg/ml W = the weight of the active, mg (label claim) R = the volume of media being withdrawn for a single sample pull, ml n = the number of the sample pull taken

14 Calculations with Media Replacement T1 % Dissolved = ((((A1 x V) /As) x C) / W) x 100 T2 % Dissolved = (((((A2 x V) + (A1 x R)) /As) x C) / W) x 100 T3 % Dissolved = (((((A3 x V) + ((A1 + A2) x R)) /As) x C) / W) x 100 Where, A1 = the absorbance of the first sample pull A2 = the absorbance of the second sample pull An = the absorbance of the nth sample pull As = the average absorbance of the standard V = the initial media volume, ml C = the concentration of the standard, mg/ml W = the weight of the active, mg (label claim) R = the volume of media being withdrawn for a single sample pull, ml n = the number of the sample pull taken

15 Check Original UV/HPLC readings Right Wavelength and/or LC conditions used? If sample/standard readings were on UV, check scan if available If sample/standard readings were on HPLC, review chromatogram for: Proper peak shape Non-parent peaks Contaminants

16 Reanalyze Samples/Standards If available, reanalyze samples in question Reanalyze standard Compare standard against a fresh standard to ensure original standard was prepared correctly Make sure dilutions and purity are properly factored Dilution error possible?

17 Standard Preparation Create new standard to compare against one used for run Ensure adjustments are made for moisture and purity Right pipet/flask used? Within expiration period?

18 Storage Were samples within expiration period? Covered? Stored Properly? Labeled?

19 Sampling and Filtration Sampling and Filtration are one of the most common causes of dissolution failures: USP Position USP Timing Filtration

20 USP Sampling Area Sampling takes practice and attention! All samples must be withdrawn: ½ way between top of paddle/basket and media No closer than 1cm from vessel wall Recommend not sampling within 1cm of spindle

21 Imagine A Donut

22 Sample Timing Must sample (and filter) within 2% of the timepoint or 15 minutes, whichever is less Paddles/baskets must still be stirring while sampling

23 Filtration Is the filter validated? Efficiency Leaching Adsorption When was the sample filtered? At time of sampling? Immediately after sampling? After time point collected?

24 Autosampling Autosampling can reduce many of the errors associated with manual sampling: Position constant Timing within 2% Consistent pull force Documentation

25 Dissolution Observations

26 Observations Observations of the dosage form are key in order to diagnose potential issues with the dosage form. Observations can tell why something happened, % dissolved issues can only tell you what happened.

27 % Dissolved When should you take observations? Ideally, observations should be taken at all critical timepoints of the run Wetting/Dose Dumping Brand to Brand Differences in Dissolution Profiles Release Profile Complete Release/Asymptote Whenever sampling manually Brand A Brand B Brand C Time (minutes)

28 What to look for Initial Stages: Floating or Moving? Coning? Dispersion Profile? Dissolved Gasses? Cross-linking? Compare appearance of all dosage forms

29 Floating/Moving Dosage Forms If a dosage form is floating or moving, it adds additional sources of energy to the dosage form and increases variability Sinkers should be used when floating, spinning, dancing, etc. occurs Use validated sinker

30 Coning Cones should be loose Cone should be in consistent location (bottom center) Look for consistent behavior across vessels, any trending noticed? In cases of severe coning, method alteration may be needed Higher RPM Peak vessel

31 Dispersion Profile and Dissolved Gasses Are air bubbles on stirring element, vessel, dosage form? How are particles distributed in vessel (top, middle, and bottom) Are particles floating? Presence of air bubbles may indicate degassing method is not sufficient

32 Cross-linking Cross-linking of capsule shells can result in hardened and chemically resistant shells. Very common to see in stability testing. Delay opening Trap Drug Product Pellicle Formation If Cross-Linking is seen, testing with pepsin or pancreatin should be performed

33 Compare Results of all 6-8 Vessels Take observations as a complete set, just like data Write observations down Is there any atypical behavior? Trending of observations?

34 Observations During Release Profile Size of dosage form/particles Distribution of dosage form Consistency of the dosage forms Anything changed?

35 Observations at End of Run Any material left? Describe material May be beneficial to assay dosage form post-mortem in formulation development or investigation of strange data or observations

36 Media Preparation Degassing Procedure validated? Time of degassing vs. time of use Pouring Technique

37 Sample Handling When exposed to air? Weighed? Inspected? Where placed? How handled? Bare hands Gloves Tweezers

38 Instrument Checks Pre-run checks of the instrument are always recommended. Pre-run checks are actually required if following the FDA or ASTM Mechanical Qualification procedure. Recheck these same parameters in the event of a run failure

39 Dissolution Mechanical Qualification Standards Prior to each dissolution test, the guidelines generally agree, requiring documentation for: Vessel examination Paddle examination Baskets and basket shaft examination Vessel medium temperature measurement Vibration evaluation Water bath (USP) Speed measurement on single position (USP)

40 Dissolution Mechanical Qualification Standards Vessel Examination, must be free from: Scratches Cracks Pits Residue Surface irregularities Centering Device, should also be: Complete Tight to Vessel Plate

41 Vessel Examination

42 Dissolution Mechanical Qualification Standards Paddle Examination, must be free of: Defects Rusting Corrosion Peeling or loose coating Knicks, dents or misshapen appearance

43 Stainless Steel Paddles Minor Rusting

44 Stainless Steel Paddles Advanced Corrosion

45 Teflon Paddles Good, Okay, and Terrible

46 Dissolution Mechanical Qualification Standards Basket and Basket Shaft, must be free from: Defects Rusting Corrosion Loose wires Clogged mesh openings Dented sides or bottom Knicks, dents or misshapen appearance O-ring are not compliant; three clips required

47 Baskets From New to Dead

48 Basket Bad Example

49 Dissolution Mechanical Qualification Standards Other Requirements: Vessel Temperature: The temperature of the medium inside each vessel is measured at the time of use. Media temperature must be ±0.5ºC from target. Vibration: There can be no significant vibration in the dissolution apparatus or medium. Possible sources of vibration are the surrounding environment, the dissolution unit itself or one of its components or an external water bath circulating heater.

50 Perform Mechanical Checks on System If no issues w/ individual components is noted, investigation of the system s overall alignment and performance should be reviewed MQ tolerances, particularly height and centering can be susceptible to change If height/centering is out, can have major impacts on hydrodynamics

51 Dissolution Mechanical Qualification Specifications and Tolerances Parameter Basket and Paddle Depth Rotational Speed Shaft Wobble Shaft Verticality ICH Harmonized (USP, JP, EP) FDA DPA-LOP.002 ASTM E ± 2 mm 25 ± 2 mm 25 ± 2 mm (or <8%) ± 4% of specified rate No significant wobble Not measured Basket Wobble ± 1 mm ± 2 rpm of target 1.0 mm total runout 0.5 from vertical (x and y 90º) 1.0 mm total runout lower rim ± 2 rpm or within 2% of target (larger) 1.0 mm total runout Within Bubble (x and y 90º) 1.0 mm total runout lower rim USP Toolkit Ver mm ± 1 rpm of target 1.0 mm total wobble Not measured 1.0 mm total wobble

52 Dissolution Mechanical Qualification Specifications and Tolerances Parameter Vessel/Shaft Centering Vessel Verticality Vessel Plate Level Performance Verification Test (PVT) ICH Harmonized (USP, JP, EP) NMT 2 mm from center axis Not Measured FDA DPA-LOP mm from center line (upper/lower) 1.0 from vertical (2-90 positions) ASTM E mm from center line (upper/lower) 1.0 from vertical (2-90 positions) USP Toolkit Ver 2.0 NMT 2.0 mm difference (4-90 positions) NMT 0.5 from vertical Not Measured Not Measured Not Measured NMT 0.5 from horizontal USP Prednisone Tablets RS Not Measured Not Measured USP Prednisone Tablets RS

53 SOP investigation In the event a failure can not be diagnosed, and several failures have occurred, SOP/validation review may be needed: Filter choice Chemicals specified Is the method over-discriminatory? Was intermediate precision evaluated for: Chemist to chemist variability Manual vs. autosampling Resident vs. non-resident probes 3-clip vs. O-ring baskets?

54 SOP Investigation Carryover? Cleaning method established? Dissolution apparatus Sampling equipment

55 Investigation Conclusion In the event of a determinant error, failing run can be aborted and test re-run In the event of a non-determinant error Multiple runs may be needed Analyst re-training may be advisable

56 Special Thanks/Reference Dissolution Aberrant Data Investigation, Bryan Crist, Pharmaceutical Canada, March 2004, vol. 4, no. 4., pp. 7-12

57 Questions? Dissolution Discussion Group

58 Upcoming Events

Frequently Asked Questions about Mechanical Qualification. Ken Boda Applications Engineer Agilent Technologies

Frequently Asked Questions about Mechanical Qualification. Ken Boda Applications Engineer Agilent Technologies Frequently Asked Questions about Mechanical Qualification Ken Boda Applications Engineer Agilent Technologies Is MQ Acceptable to the FDA? Yes Either FDA or ASTM approach is acceptable Many people (including

More information

The Transition to Enhanced Mechanical Qualification of the Dissolution Apparatus

The Transition to Enhanced Mechanical Qualification of the Dissolution Apparatus The Transition to Enhanced Mechanical Qualification of the Dissolution Apparatus Bryan Crist Scientific Affairs Manager Dissolution Systems Agilent Technologies, Inc. Disso Europe 2016 Bucharest, Romania

More information

Mechanical Qualification (MQ) for Apparatus 1 and 2. Oct 10th, 2008

Mechanical Qualification (MQ) for Apparatus 1 and 2. Oct 10th, 2008 Mechanical Qualification (MQ) for Apparatus 1 and 2 Oct 10th, 2008 Overview of Present Qualifications Installation Qualification (IQ) Operational Qualification (OQ) Performance Qualification (PQ) 2 IQ

More information

Food and Drug Administration Division of Pharmaceutical Analysis. Mechanical Qualification of Dissolution Apparatus 1 and 2

Food and Drug Administration Division of Pharmaceutical Analysis. Mechanical Qualification of Dissolution Apparatus 1 and 2 Page 1 of 10 Sections included in this document Table of Contents Document history 1. Purpose 2. Scope/Policy 3. Responsibilities 4. Background 5. References 6. Procedure 7. Records 8. Glossary - Not applicable

More information

02 November, 2017 Moderated by Bryan Crist Agilent Technologies

02 November, 2017 Moderated by Bryan Crist Agilent Technologies 02 November, 2017 Moderated by Bryan Crist Agilent Technologies 1 USP Analytical Instrument Qualification In this session we will discuss the revised USP chapter Analytical Instrument Qualification

More information

711 DISSOLUTION. Portions of the present general chapter text that are national USP text, and therefore not part of the

711 DISSOLUTION. Portions of the present general chapter text that are national USP text, and therefore not part of the Change to read: 711 DISSOLUTION This general chapter is harmonized with the corresponding texts of the European Pharmacopoeia and/or the Japanese Pharmacopoeia. The texts of these pharmacopeias are therefore

More information

DISSOLUTION TEST FOR SOLID DOSAGE FORMS

DISSOLUTION TEST FOR SOLID DOSAGE FORMS Seite 1 von 15 01/2010:20903 corrected 6.8 2.9.3. DISSOLUTION TEST FOR SOLID DOSAGE FORMS This test is provided to determine compliance with the dissolution requirements for solid dosage forms administered

More information

BRIEFING 711 DISSOLUTION

BRIEFING 711 DISSOLUTION BRIEFING 711 Dissolution, USP 28 page 2412 and page 234 of PF 30(1) [Jan. Feb. 2004]. The United States Pharmacopeia is the coordinating pharmacopeia in the efforts toward interntional harmonization of

More information

Qualification Test Specifications for Dissolution Test Systems

Qualification Test Specifications for Dissolution Test Systems 35 for Vantage Dissolution Point Drive, Rochester NY Test 14624 Systems Phone: 800-828- 1470 Fax: 800-395- 0543 Transcat.com Test Set Points/Range Acceptance Criteria Rotational Speed Runout (Wobble) Shaft

More information

International Journal of Generic Drugs ANALYTICAL METHOD PROCEDURES THIS SOP IS 'SWITCHED : OFF : ON'

International Journal of Generic Drugs ANALYTICAL METHOD PROCEDURES THIS SOP IS 'SWITCHED : OFF : ON' DISSOLUTION ASSAY Release & Stability Studies Ed. 04. General Index: 1. PRODUCT SPECIFICATIONS (USP MONOGRAPH) 2. IDENTIFICATION BY HPLC 3. ASSAY - HPLC SETUP 4. STANDARD PREPARATION 5. SYSTEM SUITABILITY

More information

USP Dissolution Instrument Calibration or PQ Standard Operating Procedure or SOP

USP Dissolution Instrument Calibration or PQ Standard Operating Procedure or SOP USP Dissolution Instrument Calibration or PQ Standard Operating Procedure or SOP Part 2: Medium Preparation and Calibrator Tablet Tests Methods, Calculations and Typical Results PHARMA TEST APPARATEBAU

More information

Comparison between Compendial and Enhanced Mechanical Calibration Specifications

Comparison between Compendial and Enhanced Mechanical Calibration Specifications Comparison between Compendial and Enhanced Mechanical Calibration Specifications Calibration Parameter Current Pharmacopoeia Enhanced USP Specification Enhanced FDA Specification Bench Horizontality 1

More information

Dissolution and Drug Release Testing. DISSOLUTION SYSTEMS SOURCE BOOK Edition

Dissolution and Drug Release Testing. DISSOLUTION SYSTEMS SOURCE BOOK Edition Dissolution and Drug Release Testing DISSOLUTION SYSTEMS SOURCE BOOK 2011-2012 Edition Welcome Agilent Technologies, Inc. is well known within the pharmaceutical community for its broad portfolio of instruments

More information

COPLEY TABLET DISSOLUTION TESTERS

COPLEY TABLET DISSOLUTION TESTERS Apparatus 1 (Basket) Apparatus 2 (Paddle) COPLEY TABLET DISSOLUTION TESTERS In the majority of cases, the effectiveness of tablets or capsules administered orally relies on the drug dissolving in the fluids

More information

Dissolution Toolkit Procedures for Mechanical Calibration and Performance Verification Test Apparatus 1 and Apparatus 2 Version 2.

Dissolution Toolkit Procedures for Mechanical Calibration and Performance Verification Test Apparatus 1 and Apparatus 2 Version 2. Dissolution Toolkit Procedures for Mechanical Calibration and Performance Verification Test Apparatus 1 and Apparatus 2 Version 2.0 March 22, 2010 Scope: The dissolution toolkit provides a description

More information

Effect of the Irregular Inner Shape of a Glass Vessel on Prednisone Dissolution Results

Effect of the Irregular Inner Shape of a Glass Vessel on Prednisone Dissolution Results dx.doi.org/10.14227/dt120405p15 Effect of the Irregular Inner Shape of a Glass Vessel on Prednisone Dissolution Results Makoto Tanaka 1, 2, Hiroyasu Fujiwara 1, and Morimasa Fujiwara 1 e-mail: takao-mt@mbj.nifty.com

More information

Effects of Dissolved Gases in Surfactant Dissolution Media

Effects of Dissolved Gases in Surfactant Dissolution Media Effects of Dissolved Gases in Surfactant Dissolution Media Kyle A. Fliszar*, Richard J. Forsyth, Zhong Li, Gregory P. Martin Pharmaceutical Analysis & Control, PR&D, Merck & Co. West Point, PA, USA. email:

More information

ANALYTICAL.

ANALYTICAL. ANALYTICAL P R O D U C T P R O G R A M 2 0 0 8 / 0 9 R PHARMA TEST PAGE 3 > EDITORIAL PAGE 4 > DISSOLUTION TESTER - PT-DT70 PAGE 5 > DISSOLUTION TESTER - PT-DT8S PAGE 6 > DISSLOUTION TESTER - PTWS 310

More information

11 May, 2017 Moderated by Bryan Crist Agilent Technologies

11 May, 2017 Moderated by Bryan Crist Agilent Technologies 11 May, 2017 Moderated by Bryan Crist Agilent Technologies 1 Welcome to the DDG Online Meeting This is our 26 th session Second meeting in 2017; now in our 7 th year of consecutive meetings Worldwide users

More information

Roller Chain and Sprocket Troubleshooting Guide

Roller Chain and Sprocket Troubleshooting Guide Roller Chain and Sprocket Troubleshooting Guide Drive System (Roller Chain and Sprockets) Troubleshooting Guide 1. General Chain is riding up on the sprocket. The roller chain and sprocket do not match.

More information

Hygea Ultrasonic Validation Kit

Hygea Ultrasonic Validation Kit Hygea Ultrasonic Validation Kit Ultrawave s Hygea Ultrasonic Validation kit contains the tools and accessories required to perform Validation and Periodic testing of your ultrasonic cleaning bath. Designed

More information

Chemical Analysis for Methyltestosterone

Chemical Analysis for Methyltestosterone APPLIED CONSUMER SERVICES, INC. Chemical Analysis for Methyltestosterone Report #: 27584 Report to: Mr. Mitch Ross 24 Dockside Lane # 131 Key Largo, Florida 33017 2016 1 1 8 9 0 N W 8 7 t h C o u r t,

More information

POOL CARE GUIDE. The E-Z Clor System To Perfect Pool Water.

POOL CARE GUIDE. The E-Z Clor System To Perfect Pool Water. POOL CARE GUIDE The E-Z Clor System To Perfect Pool Water www.ezclorchemicals.com CONTENTS E-Z Clor 3-Step System...................... 1 Circulation/Your Pump........................ 2 The Filtration

More information

Quantos Automated Dosing Solution Preparation Precise concentrations Process compliance Minimize out-of-specs

Quantos Automated Dosing Solution Preparation Precise concentrations Process compliance Minimize out-of-specs Quantos Automated Dosing Solution Preparation Precise concentrations Process compliance Minimize out-of-specs Gravimetric Sample Preparation The Alternative to Volumetric Flasks From Volumetric to Gravimetric

More information

RULES PUBLICATION NO. 52/P UNDERWATER INSPECTION OF MOBILE OFFSHORE DRILLING UNITS IN LIEU OF DRYDOCKING

RULES PUBLICATION NO. 52/P UNDERWATER INSPECTION OF MOBILE OFFSHORE DRILLING UNITS IN LIEU OF DRYDOCKING RULES PUBLICATION NO. 52/P UNDERWATER INSPECTION OF MOBILE OFFSHORE DRILLING UNITS IN LIEU OF DRYDOCKING 2001 Publications P (Additional Rule Requirements) issued by Polski Rejestr Statków complete or

More information

Peak Pressure Indicator MSI-3

Peak Pressure Indicator MSI-3 Peak Pressure Indicator MSI-3 Operating Instructions Cylinder Pressure Monitoring Digital Pressure Indicator DPI No installation Ready to use High quality sensor Large memory Easy handling Cost effective.com

More information

PROCESS CONTROL HANDBOOK Z-6

PROCESS CONTROL HANDBOOK Z-6 Reversal Bath This step prepares the film for the color-developer step. In the reversal bath, a chemical reversal agent is absorbed into the emulsion. No chemical reaction takes place until the film enters

More information

KOBOLD KSM FLOWMETER. User Instructions

KOBOLD KSM FLOWMETER. User Instructions KOBOLD KSM FLOWMETER User Instructions USA 1801 Parkway View Drive Pittsburgh, PA 15205 PH 412-788-2830 Canada 9A Aviation Point Claire, QC H9R 4Z2 PH 514-428-8090 www.koboldusa.com KSM_manual_08-10 Table

More information

MIL-STD-883H METHOD EXTERNAL VISUAL

MIL-STD-883H METHOD EXTERNAL VISUAL * EXTERNAL VISUAL 1. PURPOSE. The purpose of this test method is to verify the workmanship of hermetically packaged devices. This test method shall also be utilized to inspect for damage due to handling,

More information

Agilent 1220 Infinity II LC Mobile Upgrade Kit

Agilent 1220 Infinity II LC Mobile Upgrade Kit Agilent 1220 Infinity II LC Mobile Upgrade Kit Note 1220 Infinity II LC Mobile Upgrade Kit This note describes the procedures to install an Mobile Upgrade kit to an existing Agilent 1220 Infinity II LC.

More information

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator Procedure No: SOP-029 Revision No: 1.1 (December 29, 2010) Page No.: 1 of 7 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting, air quality

More information

Extended Application Note

Extended Application Note Extended Application Note Harmful Substances in Dietary Supplements HPLC Columns with APP A-337 www.mtc-usa.com 1-732-578-1777 INTRODUCTION A dietary supplement is a substance to be consumed for the purpose

More information

Reliable System Performance PM valves

Reliable System Performance PM valves MAKING MODERN LIVING POSSIBLE Service manual Reliable System Performance PM valves Danfoss components are designed to give a high level of performance over an extended operating life. When Danfoss develops

More information

Calcimeter Instruction Manual

Calcimeter Instruction Manual Hohner (UK - Canada - Texas) Calcimeter Instruction Manual The Hohner Calcimeter is based on industry standard versions, and is used to measure the calcium carbonate and magnesium carbonate in samples.

More information

Best Practice for Calibrating LTH Conductivity Instruments

Best Practice for Calibrating LTH Conductivity Instruments Application Note Best Practice for Calibrating LTH Conductivity Instruments As accurate process measurement becomes an everyday requirement it is vital to be able to calibrate conductivity instruments

More information

TECTYL 1422S APPLICATION PROCEDURES AND TROUBLESHOOTING GUIDE

TECTYL 1422S APPLICATION PROCEDURES AND TROUBLESHOOTING GUIDE Number 38 TECTYL 1422S APPLICATION PROCEDURES AND TROUBLESHOOTING GUIDE Tectyl 1422S is a low VOC, high solids, solventborne corrosion preventive coating. Tectyl 1422S provides excellent protection to

More information

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries. Informal Commentary Limit of Diethylene Glycol (DEG) and Ethylene Glycol (EG) in Sorbitol Solution, Sorbitol Sorbitan Solution and Noncrystallizing Sorbitol Solution December 2009 Monograph/Section(s):

More information

Right-Angle Tube Roller Model

Right-Angle Tube Roller Model SPECIFICATIONS Model no. Free Speed (RPM) Throttle Rotation Max. Torque Ft-Lb Air Pressure psi Air Inlet Hose Air Flow @Free Speed Spindle Weight 909-1700 95 Lever type Reversible 240 90 1/2 NPT 1/2 I.D.

More information

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service? Pipette specifications according to ISO 8655 Chapter 5 5.1 The ISO 8655 standard gives the accuracy and precision limits as both absolute and relative values. Specifications will depend on the technique

More information

ON WATER DISPERSIBLE GRANULES (WG)

ON WATER DISPERSIBLE GRANULES (WG) ON WATER DISPERSIBLE GRANULES (WG) Introduction Water dispersible granules are intended for application after disintegration and dispersion in water by conventional spraying equipment. WGs are formulated

More information

VITLAB micropipette. Standard Operating Procedure (SOP)

VITLAB micropipette. Standard Operating Procedure (SOP) VITLAB micropipette Standard Operating Procedure (SOP) VITLAB GmbH Linus-Pauling-Str.1 63762 Grossostheim Germany tel: +49 6026 97799-0 fax: +49 6026 97799-30 info@vitlab.com www.vitlab.com Table of contents

More information

OWNER S MANUAL NMB SERIES

OWNER S MANUAL NMB SERIES OWNER S MANUAL NMB SERIES NMB Manual V2.0.doc 1. PARTS IDENTIFICATION LIST 1. 5. 2. 4. 3. 6(a). 6(b). Figure 1.1 1. Powerpack with timer option (x1) 2. Cell (x1) 3. Cell Housing (x1) 4. O-ring (x1) 5.

More information

WARNING! DO NOT THROW AWAY THESE INSTRUCTIONS! READ AND UNDERSTAND BEFORE USING EQUIPMENT!

WARNING! DO NOT THROW AWAY THESE INSTRUCTIONS! READ AND UNDERSTAND BEFORE USING EQUIPMENT! Guardian Fall Protection Kent, WA 800-466-6385 www.guardianfall.com GENERAL SYSTEM SELECTION CRITERIA: Selection of fall protection shall be made by a Competent Person. All fall protection equipment shall

More information

Dissolution test results showing biorelevant media made from SIF Powder Original is the same as media prepared using methylene chloride

Dissolution test results showing biorelevant media made from SIF Powder Original is the same as media prepared using methylene chloride Dissolution test results showing biorelevant media made from SIF Powder Original is the same as media prepared using methylene chloride Michael Schaich, Bastian Kloefer, Mathew Leigh : info@biorelevant.com

More information

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION Application Note Doc. G8108-001 Rev. A - 23-Jul-02 ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION For more information regarding Digiquartz products contact: Paroscientific,

More information

CORESTA RECOMMENDED METHOD Nº 67

CORESTA RECOMMENDED METHOD Nº 67 CORESTA RECOMMENDED METHOD Nº 67 DETERMINATION OF WATER IN THE MAINSTREAM SMOKE OF CIGARS BY GAS CHROMATOGRAPHIC ANALYSIS (November 2005) 1. FIELD OF APPLICATION The method is applicable to the particulate

More information

Transferpette. Single and Multichannel Pipettes

Transferpette. Single and Multichannel Pipettes Transferpette Transferpette -8/-12 Transferpette Single and Multichannel Pipettes The Transferpette pipette from BRAND is designed for routine lab and research applications in a shape that is adapted to

More information

TITLE: DNA minitube - Red

TITLE: DNA minitube - Red TITLE: DNA minitube - Red Summary of Operating Conditions: Target Base Pair (Peak) 5 kb (nominal) Please see specific recommendation chapter regarding optimization of operating conditions Temperature (bath)

More information

WARNING! DO NOT THROW AWAY THESE INSTRUCTIONS! READ AND UNDERSTAND BEFORE USING EQUIPMENT!

WARNING! DO NOT THROW AWAY THESE INSTRUCTIONS! READ AND UNDERSTAND BEFORE USING EQUIPMENT! Guardian Fall Protection Kent, WA 800-466-6385 www.guardianfall.com GENERAL SYSTEM SELECTION CRITERIA: Selection of fall protection shall be made by a Competent Person. All fall protection equipment shall

More information

A NOVEL RP-HPLC METHOD FOR THE QUANTIFICATION OF ICATIBANT IN FORMULATIONS

A NOVEL RP-HPLC METHOD FOR THE QUANTIFICATION OF ICATIBANT IN FORMULATIONS A NOVEL RP-HPLC METHOD FOR THE QUANTIFICATION OF ICATIBANT IN FORMULATIONS B.Lakshmi,Kallam Haranadhareddy Institute of Technology, Guntur Prof T.V.Reddy,Prof Mallareddy College of Engineering, Secunderabad

More information

Rubber Dock Hose care, Use & Maintenance

Rubber Dock Hose care, Use & Maintenance Rubber Dock Hose care, Use & Maintenance Prepared by: Thomas J. Wise VP Corporate Development Contents Revised - July. 2014 1. Novaflex Rubber Dock Hose Testing & Inspection Program 1.1 Introduction -

More information

Installation, commissioning and servicing instructions

Installation, commissioning and servicing instructions 488.03 www.reece.com.au Pressure reducing s Installation, commissioning and servicing instructions Function Pressure reducing s are installed in residential water systems to reduce and stabilise inlet

More information

CLASS D - SENSITIVE LEAK TEST GAS AND BUBBLE METHOD. 1.1 To provide definitive requirements for PNEUMATIC pressure testing of piping systems.

CLASS D - SENSITIVE LEAK TEST GAS AND BUBBLE METHOD. 1.1 To provide definitive requirements for PNEUMATIC pressure testing of piping systems. Page 1 of 7 CLASS D - SENSITIVE LEAK TEST GAS AND BUBBLE METHOD 1. SCOPE 1.1 To provide definitive requirements for PNEUMATIC pressure testing of piping systems. 1.2 The piping system as used herein is

More information

Crusher Bearings: Knowing the Basics Leads to Better Care

Crusher Bearings: Knowing the Basics Leads to Better Care Technical Article by Brian Berg and Fabiana Maggico, Manager - Upstream Marketing at The Timken Company Table of Contents: Compression Crushers 2 Jaw Crushers 2 Cone Crushers 2 Impact Crushers 3 Better

More information

How to Achieve Optimal Weighing Performance

How to Achieve Optimal Weighing Performance How to Achieve Optimal Weighing Performance How to Achieve Optimal Weighing Performance With full-resolution 1 μg readability up to 61 g, the new Sartorius high-capacity micro balances are pushing back

More information

ENVIRONMENTAL LABORATORY APPROVAL PROGRAM CERTIFICATION MANUAL

ENVIRONMENTAL LABORATORY APPROVAL PROGRAM CERTIFICATION MANUAL 1 of 7 The quality assurance plan as presented in the Quality Manual should assure that constant and consistent test conditions are met and verified and should be tailored to the laboratory s activities.

More information

STERILIZABLE DISSOLVED OXYGEN SENSORS

STERILIZABLE DISSOLVED OXYGEN SENSORS STERILIZABLE DISSOLVED OXYGEN SENSORS USER MANUAL Table of Contents 1. Introduction 2. Operation 3. Calibration 4. Electrode Verification 5. Mounting 6. Maintenance 7. Storage 8. Instructions for Exchange

More information

User Instructions 1790 Rail Anchor

User Instructions 1790 Rail Anchor User Instructions 1790 Rail Anchor This document is intended to meet the Manufacturer s Instruction requirements as stated by ANSI Z359.1, and should be used as part of an employee training program as

More information

130 SSP Soluble Salt Profiler

130 SSP Soluble Salt Profiler 130 SSP Soluble Salt Profiler Conforms to: SSPC Guide 15, Bresle Equivalent ISO 8502-9 Test Method Page 1 of 6 The Elcometer 130 Soluble Salt Profiler provides fast and accurate measurement of the level

More information

MODEL 200 KNIFE GATE VALVES INSTALLATION & MAINTENANCE MANUAL

MODEL 200 KNIFE GATE VALVES INSTALLATION & MAINTENANCE MANUAL MODEL 200 KNIFE GATE VALVES INSTALLATION & MAINTENANCE MANUAL Index 1. List of components / General arrangement 2. Description 3. Handling 4. Installation 5. Actuators / Operation 6. Maintenance a. Changing

More information

Versatility. Accuracy. Compact. Versa. Automated Headspace Vial Sampler

Versatility. Accuracy. Compact. Versa. Automated Headspace Vial Sampler Versatility Accuracy Compact Versa Automated Headspace Vial Sampler Small Size. Big Value. Automated Static Headspace System Static headspace is one of the most popular techniques due to its versatility

More information

DETERMINING OPTIMUM RESIDUAL ASPHALT CONTENT (RAC) FOR POLYMER-MODIFIED SLURRY SEAL (MICROSURFACING) MIXTURES

DETERMINING OPTIMUM RESIDUAL ASPHALT CONTENT (RAC) FOR POLYMER-MODIFIED SLURRY SEAL (MICROSURFACING) MIXTURES Test Procedure for DETERMINING OPTIMUM RESIDUAL ASPHALT CONTENT (RAC) FOR POLYMER-MODIFIED SLURRY SEAL (MICROSURFACING) MIXTURES TxDOT Designation: Tex-240-F Effective Date: December 2004 1. SCOPE 1.1

More information

Title: Standard Operating Procedure for Elemental and Organic Carbon (EC and OC) using Non-Dispersive Infrared Detection (NDIR)

Title: Standard Operating Procedure for Elemental and Organic Carbon (EC and OC) using Non-Dispersive Infrared Detection (NDIR) Procedure No: SOP-025 Revision No: 1.0 (January 21, 2011) Page No.: 1 of 8 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting and to

More information

J138RSH and J150RSH 3/8" J138RSH 1/2" J150RSH. 3/8" Drive Sealed Head Ratchet 1/2" Drive Sealed Head Ratchet. Operation and Maintenance Manual WARNING

J138RSH and J150RSH 3/8 J138RSH 1/2 J150RSH. 3/8 Drive Sealed Head Ratchet 1/2 Drive Sealed Head Ratchet. Operation and Maintenance Manual WARNING J138RSH and J150RSH Operation and Maintenance Manual 3/8" J138RSH 1/2" J150RSH WARNING To reduce the risk of injury, read and understand these safety warnings and instructions before using the tool. Keep

More information

TITAN FLOW CONTROL, INC.

TITAN FLOW CONTROL, INC. PREFACE: This manual contains information concerning the installation, operation, and maintenance of Titan Flow Control (Titan FCI) Simplex Basket Strainers. To ensure efficient and safe operation of Titan

More information

M-06 Nitrogen Generator (Nitrogen Making Machine)

M-06 Nitrogen Generator (Nitrogen Making Machine) Guideline No.M-06 (201510) M-06 Nitrogen Generator (Nitrogen Making Machine) Issued date: 20 th October, 2015 China Classification Society Foreword This Guideline is a part of CCS Rules, which contains

More information

Research Question How does the concentration of catalase affect the speed of the decomposition reaction of Hydrogen Peroxide into oxygen and water?

Research Question How does the concentration of catalase affect the speed of the decomposition reaction of Hydrogen Peroxide into oxygen and water? Research Question How does the concentration of catalase affect the speed of the decomposition reaction of Hydrogen Peroxide into oxygen and water? Aim To observe the effect of increasing enzyme (catalase)

More information

ISO INTERNATIONAL STANDARD. Pulps Determination of drainability Part 1: Schopper-Riegler method

ISO INTERNATIONAL STANDARD. Pulps Determination of drainability Part 1: Schopper-Riegler method Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 5267-1 Second edition 1999-03-01 Pulps Determination of drainability Part 1: Schopper-Riegler method Pâtes Détermination de l'égouttabilité Partie

More information

Transferpette. Lightweight precise rugged reliable!

Transferpette. Lightweight precise rugged reliable! Lightweight precise rugged reliable! Transferpette Solutions for S cience F I R S T C L A S S B R A N D The Transferpette S microliter pipettes provide all features required for pipetting in Life Science

More information

Installation & Operation Manual Proven Quality since 1892

Installation & Operation Manual Proven Quality since 1892 Content 1. ERIKS operating companies 2. Product description 3. Requirements for maintenance staff 4. Transport and storage 5. Function 6. Application 7. Installation 8. Maintenance 9. Service and repair

More information

Air Eliminator Series

Air Eliminator Series Air Eliminator Series Air Separation Equipment FILE NO: 9-1 DATE: August 2010 SUPERSEDES: 9-1 DATE: January 2009 Air Eliminator Series Air Separation Equipment Modern sealed heating and cooling systems

More information

Transferpette. Lightweight precise rugged reliable!

Transferpette. Lightweight precise rugged reliable! Lightweight precise rugged reliable! Transferpette Solutions for S cience F I R S T C L A S S B R A N D The Transferpette S microliter pipettes provide all features required for pipetting in Life Science

More information

ST 18 Aluminium Gas Model

ST 18 Aluminium Gas Model ST 18 Aluminium Gas Model Operators Manual Issue 5 For correct operation of this autoclave it is essential that the chamber be filled with water up to the height of the studs before every use. DIXONS SURGICAL

More information

PERFORM Operating Document

PERFORM Operating Document PERFORM Operating Document Use and Maintenance of CO 2 Incubator PC-POD-CA-007-v02 Revision History Version Reason for Revision Date 01 New POD 30-Sep-13 02 Minor revisions for section 2.3, 3.1, 4.3. 14-April-16

More information

MANUFACTURING PROCEDURE SPECIFICATION INSPECTION TEST PLAN

MANUFACTURING PROCEDURE SPECIFICATION INSPECTION TEST PLAN TEST PLAN 1.0 SCOPE The purpose of this work instruction is to establish and maintain documented procedure to ensure that the magnetic particle inspection of bend body over extrados and intrados is carried

More information

OXYGEN MINERALE ABSOLUTE RANGE OWNER S MANUAL OXY MINERALE 600/1000 SERIES

OXYGEN MINERALE ABSOLUTE RANGE OWNER S MANUAL OXY MINERALE 600/1000 SERIES PURAPOOL OXYGEN MINERALE ABSOLUTE RANGE OWNER S MANUAL OXY MINERALE 600/1000 SERIES 1. PARTS IDENTIFICATION LIST 1. 5. 2. 4. 3. 6. 6. Figure 1.1 1. OXYGEN Minerale Power pack with timer option (x1) 2.

More information

HEATEC TEC-NOTE Publication No Revised

HEATEC TEC-NOTE Publication No Revised HEATEC TEC-NOTE Revised 6-17-10 Setting up Vega radar sensors for Heatec vertical asphalt tanks This document provides information for setting up Vega radar sensors installed on Heatec tanks that store

More information

Eliminating Sources of Pipetting Error in the Forensic Laboratory David M. Epstein Director of Scientific Operations. Abstract.

Eliminating Sources of Pipetting Error in the Forensic Laboratory David M. Epstein Director of Scientific Operations. Abstract. Page 1 of 6 Table of Contents Back Issues Search Editors About FSC Submissions Forensic Science Communications October 2003 Volume 5 Number 4 Research and Technology Eliminating Sources of Pipetting Error

More information

(2 pts) Draw the line of best fit through the data and estimate the concentration of Fe in your sample solution.

(2 pts) Draw the line of best fit through the data and estimate the concentration of Fe in your sample solution. Name: 1. You are asked to measure the percent iron in a steel sample you believe to be about 50% Fe by mass. You dissolve four replicate samples in nitric acid and dilute them by mass to a concentration

More information

Used with Dissolution Tester for Fraction and Flow Measurements

Used with Dissolution Tester for Fraction and Flow Measurements TOYAMA Auto Sampler W, Model PAS-615 Used with Dissolution Tester for Fraction and Flow Measurements This is a full automatic sampler to be used with 6-shaft or 8-shaft dissolution tester, with the following

More information

Series 8500 Expansion Compensators. Catalog 674H

Series 8500 Expansion Compensators. Catalog 674H Series 8500 Expansion Compensators Catalog 674H 500 Laminated Bellows Expansion J Series 8500 on Joints Expansion Compensators Sizes 3/4" through 4" Threaded, welded, flanged and grooved steel pipe joints

More information

Sterile Visual Flow Indicator

Sterile Visual Flow Indicator Sterile Visual Flow Indicator Sanitary Process Connection Installation / Operation / Maintenance Manual P.O. Box 1116 Twinsburg, OH 44087 Phone: 330/405-3040 Fax: 330/405-3070 E-mail: view@ljstar.com Web

More information

Transferpette -8/-12

Transferpette -8/-12 Transferpette -8/-12 Testing Instructions (SOP) May 2009 1. Introduction The standard DIN EN ISO 8655 describes both the design and the testing of the piston operated pipette Transferpette 8/ 12. The following

More information

PRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator.

PRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator. PRS(TC)4,8 USER MANUAL Read the complete manual before installing and using the regulator. WARNING INCORRECT OR IMPROPER USE OF THIS PRODUCT CAN CAUSE SERIOUS PERSONAL INJURY AND PROPERTY DAMAGE. Due to

More information

AUTOMATIC FLOW CARTRIDGES. Competition Analysis Not All Automatics Are Created Equal

AUTOMATIC FLOW CARTRIDGES. Competition Analysis Not All Automatics Are Created Equal - AUTOMATIC FLOW CARTRIDGES Competition Analysis Not All Automatics Are Created Equal IMI FLOW DESIGN / AUTOMATIC FLOW CARTRIDGES / F345.0 COMPETITION ANALYSIS Improve the of your Premium System Components...

More information

Spa Start Up & Maintenance

Spa Start Up & Maintenance Spa Start Up & Maintenance Your spa or hot tub is an investment in relaxation. Follow these simple steps for maximum enjoyment! Quality We offer a complete variety of products to support both of our chlorinating

More information

Comparison of the Effectiveness of Various Deaeration Techniques

Comparison of the Effectiveness of Various Deaeration Techniques Comparison of the Effectiveness of Various Deaeration Techniques Owen S. Degenhardt 1, Bonnie Waters 1, Anna Rebelo-Cameirao 1, Annelene Meyer 2, Heiko Brunner 2, and Nicholas P. Toltl 1,3 email: toltl_nick@lilly.com

More information

Needle valve. Contents. User s Manual. (1) Be sure to read the following warranty clauses of our product 1. (2) General operating instructions 2

Needle valve. Contents. User s Manual. (1) Be sure to read the following warranty clauses of our product 1. (2) General operating instructions 2 Serial No. H-V024-E-7 Needle valve User s Manual Contents (1) Be sure to read the following warranty clauses of our product 1 (2) General operating instructions 2 (3) General instructions for transportation,

More information

Improve Process Reliability

Improve Process Reliability Convenient Conductivity Calibration to Meet Pharmacopeia Regulations Pharmacopeias require that conductivity measurement systems be periodically calibrated to ensure their precision. The range of accessories

More information

OD plug gauge. for bores and internal profiles from dia. 2,5 mm. OD plug gauge in open design. Special designs Accessories. OD plug gauge for various

OD plug gauge. for bores and internal profiles from dia. 2,5 mm. OD plug gauge in open design. Special designs Accessories. OD plug gauge for various OD plug gauge for various bore diameters Cross sleeve Cross sleeve with runners OD plug gauge for bores and internal profiles from dia. 2,5 300 Oskar Schwenk GmbH & Co. Esslinger Straße 84 70736 Fellbach

More information

Hi-Temp Spa Care Guide

Hi-Temp Spa Care Guide Hi-Temp Spa Care Guide by Hasa Designed for Warm Therapuetic Waters 1 Contents Get Started................................ 3 Filling Your Spa............................. 3 Water Balance..............................

More information

TECH TIPS: ROPE DEFECTS

TECH TIPS: ROPE DEFECTS Hoist Ropes: Why must they be inspected? A hoist rope is not made to last forever. This is a useful fact, as the kinds of wear that appear on a rope can indicate areas of problems within an installation

More information

v /07/17 Page 1 of 18

v /07/17 Page 1 of 18 A. PURPOSE AND SCOPE Details the on-site measurement of free and total chlorine in water samples using the Palintest DPD,GP/HR and HACH DPD methods. B. CONTENTS A. PURPOSE AND SCOPE... 1 B. CONTENTS...

More information

ROTATING DISK VALVES INSTALLATION AND MAINTENANCE 1. SCOPE 3 2. INFORMATION ON USAGE 3 3. VALVE TYPES 3 4. OPERATORS 5 5. VALVE CONSTRUCTION 6

ROTATING DISK VALVES INSTALLATION AND MAINTENANCE 1. SCOPE 3 2. INFORMATION ON USAGE 3 3. VALVE TYPES 3 4. OPERATORS 5 5. VALVE CONSTRUCTION 6 Sub Section INDEX Page Number 1. SCOPE 3 2. INFORMATION ON USAGE 3 3. VALVE TYPES 3 4. OPERATORS 5 5. VALVE CONSTRUCTION 6 6. INSTALLATION AND OPERATION 6 7. MAINTENANCE 8 8. REPAIR 9 9. ASSEMBLY 10 10.

More information

HOW TO NOT MEASURE GAS - ORIFICE. Dee Hummel. Targa Resources

HOW TO NOT MEASURE GAS - ORIFICE. Dee Hummel. Targa Resources HOW TO NOT MEASURE GAS - ORIFICE Dee Hummel Targa Resources Introduction Measuring natural gas is both a science and an art. Guidelines and industry practices explain how to accurately measure natural

More information

28 Day Start Up Instructions July 2018

28 Day Start Up Instructions July 2018 28 Day Start Up Instructions July 2018 4322 S 80 th Street Mesa AZ 85212 877-984-4687 Wet-Edge.com Support@Wet-Edge.com 28 DAY START UP INSTRUCTIONS PROTECT YOUR INVESTMENT! 1 PRIOR TO POOL FINISH INSTALLATION

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 14 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCE AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

"HAND, WRIST AND FINGER SAFETY"

HAND, WRIST AND FINGER SAFETY PRESENTER'S GUIDE "HAND, WRIST AND FINGER SAFETY" Part of the "SAFETY MEETING KIT" Series Quality Safety and Health Products, for Today...and Tomorrow OUTLINE OF MAJOR PROGRAM POINTS OUTLINE OF MAJOR PROGRAM

More information

Transferpette. Testing Instructions (SOP) 1. Introduction. October 1998

Transferpette. Testing Instructions (SOP) 1. Introduction. October 1998 Transferpette Testing Instructions (SOP) October 1998 1. Introduction The standards ISO DIS 8655 and DIN 12650 describe both the design and the testing of the piston operated pipette Transferpette. The

More information

FRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS

FRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS FRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS Analysis Conducted BY CC RESEARCH (P) Ltd., For Nanoceutical Solutions Inc. USA 2950 Thousand Oaks drive, Suite 3, San Antonio, Texas EXECUTIVE SUMMARY

More information