Instrumental operation: ICS 3000
|
|
- Cody Short
- 5 years ago
- Views:
Transcription
1 Instrumental operation: ICS 3000 Scope The scope of this document is to present an overview of the Dionex Ion Chromatography System (ICS) 3000 and to provide the reader with guidelines for standard operations, routine maintenance and simple trouble shooting. Detailed instrument description can be found in Dionex Reference Library (CD- ROM) which contains all relevant and specific manuals. Instrument overview The Dionex ICS 3000 system is designed for ion chromatography (IC). It is compatible with high salt and alkaline solvents and therefore fittings, tubings and interior of the system is made of PEEK (polyether ether ketone) to prevent corrosion. It consists of a dual pump system with four channels on each pump. It is capable of operating two columns with two different detectors at the same time and is equipped with a pulsed amperometry detector (PAD) and a conductivity detector. In practice, the common use currently will be one column at a time and primarily on Pump 1 and the PAD detector which will be described here. The PAD is based on electrochemical detection and is in general superior for detection of carbohydrates that are traditionally difficult to detect. The system also includes an eluent generator, which can be used when ultra pure eluents are needed. In practice, the eluent generator has limitations that make it difficult to use for many applications and it will therefore be bypassed for most applications. The principle of the conductivity detector is that changes in conductivity is measured and interpreted as analytes. This requires the installation of an ion suppressor prior to the conductivity detector, which will suppress the inherently high conductivity coming from the eluents. Figure 1 shows an instrument overview. As opposed to many other HPLC systems, the system should be operated with more or less constant flow to avoid detrimental salt precipitations in the pump, column and detector. If the flow will be stopped for a prolonged time period make sure to replace salt containing eluents in all lines with e.g. 50% methanol (also to avoid microbial growth). HPLC systems are delicate equipment and should be handled accordingly. Regularly check the status of the instrument, preferably on a daily basis and always be careful when operating. Eluents should ALWAYS be prepared from fresh top quality water (Milli-Q/18.2 MΩ cm) and preferably sample preparation should also be made based on this water quality. The chromatographic system is controlled by a computer installed with Chromeleon 7 software (hereafter CM) version for instrument control and data handling. All commands and instrument handlings goes through this software.
2 The system consists of 5 compartments/modules: 1. Autosampler 2. Dual pump 3. Eluent generator 4. Column compartment (lower part) 5. Detector compartment (upper part) Figure 1: General system appearance of the ICS-3000
3 Window for starting and stopping Instrument Controller. Can be closed after use. Icon informing about Instrument Controller status. Green: Controller running. White with red cross: Controller stopped. Figure 2: Instrument Controller. Software unit that controls communication between computer and HPLC. Must be running at all times when system is operated. Communication between the computer (CM) and the HPLC is controlled by the Instrument Controller Software found under the Start Menu/Programs/Chromeleon/Service Manager. Alternatively, double click the Chromeleon Icon in the Windows task bar to open the Instrument Controller (figure 2)or right click and select start Chromeleon Instrument Controller. When the HPLC is to be controlled/operated the Instrument Controller must be running. Starting and stopping is done by pressing the command Start/Stop Instrument Controller (figure 2). The window can be closed afterwards. The Instrument Controller does not start automatically after a computer restart and it should therefore be checked that it is running, before starting operation. In the operating software (CM) each module has its own operational screen, which may be found as tabs in the top of the window (figure 3).
4 Overview Autosampler PAD detector Pump Queue System 1 PAD electrochemical detection Switching between Instrument and Data panel Figure 3: Module and system overview from controlling software Chromeleon 7. Unmarked tabs: DC ; Detector compartment. EGC1 ; Eluent generator. Audit ; Audit trail logging every event in the system. Startup ; Programmed system startup (not recommended to use unless you know what you re doing). Operating the pump The internals of the dual pump system is depicted in figure 4. The lower half of the compartment is Pump 1 and the upper half is Pump 2. The pumps are operated at flow rates between 0 and 10ml/min and max pressure is 5000 psi. Each pump is connected to four eluent channels (A, B, C and D). A Rear Seal Wash system is installed to periodically flush the backside of the pump heads in order to avoid salt precipitations and corrosion. Wash setting may be Interval, Automatic or Off. Correct setting is Interval, but the rear seal wash solutions needs to be changed on a weekly basis and consists of 20% methanol/ethanol (figure 4). Note dates of change in the equipment logbook.
5 Figure 4: Internals of Dual Pump. Pump 2 is placed in the upper half of the compartment, Pump 1 in the lower half. Note the screw valve at position 3 is the purge valve on Pump 2. Equivalent purge valve on Pump 1 found in the same position. Position 11 is the rear seal wash reservoir which should be changed on a weekly basis (20% methanol or ethanol). Preparing eluents The eluents for ion chromatography are crucial for good performance and the chromatography (column) is extremely sensitive to contaminations in the eluents. It is therefore of outmost importance that extra care is taken when eluents are prepared. Remember that the lifetime of eluents for ion chromatography is also very short compared to other types of chromatography. The biggest risk of contamination is dissolution of carbonate in the eluents and this occurs when eluents are exposed to atmospheric air. This carbonate accumulates on the columns stationary phase and reduces separation capacity and carbonate has a higher push effect as mobile phase then the others regularly applied. It is impossible to completely avoid carbonate contamination but with the proper precautions it can be minimized and allow a certain lifetime of the eluents (max. 48 hours for top performance). This includes that
6 headspace in the eluent bottles must be saturated with nitrogen to minimize the amount of carbon dioxide. Nitrogen is delivered by a nitrogen generator via a restriction valve and distributed between three of the four channels (channel D does not get nitrogen supply). Three different eluents are made to accommodate the range of analytical methods and all three are changed every time new eluents are prepared. The eluents are always prepared in the same way and always placed on the following channels: Channel A: 0.1 M NaOH Channel B: 1 M NaOAc in 0.1 M NaOH Channel C: Milli-Q water Eluent A and B are most labile for carbonate contamination and should be handled accordingly.the following is a description for generation of each eluent but should be performed alongside so that all three eluents are done approximately at the same time. All eluents are sensitive to atmospheric air. Therefore don t leave any of the newly prepared eluents standing without N 2 over pressure in the headspace for more than 15 minutes. Before any eluents are made, turn off the flow on the system and release the nitrogen pressure on the restriction valve positioned on the backside of the autosampler. Empty all eluent bottles and rinse them thoroughly in Milli-Q water. Special pressure stable eluent bottles are used since strong corrosive eluents are pressurized during running. Therefore, don t send the bottles off to washing, but rinse them properly before reuse. Eluent A: 0.1 M NaOH (2L) 1. Measure out 2 L of Milli-Q water in a measuring flask. Make sure to use clean glassware. 2. Poor the two liter of water into the eluent bottle named Eluent A (0.1 M NaOH). 3. Loosely place a lid on the bottle, but don t tighten it. 4. Place the eluent bottle in the sonication bath and adjust the water volume in the bath to the level indicated as working level. The sonication bath will fit two eluent bottles at a time. 5. Sonication for 20 min. DON T put your hands into the sonicator while it is operating and close the door to the room when you leave it. 6. When the sonication is done, close the lid securely and move the eluent bottle to the HPLC. 7. Connect channel A to the eluent bottle and start the nitrogen flow. Exchange the air in the head space by repeatedly opening and closing the lid on the bottle to relieve pressure. A total of three reliefs are sufficient. 8. Use a sterile plastic pipette with a total volume of 10 ml and a manual pipette bulb for pipetting a concentrated 50% NaOH solution (Fluka, L). The solution is stored in the exhausted waste cupboard underneath the HPLC and is extremely concentrated. ALWAYS wear
7 proper personal protection equipment (glasses, gloves and lab coat/acid rubber coat) when handling this solution. I am not kidding getting that stuff in your eyes will ruin your eyesight! 9. Check the level of NaOH solution in the bottle if the content is less than half of the original replace with new to prevent risk of carbonate contamination in the NaOH solution. Place the remainder of the solution in the toxic cupboard and mark it with COMMON USE for other purposes than HPLC. 10. Pipette 10.4 ml to the 2 L bottle of eluent. Be careful when pipetting, as the solution is very viscous. Don t dip the entire pipette into the solution only the tip underneath the surface. Close the lid to the NaOH solution immediately after use and only use the same pipette ONCE in the solution. Don t double dip take a new pipette. 11. Dispense the NaOH solution to the water in the eluent bottle underneath the water surface and be patient as the liquid will flow very slowly (viscous). Slowly rinse the inside of the pipette by gently pipetting up and down 3 times. 12. Close the lid to the eluent bottle and exchange the headspace air with nitrogen as done in step 7. Swirl the eluent bottle carefully DON T shake it, to distribute NaOH evenly. Eluent B: 1 M NaOAce in 0.1 M NaOH (1L) 1. Weigh g anhydrous sodium acetate (Sigma-Aldrich, KG,or Fluka). It is important that the correct quality of sodium acetate is used. 2. Poor this into a 1 L measuring flask (use clean glass ware) and add about 0.8L Milli-Q water. 3. Add a magnetic stir bar and place the solution on a magnetic stirrer till you have a complete dissolution. 4. Remove the magnetic stir bar (without dipping the magic stick into the solution!) and fill the remaining volume to 1 L (with freshly tapped Milli-Q water). 5. Mount a 0.22 µm sterile filter cup on top of the cleaned eluent bottle (f.ex µm stericup GP Millipore Express PLUS membrane, 250 ml funnel, 45 mm neck size). Apply vacuum to filter the eluent. 6. Loosely place a lid on top of the eluent bottle. Don t tighten it. 7. Place the eluent bottle in the sonication bath, make sure water surface in the bath is at the working level. 8. Sonicatefor 20 min. DON T put your hands into the sonicator while it is operating and close the door to the room when you leave it. 9. When the sonication is done, close the lid securely and move the eluent bottle to the HPLC. 10. Connect channel B to the eluent bottle and start the nitrogen flow. Exchange the air in the head space by repeatedly opening and closing the lid on the bottle to relieve pressure. A total of three reliefs are sufficient.
8 11. Follow the procedures for NaOH addition as described for Eluent A step EXCEPT! Only add 5.2 ml NaOH to the 1 L NaOAc-solution (ie. there should be the same NaOH concentration in both channel A and B). Eluent C: Milli-Q water 1. Fill the rinsed eluent bottle with Milli-Q water freshly tapped from the system. 2. Loosely place a lid on top of the eluent bottle. Don t tighten it. 3. Place the eluent bottle in the sonication bath and adjust the water volume in the bath to the level indicated as working level. 4. Sonicatefor 20 min. DON T put your hands into the sonicator while it is operating and close the door to the room when you leave it. 5. When the sonication is done, close the lid securely and move the eluent bottle to the HPLC. 6. Connect channel C to the eluent bottle and start the nitrogen flow. Exchange the air in the head space by repeatedly opening and closing the lid on the bottle to relieve pressure. A total of three reliefs are sufficient. End the preparation procedure by purging all three channels as described below and note in the logbook date of eluent change. Purging eluents When eluents are changed the pump needs to be purged to remove air bubble formation and old eluent from the tubes. The system is sensitive to air (as all HPLCs) and purging is therefore important. Purging the eluent channels is done be applying high flow through the specified channels and will create high back pressure on the system, if the purge valve has not been open (Open the purge valve (figure 4) by approx. half a turn). This will direct the flow path to a waste tube and relieve the pressure on the system (and the column in particular). Once the purge valve has been opened choose the appropriate channels and flow distribution to purge in CM and activate the purge/prime function (figure 5). It is possible to purge more than one channel at a time by setting the percentage distribution between the channels as desired (equal distribution, 50:50 when two channels, 33% when three and 25% when all four channels are to be purged). The software is preprogrammed to perform purging for 500 seconds at 5.7 ml/min. Repeat purging (if necessary) until air bubbles are no longer visible in the eluent channels from the bottles to the pump. IMPORTANT: Before purging the software will inform that this operation will generate a high flow through the system and that the purge valve should be open before executing. Press Execute despite warning, if the purge valve is open. If the system senses any pressure increase (even minor) during purging, the command will be terminated. Make sure that the valve has been sufficiently loosened and try again.
9 Retention time count Module connection Pump motor Flow settings Pressure indicator Prime/purge function Solvent distribution Figure 5: Pump tab. Relevant functions marked for purging, flow settings ect. Once the purging is done, close the purge valve again tightly to redirect the flow path to the column. If the valve is not tightened enough, the pressure during operation will not be sufficiently high meaning that the flow delivery to the column is insufficient. No warnings about low pressure will occur, unless pressure is completely lost (less than 200 psi) so please do observe that correct pressure is obtained during start up. Normal operating pressure for CarboPac PA1 incl. guard is at least psi (this includes a pressure restrictor coil at 1000 psi). Operating the autosampler Samples are placed in the autosampler prior to operation. The front door of the autosampler is opened (held in place by a relatively strong magnet pull it hard!) and the vials placed in the sample tray (figure 6). On the outside of the autosampler in the lowest left corner is the main power switch not a door opener! Only press this button if you really want to cut the power to the autosampler. Sample position in the tray is straight forward numbering from 1 to 100. Each position is marked on the tray. The tray may be removed from the autosampler when samples are being loaded (be careful ). The sample tray is placed in a cooled compartment and the autosampler will not operate if the front door is open. A sound alarm will occur if the door is left open (turns off again when the door is closed). Don t open the door if the robot arm is moving as this will immediately interrupt the movements. The front panel will inform
10 about the state of the autosampler: Idle means that the autosampler can be opened for vial handling (this is also the case for idle status during a sequence run). Otherwise, the front panel is used for manual handling and is not used during routine operation. All normal operational control is handled via CM. Figure 8 shows the routine operations of the autosampler from CM. The sampler has a fixed volume loop and if this needs to be changed, call a super user. Partial injection may be applied if necessary, but will result in increased standard deviations, but please consult super user or manual. NB: Be aware that the sampler consumes approx. 40 µl sample for a 2 µl injection. On top of the autosampler (towards the back) is the syringe tower (figure 6). This is where sample, buffer etc. is taken and measured before injection to the column. Make sure no air bubbles are trapped in the syringe by priming sampler before you start your analysis. The syringe is connected to a wash solution consisting of 20% methanol. The sample is introduced via the injection port, via the diverter valve (position 1 directs sample to system 1), via the loop (flow path is via the loop when the injection valve is in inject mode, meaning always except during an injection cycle) to the column. When the needle leaves the inject port (figure 7) the bed should be dry with no droplets/spillage. If not, something is obstructing the injection and poor/no signals will be observed in the following chromatogram. The needle is washed before each injection in the flush port (figure 7) next to the injection port. Here, wash solution will appear (and may spill) which is normal.
11 Diverter valve Figure 6: Autosampler with sample tray dismounted from the tray compartment Figure 7: Injection and flush port in autosampler
12 Autosampler status Sample loop size Temperature control in sample tray Prime syringe flush sample Diverter valve/inject valve Figure 8: Autosampler tab in CM. Operating the column compartment The column compartment is the module where the column is connected (figure 9). During operation the front cover of the column compartment should preferably be closed and especially if temperature control is applied (improved reproducibility is observed when temperature is set slightly above ambient). As mentioned above the flow is directed from the autosampler via the left injection valve (for pump 1) to the column. The loop is also mounted on the injection valve. Do not mess around with the tube connections here! The column is fitted in an appropriate holder in the back of the compartment.
13 Left injection valve for Pump Analytical column Note flow direction: Right to left! Analytical Column flow direction Guard column Note flow direction: Left to right! Figure 9: Internals of column and detector compartment. Blue arrow indicates flow direction on the column, which should fit with column flow directions. Upper thrid of the compartment is not present in this particular instrument set up. When a column is fitted and flow is applied please ensure that there are no leaks. Leaks at the column inlet will result in pressure decrease, whereas leaks at the column outlet may not affect the pressure but will affect the detector signal and potential flooding of compartment. If leaks are observed or registered by the system, further tighten the connections either by hand or by a wrench key (be careful, PEEK connections are weak). Column temperature is specified in CM (figure 10) and the desired temperature is written in the field for Set. Press Enter to execute the command. The field Actual will inform about the actual temperature in the column compartment (which initially is not necessarily the same as the column temperature).
14 Temperature control Figure 10: Column compartment tab for controlling column temperature. Operating the PAD detector The detector is placed as can be seen in figure 9 (item 2) and a close-up photo in figure 11. The detector consists of a gold working electrode and a ph reference electrode. The ph reference electrode (number 3 in figure 11) has very limited needs for maintenance during routine operations and should not be removed from its socket. The working electrode on the other hand is disposable and needs to be changed regularly. Routines for changing working electrode are described below. The detector can only be turned on when flow is on the system and a pump link exists where the cell voltage on the electrode is turned off, if the flow is stopped. However, cell voltage should only be turned on during sequence running and other diagnostic runs, since prolonged voltage will decrease the electrodes life time. When running, the electrode senses two signals, a total signal (a background signal that depends on the type of eluents and state of the electrode) and a baseline/real sample signal which can be autozero ed. The detector responds to any substances that are either oxidizable or reducible and the electrical output results from an electron flow caused by the chemical reaction that takes place at the surface of the electrodes. The eluents are highly charged and this can be observed as the background (Total) signal. Normal operating Total Signal should be between 15 and 35 nc. The reference electrode is capable of
15 adjusting/compensating changes in background signal when gradient elution profiles are run and in that sense maintain a stable baseline. If the reference electrode is not installed/out of order, the baseline will rise considerably when gradient profiles are used. PAD is a particularly strong detection principle for carbohydrates, because the highly alkaline eluents will charge the hydroxyl groups and therefore give rise to signals in the detector. Furthermore, this detection principle is highly sensitive and therefore makes it superior for carbohydrate analysis. From column Figure 11: PAD electrode To waste
16 Signal and Total Signal (background) Autozero (of Signal) Turning on/off cell voltage Figure 12: PAD tab in CM. Changing working electrode When the Total Signal becomes too high (above nc) or the signals during running behaves strange (extremely high signal or seemingly no signal at all/increased background signals), changing the working electrode may be needed (The electrode has a life time normally between 1 and 3 weeks continuous operation). If you suspect the electrode may be worn out, check last date of change in the logbook to verify. New users should consult a super user before changing the electrode the first time. Hereafter do as follows: 1. Turn off the flow and thereby the cell voltage. 2. Disconnect the inlet and outlet of the detector (by the incoming and out coming of the red tubing in figure 11). You may need a wrench key. 3. Unplug the blue and the yellow power plugs (figure 11). 4. Grab the electrode by two hands and pull it firmly towards you in one motion (yes, it will come off). 5. While performing the following steps, try to keep the entire electrode in an upright position. The ph reference electrode contains KCl which may leak if turned upside down. 6. Loosen the big knob (picture A, figure 13), press the two side-wings on the clips that holds the spacer block in place and take off the entire clips.
17 7. Remove the spacer block and now the individual parts of the electrode are visible (picture B, figure 13). The working electrode is the disposable one but take EXTRA care of the gasket. They are not supplied enough to be changed every time. Unless it is torn or otherwise damages it should be reused. 8. Inspect the working electrode on the gold -side. Salt precipitations or loss of connection between top and bottom is reason for loss of signal. Discard the electrode. 9. Inspect the cell body and carefully rinse it if visible salt precipitations are seen (picture B, figure 13). 10. Take a new working electrode from the package, but ONLY lift it in the snip of white tape that goes outside the actual electrode which says lift here. Use a pair of tweezers if necessary. 11. Place the working electrode and the gasket back on the cell body. Note that the two positioning holes in the gasket are different in shape, which compared to the spacer block indicates the positioning. The same goes for the working electrode. Furthermore, the electrode should be placed so that the golden parts face the cell body and so that the small circle fits over the slit in the gasket. Do not bend the working electrode in any way! Position it precisely over the two rods and slide it onto the cell body. See correct placement of the working electrode in picture C, figure Place the spacer block and the clips back on the electrode and turn the big knob until you hear a click. This is as tight as it gets. 13. Reposition the entire electrode in the detector compartment (it needs to say click), re-plug the electrical cords (yellow in yellow, blue in black!), reconnect column and waste tubings, restart flow on system (make sure there are no leaks!!) and set cell voltage on. 14. Monitor baseline for a while (approx min). The signal can be quite unstable in the beginning (and very high), but will eventually stabilize. Total signal should preferably stabilize between 15 and 30 nc. A B C Spacer block Spring- clips Working electrode Cell body Gasket Spacer block Figure 13: Changing disposable gold electrode
18 Waste collection From the detector the flow stream is passed to a waste container standing in the exhaust cabin underneath the system. This waste container needs to be checked regularly and emptied. The waste is almost exclusively consisting alkaline salts and can therefore be poured in the drain. For the same reason spillage of waste will generate vast amounts of salt precipitations and should therefore be avoided. Starting up for operation 1. Make sure that eluent level is good and check the last date for eluent preparation. If the eluents are more than 2 days old, consider if you should make new eluents before your run. If preparing new eluents, remember to purge all channels in use (A,B and C) after eluent preparation 2. Check that the correct column is on. 3. Check that flow is on and pressure correct (column flow: 0.25 ml/min, pressure: psi) 4. If column temperature control is applied make sure that the column compartment is closed (lower door). 5. Turn on the detector cell (apply cell voltage). 6. Start Baseline Monitoring (see figure 14) and watch the detector signal stabilize. 7. If the signal does not stabilize or if the total signal is very high (above 45 nc) it may be time to change the working electrode. Follow the procedure for doing so, if necessary. 8. Prime sampler, flush sample loop (make sure no air is entrapped in sample syringe) 9. Place the samples in the autosampler and prepare the sequence (see section below). If starting up after eluent change, the first sample should ALWAYS be a blank sample running the instrument method: Start after eluent change. This is to ensure regeneration of column after possible carbonate contaminations. Afterwards condition the column with 1-2 blank runs with the instrument method you intend to use before injection of the first real sample/standard. These conditioning blanks should always be run no matter if eluent has been changed or not. 10. Stop baseline monitoring. 11. Start the run either by starting the sequence directly or by queuing the sequence (add to queue in the Queue tab). Press Ready Check to get an estimate of eluent use during the entire queue.
19 Monitoring baseline. Monitor pump pressure Queue tab. For queuing several sequences Successful ready check Adding a sequence to the queue. Ready check. Testing if the system is ready for operation. Gives an estimate of eluent use. Figure 14: Queue tab. Here sequences can be queued for automatic start and general checking of system status (ready check) is performed. Setting up a sequence in CM The data structure in CM is shown in figure 15, which is initially divided between different instruments. The PAD detector is set up to system ED_1. Hereafter, a general data folder contains a folder for each user, specified by name. Maintain this order of data structure and save your sequence under your own name.
20 Switching between Instrument and Data panel Figure 15: Data structure in CM. Note the switch between Instrument and Data panel. When making a sequence, choose Create from the top menu and choose Sequence. Hereafter a wizard will start, that will guide you through setting up of a sequence: 1. Choose system: ED_1 2. Write a recognition name that it will be copied to for each sample (may be edited later), choose number of samples, number of injections pr. sample (most often 1), position of 1 st vial and injection volume (ul), normally the same as the sample loop size. 3. Choose Instrument Method (analytical run method) by accepting the lastly used method or by browsing through folders. Choose Processing Method if applicable. Can also be added later. Set default channel to ED_1. 4. Write a comment describing the particular samples (not obligatory). 5. Press finish and save the sequence to your folder under an appropriate name starting with the following date format: YYMMDD_ Figure 16 shows and example of a prepared, finished sequence. Most fields in the sequence can be edited before, during and after the run. This includes: Name, Type (blank, unknown, standard), Level (only relevant for quantification), Position (vial position in the autosampler), Volume (injection volume), Instrument Method (analytical run method, can only be edited before running) and
21 Processing Method (qualititaive/quantitative method). Some fields have scroll down menus, others are text fields. Always press Save after editing any fields. A shortcut called Fill Down is made for fast assigning sample position in the autosampler. Mark the Position field from the position of the first vial, press F9 and choose Renumber. Press enter. Alternatively, select the Fill Down menu (figure 17), choose Renumber and write the name of the first sample position. Press OK. Note that the fill down function may be used in all the other columns as well. IMPORTANT: Changing any of these fields mentioned above will not affect the results, but changing the field Status will erase data behind, if the samples have been run; Idle indicates a sample not yet analysed, Finished indicates a sample analysed and therefore a file containing data. Interrupted is a sample stopped during analysis due to some sort of error message, may also contain data, but sometimes not useful data. Figure 16: Prepared, finished sequence.
22 Fill Down menu Marked cell for 1 st vial position Figure 17: Fill down menu for fast sample position assignment. Press Fill Down or F9 to open window. Alarms and warming Pressure alarm If the pressure falls below or reaches the upper limits of the settings (which are mainly set to protect the column) a pressure sound alarm will be initiated. At the same time, all flow will be stopped. High pressure may occur if the samples contain insoluble particulate material, if the solvent concentration suddenly changes, temperature control fails or if the column is overloaded with sample due to many injections without regeneration. Loss of pressure may occur if the eluent bottles are empty, air is introduced in the system or if leaks happen on the upstream /pressure side of the column. If leakage is the cause, the system will detect the compartment where this is occurring via leak detectors and the audit trail will inform about this. Identify the source of the pressure violations, take proper action and restart the system. Leak/humidity alarm Identify the reason for leakage (which compartment) and correct if possible. Dry up any spillage in the compartments with soft tissue paper.
23 No pressure? Check that the purge valve is properly closed. Check that all fittings are securely tightened. If this is not enough purge the channels as air may have been trapped in the pump. No connection to modules? Check that the Instrument Controller is running (figure 2). When this is positivly running, reconnect the non-connected modules by pressing Connect in each module tab (see figure 5 or 8). If there is still no connection restart the computer and start the Instrument Controller again. Still no connection call security. Software error messages Close all active windows and restart CM. Still software errors, restart computer and start the Instrument Controller again. Still errors call security. No contact with Autosampler The autosampler needs a restart once in a while. If no contact can be achieved through the software, turn off the autosampler (power button lower left corner, front). Leave it off for 1-2 minutes and start again. It will now perform a few self tests (this takes a couple of minutes. When this is done, reconnect in CM. If there is still no connection, call security. Call security For more advanced guidance either contact Super users or consult the general equipment manuals found on the CD-rom: Reference Library provided by Dionex. Super users: Jane Wittrup Agger (jane.agger@umb.no; ) Bjørge Westereng (bjorge.westereng@umb.no; )
Using the Akta Prime plus October 22, 2012
Some starting precautions: 1. Vacuum filter all buffers. Removes any large particles/debris that may clog your column De-gases the buffers 2. Clarify lysates first by centrifugation and then filtration
More informationAKTA pure 25 New Owner s Intro
AKTA pure 25 New Owner s Intro The exercise below will give a quick demonstration of how easy and intuitive the AKTA pure 25 will be for you in demonstrating downstream processing to your students. Steps
More informationSOP: Buck Scientific BLC-20P HPLC Operation
Page 1 of 11 Approvals Preparer: John Buford Reviewer: Tim Kull Reviewer: Dr. Margaret Bryans Date: 16OCT13 Date: 30OCT13 Date: 31OCT13 1. Purpose 1.1. Basic operation of the Buck Scientific BLC-20P isocratic
More informationApplication Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction
Application Note Rapid performance verification of AZURA systems with refractive index detector Method Keywords ID HPLC Quality control, system verification, refractive index, AZURA Analytical HPLC Plus
More informationAKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by:
1 AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by: Materials/Reagents/Equipment Vendor 1. 0.2 um filtered Water + 0.05% sodium azide 2. 0.2 um filtered
More informationUSER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve
USER'S GUIDE Nanoflow Probe 3-4mm Red Stripe ptfe Sleeve To Injector ZU1XC Fused Silica 20µm 90µm TTF115 ptfe Sleeve Fused Silica 25µm 375µm Micromass UK Limited Floats Road Wythenshawe M23 9LZ Tel: +44
More informationSOP: Cation Exchange Chromatography of t-pa using AKTA Pure system
Page 1 of 10 Approvals Preparer: Dr. David Frank Reviewer: Jason McMillan Reviewer: Dr. Maggie Bryans Date: 20APR16 Date: 21APR16 Date: 21APR16 1. Purpose 1.1. This procedure describes the operation of
More informationManual. Online Degasser DDG-75. Manual Degasser DDG-75 April DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE Hockenheim Germany
Manual Degasser DDG-75 April 2007 1 Manual Online Degasser DDG-75 DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE-68766 Hockenheim Germany Tel. +49-6205-9450-0 Fax. +49-6205-9450-33 email info@duratec.com
More informationASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS Dionex Corporation
ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS 2000 Dionex Corporation Document No. 031277 Revision 03 April 2000 2000 Dionex Corporation All rights reserved worldwide. Printed in the United States of
More informationPipettor. User Manual
Pipettor User Manual Product Code Description Single Chanel Pipettor Variable 550.002.055 Volume 0.5 to 10ul 550.002.060 2 to 20ul 550.002.065 10 to 100ul 550.002.070 20 to 200ul 550.002.075 100 to 100ul
More informationLipodrop 2.0 device coating
Lipodrop 2.0 device coating Compiled by K.V., K.Ž, J.R AIM: To selectively coat channels in order to make them hydrophobic (pre junction channels) or hydrophilic (post junction channels). REAGENTS USED:
More informationGCMSD-Headspace Analysis SOP
Before you start GCMSD-Headspace Analysis SOP Method and Sequences names are restricted to 50 characters (MSD program will crash otherwise) The GC oven and Injection Ports need to be cooled to 50 oc to
More informationAKTA MC SOP Page 1 9/27/04 AKTA METAL CHELATING SOP
AKTA MC SOP Page 1 9/27/04 Date: 9/27/04 Author: A DeGiovanni Edited by: R Kim Reviewed by: Y. Lou AKTA METAL CHELATING SOP Materials/Reagents/Equipment Vendor 1. Water + 0.05% sodium azide 2. 0.2 um filtered
More informationUser guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection
User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection This user guide does not have the intention to introduce you to any of the theory and/or principles behind
More informationGhost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager
Ghost Peaks: How to Fix a Haunting Problem Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager Outline of Topics What is a ghost peak? How to identify ghost peaks and their source How to correct
More informationQuantitative Analysis of Hydrocarbons by Gas Chromatography
Quantitative Analysis of Hydrocarbons by Gas Chromatography Introduction Gas-liquid chromatography (GLC) accomplishes a separation by partitioning solutes between a mobile gas phase and a stationary liquid
More informationAKTA 3D SOP. Click on the System Control tab. This screen has 4 windows.
AKTA 3D SOP Page 1 9/09/04 Date: 9/09/04 Author: A DeGiovanni Reviewed by: Y. Lou AKTA 3D SOP Materials/Reagents/Equipment Vendor 1. 0.2 um filtered Water + 0.05% sodium azide 2. 0.2 um filtered buffers
More informationArmfield Distillation Column Operation Guidelines
Armfield Distillation Column Operation Guidelines 11-2016 R.Cox Safety SAFETY GLASSES ARE REQUIRED WHEN OPERATING THE DISTILLATION COLUMN Wear gloves when mixing alcohol feedstock The column will become
More informationEvaluation copy. Interdependence of Plants and Animals. computer OBJECTIVES MATERIALS
Interdependence of Plants and Animals Computer 14 Plants and animals share many of the same chemicals throughout their lives. In most ecosystems, O 2, CO 2, water, food and nutrients are exchanged between
More informationSchool of Chemistry SOP For Operation Of Glove Boxes
School SOP for Operation of Glove Boxes: The following SOP provides guidelines on how to adequately and safely operating a standard laboratory glove box and its associated devices. General Information:
More informationAgilent 1290 Infinity Pump Head Maintenance
Agilent 1290 Infinity Pump Head Maintenance Technical Note Agilent Technologies Notices Agilent Technologies, Inc. 2012-2015, 2016 No part of this manual may be reproduced in any form or by any means (including
More informationAcclaim RSLC. (Rapid Separation Liquid Chromatography)
Acclaim RSLC (Rapid Separation Liquid Chromatography) Product Manual for Acclaim RSLC Columns Page 1 of 9 Product Manual for Acclaim RSLC Columns (Rapid Separation Liquid Chromatography) Analytical Columns
More informationStandard Operational Procedure for Capillary Electrophoresis
Page 1 of 14 Standard Operational Procedure for Capillary Electrophoresis *Please ask an experienced operator for training before you use the CE if you can. A demo is most probably more efficient than
More informationAgilent 1220 Infinity II LC Mobile Upgrade Kit
Agilent 1220 Infinity II LC Mobile Upgrade Kit Note 1220 Infinity II LC Mobile Upgrade Kit This note describes the procedures to install an Mobile Upgrade kit to an existing Agilent 1220 Infinity II LC.
More informationUNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008
UNITY 2 TM Air Server Series 2 Operators Manual Version 1.0 February 2008 1. Introduction to the Air Server Accessory for UNITY 2...2 1.1. Summary of Operation...2 2. Developing a UNITY 2-Air Server method
More informationEquipment Operating Procedure Glove Box
Equipment Operating Procedure Glove Box Page 1 0.0 Changing the Compressed Gas Cylinder 1. Complete Compressed Gas Cylinder training from EHS website before manually exchanging gas cylinders. In order
More informationRead this first. Zetasizer nano series Self installation and Quick start guide MRK825-02
! Read this first Zetasizer nano series Self installation and Quick start guide I N S T R U M E N T S MRK825-02 Zetasizer Nano series Self installation and Quick start guide MAN0383 Issue 1.1 July 2007
More informationIn Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure
PPMS Service Note 1099-412 In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure Introduction: Loss of electricity to the PPMS EverCool II should not cause damage to
More informationThe instrument will process the samples one by one till all the samples have been cleaned up.
Table of Contents SPE-01 Cleanup Station... 1 LC-04SP Valve System... 5 LC-04SP Standalone Valve... 10 LC-05 Auto Injector... 12 LC-06 Dosing Pump... 16 LC-08 Column Selector... 17 SPE-01 Cleanup Station
More informationSuperconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)
Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Sensitivity: 1x10 11 A m 2 Applied DC fields: 0 T to 5 T Applied AC fields: 0 G to 3 G (zero-to-peak), 0.01 Hz to 1000 Hz Temperatures
More informationAerobic Respiration. Evaluation copy
Aerobic Respiration Computer 17 Aerobic cellular respiration is the process of converting the chemical energy of organic molecules into a form immediately usable by organisms. Glucose may be oxidized completely
More informationFROG-4000 TM Chemical Analysis System
FROG-4000 TM Chemical Analysis System User s Manual Defiant Technologies, Inc. 2015 Vol.5 This page is intentionally left blank. FROG-4000 TM Chemical Analysis System User s Manual Defiant Technologies,
More informationValve INV-907. Instructions. 1 Introduction. 2 Installation. Safety. Unpacking. Installing the valve
Instructions 1 Introduction is a motorised rotary 7 port valve. The valve is used in ÄKTA design chromatography systems. It is powered from the ÄKTAdesign P-900 series system pump, and is controlled from
More information6900 Maintenance Instruction System Flush
Equipment Required FA74005 Damper Drain Tube FA16005 Cover Removal Tool FA900005 Beaker 0.25 Litre FA900003 Solvent Cleaning Bottle FA940021 Syringe Polypropylene 50 ml as required FA999045 Gloves Latex
More informationRayPette Plus Autoclavable Pipette. User Manual
RayPette Plus Autoclavable Pipette User Manual CONTENTS 1. YOUR NEW PIPETTE... 1 1.1. Adjustable volume pipettes... 1 1.3 Fully autoclavable... 2 2. UNPACKING... 3 3. INSTALLING THE PIPETTE HOLDER... 4
More informationFigure 1: Plug in ion inlet of mass spec
Starting procedure: by Vicky - is the vacuum pump running (should run all time, otherwise it ll need 4-5 hours to have a adequate vacuum for the MS) otherwise something has gone wrong- F.T - Remove plug
More informationMB-SP Series Solvent Purification System
Operation Manual MB-SP Series Solvent Purification System Table of contents 1. System overview 2 2. Column activation.. 5 3. System purging.. 6 4. Glove box integration.... 7 5. Solvent reservoir installation....
More informationInstructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems
Corporation Issued: 5/09 P/N: 600104 Rev. 2 Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems The Gas Addition Accessory permits the
More informationArtisan Technology Group is your source for quality new and certified-used/pre-owned equipment
Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment FAST SHIPPING AND DELIVERY TENS OF THOUSANDS OF IN-STOCK ITEMS EQUIPMENT DEMOS HUNDREDS OF MANUFACTURERS SUPPORTED
More informationORB-400 BUBBLE/HAZE MACHINE. Item ref: UK User Manual
ORB-400 BUBBLE/HAZE MACHINE Item ref: 160.462UK User Manual Caution: Please read this manual carefully before operating Damage caused by misuse is not covered by the warranty Introduction Thank you for
More informationFROG-4000 TM Chemical Analysis System
FROG-4000 TM Chemical Analysis System User s Manual Defiant Technologies, Inc. 2014 Vol.4 This page is intentionally left blank. FROG-4000 TM Chemical Analysis System User s Manual Defiant Technologies,
More informationThe pipettes cover a volume range from 0.1µl to 10ml.
CONTENTS 1. YOUR NEW PIPETTE 1 1.1. Adjustable volume pipettes 1 1.2. Fixed volume pipettes 2 1.3 Fully autoclavable 3 2. UNPACKING 3 3. INSTALLING THE PIPETTE HOLDER 4 4. PIPETTE COMPONENTS 5 5. PIPETTE
More informationAutomated Liquid Handling Station
Automated Liquid Handling Station Software User s Guide Manual Part Number 32-0442-048 Rev 0 2018 Teledyne Technologies Incorporated. All rights reserved. Printed in USA. Installation See the SimPrep Quick
More informationUsER manual for Watersens ph -REDOX
UsER manual for Watersens -REDOX Cl 8 1 2 6 3 3 7 7 4 4 4 4 Parts List 1 Redox Probe 1 x 2 PH Probe 1 x 5 Tube Weight 2 x 6 Connection Valve 1 x chlorine 3 Chlorine and Pumps 2 x 7 Dosing Valve 2 x 5 5
More informationPegas 4000 MF Gas Mixer InstructionManual Columbus Instruments
Pegas 4000 MF Gas Mixer InstructionManual Contents I Table of Contents Foreword Part I Introduction 1 2 1 System overview... 2 2 Specifications... 3 Part II Installation 4 1 Rear panel connections...
More informationInstallation and Use Manual
Installation and Use Manual EMASMB & LMASMB Surface Mount Bottle Filling Stations Model EMASMB IMPORTANT THIS IS AN INDOOR APPLICATION ONLY! ALL SERVICE TO BE PERFORMED BY AN AUTHORIZED SERVICE PERSONNEL.
More informationStandard Operating Procedure
Page 1 of 10 Vigor Glove Box Standard Operating Procedure Lab: ESB 154 Department: Materials Science and Engineering PI: Paul V. Braun Written By: Subing (Alan) Qu Section 1: Overview Type of SOP: Process
More informationAutomated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:
Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: The RSK-175 standard operating procedure was developed in order to determine the amount of dissolved gas in water. Due to the expansion
More informationAppendix: Simplified Instructions for Operating the HP 5890 GC/FID and Air Sampling Systems
Appendix: Simplified Instructions for Operating the HP 5890 GC/FID and Air Sampling Systems The bucket brigade air sampler. For the square Tedlar bags with the push/pull valves: Put the valve of a tedlar
More informationSolvent System Walkthrough
v2.0 1 Solvent System Walkthrough The Contour Glass solvent system is designed to give you air-free anhydrous solvent with minimal effort. However, the system is only as good as its users so please read
More informationGlove Box Installation Manual
Glove Box Installation Manual 1998 by M. Braun Company File: GB-UNI-INS.DOC! Edition 08-00 by M. Boutin! Subject to be changed without notice Glovebox Installation Your Glove box has been fully assembled,
More informationTANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel
TANK MANAGER FOR TWO TANKS OPERATING MANUAL 10/31/11 C-More T6C L color touch panel 1 TABLE OF CONTENTS GENERAL...3 INSTALLATION...4 STONE TEST PROCEDURE...7 OPERATIONAL SUMMARY...7 AUTO CARBONATION...10
More informationSetting up and running a column
Setting up and running a column PART 1: What you need and theory PART 2: Using a bellow and packing the column PART 3: Loading sample to silica PART 4: Running column PART 5: TLC analysis and combining
More informationSRI Model H2-40 Hydrogen Generator Operation and Maintenance Nov 2006
Nov 2006 Place the H2-40 on a benchtop surface which will not be damaged by water in the event of a leak. Fill a bottle with clean water. Tapwater is OK in most large cities, but bottled drinking water
More informationBASIC RULES OF USAGE, OLIS DSM-20 CD Management The instrument is currently managed by the Analytical Biophysics Core Facility.
Olis DSM-20 instructrions and rules p. 1 BASIC RULES OF USAGE, OLIS DSM-20 CD Management The instrument is currently managed by the Analytical Biophysics Core Facility. Signup On a first-come, first-served
More informationno peaks missing peaks too small peaks increasing retention decreasing retention
GC Troubleshooting We can help you. 1. Observation: no peaks no peaks missing peaks too small peaks increasing retention decreasing retention declining baseline rising baseline bleeding plateaus interfering
More informationSOP: Anion Exchange Chromatography of Green Fluorescent Protein (GFP) using the AKTA Pure system
Page 1 of 12 Approvals Preparer: David Frank Reviewer: Hetal Doshi Reviewer: Dr. Maggie Bryans Date: 21APR16 Date: 21APR16 Date: 21APR16 1. Purpose 1.1. This procedure describes the operation of the ÄKTApure
More informationACV-10 Automatic Control Valve
ACV-10 Automatic Control Valve Installation, Operation & Maintenance General: The Archer Instruments ACV-10 is a precision automatic feed rate control valve for use in vacuum systems feeding Chlorine,
More informationPrep LC Training Manual S.V.
1234 Hach Hall 515-294-5805 www.cif.iastate.edu Prep LC Training Manual 11-6-2013 S.V. Location: Contact: 1238 Hach Hall Kamel Harrata, 1236 Hach Hall and Steve Veysey, 1234 Hach Hall Safety: All researchers
More informationLC, RSLC, Bio RS and Dual Gradient Systems
Thermo Scientific Dionex UltiMate 3000 LC, RSLC, Bio RS and Dual Gradient Systems P/N 4823.0205 Revision B December 2014 Automated Method Scouting Considerations All information in this manual is subject
More informationUWPR. LC-MS setup. Before you connect your trap and column. Connecting your trap and column
Before you connect your trap and column You need to provide your own trap and column (Note our fittings accommodate trap and column made with fused silica 360 µm OD). Check our guide for packing traps
More informationInstruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text.
Hach LDO Sensor Instruction Sheet 00745589 Safety Precautions Please read this entire instruction sheet before operating this sensor. Pay particular attention to all danger and caution statements. Failure
More informationExperiment 18 Properties of Gases
Experiment 18 Properties of Gases E18-1 E18-2 The Task In this experiment you will investigate some of the properties of gases, i.e. how gases flow, their phase changes and chemical reactivity. Skills
More informationSOP for use of the SX19 Stopped-Flow Instrument
SOP for use of the SX19 Stopped-Flow Instrument The main components of the stopped flow system are: Always book the system before use. Starting up and preparing the SX19 for use: 1. Turn on the compressed
More informationOMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT
OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT PA/PH/OMCL (16) 17 R QUALIFICATION OF EQUIPMENT ANNEX 2: QUALIFICATION OF GC EQUIPMENT Full document title and reference Document type Legislative
More informationCORESTA RECOMMENDED METHOD Nº 67
CORESTA RECOMMENDED METHOD Nº 67 DETERMINATION OF WATER IN THE MAINSTREAM SMOKE OF CIGARS BY GAS CHROMATOGRAPHIC ANALYSIS (November 2005) 1. FIELD OF APPLICATION The method is applicable to the particulate
More informationUser Manual. Quantos Automated Dosing Liquid Module
User Manual Liquid Module 1 Safety Information 1.1 Definition of warnings and symbols Signal Words WARNING for a hazardous situation with medium risk, possibly resulting in severe injuries or death if
More informationOperation of the Perkin Elmer TGA-GC/MS
Operation of the Perkin Elmer TGA-GC/MS Summary of the TGA-GC/MS: The TGA-GC/MS allows a user to decompose a sample by heating, measure its loss of mass, and simultaneously analyze the chemical composition
More informationBD Cytopeia Fluidic Kit User s Guide
BD Cytopeia Fluidic Kit User s Guide For Research Use Only 23-17618-00 9/2015 Becton, Dickinson and Company BD Biosciences 2350 Qume Drive San Jose, CA 95131 USA BD Biosciences European Customer Support
More informationAppendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation
Page 1 of 19 : SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor Description and Principle of Operation The photoacoustic multi-gas monitor (INNOVA 1412, Innova AirTech Instruments, Denmark) is a highly
More informationGeneral Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!
The portable Gas Chromatograph is now more versatile than ever. DPS Instruments is pleased to present the newest Portable Companion 2 Gas Chromatograph with room for 2 Detectors. The Companion 2 GC was
More informationCalcium Reactor. CalcFeeder PRO. User Manual 1.1a. Model. CalcFeeder AC1 PRO CalcFeeder AC2 PRO CalcFeeder AC3 PRO CalcFeeder AC4 PRO
Calcium Reactor CalcFeeder PRO User Manual 1.1a Model CalcFeeder AC1 PRO CalcFeeder AC2 PRO CalcFeeder AC3 PRO CalcFeeder AC4 PRO Page 1 Dear Customer, Thank you for purchasing an Pacific Sun CalcFeeder
More informationMASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120
MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120 QUICK START GUIDE P/N 6006154, REVISION C MAY 2013 Model 8120 Mask Integrity Tester is patented under U.S. Patent No. 8,312,761. Additional patents are pending.
More informationEmulsiflex C3 Homogenizer Protocol
Emulsiflex C3 Homogenizer Protocol Contact Annette Erbse, ex. 2-0528, erbse@colorado.edu, Office C316 Nicole Kethley, Nicole.kethley@colorado.edu, Office C316 or C370 Please contact Annette Erbse or Nicole
More informationNanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI
NanoSight NS300 Operation instructions Injection/flushing brief overview: 1. Do not exceed flow of 1 ml per 20 seconds. 2. Inject two 1 ml syringes with nano-pure or DI water. 3. If the water does not
More informationI T T Pressure Reducing Valve WARNING INSTALLATION, OPERATION, AND MAINTENANCE MANUAL
INSTALLATION, OPERATION, AND MAINTENANCE MANUAL I-867-4T 867-4T Pressure Reducing Valve HANG THESE INSTRUCTIONS ON THE INSTALLED VALVE FOR FUTURE REFERENCE WARNING Read and understand all instructions
More informationMicrofluidic Demonstration
Microfluidic Demonstration Co-flow of liquids across a microfluidic chip Create a microfluidic controller to flow liquids across a microfluidic chip. 1. Assemble fluidic circuits and attach to breadboard.
More informationSomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings
SomnoSuite FAQ V1.3 January 2015 Setup 1. How do I connect the SomnoSuite to my oxygen source? 2. Is there a way to speed up the downward movement of the pusher block when setting the empty position? 3.
More informationMix 920 ml of 87% glycerol (on shelf) with 1080 ml of Isoton II (in cabinet), stir in a flask until homogenized (the glycerol will sink).
Instructions for the care and use of the Multisizer 3 The Multisizer is intended for quantifying either cells or spheres. We are in possession of two apertures: one with a 100-µm diameter opening and the
More informationStandard Operating Procedure #COE-SOP-0001 Chemical Fume Hood Operation
Standard Operating Procedure # Chemical Fume Hood Operation Facility: NMSU College of Engineering Laboratories Written by: Juanita Miller, Safety Specialist, miljgh@nmsu.edu (575)-646-1292 Scope: This
More informationInstallation of Your SprayMaster System
Installation of Your SprayMaster System 1. At the installation site, remove all equipment from the corrugated box and the polyethylene drum and replace the drum lid. Check the picture to identify each
More informationINSTALLATION, CARE & USE MANUAL. EMASMB & LMASMB Surface Mount Bottle Filling Stations
INSTALLATION, CARE & USE MANUAL EMASMB & LMASMB Surface Mount Bottle Filling Stations IMPORTANT THIS IS AN INDOOR APPLICATION ONLY! ALL SERVICE TO BE PERFORMED BY AN AUTHORIZED SERVICE PERSONNEL. TOOLS/ITEMS
More informationEnvironmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!
After years of development and testing, DPS Instruments is pleased to present the newest, most expandable and versatile Gas Chromatography Systems in history. The DPS 600 Series GC systems are the world
More information[USING THE VITROBOT April 9, Using the Vitrobot
Using the Vitrobot The Vitrobot Mk 3 (aka Mark III, Mark 3, etc.) can control both the local temperature and the local humidity during the process of plunge-freezing a grid for cryoem. It does require
More informationElcometer 138. Bresle Kit and Patches. Operating Instructions
Elcometer 138 English Bresle Kit and Patches Operating Instructions R English This product meets the Electromagnetic Compatibility Directive. The product is Class B, Group 1 ISM equipment according to
More informationAgilent ICP-MS Interactive Troubleshooting tool for Low sensitivity
Agilent ICP-MS Interactive Troubleshooting tool for Low sensitivity Rev. B 2017,Janyary,20 Next > Preconditions 7900, 7800, 7700, or 8800 ICP-MS is used. Sample introduction type is PeriPump. If you are
More informationVersion BA D
Aurora Dispo osable Cartridge Handling Manual For Aurora cartridge 210 0001 CA D Version 1.00 106 0010 BA D 2011 Boreal Genomics, Inc. All rights reserved. All trademarks are property of their owners.
More information6800 Maintenance Instruction System Flush Procedure
Equipment Required FA74005 FA65318 FA900005 FA900003 Damper Drain Tube 6800 Cover Removal Tool Beaker 0.25 Litre Solvent Cleaning Bottle FA940021 Syringe Polypropylene 50 ml as required FA999045 Gloves
More informationInstructions for SILEX 1C Mixed bed cartridge filter
D01B-30A-UK1 January 2017 Instructions for SILEX 1C Mixed bed cartridge filter With conductivity sensor Gravity flow or pressure installation SILHORKO-EUROWATER A/S Phone +45 86 57 12 22 DK-8660 Skanderborg
More informationElcometer 138. Bresle Kit and Patches. Operating Instructions
Elcometer 138 English Bresle Kit and Patches Operating Instructions R English The Conductivity Meter model B-173 has been tested in accordance with EU regulations governing Electro-magnetic compliance
More informationOPERATING AND MAINTENANCE MANUAL
Series 4300 Engineered Performance TABLE OF CONTENTS 0 INTRODUCTION 1 1 Scope 1 2 Description 1 3 Specifications 1 0 INSTALLATION 1 1 Mounting 1 2 Piping 1 1 Connecting Process Pressure 2 2 Vent Connections
More informationARC Unofficial Manual to the Solvent Dispensing System. Designed by J. C. Meyer
RC Unofficial Manual to the Solvent Dispensing System Designed by J. C. Meyer 1 Table of Contents: I. Introduction...3 II. rgon Supply...3 Tanks Regulators Over Pressurization III. Solvent Kegs...4 Solvent
More informationMoveRoll Conveyor Operating and Maintenance Manual
MoveRoll Conveyor Operating and Maintenance Manual 1. Read this first! This manual contains information for protection of personnel in the roll handling area from possible injury and/or equipment damage.
More informationUsing the UltraRAE. Firmware 2.35
Using the UltraRAE Firmware 2.35 Training Agenda UltraRAE features Setting up the UltraRAE Turning on the UltraRAE Idle Operation RAE-Sep Tubes Prepping for a measurement Taking a measurement Alarm modes
More informationThe Univentor 1250 Anaesthesia Unit
THE UNIVENTOR 1200/1250 ANAESTHESIA UNIT The Univentor 1250 Anaesthesia Unit TABLE OF CONTENTS EDITION 1 Section 1 - WARRANTY & SERVICE 1.1. WARRANTY 2 1.2. DAMAGED SHIPMENTS 2 1.3. SERVICE 2 Section 2
More informationModel: 43T. Bermad Pressure Relief Valve
Model: 43T Bermad Pressure Relief Valve Installation Operation Maintenance Manual () Rev.C1_01.08.17 Page 1 of 10 Safety First BERMAD believes that the safety of personnel working with and around our equipment
More informationAQUARIUS. Operating Instructions for the. Type 70 Electrolytic Gas Soldering / Welding Unit. Table of Contents
AQUARIUS Operating Instructions for the Type 70 Electrolytic Gas Soldering / Welding Unit Table of Contents 1. Important General Remarks 2 2. Safety Instructions 2 3. Description of the Soldering Unit
More informationCavitron DualSelect. Introduction ENGLISH. Directions For Use. Cavitron DualSelect Dispensing System ENGLISH 1
ENGLISH Cavitron DualSelect Dispensing System Directions For Use Please read carefully and completely before operating unit. Introduction Cavitron DualSelect Dispensing System Congratulations. Your purchase
More informationJETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012
JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012 UNIVERSITY OF TEXAS AT ARLINGTON Nanofabrication Research and Teaching Facility TABLE OF CONTENTS 1. Introduction....2 1.1 Scope of Work.....2
More informationObservations Possible Causes Recommended Actions. Turquoise color
ABI PRISM 3700 Instrument-Hardware-Array View Troubleshooting Module Please contact technical support if you have any questions regarding the document. Text in blue indicates a link to another section
More information