Equipment Operating Procedure Glove Box
|
|
- Maud Horn
- 6 years ago
- Views:
Transcription
1 Equipment Operating Procedure Glove Box Page 1
2 0.0 Changing the Compressed Gas Cylinder 1. Complete Compressed Gas Cylinder training from EHS website before manually exchanging gas cylinders. In order to access gas cylinders stored in Scott Lab, request the garage key from the 3 rd floor office. Scott Lab office keys have access to cylinder storage. Cylinder storage is the 2 nd door on the left in the 1 st floor garage in West side of Scott Lab. 2. To remove current cylinder, turn valve on top of tank clockwise to close it. 3. Using a wrench, twist the bolt connecting cylinder to safety valve in an upward motion to loosen it. Continue twisting the bolt until the safety valve is removed. Replace protective valve covering on gas cylinder. 4. Important: Tear off the contents tag on the cylinder to show the tank is Empty. 5. To attach new gas cylinder, remove protective valve covering and insert the safety valve in the port on top of the cylinder. 6. Using a wrench, screw the valve bolt in a downward motion to tighten it completely. 7. Turn the control valve in top of the cylinder counter-clockwise to open flow from cylinder. 8. Tear off the contents tag on the cylinder to show the tank is In Use. 9. The cylinder is now fully attached and ready to use. 1.0 Antechamber Controls These steps assume the atmosphere in the glove box is purified and not open to the outside air. Security Code for glove box controls: Large Antechamber Inserting an Item into the Glovebox 1. Important: Verify the antechamber contains purified gas before opening the inner door. 2. Ensure the inner door and refill valve are fully closed. 3. Open the outside door and insert the object into the Antechamber. 4. Close the outside door and open the vacuum valve and evacuate to -29 in. Hg. 5. Close the vacuum valve and open the refill valve until pressure reaches -15 in. Hg. 6. Repeat evacuation and refill at least 3 times, finally reaching atmospheric pressure. 7. Close the refill valve and open the inner door to remove the object from the chamber. Page 2
3 1.1.2 Removing an Item from the Glovebox 1. Ensure that the outside door and refill valve are fully closed. 2. Open the inner door. 3. Insert the object into the Antechamber and close the inner door. 4. Open the outside door and remove the object. 5. Purge the antechamber before re-opening the inside door. 1.2 Small Antechamber Inserting an Item into the Glovebox 1. Ensure the inner door and 3-way valve for mini-chamber are fully closed. 2. Open the outside door and insert the object into the Antechamber. 3. Close the outside door and turn 3-way valve to evacuate and evacuate to -29 in. Hg. 4. Turn the 3-way valve to refill until pressure reaches -15 in. Hg. 5. Repeat evacuation and refill at least 3 times. 6. Close the 3-way valve and open the inner door to remove the object from the chamber Removing an Item from the Glovebox 1. Important: Verify the antechamber contains purified gas before opening the inner door. 2. Ensure that the outside door and 3-way valve are fully closed. 3. Open the inner door and insert the object. 4. Close the inner door. 5. Open the outer door and remove the object. 6. Purge the antechamber before re-opening the inside door. 2.0 Removing Front Panel to Expose Glove Box 1. Verify that toxic materials will not be present inside of glove box at the time. 2. If planning to purge glove box and reset inert Argon atmosphere, verify that at least six (6) full tanks of argon are available for the glove box. 3. Close valve on argon tank by turning the valve clockwise. Page 3
4 4. Turn off power to glove box my pressing switch on front-facing panel located beneath antechamber. 5. Raise light fixtures over the top of the glove box. Light fixtures rotate along hinge. 6. Remove gloves from the front panel by removing the two bands securing the glove and then pulling off the glove itself. 7. While carefully holding onto the front panel, open each safety lock sealing the front panel and the glove box. IMPORTANT: Start at the top, work down the sides, and finally unlock the bottom. 8. Warning: Avoid letting the panel edge contact the sealing on front of glove box. This may lead to a scratch that prevents the seal from forming when system restarts. Be extra careful when removing the front panel from the glove box. Set the panel aside. 3.0 Restarting Glove Box after Exposing Atmosphere 1. Make sure at least 6 full argon tanks are on hand to purge and refill the glove box. 2. Verify that all seals are secured on front panels, each antechamber door is tightly closed, exhaust system is functioning, and cooling system is functioning (if applicable). 3. Note: If a front panel was removed from the glove box at any point, clean the seal with rubbing alcohol and apply high vacuum grease. 4. Turn on argon feed to the glove box by turning the top valve clockwise. 5. Turn on glove box power using the switch located on the front panel beneath the antechambers. 6. Perform a pressure check on the glove box (see section 5.0) 7. (step 5-8 is the purging method for the antechamber) Set the Working Maximum pressure to mbar and the Working Minimum Pressure to +5.0 mbar. To do this, press the System Config. button and adjust the numeric values with the blue display. The working gas will enter the box automatically. 8. Turn on the vacuum pump by pressing the Start button on the main screen. 9. Open the evacuate valve on the Large Antechamber and allow the chamber to evacuate to evacuate to Close the evacuate valve and repeat for the mini antechamber. 11. Wait 10 minutes for the glovebox to stabilize. 12. Use the appropriate foot pedal to add/remove gas to/from the glovebox to reach desired pressure. 13. Important: See section 5.0 to perform a pressure test on the glove box and verify that all seals are secured. 14. It is extremely important to purge the glove box to displace as much air in the box as possible with inert gas. Important: Purging the glovebox requires 3 full cylinders of Page 4
5 working gas (6 cylinders for large glove box). Follow the steps in section 4.0 to purge the system. 4.0 Purging the Glovebox 1. Purging the glove box requires 3 full cylinders of working gas (6 cylinders for large glove box). Verify sufficient cylinders are in inventory (lab and 1 st floor garage). 2. Set Working Maximum pressure to mbar and the Working Minimum Pressure to +5.0 mbar (see section 3.0). 3. Open the manual purge valve located on the back panel of the glove box beneath the antechambers. 4. Adjust the flow using the purge valve so that there is a steady flow of incoming gas and the box pressure is maintained at >1 mbar. 5. Create turbulence in the box to increase purge efficiency. This can be done by pressing in the gloves or using an electric fan or vacuum inside the glove box. 6. As the working gas cylinder empties, close the purge valve and change the cylinder. Open the valve on the cylinder and then re-open the purge valve. 7. Purge the system with the required number of cylinders or until Oxygen content is 25 ppm or less. Then close the manual purge valve. 8. The oxygen content might increase once the blower has been turned on (Main Screen -> Blower Control). If oxygen level rises above 100 ppm, stop the blower and continue to purge. 9. The blower system circulates the atmosphere through the purification system. It can be set to low speed during weighing, or set to high speed to recover from operator error. 10. Set the blower to 50% during normal operation. (100% = full blower speed) 5.0 Preforming a Pressure Test (Check Seal Quality) 1. Perform a pressure test anytime the front panel of the glove box is removed and replaced. A pressure test should also be performed periodically for safety and quality assurance. 2. Close purge valve located on back of glove box panel beneath antechamber. 3. Turn compressed cylinder on by rotating the valve on top of the cylinder counterclockwise. 4. Set Maximum Working pressure to 12.0 mbar and the Minimum Working pressure to 10.0 mbar. Page 5
6 5. Allow 10 minutes for the glovebox to stabilize at 11.0 mbar. Use the foot pedals to finely adjust the pressure. 6. Watch the current pressure in the glove box carefully for 25 minutes. If the pressure fluctuates more than 1 mbar, the system is not secured. Troubleshoot the seals/connections or contact Innovative Technology Inc. 6.0 Regenerating Purification Columns (Single-Column, offline) 1. Make sure that regen gas tank is sufficiently full. Column regeneration consumes approx. one-half of the tank. 2. Unless a second purification column is installed, the atmosphere of the glove box will be jeopardized so all experiments should be stopped and all chemical containers should be closed tightly. 3. Connect a cylinder of regenerating gas (3-7% Hydrogen in Nitrogen or Argon) to the regen gas connections (cylinder in left position if looking at them). Ensure connections are tightly secured. 4. Open the valve on top of the tank, allow the gas line to fill. Set the regulator pressure to 7 psi. 5. Turn off the blower on the main screen. 6. Press the Start button in the Regen Controls menu. The screen will prompt the user with two questions. 7. Are the column valves closed? Make sure that column valves are closed. These are located on top of the purification column inside the purification box. If the yellow valves are visible, the column valves are not closed. 8. Click Yes in the blue box next to the question. A click will be heard; the solenoid valves are opening to allow regen gas to flow. 9. Is Flow OK? At this point, the regen gas should have built up pressure and is ready to flow through the purification chamber. Adjust the needle valve on the front of the purification box to set the flow between scfh. The flow indicator should be held steady in between these two value. 10. Click Yes in the blue box next to the question. Now the regen gas will stop flowing and the flow indicator will drop to 0. The flow will automatically restart after the column is sufficiently heated (default = 3 hours). 11. The remainder of the column regeneration is automatically controlled by the glove box. The default time is 14.5 hours. At the end of the cycle, the blower will automatically turn back on. Page 6
Glove Box Installation Manual
Glove Box Installation Manual 1998 by M. Braun Company File: GB-UNI-INS.DOC! Edition 08-00 by M. Boutin! Subject to be changed without notice Glovebox Installation Your Glove box has been fully assembled,
More informationStandard Operating Procedure
Page 1 of 10 Vigor Glove Box Standard Operating Procedure Lab: ESB 154 Department: Materials Science and Engineering PI: Paul V. Braun Written By: Subing (Alan) Qu Section 1: Overview Type of SOP: Process
More informationUniversity of Minnesota Nano Center Standard Operating Procedure
University of Minnesota Nano Center Standard Operating Procedure Equipment Name: Controlled Atmosphere Glove Box Model: Labconco Protector Location: PAN 185 Badger Name: Not on Badger Revision Number:
More informationSchool of Chemistry SOP For Operation Of Glove Boxes
School SOP for Operation of Glove Boxes: The following SOP provides guidelines on how to adequately and safely operating a standard laboratory glove box and its associated devices. General Information:
More informationGlove Box with Gas Purification System and Digital Control VGB-6. MTI Corporation
Glove Box with Gas Purification System and Digital Control VGB-6 MTI Corporation 860 S 19 th Street, Richmond, CA 94804, USA Tel: 510-525-3070 Fax: 510-525-4705 E-mail: info@mtixtl.com Web site: www.mtixtl.com
More informationMB-SP Series Solvent Purification System
Operation Manual MB-SP Series Solvent Purification System Table of contents 1. System overview 2 2. Column activation.. 5 3. System purging.. 6 4. Glove box integration.... 7 5. Solvent reservoir installation....
More informationUsage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System)
Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System) Revision date October 2014 2 Emergency Plan for Nano Furnace Standard Operating Procedures for Emergencies Contact information Person
More informationChromat Calibration Updated October 27th, 2017
Chromat Calibration Updated October 27th, 2017 Calibrating the Chromatograph Performing the Chromat Calibration is highly recommended when starting a new well. The MLogger already has a default calibration
More informationOperating Instructions
Operating Instructions Before operating the thin film evaporator, please be aware of all safety concerns associated with this experiment: Burn hazard from the column and steam lines, Chemical hazards associated
More informationwith O 2 Controller Instruction Manual
14500 Coy Drive, Grass Lake, Michigan 49240 734-475-2200 E-mail: sales@coylab.com www.coylab.com Hypoxic Cabinets (In-Vivo / In-Vitro) with O 2 Controller Instruction Manual Index Page Warranty 2 Warnings
More informationMJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone:
MJB4 Mask Aligner Operating Procedure Effective Date: 07/12/2012 Author(s): Jiong Hua Phone: 402-472-3773 Email: jhua2@unl.edu 1 1 Introduction 1.1 Key Words Karl Suss MJB4 Mask Aligner, Optical Lithography,
More information256 Pneumatic Pressure Indicator
256 Pneumatic Pressure Indicator 51425699 Copyright 2002 Slope Indicator Company. All Rights Reserved. This equipment should be installed, maintained, and operated by technically qualified personnel. Any
More informationGETZ EQUIPMENT INNOVATORS RECOVERY SYSTEM CLEAN AGENT FE-36 / HALOTRON 1 PART NO. 4G59751 OPERATIONS MANUAL
GETZ EQUIPMENT INNOVATORS CLEAN AGENT FE-36 / HALOTRON 1 PART NO. 4G59751 OPERATIONS MANUAL GETZ EQUIPMENT INNOVATORS PEKIN, ILLINOIS, U.S.A. PHONE: (888) 747-4389 FAX: (309) 495-0625 Limited Warranty
More informationIMPORTANT CO2/ HPA AIR TANK SAFETY INSTRUCTION AND GUIDELINES. Tank valves must be installed or removed by qualified personnel.
!WARNING! IMPORTANT SAFETY INSTRUCTION AND GUIDELINS!WARNING! IMPORTANT CO2/ HPA AIR TANK SAFETY INSTRUCTION AND GUIDELINES GETTING STARTED This Paintball Marker is NOT A TOY. Misuse can cause serious
More informationAX5000 Operational Manual
MIYACHI AMERICA CORPORATION The World Leader in Hermetic Sealing Systems AX5000 Operational Manual 0 Document #107-00092-001 Dec, 2013 AX5000 Operational Manual Miyachi America Corporation 1820 S. Myrtle
More informationAcornVac Vacuum Plumbing Systems - Trouble Shooting Guide
AcornVac Vacuum Plumbing Systems - Trouble Shooting Guide 1. Accumulators Problem Accumulator is overflowing If Accumulator continues to overflow Correction 1. Push and hold the manual activation button
More informationIMPORTANT CO2/ HPA AIR TANK SAFETY INSTRUCTION AND GUIDELINES. Tank valves must be installed or removed by qualified personnel.
!WARNING! IMPORTANT SAFETY INSTRUCTION AND GUIDELINS!WARNING! IMPORTANT CO2/ HPA AIR TANK SAFETY INSTRUCTION AND GUIDELINES GETTING STARTED This Paintball Marker is NOT A TOY. Misuse can cause serious
More informationOperations Manual. Internal automatic Pressure control for orbital tube welding. By Exel Orbital Systems, Inc. Page
Operations Manual Internal automatic Pressure control for orbital tube welding. By Exel Orbital Systems, Inc. Page SG1 Overview and Dimensions 5.25 / 133 mm PS Power Supply Connection 4.06 / 103 mm Top
More informationTitle: Xactix XeF2 Etcher Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 03/23/2016
Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Xactix XeF2 Etcher. All users are expected to have read and understood this document.
More informationCalcium Reactor. CalcFeeder PRO. User Manual 1.1a. Model. CalcFeeder AC1 PRO CalcFeeder AC2 PRO CalcFeeder AC3 PRO CalcFeeder AC4 PRO
Calcium Reactor CalcFeeder PRO User Manual 1.1a Model CalcFeeder AC1 PRO CalcFeeder AC2 PRO CalcFeeder AC3 PRO CalcFeeder AC4 PRO Page 1 Dear Customer, Thank you for purchasing an Pacific Sun CalcFeeder
More informationUniversity of MN, Minnesota Nano Center Standard Operating Procedure
Equipment Name: Image Reversal Oven Badger name: ir-oven Revision #: 2 Model: YES 310 Revisionist: Paul Kimani Location: Bay 2 Date: October 29, 2013 1. Description The Yield Engineering Systems YES-310
More informationSomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings
SomnoSuite FAQ V1.3 January 2015 Setup 1. How do I connect the SomnoSuite to my oxygen source? 2. Is there a way to speed up the downward movement of the pusher block when setting the empty position? 3.
More informationSedan Range. Evaporative Loss System Check Clarified Procedure
SERVICE Sedan Range DATE 3/98 TECHNICAL BULLETIN Evaporative Loss System Check Clarified Procedure 05.1-29 MODEL 1996-97 MY Sedan (4.0L) Range VIN 772231-812255 SWB 776271-812255 LWB ISSUE: A mandatory
More informationInstruction Manual VCG-6
Specifications Amps / Volts requirements 400 ma @ 24 VDC Propane pressure (VCG-6LP) 11 WC / 2.8 kpa Natural Gas pressure (VCG-6NG) 4.5 WC / 1.15 kpa Cu ft per hour / CO2 3-6 SCFH BTU Rating (Variable)
More information14500 Coy Drive, Grass Lake, Michigan Vinyl Anaerobic Airlock Chamber Instruction Manual
14500 Coy Drive, Grass Lake, Michigan 49240 734-475-2200 E-mail: sales@coylab.com www.coylab.com Vinyl Anaerobic Airlock Chamber Instruction Manual 1.0 DESCRIPTION 1.3 CHAMBER ASSEMBLY 2.0 ERRORS MESSAGES
More informationInstallation Instructions
Page 5750-S-1 Installation Instructions General These mounting instructions for Circular INCINO- PAK Burners are in addition to the specific instructions offered for other Maxon component items: Shut-Off
More informationIn Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure
PPMS Service Note 1099-412 In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure Introduction: Loss of electricity to the PPMS EverCool II should not cause damage to
More informationJETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012
JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012 UNIVERSITY OF TEXAS AT ARLINGTON Nanofabrication Research and Teaching Facility TABLE OF CONTENTS 1. Introduction....2 1.1 Scope of Work.....2
More informationAPPS-160 Gen 2 User Manual
APPS-160 Gen 2 User Manual Applicable Products: PN 11914406 - APPS-160 10 to 50 PSIG operating pressure PN 11918618 - APPS-160 40 to 85 PSIG operating pressure Rev 9/16/2004 Chart Ind. www-chart-ind.com
More informationBasic & Controlled Atmosphere Glove Boxes
I N T R O D U C T I O N Precise Basic and Controlled Atmosphere Glove Boxes provide a leak-tight environment for work with contamination-sensitive materials. Precise Basic Glove Boxes are simple, economical
More informationEndo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax
OPERATION MANUAL Zutron Medical, LLC 17501 W 98 th St #40-27 Lenexa, KS 66219 Phone 877-343-5873 Fax 913-967-5944 ZUT-Lab-004-30004 REV. 03312017 Table of Contents 2 Introduction 1. Intended Use 2. Labels,
More informationSPUTTER STATION STANDARD OPERATING PROCEDURE
SPUTTER STATION STANDARD OPERATING PROCEDURE Purpose of this Instrument: This instrument is used for deposition of thin metal or oxide films. Source materials supplied by WVU Shared Research Facilities:
More informationRevised: June 7, 2017
LC Technologies Thermal Evaporator Standard Operating Procedure Faculty Supervisor: Prof. Robert White, Mechanical Engineering (x72210) Safety Office: Peter Nowak x73246 (Just dial this directly on any
More informationTrion PECVD SOP IMPORTANT: NO PLASTIC, TAPE, RESISTS, OR THERMAL PASTE ARE ALLOWED IN THE CHAMBER
Trion PECVD SOP IMPORTANT: NO PLASTIC, TAPE, RESISTS, OR THERMAL PASTE ARE ALLOWED IN THE CHAMBER CAUTION: THE CHAMBER PLATE GETS EXTREMELY HOT Start Up Procedure 1) Open bottle and regulator for Helium,
More informationNTF-515. Operation Manual
NTF-515 Operation Manual RTI Technologies, Inc. 10 Innovation Drive York, Pennsylvania 17402 USA Phone: 717-840-0678 Toll Free: 800-468-2321 Web-site: www.rtitech.com Manual P/N: 035-81474-00 (RevA) TABLE
More informationManual Solvent Purification System. Innovative Technology, Inc.
Innovative Technology, Inc. 2 New Pasture Road, Newburyport, MA 01950-4054 USA Email: Info@gloveboxes.com TEL: +1 978 462 4415 FAX: +1 978 462 3338 Manual Solvent Purification System Copyright Innovative
More informationSuperconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)
Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Sensitivity: 1x10 11 A m 2 Applied DC fields: 0 T to 5 T Applied AC fields: 0 G to 3 G (zero-to-peak), 0.01 Hz to 1000 Hz Temperatures
More informationWelker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual
Installation, Operation, and Maintenance Manual Welker Sampler Model GSS-1 The information in this manual has been carefully checked for accuracy and is intended to be used as a guide to operations. Correct
More informationUsage Policies Notebook for Trion RIE / ICP Dry Etch
Usage Policies Notebook for Trion RIE / ICP Dry Etch Revision date September 2014 2 Emergency Plan for Trion RIE/ICP Dry Etch Standard Operating Procedures for Emergencies Contact information Person Lab
More informationInstructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems
Corporation Issued: 5/09 P/N: 600104 Rev. 2 Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems The Gas Addition Accessory permits the
More informationSolvent System Walkthrough
v2.0 1 Solvent System Walkthrough The Contour Glass solvent system is designed to give you air-free anhydrous solvent with minimal effort. However, the system is only as good as its users so please read
More informationPROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road
PROPORTIONING VALVE Model 150 INSTRUCTION MANUAL March 2017 IMS Company 10373 Stafford Road Telephone: (440) 543-1615 Fax: (440) 543-1069 Email: sales@imscompany.com 1 Introduction IMS Company reserves
More informationSeries IRV1000/2000/3000
CAT.ES60-16 A Regulator Series /2000/3000 Allows adjustment of vacuum line pressure Regulator Series /2000/3000 3 sizes offered in the series Variations have been expanded to three sizes from only one
More informationAppendix: Simplified Instructions for Operating the HP 5890 GC/FID and Air Sampling Systems
Appendix: Simplified Instructions for Operating the HP 5890 GC/FID and Air Sampling Systems The bucket brigade air sampler. For the square Tedlar bags with the push/pull valves: Put the valve of a tedlar
More informationKarl Suss MJB4 Mask Aligner
Karl Suss MJB4 Mask Aligner Tool Manager: Yong Sun ( yongs@princeton.edu; Office 8-8234; Cell 609-917-5076 ) Backup: George Watson ( gwatson@princeton.edu; Office 8-4626; Cell 732-996-2713 ) ******************************************************************************
More informationFLUSHMATE III FLUSHMATE FLUSHOMETER - TANK SYSTEM. Owner s Service Manual. 503 Series. 503 Series
Owner s Service Manual 503 Series FLUSHMATE III FLUSHOMETER - TANK SYSTEM 503 Series FLUSHMATE A Division of Sloan Valve Company 30075 Research Drive New Hudson, MI 48165 800-533-3450 248-446-5300 http://www.flushmate.com
More informationRejuvenation Instructions
Rejuvenation Instructions #401 Air Systems UPR This NRI covers the following: Understanding the applications and operation of flow meters. Understand the application and operation of test pressure gauges.
More informationINSTRUCTION MANUAL FOR POLYKEG DISPOSABLE KEG
TABLE OF CONTENTS OF MANUAL Chapters 1.0 Product description 2.0 Safety rules for the utilisation of the keg 3.0 Instructions for manual filling 4.0 Instructions for automatic filling 5.0 Instructions
More informationBasic ICP Operating Procedures
Center for High Technology Materials 2 February, 2009 University of New Mexico Created by Beth Fuchs Basic ICP Operating Procedures INTRODUCTION: The ICP is an inductively coupled plasma etching system,
More information! WARNING! IMPORTANT HPA AIR TANK SAFETY INSTRUCTION AND GUIDELINES ! WARNING! IMPORTANT SAFETY INSTRUCTION AND GUIDELINES
! WARNING! IMPORTANT SAFETY INSTRUCTION AND GUIDELINES! WARNING! IMPORTANT HPA AIR TANK SAFETY INSTRUCTION AND GUIDELINES This Paintball Marker is NOT A TOY. Misuse can cause serious injury or death. It
More informationSSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA
Page 1 of 9 SSI Solaris 150 RTA The Solaris 150 RTA is a rapid thermal annealing system capable of handling sample sizes up to 100mm (4 diameter) or smaller. The system can anneal in N 2 and Forming gas
More informationOperating Procedures for Metal Evaporator I
Operating Procedures for Metal Evaporator I Metal Evaporator I is intended as a tool and a training device. Understanding the operation of this equipment should give you a basic knowledge of vacuum and
More informationWARNING OLED & BASIC MENU QUICK START
Getting Started QUICK START WARNING The VANGUARD MARKER is not a toy. Careless or improper use, including failure to follow instructions and warnings within this Operator Manual and attached to the VANGUARD
More informationMINIBRUTE ANNEAL TUBE STANDARD OPERATION PROCEDURE
Arizona State University NanoFab MINIBRUTE ANNEAL TUBE STANDARD OPERATION PROCEDURE Rev A Title: MINIBRUTE ANNEAL TUBE STANDARD OPERATION PROCEDURE Table of Contents Issue: Rev A Contents Table of Contents...1
More informationStandard Operating Procedure Inert Vacuum Line. Ryan Mulvenna, May 2012
Standard Operating Procedure Inert Vacuum Line Ryan Mulvenna, May 2012 Description of Process The inert vacuum lines located in the fume hoods one (Conjugated Polymer Synthesis: CPS) and two (ATRP polymer
More informationUsing the Akta Prime plus October 22, 2012
Some starting precautions: 1. Vacuum filter all buffers. Removes any large particles/debris that may clog your column De-gases the buffers 2. Clarify lysates first by centrifugation and then filtration
More informationCombination Air Valve
Combination Air Valve For Sewage and Wastewater Model C50 Installation, Operation and Maintenance Manual (IOM) Table of Contents General... Page 2 Safety... Page 2 Operational Data... Page 3 Materials
More informationInert Air (N2) Systems Manual
INSTRUCTION MANUAL Inert Air (N2) Systems Manual N2-MANUAL 2.10 READ AND UNDERSTAND THIS MANUAL PRIOR TO OPERATING OR SERVICING THIS PRODUCT. GENERAL INFORMATION Positive pressure nitrogen gas pressurizing
More informationBirck Nanotechnology Center
EFFECTIVE DATE: October 20, 2016 PAGE 1 of 8 This instruction covers the set-up and use of the bench-top for annealing and oxidizing samples within the cleanroom. 1. SAFETY REQUIREMENTS 1.1 Safety glasses
More informationSyringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT
ATTENTION SYRINGE REPLACEMENT Please read through the document completely before starting any repairs. Refer to the proper section in the service manual for complete removal and replacement procedures.
More informationApproved by Principal Investigator Date: Approved by Super User: Date:
Approved by Principal Investigator Date: Approved by Super User: Date: Standard Operating Procedure BNC Commonwealth Dual Ion Beam Deposition System (CDIBS) Version 2010 February 14 I. Purpose This Standard
More informationOPERATION MANUAL NTF-15
OPERATION MANUAL NTF-15 Nitrogen Tire Filling Valve Stem Caps (Qty=200) Order P/N 436075 RTI Technologies, Inc 10 Innovation Drive York, PA 17402 800-468-2321 www.rtitech.com 035-81235-00 (Rev B) TABLE
More informationPressure Dump Valve Service Kit for Series 2300 Units
Instruction Sheet Pressure Dump Valve Service Kit for Series 00 Units. Overview The Nordson pressure dump valve is used to relieve hydraulic pressure instantly in Series 00 applicator tanks when the unit
More informationInstallation Instructions
Model: ACT-4100 ACT-4100-3 ACT-4100-6 SUN ELECTRIC CORPORATION Page: 1 of 5 Installation Instructions INSTALLATION MUST BE PERFORMED BY QUALIFIED SUN PERSONNEL ONLY INSTALLATION OVERVIEW: The Installation
More informationUser Manual. Quantos Automated Dosing Powder Module
User Manual Powder Module 1 Introduction Thank you for purchasing the Quantos dosing system. Quantos is based on the renowned and reliable XPE series balances from METTLER TOLEDO. Besides the highly accurate
More informationOXFORD PLASMALAB 80PLUS (CLOEY)
Arizona State University NanoFab OXFORD PLASMALAB 80PLUS (CLOEY) Rev D Table of Contents Contents Table of Contents...1 1. Purpose / Scope...2 2. Reference Documents...2 3. Equipment / Supplies / Material...2
More informationOPERATION MANUAL NTF-60 Plus
OPERATION MANUAL NTF-60 Plus Nitrogen Tire Filling Valve Stem Caps (Qty=200) Order P/N 436075 RTI Technologies, Inc 10 Innovation Drive York, PA 17402 800-468-2321 www.rtitech.com 035-81264-00 (Rev A)
More informationUM-010 Invivo User Manual
UM-010 Invivo 2 300 User Manual Contents 1. Overview 5 1.1 Safety instructions 5 1.2 Regulatory compliance 6 1.3 Symbols 6 1.4 Installation and relocation 7 1.5 Weight and dimensions 8 2. Gas and electrical
More informationArizona State University Center for Solid State Electronics Research Issue: E Title: Heat Pulse 610 Operating Procedure Page 1 of 7
Title: Heat Pulse 610 Operating Procedure Page 1 of 7 Table of Contents 1.0 Purpose/Scope... 2 2.0 Reference Documents... 2 2.1 Heat Pulse Operations Manual... 2 2.2 Solaris software Operations Manual...
More informationELECTRONIC PLC SERIES Test Benches
Whether you are producing a single assembly or a production run, CustomCrimp has the test equipment to give both you and your customer the assurance that the product has been assembled correctly. A programmable
More informationFAULT CODE TROUBLESHOOTING INDEX
FAULT CODE TROUBLESHOOTING INDEX 1. Display indicates Change Filters 2. Display indicates Drip Tray Full Continuous Alarm will Sound 3. Display indicates Cold Fault 4. Display indicates Hot Fault 5. Display
More informationuser manual .68 CALIBER ELECTRONIC PAINTBALL MARKER Meets or exceeds ASTM standards.
user manual TM TM TABLE OF CONTENTS: Page Topic(s) 1 Warning and Rules Safe Marker Handling 2 Warranty Information 3 Welcome 4 Initial Assembly Battery Installation 9 Attaching Propellant Source Attaching
More informationP5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.
Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 TestEquipmentDepot.com P5513 Pneumatic Comparison Test Pump Users Manual PN 3963372 November 2010 2010 Fluke Corporation. All
More informationAmmonia Treatment Instructions Page 1 of 5
Page 1 of 5 Original version by Douglas Call Updated January 21, 2009 by David Jones Department: Chemical Engineering Location: Fenske 102 (access thru room 103) Contact: BP Matthew Holbrook 161 Fenske
More informationNanofabrication Facility: PECVD SOP Rev. 00, April 24
Author: Charlie Yao & Mario Beaudoin Email: charlieyao@gmail.com; Beaudoin@physics.ubc.ca Phone: 604-822-1853(MB). Purpose This document outlines the standard operation for the Trion Plasma Enhanced Chemical
More informationSmooth Vent Valve. Series XVD XL XL Q D- XVD XGT CYV
Smooth Vent Valve Series Valve / needle valve integrated construction requires only 1/4 the piping space of previous models. Particulates significantly reduced through the use of a metal diaphragm in the
More informationStandard Operating Manual
Standard Operating Manual ARC12M Sputter Copyright 11.2015 by Hong Kong University of Science & Technology. All rights reserved. Page 1 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness
More informationUser Instruction Manual
User Instruction Manual 4500 psi Air Compressor Ver 2, 1.18 Contents Parts Included...3 Assembly Instructions...3-5 Operation Instructions...6-7 Oil Change Intervals...8 Air Filter Replacement...9 Setting
More information3M Liqui-Cel EXF-10x28 Series Membrane Contactor
Membrane Contactors 3M Liqui-Cel EXF-10x28 Series Membrane Contactor Assembly and Disassembly Instructions 3M.com/Liqui-Cel TABLE OF CONTENTS I. Safety and Warning 3 II. Assembly Parts 4 III. Part Orientation
More informationMANUAL BE SERIES Test Benches
The CustomCrimp Manual BE Series Test Benches are designed with features that make proof and burst testing of hydraulic hose assemblies a quick and easy procedure. CUSTOMIZED AND SPECIAL DESIGN BENCHES
More information5.1.3 Mechanical Hazards Drive assemblies have sufficient power to cause injury. Keep hands, fingers, clothing and tools clear of moving parts.
Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the PE4400. All users are expected to have read and understood this document. It is
More informationCover Page for Lab Report Group Portion. Flow Visualization in a Water Channel
Cover Page for Lab Report Group Portion Flow Visualization in a Water Channel Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 08 September 2017 Name 1: Name 2: Name 3: [Name
More informationStandard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300
Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300 The Thermo Scientific icap 6300 Inductively Coupled Plasma Optical Emission
More informationPressure Dump Valve Service Kit for Series 3000 Units
Instruction Sheet Pressure Dump Valve Service Kit for Series 000 Units. Overview The Nordson pressure dump valve is used to relieve hydraulic pressure instantly in Series 00, 400, 500, and 700 applicator
More informationSINORIX Valve Reconditioning
Document No. 129-097 SINORIX Valve Reconditioning Product Description These instructions explain how to recondition and fill 1-inch through 4-inch SINORIX valves. Product Numbers Verify that the appropriate
More informationNorton Group Toepler Pump Operating Instructions. Version 3.1
Norton Group Toepler Pump Operating Instructions Version 3.1 I. Startup 1 Take the oil diffusion pump out of line. Valves A and H should be closed but connected to dynamic vacuum. 2 Attach a collection
More informationACCUMULATOR OPERATING & MAINTENANCE INSTRUCTIONS
ACCUMULATOR OPERATING & MAINTENANCE INSTRUCTIONS READ ALL INSTRUCTIONS PRIOR TO INSTALLATION AND OPERATION TO AVOID POSSIBLE INJURY Warning: Always consider any accumulator to contain pressure until proven
More informationB. Assemble the Method-25 sample train with a heated probe/filter.
6/30/2015 VOC Reporting, Inc. Method 25 Heated Controller Operating Instructions. pg. 1 of 5 INSTRUCTIONS FOR EPA METHOD-25 FIELD SAMPLE COLLECTION USING A VRi HEATED PROBE AND FILTER ASSEMBLY Introduction:
More informationPerforma Cv Twin Alternating and High Flow Systems. Manual Supplement
Performa Cv Twin Alternating and High Flow Systems Manual Supplement Table of Contents 1.0 Installation and Start-Up.............. 3 Water Line Connection Brine Tank Turbine Connection Connecting Manifold
More informationSafety, Operating and Maintenance Manual. TRAIN Rat DS9-TRAIN
Safety, Operating and Maintenance Manual TRAIN Rat DS9-TRAIN Industry s first air powered entry tool! Quickly, quietly and reliably opens passenger train doors! Introducing the TRAIN Rat. This innovative
More informationProcedure for Operating Columbia (Nevis) LHe Electron Bubble Chamber Cryostat. Version 5.1
Project: Novel Electron Bubble Particle Detector Procedure for Operating Columbia (Nevis) LHe Electron Bubble Chamber Cryostat Version 5.1 (Using side filling line) Hand Processed Changes HPC No. Date
More informationUsage Policies Notebook for Xenon Difluoride (XeF 2 ) Isotropic Si Etch
Usage Policies Notebook for Xenon Difluoride (XeF 2 ) Isotropic Si Etch Revision date September 2014 2 Emergency Plan for XeF 2 Si Etcher Standard Operating Procedures for Emergencies Contact information
More informationPCS 340 Core Saturator : 115 Volt : 230 Volt. Instruction Manual
PCS 340 Core Saturator 127-70: 115 Volt 127-70-1: 230 Volt Instruction Manual Updated 4/30/2015 Ver. 1.2 OFI Testing Equipment, Inc. 11302 Steeplecrest Dr. Houston, Texas 77065 U.S.A. Tele: 832.320.7300
More informationCover Page for Lab Report Group Portion. Head Losses in Pipes
Cover Page for Lab Report Group Portion Head Losses in Pipes Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 February 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section
More informationUNIVERSITY OF ROCHESTER ENVIRONMENTAL HEALTH & SAFETY
Revision No.: 1 Page 1 of 6 I. PURPOSE This policy/procedure establishes the proper steps for set up, and operation of the BullEx I.T.S Xtreme training system. II. PERSONNEL AFFECTED Fire Safety staff
More informationUser's Manual. MixRite TF 10. Edition 05.08
User's Manual MixRite TF 10 Edition 05.08 1 Tefen MixRite TF 10 fertilizer and chemicals Injector Congratulations on your purchase of one of Tefen s high quality products. To get the best results from
More informationTECHNICAL INFORMATION
TECHNICAL INFORMATION Installation, Operation, and Maintenance Balston 75-96 and 75-98 Nitrogen Generation System Tower Pressure Gauge 5'-1" 3'-6" 7'-4" 6'-10 1/2" PLAN VIEW Figure 1 - Balston 75-96 and
More informationNTF-230 OPERATION MANUAL
NTF-230 OPERATION MANUAL MAHLE Aftermarket Inc., Service Solutions 10 Innovation Drive York, Pennsylvania 17402 USA Phone: 717-840-0678 Toll Free: 800-468-2321 Web-site: www.servicesolutions.mahle.com
More informationGUIDELINES FOR CONDUCTING HYDROGENATION REACTIONS
Page 1 of 5 GUIDELINES FOR CONDUCTING 1. Definition of Hydrogenation and Hydrogenolysis Reactions Hydrogenation reactions are synthetic transformations involving hydrogen, metallic catalysts (such as nickel,
More informationIs there any visible damage to the machine? If yes, describe damage:
CH-MG Illustrated Start-Up Procedure, Rev 2A Pre-Start-Up Checks Is there any visible damage to the machine? If yes, describe damage: Is the machine installed outdoors? If outdoors, is the machine on the
More information