MASSACHUSETTS STATE POLICE CRIME LABORATORY. Certificate of Calibration Procedure for the Alcotest 9510

Size: px
Start display at page:

Download "MASSACHUSETTS STATE POLICE CRIME LABORATORY. Certificate of Calibration Procedure for the Alcotest 9510"

Transcription

1 MASSACHUSETTS STATE POLICE CRIME LABORATORY Certificate of Calibration Procedure for the Alcotest 9510 Effective Date: May 10, 2017 Page 1 of 23

2 TABLE OF CONTENTS 1 INTRODUCTION ABBREVIATIONS EQUIPMENT REAGENTS AND STANDARDS PREPARATION ADJUSTMENT PROCEDURE CALIBRATION QUALITY CONTROL TEST (0.080% BREATH HOSE) SET BT CERTIFICATION DATE POST-CALIBRATION REVISION HISTORY MANUAL REVIEW AUTHORIZATION Effective Date: May 10, 2017 Page 2 of 23

3 1 INTRODUCTION 1.1 The Draeger Alcotest 9510 breath test instrument (BT) requires an annual certification. This is to be performed by the Office of Alcohol Testing (OAT). 2 ABBREVIATIONS 2.1 OAT: Office of Alcohol Testing 2.2 BT: Draeger Alcotest 9510 Breath Test instrument 2.3 IR: Infrared 2.4 EC: Electrochemical (aka Fuel Cell) 2.5 QC: Quality Control 3 EQUIPMENT 3.1 Alcotest Guth 2100 Simulator, Alcosim Simulator, Draeger Alcotest CU34 Simulator or equivalent 3.3 Dry Gas Regulator Clear tubing Adaptor 3.4 Keyboard 3.5 Printer with USB connection 3.6 CAT 5 Cable 3.7 Barometer 3.8 Draeger Alcotest 9510 USB Access Key Effective Date: May 10, 2017 Page 3 of 23

4 3.9 Phillips Head Screwdriver 3.10 Short Silicone Hose 3.11 NIST traceable clock 4 REAGENTS AND STANDARDS 4.1 Reagents (equivalents may be substituted) % Purchased Ethanol Solution % Purchased Ethanol Solution % and 0.100% OAT Manufactured Ethanol Solution, see OAT Solution Protocol 4.2 Standards (equivalents may be substituted) % Dry Gas cylinder, Intoximeters % Dry Gas cylinder, Intoximeters % Dry Gas cylinder, Intoximeters % Dry Gas cylinder, Intoximeters PREPARATION 5.1 Adjustment and Verification Solutions Fill simulators with approximately 500 ml of the designated OAT manufactured or purchased ethanol solution and check for air leaks The simulators are used in tandem. The solutions contained in each pair of simulators must be from the same lot number Turn on all simulators for at least thirty minutes to prior to testing. 5.2 BT Certification Form (OAT-F008) After solutions, dry gas cylinders, and/or simulators are replaced; verify the lot numbers, expiration dates, model numbers, and serial numbers of all the Effective Date: May 10, 2017 Page 4 of 23

5 solutions, all the dry gas cylinders, and all the simulators at the appropriate stations on the Form. This shall be properly documented on the Lot Number chart Print the Form for the appropriate Police Department or Agency and the appropriate station where the certification will take place Station 1 Certification Form Station 2 Certification Form BP/OAT/BATMOBILE Station 1 Certification Form BP/OAT/BATMOBILE Station 2 Certification Form Document on the Form whether the BT that has been received from the police department or agency has arrived with a top cover, side cover, gas enclosure key, power cord, keyboard, scanner with USB cord, printer USB cord, and CAT All printouts generated during the certification process are retained with the Form for review. 5.3 Removal of Stand and Gas Enclosure (if applicable) Stand Remove gas enclosure cover with the gas enclosure cover key If the police department brought the gas enclosure cover key, leave the key in the keyhole of the enclosure cover to ensure the return of the key If the stand is attached to the BT, remove the bottom cylinder (cylinder 2) to access and remove the two screws Remove the side panel from BT Unplug all cords and cables from the BT Push instrument back before lifting off the stand Gas Enclosure Unplug the two cylinder hoses and the transducer cable Remove the five screws on the bottom of the enclosure to remove the BT from the enclosure. Effective Date: May 10, 2017 Page 5 of 23

6 5.4 Instrument Set-Up At the laboratory bench station printed on the Form, connect the printer, keyboard, CAT 5 cable, and power cord to the BT and power on Insert the Draeger Alcotest 9510 USB Access Key Change the Police Department or Agency network to the OAT Network if the network was not changed through Initial Set-Up Procedure for Draeger Alcotest Use the appropriate IP address, subnet mask, and gateway assigned to that specific laboratory bench station Enter Menu Enter Settings Enter Network Type in the IP address, Subnet Mask, and Gateway Select Save and Press OK. BT will restart to save the settings Remove the black rubber cuvette cap from the Cuvette Inlet (Diagram 1). Diagram 1 Draeger 9510: Back of Instrument Clear Hose Simulator to Cuvette Inlet Effective Date: May 10, 2017 Page 6 of 23

7 6 ADJUSTMENT PROCEDURE 6.1 Database Management Enter Menu Enter Settings Enter Database Management Check Evid90-24 Number of records and record the number in the box on the Form as well as for the 90B, PC, and Training Tests If greater or equal to 150 tests for any of the categories, verify and document all records have sent to CJIS server. When confirmed, delete records Less than 150 tests, check Evid90B in dropdown and follow same process as Follow this process for PC, Train, and Periodic Test Press Cancel when finished. 6.2 Set Date and Time Enter Set Date and Time Set the date and time using a NIST traceable clock Save settings, then x to leave menu. 6.3 Check Internal Printer Paper Press the black button on the paper flap of the BT to open the Internal Printer to check if the paper needs replacement If the paper does not need to be replaced, continue to Check Firmware Version If the paper needs to be replaced, remove the current paper roll and replace with a new roll Ensure the paper is fed through the compartment of the paper flap. 6.4 Check Firmware Version Enter Menu. Effective Date: May 10, 2017 Page 7 of 23

8 6.4.2 Enter About Verify the M16, WinCE, and Cfg Versions with the Form Note: If the version is not correct, the updated version will need to be downloaded. Follow Initial Set-Up Procedure for Draeger Alcotest 9510 (OAT-D004) to update the firmware Print, then x to leave the menu. 6.5 Update Barometric Pressure Enter Menu Enter Maintenance Enter Calibration Enter Ambient Pressure Correction Record BT Ambient Pressure readingas Initial BT Ambient Pressure on the Form Record the MKS Baratron Pressure Sensor reading as External Ambient Pressure, and circle which MKS Baratron Pressure Sensor is being used on the Form Calculate the acceptable range from the MKS Baratron Pressure Sensor, truncating to one decimal place. Acceptable range is determined as +/- 0.25% of the MKS Baratron Pressure Sensor reading. Record this range on the Form as Acceptable Range Determine if BT Ambient Pressure is within the acceptable range of the MKS Baratron Pressure Sensor Since the first decimal is not known on the BT reading, then assume that it is a zero If the BT Ambient Pressure is within the acceptable range of the MKS Baratron Pressure Sensor, then no adjustment is necessary on the BT. If the BT Ambient Pressure is not within the acceptable range, then adjust the External Ambient Pressure to the MKS Baratron Pressure Sensor reading and record on the form, truncating the decimal places Record on the Form the Final BT Ambient Pressure Save settings Effective Date: May 10, 2017 Page 8 of 23

9 Cancel to exit screen, then x to leave menu. 6.6 Adjustment (0.100% Wet) Once instrument is Ready, enter Menu Enter Maintenance Enter Calibration Procedure Enter assigned username (no password), then OK Verify your user information. If information is correct, press OK If information is incorrect, answer No to the prompt and make the appropriate edits to the operator database and then send the operator database to BT Follow the screen prompts along with the Form to enter information including, but not limited to: adjustment and adjustment check lot numbers and solution concentrations with expiration dates, and simulator serial and model numbers with expiration dates that are connected to the BT. Document all information on the Form Verify if entered information is correct on the screen. If it is not correct, then double click on the incorrect information and make the necessary changes. If the entered information is correct, then enter Save On the Form, record the temperature of the simulators that will be connected to the BT Attach 0.100% purchased solution tandem simulators. Connect the short silicone hose to the Cuvette Inlet (Diagram 1) and insert the adaptor connected to the clear hose to the Simulator Standard Pump Outlet Port (Diagram 1) Initiate adjustment process. 6.7 Adjustment Check (0.100% Wet) When BT screen prompts for the calibration check, remove simulators and attach 0.100% OAT manufactured solution tandem simulators. Connect the short silicone hose to the Cuvette Inlet (Diagram 1) and insert the adaptor connected to the clear hose to the Simulator Standard Pump Outlet Port (Diagram 1) Initiate adjustment check process. Effective Date: May 10, 2017 Page 9 of 23

10 6.7.3 Printout appears when the procedure is complete. Review the report to determine if results are within the following acceptable tolerances: The IR and EC results must fall between 0.097%-0.103% alcohol content The IR and EC results (n=6) must agree within +/ alcohol content of each other If the results are within the acceptable tolerances, remove simulators from the BT and proceed to Verification Test and Manual Adjustment (6.8) Repeat the adjustment (6.6) and adjustment check (6.7) once if the adjustment check results are outside the acceptable tolerances. If the results still do not meet acceptance tolerances, consult with a supervisor to determine if further testing is required or if the BT should be set aside until a later date and the Certification repeated. If the results are still unacceptable, mark the instrument as Out of Service. 6.8 Verification Test and Manual Adjustment (0.100% Dry, Inlet 1) Activate Cylinder Enter Menu Enter Maintenance Enter Set Active Cylinder Set Cyl 1 Active OK Insert 0.100% dry gas cylinder adaptor hose into Dry Gas Inlet 1 (Diagram 2) of BT. Effective Date: May 10, 2017 Page 10 of 23

11 Clear Hose of Dry Gas Inlet 1 Diagram 2 Draeger 9510: Back of Instrument Enter Menu Enter Maintenance Enter Verification Test Enter assigned username (no password), then OK Follow screen prompts along with the Form to enter information including dry gas concentration, lot number, and expiration date. Document all information on the Form Verify if entered information is correct on the screen. If it is not correct, then double click on the incorrect information and make the necessary changes. If the entered information is correct, then enter Save. Effective Date: May 10, 2017 Page 11 of 23

12 6.8.8 When prompted, set Test-Gas to Dry and Gas-Input to Cal Gas Inlet 1 then Save Set Number of Tests to 1 then Save Review the Verification Report to determine if the results are within the following acceptable tolerances: The IR and EC results must fall between 0.099%-0.101% alcohol content The IR and EC results must agree within +/ alcohol content of each other If the results are within the acceptable tolerances, remove adaptor hose for cylinder for Dry Gas Inlet 1 (Diagram 2) and proceed to the next verification test (6.9) If the results of the 0.100% Verification Test are outside the acceptable tolerances perform the following manual adjustments: Answer No to Perform another verification test Enter Menu Enter Maintenance Enter Calibration Enter EC Sensor Configuration Document the EC dry gas offset Initial value number on the Form. If the IR and EC results do not match, adjust the EC Drygas Offset to match the IR result. For example, if the EC result is 0.098%, IR result is 0.101% and the EC Drygas Offset is 6, the EC Dry gas Offset would need to be changed to 9 (0.099, 0.100, and is a difference of 3, 6 plus 3 is 9). Record the changed number on the Form as the EC drygas offset Final. Save the changes Enter IR/EC Correction Document the Calgas Inlet Initial value on the Form. If the IR result is not 0.100%, adjust Drygas % for Calgas Inlet. For example, if the IR result is 0.101% and the Drygas % is set to 3, the Drygas % would need to be changed to 2 (0.101% down to 0.100% is 1, 3 minus 1 is 2). Record the Dry gas % for Calgas Inlet Final number on Form. Save the Effective Date: May 10, 2017 Page 12 of 23

13 changes Repeat the above steps for 0.100% Verification Test Review the Verification Report to determine if the results are within the acceptable tolerances (0.099%-0.101%) If the results are acceptable, proceed to the 0.400% Verification Test (6.9) If the results are not within acceptable tolerances, repeat the adjustment process one more time. If the results still do not meet the acceptance tolerances, consult with a supervisor to determine if further testing is required or if the BT should be set aside for a later date and repeat the Certification Remove cylinder from Dry Gas Inlet 1 (Diagram 2). 6.9 Verification Test and Manual Adjustment (0.400% Wet) Attach 0.400% purchased solution tandem simulators. Connect the short silicone hose to the Cuvette Inlet (Diagram 1) and insert the clear hose with adaptor to the Simulator Standard Pump Outlet Port (Diagram 1) Answer Yes to the prompt Perform another verification test Follow the screen prompts along with the Form to enter information including, but not limited to, solution lot number and solution concentration with expiration date, and simulator serial and model number with expiration date that are connected to the BT. Document all information on the Form Verify if entered information is correct on the screen. If it is not correct, then double click on the incorrect information and make the necessary changes. If the entered information is correct, then enter Save On the Form, record the temperature of the simulator connected to the BT When prompted, set Test-Gas to Wet and Gas-Input to Cuvette Inlet then Save Set Number of Tests to 1 then Save Review the Verification Report to determine if the results are within the following acceptable tolerances: The IR and EC results must fall between 0.392%-0.408% alcohol content. Effective Date: May 10, 2017 Page 13 of 23

14 The IR and EC results must agree within +/ alcohol content of each other If the results are within the acceptable tolerances, continue to the Calibration (7.0) If the results of the 0.400% Verification Test are outside the acceptable tolerances perform the following manual adjustments: Answer No to the prompt Perform another verification test Enter Menu Enter Maintenance Enter Calibration Enter EC Sensor Configuration On the Form, document current EC Quadratic Correction Factor as the initial value. If the value is a zero or a negative number, enter 1 on the Form Press Cancel to Exit the screen Enter IR/EC Correction On the Form, document current IR Slope Multiplier as the Initial value Using the Draeger Quadr computer program; enter IR Slope Multiplier in the quadratic IR field, EC Quadratic Correction Factor in the quadratic EC field, Result IR and Result EC Enter Calculate tab On the Form, document the New IR-corr. in Final IR Slope Multiplier and New EC-corr. in Final EC Quadratic Correction Factor Enter the new EC Quadratic Correction Factor and IR Slope Multiplier in the proper screens in the BT and save settings Repeat steps for 0.400% Verification Test If results are acceptable, proceed to the Calibration (7). If the desired result is still not obtained, consult with a supervisor to determine if further testing is required or if the BT should be set aside until a later date and repeat the Certification. Effective Date: May 10, 2017 Page 14 of 23

15 7 CALIBRATION Remove simulators from the BT. 7.1 Calibration 0.200% Dry Gas Insert 0.200% dry gas cylinder hose adaptor into Dry Gas Inlet 1 (Diagram 2) of BT Answer Yes to the prompt Perform another verification test Follow screen prompt requests along with the Form to enter information including concentration, dry gas lot number, and expiration date Verify if entered information is correct on the screen. If it is not correct, then double click on the incorrect information and make the necessary changes. If the entered information is correct, then enter Save When prompted, set Test-Gas to Dry and Gas-Input to Cal Gas Inlet 1 then Save Set the number of tests to ensure that at a minimum three tests are completed Review the Verification Report to determine results are within the following acceptable tolerances: The IR and EC results must fall between 0.192%-0.208% alcohol content The IR and EC results (n=6) must agree within +/ alcohol content of each other If the results are within the acceptable tolerances, proceed to the next calibration (7.2) test If the results are not within acceptable tolerances, repeat the 0.200% calibration. (7.1) If the results still do not meet acceptance tolerances, consult with a supervisor to determine if further testing is required or if the BT should be set aside until a later date and repeat the Certification Remove cylinder from Dry Gas Inlet 1 (Diagram 2). 7.2 Calibration 0.080% Dry Gas Effective Date: May 10, 2017 Page 15 of 23

16 7.2.1 Repeat the steps for the above 0.200% verification using a 0.080% cylinder Review the Verification Report to determine if the results are within the following acceptable tolerances: The IR and EC results must fall between 0.074%-0.086% alcohol content The IR and EC results (n=6) must agree within +/ alcohol content of each other If the results are within the acceptable tolerances, proceed to the next calibration (7.3) Because the next calibration concentration uses a different inlet, answer No to Perform another verification test. Remove cylinder from Dry Gas Inlet 1 (Diagram 2) If the results are not within acceptable tolerances, repeat the 0.080% calibration (7.2) If the results still do not meet acceptance tolerances, consult with a supervisor to determine if further testing is required or if the BT should be set aside until a later date and repeat the Certification. 7.3 Calibration 0.020% Dry Gas Activate Cylinder Enter Menu Enter Maintenance Enter Set Active Cylinder Set Cyl 2 Active OK Insert 0.020% dry gas cylinder hose into Dry Gas Inlet 2 (Diagram 3) of BT. Effective Date: May 10, 2017 Page 16 of 23

17 Clear Hose of Dry Gas, Inlet 2 Diagram 3 Draeger 9510: Back of Instrument Enter Menu Enter Maintenance Enter Verification Test Enter assigned username (no password), then OK if information is correct Follow screen prompts along with the Form to enter information including concentration, dry gas lot number, and expiration date Verify if entered information is correct on the screen. If it is not correct, then double click on the incorrect information and make the necessary changes. If the entered information is correct, then enter Save When prompted, set Test-Gas to Dry and Gas-Input to Cal Gas Inlet 2 then Save Set the number of tests to ensure that at a minimum, three tests are completed Review the Verification Report to determine if the results are within the following acceptable tolerances: Effective Date: May 10, 2017 Page 17 of 23

18 The IR and EC results must fall between 0.016%-0.024% alcohol content The IR and EC results (n=6) must agree within +/ alcohol content of each other If the results are within the acceptable tolerances, proceed to the Quality Control Test (8) If the results are not within acceptable tolerances, repeat the 0.020% calibration (7.3) If the results still do not meet acceptance tolerances, consult with a supervisor to determine if further testing is required or if the BT should be set aside until a later date and repeat the Certification Remove cylinder from Dry Gas Inlet 2 (Diagram 3). 8 QUALITY CONTROL TEST (0.080% BREATH HOSE) 8.1 Attach 0.080% OAT manufactured solution tandem simulators. Connect the clear hose in the Simulator Standard Pump Outlet Port (Diagram 1). 8.2 Answer Yes to the prompt Perform another verification test. 8.3 Follow the screen prompts along with the Form to enter information including, but not limited to, solution lot number and solution concentration with expiration date, and simulator serial and model number with expiration date that are connected to the BT. Document all information on the Form Verify if entered information is correct on the screen. If it is not correct, then double click on the incorrect information and make the necessary changes. If the entered information is correct, then enter Save. 8.4 On the Form, record the temperature of the simulator connected to the BT. 8.5 When prompted, set Test-Gas to Wet and Gas-Input to Breath Hose then Save. Effective Date: May 10, 2017 Page 18 of 23

19 8.6 Set the number of tests to ensure at minimum three tests are completed. 8.7 When prompted by the prompts on the display, connect and disconnect the breath hose to the output port on the simulator. 8.8 Review the Verification Report to determine if the results are within the following acceptable tolerances: The IR and EC results must fall between 0.074%-0.086% alcohol content The IR and EC results (n=6) must agree within +/ alcohol content of each other If the results are within the acceptable tolerances, then answer No to Perform another verification test and proceed to Set BT Certification Date (9) If the results are not within the acceptable tolerances, repeat quality control breath hose test If the results still do not meet acceptance tolerances, consult with a supervisor to determine if further testing is required or if the BT should be set aside until a later date and repeat the Certification. 9 SET BT CERTIFICATION DATE 9.1 Replace Cuvette Cap Replace black rubber cuvette cap on the Cuvette Inlet (Diagram 1). 9.2 Cylinder Check Connect the transducer cable on the gas enclosure to BT. Make sure the transducer plugs in smoothly and check BT display for connectivity. Connect the gas hoses to the proper inlets Attach a 0.080% dry gas cylinder to the BT via gas enclosure on each inlet. Make sure the BT screen displays the pressures of each inlet Remove cylinders. 9.3 Set BT Certification Date After Technical and Administrative review, the analyst that performed the certification or that supervised the analyst in training performing the certification will set the BT Certification Date. Effective Date: May 10, 2017 Page 19 of 23

20 9.3.2 Enter Menu Enter Maintenance Enter Set BT Certification Start Date The analyst (and the analyst in training, if applicable) must sign the printout The BT Certification Date must be done on the same day the BT was certified Place a label on the BT that includes: Police Department or Agency Model Number Serial Number Expiration Date 9.4 Department Network Enter Menu Enter Settings Enter Network Adjust the IP address, subnet mask, and gateway settings back to the Police Department or Agency s information After verification of the network settings by another analyst (who did not perform the calibration), press Save and restart BT The BT display should read PER TEST REQUIRED upon restart. 10 POST-CALIBRATION 10.1 Power down the BT Unplug all cords and cables Remove Draeger Alcotest 9510 USB Access Key Attach the gas enclosure to the BT using five screws. Effective Date: May 10, 2017 Page 20 of 23

21 10.5 Reassemble the BT and gas enclosure with the stand and reconnect all the cords and cables that were provided by the police department or agency Call Police Department or Agency to notify that the instrument is ready Document the use of the tandem simulators on the BT Solution Use Log. The solutions may be used up to 30 times. Effective Date: May 10, 2017 Page 21 of 23

22 11 REVISION HISTORY REVISION DATE VERSION REVISED BY TOTAL PAGES REVISION 9/15/ OAT 21 Original 5/18/ OAT 19 Updated acceptance criteria 5/10/ OAT 23 Minor updates throughout, see tracked changes in QA Effective Date: May 10, 2017 Page 22 of 23

23 12 MANUAL REVIEW ALL SIGNATURES ON FILE IN QUALITY ASSURANCE Melissa O Meara SECTION SUPERVISOR OAT DATE Nancy W. Brooks SECTION MANAGER OF FORENSIC CHEMISTRY DATE Albert A. Elian DEPUTY DIRECTOR OF FORENSIC CHEMISTRY DATE Kim Anderson QUALITY ASSURANCE MANAGER DATE 13 AUTHORIZATION Kristen L. Sullivan LABORATORY DIRECTOR DATE Effective Date: May 10, 2017 Page 23 of 23

STANDARD OPERATING PROCEDURES. Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2.

STANDARD OPERATING PROCEDURES. Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2. STANDARD OPERATING PROCEDURES Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2.5 ) Sampler (RAAS) AMBIENT AIR MONITORING PROGRAM for the 130 LIBERTY STREET DECONSTRUCTION

More information

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems Corporation Issued: 5/09 P/N: 600104 Rev. 2 Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems The Gas Addition Accessory permits the

More information

TEXAS DEPARTMENT OF PUBLIC SAFETY Breath Alcohol Laboratory

TEXAS DEPARTMENT OF PUBLIC SAFETY Breath Alcohol Laboratory TEXAS DEPARTMENT OF PUBLIC SAFETY Breath Alcohol Laboratory 5805 N LAMAR BLVD BOX 4087 AUSTIN, TEXAS 78773 512/424-5200 STEVEN C. McCRAW DIRECTOR DAVID G. BAKER CHERYL MacBRIDE DEPUTY DIRECTORS January

More information

Instruction Manual Dräger MSI P7 and MSI P7 plus

Instruction Manual Dräger MSI P7 and MSI P7 plus Dräger MSI GmbH Rohrstraße 32 58093 Hagen Tel.: +49-2331 / 9584-0 Fax: +49-2331 / 9584-29 e-mail: info@draeger-msi.de D 923; Edition 2011-01-01 Content 1. General Hints Page 4 2. The Instrument 2.1 Front

More information

Technical Procedure for General Laboratory Equipment

Technical Procedure for General Laboratory Equipment Technical Procedure for General Laboratory Equipment 1.0 Purpose - This procedure specifies the required elements for the use of general laboratory equipment. 2.0 Scope - This procedure applies to all

More information

Using the UltraRAE. Firmware 2.35

Using the UltraRAE. Firmware 2.35 Using the UltraRAE Firmware 2.35 Training Agenda UltraRAE features Setting up the UltraRAE Turning on the UltraRAE Idle Operation RAE-Sep Tubes Prepping for a measurement Taking a measurement Alarm modes

More information

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120 MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120 QUICK START GUIDE P/N 6006154, REVISION C MAY 2013 Model 8120 Mask Integrity Tester is patented under U.S. Patent No. 8,312,761. Additional patents are pending.

More information

Air Bubbler Depth Gauge DG2200 Installation and Reference Manual

Air Bubbler Depth Gauge DG2200 Installation and Reference Manual Air Bubbler Depth Gauge DG2200 Installation and Reference Manual Rev. 3.2.2 06/2011 Installation and Reference Manual Page 2 Contents Chapter 1: Installation and Overview... 3 1.1 DG2200 Location Diagram...

More information

HPICAL Operation & Data Logging Procedures. Click spacebar to advance through slides 1

HPICAL Operation & Data Logging Procedures. Click spacebar to advance through slides 1 HPICAL-15000 Operation & Data Logging Procedures Click spacebar to advance through slides 1 WARNING Always wear proper safety equipment when using high pressure equipment. Do not exceed 125 psi air pressure.

More information

BIOGAS COMBUSTION PLANT MAINTENANCE MANUAL

BIOGAS COMBUSTION PLANT MAINTENANCE MANUAL BIOGAS COMBUSTION PLANT MAINTENANCE MANUAL (PLANT MAINTENANCE MANUAL) Puente Gallego Landfill, Rosario (Argentina) ARIA.BIZ S.A. Montevideo 589, 7 floor 1019 Capital federal Argentina Tel: +54 11 5171

More information

Breath Alcohol Calibration Test No Summary Report

Breath Alcohol Calibration Test No Summary Report Colorative Testing Services, Inc FORENSIC TESTING PROGRAM Breath Alcohol Calibration Test No. 16-569 Summary Report This test was sent to 33 participants. Each sample pack consisted of four 34L certified

More information

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation Page 1 of 19 : SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor Description and Principle of Operation The photoacoustic multi-gas monitor (INNOVA 1412, Innova AirTech Instruments, Denmark) is a highly

More information

XC2 Client/Server Installation & Configuration

XC2 Client/Server Installation & Configuration XC2 Client/Server Installation & Configuration File downloads Server Installation Backup Configuration Services Client Installation Backup Recovery Troubleshooting Aug 12 2014 XC2 Software, LLC Page 1

More information

Met One E-BAM Particulate Monitor

Met One E-BAM Particulate Monitor STANDARD OPERATING PROCEDURES Met One E-BAM Particulate Monitor AMBIENT AIR MONITORING PROGRAM for the 130 LIBERTY STREET DECONSTRUCTION PROJECT LOWER MANHATTAN DEVELOPMENT CORPORATION 1 Liberty Plaza

More information

Read this first. Zetasizer nano series Self installation and Quick start guide MRK825-02

Read this first. Zetasizer nano series Self installation and Quick start guide MRK825-02 ! Read this first Zetasizer nano series Self installation and Quick start guide I N S T R U M E N T S MRK825-02 Zetasizer Nano series Self installation and Quick start guide MAN0383 Issue 1.1 July 2007

More information

GOLF LIFE MANAGEMENT SYSTEM Member Manual

GOLF LIFE MANAGEMENT SYSTEM Member Manual GOLF LIFE MANAGEMENT SYSTEM Member Manual CONTENTS: 1.Activate your account 2.Add Courses to Your List 3.Post Scores 4.Edit Scores 5.Frequently Asked Questions HANDICAP ACCOUNT ACTIVATION Before using

More information

Overview. Front Panel: Keypad and Display

Overview. Front Panel: Keypad and Display Overview The GA-200B is an analyzer that integrates a gas sampling system with sensors to measure and display the concentrations of oxygen and carbon dioxide in a sample as the percentage of a gas in the

More information

Setting Up the Ingenico isc250 Pinpad via USB

Setting Up the Ingenico isc250 Pinpad via USB One Blue Hill Plaza, 16 th Floor, PO Box 1546 Pearl River, NY 10965 1-800-PC-AMERICA, 1-800-722-6374 (Voice) 845-920-0800 (Fax) 845-920-0880 Setting Up the Ingenico isc250 Pinpad via USB The following

More information

Breath Alcohol Simulator Solution Analysis No Summary Report

Breath Alcohol Simulator Solution Analysis No Summary Report Collaborative Testing Services, Inc FORENSIC TESTING PROGRAM Breath Alcohol Simulator Solution Analysis No. 06-568 Summary Report This test was sent to 63 participants. Each sample pack consisted of two

More information

PGA3510 Portable 3-Gas IR Analyzer. Operations Manual

PGA3510 Portable 3-Gas IR Analyzer. Operations Manual PGA3510 Portable 3-Gas IR Analyzer Operations Manual Please read, understand, and follow these instructions before operating this equipment. Super Systems, Inc. is not responsible for damages incurred

More information

Technical Procedure for Instrument Calibration and Maintenance

Technical Procedure for Instrument Calibration and Maintenance Technical Procedure for Instrument Calibration and Maintenance 1.0 Purpose To outline the schedules and procedures involved in the maintenance and verification of critical instrument calibration in the

More information

20. Technical Equipment. Archived. This section applies to all equipment used in forensic examinations, such as

20. Technical Equipment. Archived. This section applies to all equipment used in forensic examinations, such as Instruments and Equipment Introduction 5.5. Authorization to use technical equipment 5.5.3 Use of instruments and equipment Instruction and maintenance manuals 5.4. 5.5.3 Use of nonlaboratory equipment

More information

Thermo Environmental Model 146C Calibrator Certification. Section Revision No September 17, 2014 Page 1 of 54

Thermo Environmental Model 146C Calibrator Certification. Section Revision No September 17, 2014 Page 1 of 54 Thermo Environmental Model 146C Calibrator Certification Section 2.3.4 Page 1 of 54 Page 3 of 54 Table of Contents 2.3.4 Thermo Environmental Model 146C Calibrator Certification QA Procedure... 4 2.3.4.1

More information

Setting up the Ingenico isc250 Pinpad via USB in Windows 8

Setting up the Ingenico isc250 Pinpad via USB in Windows 8 One Blue Hill Plaza, 16 th Floor, PO Box 1546 Pearl River, NY 10965 1-800-PC-AMERICA, 1-800-722-6374 (Voice) 845-920-0800 (Fax) 845-920-0880 Setting up the Ingenico isc250 Pinpad via USB in Windows 8 The

More information

5000TOC Sensor Standard Operating Procedure for TOC Calibration Conductivity Calibration Temperature Calibration Flow Rate Calibration

5000TOC Sensor Standard Operating Procedure for TOC Calibration Conductivity Calibration Temperature Calibration Flow Rate Calibration Part No. 84468 5000TOC Sensor Standard Operating Procedure for TOC Calibration Conductivity Calibration Temperature Calibration Flow Rate Calibration This document contains proprietary information, which

More information

GMA FUNCTIONAL TEST SETTING & TESTING OF SPIN UP PARAMETERS. P0586 REV -- 20, June 2000 GPB SCIENCE MISSION PROCEDURE

GMA FUNCTIONAL TEST SETTING & TESTING OF SPIN UP PARAMETERS. P0586 REV -- 20, June 2000 GPB SCIENCE MISSION PROCEDURE P0586 Rev. June 20, 2000 STANFORD UNIVERSITY W.W. HANSEN EXPERIMENTAL PHYSICS LABORATORY GRAVITY PROBE B, RELATIVITY GYROSCOPE EXPERIMENT STANFORD, CALIFORNIA 94305-4085 GMA FUNCTIONAL TEST SETTING & TESTING

More information

Running the LiCor 6400 Infra Red Gas Analyzer (IRGA) with 4 whole-plant cuvettes attached

Running the LiCor 6400 Infra Red Gas Analyzer (IRGA) with 4 whole-plant cuvettes attached Running the LiCor 6400 Infra Red Gas Analyzer (IRGA) with 4 whole-plant cuvettes attached Note: You must read the Primer manual Sections 1-6 before using the IRGA. The manual is located in the wall cabinet

More information

OMNISPORT 2000 SWIMMING QUICK REFERENCE

OMNISPORT 2000 SWIMMING QUICK REFERENCE Before During After OMNISPORT 2000 SWIMMING QUICK REFERENCE 1. Confirm that the timer is reset. 2. Set the timer to the correct event and heat. The right LCD, line 2 displays E:(event) H:(heat) R:(round)

More information

Operating Instructions Part No

Operating Instructions Part No DIGITAL AUTOMATIC TYRE INFLATOR Operating Instructions Part No. 11.0545 Thank you for selecting this Jamec Pem Automatic Tyre Inflator. Please read this manual before carrying out any installation or service

More information

Description. Measuring Device 3760XBI USP. Balancing

Description. Measuring Device 3760XBI USP. Balancing Measuring Device 3760XBI Description 3760XBI comprises from a differential pressure unit with an integrated true differential pressure sensor for the exact measuring of both differential and static pressures

More information

OPTIMA 7 - BIOGAS Calibrating the gas sensors SENSOR ADJUSTMENT. Diamond Scientific 625 Peachtree St., Cocoa, Florida /Tel:

OPTIMA 7 - BIOGAS Calibrating the gas sensors SENSOR ADJUSTMENT. Diamond Scientific 625 Peachtree St., Cocoa, Florida /Tel: SENSOR ADJUSTMENT The calibration of your OPTIMA 7 should be performed by experienced service personal only! MRU will not be responsible for any misuse or wrong interpretation of this calibration instruction.

More information

X-dock Software History

X-dock Software History X-dock Software History SW Version Bugfixes Features 01.01.01 01.02.2013 01.01.02 01.02.2013 01.02.05 04.03.2013 01.02.10 Start of history Pac 7000 support with FW 4.0 (older firmware locked) Improvements

More information

DIVE MEET SCORING INSTRUCTIONS. Day before meet

DIVE MEET SCORING INSTRUCTIONS. Day before meet DIVE MEET SCORING INSTRUCTIONS Day before meet Physical set up Set up registration/scoring table #1 on 1 meter side and scoring table #2 on 3 meter side of diving well, judges chairs, and award stand as

More information

Forensic Sciences Chapter ALABAMA DEPARTMENT OF FORENSIC SCIENCES ADMINISTRATIVE CODE CHAPTER CHEMICAL TEST FOR INTOXICATION

Forensic Sciences Chapter ALABAMA DEPARTMENT OF FORENSIC SCIENCES ADMINISTRATIVE CODE CHAPTER CHEMICAL TEST FOR INTOXICATION Forensic Sciences Chapter 370-1-1 ALABAMA DEPARTMENT OF FORENSIC SCIENCES ADMINISTRATIVE CODE CHAPTER 370-1-1 CHEMICAL TEST FOR INTOXICATION TABLE OF CONTENTS 370-1-1-.01 Admissibility Rule For The Evidential

More information

Armfield Distillation Column Operation Guidelines

Armfield Distillation Column Operation Guidelines Armfield Distillation Column Operation Guidelines 11-2016 R.Cox Safety SAFETY GLASSES ARE REQUIRED WHEN OPERATING THE DISTILLATION COLUMN Wear gloves when mixing alcohol feedstock The column will become

More information

D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA

D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA Software version 2.6 Table of contents 1 Introduction...3 1.1 Optical Sensor Technology...3 1.2 Installation...4 1.3 Operation...5

More information

FireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS

FireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS FireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS WARNING THE WARRANTIES MADE BY MSA WITH RESPECT TO THE PRODUCT ARE VOIDED IF THE PRODUCT IS NOT USED AND MAINTAINED IN ACCORDANCE

More information

ACV-10 Automatic Control Valve

ACV-10 Automatic Control Valve ACV-10 Automatic Control Valve Installation, Operation & Maintenance General: The Archer Instruments ACV-10 is a precision automatic feed rate control valve for use in vacuum systems feeding Chlorine,

More information

SDM-2012 Docking Station Standalone Configuration Operator s Manual

SDM-2012 Docking Station Standalone Configuration Operator s Manual SDM-2012 Docking Station Standalone Configuration Operator s Manual Part Number: 71-0254RK Revision: P5 Released: 10/5/12 www.rkiinstruments.com Warranty RKI Instruments, Inc. warrants gas alarm equipment

More information

PERFORM Operating Document

PERFORM Operating Document PERFORM Operating Document Use and Maintenance of CO 2 Incubator PC-POD-CA-007-v03 Revision History Version Reason for Revision Date 01 New POD 30-Sep-13 02 Minor revisions for section 2.3, 3.1, 4.3. 14-April-16

More information

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER)

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER) Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER) Background There are two main sources of energy available for animal metabolism: carbohydrates (CHO) and fats. These molecules are broken

More information

RAM 4021-DPX Operation Manual

RAM 4021-DPX Operation Manual RAM 4021-DPX Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS ABL 4021-DPX / RAM 4021-DPX For Your Safety... 3 Description... 3 Setup Mode... 4 Lights/Alarms... 4 Operation...

More information

Cover Page for Lab Report Group Portion. Head Losses in Pipes

Cover Page for Lab Report Group Portion. Head Losses in Pipes Cover Page for Lab Report Group Portion Head Losses in Pipes Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 February 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section

More information

STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA SOP#: 1157.1 Date Issued: 05/14 Date Revised: 5/15 Page 1 of 6 TITLE: SCOPE: RESPONSIBILITY: PURPOSE: SurgiVet

More information

WELCOME TO THE FUTURE OF TELEVISION. User Manual

WELCOME TO THE FUTURE OF TELEVISION. User Manual WELCOME TO THE FUTURE OF TELEVISION User Manual Welcome to BTV Each living room and family sharing our technology resonate our commitment to excellence. In our responsibility to offer the best products,

More information

Dräger X-dock Frequently Asked Questions

Dräger X-dock Frequently Asked Questions COSTS Why do I save costs when using Dräger X-dock? SOFTWARE What are the benefits of using a central database? What is the difference between the X-dock Manager Standard and the Professional Version?

More information

Model of a Standard Operating Procedure (SOP) for SO 2 measurements with a UV fluorescence analyzer

Model of a Standard Operating Procedure (SOP) for SO 2 measurements with a UV fluorescence analyzer Model of a Standard Operating Procedure (SOP) for SO 2 measurements with a UV fluorescence analyzer How to use this specimen SOP This model SOP aims at giving support to network operators in setting-up

More information

AX5000 Operational Manual

AX5000 Operational Manual MIYACHI AMERICA CORPORATION The World Leader in Hermetic Sealing Systems AX5000 Operational Manual 0 Document #107-00092-001 Dec, 2013 AX5000 Operational Manual Miyachi America Corporation 1820 S. Myrtle

More information

Mr Roger Wood Measurement & Process Group (M&P) Network Strategy National Grid Gas plc Brick Kiln Street Hinckley Leicestershire LE10 0NA

Mr Roger Wood Measurement & Process Group (M&P) Network Strategy National Grid Gas plc Brick Kiln Street Hinckley Leicestershire LE10 0NA Mr Roger Wood Measurement & Process Group (M&P) Network Strategy National Grid Gas plc Brick Kiln Street Hinckley Leicestershire LE10 0NA Dear Roger, Your Ref: Our Ref: Direct Dial: 020 7901 7468 Email:

More information

Operating the DataMaster DMT. The DataMaster DMT

Operating the DataMaster DMT. The DataMaster DMT Version 1.0 Division of Criminal Investigation Alcohol Section July 2009 The DataMaster DMT 1 The DataMaster DMT is a scientific instrument designed to analyze a sample of a person s s breath and determine

More information

Maintenance of the Electronic ( EPC ) Pressure Controllers on GCs

Maintenance of the Electronic ( EPC ) Pressure Controllers on GCs Gas flows in all GCs are controlled by Electronic Pressure Controllers ( EPC ). There may be up to 7 EPCs mounted in one GC depending on which combination of injectors and detectors are installed. Each

More information

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4 Calibration....

More information

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012 JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012 UNIVERSITY OF TEXAS AT ARLINGTON Nanofabrication Research and Teaching Facility TABLE OF CONTENTS 1. Introduction....2 1.1 Scope of Work.....2

More information

IEP SENSOR BRACKET CHECKOUT AND VERIFICATION. This is a non-hazardous test.

IEP SENSOR BRACKET CHECKOUT AND VERIFICATION. This is a non-hazardous test. Page 1 of 20 NEXT GENERATION GAS SYSTEM (NGGS ) IEP SENSOR BRACKET CHECKOUT AND VERIFICATION PROCEDURE 870-PROC-623 Revision A Goddard Space Flight Center Greenbelt, Maryland 20771 FEBRUARY 13, 2002 This

More information

Boyle s Law: Pressure-Volume Relationship in Gases

Boyle s Law: Pressure-Volume Relationship in Gases Boyle s Law: Pressure-Volume Relationship in Gases The primary objective of this experiment is to determine the relationship between the pressure and volume of a confined gas. The gas we will use is air,

More information

CircuFlow Quick Setup Guide

CircuFlow Quick Setup Guide CircuFlow 5200 Quick Setup Guide LB04.0001 Rev B 20140314 TABLE OF CONTENTS Setup Therapy Setup Gradient Mode Pressure Mode Lock Device Unlock Device Treatment 03 04 05 08 11 12 13 02 SETUP 01 01 Open

More information

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021-PR Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021-PR For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4

More information

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments Pegas 4000 MF Gas Mixer InstructionManual Contents I Table of Contents Foreword Part I Introduction 1 2 1 System overview... 2 2 Specifications... 3 Part II Installation 4 1 Rear panel connections...

More information

APA software instruction manual

APA software instruction manual 1. Starting the program In order to start the control software for the APA device press APAxx shortcut located on the desktop of the supplied computer. XX corresponds to the current software version. When

More information

D-Opto. Dissolved Oxygen Sensor. Operation Manual (Software Version 1.0)

D-Opto. Dissolved Oxygen Sensor. Operation Manual (Software Version 1.0) D-Opto Dissolved Oxygen Sensor Operation Manual (Software Version 1.0) Table of contents 1 Introduction...3 1.1 Optical Sensor Technology...3 2 Software...4 2.1 Installation...4 2.2 Operation...4 3 Operation...9

More information

WSN: XXXXXX 101 ROME COURT FORT COLLINS, CO, PHONE #: (970) FAX #: (970) TO-Clean Manual

WSN: XXXXXX 101 ROME COURT FORT COLLINS, CO, PHONE #: (970) FAX #: (970) TO-Clean Manual WSN: XXXXXX 101 ROME COURT FORT COLLINS, CO, 80524 PHONE #: (970) 221-9179 FAX #: (970) 221-9364 TO-Clean Manual Copyright 2008 by Wasson ECE Instrumentation, Inc. (SJC 02/05/2008) All rights reserved

More information

Document Number: 302/34000/76 SV/sv Revision: (B) 2009/02/05

Document Number: 302/34000/76 SV/sv Revision: (B) 2009/02/05 Product Calibration Procedure For TriStar II 3020 APPROVALS: Director, Engineering Stefan Koch Manager, Product Integrity and Performance Tony Thornton Senior Project Manager Patrick Wommack Manager, International

More information

Liquid Handling Systems

Liquid Handling Systems Liquid Handling Systems 1.0 Purpose - This procedure specifies the required elements for the maintenance and use of the Hamilton- Microlab 1000 Plus and the Microlab 500 Liquid Handling Systems. 2.0 Scope

More information

Diver-Office. Getting Started Guide. 2007, Schlumberger Water Services

Diver-Office. Getting Started Guide. 2007, Schlumberger Water Services Diver-Office Getting Started Guide 2007, Schlumberger Water Services Copyright Information 2007 Schlumberger Water Services. All rights reserved. No portion of the contents of this publication may be reproduced

More information

Section II: Operator Responsibilities Sulfur Dioxide Standard Operating Procedure. Revision 12. Effective Date: November 1, 2016

Section II: Operator Responsibilities Sulfur Dioxide Standard Operating Procedure. Revision 12. Effective Date: November 1, 2016 Page 1 of 34 Section II: Operator Responsibilities Sulfur Dioxide Standard Operating Procedure Revision 12 Effective Date: Section No. 2.8.2 Page 2 of 34 Approval Sign Off-Sheet I certify that I have read

More information

HiVolCal User Manual High Volume Field Calibrator Model HVC1 (7 to 54 CFM) NIST Traceable ISO 9001:2008

HiVolCal User Manual High Volume Field Calibrator Model HVC1 (7 to 54 CFM) NIST Traceable ISO 9001:2008 HiVolCal User Manual High Volume Field Calibrator Model HVC1 (7 to 54 CFM) NIST Traceable ISO 9001:2008 10 Park Pl. Butler, NJ 07405 USA www.mesalabs.com MK101-20 Rev. A Table of Contents. Section Topic

More information

Basketball Instructions

Basketball Instructions Quality Family Games Basketball Instructions Model 752.69 Product Registration 752.69 08/14/14 Visit www.carrom.com to register your product to receive the latest product updates, special offers, and other

More information

Fencing Time Version 4.3

Fencing Time Version 4.3 Fencing Time Version 4.3 Upgrading your Fencing Time Server October 2017 Copyright 2017 by Fencing Time, LLC. All rights reserved. Overview Periodically, a new version of Fencing Time is released. In most

More information

OPERATION MANUAL DIVE CALIPER NOT SCALE / IF IN DOUBT ASK* *ALL UNITS IN MM/DEGREES UNLESS SPECIF. Dive Caliper General DC-051

OPERATION MANUAL DIVE CALIPER NOT SCALE / IF IN DOUBT ASK* *ALL UNITS IN MM/DEGREES UNLESS SPECIF. Dive Caliper General DC-051 NOT SCALE / IF IN DOUBT ASK* DIVE CALIPER OPERATION MANUAL *THIS INFORMATION IS THE PROPERTY OF ZEBRA-TECH LTD. ALL RIGHTS RESERVED.* *ALL UNITS IN MM/DEGREES UNLESS SPECIF SURFACE F 3.2 Ra UNLESS S Part

More information

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300 Technical Note GA-300 Overview The GA-300 CO2 and O2 Gas Analyzer is easy to use, robust, and adaptable to human, animal, and plant applications. The GA-300 has two analog outputs to allow recording and

More information

ANALOG PRESSURE GAUGE

ANALOG PRESSURE GAUGE 304-00103 ANALOG PRESSURE GAUGE CALIBRATION PROCEDURE BARFIELD P/N 304-00103 20-304-00103 August 6, 2013 BARFIELD, INC. Corporate Headquarters 4101 Northwest 29th Street Miami, Florida 33142 www.barfieldinc.com

More information

For running only the scoresheet application without any video features only some very basic hardware / software requirements have to be fulfilled:

For running only the scoresheet application without any video features only some very basic hardware / software requirements have to be fulfilled: Digital Scoresheet user manual Requirements For running only the scoresheet application without any video features only some very basic hardware / software requirements have to be fulfilled: Laptop, preferably

More information

D-Opto Dissolved Oxygen Sensor Operation Manual

D-Opto Dissolved Oxygen Sensor Operation Manual D-Opto Dissolved Oxygen Sensor Operation Manual Software version 2.6 Contents 1. Introduction..... 1 Optical Sensor Technology.... 1 Features...... 1 2. D-Opto Specifications. 2 Specifications... 2 3.

More information

BGA244 Binary Gas Analyzer

BGA244 Binary Gas Analyzer Quick Start Guide Revision 1.0 Certification Warranty Service certifies that this product met its published specification at the time of shipment. This product is warranted against defects in materials

More information

Electronic Commissioning Unit. Instructions

Electronic Commissioning Unit. Instructions Electronic Commissioning Unit Instructions 1. The Comdronic AC6 Electronic Commissioning Unit 2. Health and Safety 3. Description of Unit 4. Connection and Set-up Procedure 5. Menu System a. Main Menu

More information

Technical Bulletin. Seametrics Smart Sensors: Barometric Compensation (with optional DTW setting) Introduction. How Pressure is Measured

Technical Bulletin. Seametrics Smart Sensors: Barometric Compensation (with optional DTW setting) Introduction. How Pressure is Measured Seametrics Smart Sensors: Precision Environmental Sensors An ONICON Brand The Barometric Compensation Utility is specifically for the PT2X and LevelSCOUT sensors. It uses barometric data to compensate

More information

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008 UNITY 2 TM Air Server Series 2 Operators Manual Version 1.0 February 2008 1. Introduction to the Air Server Accessory for UNITY 2...2 1.1. Summary of Operation...2 2. Developing a UNITY 2-Air Server method

More information

Multiplate Platelet Function Analyzer: An Overview

Multiplate Platelet Function Analyzer: An Overview Multiplate Platelet Function Analyzer: An Overview The Multiplate is for research use only in the US and Canada Scope: This is a summary of most aspects of platelet function testing using the Multiplate

More information

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument OxyScan Graphic Operating Instructions UMS Micro-oxygen sensor 501 Microprocessor instrument Introduction Thank you for choosing the UMS Micro Oxygen Sensor 501 - a highly advanced product! Please read

More information

Touch Screen Guide. OG-1500 and OG Part # T011

Touch Screen Guide. OG-1500 and OG Part # T011 Touch Screen Guide OG-1500 and OG-2000 Part # 9000000.T011 Effective 11/2010 External View Internal View 1. Transducer Banks 2. Oxygen Sensor 3. PLC These are the two manifolds with three (3) transducers

More information

Scoreboard Operator s Instructions MPC Control

Scoreboard Operator s Instructions MPC Control Scoreboard Operator s Instructions MPC Control Some features on the keyboard overlay may not be included on the particular model being operated. Since 1934 Retain this manual in your permanent files 1/21/2011

More information

The ICC Duckworth-Lewis-Stern calculator. DLS Edition 2016

The ICC Duckworth-Lewis-Stern calculator. DLS Edition 2016 The ICC Duckworth-Lewis-Stern calculator DLS Edition 2016 (DLS2-2016) Installation and operating instructions Queries about program operation should be sent to: Steven.Stern@qut.edu.au 2016 International

More information

RAM Operation Manual

RAM Operation Manual RAM 4021-1 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021-1 For Your Safety... 2 Description... 2 Setup Mode... 3 Lights/Alarms... 3 Operation... 4 Calibration...

More information

The following documentation is an electronicallysubmitted vendor response to an advertised solicitation from the West Virginia Purchasing Bulletin

The following documentation is an electronicallysubmitted vendor response to an advertised solicitation from the West Virginia Purchasing Bulletin The following documentation is an electronicallysubmitted vendor response to an advertised solicitation from the West Virginia Purchasing Bulletin within the Vendor Self Service portal at wvoasis.gov.

More information

McKnight Hockey Association

McKnight Hockey Association McKnight Hockey Association Electronic Evaluation Tool Manual 2013-2014 Table of Contents Introduction...3 Evaluation Tool...3 Login to OneClickIce...3 Evaluations...4 PROCESS...4 Evaluation Procedure...5

More information

!!!DANGER!!! DO NOT USE THIS EQUIPMENT TO PURGE TOXIC OR FLAMMABLE GAS AND

!!!DANGER!!! DO NOT USE THIS EQUIPMENT TO PURGE TOXIC OR FLAMMABLE GAS AND !!!DANGER!!! DO NOT USE THIS EQUIPMENT TO PURGE TOXIC OR FLAMMABLE GAS AND DO NOT USE THIS EQUIPMENT UNDER FLAMMABLE, VOLATILE OR TOXIC ENVIRONMENTAL CONDITIONS 1 2 Recortest III to Rec4 Upgrade to Windows

More information

Software for electronic scorekeeping of volleyball matches, developed and distributed by:

Software for electronic scorekeeping of volleyball matches, developed and distributed by: Software for electronic scorekeeping of volleyball matches, developed and distributed by: Developed for the rules of USports 2017-18 As adopted by Ontario University Athletics for Men s & Women s Volleyball

More information

4.0 Audit Procedure for the MetOne SASS Speciation Sampler

4.0 Audit Procedure for the MetOne SASS Speciation Sampler 4.0 Audit Procedure for the MetOne SASS Speciation Sampler Table of Contents 4.0 Audit Procedure for the MetOne SASS Speciation Sampler... 1 4.1 Background...4 4.2 Calibration Equipment...4 4.3 Audit Procedure

More information

Lifeloc FC10 Plus. Operations Manual. Unlock the Power of Alcohol Testing

Lifeloc FC10 Plus. Operations Manual. Unlock the Power of Alcohol Testing Operations Manual Unlock the Power of Alcohol Testing Table of Contents 2 Introduction Congratulations...4 Front View...5 Features...6 Preparation Installing Batteries... 7 Turning the FC On and Off...

More information

PERFORM Operating Document

PERFORM Operating Document PERFORM Operating Document Use and Maintenance of CO 2 Incubator PC-POD-CA-007-v02 Revision History Version Reason for Revision Date 01 New POD 30-Sep-13 02 Minor revisions for section 2.3, 3.1, 4.3. 14-April-16

More information

Manual Weighingblock VB2 series and Uniscale

Manual Weighingblock VB2 series and Uniscale Manual Weighingblock VB2 series and Uniscale Note: At page 8 in this manual you will find a short form instruction. Normally the only instruction shipped together with the Scale. Overview different ranges.

More information

DeWiggler. Get your calibrations under control

DeWiggler. Get your calibrations under control DeWiggler Get your calibrations under control DeWiggler Pro DeWiggler Pro comes aboard with a calibration professional. Calibrations are determined and set in real-time. When the calibrator leaves, Pro

More information

STATE OF CALIFORNIA AIR RESOURCES BOARD AIR MONITORING QUALITY ASSURANCE VOLUME V AUDIT PROCEDURES MANUAL FOR AIR QUALITY MONITORING APPENDIX E

STATE OF CALIFORNIA AIR RESOURCES BOARD AIR MONITORING QUALITY ASSURANCE VOLUME V AUDIT PROCEDURES MANUAL FOR AIR QUALITY MONITORING APPENDIX E STATE OF CALIFORNIA AIR RESOURCES BOARD AIR MONITORING QUALITY ASSURANCE VOLUME V AUDIT PROCEDURES MANUAL FOR AIR QUALITY MONITORING APPENDIX E SYSTEM AUDIT PROCEDURES FOR THROUGH-THE-PROBE CRITERIA AUDITS

More information

Operating Instructions Part No

Operating Instructions Part No DIGITAL AUTOMATIC TYRE INFLATOR Operating Instructions Part No. 11.0578 Thank you for selecting this Jamec Pem Automatic Tyre Inflator. Please read this manual before carrying out any installation or service

More information

REMOTE CLIENT MANAGER HELP VERSION 1.0.2

REMOTE CLIENT MANAGER HELP VERSION 1.0.2 VERSION 1.0.2 MERCHANT SALES: 800-637-8268 New Merchant Accounts PARTNER PROGRAMS: 800-637-8268 New and existing partnerships CUSTOMER CARE: 800-338-6614 Existing merchant account support Statements and

More information

Operating Manual. SUPREMA Calibration. Software for Fire and Gas Warning Units. Order No.: /01. MSAsafety.com

Operating Manual. SUPREMA Calibration. Software for Fire and Gas Warning Units. Order No.: /01. MSAsafety.com Operating Manual Software for Fire and Gas Warning Units Order No.: 10154656/01 MSAsafety.com MSA Europe GmbH Schlüsselstrasse 12 8645 Rapperswil-Jona Switzerland info.ch@msasafety.com www.msasafety.com

More information

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure PPMS Service Note 1099-412 In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure Introduction: Loss of electricity to the PPMS EverCool II should not cause damage to

More information

ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS Dionex Corporation

ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS Dionex Corporation ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS 2000 Dionex Corporation Document No. 031277 Revision 03 April 2000 2000 Dionex Corporation All rights reserved worldwide. Printed in the United States of

More information

Environmental Protection Division

Environmental Protection Division Operational Policy Manual Section Subsection Environmental Protection Division 6.0 6.01.02 Name of Policy: Replaces: Standard Auditing Procedure for Continuous Emission Monitors and Ambient Air Monitoring

More information

How to calibrate the formaldehyde sensor within a PPMonitor Wireless Unit:

How to calibrate the formaldehyde sensor within a PPMonitor Wireless Unit: How to calibrate the formaldehyde sensor within a PPMonitor Wireless Unit: Please read this document thoroughly before attempting to check or adjust the calibration of the formaldehyde sensor within the

More information