5000TOC Sensor Standard Operating Procedure for TOC Calibration Conductivity Calibration Temperature Calibration Flow Rate Calibration

Size: px
Start display at page:

Download "5000TOC Sensor Standard Operating Procedure for TOC Calibration Conductivity Calibration Temperature Calibration Flow Rate Calibration"

Transcription

1 Part No TOC Sensor Standard Operating Procedure for TOC Calibration Conductivity Calibration Temperature Calibration Flow Rate Calibration This document contains proprietary information, which is protected by copyright. All rights are reserved. No part of this document may be photocopied (other than where specifically noted), reproduced or translated, into another language without the prior written consent of Mettler-Toledo Thornton, Inc.

2 Mettler-Toledo Thornton, Inc No part of this manual may be reproduced or transmitted in any form or by any means, electronic or mechanical, including photocopying (other than where specifically noted), for any purpose without the express written permission of Mettler-Toledo, Inc. U.S. Government Restricted Rights: This documentation is furnished with Restricted Rights. METTLER TOLEDO THORNTON RESERVES THE RIGHT TO MAKE REFINEMENTS OR CHANGES WITHOUT NOTICE.

3 IMPORTANT SAFETY INFORMATION Please read thoroughly before operating the 5000TOC System Suitability Test Kit - Follow all warnings, cautions, and instructions indicated on and supplied with this product. Install equipment as specified in this instruction manual. Follow appropriate local and national codes. Use only factory documented components for repair. Tampering or unauthorized substitution of parts and procedures can affect the performance and cause unsafe operation of your process as well as void factory warranties. Protective covers must be in place unless qualified personnel are performing maintenance. If this equipment is used in a manner not specified by the manufacturer, the protection provided by it against hazards may be impaired. Prior to shipping the sensor back to the factory for repair or re-calibration, water MUST be drained from sensor to avoid damage due to freezing. WARNINGS: Installation of cable connections and servicing of this product require access to shock hazard voltage levels. Main power must employ a switch or circuit breaker as the disconnecting device for the equipment. Electrical installation must be in accordance with the National Electrical Code and/or any other applicable national or local codes. Safety and performance require that this instrument be connected and properly grounded through a three-wire power source. This manual includes safety information with the following designations and formats: WARNING: POTENTIAL FOR PERSONAL INJURY. CAUTION: possible instrument damage or malfunction. NOTE: important operating information. Definition of Equipment Symbols On the instrument indicates: Warning Caution, risk of electric shock. ~ On On the instrument indicates: Caution (refer to accompanying documents). the instrument indicates: There is alternating current present.

4

5 TABLE OF CONTENTS 1. INTRODUCTION SCOPE CALIBRATION SELECTION AND SEQUENCE EQUIPMENT REQUIRED EQUIPMENT DATA FLOW RATE CALIBRATION Flow Rate Verification Before Calibration Flow Rate Calibration Flow Rate Verification After Calibration EQUIPMENT PREPARATION Installing the Thornton 5000TOC Calibration Test Kit Brackets Installing the Thornton 5000TOC Calibration Kit Tubing Final Installation of the Thornton 5000TOC Calibration Kit MAX CONFIGURATION TEMPERATURE CALIBRATION T1 Temperature Verification Before Calibration T1 Temperature Calibration T2 Temperature Verification Before Calibration T2 Temperature Calibration CONDUCTIVITY CALIBRATION C1 Conductivity Verification Before Calibration C1 Conductivity Calibration C2 Conductivity Verification Before Calibration C2 Conductivity Calibration TOC CALIBRATION TOC As Found TOC Calibration Procedure TOC Calibration Verification Procedure RESTORE TO OPERATION SETTING SAMPLE FLOW RATE Flow Rate Adjustment Procedure Flow Rate Indication WORKSHEET 1: PART NUMBER AND SERIAL NUMBER DATA WORKSHEETS FOR CALIBRATION AND VERIFICATION... 17

6

7 1. Introduction The Thornton 5000TOC Total Organic Carbon Sensor and 770MAX meter measures the amount of organic carbon in high purity waters by oxidizing organic carbon to CO 2 with appropriate UV radiation. The resulting increase between two temperaturecompensated conductivity measurements of the sample flow stream at points before and after oxidation is used to calculate the amount of organic carbon present. The complete calibration of the 5000TOC Sensor consists of calibration of the following. flow rate sensor measurement two temperature sensor measurements two conductivity sensor measurements TOC measurement Throughout the test procedure, the units ppb or ppb Carbon and µg Carbon/L will appear. They are identical for this procedure. ppb (or ppm) is the common terminology used throughout the industry, though it is not a recognized SI unit. 2. Scope This document provides procedures to calibrate the Thornton 5000TOC Total Organic Carbon Sensor (PN or PN and related part numbers). The 5000TOC Sensor is used with the Thornton 770MAX meter. Calibration of the 770MAX meter is covered in separate documentation. 3. Calibration Selection and Sequence Calibration of the 5000TOC Sensor consists of four sequential calibration procedures; flow rate, temperature, conductivity, and TOC. To perform a complete calibration of every component of the 5000TOC Sensor, these procedures must be performed in this order. If you choose to calibrate the TOC measurement more frequently than the other measurements, please refer to Section 11 of this manual. 4. Equipment Required Thornton 5000TOC Calibration Test Kit (Thornton PN ). This consists of a calibrated conductivity sensor ( ), 5 foot patch cable (Thornton PN ), 316L SS flow housing (Thornton PN ), holding brackets, SOP for Calibration, pump, power supply, and miscellaneous connectors and tubing. Thornton 5000TOC Calibration Solution Set (Thornton PN ). Stop watch or other timing device with 1 second resolution. 100 ml or larger volumetric container such as graduated cylinder with <3% tolerance. Note: Alternatively, item 1 can be replaced with a Thornton System Suitability Test Kit (Thornton PN ) in conjunction with a Thornton Calibration Test Accessory Kit (Thornton PN ) to execute this procedure. Note: Use a hold time to freeze the relay and analog outputs so that alarms are not triggered during any service procedures. Normal relay and analog output operation resumes when the set period has elapsed or when the hold time is reset to zero. Set hold time by pressing Menu from the 770MAX main display. Scroll to Set Hold time by using the up/down arrows. Press Enter, then enter a time between 1 to 99 minutes, and press Menu. The hold time is now set. Caution: Be sure to adjust needle valve to less than 3 turns from closed position before re-installing into process line! 5. Equipment Data Record all relevant data identified in Worksheet 1. Page 1

8 6. Flow Rate Calibration This procedure uses a volume and time method for calibrating the flow sensor. Process water is sufficient to use for this calibration. The end of the drain tube at the sample outlet is the measurement point. NOTE: Since it is possible for the person performing this calibration to come into contact with the process water, use of protective gloves, or other suitable safety equipment is recommended. Caution: If the process water is hot, protection from direct contact should be used, or an alternate source of water must be provided Flow Rate Verification Before Calibration 1. Adjust the 5000TOC flow rate to 20 ± 1 ml/min as described in Section 13. Setting Sample Flow Rate. 2. Arrange the drain tube at the atmospheric drain such that the water dripping from it may be collected into the volumetric container. 3. On the 770MAX, press Menu (exit). 4. Press or to select Calibrate. Press Enter (or Page Down). 5. Press or to select Sensor. Press Enter (or Page Down). 6. Enter Output Hold Time. Press Enter (or Page Down). Output Hold Time is the amount of time that you want the present relay status and analog output value to be held (unchanged) while calibration is performed. 7. Press or to select the letter for the measurement of the 5000TOC Sensor being calibrated. Any of the various sub-sensors within the 5000TOC Sensor may be calibrated from any other measurement letter. For example, if Measurement C is the TOC from the unit being calibrated, and Measurement D is its conductivity, then any of the subsensors within the unit may be calibrated from either Measurement C or D. Press Enter. 8. Press or to select 2-point. Press Enter or. 9. Press or to select the Flow. Press Page Down. The 770MAX should now be displaying the 5000TOC Sensor flow rate. 10. Place a dry volumetric container under the drain line to capture the water, and begin the time measurement using the stop watch. It is important that the water collection and time start be closely matched. 11. Collect a known volume of water with the volumetric container for a specific time, at least 3 minutes. 12. Record in Table 1 Section A the volume of water in ml and the elapsed time in seconds. These two values determine the reference flow rate. 13. Calculate the reference flow rate and record as F ref in Table 1 Section A according to the equation below. The volume may also be determined by weight. Measure the mass increase of the container in grams, and divide the mass by the density of the water (typically, 1.00 g/ml at room temperature). Flow Rate (ml/min) = F ref 60 sec/min volume collected (ml) = Elapsed Time (sec) eq Record the flow rate as shown on the display of the 770MAX as the F before value in Table 1 Section A. 15. Subtract the measured flow rate from the reference flow rate (F ref F before ). 16. If the absolute value of the difference is less than the limit, then flow rate adjustment is not required. Press Menu to exit and proceed to Section 7. Equipment Preparation. Otherwise, leave the 770MAX in its current state and proceed to the next section Flow Rate Calibration NOTE: All flow adjustments are to be made using the needle valve on the inlet of the 5000TOC Sensor.. 1. Using the display on the 770MAX (in the calibration mode as described in Section 6.1), adjust the flow rate through the 5000TOC Sensor to about 24 ± 1 ml/min as indicated on the 770MAX display. 2. Place a dry volumetric container under the drain line to capture the water, and begin the time measurement using the stop watch. It is important that the water collection and time start be closely matched. Page 2

9 3. Collect a known volume of water with the volumetric container for a specific time, at least 3 minutes. 4. Record in Table 1 Section B the volume of water in ml and the elapsed time in seconds. 5. Using eq 1, calculate the water flow rate and record as F ref1 in Table 1 Section B. 6. Enter the reference flow rate into the 770MAX and press Page-Down. 7. Wait while the calibration is being performed. 8. Adjust the flow rate to about 16 ± 1 ml/min. 9. Empty the volumetric container. Repeat steps Record data as F ref2 in Table 1 Section B. 11. After the flow sensor calibration is completed, enter the correct date. Press Page-Down. 12. Select YES to save the new calibration data within the 5000TOC Sensor. Note: if NO is selected, the data will only be temporarily saved, and will revert back to the previous value should power be lost or the 5000TOC Sensor patch cable is unplugged Flow Rate Verification After Calibration 1. Repeat steps 1-11 in Section 6.1. to verify the flow sensor is in calibration. 2. Record the volume of water in ml and the elapsed time in seconds in Table 1 Section C. These two values determine the reference flow rate. 3. Using eq1, calculate the reference flow rate and record in Table 1 Section C as F ref. 4. Record the flow rate as shown on the display of the 770MAX as F after in Table 1 Section C. 5. Subtract reference flow rate from the after flow rate (F ref F after ). 6. If the absolute value of the difference is less than the limit, then flow rate calibration is complete. Otherwise, return to Section Equipment Preparation Flow Calibration is complete Installing the Thornton 5000TOC Calibration Test Kit Brackets The Thornton 5000TOC Calibration Test Kit provides an easy means to introduce calibration solutions into the 5000TOC Sensor. This kit is also used for Temperature, Conductivity and TOC calibrations. Components in this kit include a bracket assembly, transfer cap assembly, assorted fittings, 30 cc plastic syringe, positive displacement pump and universal AC power supply. 1. Remove the Main Bracket A from the Calibration Kit and install on top of the 5000TOC Sensor as illustrated in Figure Remove the Ref. Sensor Holder from the Calibration Kit and install it on the above assembly as illustrated in Figure Remove the Main Bracket B from the Calibration Kit and install as illustrated in Figure 1. Page 3

10 REF SENSOR HOLDER MAIN BRACKET MAIN BRACKET "B" THUMB NUTS Figure 1 4. Remove the Pump from the Calibration Kit and install onto the above assembly as illustrated in Figure 2. Figure 2 5. Remove the Bottle Holder Bracket from the Calibration Kit and install onto the above assembly as illustrated in Figure Remove the Ref. Sensor Assembly from the Calibration Kit and install onto the Ref. Sensor Holder as illustrated in Figure Secure the entire assembly by tightening the Thumb Screws. The entire bracket assembly is now complete and should look like Figure 3. Page 4

11 Figure Installing the Thornton 5000TOC Calibration Kit Tubing 1. Remove the Transfer Cap assembly from the 5000TOC Calibration Kit and clean it using de-ionized or process water (water to be tested). 2. Take one of the Blank Water bottles and shake it to homogenize the contents. 3. Remove the Blank Water bottle cap and replace it with the cleaned Transfer Cap. 4. Place the Blank Water bottle onto the assembled Bottle Holder Bracket as illustrated in Figure 4. Figure 4 5. Remove Tube A from the 5000TOC Calibration Kit and clean it using de-ionized or process water (water to be tested). Page 5

12 6. Insert one end of Tube A into the Transfer Cap until it comes to within 1 inch (3 cm) of the bottom of the Blank Water bottle. 7. Place the other end of Tube A into the IN port of the Pump, see Figure 5. B A C D NEEDLE VALVE Figure 5 8. Connect Tube B from the OUT of the Pump to the input of the Ref. Sensor Assembly according to Figure Remove Tube C from the Calibration Kit and connect one end to the output of the Ref. Sensor Assembly according to Figure Connect the other end of Tube C to the inlet filter on the Sample Inlet of the 5000TOC Sensor according to Figure Remove Tube D from the 5000TOC Calibration Kit and clean it using de-ionized or process water (water to be tested). 12. Connect one end of Tube D to the Sample Outlet of the 5000TOC Sensor (do not connect the other end at this time). 13. Place the other end of Tube D over a container that can hold at least 100 ml of liquid. Do not allow the lower end of the tube to become contaminated. 14. Verify the pump switch is in the OFF position. 15. Connect the Universal Power Supply to the pump and to VAC Hz power. The kit includes a North American AC power cord. Replace or adapt the power cord as necessary and in accordance with local regulations. 16. Turn the pump power switch to ON and operate pump for approximately 2 minutes to remove any residual water from the pump and the 5000TOC Sensor. 17. Turn the pump OFF. 18. Once the residual water has been successfully purged, insert about 4 inches (10 cm) of the exposed end of Tube D into the Transfer Cap. The assembly is now complete and should now appear as per Figure 6. IMPORTANT NOTE: When the pump is first put into service, or after extended storage, it may need to be primed by injecting reagent water into the inlet fitting with the provided syringe. Place the pump in its normal vertical position with the inlet tube from the bottle removed from the inlet fitting and the outlet tube placed to discharge to a nearby drain or into a container. Draw reagent or clean process water into the syringe, place the tip of the 30cc syringe into the pump inlet fitting, turn on the pump and push about 10 ml of water into the fitting or until water appears at the outlet fitting. Excess water will run from the inlet fitting. Turn off the pump. Reattach the tube from the bottle to the pump inlet fitting, and proceed. Reagent water, known to be low TOC, is ideal to be utilized for flushing the 5000TOC Sensor. Page 6

13 Figure Final Installation of the Thornton 5000TOC Calibration Kit 1. Connect the Reference Conductivity sensor smart connector to the 770MAX patch cable ( ). 2. Connect the other end of the 770MAX patch cable to an unused calibrated channel on the 770MAX meter. If no channel is available, then another calibrated 770MAX should be utilized. 3. Set up a measurement to display the Reference Sensor temperature and temperature compensated conductivity. For example, Measurement B could be set to display conductivity and Measurement C could be set to display temperature (see suggested 770MAX configuration in Section 8.). 4. Turn on the pump. The setup should now be such that water continuously circulates through the Blank Water bottle and the 5000TOC Sensor. 5. Adjust the flow rate to its maximum rate. This may be done by adjusting the needle valve knob fully counterclockwise until it stops. 6. The system is now ready for temperature calibration. Turn off the pump and proceed to the next section MAX Configuration The 770MAX accepts input from multiple sensors from which it computes values in many different units. This extensive variety is supported by a multi-line scrolling display and the instrument's "custom names" feature, which allows labeling of each measurement. To simplify the calibration process, it is recommended that the 770MAX be configured to display the following measurements and units (see, 770MAX Multiparameter Analyzer/Transmitter Instruction Manual, part number 84372): Measurement A: TOC displayed in ppb (parts per billion) Measurement B: Conductivity displayed in µs/cm (resolution: 3 decimal places) Measurement C: Temperature displayed in degrees Celsius ( C) (resolution: 2 decimal places) Measurement D: Reference Conductivity displayed in µs/cm (resolution: 3 decimal places) Measurement E: Reference temperature displayed in degrees Celsius ( C) (resolution: 2 decimal places) Page 7

14 To simplify the calibration process, this procedure will use conductivity in units of µs/cm and temperature in degrees Celsius. 9. Temperature Calibration This procedure is to be performed on T1 and T2, which correspond to the temperature sensors embedded in the two conductivity sensors. If conductivity or TOC calibrations are to be done, temperature calibration must be done first T1 Temperature Verification Before Calibration 1. Turn off the UV lamp (if not already off) by pressing the lamp button, located on the front panel of the 5000TOC Sensor, so that it s green LED is off. The lamp shall remain off throughout this procedure. 2. Turn on the pump. Water should now be continuously circulating through the pump, 5000TOC Sensor, Reference Conductivity Sensor, the water container, and back through the pump. 3. Allow the pump to run until thermal equilibrium is achieved, typically 1/2 hour (less if the lamp has been off). 4. Record into Table 2 the Reference Sensor temperature (to two decimal places) displayed on the 770MAX, T ref. 5. On the 770MAX, press Menu (exit). 6. Press or to select Calibrate. Press Enter (or Page Down). 7. Press or to select Sensor. Press Enter (or Page Down). 8. Enter Output Hold Time. Press Enter (or Page Down). 9. Press or to select the letter for the measurement of the 5000TOC Sensor being calibrated. Press Enter. 10. Press or to select 1-point. Press Enter or. 11. Press or to select T1. Press Page Down. 12. Record the temperature of T1 into Table 2, T before. 13. Compute the difference, T ref - T before. 14. If the absolute value of the difference is less than the limit, then a temperature calibration is not required. Press Menu (exit) to return to measurement mode and proceed to Section 9.3. T2 Temperature Verification Before Calibration. 15. If T1 calibration is required, remain in this Calibration menu and proceed immediately to Section 9.2. T1 Temperature Calibration T1 Temperature Calibration 1. If T1 requires adjustment, enter the temperature of the Reference Sensor into the 770MAX CALIBRATE SENSOR MENU where it displays Cal point 1= To avoid inaccuracy due to changes in ambient temperature, this value should be entered as soon as practical after reading the Reference Sensor. 2. Press Page Down. 3. Wait for calibration to complete. 4. Enter the correct date (if needed). Press Page-Down. 5. Select YES to save the new calibration data within the 5000TOC Sensor. Press Enter. Note: if NO is selected, the data will only be temporarily saved, and will revert back to the previous value should power be lost, or the 5000TOC Sensor patch cable is unplugged. 6. Press Menu (exit) to return to measurement mode. 7. Record the temperature of T1, T after in Table Verify that T ref T after is within limits. 9. Verify that the new computed temperature factor is within acceptable limits by performing the following steps: 10. Press Menu (exit) to go to MAIN MENU. 11. Press or to select Measurement. Press Enter (or Page Down). 12. Press or to select the letter for the measurement of the 5000TOC Sensor displaying temperature. 13. Press Page Down. 14. Record Adder1 in Table 2. Page 8

15 15. Verify that this value is within acceptable limits. If this value exceeds limits, repeat Section Return the 770MAX to measurement mode by pressing the Menu (exit) key T2 Temperature Verification Before Calibration 1. Record into Table 2 the Reference Sensor temperature (to two decimal places) displayed on the 770MAX, T ref. 2. On the 770MAX, press Menu (exit). 3. Press or to select Calibrate. Press Enter (or Page Down). 4. Press or to select Sensor. Press Enter (or Page Down). 5. Enter Output Hold Time. Press Enter (or Page Down). 6. Press or to select the letter for the measurement of the 5000TOC Sensor being calibrated. Press Enter. 7. Press or to select 1-point. Press Enter or. 8. Press or to select T2. Press Page Down. 9. Record the temperature of T2 into Table 2, T before. 10. Compute the difference, T ref - T before. 11. If the absolute value of the difference is less than the limit, then temperature calibration is not required, and proceed to Section 10. Conductivity Calibration. 12. If T2 calibration is required, remain in this Calibration menu and proceed immediately to Section 9.4. T2 Temperature Calibration T2 Temperature Calibration 1. If T2 requires adjustment, enter the temperature of the Reference Sensor into the 770MAX. To avoid inaccuracy due to changes in ambient temperature, this value should be entered as soon as practical after reading the Reference Sensor. 2. Press page down. 3. Wait for calibration to complete. 4. Record the temperature of the T2, T after in Table Enter the correct date (if needed). Press Page-Down. 6. Select YES to save the new calibration data within the 5000TOC Sensor. Press Enter. Note: if NO is selected, the data will only be temporarily saved, and will revert back to the previous value should power be lost, or the 5000TOC Sensor patch cable is unplugged. 7. Press Menu (exit) to return to measurement mode. 8. Verify that T ref T after is within limits. 9. Verify that the new computed temperature factor is within acceptable limits by performing the following steps: 10. Press Menu (exit) to go to MAIN MENU. 11. Press or to select Measurement. Press Enter (or Page Down). 12. Press or to select the letter for the measurement of the 5000TOC Sensor displaying temperature. 13. Press Page Down. 14. Press Enter until the curser is under the 1 on the Adder1 field (example: Adder1 ). 15. Press or to change the field to Adder Record Adder2 value on Table Verify that this value is within acceptable limits. If this value exceeds limits, repeat Section Return the 770MAX to measurement mode by pressing the Menu (exit) key. 10. Conductivity Calibration Temperature Calibration is complete. This procedure is to be performed on C1 and C2 which correspond to the two conductivity sensors within the unit. Page 9

16 10.1. C1 Conductivity Verification Before Calibration 1. Verify the UV lamp is off. 2. Continuing with the same Blank Water from the Temperature Calibration, verify that the water quality is <1.5 µs/cm (>0.67 MΩ-cm) as reported by Reference Conductivity sensor. If the water conductivity has increased above 1.5 µs/cm, then turn off the pump, install a fresh bottle of Blank Water, and restart the pump. Note that the conductivity sensors can be calibrated above 1.5 µs/cm with no adverse measurement impact, but we recommend a lower conductivity calibration to emulate process water. 3. If the same bottle of Blank Water is still being used, then continue to the next step. Otherwise, allow the pump to run until complete mixing is achieved, typically 30 minutes. 4. Record in Table 3 the conductivity (at least three significant figures) of the Reference Conductivity Sensor displayed on the 770MAX, C ref. 5. Press Menu (exit) to go to MAIN MENU. 6. Press or to select Calibrate. Press Enter (or Page Down). 7. Press or to select Sensor. Press Enter (or Page Down). 8. Enter Output Hold Time. Press Enter (or Page Down). 9. Press or to select the letter for the measurement of the 5000TOC Sensor being calibrated. Press Enter. 10. Press or to select 1-point. Press Enter or. 11. Press or to select C1. Press Page Down. 12. Record the conductivity of C1 into Table 3, C before. 13. Compute the % difference according to eq ( C - C ) before C ref ref 14. If the absolute value of the % difference is less than the limit, then a C1 conductivity calibration is not required. Press Menu (exit) to return to measurement mode and proceed to Section C2 Conductivity Verification Before Calibration. 15. If C1 calibration is required, proceed to Section C1 Conductivity Calibration C1 Conductivity Calibration 1. If C1 requires adjustment, enter the conductivity of the Reference Sensor into the 770MAX. This value should be entered as soon as practical after reading the Reference Sensor. 2. Press page down. 3. Wait for calibration to complete. 4. Enter the correct date (if needed). Press Page-Down. 5. Select YES to save the new calibration data within the 5000TOC Sensor. Press Enter. Note: if NO is selected, the data will only be temporarily saved, and will revert back to the previous value should power be lost, or the 5000TOC Sensor patch cable is unplugged. 6. Press Menu (exit) to return to measurement mode. 7. Record the conductivity of the C1, C after in Table Verify that the % difference is within limits. 9. Verify that the new computed conductivity factor is within acceptable limits by performing the following steps: 10. Press Menu (exit) to go to MAIN MENU. 11. Press or to select Measurement. Press Enter (or Page Down). 12. Press or to select the letter for the measurement of the 5000TOC Sensor displaying conductivity. 13. Press Page Down. 14. Record Multiplier1 in Table 3. eq 2 Page 10

17 15. Verify that the Multiplier1 is within acceptable limits (noted in Table 3). If Multiplier1 exceeds limits, repeat the Section Return the 770MAX to measurement mode by pressing the Menu (exit) key C2 Conductivity Verification Before Calibration 1. Using Table 3, record the exact conductivity (at least three significant figures) of the Reference Conductivity Sensor displayed on the 770MAX, C ref. 2. Press Menu (exit) to go to MAIN MENU. 3. Press or to select Calibrate. Press Enter (or Page Down). 4. Press or to select Sensor. Press Enter (or Page Down). 5. Enter Output Hold Time. Press Enter (or Page Down). 6. Press or to select the letter for the measurement of the 5000TOC Sensor being calibrated. Press Enter. 7. Press or to select 1-point. Press Enter or. 8. Press or to select C2. Press Page Down. 9. Record the conductivity of C2 into Table 3, C before. 10. Compute the % difference according to eq If the absolute value of the % difference is less than the limit, then a C2 conductivity calibration is not required. Press Menu (exit) to return to measurement mode and proceed to Section 11. TOC Calibration. 12. If C2 calibration is required, proceed to Section C2 Conductivity Calibration 1. If C2 requires adjustment, enter the conductivity of the Reference Sensor into the 770MAX. This value should be entered as soon as practical after reading the Reference Sensor. 2. Press page down. 3. Wait for calibration to complete. 4. Record the conductivity of the C2, C after in Table Enter the correct date (if needed). Press Page-Down. 6. Select Yes to save the new calibration data within the 5000TOC Sensor. Press Enter. Note: if No is selected, the data will only be temporarily saved, and will revert back to the previous value should power be lost, or the 5000TOC Sensor patch cable is unplugged. 7. Press Menu (exit) to return to measurement mode. 8. Verify that the % difference is within limits. 9. Press Menu (exit) to go to MAIN MENU. 10. Verify that the new computed conductivity factor is within acceptable limits by performing the following steps: 11. Press Menu (exit) to go to MAIN MENU. 12. Press or to select Measurement. Press Enter (or Page Down). 13. Press or to select the letter for the measurement of the 5000TOC Sensor displaying conductivity and press Page Down. 14. Press Enter until the curser is under the 1 on the Multiplier1 field (example: Multiplier1). 15. Press or to change the field to Multiplier Record the Multiplier2 value in Table Verify that the Multiplier2 is within acceptable limits (noted in Table 3). If Multiplier2 exceeds limits, repeat the Section Return the 770MAX to measurement mode by pressing the Menu (exit) key. 19. Turn off the pump. Conductivity Calibration is now complete. Page 11

18 11. TOC Calibration Slope calibration is performed using a solution of known TOC concentration. Mettler-Toledo Thornton provides the 5000TOC Calibration Solution Set (P/N ) with a bottle of 500 ppb (sucrose) calibration solution for this purpose. The kit also contains a bottle of 250 ppb (sucrose) for verification TOC As Found 1. Disconnect Tube D and remove it completely from the assembly. See Figure Disconnect Tube C and remove it completely from the assembly. 3. Disconnect the end of Tube B that is connected to the Reference Sensor assembly and connect it to the needle valve on the Sample Inlet of the 5000TOC Sensor. B A C D NEEDLE VALVE Figure 7 4. Remove the Transfer Cap from the Blank Water bottle and place it on the 500 ppb Sucrose Solution bottle. 5. Install the 500 ppb Sucrose Solution bottle on the Bottle Holder bracket. 6. Connect the original drain tube to the Sample Outlet of the 5000TOC Sensor and run it to atmospheric drain. It is critical not to re-circulate this solution. 7. Turn on the pump. Water should now be visible at the drain line at the atmospheric drain. Note: the 500 ml volume of the calibration solution will provide more than 20 minutes of operation at the nominal flow rate of 20 ml/min. 8. Adjust the flow rate to 20 ml/min per Section 13. Setting Sample Flow Rate. 9. Verify that UV Lamp is on. The lamp must be on to measure and display TOC. 10. Allow the 500 ppb solution to flow through the 5000TOC sensor until the readings stabilize, approximately 3 to 5 minutes. 11. Record the TOC of the Reference Solution into Table 4, TOC ref. 12. Record the TOC of the 5000TOC Sensor into Table 4, TOC before. 13. Compute the difference, TOC ref - TOC before. 14. If the absolute value of the difference is less than the limit, then a TOC calibration is not required. Proceed to Section 12. Restore to Operation. 15. If TOC calibration is required, proceed to Section TOC Calibration Procedure. Page 12

19 11.2. TOC Calibration Procedure 1. If the TOC calibration requires adjustment, enter the TOC of the Reference Solution into the 770MAX. This value should be entered as soon as practical after reading the Reference Sensor. 2. Press Menu (exit) to go to MAIN MENU. 3. Press or to select Calibrate. Press Enter (or Page Down). 4. Press or to select Sensor. Press Enter (or Page Down). 5. Enter Output Hold Time. Press Enter (or Page Down). 6. Press or to select the letter for the measurement of the 5000TOC Sensor being calibrated. Press Enter. 7. Press or to select 1-point. Press Enter or. 8. Press or to select TOC. Press Page Down. 9. Press or to select Slope. Press Page Down. 10. Enter the TOC of the Reference Solution into the 770MAX. 11. Press page down. 12. Wait for calibration to complete. 13. Enter the correct date (if needed). Press Page-Down. 14. Select YES to save the new calibration data within the 5000TOC Sensor. Press Enter. Note: if NO is selected, the data will only be temporarily saved, and will revert back to the previous value should power be lost, or the 5000TOC Sensor patch cable is unplugged. 15. Return the 770MAX to measurement mode by pressing the Menu (exit) key. 16. Record the TOC, TOC after in Table Verify that difference is within limits. 18. If difference is acceptable; proceed to Section 11.3 TOC Calibration Verification Procedure. 19. If the difference is not within acceptable limits, repeat Section TOC Calibration Procedure TOC Calibration Verification Procedure 1. Turn off the pump. 2. Remove the transfer cap from the Reference Solution, install the 250 ppb Sucrose Solution bottle on the bracket, and attach cap to the new solution. 3. Turn on the pump. 4. Verify the flow is 20 ml/min. 5. Wait until the displayed reading stabilizes, typically 3 to 5 minutes. 6. Record the TOC of the 250 ppb solution in Table Record the measured TOC displayed on the 770MAX in Table Verify that the system response matches the TOC concentration listed on the container within the limits stated in Table If the limits are not acceptable, repeat sections 11.2 and If the limits are acceptable, the procedure is completed. Proceed to Section 12. Restore to Operation. 12. Restore to Operation TOC Calibration is complete. Caution: Be sure to adjust needle valve to less than 3 turns from closed position before re-installing into process line! 1. Restore the connections to the unit in the same operational configuration prior to calibration. 2. Remove the TOC System Suitability Test Kit. 3. Connect the process water line to the 5000TOC Sensor sample inlet. Page 13

20 4. Turn on the process water supply to the 5000TOC Sensor using the shut off valve (recommended in the 5000TOC Sensor Instruction Manual). Insure the process water supply is turned on by viewing the discharge line at the atmospheric drain. 5. Adjust the flow rate to 20 ml/min per Section 13 Setting Sample Flow Rate. 6. Turn on the UV Lamp if not already on. 13. Setting Sample Flow Rate At various times during the calibration of TOC the flow rate needs to be set using this procedure Flow Rate Adjustment Procedure 1. Press Menu (exit) to go to MAIN MENU. 2. Scroll to the Measurements menu and press Enter. 3. Select the measurement letter (A, B, C ) for the TOC measurement of interest. 4. Press Page Down (twice) until the Press 5 for TOC menu appears. 5. Press enter to activate Set Flow Rate option. Change this setting to Yes. This immediately initiates the Set Flow Rate mode at the sensor. It will stay in this mode until the user changes the Set Flow Rate: Yes back to Set Flow Rate: No. If the user does not set the flow rate once the Set Flow Rate mode has been initiated, the 770MAX will time out, at which time the mode automatically goes back to Set Flow Rate: No and the instrument returns to the measurement display screen. 6. The flow rate is set by adjusting the needle valve installed on the sample inlet connection of the 5000TOC Sensor (see Figure 8). NOTE: Clockwise (CW) to decrease flow, counterclockwise (CCW) to increase flow. Figure 8. Needle Valve Location 7. In this mode, the LEDs on the front cover of the 5000TOC Sensor act as a bar-graph indicator as described according to Section and Figure Adjust the inlet Needle Valve s knob until all four LEDs light on the 770MAX. 9. Go to the 770MAX and change Set Flow Rate: Yes to Set Flow Rate: No. The flow rate setting operation is now complete. Page 14

21 13.2. Flow Rate Indication The flow rate value is indicated by the four LEDs (lights) on the front of the 5000TOC Sensor. Fault Error Sensor Status UV Lamp On = LED illuminated = LED not illuminated The flow rate is set properly when all LEDs are illuminated. When the flow rate is at 20 ml/min (±2.5%), all four LEDs will be on. If the flow rate is too high, the top LEDs will be on. Turn the adjustment knob clockwise. If the flow rate is too low, the bottom LEDs will be on. Turn the adjustment knob counterclockwise. If the flow rate is far from 20 ml/min, only the top (Fault) LED or the bottom (UV Lamp) LED will be on. As the flow rate approaches the correct value, more LEDs will turn on. For example, if the flow rate is too high, as the adjustment knob is turned clockwise, first the Fault LED will be on, then the Fault and Error LEDs will be on, then the Fault, Error and Sensor Status LEDs will be on and finally all four LEDs will be on when the flow rate is 20 ml/min. Very LOW LOW NEAR SET SET NEAR SET HIGH Very HIGH Fault Error Sensor Status UV Lamp On = LED illuminated = LED not illuminated Figure 9. LED Indicator for Flow rate Page 15

22 14. Worksheet 1: Part Number and Serial Number Data Use the table below to record part number, serial number, of the Unit Under Test (UUT) and Reference equipment. Write N/A (not applicable) in spaces not used. 5000TOC Sensor Unit Under Test (UUT) Data Part Number Serial Number ID Tag (if available) 770MAX Meter Reference Data Part Number Serial Number ID Tag (if available) Date Last Calibrated Calibrated Due Date Reference Conductivity/Temperature Sensor Data Part Number Serial Number Date Last Calibrated Calibrated Due Date 500 ppb TOC Solution 250 ppb TOC Solution Reference TOC Solution Manufacturing Dates Signature Date Comments Page 16

23 15. Worksheets for Calibration and Verification Table 1. Flow Sensor Verification and Calibration Section A: Flow Sensor Verification Before Calibration Calibration point and conditions As found ~20 ml/min Volume (ml) Elapsed Time (sec) Reference Flow Rate F ref (ml/min) As Found Flow Rate F before (ml/min) F ref - F before (ml/min) Limit (ml/min) ±1 Within Limits or Adjustment Required Calibration point and conditions Volume (ml) Section B: Flow Sensor Calibration Elapsed Time (sec) Reference Flow Rate (ml/min) #1 ~24 ml/min F ref1 #2 ~16 ml/min F ref2 Section C: Flow Sensor Verification After Calibration Calibration point and conditions After ~20 ml/min Volume (ml) Elapsed Time (sec) Reference Flow Rate F ref (ml/min) After Flow Rate F after (ml/min) F ref - F after (ml/min) Limit (ml/min) ±1 Within Limits or Adjustment Required Signature Date Comments Page 17

24 Table 2. Temperature Sensor Verification and Calibration Reference Temperature T ref, ( C) As Found Temperature T before ( C) Measurement Type and Sensor # T ref - T before ( C) Limit ( C) Temperature T1 ±0.25 Temperature T2 ±0.50 Within Limits or Adjustment Required Temperature After T after ( C) Temperature Calibration Factor Limits Calibration Factors Adder Adder Limit T1 Adder1-5.0 to +5.0 T2 Adder2-5.0 to +5.0 Signature Date Comments Within Limits (Y or N) Page 18

25 Table 3. Conductivity Sensor Verification and Calibration Reference Conductivity C ref (µs/cm) As Found Conductivity C before (µs/cm) 100 ( C - C ) before Measurement Type and Sensor # Cref (%) Limit (%) Conductivity C1 ±2.0 Conductivity C2 ±5.0 ref Within Limits or Adjustment Required Conductivity After C after (µs/cm) Conductivity Calibration Factor Limits Calibration Factors Multiplier Multiplier Limit C1 Multiplier C2 Multiplier Signature Date Comments Within Limits (Y or N) Page 19

26 Table 4. TOC As Found Verification Reference TOC ref (ppb) As Found TOC before (ppb) TOC ref - TOC before (ppb) Limit (ppb TOC, 500 ppb ±50 Signature Date Comments Within Limits or Adjustment Required Page 20

27 Table 5. TOC Calibration and Verification Reference TOC ref (ppb) TOC after TOC ref TOC after (ppb) Limit(ppb) (ppb) TOC, 500 ppb ±50 TOC, 250 ppb ±25 Signature Date Comments Within Limits? (Y or N) Page 21

28 Page 22 This page is intentionally blank.

29

30 Mettler-Toledo Thornton, Inc. 36 Middlesex Turnpike Bedford, MA Tel Fax Toll Free PURE Part Rev.B 01/07

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin Electronic Control Manual 216 Sensors and Stats Section S Technical Bulletin CDS-2000 Issue Date 0393 CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin Introduction 3 Pre-Verification

More information

Table of Contents. Sensor Calibration and Troubleshooting CDS4000 CO 2. Introduction 1. Handling Information. Calibration 2.

Table of Contents. Sensor Calibration and Troubleshooting CDS4000 CO 2. Introduction 1. Handling Information. Calibration 2. FANs 216, 1628.3 Technical Bulletin CDS4000 Issue Date 0797 CDS4000 CO 2 Sensor Calibration and Troubleshooting Table of Contents Introduction 1 Handling Information 1 Calibration 2 Preparation 2 Cautions

More information

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel TANK MANAGER FOR TWO TANKS OPERATING MANUAL 10/31/11 C-More T6C L color touch panel 1 TABLE OF CONTENTS GENERAL...3 INSTALLATION...4 STONE TEST PROCEDURE...7 OPERATIONAL SUMMARY...7 AUTO CARBONATION...10

More information

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax OPERATION MANUAL Zutron Medical, LLC 17501 W 98 th St #40-27 Lenexa, KS 66219 Phone 877-343-5873 Fax 913-967-5944 ZUT-Lab-004-30004 REV. 03312017 Table of Contents 2 Introduction 1. Intended Use 2. Labels,

More information

ACV-10 Automatic Control Valve

ACV-10 Automatic Control Valve ACV-10 Automatic Control Valve Installation, Operation & Maintenance General: The Archer Instruments ACV-10 is a precision automatic feed rate control valve for use in vacuum systems feeding Chlorine,

More information

Model 130M Pneumatic Controller

Model 130M Pneumatic Controller Instruction MI 017-450 May 1978 Model 130M Pneumatic Controller Installation and Operation Manual Control Unit Controller Model 130M Controller is a pneumatic, shelf-mounted instrument with a separate

More information

Overview. Front Panel: Keypad and Display

Overview. Front Panel: Keypad and Display Overview The GA-200B is an analyzer that integrates a gas sampling system with sensors to measure and display the concentrations of oxygen and carbon dioxide in a sample as the percentage of a gas in the

More information

STANDARD OPERATING PROCEDURES. Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2.

STANDARD OPERATING PROCEDURES. Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2. STANDARD OPERATING PROCEDURES Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2.5 ) Sampler (RAAS) AMBIENT AIR MONITORING PROGRAM for the 130 LIBERTY STREET DECONSTRUCTION

More information

ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS Dionex Corporation

ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS Dionex Corporation ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS 2000 Dionex Corporation Document No. 031277 Revision 03 April 2000 2000 Dionex Corporation All rights reserved worldwide. Printed in the United States of

More information

Reverse Osmosis System Installation Guide and Operation Manual.

Reverse Osmosis System Installation Guide and Operation Manual. Reverse Osmosis System Installation Guide and Operation Manual. Table of Contents 1 Introduction...2 2 Installation...3 2.1 Feed water connection...3 2.2 Permeate and concentrate plumbing...3 2.3 Electrical

More information

Met One E-BAM Particulate Monitor

Met One E-BAM Particulate Monitor STANDARD OPERATING PROCEDURES Met One E-BAM Particulate Monitor AMBIENT AIR MONITORING PROGRAM for the 130 LIBERTY STREET DECONSTRUCTION PROJECT LOWER MANHATTAN DEVELOPMENT CORPORATION 1 Liberty Plaza

More information

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems Corporation Issued: 5/09 P/N: 600104 Rev. 2 Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems The Gas Addition Accessory permits the

More information

Operation Manual - PN A MENSOR MODEL 73 SHOP AIR BOOSTER

Operation Manual - PN A MENSOR MODEL 73 SHOP AIR BOOSTER Operation Manual - PN 0017946001 A MENSOR MODEL 73 SHOP AIR BOOSTER Mensor Model 73 Shop Air Booster System (750 psi Version) April 23, 2012 Trademarks / Copyright Mensor is a registered trademark of Mensor

More information

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4 Calibration....

More information

Using the UltraRAE. Firmware 2.35

Using the UltraRAE. Firmware 2.35 Using the UltraRAE Firmware 2.35 Training Agenda UltraRAE features Setting up the UltraRAE Turning on the UltraRAE Idle Operation RAE-Sep Tubes Prepping for a measurement Taking a measurement Alarm modes

More information

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings SomnoSuite FAQ V1.3 January 2015 Setup 1. How do I connect the SomnoSuite to my oxygen source? 2. Is there a way to speed up the downward movement of the pusher block when setting the empty position? 3.

More information

RAM 4021 Operation Manual

RAM 4021 Operation Manual RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For your safety...3 Description...3 Set-up mode...4 Annunciator lights/alarms...4 Operation...5 Calibration...6

More information

Columbus Instruments

Columbus Instruments 0215-003M Portable O 2 /CO 2 /CH 4 Meter User s Manual Columbus Instruments 950 NORTH HAGUE AVENUE TEL:(614) 276-0861 COLUMBUS, OHIO 43204, USA FAX:(614) 276-0529 1 www.colinst.com TOLL FREE 1-800-669-5011

More information

with O 2 Controller Instruction Manual

with O 2 Controller Instruction Manual 14500 Coy Drive, Grass Lake, Michigan 49240 734-475-2200 E-mail: sales@coylab.com www.coylab.com Hypoxic Cabinets (In-Vivo / In-Vitro) with O 2 Controller Instruction Manual Index Page Warranty 2 Warnings

More information

FAULT CODE TROUBLESHOOTING INDEX

FAULT CODE TROUBLESHOOTING INDEX FAULT CODE TROUBLESHOOTING INDEX 1. Display indicates Change Filters 2. Display indicates Drip Tray Full Continuous Alarm will Sound 3. Display indicates Cold Fault 4. Display indicates Hot Fault 5. Display

More information

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021-PR Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021-PR For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4

More information

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator Procedure No: SOP-034 Revision No: 1.0 Revised Dec. 29, 2010 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE This procedure is intended to describe the operations of the Dasibi model 5008 calibrator. The Dasibi

More information

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler.

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler. Exercise 5-2 Bubblers EXERCISE OBJECTIVE Learn to measure the level in a vessel using a bubbler. DISCUSSION OUTLINE The Discussion of this exercise covers the following points: Bubblers How to measure

More information

Model GFC 7000T CO 2 Analyzer (Addendum to GFC 7000TA Manual, PN 07272)

Model GFC 7000T CO 2 Analyzer (Addendum to GFC 7000TA Manual, PN 07272) Manual Addendum Model GFC 7000T CO 2 Analyzer (Addendum to GFC 7000TA Manual, PN 07272) P/N M07273 DATE 06/11/13 TELEDYNE ELECTRONIC TECHNOLOGIES Analytical Instruments 16830 Chestnut Street City of Industry,

More information

SPECIFICATIONS APCEPH1

SPECIFICATIONS APCEPH1 APCEPH1 ph CONTROLLER SPECIFICATIONS APCEPH1 Input voltage 120 Volts AC Maximum amperage 14.5 amps @ 120 VAC ph Accuracy +/- 0.2 ph ph Control range Adjustable 4.5 8.5 ph Weight < 1 lbs Dimensions 3" x

More information

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Sensitivity: 1x10 11 A m 2 Applied DC fields: 0 T to 5 T Applied AC fields: 0 G to 3 G (zero-to-peak), 0.01 Hz to 1000 Hz Temperatures

More information

1020 Industrial Drive, Orlinda, TN fax

1020 Industrial Drive, Orlinda, TN fax Operation Manual Tank Distribution System A-UPT Series 615-654-4441 sales@specialtyh2o.com 615-654-4449 fax TABLE OF CONTENTS Section 1 GENERAL 1.2 Warnings and Cautions... 1 1.2 Theory of Operation...

More information

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4 Calibration....

More information

RAM 4021-DPX Operation Manual

RAM 4021-DPX Operation Manual RAM 4021-DPX Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS ABL 4021-DPX / RAM 4021-DPX For Your Safety... 3 Description... 3 Setup Mode... 4 Lights/Alarms... 4 Operation...

More information

Gas Diluter Portable Gas Dilution System User s Manual

Gas Diluter Portable Gas Dilution System User s Manual Gas Diluter Portable Gas Dilution System User s Manual Defiant Technologies, Inc. 2015 Vol.1 This page is intentionally left blank. Gas Diluter Portable Gas Dilution System User s Manual Defiant Technologies,

More information

JOLLY2. Installation user s manual. 6 different operating modes selectable. version 3.3. DATA TO BE FILLED OUT BY THE INSTALLER (Page 1)

JOLLY2. Installation user s manual. 6 different operating modes selectable. version 3.3. DATA TO BE FILLED OUT BY THE INSTALLER (Page 1) ENGLISH ENGLISH ENGLISH ENGLISH Installation user s manual Warning! electrical scheme modified JANUARY 2005 version 3.3 JOLLY2 DATA TO BE FILLED OUT BY THE INSTALLER (Page 1) 6 different operating modes

More information

700 SERIES BMS II BUBBLER MONITORING SYSTEM

700 SERIES BMS II BUBBLER MONITORING SYSTEM 1 700 SERIES BMS II BUBBLER MONITORING SYSTEM VISIT OUR WEBSITE AT SIGMACONTROLS.COM 700 SERIES BMS II MANUAL 062114 2 TABLE OF CONTENTS INTRODUCTION 3 Physical Specifications WIRING 5 PROGRAMMING & SETUP..

More information

SPECIFICATIONS ATTENTION

SPECIFICATIONS ATTENTION VPS 504 S06 Installation Manual - P/N 80122 - Ed. 01/09 VPS 504 S06 and S05 Valve Proving System Installation Instructions VPS 1 6 Gases Natural gas, air and other inert gases. NOT suitable for butane

More information

Water Weir Flow Controller. Introduction. Safety Precautions. Mounting the Hardware

Water Weir Flow Controller. Introduction. Safety Precautions. Mounting the Hardware 57007-88 Introduction Safety Precautions This instruction sheet describes how to set up and use the Hach (Figure 1). A water weir is a device that raises or diverts water to regulate the flow. Hach s water

More information

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments Pegas 4000 MF Gas Mixer InstructionManual Contents I Table of Contents Foreword Part I Introduction 1 2 1 System overview... 2 2 Specifications... 3 Part II Installation 4 1 Rear panel connections...

More information

3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS. Item #31289

3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS. Item #31289 3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS Item #31289 The EASTWOOD 3 GALLON, OILLESS PANCAKE COMPRESSOR, with an Integral Air Regulator, efficiently supplies all compressed air requirements for

More information

O2100C Oxygen Measurement Module Technical Use Notes do not use other wall adapters with the O2100Cmodule. 10% / V 5% / V 2% / V 1% / V 10% / V

O2100C Oxygen Measurement Module Technical Use Notes do not use other wall adapters with the O2100Cmodule. 10% / V 5% / V 2% / V 1% / V 10% / V O2100C Oxygen Measurement Module The O2100C module measures the partial pressure of O2 and thus the module output is proportional to the pressure in the sample cell. Gas sampled must be free of liquids

More information

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300 Technical Note GA-300 Overview The GA-300 CO2 and O2 Gas Analyzer is easy to use, robust, and adaptable to human, animal, and plant applications. The GA-300 has two analog outputs to allow recording and

More information

RAM Operation Manual

RAM Operation Manual RAM 4021-1 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021-1 For Your Safety... 2 Description... 2 Setup Mode... 3 Lights/Alarms... 3 Operation... 4 Calibration...

More information

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure PPMS Service Note 1099-412 In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure Introduction: Loss of electricity to the PPMS EverCool II should not cause damage to

More information

User Manual. Quantos Automated Dosing Liquid Module

User Manual. Quantos Automated Dosing Liquid Module User Manual Liquid Module 1 Safety Information 1.1 Definition of warnings and symbols Signal Words WARNING for a hazardous situation with medium risk, possibly resulting in severe injuries or death if

More information

OPERATION MANUAL RTI RHS650 RTI TECHNOLOGIES, INC East Market Street York, PA Manual P/N

OPERATION MANUAL RTI RHS650 RTI TECHNOLOGIES, INC East Market Street York, PA Manual P/N OPERATION MANUAL RTI RHS650 RTI TECHNOLOGIES, INC. 4075 East Market Street York, PA 17402 Manual P/N 035-80589-02 Table of Contents Components... 2 Test... 4 Recovery/Recycling... 5 Evacuation... 7 Charging...

More information

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc. The HumiSys RH Generator Designed, built, and supported by InstruQuest Inc. Versatile Relative Humidity Generation and Multi-Sensor System The new HumiSys with single or dual RH probes capabilities is

More information

Model 106 DPI "Micro-switch" Installation and Operating Instructions

Model 106 DPI Micro-switch Installation and Operating Instructions Mid-West Instrument Model 106 DPI "Micro-switch" Installation and Operating Instructions BULLETIN NO. IM116DPImicro/09A Replaces --- INSPECTION Before installation carefully check the Model Number on each

More information

BI-680 Online Dissolved Oxygen Controller Instruction Manual

BI-680 Online Dissolved Oxygen Controller Instruction Manual BI-680 Online Dissolved Oxygen Controller Instruction Manual BANTE INSTRUMENTS CO., LTD BI-680 Online Dissolved Oxygen Controller 1 Introduction Thank you for selecting the BI-680 online dissolved oxygen

More information

Installation of Your SprayMaster System

Installation of Your SprayMaster System Installation of Your SprayMaster System 1. At the installation site, remove all equipment from the corrugated box and the polyethylene drum and replace the drum lid. Check the picture to identify each

More information

OC Panel High Limit Aquastat Kit, Manual Reset p/n

OC Panel High Limit Aquastat Kit, Manual Reset p/n OC Panel High Limit Aquastat Kit, Manual Reset p/n 233202 Instruction Sheet APPLICATION The OC (Option Control) Panel High Limit Aquastat Kit provides electronic temperature sensing in a UL limit-rated

More information

Agilent Auxiliary Gas Module AGM 2. User s Guide

Agilent Auxiliary Gas Module AGM 2. User s Guide Agilent Auxiliary Gas Module AGM 2 User s Guide Notices Agilent Technologies, Inc. 1992, 1996, 1999, 2010 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information

256 Pneumatic Pressure Indicator

256 Pneumatic Pressure Indicator 256 Pneumatic Pressure Indicator 51425699 Copyright 2002 Slope Indicator Company. All Rights Reserved. This equipment should be installed, maintained, and operated by technically qualified personnel. Any

More information

TEL/jlRE" Introduction. Display Features and Modes. Startup Procedure. Power-Up Procedure. Adjustment Modes

TEL/jlRE Introduction. Display Features and Modes. Startup Procedure. Power-Up Procedure. Adjustment Modes TEL/jlRE" Introduction The Telaire 7001 CO 2 /T emperature monitor (shown in Fi gt u e 1 below) is an easy to use hand-held instnunent, which provides stable and highly accurate readings due to Telaire

More information

Approved by Principal Investigator Date: Approved by Super User: Date:

Approved by Principal Investigator Date: Approved by Super User: Date: Approved by Principal Investigator Date: Approved by Super User: Date: Standard Operating Procedure BNC OAI 200 Lithographic Mask Aligner (Aligner 3) Version 2011 June 2 I. Purpose This Standard Operating

More information

The Discussion of this exercise covers the following points:

The Discussion of this exercise covers the following points: Exercise 5-3 Wet Reference Leg EXERCISE OBJECTIVE Learn to measure the level in a vessel using a wet reference leg. DISCUSSION OUTLINE The Discussion of this exercise covers the following points: Measuring

More information

TR Electronic Pressure Regulator. User s Manual

TR Electronic Pressure Regulator. User s Manual TR Electronic Pressure Regulator Page 2 of 13 Table of Contents Warnings, Cautions & Notices... 3 Factory Default Setting... 4 Quick Start Procedure... 5 Configuration Tab... 8 Setup Tab... 9 Internal

More information

THE UNIVENTOR 400 ANAESTHESIA UNIT

THE UNIVENTOR 400 ANAESTHESIA UNIT THE UNIVENTOR 400 ANAESTHESIA UNIT TABLE OF CONTENTS EDITION 3 Section 1 - WARRANTY & SERVICE 1.1. WARRANTY 2 1.2. DAMAGED SHIPMENTS 2 1.3. SERVICE 2 Section 2 - INTRODUCTION 2.1. INTRODUCTION 3 Section

More information

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator Procedure No: SOP-029 Revision No: 1.1 (December 29, 2010) Page No.: 1 of 7 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting, air quality

More information

R E D I C O N T R O L S

R E D I C O N T R O L S R E D I C O N T R O L S Operation & Maintenance Manual Portable Service Purger for Low Pressure Chillers Model: PSP-LP-1B For Refrigerants R-11, R-113, R-114 & R-123 & Other Similar Refrigerants File Literature

More information

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road PROPORTIONING VALVE Model 150 INSTRUCTION MANUAL March 2017 IMS Company 10373 Stafford Road Telephone: (440) 543-1615 Fax: (440) 543-1069 Email: sales@imscompany.com 1 Introduction IMS Company reserves

More information

AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST

AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST OIL-LESS TYPE IMPORTANT: PLEASE READ CAREFULLY BEFORE STARTING OPERATIONS. THE CONTENTS ARE FOR GENERAL INFORMATION OF ALL THE SIMILAR MODELS. Record

More information

HumiSys HF High Flow RH Generator

HumiSys HF High Flow RH Generator HumiSys HF High Flow RH Generator Designed, built, and supported by InstruQuest Inc. Versatile Relative Humidity Generation and Multi-Sensor System The HumiSys HF is a high flow version of the previously

More information

In Vivo Scientific, LLC INSTRUCTION MANUAL

In Vivo Scientific, LLC INSTRUCTION MANUAL CO 2 Controller In Vivo Scientific, LLC INSTRUCTION MANUAL CONTENTS CONTENTS...1 ABOUT THIS MANUAL...2 INTRODUCTION...2 Cautions and Warnings...2 Parts List...2 Unpacking...2 INSTRUMENT DESCRIPTION...3

More information

User s Guide. Pressure Controller for liquid delivery systems

User s Guide. Pressure Controller for liquid delivery systems Flow Control User s Guide Pressure Controller for liquid delivery systems Precise Pressure Control throughout the experiment Flow control Compatible with any perfusion system Ideal for Small Volume Delivery

More information

Operations Manual. Internal automatic Pressure control for orbital tube welding. By Exel Orbital Systems, Inc. Page

Operations Manual. Internal automatic Pressure control for orbital tube welding. By Exel Orbital Systems, Inc. Page Operations Manual Internal automatic Pressure control for orbital tube welding. By Exel Orbital Systems, Inc. Page SG1 Overview and Dimensions 5.25 / 133 mm PS Power Supply Connection 4.06 / 103 mm Top

More information

FX134, FX1234, FX3030 Problem and Solutions

FX134, FX1234, FX3030 Problem and Solutions FX134, FX1234, FX3030 Problem and Solutions Unit Will Not Power Up Turn the power switch on. The unit LCD will display revision program and filter life within 3 to 5 seconds after turning on unit. If this

More information

BUBBLER CONTROL SYSTEM

BUBBLER CONTROL SYSTEM BUBBLER CONTROL SYSTEM Description: The HDBCS is a fully automatic bubbler system, which does liquid level measurements in water and wastewater applications. It is a dual air compressor system with, air

More information

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved

More information

UBEC 1AT. AUTO TANK Fill System Installation, Operation, & Setup Instructions

UBEC 1AT. AUTO TANK Fill System Installation, Operation, & Setup Instructions Document Number: XE-ATA5PM-R1A UBEC 1AT AUTO TANK Fill System 08899155 Installation, Operation, & Setup Instructions Rev170906-EB-FRC PHYSICAL: 1302 WEST BEARDSLEY AVE ELKHART, IN 46514 WWW.ELKHARTBRASS.COM

More information

- Repeat calibration until desired results are reached otherwise set SW2 and SW1-8 open.

- Repeat calibration until desired results are reached otherwise set SW2 and SW1-8 open. Calibration Procedures on Cobe Centrysystem 3: I) Balance Chamber Calibration: - Verify that pressure regulators PRV and PRV2 are calibrated as well as pressure transducers PDI and PDO with switches SW2

More information

MASSACHUSETTS STATE POLICE CRIME LABORATORY. Certificate of Calibration Procedure for the Alcotest 9510

MASSACHUSETTS STATE POLICE CRIME LABORATORY. Certificate of Calibration Procedure for the Alcotest 9510 MASSACHUSETTS STATE POLICE CRIME LABORATORY Certificate of Calibration Procedure for the Alcotest 9510 Effective Date: May 10, 2017 Page 1 of 23 TABLE OF CONTENTS 1 INTRODUCTION... 3 2 ABBREVIATIONS...

More information

FOR INSTALLING CO 2 BLENDER KIT (P/N IN BEER SYSTEM

FOR INSTALLING CO 2 BLENDER KIT (P/N IN BEER SYSTEM IMI CORNELIUS INC One Cornelius Place Anoka, MN 55303-623 Telephone (800) 238-3600 Facsimile (612) 22-326 INSTALLATION INSTRUCTIONS FOR INSTALLING CO 2 BLENDER KIT (P/N 111612000 IN BEER SYSTEM SECONDARY

More information

ID-Cube. Installation and Operations Manual For ID-Cube Source and Power Supply. Revision 2 Document #:

ID-Cube. Installation and Operations Manual For ID-Cube Source and Power Supply. Revision 2 Document #: ID-Cube Installation and Operations Manual For ID-Cube Source and Power Supply Revision 2 Document #: 7.1.002 IonSense Inc. 999 Broadway Suite 404 Saugus, MA 01906 Copyright 2014 by IonSense Inc. All rights

More information

User Manual. Quantos Automated Dosing Powder Module

User Manual. Quantos Automated Dosing Powder Module User Manual Powder Module 1 Introduction Thank you for purchasing the Quantos dosing system. Quantos is based on the renowned and reliable XPE series balances from METTLER TOLEDO. Besides the highly accurate

More information

Heavy Duty Dissolved Oxygen Meter

Heavy Duty Dissolved Oxygen Meter User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 FAX 781.665.0780 - TestEquipmentDepot.com Introduction Congratulations

More information

Aquavar SOLO 2 Frequently Asked Questions

Aquavar SOLO 2 Frequently Asked Questions Aquavar SOLO 2 Frequently Asked Questions How do I size the Aquavar SOLO 2 for the appropriate pump/motor combination? Can I use a 208 Volt motor? Can I run the Aquavar SOLO 2 up to 80HZ? What are the

More information

Supplementary Operator s Manual 42/24-14 EN Rev. 3

Supplementary Operator s Manual 42/24-14 EN Rev. 3 Advance Optima Continuous Gas Analyzers AO2000 Series with Uras26, Magnos206, Caldos25, Caldos27 Designed per Category 3G for Measurement of Flammable Gases ( Safety Concept ) and Non-flammable Gases Supplementary

More information

1020 Industrial Drive, Orlinda, TN fax

1020 Industrial Drive, Orlinda, TN fax Operation Manual Ultrafiltration for High Purity Distribution K-A-HPTUF Series 615-654-4441 sales@specialtyh2o.com 615-654-4449 fax TABLE OF CONTENTS Section 1 GENERAL 1.1 Warnings and Cautions... 1 1.2

More information

Quick Installation Guide Doc nr.: C Date:

Quick Installation Guide Doc nr.: C Date: Quick Installation Guide Doc nr.: 9.17.093C Date: 13-05-2016 SCOPE OF THIS GUIDE mini CORI-FLOW instruments are highly accurate instruments for measuring and controlling the mass flow rate of liquids and/or

More information

RADIATION PROCEDURES MANUAL Procedure Cover Sheet

RADIATION PROCEDURES MANUAL Procedure Cover Sheet RADIATION PROCEDURES MANUAL Procedure Cover Sheet Procedure Title: Calibration of Eberline Portable Particulate Noble Gas Monitor Procedure Number: TSO-08-13-REV 1 Effective Date: July 1, 2008 Approved

More information

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante810 Benchtop Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante810 benchtop dissolved oxygen

More information

6800 Maintenance Instruction System Flush Procedure

6800 Maintenance Instruction System Flush Procedure Equipment Required FA74005 FA65318 FA900005 FA900003 Damper Drain Tube 6800 Cover Removal Tool Beaker 0.25 Litre Solvent Cleaning Bottle FA940021 Syringe Polypropylene 50 ml as required FA999045 Gloves

More information

Blue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION

Blue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION Blue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION Install your Blue River Technologies Port-A-Poly mixing system in a clean dry area.

More information

Budget Range Operators Handbook

Budget Range Operators Handbook Budget Range Operators Handbook BAMBI AIR COMPRESSORS LTD 152 Thimble Mill Lane Heartlands Birmingham B7 5HT United Kingdom Tel: 0121 322 2299 Fax: 0121 322 2297 Email: sales@bambi-air.co.uk www.bambi-air.co.uk

More information

MP15 Jockey Pump Controller

MP15 Jockey Pump Controller Setup and Operating Instructions MP15 Jockey Pump Controller This manual provides general information, installation, operation, maintenance, and system setup information for Metron Model MP15 Jockey Pump

More information

SCOPE OF THIS GUIDE STARTING-UP. 1 Check mini CORI-FLOW functional properties

SCOPE OF THIS GUIDE STARTING-UP. 1 Check mini CORI-FLOW functional properties Quick Installation Guide Doc nr.: 9.17.052H Date: 10-12-2015 SCOPE OF THIS GUIDE mini CORI-FLOW instruments are highly accurate instruments for measuring and controlling the mass flow rate of liquids and/or

More information

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation Page 1 of 19 : SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor Description and Principle of Operation The photoacoustic multi-gas monitor (INNOVA 1412, Innova AirTech Instruments, Denmark) is a highly

More information

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y INSTALLATION and INSTRUCTION MANUAL for QUALITY AIR BREATHING SYSTEMS Model 50 Systems Outfitted with ABM-725 Monitor M A R T E C H S E R V I C E S C O M P A N Y OFFICE: (507) 843-4700 P.O. BOX 7079 Toll

More information

6900 Maintenance Instruction System Flush

6900 Maintenance Instruction System Flush Equipment Required FA74005 Damper Drain Tube FA16005 Cover Removal Tool FA900005 Beaker 0.25 Litre FA900003 Solvent Cleaning Bottle FA940021 Syringe Polypropylene 50 ml as required FA999045 Gloves Latex

More information

OWNER S MANUAL NMB SERIES

OWNER S MANUAL NMB SERIES OWNER S MANUAL NMB SERIES NMB Manual V2.0.doc 1. PARTS IDENTIFICATION LIST 1. 5. 2. 4. 3. 6(a). 6(b). Figure 1.1 1. Powerpack with timer option (x1) 2. Cell (x1) 3. Cell Housing (x1) 4. O-ring (x1) 5.

More information

PremAire INSTRUCTIONS FOR VORTEX TUBE MODE OF OPERATION

PremAire INSTRUCTIONS FOR VORTEX TUBE MODE OF OPERATION DUAL SUPPLY REGULATOR ESCAPE CYLINDER VORTEX MAIN INLET PremAire INSTRUCTIONS FOR VORTEX TUBE MODE OF OPERATION WARNING THIS MANUAL MUST BE READ CAREFULLY BY ALL PERSONS WHO HAVE OR WILL HAVE THE RESPONSIBILITY

More information

TECHNICAL INFORMATION

TECHNICAL INFORMATION TECHNICAL INFORMATION Installation, Operation, and Maintenance Balston 75-96 and 75-98 Nitrogen Generation System Tower Pressure Gauge 5'-1" 3'-6" 7'-4" 6'-10 1/2" PLAN VIEW Figure 1 - Balston 75-96 and

More information

NIV EAU MATIC ST User Manual Made in Canada

NIV EAU MATIC ST User Manual Made in Canada NIV EAU MATIC ST User Manual Made in Canada www.niveaumatic.com info@niveaumatic.com -2- Table of contents Introduction Section 1 Page 4 Installation Section 2 Pages 5 to 10 2.1 Control box 2.2 Sensor

More information

Digital Vacuum Regulator

Digital Vacuum Regulator Temperature Control for Research and Industry Digital Vacuum Regulator User s Manual Model 300 INDEX SECTION PAGE 1. QUICK OPERATING INSTRUCTIONS........................... 3 Safety Notices.................................................

More information

Propane Conversion Kit Instructions

Propane Conversion Kit Instructions 604 8 0/ US/CA For heating engineers Propane Conversion Kit Instructions Logano G4 X gas-fired boiler This conversion kit and the accompanying instructions are for conversion of G4 X gas-fired boilers

More information

Agilent 1290 Infinity Pump Head Maintenance

Agilent 1290 Infinity Pump Head Maintenance Agilent 1290 Infinity Pump Head Maintenance Technical Note Agilent Technologies Notices Agilent Technologies, Inc. 2012-2015, 2016 No part of this manual may be reproduced in any form or by any means (including

More information

accidents which arise due to non-observance of these instructions and the safety information herein.

accidents which arise due to non-observance of these instructions and the safety information herein. 3 GALLON PANCAKE COMPRESSOR Model: 50959 CALIFORNIA PROPOSITION 65 WARNING: You can create dust when you cut, sand, drill or grind materials such as wood, paint, metal, concrete, cement, or other masonry.

More information

Instruction Manual No. 742, 8/98

Instruction Manual No. 742, 8/98 Instruction Manual No. 742, 8/98 Eclipse Ratio Regulators ES Series COPYRIGHT Copyright I997 by Eclipse Combustion, Inc. All rights reserved worldwide. This publication is protected by federal regulation

More information

Copyright 2004 by the Thomas G. Faria Corporation, Uncasville CT No part of this publication may by reproduced in any form, in an electronic

Copyright 2004 by the Thomas G. Faria Corporation, Uncasville CT No part of this publication may by reproduced in any form, in an electronic Copyright 2004 by the Thomas G. Faria Corporation, Uncasville CT No part of this publication may by reproduced in any form, in an electronic retrieval system or otherwise, without the prior written permission

More information

OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE

OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE 1 These instructions generally describe the installation, operation, and maintenance of subject equipment. The manufacturer reserves the right

More information

Exercise 2-2. Second-Order Interacting Processes EXERCISE OBJECTIVE DISCUSSION OUTLINE. The actual setup DISCUSSION

Exercise 2-2. Second-Order Interacting Processes EXERCISE OBJECTIVE DISCUSSION OUTLINE. The actual setup DISCUSSION Exercise 2-2 Second-Order Interacting Processes EXERCISE OBJECTIVE Familiarize yourself with second-order interacting processes and experiment with the finer points of controller tuning to gain a deeper

More information

Constant Pressure Inlet (CCN) Operator Manual

Constant Pressure Inlet (CCN) Operator Manual Constant Pressure Inlet (CCN) Operator Manual DOC-0125 Revision J 2545 Central Avenue Boulder, CO 80301-5727 USA C O P Y R I G H T 2 0 1 1 D R O P L E T M E A S U R E M E N T T E C H N O L O G I E S, I

More information

Armfield Distillation Column Operation Guidelines

Armfield Distillation Column Operation Guidelines Armfield Distillation Column Operation Guidelines 11-2016 R.Cox Safety SAFETY GLASSES ARE REQUIRED WHEN OPERATING THE DISTILLATION COLUMN Wear gloves when mixing alcohol feedstock The column will become

More information