ASHRAE-110 Type-Test Report for 1.5m Wide Bench-Type fumetop1500 FH-150 Fume Hood of TopAir, USA Ref: ANSI/ASHRAE

Similar documents
Fumehood Performance Test Report

Mott Manufacturing Limited

Statesville, NC. Results of AM ASHRAE 110 Laboratory Fume Hood Performance Evaluations Venturi Hood

BS EN :2003 Containment Factor deciphered

Mott Manufacturing Limited

APPENDIX A SAMPLE FHT SPECIFICATION (This recommended FHT specification is available from

University of North Florida Performance Requirements For Laboratory Fume Hoods. As Manufactured Performance Tests. As Installed Performance Tests

Testing the Vented, 7-Liter Induction Chamber. Report 6 April 2008

Development and Evaluation of an Air-Curtain Fume Cabinet with Considerations of its Aerodynamics

Testing the Vented, 2-Liter Induction Chamber. Report 14 October 2004

Improving the Effectiveness of LEV A HSL perspective

Validation of a Neutral Pressure Isolation Room

Nitrous Oxide as a Tracer Gas in the ASHRAE Standard

Measuring range Δp (span = 100%) Pa

FUME CUPBOARD PERFORMANCE CHECK

Kansas State University Fume Hood Operation

H V A C C o n t r o l s L a b o r a t o r y. Application Note

SURROGATE POWDER AND ASHRAE-110 TESTING OF A MULTI-PURPOSE ENCLOSURE FOR WEIGHING/SIEVING/FLUID BED DRYER APPLICATIONS

Laboratory Guidelines and Standards Application Note LC-125

LEV Inspection & Test Report

MANAGEMENT PROCEDURE FOR LABORATORY VENTILATION

OCCUPATIONAL SAFETY AND ENVIRONMENTAL HEALTH GUIDELINE

Performance Overview calibration Laboratory EP Instruments Messtechnik und Kalibrierung GmbH

AF100. Subsonic Wind Tunnel AERODYNAMICS. Open-circuit subsonic wind tunnel for a wide range of investigations into aerodynamics

Table of contents. I Drawings 33. A1 Design criteria 1 A1.1 Local thermal discomfort indices... 1

CALIBRATION SERVICES HUMIDITY DEW POINT AIR VELOCITY MASS FLOW TEMPERATURE CO 2 PRESSURE. (+34)

Group Quality Certification System of Performance Test for Clean Rooms (CR Mark)

THE AIRCRAFT SEAT AS INDOOR AIR QUALITY AND TEMPERATURE CONTROL SYSTEM

Dräger MSI P7 Combustion Gas Analysis

by Rob Chopowick General Manager, Con- Test

AIR-X COMPACT AIR-X ON-LINE LUBRICANT AERATION MEASUREMENT. Technical Brochure ON RUNNING ENGINES

S-Type Pitot Tube. Operating Instructions

P06 & P012 Pitot-Static Probes

1. Scope: This document describes the procedure for verifying the performance of vaccine vial monitors.

P8 to P60 Pitot-Static Probes

Astudy of the air quality in the breathing zone in a room with displacement ventilation

Agilent 7800 ICP-MS. Specifications and Typical Performance. Fast track metals analysis with Solution-Ready ICP-MS

FPG8601 Force Balanced Piston Gauge

Datasheet: K-30 ASCII Sensor

ESTATES OFFICE OPERATIONS WORKS INSTRUCTIONS WI 6. Fume Cupboards, Ventilated Cabinets, Local Exhaust Ventilation

Operating instructions Pitot Static Tube

LABORATORY SAFETY SERIES: Laboratory Hoods

FLOW CONSIDERATIONS IN INDUSTRIAL SILENCER DESIGN

How to Test Air Change Effectiveness

rpbsafety.com Samoset Rd Royal Oak, MI USA OF THE NOVA 3 ABRASIVE BLASTING SUPPLIED AIR RESPIRATOR

Subsonic Wind Tunnel 300 mm

Air Leakage Detailing & Awareness Services

"LABORATORY SAFETY SERIES"

Aqua-DP. Aqua-DP Waterproof Dissolved Oxygen - ph -mv Temperature Meter. TPS Australia - Quality hand-made instruments since 1968.

SafidDampers. Class 1A & Class 1 Low Leakage Volume Control Damper ( Flange Type ) SPK Standard Construction. Description SPK

CALMO

The primary purpose of laboratory fume hoods is to keep toxic or irritating vapors out of the general laboratory working area.

Report No. 371/ This report consists of 6 pages

Testing of Active Flow Control actuators at harsh environment

Single Phase Pressure Drop and Flow Distribution in Brazed Plate Heat Exchangers

Catalogue Extract RCP4 Cleanroom Type with Pulse Motor. 42P Motor type. size : 12mm 6 : 6mm 3 : 3mm

FM Series Low and Medium Velocity Gas, Oil and Combination Burners

Thorough Examination and Test of Local Exhaust Ventilation Systems to Regulation 9 of the COSHH Regulations 2002 (as amended)

Vaisala Pressure, RH & Temp. Transmitter. The instrument was operational upon receipt. The instrument was adjusted and calibrated.

Ribas 3-Tined & 5-Tined Cultivator Attachments. Manufactured by Ribas Agromecánica to fit. BCS Power Units. Operating Instructions

Electrochemical Sensor

EXPERIMENTAL ANALYSIS OF THE CONFLUENT BOUNDARY LAYER BETWEEN A FLAP AND A MAIN ELEMENT WITH SAW-TOOTHED TRAILING EDGE

ACCUMULATION OF HYDROGEN RELEASED INTO A VENTED ENCLOSURE - EXPERIMENTAL RESULTS

Tank Inspection Report

CTAG:e43bda4f8e48714fbb aefb192 Certificate of Calibration

Subsonic Wind Tunnel 300 mm

Dräger P7 Pressure, Leak and Leak Rate Measurement

TExT Thorough Examination & Test of LEV Local Exhaust Ventilation Systems

Electrochemical Sensor

SIMULATION OF ENVIRONMENTAL FLIGHT CONDITIONS

Laboratory Ventilation. Environmental, Health, and Safety

AF101 to AF109. Subsonic Wind Tunnel Models AERODYNAMICS. A selection of optional models for use with TecQuipment s Subsonic Wind Tunnel (AF100)

PLEA th Conference, Opportunities, Limits & Needs Towards an environmentally responsible architecture Lima, Perú 7-9 November 2012

Evaluation of Consumer Personal Protective Equipment: EMERGENCY ESCAPE MASKS

Influence of the external dynamic wind pressure on the ventilation of double facades

Dräger-Tube pump accuro Dräger Tube Pump

LABORATORY FUME HOOD SPECIFICATION AND MAINTENANCE PROGRAM

Aerodynamic Shape Design of the Bow Network Monitoring Equipment of High-speed Train

LEV GUIDANCE FOR SAFE USAGE, TESTING AND MAINTENANCE OF LOCAL EXHAUST VENTILATION SYSTEMS WITHIN ANGLIA RUSKIN UNIVERSITY RISK MANAGEMENT

Vibration isolated tables. Individually tailored to your requirements.

Draft. Equipment/Instrument

Navy Guidance and Tips for Using DOEHRS-IH Ventilation NAVY & MARINE CORPS PUBLIC HEALTH CENTER

Model NDA Differential pressure transmitter. Deltapi N Series A complete range of pneumatic pressure transmitters

Earthworks X200 Quick Reference Guide

MTS Restan. Automatic System for Residual Stress Measurement by Hole-Drilling

Impact Fatigue on Suction Valve Reed: New Experimental Approach

Sprinkler Protection for Cloud Ceilings. SUPDET 2013 Jason Floyd, Ph.D. 28 February 2013

Feltest AirSpeed/2 Instruction manual

Using the Models TA-460, PVM610, and PVM620 to Measure Pressure

WP2 Fire test for toxicity of fire effluents

Models EXN and OLG. General Description. Features and Benefits. Pressurizer for Vertical Air Conditioners Increased Safety for Hazardous Locations

Ozone. Sensoric O3 3E 1

Study by numerical simulations on the breathing effect in a semi-underground highway with beams and a roof above it

Flow and Performance of an Air-curtain Biological Safety Cabinet

for Pleasant & Healthy Air-Distribution ISO 9001:2000 Linear Slot Diffusers Product Bulletin 133-D

CO2 incubator with additional process controls

Digital pressure gauge Model CPG500

Instruction Manual Dräger MSI P7 and MSI P7 plus

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

ANALYSES OF THERMAL COMFORT AND INDOOR AIR QUALITY UNDER STRATUM, DISPLACEMENT, AND MIXING VENTILATION SYSTEMS

Transcription:

ASHRAE-110 Type-Test Report for 1.5m Wide Bench-Type FH-150 Fume Hood of TopAir, USA Ref: ANSI/ASHRAE 110-1995 by Dr A F Bicen Invent UK Ltd 85 Southdown Road Harpenden AL5 1PR Tel: +44 (0) 1582 461666 Fax:+44 (0) 1582 460077 VelociCal c Tested by: Dr A F Bicen Date of Tests: 28th June - 1st July 2011 INV/ASHRAE110/557-A Issue Date: 23rd April 2013

1. INTRODUCTION ANSI/ASHRAE 110-1995 type tests carried out for 1.5m wide bench-type FH- 150 fume hood of TopAir, USA are reported. The detailed information on test methods and procedures can be found in reference 1. 2. DESCRIPTION OF FUME HOOD The fume hood tested was 1.5m wide bench-type designed & built by TopAir, USA. The sash opening width was 1250mm and the test height 500mm from the bottom cill airfoil. Other geometric details of the hood are shown in Figure 1. 3. DESCRIPTION OF TEST ROOM FACILITIES The test room was at least 7.0m long, 6.0m wide and 4.0m high. The tests facilities include a variable-volume extract air system to adjust the extract volume flow rate to the required value. The make-up air was brought in through the perforated wall/ceiling tiles opposite the tested unit so as to allow a test room pressure in the range of typically -1Pa to -3Pa. The test room differential pressure, temperature, relative humidity and velocity during tests were: Room differential pressure: -2Pa Room air temperature: 25 C - 27 C Room air relative humidity: 50% Room air velocity: much less than 0.1 m/s 4. VELOCITY TESTS Velocity tests were performed with the sash set at 500mm from the bottom cill airfoil. The velocity type-test grid for the test opening is shown in Figure 2 which also shows the velocity test results. 5 FLOW VISUALISATION TESTS 5.1 Local Visualisation (Low Volume Smoke) Tests The following observations were made: Along the opening edge = GOOD Top LHS corner = GOOD Top RHS corner = GOOD Bottom LHS corner = GOOD Bottom RHS corner = GOOD 5.2 Large Volume Smoke Test The entry flow to the hood is good. The internal smoke clears within less than 30 sec. 6 CONTAINMENT TESTS 6.1 Static Sash Tests The containment tests were performed for the same opening as in the velocity tests. Figure 3 shows the positioning of the containment test system with respect to the test opening. Figure 3 also summarises the containment results and show that the fume hood containment performance is very good. 6.2 Probe Traversing Tests at Static Sash Openings Traversing of the sampling probe along the edges of the test opening resulted in SF6 levels less than 0.010ppm.

REFERENCES 1. ANSI/ASHRAE 110-1995, ASHRAE Guideline, Method of Testing Performance of Laboratory Fume Hoods, 1995. List of instrumentation used during tests: 1. Miran 203 infrared gas analyser - SN: 8278 (containment tests) 2. VelociCalc 9545-A - SN: 0713014 (velocity tests) 3. Drager Flow Check tracer smoke generator (smoke visualisation)

sash ~15mm ~20mm ~23mm ~7 25mmR ~15mm SASH AIRFOIL sash track side cheek 25mmR ~40mm ~7 ~25mm ~20mm worktop EXTERNAL DIMENSIONS: Height = 1500mm + 800mm Width = 1500mm Depth = 800mm INTERNAL DIMENSIONS: Height = 1050mm Width = 1250mm Depth = 670mm (wall-sash) 620mm (baffle-sash) Baffle top gap: 40mm Baffle bottom gap: 140mm - 240mm ~25mm BOTTOM CILL AIRFOIL Figure 1 Geometric features of 1.5m wide bench-type FH-150 fume hood of TopAir.

VelociCalc 01 03 05 07 09 02 04 06 08 10 Grid point Mean velocity (m/s) 01 0.53 03 0.51 05 07 09 Grid point Mean velocity (m/s) 02 0.55 04 0.49 06 0.49 08 0.49 10 Average Face Velocity = 0.51m/s Figure 2 Velocity type-test grid and results.

The mannequin is at 100mm away from the sash plane and positioned at locations shown below. The sampling probe is at the breathing zone of the mannequin and 75mm away from the sash plane. The test gas injector is positioned on the work top at locations shown below. At all locations, the injector is at 150mm in from the sash plane. The test gas is 100% SF6. The test gas flow rate is 4.0lt/min. 300mm C L 300mm P1 P2 P3 POSITION P1 Mean SF6 Concentration <0.005ppm* POSITION P2 Mean SF6 Concentration <0.005ppm* <0.005ppm* * the minimum detection level of the test system used. POSITION P3 Mean SF6 Concentration Figure 3 Positioning of ASHRAE containment test system and results.

ANSI/ASHRAE 110-1995 Flow Containment & Instrumentation CERTIFICATE OF TYPE TESTING IN ACCORDANCE WITH ASHRAE 110-1995 CERTIFICATE & REPORT NO: INV/ASHRAE110/557-A ISSUE DATE: 23rd April 2013 Fume Hood Type: FH-150 bench type External Dimensions: Height = 2300mm (1500mm + 800mm) Width = 1500mm Depth = 800mm Internal Dimensions: Height = 1050mm Width = Depth = 1250mm 670mm (wall-sash) 620mm (baffle-sash) Baffle Gap Dimensions: Top gap: 40mm Bottom gap: 140mm - 240mm Fume Hood Manufacturer: TopAir USA 3182 Monterey Drive Merrick NY 11566 USA 23/04/13 ED CERTIFI Fume Hood Flow: Test opening width: 1250mm Test opening height: 500mm Face velocity: 0.51m/s This is to certify that the fume hood described above has been typetested in accordance with ASHRAE 110-1995 and resulted in performance characteristics given in the corresponding test report. Tested and Certified by: Dr A F Bicen Invent UK Ltd, 85 Southdown Road, Harpenden, Herts AL5 1PR Tel: 01582 461666 : Fax: 01582 460077 : E-mail: info@invent-uk.com