March CS-1701F Reactive Ion Etcher

Similar documents
1.1 Equipment: substrate, wafer tweezers, metal targets 1.2 Personal Protective Equipment: nitrile gloves, safety glasses 1.

Revised: June 7, 2017

NORDSON MARCH PX-1000 PLASMA ASHER STANDARD OPERATING PROCEDURE Version: 1.0 July 2016

Warnings: Notes: Revised: January 8,

Plasma Asher: March PX-500 User guide (May-30, 2017)

Notes-PECVD: Chamber 1

Warnings: Notes: Revised: October 5, 2015

STS ICP-RIE. Scott Munro (2-4826,

Arizona State University NanoFab PLASMATHERM 790 RIE. Version A

Usage Policies Notebook for Xenon Difluoride (XeF 2 ) Isotropic Si Etch

University of MN, Minnesota Nano Center Standard Operating Procedure

5.1.3 Mechanical Hazards Drive assemblies have sufficient power to cause injury. Keep hands, fingers, clothing and tools clear of moving parts.

Photolithography. Operating Instructions

Plasma 1 (Technics) Standard Operating Procedure. Revision: 1.0 Last Updated: Feb.5/2013, Revised by Grace Li

Usage Policies Notebook for STS DRIE System

Plasma-Therm PECVD. Operating Characteristics. Operating Instructions. Typical Processes. I. Loading. II. Operating

University of MN, Minnesota Nano Center Standard Operating Procedure

Plasma II (AXIC) Standard Operating Procedure. Revision: 1.0 Last Updated: Feb.6/2013, Revised by Grace Li

PLASMA ENHANCED CHEMICAL VAPOR DEPOSITION (PECVD) SOP OXFORD PLASMALAB SYSTEM 100

Trion PECVD SOP IMPORTANT: NO PLASTIC, TAPE, RESISTS, OR THERMAL PASTE ARE ALLOWED IN THE CHAMBER

Standard Operating Manual

Edge Isolation Tool. Standard Operating Procedure. Version 1.1. Date: Prepared by, Sandeep S S. Department of Electrical Engineering

Arizona State University Center for Solid State Electronic Research. Table of Contents. Issue: C Title: Oxford Plasmalab 80plus (Floey) Page 1 of 8

March Asher Operation

Nordiko Metal Sputtering System Standard Operating Procedure

Unaxis PECVD. SiH4 (5% in He)

Nanofabrication Facility: ECR Etcher SOP Rev. 01b, March 06. Standard Operating Procedure for PlasmaQuest ECR II Etching

COMELEC C-30-S Parylene Coating System

Usage Policies Notebook for Trion RIE / ICP Dry Etch

R I T. Title: STS ASE Semiconductor & Microsystems Fabrication Laboratory Revision: Original Rev Date: 01/21/ SCOPE 2 REFERENCE DOCUMENTS

OXFORD PLASMALAB 80PLUS (CLOEY)

Unaxis ICP/RIE SOP Revision 8 09/30/16 Page 1 of 5. NRF Unaxis ICP/RIE Etch SOP

STS DRIE User SOP Location: NRF Cleanroon, Dry Etch Bay

Usage Policies Notebook for AMST Molecular Vapor Deposition System MVD 100

Plasma Cleaner. Yamato Scientific America. Contents. Innovating Science for Over 125 Years. Gas Plasma Dry Cleaner PDC200/210/510 PDC610G.

Xactix XeF2 OPERATION MANUAL

MANUAL FOR SPTS APS (DIELECTRICS ETCHER)

Standard Operating Manual

STS PECVD Instructions

Standard Operating Manual

Operating Procedures for the. SAMCO ICP RIE System

Usage Policies Notebook for Parylene Coating System

Nanofabrication Facility: PECVD SOP Rev. 00, April 24

Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System)

Equipment Standard Operating Procedure Greg Allion and Kimberly Appel

Title: Xactix XeF2 Etcher Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 03/23/2016

Xactix Xenon Difluoride Etcher

SPUTTER STATION STANDARD OPERATING PROCEDURE

KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE

Issue: H Title: CHA E-Beam Evaporator Page 1 of 7. Table of Contents

Approved by Principal Investigator Date: Approved by Super User: Date:

ABM MASK ALIGNERS. NanoFab 26 March 2009 A Micro Machining & Nanofabrication Facility

Karl Suss MA6 Mask Aligner SOP

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone:

Standard Operating Procedure. For. PVD E-Beam

STS Advanced Oxide Etch DRIE System Trends

MINIBRUTE ANNEAL TUBE STANDARD OPERATION PROCEDURE

Arizona State University Center for Solid State Electronics Research Issue: E Title: Heat Pulse 610 Operating Procedure Page 1 of 7

Standard Operating Manual

Approved by Principal Investigator Date: Approved by Super User: Date:

Standard Operating Manual

Standard Operating Manual

LESKER #1 STANDARD OPERATION PROCEDURE

Unifilm Technology PVD-300 Sputter Deposition Operation Instructions

University of Minnesota Nano Fabrication Center

University of Minnesota Nano Center Standard Operating Procedure

EE 432 Lab 3 PMOS source/drain lithography and diffusion

NNCI ETCH WORKSHOP SI DRIE IN PLASMATHERM DEEP SILICON ETCHER. Usha Raghuram Stanford Nanofabrication Facility Stanford, CA May 25, 2016

RAPID THERMAL PROCESSOR (Annealsys AS-ONE 150) Lab Manual

Usage Policies Notebook for Karl Suss MA6 Mid / Deep UV Mask Aligner

Chapter 2 General description of the system

NRF Suss Delta 80 Spinner SOP Revision /14/2016 Page 1 of 11. Suss Delta 80 Spinner SOP

THERMAL EVAPORATION UNIT (for Al evaporation)

Standard Operating Manual

Operating Procedures for Metal Evaporator I

University of MN, Minnesota Nano Center Standard Operating Procedure

Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300

Birck Nanotechnology Center

Usage Policies Notebook for Thermco Atmospheric Diffusion Furnace system

COBILT CA-800 Mask Aligner Equipment Operation

University of Minnesota, MN Nano Center Standard Operating Procedure

Oerlikon Sputtering Evaporator SOP

University of Minnesota Nano Fabrication Center

Approved by Principal Investigator Date: Approved by Super User: Date:

MJB-3 Mask Aligner Property of TAU MNCF August 2012

OPERATION OF THE DIMPLER

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA

CHEMICAL ENGINEERING SENIOR LABORATORY CHEG Initiated Chemical Vapor Deposition

Basic ICP Operating Procedures

Operation of the contact mask aligner Canon PPC 210 Projection Print Camera

REACTIVE ION ETCHING OF SILICON DIOXIDE USING BOTH OXYGEN AND CARBON DIOXIDE AS GAS ADDITIVES. dames E. Constantino ABSTRACT

Plasma Etching: Safety Operating Procedures

SOP for Karl Suss MJB3 #1 Mask Aligner

Inert Atmosphere Guide

KARL SUSS MJB3 UV400 Mask Aligner Standard Operating Procedure

The MRL Furnaces USED FOR THIS MANUAL COVERS

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

[USING THE VITROBOT April 9, Using the Vitrobot

Indian Institute of Technology Kanpur Samtel Centre for Display Technologies

Standard Operating Procedure

Transcription:

March CS-1701F Reactive Ion Etcher Standard Operating Procedure Faculty Supervisor: Prof. Robert White, Mechanical Engineering (x72210) Safety Office: Peter Nowak x73246 (Just dial this directly on any campus phone.) (617)627-3246 (From off-campus or from a cell phone) Tufts Emergency Medical Services are at x66911. Revised: January 6, 2016 Note: This CS-1701F was modified by March Plasma to include a turbopump for lower pressure operation. Standard CS-1701F units without a turbopump will operate at higher pressures than indicated in this document. Warnings: The tool uses high power and low vacuum during deposition. Do not attempt to open the chamber until the etch cycle is complete. The gases used in the tool, SF 6, CF 4 and O 2, are not toxic. O 2 is the most dangerous of the three because it can cause fire. Some of the byproducts of the etch process may be toxic (depending on what materials are being etched). Therefore, it is important that the tool pumps the chamber down below 80 mt before venting to atmosphere. Required Checkout: You must be formally tested by the faculty supervisor or lab manager before you may use this tool unsupervised. When you are first learning procedures, work with experienced lab users (students, post docs, faculty or staff) to become familiar with the tool before requesting a formal checkout procedure. 1.0 Material Requirements: 1.1 Equipment: substrate, wafer tweezers 1.2 Personal Protective Equipment: nitrile gloves, safety glasses 1.3 Chemicals: none 1.4 Gases: Oxygen, Nitrogen, Sulfurhexaflouride (SF 6 ), Tetrafluoromathane (CF 4, Freon 14) 2.0 Procedure: 2.1 Turn on the tool: 2.1.1 Turn on the RIE by pressing the large red Power button located on the front panel. Check to make sure that all front panel displays and switches light up. 2.1.2 Turn on the chiller located on the floor to the right of the tool. Set the chiller temperature to 20 C (to change the chiller setpoint, hold down the square black button and turn the knob). 2.1.3 Turn on the RF Power Generator using the toggle switch on the back side of the RF Generator. 2.1.4 Turn on the mechanical vacuum pump by flipping the toggle switch located under the chrome panel on the top of the pump. 2.1.5 Turn on the turbomechanical pump controller using the toggle switch located on the back. 2.1.6 Open all four gas cylinders. Make sure the outlet pressures of the regulators are as follows: Oxygen: 15 psi ± 5 psi Nitrogen: 60 psi ± 5 psi SF 6 : 15 psi ± 5 psi CF 4 : 15 psi ± 5 psi

2.2 Load the substrate: 2.2.1 After ensuring that the chamber pressure is at atmosphere, lift the camber lid using the chamber lid handle. Once in the open position the lid will settle into slots which keeps it open. 2.2.2 Clean any particles from the chamber surfaces and seals using IPA and a fab wipe. Blow the surfaces clean using clean dry air. 2.2.3 Place substrate on the center of the etching platen with wafer tweezers. 2.2.3.1 During pumpdown, air currents in the chamber can move the wafer. It is advisable to use 3 glass microscope slides positioned 120 degrees apart pushed up against the outer edges of the wafer to hold it in place. Put one slide at the back of the chamber (far from you) and one slide 120 degrees up on either side, so there is no slide at the front of the chamber (near the viewport it appears that during pumpdown the slug of air from the viewport sometimes flips the slide onto the wafer it is directly in front of the viewport). 2.2.4 Make sure both chamber seals are free of particles. This will ensure a proper chamber seal and low base pressure. 2.2.5 Lift the lid out of the slots by raising chamber lid handle approximately 0.5. (You must lift it up before closing it!!) Lower the lid into the closed position. Take care not to the drop the lid on your fingers! Make sure it seats well and is centered. 2.3 Program etch recipe: Create and save a recipe or sequence that can be run automatically 2.3.1 Toggle the program button to select the desired Program #. 2.3.2 Program the gas settings: 2.3.2.1 Toggle the gas SET/READ button for the R DISP menu group to SET position (LED illuminated) 2.3.2.2 Toggle the R DISP button to each of the gas 1, gas 2, and gas 3 channels in turn. 2.3.2.3 Adjust the gas mass flow set point for each gas using the :INCR/DECR button. 2.3.2.4 Toggle the gas SET/READ button for the R DISP menu group to READ position (LED illuminated). You can now monitor the current mass flow controller settings for each gas. NOTE: The gas flow setpoint reads in percentage of maximum flow (0-100). To compute the actual flow rate in sccm, use the following formulae: Gas 1 (SF 6 ): Flow rate in sccm = (setpoint/100) x 50 sccm x 0.27 Gas 2 (CF 4 ): Flow rate in sccm = (setpoint/100) x 250 sccm x 0.41 Gas 3 (O 2 ): Flow rate in sccm = (setpoint/100) x 100 sccm x 1

2.3.3 Program remaining process settings: 2.3.3.1 Toggle the SET/READ button for the L DISP program group to set position (LED illuminated) 2.3.3.2 Toggle the L DISP button to set each of the proper process parameters in sequence. 2.3.3.3 Adjust the selected parameters setpoint using the INCR/DECR buttons 2.3.3.4 Here is some useful information about these parameters PRESS: Enables Independent Pressure Control. Set to zero to disable. Units are mt. If you want a high pressure etch (>200 mt) this may be useful, but most of the time it will be set to 0 and the pressure will be controlled by the gas flow rate. POWER: Process power in Watts. (max 600) ENDPT: UV based endpoint detection feature. Refer to manual before using. Set to 100 to disable. TIME: Process time in seconds. Do not run etches longer than 10 minutes = 600 seconds, because the tool will become too hot. TEMP: Our tool is not equipped with this feature. Platen temperature is controlled by the chiller. BP/RP: Base pressure in mt. Set to 80 at highest, 30 at least. The tool will pump down to this pressure before running the etch and before venting. 2.3.4 Select the next program step, if any, you wish to program using the PROGRAM button. 2.3.5 **When you are finished programming your recipe, advance to the next program # slot and set it s time to 0.** Otherwise, the programs will be executed in sequence!!!! Note that this may be a useful feature if you are trying to do a Time Multiplexed etch. (for advanced users only) 2.3.6 Toggle the SET/READ button to read (LED extinguished.) This stores the parameters in the process controller memory. You may now monitor the current value of any of the process parameters on the L DISP menu by cycling through them. 2.4 Execute the recipe: 2.4.1 Make sure the tool is in Auto Mode. If the Man Op LED is OFF it is in Auto mode. If the Man Op LED is on, then turn off Man OP (The LED should extinguish) 2.4.2 Toggle the PROGRAM button to the program number for your program (see programming in section 2.3 above). 2.4.3 Push START. 2.4.3.1 The tool will now go through an automatic sequence of pumping down below the base pressure, then turning on the gas flow, waiting for pressure to stabilize, and turning on the RF power. 2.4.3.2 When RF power turns on, check that you have a stable, uniform plasma in the chamber. If the plasma does not stabilize within 10 seconds, push STOP to end the sequence. 2.4.3.3 If everything looks good, the etch will proceed for the programmed length of time. 2.4.4 When the recipe is complete, the tool will automatically pump out the etch products and then bleed back to atmosphere. 2.4.5 Once the bleed step is complete, open the chamber and remove your sample. 2.4.6 Push STOP.

2.5 Ashing of Photoresist: After exposure to plasma, some photoresists become hardened and require an oxygen clean prior to resist stripping. This is a lower intensity plasma used to remove hardened resist at the surface. 2.5.1 Low power O 2 clean recipe: Power: 200 watts, Oxygen flowrate (gas 3): 100%, All other gas flowrates 0%, Base pressure 80 mt, Time 30 seconds. 2.5.2 Wet strip remaining photoresist: Remove remaining resist with acetone, IPA and remover 1165 if necessary 2.6 Chamber Clean 2.6.1 After you are finished processing, you are required to run a high power oxygen clean on the empty chamber: 2.6.2 High power O 2 clean recipe: Power: 300 watts Oxygen flowrate (gas 3) 100%, All other gas flowrates 0%, Base pressure 80 mt, Time: 300 seconds. 2.7 Shutdown: 2.7.1 WAIT FOR TURBOPUMP TO REACH A SPEED OF LESS THAN 10,000 RPM BEFORE PROCEEDING WITH SHUTDOWN. (Turbopump speed can be seen on the front panel of the turbopump controller) 2.7.2 **Turn off all gas bottles** 2.7.3 Turn off power to the tool, the RF power supply and the turbo controller. 2.7.4 Turn off the mechanical vacuum pump. 2.7.5 Turn off the chiller. 2.7.6 **Check to make sure all four gas bottles are off** 3.0 Emergency Shutdown: 3.1 Processes can be interrupted by toggling the STOP button. This should turn off the RF power and pump the tool down. 3.2 If for some reason STOP is not working and something very bad is happening, pushing the large red power button will turn off all power to the tool. If you push this button to shut of power to the tool and there are etch gases in the chamber, immediately evacuate the lab. Although SF 6 and CF 4 are not themselves dangerous, it is possible that etch products (results of reactions in the chamber) could be hazardous, and if you shut of power the tool will not pump the chamber out. If at any time you feel a situation is dangerous, do not hesitate to call the safety office (x73246, Peter Nowak) the faculty supervisor/lab manager (x72210, Robert White), or Tufts Emergency Services (Police/Fire/Ambulance at x66911). Report all accidents (injuries, major spills, fires) to the safety office at x73246 (Peter Nowak) and Prof. White at x72210. For emergencies, call Tufts Emergency Services at x66911.

RECOMMENDED RECIPES: Anisotropic Silicon Etch Can use SPR, AZ or 1800 series resists as an etch mask. SF 6 (gas 1) flowrate: 100% (13.5 sccm) Power: 150 W Base pressure: 40 mt O 2 (gas 3) flowrate: 0% Pressure setpoint: 0 Actual pressure during etch: 65 mt <100> silicon etch rate: 240 nm/min SPR220 resist etch rate: 120 nm/min Silicon lateral etch rate (undercut):25 nm/min Silicon Dioxide Etch Can use SPR, AZ or 1800 series resists as an etch mask. Power: 300 W CF 4 (gas 2) flowrate: 20% (20.5 sccm) Base pressure: 40 mt O 2 (gas 3) flowrate: 0% Pressure setpoint: 0 Actual pressure during etch: 140 mt Silicon dioxide etch rate: 100 nm/min (Pyrex glass) Silicon Nitride Etch Can use SPR, AZ or 1800 series resists as an etch mask. Power: 200 W CF 4 (gas 2) flowrate: 100% (100 sccm) O 2 (gas 3) flowrate: 1% (1 sccm) Pressure setpoint: 0 Actual pressure during etch: 370-400 mt PECVD Si 3 N 4 etch rate: 180 nm/min <100> silicon etch rate: 60 nm /min SPR photoresist: 90 nm/min O2 Plasma Clean / Organic Etch Power: 200 W O 2 (gas 3) flowrate: 100% (100 sccm) Pressure setpoint: 0 Actual pressure during etch: 300 mt SPR220 resist etch rate: 0.5 1 um/min AZ9260 etch rate: 0.3-0.4 um/min Organic Polymer Fast Isotropic Etch Power: 300 W Base pressure: 40 mt O 2 (gas 3) flowrate: 100% (100 sccm) Pressure setpoint: 0 Actual pressure during etch: 300 mt SPR220 resist etch rate: 1-1.5 micron/min Parylene C etch rate: 0.5-2 micron/min (variable) Silk Fibroin thin film : > 1micron/min

Polyimide Etch Power: 300 W CF 4 (gas 2) flowrate: 12% (12.3 sccm) Base pressure: 40 mt O 2 (gas 3) flowrate: 50% (50 sccm) Pressure setpoint: 0 Actual pressure during etch: 200 mt PI-2610 polyimide etch rate: 0.7-1.2 micron/min Use metal hardmask (aluminum or chromium thin film) Fast Silicon Etch Recommend Nickel or SU8 as an etch mask. Do not try using AZ9260 as an etch mask it does not survive! SF 6 (gas 1) flowrate: 100% (13.5 sccm) Power: 400 W O 2 (gas 3) flowrate: 2% (2 sccm) Pressure setpoint: 0 Actual pressure during etch: 65-100 mt <100> silicon etch rate:1.5-3 um/min SU8 resist etch rate: 1-2 um/min SU8 Etch Recommend Nickel as an etch mask. SF 6 (gas 1) flowrate: 15% (2 sccm) Power: 400 W O 2 (gas 3) flowrate: 50% (50 sccm) Pressure setpoint: 0 Actual pressure during etch: 170-200 mt SU8 resist etch rate: 3 um/min <100> silicon etch rate:< 100 nm/min PDMS/Glass Bonding Note: consider using the ETP corona discharge wand for treating PDMS for PDMS/glass bonding it is easier and more effective. See the separate SOP on that process. Power: 40 W O 2 (gas 3) flowrate: 100% (100 sccm) Pressure setpoint: 0 Time: 10 seconds Actual pressure during etch: 300 mt Place surfaces to be bonded face up during plasma treatment. After plasma treatment, place glass and PDMS surface into contact. Press together gently and try to avoid surface bubbles. Cure on hotplate at 80 o C for 10 mins to 12 hours. (No systematic study has been conducted of the effect of the cure time different users have different protocols)