Manual for continuous distillation

Similar documents
1. Study the performance of a binary distillation column operated in batch mode.

Operating Instructions

Armfield Distillation Column Operation Guidelines

Distillation Design The McCabe-Thiele Method

CHE 4115 Chemical Processes Laboratory 2 Experiment 1. Batch Distillation

The Discussion of this exercise covers the following points:

CPE562 Chemical Process Control HOMEWORK #1 PROCESS & INSTRUMENTATION DIAGRAM

Operating Instructions

CHE 306 Stagewise Operations

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler.

Felles Lab: Distillation Columns

DISTILLATION POINTS TO REMEMBER

Cover Page for Lab Report Group Portion. Pump Performance

CHEMICAL ENGINEERING LABORATORY CHEG 239W. Control of a Steam-Heated Mixing Tank with a Pneumatic Process Controller

2. Determine how the mass transfer rate is affected by gas flow rate and liquid flow rate.

Lab 1. Adiabatic and reversible compression of a gas

Dynamics and Control of Chemical Processes Solution to Lab #8 Shutdown of C4-C5 separation section

LABORATORY EXERCISE 1 CONTROL VALVE CHARACTERISTICS

Cover Page for Lab Report Group Portion. Head Losses in Pipes

This portion of the piping tutorial covers control valve sizing, control valves, and the use of nodes.

Exercise 2-2. Second-Order Interacting Processes EXERCISE OBJECTIVE DISCUSSION OUTLINE. The actual setup DISCUSSION

GCMSD-Headspace Analysis SOP

CHM Gas Chromatography: Our Instrument Charles Taylor (r10)

Cover Page for Lab Report Group Portion. Compressible Flow in a Converging-Diverging Nozzle

FROG-4000 TM Chemical Analysis System

1.2 Example 1: A simple hydraulic system

Heat Engine. Reading: Appropriate sections for first, second law of thermodynamics, and PV diagrams.

Cover Page for Lab Report Group Portion. Compressible Flow in a Converging-Diverging Nozzle

Quantitative Analysis of Hydrocarbons by Gas Chromatography

5890II GC Standard Operating Procedure 9/2/2005

Gas Absorption Draft Standard Operating Procedure

Boyle s Law: Pressure-Volume Relationship in Gases

Start up XXX from normal shutdown Introduction

AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by:

16. Studio ScaleChem Calculations

Vapor Pressure of Liquids

FROG-4000 TM Chemical Analysis System

DSTWU A Shortcut Distillation Model in Aspen Plus V8.0

TEMPERATURE S RELATIONSHIP TO GAS & VAPOR PRESSURE

DISTILLATION PRESSURE CONTROL TROUBLESHOOTING THE HIDDEN PITTFALLS OF OVERDESIGN

product manual HM-4140, HM-4150, HM-4160 HM-4160A HM-4150 Humboldt FlexPanels

The Gas Laws: Boyle's Law and Charles Law

The Discussion of this exercise covers the following points: Range with an elevated or suppressed zero Suppressed-zero range Elevated-zero range

Appendix: Simplified Instructions for Operating the HP 5890 GC/FID and Air Sampling Systems

Countercurrent Stagewise Operations Equilibrium Stage Number of Actual Stages Definition of The Design Problem...

Liquid -Vapor. Contacting columns

UBEC 1AT. AUTO TANK Fill System Installation, Operation, & Setup Instructions

Boyle s Law VC 09. Experiment 9: Gas Laws. Abstract

Introduction. Part one: Identify the Hydraulic Trainer Components

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

FROG-5000TM Chemical Analysis System User s Manual

LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Gas Diluter Portable Gas Dilution System User s Manual

SOP: Buck Scientific BLC-20P HPLC Operation

PerkinElmer Aanalyst100

OPERATION MANUAL RTI RHS650 RTI TECHNOLOGIES, INC East Market Street York, PA Manual P/N

GAS DEHYDRATION SYSTEM

ERTC PETROCHEMICAL Conference 20 st -22 nd February 2002, Amsterdam, The Netherlands

R E D I C O N T R O L S

LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary

The Discussion of this exercise covers the following points:

Gas Laws: Boyle s and Amonton s Laws Minneapolis Community and Technical College v.9.08

AKTA 3D SOP. Click on the System Control tab. This screen has 4 windows.

The water supply for a hydroelectric plant is a reservoir with a large surface area. An outlet pipe takes the water to a turbine.

GAS DEHYDRATION SYSTEM

Separation of Acetone-Water with Aspen HYSYS V8.0

AKTA MC SOP Page 1 9/27/04 AKTA METAL CHELATING SOP

Fun with Gas Laws. Prepared by Vance O. Kennedy and Ross S. Nord, Eastern Michigan University PURPOSE

Exercise 2-3. Flow Rate and Velocity EXERCISE OBJECTIVE C C C

Exercise 4-2. Centrifugal Pumps EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION. Pumps

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Part 1: Distillation and Analysis of an Unknown Alcohol Mixture

Experiment GC : Analysis of BTEX by GC-FID

INSTRUCTOR GUIDE REFERENCES: PUMPING APPARATUS DRIVER/OPERATOR HANDBOOK, FIRST EDITION, IFSTA

Model 130M Pneumatic Controller

Lab # 03: Visualization of Shock Waves by using Schlieren Technique

Removing nitrogen. Nitrogen rejection applications can be divided into two categories

Injection Systems INSTALLATION AND OPERATING GUIDE HI FLO VERTICAL TANK SYSTEMS

By: Eng. Ahmed Deyab Fares - Mobile:

Mass Spec will not Autotune

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

Another convenient term is gauge pressure, which is a pressure measured above barometric pressure.

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

Introductory Lab: Vacuum Methods

Automated Liquid Handling Station

Analysis and Modeling of Vapor Recompressive Distillation Using ASPEN-HYSYS

Vapor Pressure of Liquids

Gerald D. Anderson. Education Technical Specialist

Exercise 8. Closed-Loop Pressure Control, Proportional-Plus-Integral Mode EXERCISE OBJECTIVE

Workshop 302-compressor-anti-surge

Using the Akta Prime plus October 22, 2012

Trion PECVD SOP IMPORTANT: NO PLASTIC, TAPE, RESISTS, OR THERMAL PASTE ARE ALLOWED IN THE CHAMBER

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI

UNLOADING PROCEDURE OF AN AIRFLOW CONTAINER

Diffusivity in Gases Experiment ver 05.10a

Vapor Pressure of Liquids

Calibration of a GC955

AKTA pure 25 New Owner s Intro

Transcription:

Manual for continuous distillation 1. Week 1: Objectives: Run the column at total reflux. When steady state is reached, take the sample from the top and bottom of the column in order to determine the overall efficiency. Based on the overall efficiency, draw the quasi equilibrium curve. Compare the tray numbers you get from the quasi equilibrium line with the actual tray numbers. Take the samples from consecutive trays to determine the local efficiency.

Operation procedure: Preparation for Start-up: Key purposes: Get familiar with the system; locations of the valves and sample taps Check the liquid level in the reboiler Check the composition in the reboiler and feed tank Turn the Siemens system on Check if the valves are in the correct position 1. Using the sight-glass on the reboiler to assess its level of the liquid (in the range of around 70%). 2. Prepare to mix the contents of the reboiler. a. Close the red gauge valve on the bottoms line.(see graph below) b. Open the green valve on the same line just after it exits the reboiler. Follow the line down to the bottom of the reboiler. Open the orange ball

valve and make sure that the blue ball valve on the sample port is closed 3. Ask the TA/Lab Assistant to turn on the computer system hooked to the East column. 4. Using the computer control system, turn on the reboiler-mixing pump on the 1st floor, as shown below to mix the contents of the reboiler. Let mix for at least twenty minutes. 5. While you are waiting, check that the manual by-pass valves are closed. a. Reflux red valve parallel to the Reflux control valve b. Distillate red valve parallel to the Distillate control c. Bottoms blue valve parallel to the Bottoms control d. Water blue valve parallel to the Water control valve

Red gauge valve on Bottoms line Blue ball valve on line exiting reboiler Orange valve at bottom of reboiler open while mixing Figure 1. East Column Reboiler Pumping Line Figure 2. East column second floor control valves

Start up: 1. Ensure that the reboiler has enough material and then turn on the air for the 1st floor by manually opening the air valve fully. The exact position of the valve is A of the graph shown below. Air is needed to operate pneumatic control valves. 2. Turn the water on for the 1st floor by manually opening the water valve fully. The exact position of the valve is C in the picture below. Water is needed for cooling purposes. Water valves by convention are blue in color. 3. Kindly note the sequence of turning on the utilities, it is air first (no cost) followed by water (somewhat expensive) followed by steam (highly expensive). The sequence will be exactly reversed while turning them off. The rationale is to cut-off the steam and save the operating cost as much as possible. 4. Ask the Lab Assistant/TA to purge the steam lines using the steam purge valve located on the 1st floor. While the lines are being purged, the students should have fingers in their ears and should not be directly facing the outlet of the purge line. The Lab Assistant/TA should have ear plugs. The ear plugs should necessarily be discarded after one use. 5. Repeat steps 1 and 2 on 2nd floor. i.e. turn on the air for the 2nd floor by manually opening the air valve fully. The exact position of the

valve is B. Also, turn on the water for the 2nd floor by manually opening the water valve fully. This valve is a large blue valve located at the base of the water rotameter, which is D shown in the picture. 6. Then go to the Control Room. One of the computer systems is hooked to the West Column. The system would have already been made on by the Lab Assistant/TA. The computer screen will indicate values of different process parameters like temperature, pressure and flow rates as well as the positioning of different valves. Note that the values of some of the parameters might not be correctly shown on the screen due to some or the other reason and typically such values will be grayed out. Set the Steam/PID on the screen. Usually 40% is a good start for batch column. You will slowly see a rise in the temperature of the liquid in the reboiler and the pressure. You should see the boiling of the liquid in some time. Allow a stable level to be attained in the sight glass. Set the Reflux/PID to a 100%. Now, it essentially means that you are operating the column at total reflux. In case, you see the column flooding meaning the sight glass getting overfilled, turning on the Reflux Pump momentarily helps it to get stable. When the Reflux Pump turns on, it will be indicated in green. Reducing the % in the Steam/PID also helps to lower down the level of the material in the sight glass. 7. After the level of the material in the stand pipe is at a fairly stable for

some time, check if the system has attained steady state or not. This can be done by measuring the composition of one of the streams (reflux is the easiest because it is located on the second floor) as a function of time. When the composition changes by less than 5% with time, the system can be considered in a steady state condition. The reflux sample port is located on the second floor. A B

C D A- Air supply on the first floor B- Air supply on the second floor C- Cooling water on the first floor D- Cooling water on the second floor Shutting Down 1. As mentioned before, the utilities will be turned off exactly in the reverse order meaning steam first, then water and lastly air. Use the computer system to close the Steam control valve. To close the steam valve, click on the display box labeled STEAM/PID located above the valve symbol on the reflux line. Click on the box to the right of MAN. A small window will pop up. Change the opening of the valve to 0 and

click OK. 2. Turn off the reboiler mixing pump (EBTP/PUMP): a. To turn off the reboiler-mixing pump: click on Bottoms/PUMP or the pump icon; if not yet in manual mode, click on Manual ; click Stop and close the window. 3. With the reflux control valve (Reflux/PID) fully open, use the reflux pump (Reflux/PUMP) to drain the standpipe. Turn pump off. a. To open the reflux valve: Click on the display box labeled Reflux/PID located above the valve symbol on the reflux line. Click on the box to the right of MAN. A small window will pop up. Change the opening of the valve to 100 and click OK. 4. Open the red reflux valve (also serves as a manual bypass valve) (see Figure 1). 5. Close manual valves associated with the distillate tank (Figure 3). 6. Turn off the water and air valves by manually turning them fully in the clockwise direction on both floors.

Figure 3. Tanks on the first floor. Calculation: Overall efficiency can be calculated by comparing the actual number of plates in the column with the number obtained from McCabe-Thiele diagram when the column operates at total reflux. Local or point efficiencies can be calculated by measuring the compositions of the liquid streams corresponding to two successive plates. The local efficiency is typically close to the overall efficiency. E overall = N theoretical N actual 100%

The theoretical tray numbers can be obtained as follows: Generate the equilibrium curve with relative volatility; Draw horizontal and vertical lines between the equilibrium line and y=x from the top product composition to bottom product composition; Tray numbers can be counted in the graph. An example drawing of the procedure is shown in Fig 1. From this graph, N theoretical = 7 1 = 6 Assuming the actual tray number is 12, E overall = N theoretical N actual 100% = 6 12 1 = 54.5% 1.0 0.8 equilibrium line 2 1 0.6 3 y A 0.4 4 0.2 7 6 0.0 0.0 0.2 0.4 0.6 0.8 1.0 x B x A Fig 1. An illustration of the way to get theoretical tray numbers x D

Given the analytical equation of equilibrium line, y = f(x), and the overall efficiency of the column, E overall, we can get the analytical expression of quasi equilibrium line as blow: y = x + E overall (f(x) x) Compare the tray numbers getting between the quasi equilibrium line and the y=x to the actual numbers to verify the validation of the McCabe-Thiele diagram. The way to get quasi equilibrium line is shown in Fig 2. In Fig 2, b represents the driving force at a certain tray if at ideal case, a is the driving force in reality in the column we run. 1.0 0.8 0.6 equilibrium line b a y A 0.4 0.2 a/b=e overall quasi equilibrium line 0.0 0.0 0.2 0.4 0.6 0.8 1.0 x A Fig 2. An illustration of the way to get quasi equilibrium line

1.0 0.8 equilibrium line 0.6 y A 0.4 0.2 quasi equilibrium line 0.0 0.0 0.2 0.4 0.6 0.8 1.0 x B x A Fig 3. An illustration of the way to count the number of trays x D In Fig 3, the tray numbers getting from the quasi equilibrium line is 10 and the actual tray numbers is 11. Error = 11 10 11 100% = 9.1% By taking the sample from successive trays, the local efficiency can be worked out as follows: Fig 4 is a graphical description. E local = y n y n+1 y n y n+1 = x n 1 x n y n x n

1 equilibrium line y * n y A y n y n+1 y=x 0 0 x n x n-1 1 x A Fig 4. An illustration of the way to get local efficiency 2. Week 2: Run the column at different reflux ratios. Study the effect of reflux ratio on the concentrations of top and bottom product. Operation procedure:

From week 2 onwards, it is suggested to check if the reboiler has enough material from the control panel before starting the experiment. If the liquid level in the reboiler is less than 15 in, the liquid in the distillate tank should be pumped back to the. To pump the distillate back, you need to follow the procedure below. First open two valves located below the distillate tank, which are shown in graph E. Then open the bypass valve which is indicated on graph C. Finally open the pump below the reboiler on graph F. When you pump the liquid back to the reboiler, make sure someone checks the level in the distillate tank, as you may run the risk of burning the pump if the distillate tank is pump to empty. The general start up procedure followed is the same as the first week during all the three weeks, so repeat procedure 1~7 at first for week 2 and week 3. Open the valve beside the rotameter on B to change the distillate flow rate in order to change the reflux ratio. The distillate will go through the rotameter on the 2 nd floor down to the tank located on the 1 st floor, which is shown in graph E. Notes: All composition measurements are done with a Perkin Elmer Gas Chromatography instrument in the Lab. Get help from the lab assistant on using it.

Shutting Down Turn off the valve for distillate first. Then shut down the steam on the control panel, which means set the PID of steam to 0%. Wait for 1~2min, as after shutting off the stream, the vapor may still go to the top and that means the water is still needed to condense the top product. Then close the air and water valves on the first and second floor, just like the shutting down procedure for the first week. Reflux rotameter Reflux sampling port A Distillate rotameter B

Sight glass Bypass valve Bottom product sampling port C D

2 1 E F A- Reflux rotameter B- Distillate rotameter C- East column 1 st floor reboiler D- East column second floor control valves.

E- Distillate tank for east column F- Control panel overview of the east column 3. Week 3: Run the column at different reflux ratios. If there is still time left, run the column with a reflux ratio smaller than the minimum reflux ratio. GC (gas chromatography)

For our GC system, the absolute value of the peak area will vary a lot even if the same sample is injected for several times, which can be observed with the error bar in the graph. For this graph, four standard samples are tested and each concentration is calibrated for three times. The absolute peak area and peak area percentage is taken out to draw the curve. In the graph, the black points are absolute value of the peak area of ethanol, while the blue points are the peak area of ethanol over the total area, which is a relative number. The error bar for the absolute area is huge, especially for the mole fraction around 0.6, but for the relative area, the error bar can be barely seen, indicating a more reliable measurement for our GC system.

Peak area (V*s) Peak area percentage(%) 2.0 1.5 1.0 Equation y = a + b*x Weight Instrumental Residual Sum of 18.80314 Squares Adj. R-Square 0.98832 Mean Value absolute area relative area Standard Error Intercept -316738.0310 42671.23728 Slope 2.11196E6 43147.6728 150 100 0.5 Equation y = a + b*x Weight Instrumental Residual Sum of 6921.93153 Squares Adj. R-Square 0.98842 Slope 116.4014 0.12339 0.0 0 0.00 0.25 0.50 0.75 1.00 1.25 Mean Mole fraction Value Standard Error Intercept -21.25416 0.09766 50 For this method, there is not much calculation going on. The students need to calibrate the curve with relative peak area as the graph above. As the peak area percentage of a same sample will not vary much for different runs, the students just need to run each concentration once. After getting the calibration curve, they can use this curve to get their mole fraction directly from the relative peak area for the three-week experiment.