PERFORMANCE QUALIFICATION AND OCCUPATIONAL EXPOSURE LIMIT (OEL) FOR ISOLATOR

Similar documents
CTG Certification Protocol -- Version 3.0 SUMMARY

HPI-4N1 NEGATIVE PRESSURE COMPOUNDING ASEPTIC ISOLATOR PERFORMANCE EVALUATION

R2S Standard Sampling Assembly OPERATION PROCEDURE

EMT EnvironmentalMonitoringTECHNOLOGIES

NuAire Model NU PR797 Positive Pressure Recirculating Compounding Aseptic Isolator Performance Evaluation Compliance to USP797

Draft. Equipment/Instrument

NuAire Model NU NTE797 Negative Pressure Total Exhaust Compounding Aseptic Containment Isolator Performance Evaluation Compliance to USP797

NuAire Model NU NR797 Negative Pressure Recirculating Compounding Aseptic Containment Isolator Performance Evaluation Compliance to USP797

Regulatory Basics for Facility Design (WHO GMP): Current GMP Requirements

CLASS D - SENSITIVE LEAK TEST GAS AND BUBBLE METHOD. 1.1 To provide definitive requirements for PNEUMATIC pressure testing of piping systems.

PERFORM Operating Document

CentriFlow Meter Integrated Air Entrainment System

H 2 O 2 Bio-Decontamination and Aseptic Transfer of Heat Sensitive Materials Sergio Mauri, Director, Global Marketing and Business Intelligence

Kansas State University Fume Hood Operation

Whitepaper Technical Opinion: Grade A - Air Supply and Local Protection

Approval Block. Prepared by: Signature Date. Evan Parnell 08 NOV Reviewed by: Signature Date. Approved by: Signature Date

ProSys Laboratory Analytical Isolators

RE: PROPANE SYSTEM & REGULATOR TESTING/INSPECTION PROCEDURE

STANDARD OPERATING PROCEDURES. Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2.

OEL LIMITS: Products are classified according to hazards levels that correspond to these five OEL levels as shown in the chart below :

DESIGN AND INSTALLATION CONSIDERATIONS FOR TSI AEROTRAK VHP-RESISTANT REMOTE PARTICLE COUNTERS

PERFORM Operating Document

CETA Compounding Isolator Testing Guide CAG Revised December 8, 2008

Microbial Impaction Sampler for Clean Room Gas Monitoring TEST AIR

"LABORATORY SAFETY SERIES"

OPERATOR, PRODUCT AND ENVIRONMENTAL PROTECTION

OPERATOR, PRODUCT AND ENVIRONMENTAL PROTECTION

RADIATION PROCEDURES MANUAL Procedure Cover Sheet

Air Handling Unit (AHU) Air handling unit which serves to condition the air and provide the required air movement within a facility.

Fumehood Performance Test Report

Annex 8. Guidelines on heating, ventilation and air-conditioning systems for non-sterile pharmaceutical products. Background. Summary of main changes

FUNDAMENTAL SAFETY OVERVIEW VOLUME 2: DESIGN AND SAFETY CHAPTER I: AUXILIARY SYSTEMS. A high-capacity EBA system [CSVS] [main purge]

SURROGATE POWDER AND ASHRAE-110 TESTING OF A MULTI-PURPOSE ENCLOSURE FOR WEIGHING/SIEVING/FLUID BED DRYER APPLICATIONS

Statesville, NC. Results of AM ASHRAE 110 Laboratory Fume Hood Performance Evaluations Venturi Hood

TECHNICAL DATA. Page 1 of 12

Exhibit 4. Determination of Static Pressure Performance of the Healy Clean Air Separator (Executive Orders VR-201-N and VR-202-N)

TExT Thorough Examination & Test of LEV Local Exhaust Ventilation Systems

STANDARDS FOR COMPRESSED GAS TESTING

TECHNICAL DATA. Pressure Regulation 531a. April 24, 2009

VALITEQ DMS v2.o Document Management Software Tutorial

The primary purpose of laboratory fume hoods is to keep toxic or irritating vapors out of the general laboratory working area.

Group Quality Certification System of Performance Test for Clean Rooms (CR Mark)

Cover Page for Lab Report Group Portion. Pump Performance

AIR-OPERATED DOUBLE DIAPHRAGM PUMP USER S MANUAL

LABORATORY SAFETY SERIES: Laboratory Hoods

Installation Instructions

Toolkit for safe Use of Microbiological Safety Cabinets(MSC s)

NEW: CO2 incubator with sterilizable sensor

Perform Pressure & Leak Test, Tubing & Piping. Module 12306

Oxygen Usage Best Practice Guide

Met One E-BAM Particulate Monitor

TWC Services, Inc. Structured On-The-Job Training for JT- 1, 2, 3

Building Official's Guide to Duct Testing based on 2009 IECC

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

Preparation and Installation of the Sanitary BDI-FLX Sensor and Connection to the BDI-FLX Interface Cable

RS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.

Equipment Operating Procedure Glove Box

Mott Manufacturing Limited

Instruction Manual. Pipe Friction Training Panel

IC Sentinel Applications Guide for Healthcare Environmental Air Quality Monitoring

GM Series Dual-Block Multi-Function Gas Control Valves

ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS

TS1. Ultrasonic Tank Sender. Installation and Operating Instructions. For TS1 Firmware v3.8. Page 1 INST-TS1-V13 18/11/10

Overview ASTATT3. Carry out air tightness testing for single dwellings and other small buildings

Facility design for containment. Gordon Farquharson, July 2016

This test shall be carried out on all vehicles equipped with open type traction batteries.

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the

ATTACHMENT O WIPP MINE VENTILATION RATE MONITORING PLAN

Standard Operating Procedure #COE-SOP-0001 Chemical Fume Hood Operation

Cover Page for Lab Report Group Portion. Head Losses in Pipes

1.0 PURPOSE 2.0 SCOPE 3.0 DEFINITIONS. ANSI: American National Standards Institute. CCC: Chemical Control Centre

CORESTA RECOMMENDED METHOD N 6

LAKOS Waterworks. PWC Series Sand Separators. Installation & Operation Manual LS-829 (10/12)

Attachment C. Standard Operating Procedures. AMBIENT AIR MONITORING PROGRAM for the 130 LIBERTY STREET DECONSTRUCTION PROJECT

Constant Pressure Inlet (CCN) Operator Manual

SAFETY DATA SHEET. In accordance with 29 CFR Section 1. Identification of the substance or the mixture and of the supplier

General Accreditation Guidance. User checks and maintenance of laboratory balances

LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12

SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL

SAFETY DATA SHEET. In accordance with 29 CFR Section 1. Identification of the substance or the mixture and of the supplier

VINAYAK PHARMA TECHNOLOGY

Design of a Solid Wall Transonic Wind Tunnel

Inspection and Approval of Asphalt Mix Design Laboratories

Testing the Vented, 7-Liter Induction Chamber. Report 6 April 2008

USP <797> LABELING ESTABLISHING BEYOND-USE DATES AND IN-USE TIMES

CSU Channel Islands Lead Operations and Maintenance Program

CHECK-VALVES Automatically Position Themselves by the Natural Forces of the Conveying Air. A PNEUMATIC CONVEYING SYSTEM PATENT #: 5,160,222

Spray Gun 134-A. 1 Liter High- Pressure Gel Coat. Usage Instructions & Parts List

University of North Florida Performance Requirements For Laboratory Fume Hoods. As Manufactured Performance Tests. As Installed Performance Tests

OPERATION OF LIQUID PARTICLE COUNTER. GP-B P0045 Rev -

SERVICE MANUAL MODEL BA020AMST BREATHING AIR PANEL

Space Pressurization: Concept and Practice ASHRAE Distinguished Lecture Series

LRS(H)4 USER MANUAL. Read the complete manual before installing and using the regulator.

P5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

Operating Instructions for Cobra and Lynx Duct Leakage Testers

Proficiency Module Syllabus. P601 - Thorough Examination and Testing of Local Exhaust Ventilation Systems

PRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator.

SAFETY DATA SHEET. In accordance with 29 CFR Section 1. Identification of the substance or the mixture and of the supplier

Standard Operating Procedures (v9 5/1/17)

GREIG FILTERS, INC. OPERATION MANUAL

Transcription:

wjpmr, 2017,3(7), 157-162 SJIF Impact Factor: 4.103 WORLD JOURNAL OF PHARMACEUTICAL Ratnesh and Vivek et al. AND MEDICAL RESEARCH www.wjpmr.com Research Article ISSN 2455-3301 WJPMR PERFORMANCE QUALIFICATION AND OCCUPATIONAL EXPOSURE LIMIT (OEL) FOR ISOLATOR Natrajan S. Iyer, Dr. Aditya P. Sharma, Arun Kumar, Ratnesh Verma*, Vivek Kumar Sharma* and Rekha Mehta VHB Medisciences Ltd. Rdurapur, Pant Nagar, Uttarakhand. *Corresponding Author: Ratnesh Verma and Vivek Kumar Sharma VHB Medisciences Ltd. Rdurapur, Pant Nagar, Uttarakhand. Article Received on 07/06/2017 Article Revised on 28/06/2017 Article Accepted on 19/07/2017 ABSTRACT Isolators are widely used in Pharmaceutical Industries for aseptic compounding applications of sensitive and hazardous substances. Isolator is designed to provide continuous and complete isolation of the inside environment of the isolator from the external room environment (including its operators).only the gloves provided in isolators or robotic arms are used to compounding or dispensing the product. This ensures that environment is maintained as grade A and contamination-free. Present article can be used to check the isolators when gaskets are changed, glove ports are repaired or new product is taken and to determine the Occupational exposure limit (OEL). KEY WORDS: OEL: Occupational Exposure Limit. INTRODUCTION Determination of OEL (Occupational Exposure Limit) is to identify the amount of drug substance which comes out from the isolator and exposed to the external environment (including its operator) during compounding. The series of experiments has been performed to verify the isolator for its performance, which are: Light intensity, Air Velocity, Differential pressure across the chamber, Filter integrity test, Temperature & RH monitoring, Viable particle count, Non viable particle count, Recovery test, Air flow pattern, Chamber leak test, Leak test of gloves and Occupational exposure limit determination. Isolator Description The Isolator used in pharmaceutical industries comprise an aseptic processing chamber, the Isolator maintained as Grade A micro-environment in which the various open processing stages of individual aseptic batches can be carried out. By design isolators are of two group, one group has the main objective of providing containment for the handling of hazardous materials either aseptically or not. Another group has the main objective of providing a microbiologically controlled environment within which aseptic operations can be carried out. Containment isolators operates on negative pressure inside the isolation chamber and most isolators used for aseptic processing works on positive pressure. The Isolator designed to stand alone, unit comprising process chamber, transfer hatch, air handling units (complete with fans and HEPA filters) and Suitable sterilizing methods. After switching ON the Isolator, switch ON the lights and blower of isolator and allow to stabilize the differential pressure according to design of isolator. Material to be dispense or manufacturing is transferred to pass box with other dispensing tools. After attaining the required pressure, material is transferred to the weighing chamber. Now the double poly bag covered material is opened and dispensing or compounding of material is done by using glove port. After dispensing or compounding the remaining material and empty poly bags are then taken out from the out pass box in intact condition. Experimental Plan 1. Light Intensity 2. Air Velocity 3. Differential pressure across the chamber 4. Filter integrity test 5. Temperature & RH monitoring 6. Viable particle count 7. Non viable particle count 8. Recovery test of Chamber 9. Air flow pattern 10. Chamber leak Test 11. Leak test of gloves 12. Occupational Exposure Limit (OEL) determination www.wjpmr.com 157

EXPERIMENTAL PROCEDURE 1. Light Intensity Test 1.1 Light intensity of Isolator chamber shall be verified with calibrated LUX meter (certified by NABL), to be sufficient that provides proper lighting for dispensing and to avoid any discrepancy. 1.2 Light intensity shall be measured at the base of isolator s working place at three location i.e A- Left corner, B- Center & C- Right corner of chamber via Lux meter. Top view of isolator depicting different Location for light intensity verification. Performance of Light Intensity Test Pictorial Presentation of Light Intensity Measurement. 2. Air Velocity Air velocity shall be measured by Calibrated hot wire Anemometer (certified by NABL) in the Chamber at different 5 locations for both Supply duct and Return duct of the Isolator chamber. 2.1 Air Change Rate 2.1.1 Start the Isolator and on Normal running conditions, monitor the airflow into or exhausting from the chamber by using a Hot Wire Anemometer. 2.1.2 From this calculate the volume of air through the chamber and then calculate the Air Change Rate. 2.1.3 Measure the air velocity two inches below the Inlet, at 5 locations (Four corners & One Center) with the Anemometer and record. 2.1.4 Calculate the average velocity of the air coming from supply Inlet. 2.1.5 Calculate the airflow by multiplying the average velocity with the effective filter area. Air flow = Average Velocity (Ft/Min) x Area of the Air Inlet Filter (Ft 2 ) = Ft / Min. x Ft 2 = Ft 3 / Min. or CFM 2.1.6 Calculate the airflow to the entire supply Inlet in the Chamber and add up the values to get the total airflow in the Chamber (CFM) 2.1.7 Calculate the number of air changes per hour by using the formula Air Changes / hour = Total air flow in the Chamber (CFM) x 60 Chamber Volume (Ft 3 ) 3. Differential Pressure Across The Chamber The purpose of differential air pressure test across the chamber is to verify the capability of the isolator system to maintain the specified pressure differential between the room and of the inside of chamber of isolator. 3.1 When all doors closed, pressure difference shall be measured and recorded. 3.2 The differential pressure of isolator shall be monitored visually on a mgnehelic gauge at least three times in a shift. 4. Filter Integrity 4.1 Start the isolator and ensure the fan running at normal conditions/ flow rates. 4.2 Position the smoke generator and introduce PAO smoke into the air stream, ahead of the HEPA filters, at the concentration of 80 100 micro gram per liter of air at the filter s designed airflow rating and set the instrument at 100% concentration. 4.3 Scan the downstream side of the filter with an photometer probe. 4.4 The probe should scan the entire filter face and frame at a position about 1 to 2 inches from the face of the filter. Pictorial presentation of Velocity measurement. www.wjpmr.com 158

Photograph of filter of Isolator. 5. Temperature & RH monitoring The purpose of this test is to demonstrate the capability of isolator systems to maintain air temperature and relative humidity levels within the control limits over the time periods. 5.1 Keep the calibrated digital thermo hygrometer in the isolator chamber (at center). 5.2 Measure at least 03 readings for a shift after the stabilization of the temperature and relative humidity. 6. Viable Particulate Count Viable particulate count shall be taken by two-method i.e. Passive Air Sampling (Settle plate method) and Active air Sampling (Volumetric air sampling) for isolator chamber. 6.1 Passive Air Sampling (By Settle Plate) 6.1.1 Soyabean casein digest agar plates shall be exposed for fungal and bacterial monitoring respectively. 6.1.2 All the plates which shall be incubate, must be pre incubated at 30 C-35 C for 24 hours and one unexposed plate must be kept as a negative control. Keep one agar plate in uncontrolled area. This plate serves as Positive control to check the growth effectiveness of the media. 6.1.3 The lid of the Petri plates shall be placed under the plate so as to incline them towards the flow of air. The Petri plates shall be exposed for 4 hours. 6.1.4 After completion of exposure, these exposed plates shall be transferred to the micro lab. 6.1.5 Incubate the exposed plates at 20.0 C to 25.0 C for 72 hours and count the number of colonies without opening the plates. Further incubate the plates at 30.0 C to 35.0 C for 48 hours and again count the colonies. 6.2 Active Air Sampling 6.2.1 Active Air Sampling shall be done by using the Air sampler. 6.2.2 Switch on the instrument, set the parameters like time and air flow and place the Soyabean casein digest agar media plate carefully to avoid the external contamination. 6.2.3 1000 liters of air shall be sampled by the air sampler. After completion of air sampling, remove the sieve and aseptically place the cover lid on the plate. 6.2.4 Plate shall be recovered after sampling and shall be incubated at 20.0 C to 25.0 C for 72 hours and count the number of colonies without opening the plates. Further incubate the plates at 30.0 C to 35.0 C for 48 hours and again count the colonies. 7. Non Viable Particle count 7.1 Calibrated air borne particle counter shall be used for air borne particle count test. 7.2 This test shall be conducted at rest & dynamic both condition. 7.3 The Isolator chamber must be start before 30 mins of activity to stabilize the chamber. 7.4 Derive the minimum no. of sampling point locations from the formula Minimum No. of sampling Location = Area of the chamber in Sq.m Establishment of single sample volume per location The single sample volume Vs per location is determined by the formula Vs= 20 x 1000 Cn.m Where, Vs = Minimum single sample volume per location expressed in liters Cn.m = is the class limit (number of particles per meter cube) 20 = is the defined number of particles that could be counted if the particle concentration where at the class limit. 8. Recovery test 8.1 Switch on the power of isolator blower from the control panel. 8.2 Open the isolator door for 2 minutes then close it and take the particle count. 8.3 Take the particle count inside the chamber for every minute till the chamber reaches to stable condition(grade A limit). 8.4 Note the time when particles come within acceptance criteria. www.wjpmr.com 159

9. Air Flow Pattern test Airflow pattern test is to confirm either airflow direction or visualization with respect to the design and performance specifications to verify the airflow throughout chamber and the capability of the HEPA Filter unit to limit the dispersion and turbulence of air so as to maintain classified environmental conditions. 9.1 Set the fogger used for creation of smoke of Poly Ethylene glycol (PEG) or suitable substance pointing in the direction of airflow in the center of HEPA filter in flute type pipe in which fog shall be dispersed. 9.2 Smoke is provide for visual indication of the airflow direction and fluctuations due to turbulence 9.3 Verify that the path of smoke is parallel to the airflow. 10. Chamber Leak Test: Chamber leak test is done to check the integrity of the chamber i.e to determine the chamber is properly sealed or not. 10.1 Persian Blue Sealing check Method 10.1.1 A thin layer of Persian blue shall be applied all over the gasket of the chamber. Chamber shall be then sealed tightly with nut-bolt provided on chamber. (Refer Picture-A). 10.1.2 After opening the chamber door, the door shall be checked to verify presence of Persian blue on door surface. (Refer Picture-B). 10.1.3 If the Persian blue found to be spread all over the surface of the door where the gasket was touched, it confirms that there is no leakage from the door of chamber. 10.1.4 Put a mark on bolt at which chamber door was sealed for regular sealing purpose. 10.1.5 If Persian blue not found to be spread all over the surface, it means leakage is present. Re perform the sealing check method by means of more tightening of bolt on the door to provide high degree of sticking of door to gasket. 11. Leak Test of hand Gloves Leak test of hand glove is done to verify whether hand glove is capable of acting as a barrier between human and internal environment. 11.1 2 kg pressure of Compressed air shall be supplied to create the positive pressure in the chamber. 11.2 When positive pressure attains in chamber the gloves got swell up. 11.3 To check Integrity of gloves apply Soap solution(t- Pol) on whole glove, specially at all the joints. 11.4 If no air bubble got generated on surface of glove, it means no leakage is present on the glove. 12. Occupational Exposure Limit (OEL) Determination OEL is determined to verify the amount of the material which is exposed to external environment by the Isolator chamber. Air sampler is used to take the sample from all over the expected joints present in chamber. Air Sampler Setting (Sample Volume) To calculate the sample volume to be sucked by the Air sampler, first needs to determine the ACPH of the Room where Isolator has been kept. For Example Volume of air to be sampled is determined according to Room ACPH. Capacity of Dispensing Room AHU is :- 2372 CFM Room Volume is :- 25.88 m 3 Now for setting up the volume to be sucked by air sampler:- AHU capacity = 2372 CFM = 2.372 m 3 /min Air change/hour = 2.372*60 = 142.32 m 3 /hr Volume to be sampled by air sampler = 142.32/3600 = 0.039m 3 /sec Total Volume scanned from all various location 15 minutes = 0.039*15*60 = 35.1 m 3 www.wjpmr.com 160

12.1 Determination of OEL (Occupational Exposure Limit) 12.1.1 Take a suitable ingredient which OEL has to be determined (For performance qualification low density powder shall be used). 12.1.2 Take ingredient in isolator and keep into the passbox by means of lock door as per respective SOP. 12.1.3 Start the blower to stabilize the isolator by keeping it in static condition for few minutes as per defined procedure (isolator manual) to attain the said differential pressure. 12.1.4 Transfer the material into the chamber by opening door between the chamber and pass box. 12.1.5 Weight the 1 kg of ingredient and transfer it to the double poly bag by means of glove port. 12.1.6 Air sampler having whatman filter of 0.4 micron wetted with 70% of IPA is moved through all the joints of isolator along the inlet pass box window, gloves port and outlet pass box window to take the air sample. 12.1.7 After air sampling the sampled filter paper is properly sealed in a sampling bag and the filters are immediately sent to external laboratory for determination (By LCMS) of amount of ingredient sampled on it to determine the exposed amount of ingredient. 2. Sample Exposed in Chamber. 3. Air Sampler having whatman filter of 0.4 micron. Pictorial presentation of Sampling Location. *Note: Location 1, 2, 3 and 4 are joint locations for air sampling. 4. Air Sampling By Air Sampler. 1. Chamber in Stabilize condition. www.wjpmr.com 161

Acceptance Criteria S.No. Test Parameter Acceptance Criteria 01. Light Intensity Test NLT 500 Lux 02. Air Velocity Air Changes As per design and Vendor specification 03. Differential pressure 04. Filter integrity NMT 0.003% 05. Temperature NMT 25 0 C RH NMT 55% 06. Viable Particle Count Active Air Sampling Less than 1 cfu /m 3 Settle plate monitoring Less than 1 cfu/plate 07. Non Viable Particle Count Static & Dynamic Condition: 0.5 micron 3520 particle 5.0 micron 20 particle 08. Recovery test NMT 15 min 09. Air flow pattern As per design data 10. Chamber Leak Test No Leakage should be observed 11. Leak Test of hand Glove No Leakage should be observed 12. Determination of Occupational Exposure Limit(OEL) NMT 0.3 microgram CONCLUSION The procedure used for verification of isolator performance and determination of occupational exposure limit will help to: 1. Determine the Occupational exposure limit (amount of drug substance exposed out from isolator) 2. Determine and verify the internal parameters like light intensity, air velocity, chamber leak test and leak test of gloves. ACKNOWLEDGEMENT Authors of this study are grateful to management of VHB Medisciences Limited, Rudapur for giving all kind of support in terms of materials and time. Authors also deeply immense to Mr. Baban Jogale, plant head and Mr. Ashok jain, owner,vhb Medisciences limited, Rudrapur for providing such kind of environment. REFERENCES 1. WHO Technical Report Series No. 961., 2011 Annex-6. 2. PIC/S PI 014-3 25 September 2007. www.wjpmr.com 162